Fluke Documentation
Fluke Documentation
Quality Assurance:
Inspection Program Development
and Procedures
J. Tobey Clark
Director, Instrumentation & Technical Service,
Faculty, Biomedical Engineering/School of Engineering
Michael Lane
Associate Director, Instrumentation & Technical Services
Leah Rafuse
Clinical Engineer, Instrumentation & Technical Services
The purpose of this document is to provide The publication is organized into four primary
guidance in establishing and managing a medi- sections:
cal equipment quality assurance (QA) program • A glossary of terms
and to present detailed procedures for inspec- • Background information on establishing and
tion, preventive maintenance, safety evaluation, managing a program including goals, meth-
and performance testing. ods, device inclusion and rating criteria, and
The target audiences for this publication standards
are those responsible for establishing and • General and device-specific procedures and
managing medical equipment QA program, forms for inspection, preventive mainte-
and staff performing inspections and device nance, safety evaluation, and performance
testing. Readers should have a basic techni- testing
cal background and some exposure to medical • An appendix with reference information
equipment in healthcare. The book does not Forms, tables, diagrams, illustrations, and photos
provide background information on clinical or are used to aid in the understanding of the
technical concepts or medical equipment prin- content.
ciples of operation. The publication takes into consideration
Medical equipment QA is part of an over- the advances in device reliability, reduced
all medical equipment management program preventive maintenance requirements, and
for a healthcare facility or system. A complete internal device surveillance (self test) along
program also includes corrective maintenance with changes in standards. Due to the ongoing
or repair, equipment control, asset management, efforts at global harmonization, international
health care technology planning, education, standards are used and referenced where
and activities directed toward improving medi- applicable, such as electrical safety testing
cal device-related patient safety. references IEC 62353.
Acknowledgements
The publication was funded by a grant from Fluke
Electronics Corporation (FEC). A number of diagrams,
illustrations and figures were provided by FEC as was
test equipment used in many of the photographs.
Complete
Maintenance
Strategy
Worksheet
The risk assessment should be done for
each new device type during the incoming
inspection when the device is added to the
inventory. The device will then have a test- Assign
ing frequency assigned. After this is done, inspection
frequency
the maintenance history of the device should
be monitored in order to evaluate the effec-
tiveness of the maintenance program. The
process is shown in Figure 1.
Inspection Accumulate Inspection
frequency may device frequency may
be too low history be too high
Evaluate
device
history
Yes
Frequent failures or
problems found
during testing?
Yes
No No
Infrequent failures
and no problems found
during testing?
No
Infrequent
failures, some minor
repairs or adjustments
needed during
testing?
Yes
Figure 1.
Most device types have been evaluated and classified for test frequency already.
For new device types, use the scoring system to evaluate the frequency of testing.
Criteria: Choose one rating from each category Weight Score
Clinical function
No patient contact 1
Device may make contact with patient but function is non-critical 2
Device is used for patient diagnosis, or direct monitoring 3
Device is used to deliver direct treatment to the patient 4
Device is used for a life support 5
Physical risk
Device poses no appreciable risk due to failure 1
Device failure will result in low risk 2
Device failure will result in inappropriate therapy, misdiagnosis, or loss of monitoring 3
Device failure could result in severe injury to, or death of, patient or user 4
Problem avoidance probability
Maintenance or inspection would not impact reliability of the device 1
Common device failure modes are unpredictable or not very predictable 2
While common device failure modes are not very predictable, device history indicates that 3
TSP testing frequently detects problems
Common device failure is predictable and can be avoided by preventive maintenance 4
Specific regulatory or manufacturers requirements dictate preventive maintenance or testing 5
Incident history
No significant history 1
A significant history of incidents exists 2
Manufacturers/regulatory requirements for specific schedules
No requirements 1
There are requirements for testing independent of a numerical rating system 2
Total Score:
Assignment: 0.0x 0.5x 1x 2x 3x 4x (times per year tested)
A combined score of 13 or more is justification for semiannual testing.
A combined score of 9-12 is justification for annual testing.
A combined score of 8 or less is justification for less than annual testing (either bi-annual or no scheduled testing, depending on clinical application).
Anesthesia machines and vaporizers are recommended for testing three times per year.
Some blood delivery devices such as warmers may be required to be tested four times per year based on AABB or CAP requirements.
Problem Manufacturer
Clinical Physical Avoidance Incident or Regulatory Total Testing
Device Type Function Risk Probability History Requirements Score Frequency
Ventilator 5 4 4 2 2 17 Semiannual
Electrosurgical unit 4 4 2 2 1 13 Semiannual
Infusion pump 4 3 2 2 1 12 Annual
Pulse oximeter 3 3 2 1 1 10 Annual
Exam table 2 2 2 1 1 8 Bi-annual
13.00 %
12.00 %
11.00 %
10.00 %
9.00 %
8.00 %
7.00 %
6.00 %
5.00 %
4.00 %
3.00 %
2.00 %
1.00 %
0.00 %
p
De nito ph
m
A
ng
er
r
O2
l M nt
it
nt
it
r
it
BP
to
m
ito
PC
Un
Un
Un
te
ac
a
fa
a
di
ila
Pu
SP
NI
f
on
s
gr
In
In
,P
ee
y
p,
al
En sion
ia
nt
io
S
r,
on
r,
m
r,
ic
t,
r
lF
m
Ve
ito
to
rd
g
an
to
rg
Pu
si
er
rin
es
ta
ra
la
ca
ba
on
fu
su
di
th
Fe
te
pr
ril
on
ito
tro
In
M
cu
Ra
tro
er
fib
m
si
on
ec
In
yp
Co
ec
,
fu
er
p,
M
El
/H
El
In
m
n
po
ar
Pu
tio
W
Hy
ta
lS
ra
nt
Ce
Figure 2.
Device information
Device type Owner Hosp Vendor MD Other
Manufacturer Department
Model Tester Hosp TSP Mfr User
Serial number Prev maint Hosp TSP Mfr User
Control number Tests/year 1x 2x 3x 4x
Inclusion assessment
The intent is to capture all powered devices that by function, physical risk, maintenance requirements, or a history
of incidents or safety problems should be managed as a part of the medical equipment management system. Some
devices have a borderline inclusion between medical equipment and utilities, so tests for inclusion in the utilities
management program are included.
1: Life support equipment
Would failure of this device result in immediate death of the patient? And Yes No
Is the powered device used for direct patient treatment or care? Yes No
An answer of “Yes” to both of the above questions indicates that the device should be managed in the medical equipment inventory as a life support device.
2: Medical equipment
Is the powered device used for direct patient treatment or care? Yes No
Does the powered device provide diagnostic/monitoring information used in treatment? Yes No
Does this powered device come in contact with the patient? Yes No
An answer of “Yes” to any of the above three questions indicates that the device should be included in the medical equipment management program and
be inventoried under those provisions.
3: Utilities equipment
Does this device facilitate life support functions? Yes No
Does this device support infection control systems? Yes No
Does this device support facility environmental systems? Yes No
Does this device support critical facility utility systems? Yes No
Does this device support essential communications systems? Yes No
If the device fails to meet the medical equipment requirements, but there are any “Yes” answers to the utilities equipment questions, the device should be included
in the utilities equipment management program.
4: Clinical and physical risk
Does the device pose risk to the patient or staff when used in the facility? Yes No
Would failure or loss of use of the device adversely affect the deliver of health care? Yes No
Does this product or class of device have a history of incidents or safety recalls? Yes No
5: Maintenance requirements
Does the device require periodic inspection in order to ensure safe delivery of care? Yes No
Does the device require periodic performance testing to ensure safe delivery of care? Yes No
Does the device require periodic preventive care to ensure safe delivery of care? Yes No
If the device fails to meet the requirements for Medical or Utilities Equipment, but there are “YES” answers to the above questions in Clinical and Physical risk or
Maintenance Requirements, the device should be managed on a general equipment inventory with preventive maintenance or testing as appropriate.
Management program assignment
This device will be assigned as Life Support Equipment Yes No
This device will be assigned as Medical Equipment Yes No
This device will be assigned as Utilities Equipment Yes No
This device will be assigned as General Equipment Yes No
This device will be NOT be included in any equipment management program Yes No
Department:____________________________________________________________________________
Manager:________________________________________________ Date:_________________________
����� This vendor certifies this device is provided to the facility in safe and useable condition, and is FDA
approved for the intended clinical procedure(s). This vendor has or will test the device for proper
function prior to clinical use, but after arrival at the facility. Training will be provided to staff on
safe use and potential risks.
����� This device is an investigational device for which FDA approval does not exist as of yet. The device
will not be used clinically until all hospital investigational review board approval is received.
Training will be provided on safe use and potential risks.
����� This equipment is provided repeatedly to the hospital and is maintained by the vendor. It is
checked between assignments to different facilities, and records of maintenance are provided
annually to the facility. Ongoing performance and safety testing is provided, and the vendor
certifies that the device is safe for use as provided at delivery.
����� Non-hospital employees who will be delivering and/or operating the equipment are appropriately
trained and qualified to be transporting, setting up, and operating the equipment.
_______________________________________________________________________________________
HNICAL
HNICAL
HNICAL
PR
PR
PR
The University of Vermont uses a variety of
O G RA
O G RA
O G RA
TESTED BY TESTED BY TESTED BY
labels to indicate specific data necessary for
EC
EC
EC
B -ANNUAL ANNUAL SEMI ANNUAL
technology management, regulatory require-
M
M
T T
• U • U
T
VM • • U VM •
ments, and for safety. Labels may be used for VM •
performance inspections—bi-annual, annual,
semiannual and general, battery installa-
tion, specific calibration, warranty, hazards RVICES RVICES
SE SE
and warnings, and upgrade or recall data. In DATE DUE
HNICAL
HNICAL
DA E DUE
PR
PR
addition, a unique identifier control number is
O G RA
O G RA
TESTED BY
ES ED BY
placed on each piece of equipment for ease of
EC
EC
QUARTERLY
M
T T
device tracking. The following are examples of • U • U
VM • VM •
these labels:
Control number tags are made of aluminum
and attached with a permanent adhesive. Biomedical Engineering Dept
PERFORMANCE TESTED BY
These tags need to be rugged in order to stand TECHNICAL SERVICES PROGRAM ELECTRICAL SAFETY TEST
up to repeated cleaning and disinfecting of the Name________________ Date__________
Completed by__________________________
Date__________________________________
equipment. Other labels are made of a material Inspection Due_________________________
UAL BE269
that allows them to be peeled off and replaced.
Labels that require written information should
be filled in using permanent marker, to avoid
fading of the ink. NON-HOSPITAL OWNED BATTERY REPLACED
EQUIPMENT Date____________ By____________
Date Due_______________________
Electrical Safety Tested
Biomedical Engineering Dept
Date______________ By_____________ BE702
Next Inspection Due_____________________
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer 20 cc or larger syringe
IDA4 Plus Infusion Device Analyzer 3 way stopcock
Tubing set for infusion pump Tubing and connectors to connect to IDA 4 Plus
Reservoir to connect to tubing set (bag or bottle)
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Preventive maintenance
Clean flow detector
Replace battery every 24 months
Complete model-specific preventive maintenance
Performance testing
Verify unit operates on battery
Pole clamp function
Flow rate accuracy ± 10 %
Volume accuracy ± 10 %
Infusion complete/KVO
Occlusion detection pressure ± 1 psi
Piggyback infusion
Alarm function
Complete model-specific performance testing
Burns
Cardiac fibrillation
Let-go current
Sensitivity limit
1 mA 10 mA 100 mA 1A 10 A 100 A
Figure 4. Effects of current flowing from one skin contact point to another.
230 V
50 Hz
Fuses 230 V
10 Ω
~22 A
~11 V
Figure 5. Hazard current from electrical failure being safety shunted to ground through an
alternative pathway.
Other important points about IEC 60601.1 2. Association for the Advancement of Medi-
are the use of up to 25 amperes AC for protec- cal Instrumentation (AAMI) - ANSI/AAMI
tive earth testing, leakage current is measured ES1, Safe Current Limits for Electromedical
at 110 % of mains voltage, and performance of Apparatus is another commonly accepted
dielectric strength/insulation testing. standard.
A new IEC standard is used for medical 3. Underwriters Laboratories (UL) - UL544,
device testing in hospitals. IEC standards 62353 Medical Equipment requirements is a stan-
applies to testing of medical equipment and dard for manufacturers, not hospitals.
medical electrical systems, which comply with These standards may be referenced by accredi-
IEC 60601-1. IEC 62353 was developed because tation, code or regulatory organizations such as
IEC 60601.1 is a type-testing standard with no the Joint Commission, Occupational Health and
risk management criteria and is impractical for Safety Administration or other organizations
testing in the hospital environment. monitoring healthcare institutions in the United
IEC 62353 tests include those prior to use on States. The Appendices describe the above
patients, during schedule periodic testing, and standards and test setups.
after repair. Thus, this standard is for hospi- Global harmonization of standards has lead to
tals and does not address equipment design. the development of world wide standards. After
In Annex E of the document, the manufacturer the deadlines below, equipment must be certi-
is requested to provide information on testing fied to the IEC60601-1 standard or the device
interval and procedure based on risk, typical cannot be sold in that country.
usage, and device history. Minimum interval
• USA uses UL2601-1
requirements for life support and other critical
– The deadline was December 31, 2004
equipment is set at 24 months.
• Europe uses EN60601-1
In the United States, there are several primary
– The deadline was June 13, 1998
and secondary organizations setting standards:
• Canada uses CAN/CSA-C22.2 No. 601.1-M90
1. National Fire Protection Association
– The deadline was January 1, 2000
(NFPA) - NFPA 99, Standard for Healthcare
Facilities is the primary standard addressing
electrical safety testing affecting healthcare
institutions. Other publications are NFPA 70,
National Electrical Code, and NFPA 70E,
Electrical Safety in the Workplace.
To protective
earth or
N
R
enclosure
S
Pro ON
01 RN
N
2
5
8
0
3
6
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: Fluke Biomedical ESA620 or equivalent
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Fuse rating is appropriate
Power cord, accessory cables, charger, patient cables, connectors
Integrity of mechanical parts
Electrical safety Criteria IEC 62353
Ground wire resistance 0.3 W
Current in µA Applied part
Type Type Type
B BF CF
Equipment leakage – alternative method
For accessible conductive parts of Class I equipment con-
1000 1000 1000
nected or not connected to the protective earth conductor
For Class II ME equipment 500 500 500
Equipment leakage – direct or different method
Equipment leakage current for accessible conductive parts
of Class I ME equipment connected or not connected to the 500 500 500
protective earth conductor
Equipment Leakage current for Class II ME equipment (NC) 100 100 100
Applied part leakage current – alternative method (a.c.)
Applied part leakage current of applied part — < 5000 < 50
Applied part leakage current – direct method (a.c.)
Total patient leakage current mains voltage on applied part — 5000 100
Insulation test (optional) 500 V dc applied < 2 MW
Physical condition
Verify the case integrity and look for damage.
Ensure the device is not contaminated. Check
the controls, indicators, and displays. Verify
the labeling is appropriate, not damaged and
legible. Check the fuse has the proper rating.
Visually inspect the power cord, plug, any
cables, connectors, chargers, or other external
connections. Verify any mechanical parts are in
good condition.
L1
APPLIED
PART
MAINS L2
OPEN DUT_L2
L2
S1 S2
EARTH
OPEN GREEN
TERMINAL DUT_PE FE
PE S3
CONDUCTIVE PART
_ TEST LEAD
CURRENT
SOURCE OHM
1 A DC
+ RED TERMINAL
APPLIED
PART
+ DUT_L2
500 V
_ DC
MD
MEG
OHMS
DUT_PE
FE
EARTH
OPEN
CONDUCTIVE PART
S3
PE
L1
APPLIED
PART
MAINS L2
OPEN DUT_L2
L2
S1 S2
EARTH
OPEN
DUT_PE
FE
S3
CONDUCTIVE PART
MD
µA
APPLIED PART
PE S4
L1
APPLIED
PART
MAINS L2
OPEN DUT_L2
L2
S1 S2
EARTH
OPEN
DUT_PE
FE
S3
CONDUCTIVE PART
MD TEST LEAD
µA
APPLIED PART
PE S4
L1
APPLIED
PART
MAINS L2
OPEN DUT_L2
L2
S1 S2
EARTH
OPEN
DUT_PE
FE
S3
CONDUCTIVE PART
LEAD SELECT
MD
µA
PE S4
L1
APPLIED
PART
MAINS L2
OPEN DUT_L2
L2
S1 S2
EARTH OPEN
DUT_PE
FE
S3
CONDUCTIVE PART
TEST LEAD
RED TERMINAL
PE S6
1:1.1
MD LEAD SELECT
L1 µA
L2
ISOLATION S4
TRANSFORMER
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent)
Other equipment as necessary
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Preventive maintenance
Calibrate to manufacturer’s specifications
Check all fluid levels
Replace battery every 24 months
Clean exterior
Lubricate as required
Complete model-specific preventive maintenance
Performance testing
Verify unit operates within manufacturer’s specifications
Operates on battery power
Audible alarms
Visual alarms
Remote alarms
Complete model-specific performance testing
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent)
MPS450 Multiparameter Simulator (or equivalent)
Stopwatch or watch with a second hand
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Preventive maintenance
Replace battery every 24 months
Complete model-specific preventive maintenance
Performance testing
Verify unit operates on battery
Heart rate accuracy ±5%
Respiratory rate accuracy ±5%
Apnea alarm function
Apnea alarm delay time ± 20 %
60 BPM rejection of ECG artifact
Alarm function
Complete model-specific performance testing
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent) DPM 4 Pressure Meter (or equivalent)
Stopwatch or watch with a second hand Tubing and connectors to connect to DPM 4
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Preventive maintenance
Replace battery every 24 months
Replace filters
Lubricate motor
Complete model-specific preventive maintenance
Performance testing
Verify unit operates on battery
Vacuum gauge accuracy ± 10 %
Maximum vacuum Thoracic, low volume: > 40 mmHg
Emergency, surgical, tracheal, uterine: > 400 mmHg
Breast pump: > 200 mmHg
Vacuum rise time Thoracic: < 4 sec/30 mmHg
Emergency, Surgical, Tracheal: < 4 sec/300 mmHg
Uterine: < 3 sec/30 mmHg
Breast pump: < 2 sec/150 mmHg
Complete model-specific performance testing
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent)
MPS450 Multiparameter Simulator (or equivalent)
Cardiac output adapter box
Cables to connect to cardiac output monitor
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Performance testing
Verify accuracy of blood temperature 37 °C ± 0.2 °C
Verify cardiac output accuracy
Complete model-specific performance testing
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent)
MPS450 Multiparameter Simulator (or equivalent)
Index 2 SpO2 Simulator (or equivalent)
Vacuum or canned air for clearing dust from cooling fans
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Preventive maintenance
Clean dust from cooling fans
Complete model-specific preventive maintenance
Performance testing
Verify display is clear and legible
Verify monitoring capabilities of hard wired monitors
Verify monitoring capabilities of telemetry transmitters
Verify recorder accuracy ±4%
Verify operation of alarms
Complete model-specific performance testing
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent)
DPM 4 Pressure Meter (or equivalent)
Stopwatch or watch with a second hand
Tubing and connectors to connect to DPM 4
Compression set
PVC pipe to attach compression set to
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Preventive maintenance
Replace battery every 24 months
Complete model-specific preventive maintenance
Performance testing
Verify unit operates on battery
Pressure accuracy ±2%
Timing cycle accuracy ±2%
Alarm function
Complete model-specific performance testing
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent)
Impulse 4000 Defibrillator and Pacer Analyzer (or equivalent)
MPS450 Multiparameter Simulator (or equivalent)
Ohmmeter (Can be part of a multimeter, such as the Fluke 73 series digital multimeter)
Cables and connectors to connect defibrillator to analyzer
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC, < 500 µA SFC
Patient leakage current < 100 µA B and BF, < 10 µA CF
Patient lead leakage current – isolation test (mains on patient applied part) < 100 µA BF, < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Preventive maintenance
Replace battery every 24 months
Verify electrodes, gel and paddles are stored with the defibrillator and are within expiration dates
Verify proper time and date, correct if necessary
Complete model-specific preventive maintenance
continued on page 61
Test Result
Pass Fail N/A
Performance testing
Verify unit operates on battery
Paddle continuity ≤ 0.15 W
Heart rate accuracy ±5%
Recorder speed ±4%
Verify operation of alarms
Output accuracy ± 15 %
Output energy at maximum setting for 10 charge cycles ± 15 %
Charge time after 10 discharge cycles ≤ 15 sec
Energy after 60 sec of full charge ≥ 85 %
Internal discharge function
Synchronizer operation ≤ 60 msec
Pacer output accuracy ± 10 %
Pacer rate accuracy ±5%
Demand-mode sensitivity
Complete model-specific performance testing
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent)
MPS450 Multiparameter Simulator (or equivalent)
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Preventive maintenance
Replace battery every 24 months
Clean rollers and paper guides
Lubricate motor and paper drive mechanism
Verify proper time and date. Correct if necessary
Complete model-specific preventive maintenance
Performance testing
Verify unit operates on battery
Heart rate accuracy ±5%
Amplitude accuracy ±5%
Recorder speed ±4%
Paper cue
Alarm function
Complete model-specific performance testing
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent)
QA-ES II Electrosurgery Analyzer (or equivalent)
Cables and connectors to connect ESU to QA-ES II
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Preventive maintenance
Replace filters as necessary
Complete model-specific preventive maintenance
Performance testing
Inspect dispersive electrode
Operation of footswitch
Output power ± 15 %
Return electrode monitor
Alarms
Complete model-specific performance testing
Remove the monopolar electrode from red the return electrode alarm. Be sure the disper-
‘VAR. LOAD’ connector on the QA-ES II and the sive electrode is placed on a non-conductive
dispersive electrode from the black connector. surface where the electrode will not come
Connect the ESU bipolar active electrode to the into contact with anyone. The load resistance
red ‘VAR LOAD’ connector. Connect the return may need to be adjusted for bipolar operation.
of the bipolar electrode to the black ‘VAR LOAD’ Repeat the output measurements detailed above
connector. The dispersive cable will need to for both cut and coagulation modes according
remain connected to the ESU to avoid activating to the table below.
Return electrode monitor: Connect the the alarm sounds. The return electrode alarm
dispersive electrode to the ESU and connect should also sound if the dispersive cable is
the two wires of the dispersive electrode to disconnected from the ESU. The ESU should not
the red and black ‘VAR.LOAD’ connectors of the activate when the return electrode monitor is
QA-ES II. To make a quick test set from a dis- alarming.
posable dispersive electrode, cut the electrode Verify operation of alarms: Simulate any
from the cable and separate the two wires in alarm conditions. Check that all alarms are
the dispersive cable. Strip the wires and attach functional and that the volume is adequately
banana plug connectors. loud. Ensure that appropriate visual indicators
Press F2 ‘KNOB PARAM.’ on the QA-ES II until are functioning.
the asterisk appears next to Mode. Use the Complete model-specific performance
encoder knob to select REM test and then press testing: Refer to the service manual for perfor-
enter to select the test. Press F2 ‘KNOB PARAM’ mance inspection tasks specific to the device.
to select Delay. Turn the encoder knob to set the Complete the performance inspection per
delay to 3000 ms. manufacturer’s procedure.
Press F3 ‘START’ on the QA-ES II. The load Return to service: Before returning to use,
resistance will start at 10 W and gradually return any settings that were adjusted to their
increase. Press F3 ‘STOP’ to stop the test when original settings.
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent) 20 cc or larger syringe
IDA 4 Plus Infusion Device Analyzer (or equivalent) 3 way stopcock
Tubing set for feeding pump Tubing and connectors to
Reservoir to connect to tubing set (bag or bottle) connect to IDA 4 Plus
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Preventive maintenance
Clean flow detector
Replace battery every 24 months
Complete model-specific preventive maintenance
Performance testing
Verify unit operates on battery
Pole clamp function
Flow rate accuracy ± 10 %
Volume accuracy ± 10 %
Occlusion detection pressure ± 1 psi
Alarm function
Complete model-specific performance testing
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent)
SigmaPace 1000 External Pacemaker Analyzer (or equivalent)
Cables and connectors to connect pacemaker to analyzer
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Preventive maintenance
Replace battery every 24 months
Complete model-specific preventive maintenance
Performance testing
Verify unit operates on battery
Rate accuracy ±5%
Output accuracy ± 10 %
Pulse width ± 10 %
Alarm function
Complete model-specific performance testing
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent) Cables to connect fetal monitor to
PS320 Fetal Simulator (or equivalent) analyzer
MFH-1 Mechanical Fetal Heart (or equivalent) Stopwatch or clock with second hand
Ultrasound gel
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Preventive maintenance
Replace battery every 24 months
Clean rollers and paper guides
Lubricate motor and paper drive mechanism
Verify proper time and date, correct if necessary
Complete model-specific preventive maintenance
Performance testing
Verify unit operates on battery
Fetal heart rate accuracy ±5%
Maternal heart rate accuracy ±5%
Intrauterine pressure accuracy ±2%
Recorder speed (3 cm/min) ±4%
Alarm function
Complete model-specific performance testing
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent)
DPM 4 with temperature probe (or equivalent)
Water flow meter
Tubing and connectors to connect flow meter
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Preventive maintenance
Inspect and clean reservoir
Lubricate motor
Complete model-specific preventive maintenance
Performance testing
Fluid level
Flow rate
Temperature accuracy ± 1 ºC
High temperature protection ≥ 43 ºC
Low temperature protection ≤ 1 ºC
Temperature probe accuracy ± 1 ºC
Alarms
Complete model-specific performance testing
Performance inspection
Fluid level: Ensure there is an adequate water
level in the reservoir. There should be sufficient
water for circulating throughout the blanket, but
the water reservoir should not be overfilled.
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent)
INCU incubator analyzer (or equivalent)
Stopwatch or watch with a second hand
Heat Gun
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Preventive maintenance
Clean cooling vents and filters
Inspect and clean ducts, heater, and fans
Inspect gaskets for signs of deterioration
Inspect port closures and port sleeves
Replace battery every 24 months
Complete model-specific preventive maintenance
continued on page 90
Test Result
Pass Fail N/A
Performance testing
Verify unit operates on battery
Fan operation
Warm up time ± 20 %
Air temperature accuracy ± 1 ºC
Skin temperature accuracy ± 0.3 ºC
Temperature overshoot ± 2 ºC
Relative humidity ± 10 %
Air flow ≤ 0.35 m/s
Air temperature alarms
Skin temperature alarms
High temperature protection ≤ 40 °C
Noise level ≤ 60 dB normal conditions
≤ 80 dB alarm activated
≥ 80 dB alarm activated,
3 m from incubator
Alarm function
Complete model-specific performance testing
***CAUTION***
DO NOT USE THE AIR FLOW SENSOR IN
THE PRESENCE OF OXYGEN. THE SENSOR
USES A HOT-WIRE TECHNIQUE FOR AIR
VELOCITY MEASUREMENT AND MAY
BECOME A SOURCE OF IGNITION.
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent) Tubing set for infusion pump
IDA 4 Plus Infusion Device Analyzer (or equivalent) 20 cc or larger syringe
Reservoir to connect to tubing set (bag or bottle) 3 way stopcock
Tubing and connectors to connect to IDA 4 Plus
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Preventive maintenance
Clean flow detector
Replace battery every 24 months
Complete model-specific preventive maintenance
Performance testing
Verify unit operates on battery
Pole clamp function
Flow rate accuracy ± 10 %
Volume accuracy ± 10 %
Infusion complete/KVO
Occlusion detection pressure ± 1 psi
Piggyback infusion
Alarm function
Complete model-specific performance testing
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent) BP Pump2 NIBP Analyzer (or equivalent)
Stopwatch or watch with a second hand PVC pipe to attach tourniquet cuff to
Tubing and connectors to connect to DPM 4
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC, < 500 µA SFC
Patient leakage current < 100 µA B and BF, < 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF, < 10 µA CF
(mains on patient applied part)
Insulation test (optional) 500 V < 2 MW
Preventive maintenance
Check condition of tubing, cuffs, and hoses
Clean recorder paper compartment, rollers and paper guides
Lubricate motor and paper drive mechanism
Verify proper time and date, correct if necessary
Replace battery every 24 months
Complete model-specific preventive maintenance
Performance testing
Verify unit operates on battery
Leak test ≤ 15 mmHg/min
Static pressure accuracy ± 3 mmHg
Pressure relief test ≤ 330 mmHg
Dynamic pressure accuracy ± 10 mmHg
Heart rate accuracy ±5%
Auto interval time ± 10 %
Stop/Cancel/Deflate ≤ 10 sec
Recorder operation
Alarm function
Complete model-specific performance testing
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent)
MPS450 Multiparameter Simulator (or equivalent)
BP Pump 2 NIBP Analyzer (or equivalent)
Index 2 SpO2 Analyzer (or equivalent)
Stopwatch or watch with a second hand
Cables to connect to MPS450
Tubing and connectors to connect to BP Pump 2
PVC pipe to attach BP cuff to
Gas with a known quantity of CO2
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Test Result
Pass Fail N/A
Preventive maintenance
Check condition of tubing, cuffs, and hoses
Clean recorder paper compartment, rollers and paper guides
Lubricate motor and paper drive mechanism
Verify proper time and date, correct if necessary
Replace battery every 24 months
Complete model-specific preventive maintenance
Performance testing
Verify unit operates on battery
Heart rate accuracy ±5%
Amplitude accuracy ±5%
Recorder speed ±4%
Respiration rate accuracy ±5%
Leak test ≤ 15 mmHg/min
Static pressure accuracy ± 3 mmHg
Pressure relief test ≤ 330 mmHg
Dynamic pressure accuracy ± 10 mmHg
Auto interval time ± 10 %
Stop/Cancel/Deflate ≤ 10 sec
SpO2 accuracy ±3%
Invasive pressure accuracy ±5%
Temperature accuracy ± 0.3 C
Carbon dioxide concentration accuracy ± 0.4 vol %
Alarm function
Complete model-specific performance testing
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent)
IDA 4 Plus Infusion Device Analyzer (or equivalent)
PCA trigger interface
Tubing set for PCA pump
Reservoir to connect to tubing set (bag, bottle, or syringe)
20 cc or larger syringe
3 way stopcock
Cable to connect PCA trigger interface to PCA pump
Tubing and connectors to connect infusion set to IDA4 Plus
Tubing and connectors to connect to IDA 4 Plus
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Test Result
Pass Fail N/A
Preventive maintenance
Clean flow detector
Replace battery every 24 months
Complete model-specific preventive maintenance
Performance testing
Verify unit operates on battery
Door lock
Pole clamp function
Load dose ± 10 %
Flow rate accuracy ± 10 %
Volume accuracy ± 10 %
PCA dose ± 10 %
Lock out interval ±5%
Dose limit
KVO rate ± 10 %
Occlusion detection pressure ± 1 psi
Alarm function
Complete model-specific performance testing
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent)
DALE40 Phototherapy Radiometer (or equivalent)
Stopwatch or watch with a second hand
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Preventive maintenance
Inspect bulbs
Complete model-specific preventive maintenance
Performance testing
Timer accuracy ± 0.5 %
Output accuracy ≥ 4.5 µW/cm2/nm
(≥ 198 µW/cm2 @ 44 nm bandwidth)
≤ 40 µW/cm2/nm
(≤ 1760 µW/cm2 @ 44 nm bandwidth)
Alarm function
Complete model-specific performance testing
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent) Squeeze bulb with bleed valve
DPM 4 Pressure Meter (or equivalent) Tubing and connectors to connect
Stopwatch or watch with a second hand to DPM 4
PVC pipe to attach tourniquet cuff to
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Preventive maintenance
Check condition of tubing, cuffs, and hoses
Replace battery every 24 months
Complete model-specific preventive maintenance
Performance testing
Verify unit operates on battery
Verify function of control valve
Controller stability ± 10 mmHg after 15 min
Cuff pressure accuracy ±5%
Timer accuracy ± 2 min after 15 min
Maximum cuff pressure ≤ 550 mmHg or
manufacturer’s specification
Complete model-specific performance testing
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent)
Index 2 SpO2 Analyzer (or equivalent)
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Preventive maintenance
Clean recorder paper compartment, rollers and paper guides
Lubricate motor and paper drive mechanism
Verify proper time and date, correct if necessary
Replace battery every 24 months
Complete model-specific preventive maintenance
Performance testing
Verify unit operates on battery
Heart rate accuracy ±5%
SpO2 accuracy ±3%
Recorder operation
Alarm function
Complete model-specific performance testing
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent)
INCU incubator analyzer (or equivalent)
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Preventive maintenance
Clean vents and filters
Replace battery every 24 months
Complete model-specific preventive maintenance
Performance testing
Verify unit operates on battery
Fan operation
Temperature accuracy ± 0.3 ºC
Temperature alarms
Alarm function
Complete model-specific performance testing
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: DPM 4 Pressure Meter (or equivalent)
Stopwatch or watch with a second hand
Tubing and connectors to connect to DPM 4
PVC pipe to attach cuff to
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Preventive maintenance
Check condition of tubing, cuffs, and hoses
Complete model-specific preventive maintenance
Performance testing
Gauge zero ± 1 mmHg
Leak test ≤ 15 mmHg/min
Pressure accuracy ± 3 mmHg
Complete model-specific performance testing
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent)
Oscilloscope such as Fluke 199XRAY
Stopwatch or watch with a second hand
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Preventive maintenance
Inspect leads and electrodes
Complete model-specific preventive maintenance
Performance testing
Output accuracy ± 10 %
Timer accuracy ± 10 sec
Alarm function
Complete model-specific performance testing
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent)
UW5 Ultrasound wattmeter (or equivalent)
Stopwatch or watch with a second hand
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Preventive maintenance
Inspect sound head
Complete model-specific preventive maintenance
Performance testing
Verify unit operates on battery
Output accuracy ± 20 %
Duty cycle
Timer accuracy ± 10 sec
Alarm function
Complete model-specific performance testing
Equipment information
Control number:____________________________________ Hospital:_ __________________________________________
Manufacturer:______________________________________ Model:_____________________________________________
Serial number:_ ____________________________________ Location:___________________________________________
Test information
Technician:________________________________________ Date:_______________________________________________
Test type: Incoming_ ________ Post repair_________
Test equipment needed: ISA 601 Electrical Safety Analyzer (or equivalent)
VT PLUS HF Ventilator Analyzer (or equivalent)
Test lung (such as ACCULUNG)
Hoses and connectors to connect to VT PLUS HF
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety
Ground wire resistance < 0.3 W
Chassis leakage < 100 µA NC,
< 500 µA SFC
Patient leakage current < 100 µA B and BF
< 10 µA CF
Patient lead leakage current – isolation test < 100 µA BF
(mains on patient applied part) < 10 µA CF
Insulation test (optional) 500 V < 2 MW
Preventive maintenance
Clean vents and filters
Replace tubing
Replace battery every 24 months
Complete model-specific preventive maintenance
Test Result
Pass Fail N/A
Performance testing
Verify unit operates on battery
Gas cylinders and regulators
Hoses, tubing, and connectors
Volume accuracy ± 10 %
Respiration rate
I:E ratio
Pressure accuracy ± 10 %
PEEP
O2 accuracy ±2%
Alarm function
Complete model-specific performance testing
Performance inspection
Verify unit operates on battery: Check that
the ac power indicator is lit when the power
cord is plugged into an outlet. Unplug the ac
power cord and perform the remainder of the
functional test on battery power. The ac power
indicator should go out when the power cord
is unplugged and the battery indicator should
light. Be sure to plug the power cord in at the
conclusion of the test.
Gas cylinders and regulators: Check the
condition of gas cylinders and regulators.
Remove each cylinder and verify that the index
pins are present on the cylinder yoke. Verify
that cylinders have the correct color coding and
labeling and that the cylinders are within their
expiration dates. Replace the cylinders in their
performance that needs to be met in order to requirements in addition to the parent standard.
ensure the safe use of medical equipment. These standards are general in nature, like the
A medical equipment maintenance program parent standard, and are applicable to all medi-
needs to meet or exceed all local standards. cal equipment. Particular standards contain
All relevant standards should be reviewed to requirements that are exceptions to the parent
ensure program compliance. The following and collateral standards. These types of stan-
Chapter discusses some standards that biomed- dards are specific to a device type. IEC 60601
ical personnel should be familiar with. is the parent standard. Collateral standards are
labeled as 60601-1-xx and particular standards
IEC 60601-1 are labeled as 60601-2-xx, with xx represent-
The International Electrotechnical Commission, ing a specific document.
IEC, is a worldwide organization that pro- IEC 60601-1 is mainly used by manufacturers
motes global standardization in the electronics of medical equipment. Medical equipment that
industry. IEC 60601-1, titled Medical electri- is manufactured to this standard has been sub-
cal equipment - Part 1: General requirements jected to rigorous safety and performance tests
for basic safety and essential performance, and has met quality assurance specifications.
addresses the issues of safely designing medi-
cal equipment and serves as the foundation for IEC 62353
safe manufacturing practices. This standard is IEC 62353 is an international standard pub-
mainly used in the design and manufacture of lished by the International Electrotechnical
medical equipment. Commission, a worldwide organization that
In 2005, the third edition of 60601-1 was promotes global standardization in the elec-
published. The object of the standard is to pro- tronics industry. The standard deals with the
vide general requirements for safety of medical testing of medical equipment before first use,
devices and to provide the basis for more spe- after servicing, or periodic safety inspections.
cific standards. This edition combines product The standard specifies how to test for elec-
requirements with manufacturing processes trical safety and gives limits for acceptable
such as risk management. This edition also measurements. Specific tests for measuring the
addresses the concept of essential performance, protective earth resistance, leakage current,
parts of the equipment operation that directly applied part leakage current, and insulation
affect the safety of the patient and operators. resistance are outlined. These terms are
The rationale behind the standard is to iden- defined as:
tify specific hazards associated with medical Protective earth resistance: Sometimes
equipment and to define an acceptable level of referred to as ground wire resistance. Resis-
risk for each hazard. Additionally, it provides tance between any conductive part of the
an objective test to determine if the risks have equipment and the protective connector of
been acceptably minimized, while avoiding the main power supply plug, the protective
requirements defining how to minimize risks. connector of the appliance inlet, or the protec-
This standard is not intended to be used tive conductor permanently connected to the
alone, as it addresses general safety issues supply mains.
applied broadly across medical equipment. More Equipment leakage current: Current flow-
specific standards need to be applied to specific ing from the supply mains to earth through
types of medical equipment. The 60601 family the protective earth conductor and accessible
of standards contains collateral and particu- conductive parts.
lar standards. Collateral standards contain
J. Tobey Clark, MSEE CCE, is the Director, Michael W. Lane, MBA, is the Associate
Instrumentation and Technical Service, at Director, Instrumentation and Technical
the University of Vermont. He leads the Services, at the University of Vermont.
Technical Services Program, a 26 hospital He manages the operations of Technical
shared service clinical engineering program Services Program, a 26 hospital shared
serving Vermont, upstate New York, and service clinical engineering program serving
northern New Hampshire. Tobey also directs Vermont, upstate New York, and northern
the Instrumentation & Model Facility (IMF) New Hampshire. He holds Certification as a
which designs, develops, fabricates and Quality Manager from the American Society
services custom research instruments for for Quality. Michael is a member of the
the University of Vermont community. He Vermont Council for Quality and serves as a
has a faculty appointment in the School of state examiner for Performance Excellence.
Engineering and the College of Nursing Michael is a member of the Association for the
and Health Sciences where he teaches Advancement of Medical Instrumentation
medical instrumentation courses. Tobey is and of the American Society for Field
involved in a number of professional activities Service Managers.
including serving as a board member of the
ACCE Healthcare Technology Foundation
and as an advisor to the World and Pan
American Health Organizations. He was the
2002 recipient of the Clinical/Biomedical
Engineering Career Achievement award
from the Association for the Advancement
of Medical Instrumentation and the 2008
Professional Achievement in Management
award from the American College of Clinical
Engineering. Tobey is currently supported by
Leah Rafuse, BSME, is a clinical engineer
with Technical Services Program, a 26
several grants related to medical technology
hospital shared service clinical engineering
education and international clinical
Program, at the University of Vermont. Leah is
engineering exchange.
responsible for clinical engineering services
for eight hospitals in upstate New York. Leah
is a graduate of the University of Vermont’s
engineering program. Prior to taking over
clinical engineering services in New York,
Leah worked with Technical Services Program
as a biomedical equipment technician,
specializing in anesthesia equipment.
©2009 Fluke Biomedical.
1/2009 3276553 C-EN-N Rev A