Cermax ®

Products and Specifications

.
Short Arc:
Xenon lamps, modules

Medical Endoscopy

Video Projection

Surgical Headlamps

Surgical Microscopy

Dental Curing & Bleaching

Phototherapy

Biofluerescence

Excelitas Technologies’ Cermax®


xenon short arc lamps and associated operating
equipment are a unique and innovative approach to
many challenging and demanding lighting applications.
Cermax® lamp technology is the leading technology
for use in diagnostic and surgical endoscopes,
surgical microscopes, surgical headlamps as well
as a variety of high performance video and home
theatre projectors.
and power supplies.
Safe and compact solution Complete solutions
Utilizing an integrated parabolic or ellipsoidal reflector, Excelitas Technologies lamp operating
Cermax® lamps produce high intensity, collimated or equipment, including DC power supplies, lamp holders
focused output of light. Due to the xenon lamps broad (modules and heat sinks), light engines and complete fiber-
color spectrum, the lamp is filtered to emit either optic illuminators are also included in this
visible, UV or IR light depending on application or specification sheet. All operating equipment has been
usage. With their internal reflector and rugged ceramic specifically designed to work perfectly with Cermax® lamps
body and seal construction, Cermax® lamps are a providing the correct amount of current, voltage, ignition
safe and compact alternative to conventional quartz pulses, and cooling.
xenon lamps making them ideal for such applications
as medical endoscopy, fiber optic illumination and High performance
video projection. Although the overall efficacy of Cermax® lamps is
lower than some alternative technologies, Cermax®
Due to it’s new extended operating life, output xenon lamps provide extremely high brightness due to
degradation curve, instant on-off, DC operation and their DC operation, high pressure and ultra-short arc
mercury-free content, Excelitas’ new Cermax® lamps are gaps. They also don’t require external reflectors, optics
an excellent alternative solution to high pressure or color balancing (to achieve 6000 K) which greatly
mercury lamps for the newly emerging home theatre lowers the usable efficacy of other lamp technologies.
and RPTV markets. It is the usable, focused, color balance light that is
important and Cermax® lamps achieve tremendous
Flexible power ranges performance parameters as compared to any other
Available in two configurations —parabolic and lamp technology available today.
ellipsoidal, Standard Cermax® lamps have operating
power ranges from 80W to 1000W. The newly designed
Cermax® lamps are available only with ellipsoidal
reflectors and are offered with operating power ranges
from 270W to 2700W.

Solid construction
Lamp construction for all lamps is entirely of metal,
sapphire and ceramic. No organic (carbon-based)
materials, mercury, rare-earth elements, or any other
materials with disposal problems are used it the
construction of Cermax® lamps. The xenon fill gas is an
inert gas and is non-toxic.
4 Type A Elliptical
PE80A-10F PE80A-10UV PE80A-13F PE80A-13UV PE150A-10F PE150A-10UV PE150A-13F PE150A-13UV
Power (Watts)_________________________ Nominal 85 85 85 85 150 150 150 150
Range 85 - 100 85 - 100 85 - 100 85 - 100 100 - 150 100 - 150 100 - 150 100 - 150
Current (Amps DC)_____________________ Nominal 7.7 7.7 7.7 7.7 12.5 12.5 12.5 12.5
Range 7.0- 10.0 7.0- 10.0 7.0- 10.0 7.0- 10.0 10.0- 14.0 10.0- 14.0 10.0- 14.0 10.0- 14.0
Voltage (Volts DC)______________________ Nominal 11.0 11.0 11.0 11.0 11.7 11.7 11.7 11.7
Range 10.5- 13.5 10.5- 13.5 10.5- 13.5 10.5- 13.5 10.0- 13.6 10.0- 13.6 10.0- 13.6 10.0- 13.6
Maximum Top Ceramic Operating Temperature (°C)1 120 120 120 120 120 120 120 120
Minimum Ignition Voltage at Lamp (kV)2 28 28 28 28 28 28 28 28
Minimum Ignition Pulse Duration (nS) 100 100 100 100 100 100 100 100

Radiant Output (Watts)3 7.0 7.0 7.0 7.0 21.0 21.0 21.0 21.0
UV Output, <390nm (Watts)3 0.4 0.9 0.4 0.9 1.1 2.5 1.1 2.5
IR Output, >770nm (Watts)3 4.0 3.6 4.0 3.6 12.0 10.8 12.0 10.8
Visible Output, 390-770nm (Lumens)3 800 720 800 720 1800 1620 1800 1620
Color Temperature (Kelvin)4 5900 5050 5900 5050 5900 5050 5900 5050
Spot Size at Crossover_____________ @ 50% points4 .10 / 2.5 .10 / 2.6 .11 / 2.8 .11 / 2.9 .10 / 2.5 .10 / 2.6 .11 / 2.8 .11 / 2.9
(inches / mm) @ 10% points4 .25 / 6.4 .25 / 6.5 .25 / 6.4 .25 / 6.5 .25 / 6.4 .25 / 6.5 .25 / 6.4 .25 / 6.5

f-Number 1.0 1.0 1.3 1.3 1.0 1.0 1.3 1.3


Focal Distance “A” (inches / mm) .242 / 6.15 .242 / 6.15 .490 / 12.45 .490 / 12.45 .242 / 6.15 .242 / 6.15 .490 / 12.45 .490 / 12.45
Cathode Defocus “D” (inches / mm) .014 / .36 .014 / .36 .014 / .36 .014 / .36 .014 / .36 .014 / .36 .014 / .36 .014 / .36
Arc Gap (inches / mm) .025 / .64 .025 / .64 .025 / .64 .025 / .64 .032 / .81 .032 / .81 .032 / .81 .032 / .64
Window Diameter (inches / mm) .770 / 19.56 .770 / 19.56 .770 / 19.56 .770 / 19.56 .770 / 19.56 .770 / 19.56 .770 / 19.56 .770 / 19.56

Focused Output____________________ 6mm Aperture 920 / 8.4 - 660 / 6.0 - 1500 / 14.0 - 1100 / 10.0 -
(Lumens / Watts) 3mm Aperture 590 / 5.2 - 420 / 3.7 - 1000 / 8.5 - 700 / 6.3 -
1
Optimum operating temperature 120-130 (°C). 2 With a PKI approved power supply. 3
Total nominal output from lamp in all directions. 4
Nominal value. All specifications subject to change without notice.

Spectral Output

Relative Spectral Distribution


1.0 UV F

0.8

0.6

0.4

0.2

0.0
200 300 400 500 600 700 800 900 1000 1100
Wavelength (nm)

Type A Elliptical Dimensions Type A Elliptical Reflector

NOTES: A. Lamp must not be operated with window facing upwards within 45° of vertical. B. Seal temperature must not exceed 150°C. C. Current/power regulated power supplies and Excelitas lamp housing units are recommended. D. Lamp must be operated
within recommended current and power range. Over powering may lead to arc instability, hard starting and premature aging. E. Hot mirror assembly is available for IR filtering for 300W lamps and below. F. Cermax® lamps are much safer lamps to use than their quartz
xenon arc lamp equivalents. However, caution must be practiced when operating lamps because they are under high pressure, require high voltage, reach temperatures up to 200°C, and their IR and UV radiation can cause skin burns and eye damage. Read hazard sheet
included with each lamp shipment.
Type B Elliptical 5

PE125B-10F PE125B-10UV PE175B-10F PE175B-10UV PE300B-10F PE300B-10UV


Power (Watts)_________________________ Nominal 125 125 175 175 300 300
Range 75 - 150 75 - 150 150 - 200 150 - 200 175 - 305 175 - 305
Current (Amps DC)______________________ Nominal 11 11 14 14 22 22
Range 7 - 14 7 - 14 12 - 16 12 - 16 13 - 23 13 - 23
Voltage (Volts DC)______________________ Nominal 11.0 11.0 12.5 12.5 13.5 13.5
Range 10.0 - 12.5 10.0 - 12.5 10.5 - 13.5 10.5 - 13.5 11.5 - 15.0 11.5 - 15.0
Maximum Top Ceramic Operating Temperature (°C)1 150 150 150 150 150 150
Minimum Ignition Voltage at Lamp (kV)2 23 23 23 23 23 23
Minimum Ignition Pulse Duration (nS) 100 100 100 100 100 100

Radiant Output (Watts)3 18.5 18.5 34.0 34.0 65.0 65.0


UV Output, <390nm (Watts)3 1.1 2.7 1.5 4.9 3.3 8.5
IR Output, >770nm (Watts)3 10.3 9.7 18.8 18.0 32.0 30.0
Visible Output, 390-770nm (Lumens)3 1570 1400 2740 2460 6900 5850
Color Temperature (Kelvin)4 5900 5050 5900 5050 5900 5050
Spot Size at Crossover_____________ @ 50% points4 .09 / 2.3 .09 / 2.3 .09 / 2.3 .09 / 2.3 .09 / 2.3 .09 / 2.3
(inches / mm) @ 10% points4 .23 / 5.8 .23 / 5.8 .23 / 5.8 .23 / 5.8 .23 / 5.8 .23 / 5.8

f-Number 1.0 1.0 1.0 1.0 1.0 1.0


Focal Distance “A” (inches / mm) .550 / 13.97 .550 / 13.97 .550 / 13.97 .550 / 13.97 .550 / 13.97 .550 / 13.97
Cathode Defocus “D” (inches / mm) .008 / .20 .008 / .20 .008 / .20 .008 / .20 .008 / .20 .008 / .20
Arc Gap (inches / mm) .038 / .97 .038 / .97 .038 / .97 .038 / .97 .038 / .97 .038 / .97
Window Diameter (inches / mm) 1.00 / 25.4 1.00 / 25.4 1.00 / 25.4 1.00 / 25.4 1.00 / 25.4 1.00 / 25.4

Focused Output___________________ 6mm Aperture 1250 / 11 - 1930 / 18.5 - 3500 / 32 -


(Lumens / Watts) 3mm Aperture 650 / 6.4 - 1170 / 9 - 2000 / 20 -
1
Optimum operating temperature 120-130 (°C). 2 With a PKI approved power supply. 3
Total nominal output from lamp in all directions. 4
Nominal value. All specifications subject to change without notice.

Spectral Output

Relative Spectral Distribution


1.0 UV F

0.8

0.6

0.4

0.2

0.0
200 300 400 500 600 700 800 900 1000 1100
Wavelength (nm)

Type B Dimensions Type B Reflector

NOTES: A. Lamp must not be operated with window facing upwards within 45° of vertical. B. Seal temperature must not exceed 150°C. C. Current/power regulated power supplies and Excelitas lamp housing units are recommended. D. Lamp must be operated
within recommended current and power range. Over powering may lead to arc instability, hard starting and premature aging. E. Hot mirror assembly is available for IR filtering for 300W lamps and below. F. Cermax® lamps are much safer lamps to use than their quartz
xenon arc lamp equivalents. However, caution must be practiced when operating lamps because they are under high pressure, require high voltage, reach temperatures up to 200°C, and their IR and UV radiation can cause skin burns and eye damage. Read hazard sheet
included with each lamp shipment.
6 Type C Elliptical
PE300C-10F PE300C-10UV PE300C-13F PE300C-13UV PE500C-10F PE500C-10UV PE500C-13F PE500C-13UV
Power (Watts)_________________________ Nominal 300 300 300 300 500 500 500 500
Range 175 - 300 175 - 300 175 - 300 175 - 300 350 - 525 350 - 525 350 - 525 350 - 525
Current (Amps DC)______________________ Nominal 22.0 22.0 22.0 22.0 32.0 32.0 32.0 32.0
Range 13.0 - 23.0 13.0 - 23.0 13.0 - 23.0 13.0 - 23.0 23.0 - 35.0 23.0 - 35.0 23.0 - 35.0 23.0 - 35.0
Voltage (Volts DC)______________________ Nominal 13.5 13.5 13.5 13.5 15.5 15.5 15.5 15.5
Range 11.5 - 15.0 11.5 - 15.0 11.5 - 15.0 11.5 - 15.0 14.0 - 16.5 14.0 - 16.5 14.0 - 16.5 14.0 - 16.5
Maximum Top Ceramic Operating Temperature (°C)1 150 150 150 150 150 150 150 150
Minimum Ignition Voltage at Lamp (kV)2 23 23 23 23 30 30 30 30
Minimum Ignition Pulse Duration (nS) 100 100 100 100 100 100 100 100

Radiant Output (Watts)3 75 75 75 75 112 112 112 112


UV Output, <390nm (Watts)3 3.8 9.7 3.8 9.7 5.0 10.3 5.0 10.3
IR Output, >770nm (Watts)3 37.0 34.5 37.0 34.5 65.0 63.0 65.0 63.0
Visible Output, 390-770nm (Lumens)3 7475 6725 7475 6725 10500 9550 10500 9550
Color Temperature (Kelvin)4 5900 5050 5900 5050 5900 5050 5900 5050
Spot Size at Crossover_____________ @ 50% points4 .06 / 1.5 .06 / 1.5 .07 / 1.8 .07 / 1.8 .07 / 1.8 .07 / 1.8 .07 / 1.8 .07 / 1.8
(inches / mm) @ 10% points4 .21 / 5.3 .21 / 5.3 .23 / 5.8 .23 / 5.8 .21 / 5.3 .21 / 5.3 .23 / 5.8 .23 / 5.8

f-Number 1.0 1.0 1.3 1.3 1.0 1.0 1.3 1.3


Focal Distance ”A” (inches / mm) .75 / 19.1 .75 / 19.1 1.10 / 27.9 1.10 / 27.9 .75 / 19.1 .75 / 19.1 1.10 / 27.9 1.10 / 27.9
Cathode Defocus “D” (inches / mm) .005 / .13 .005 / .13 .005 / .13 .005 / .13 .005 / .13 .005 / .13 .005 / .13 .005 / .13
Arc Gap (inches / mm) .045 / 1.14 .045 / 1.14 .045 / 1.14 .045 / 1.14 .045 / 1.14 .045 / 1.14 .045 / 1.14 .045 / 1.14
Window Diameter (inches / mm) 1.00 / 25.4 1.00 / 25.4 1.00 / 25.4 1.00 / 25.4 1.00 / 25.4 1.00 / 25.4 1.00 / 25.4 1.00 / 25.4

Focused Output___________________ 12mm Aperture - - - - - - - -


(Lumens / Watts) 9mm Aperture - - - - - - - -
6mm Aperture 4500 / 37 - 3100 / 28.5 - 7700 / 75 - 7200 / 70 -
3mm Aperture 2300 / 23 - 1750 / 17.5 - 4085 / 39 - 3600 / 35 -

1
Optimum operating temperature 120-130 (°C). 2 With a PKI approved power supply. 3 Total nominal output from lamp in all directions. 4
Nominal value. All specifications subject to change without notice.

Spectral Output

Relative Spectral Distribution


1.0 UV F

0.8

0.6

0.4

0.2

0.0
200 300 400 500 600 700 800 900 1000 1100
Wavelength (nm)
Type C Elliptical 7

PE700C-10F PE700C-10UV PE700C-13F PE700C-13UV


Power (Watts)_________________________ Nominal 700 700 700 700
Range 600 - 750 600 - 750 600 - 750 600 - 750
Current (Amps DC)______________________ Nominal 41.5 41.5 41.5 41.5
Range 35.0 - 46.0 35.0 - 46.0 35.0 - 46.0 35.0 - 46.0
Voltage (Volts DC)______________________ Nominal 16.85 16.85 16.85 16.85
Range 15.5 - 18.0 15.5 - 18.0 15.5 - 18.0 15.5 - 18.0
Maximum Top Ceramic Operating Temperature (°C)1 180 180 180 180
Minimum Ignition Voltage at Lamp (kV)2 38 38 38 38
Minimum Ignition Pulse Duration (nS) 100 100 100 100

Radiant Output (Watts)3 190 190 190 190


UV Output, <390nm (Watts)3 8.5 17.0 8.5 17.0
IR Output, >770nm (Watts)3 100.0 97.0 100.0 97.0
Visible Output, 390-770nm (Lumens)3 21000 18600 21000 18600
Color Temperature (Kelvin)4 5900 5050 5900 5050
Spot Size at Crossover_____________ @ 50% points4 .10 / 2.5 .10 / 2.5 .10 / 2.5 .10 / 2.5
(inches / mm) @ 10% points4 .21 / 5.3 .21 / 5.3 .23 / 5.8 .23 / 5.8

f-Number 1.0 1.0 1.3 1.3


Focal Distance ”A” (inches / mm) .75 / 19.1 .75 / 19.1 1.10 / 27.9 1.10 / 27.9
Cathode Defocus “D” (inches / mm) .005 / .13 .005 / .13 .005 / .13 .005 / .13
Arc Gap (inches / mm) .045 / 1.14 .045 / 1.14 .045 / 1.14 .045 / 1.14
Window Diameter (inches / mm) 1.00 / 25.4 1.00 / 25.4 1.00 / 25.4 1.00 / 25.4

Focused Output___________________ 12mm Aperture 20000 / 193 17800 / 172 18500 / 180 16500 / 175
(Lumens / Watts) 9mm Aperture 18000 / 175 16800 / 162 16500 / 160 14650 / 150
6mm Aperture 12100 / 117 10850 / 105 11232 / 109 10200 / 109
3mm Aperture 7300 / 65 6500 / 62 6500 / 60 5700 / 54

1
Optimum operating temperature 120-130 (°C). 2 With a PKI approved power supply. 3 Total nominal output from lamp in all directions. 4
Nominal value. All specifications subject to change without notice.

Type C Dimensions Type C Reflector

NOTES: A. Lamp must not be operated with window facing upwards within 45° of vertical. B. Seal temperature must not exceed 150°C. C. Current/power regulated power supplies and Excelitas lamp housing units are recommended. D. Lamp must be operated
within recommended current and power range. Over powering may lead to arc instability, hard starting and premature aging. E. Hot mirror assembly is available for IR filtering for 300W lamps and below. F. Cermax® lamps are much safer lamps to use than their quartz
xenon arc lamp equivalents. However, caution must be practiced when operating lamps because they are under high pressure, require high voltage, reach temperatures up to 200°C, and their IR and UV radiation can cause skin burns and eye damage. Read hazard sheet
included with each lamp shipment.
8 Type D Elliptical
PE1000D-10F PE1000D-10UV PE1000D-13F PE1000D-13UV PE1200D-10F PE1200D-10UV PE1200D-13F PE1200D-13UV
Power (Watts)_________________________ Nominal 1000 1000 1000 1000 1200 1200 1200 1200
Range 600 - 1050 600 - 1050 600 - 1050 600 - 1050 950 - 1250 950 - 1250 950 - 1250 950 - 1250
Current (Amps DC)______________________ Nominal 50 50 50 50 63 63 63 63
Range 30 - 56 30 - 56 30 - 56 30 - 56 45 - 70 45 - 70 45 - 70 45 - 70
Voltage (Volts DC)______________________ Nominal 20.0 20.0 20.0 20.0 19.0 19.0 19.0 19.0
Range 15.0 - 24.0 15.0 - 24.0 15.0 - 24.0 15.0 - 24.0 17.0 - 22.0 17.0 - 22.0 17.0 - 22.0 17.0 - 22.0
Maximum Top Ceramic Operating Temperature (°C)1 150 150 150 150 160 160 160 160
Minimum Ignition Voltage at Lamp (kV)2 28 28 28 28 28 28 28 28
Minimum Ignition Pulse Duration (nS) 100 100 100 100 100 100 100 100

Radiant Output (Watts)3 245 245 245 245 273 273 273 273
UV Output, <390nm (Watts)3 13.0 24.0 13.0 24.0 14.6 37.0 14.6 37.0
IR Output, >770nm (Watts)3 135 132 135 132 150 146 150 146
Visible Output, 390-770nm (Lumens)3 23500 21000 23500 21000 30000 27000 30000 27000
Color Temperature (Kelvin)4 5900 5050 5900 5050 5900 5050 5900 5050
Spot Size at Crossover_____________ @ 50% points4 .14 / 3.5 .14 / 3.5 .15 / 3.8 .15 / 3.8 .15 / 3.8 .15 / 3.8 .16 / 4.1 .16 / 4.1
(inches / mm) @ 10% points4 .33 / 8.4 .33 / 8.4 .35 / 8.9 .35 / 8.9 .40 / 10.3 .40 / 10.3 .53 / 13.5 .53 / 13.5

f-Number 1.0 1.0 1.3 1.3 1.0 1.0 1.3 1.3


Focal Distance “A” (inches / mm) .946 / 24.03 .946 / 24.03 1.389 / 35.28 1.389 / 35.28 .946 / 24.03 .946 / 24.03 1.389 / 35.28 1.389 / 35.28
Cathode Defocus “D” (inches / mm) .008 / .20 .008 / .20 .008 / .20 .008 / .20 .008 / .20 .008 / .20 .008 / .20 .008 / .20
Arc Gap (inches / mm) .062 / 1.57 .062 / 1.57 .062 / 1.57 .062 / 1.57 .067 / 1.70 .067 / 1.70 .067 / 1.70 .067 / 1.70
Window Diameter (inches / mm) 1.375 / 34.92 1.375 / 34.92 1.375 / 34.92 1.375 / 34.92 1.375 / 34.92 1.375 / 34.92 1.375 / 34.92 1.375 / 34.92

Focused Output___________________ 12mm Aperture 21000 / 200 18900 / 180 18000 / 163 16200 / 147 23250 / 208 21000 / 187 22000 / 195 19725 / 176
(Lumens / Watts) 9mm Aperture - - - - 19500 / 185 17250 / 164 18750 / 178 16875 / 158
6mm Aperture 13000 / 120 11700 / 108 12000 / 110 10800 / 99 13500 / 128 12000 / 113 12000 / 108 10800 / 97
3mm Aperture 4800 / 47 4320 / 42 3500 / 33 3150 / 30 5700 / 54 5100 / 49 4200 / 40 3780 / 36

1
Optimum operating temperature 120-130 (°C). 2 With a PKI approved power supply. 3 Total nominal output from lamp in all directions. 4
Nominal value. All specifications subject to change without notice.

Spectral Output

Relative Spectral Distribution


1.0 UV F

0.8

0.6

0.4

0.2

0.0
200 300 400 500 600 700 800 900 1000 1100
Wavelength (nm)
Type D Elliptical 9

PE1500D-10F PE1500D-10UV PE1500D-13F PE1500D-13UV


Power (Watts)_________________________ Nominal 1500 1500 1500 1500
Range 900 - 1550 900 - 1550 900 - 1550 900 - 1550
Current (Amps DC)______________________ Nominal 65 65 65 65
Range 45 - 80 45 - 80 45 - 80 45 - 80
Voltage (Volts DC)______________________ Nominal 18.5 18.5 18.5 18.5
Range 17.0 - 23.0 17.0 - 23.0 17.0 - 23.0 17.0 - 23.0
Maximum Top Ceramic Operating Temperature (°C)1 160 160 160 160
Minimum Ignition Voltage at Lamp (kV)2 28 28 28 28
Minimum Ignition Pulse Duration (nS) 100 100 100 100

Radiant Output (Watts)3 365 365 365 365


UV Output, <390nm (Watts)3 19.5 49.0 19.5 49.0
IR Output, >770nm (Watts)3 200 195 200 195
Visible Output, 390-770nm (Lumens)3 40000 36000 40000 36000
Color Temperature (Kelvin)4 5900 5050 5900 5050
Spot Size at Crossover_____________ @ 50% points4 .15 / 3.8 .15 / 3.8 .16 / 4.1 .16 / 4.1
(inches / mm) @ 10% points4 .40 / 10.3 .40 / 10.3 .53 / 13.5 .53 / 13.5

f-Number 1.0 1.0 1.3 1.3


Focal Distance “A” (inches / mm) .946 / 24.03 .946 / 24.03 1.389 / 35.28 1.389 / 35.28
Cathode Defocus “D” (inches / mm) .008 / .20 .008 / .20 .008 / .20 .008 / .20
Arc Gap (inches / mm) .075 / 1.91 .075 / 1.91 .075 / 1.91 .075 / 1.91
Window Diameter (inches / mm) 1.375 / 34.92 1.375 / 34.92 1.375 / 34.92 1.375 / 34.92

Focused Output___________________ 12mm Aperture 31000 / 278 28000 / 250 29000 / 280 26000 / 235
(Lumens / Watts) 9mm Aperture 26000 / 247 23000 / 225 25000 / 238 22500 / 211
6mm Aperture 18000 / 171 16000 / 154 16000 / 145 14400 / 130
3mm Aperture 7500 / 72 6800 / 65 5600 / 53 5040 / 48

1
Optimum operating temperature 120-130 (°C). 2 With a PKI approved power supply. 3 Total nominal output from lamp in all directions. 4
Nominal value. All specifications subject to change without notice.

Type D Dimensions Type D Reflector

NOTES: A. Lamp must not be operated with window facing upwards within 45° of vertical. B. Seal temperature must not exceed 150°C. C. Current/power regulated power supplies and Excelitas lamp housing units are recommended. D. Lamp must be operated
within recommended current and power range. Over powering may lead to arc instability, hard starting and premature aging. E. Hot mirror assembly is available for IR filtering for 300W lamps and below. F. Cermax® lamps are much safer lamps to use than their quartz
xenon arc lamp equivalents. However, caution must be practiced when operating lamps because they are under high pressure, require high voltage, reach temperatures up to 200°C, and their IR and UV radiation can cause skin burns and eye damage. Read hazard sheet
included with each lamp shipment.
10 Type A Parabolic
PE80AF PE80AUV PE150AF PE150AUV
Power (Watts)_________________________ Nominal 85 85 150 150
Range 85 - 100 85 - 100 100 - 150 100 - 150
Current (Amps DC)______________________ Nominal 7.7 7.7 12.5 12.5
Range 7.0- 10.0 7.0- 10.0 10.0- 14.0 10.0- 14.0
Voltage (Volts DC)______________________ Nominal 11.0 11.0 11.7 11.7
Range 10.5- 13.5 10.5- 13.5 10.0- 13.6 100 - 13.6
Maximum Top Ceramic Operating Temperature (°C)1 120 120 120 120
Minimum Ignition Voltage at Lamp (kV)2 28 28 28 28
Minimum Ignition Pulse Duration (nS) 100 100 100 100

Radiant Output (Watts)3 6.0 6.0 16.0 16.0


UV Output, <390nm (Watts)3 0.3 0.7 0.9 2.0
IR Output, >770nm (Watts)3 3.2 2.9 8.0 7.2
Visible Output, 390-770nm (Lumens)3 650 580 1350 1210
Color Temperature (Kelvin)4 5900 5050 5900 5050

Cathode Defocus “D” (inches / mm) .014 / .36 .014 / .36 .014 / .36 .014 / .36
Arc Gap (inches / mm) .025 / .64 .025 / .64 .032 / .81 .032 / .81
Window Diameter (inches / mm) .770 / 19.56 .770 / 19.56 .770 / 19.56 .770 / 19.56

Focused Output with f/1.0 lens________ 6mm Aperture 540 / 4.8 - 900 / 8.0 -
(Lumens / Watts) 3mm Aperture 350 / 3.3 - 700 / 5.5 -
1
Optimum operating temperature 120-130 (°C). 2 With a PKI approved power supply. 3 Total nominal output from lamp in all directions. 4
Nominal value. All specifications subject to change without notice.

Spectral Output

Relative Spectral Distribution


1.0 UV F

0.8

0.6

0.4

0.2

0.0
200 300 400 500 600 700 800 900 1000 1100
Wavelength (nm)

Type A Parabolic Dimensions Type A Parabolic Reflector

NOTES: A. Lamp must not be operated with window facing upwards within 45° of vertical. B. Seal temperature must not exceed 150°C. C. Current/power regulated power supplies and Excelitas lamp housing units are recommended. D. Lamp must be operated
within recommended current and power range. Over powering may lead to arc instability, hard starting and premature aging. E. Hot mirror assembly is available for IR filtering for 300W lamps and below. F. Cermax® lamps are much safer lamps to use than their quartz
xenon arc lamp equivalents. However, caution must be practiced when operating lamps because they are under high pressure, require high voltage, reach temperatures up to 200°C, and their IR and UV radiation can cause skin burns and eye damage. Read hazard sheet
included with each lamp shipment.
Type B Parabolic 11

PE125BF PE125BUV PE175BF PE175BUV PE300BF PE300BUV


Power (Watts)_________________________ Nominal 125 125 175 175 300 300
Range 75 - 150 75 - 150 150 - 200 150 - 200 180 - 320 180 - 320
Current (Amps DC)______________________ Nominal 12 12 14 14 21 21
Range 6 - 14 6 - 14 12 - 16 12 - 16 10 - 22 10 - 22
Voltage (Volts DC)______________________ Nominal 11.0 11.0 12.5 12.5 14.0 14.0
Range 9.5 - 12.5 9.5 - 12.5 11.0 - 14.0 11.0 - 14.0 13.0 - 16.0 13.0 - 16.0
Maximum Top Ceramic Operating Temperature (°C)1 150 150 150 150 150 150
Minimum Ignition Voltage at Lamp (kV)2 23 23 23 23 23 23
Minimum Ignition Pulse Duration (nS)

Peak Intensity (Candelas) 3.0 x 105 2.8 x 105 3.5 x 105 3.3 x 105 5.2 x 105 4.6 x 105
Radiant Output (Watts)3 17 17 25 25 50 50
UV Output, <390nm (Watts)3 0.8 2.2 1.2 3.5 2.6 6.6
IR Output, >770nm (Watts)3 10.0 9.0 14.0 13.0 28.8 26.8
Visible Output, 390-770nm (Lumens)3 1500 1350 2200 2000 5000 4500
Color Temperature (Kelvin)4 5600 5050 5600 5050 5600 5050

Beam Geometry5______________________ @ 1 Hour 4.5 4.5 4.5 4.5 5.0 5.0


(Degrees) @ 100 Hours 5.0 5.0 5.0 5.0 6.0 6.0
@ 1000 Hours 6.0 6.0 6.0 6.0 7.0 7.0

Cathode Defocus “D” (inches / mm) .003 / .08 .003 / .08 .003 / .08 .003 / .08 .003 / .08 .003 / .08
Arc Gap (inches / mm) .045 / 1.14 .045 / 1.14 .045 / 1.14 .045 / 1.14 .049 / 1.24 .049 / 1.24
Window Diameter (inches / mm) 1.00 / 25.4 1.00 / 25.4 1.00 / 25.4 1.00 / 25.4 1.00 / 25.4 1.00 / 25.4

Focused Output with f/1.0 lens________ 8mm Aperture 1050 / 9.5 - 1400 / 13.0 - 3130 / 29.0 -
(Lumens / Watts) 6mm Aperture 620 / 5.6 - 830 / 7.8 - 1410 / 14.0 -
1
Optimum operating temperature 120-130 (°C). 2 With a PKI approved power supply. 3 Total nominal output from lamp in all directions. 4 Nominal value. 5 Half angle at 10% points. All specifications
subject to change without notice.

Spectral Output

Relative Spectral Distribution


1.0 UV F

0.8

0.6

0.4

0.2

0.0
200 300 400 500 600 700 800 900 1000 1100
Wavelength (nm)

Type B Parabolic Dimensions Type B Parabolic Reflector

NOTES: A. Lamp must not be operated with window facing upwards within 45° of vertical. B. Seal temperature must not exceed 150°C. C. Current/power regulated power supplies and Excelitas lamp housing units are recommended. D. Lamp must be operated within
recommended current and power range. Over powering may lead to arc instability, hard starting and premature aging. E. Hot mirror assembly is available for IR filtering for 300W lamps and below. F. Cermax® lamps are much safer lamps to use than their quartz xenon arc lamp
equivalents. However, caution must be practiced when operating lamps because they are under high pressure, require high voltage, reach temperatures up to 200°C, and their IR and UV radiation can cause skin burns and eye damage. Read hazard sheet included with each
lamp shipment.
12 Type D Parabolic
PE500DF PE500DUV PE1000DF PE1000DUV
Power (Watts)_________________________ Nominal 500 500 1000 1000
Range 340 - 575 340 - 575 850 - 1050 850 - 1050
Current (Amps DC)______________________ Nominal 27 27 51 51
Range 20 - 32 20 - 32 46 - 54 46 - 54
Voltage (Volts DC)______________________ Nominal 18.5 18.5 19.5 19.5
Range 16.0- 21.0 16.0- 21.0 18.5- 22.0 18.5- 22.0
Maximum Top Ceramic Operating Temperature (°C)1 150 150 150 150
Minimum Ignition Voltage at Lamp (kV)2 35 35 35 35
Minimum Ignition Pulse Duration (nS) 100 100 100 100

Peak Intensity (Candelas) 2.4 x 106 1.7 x 106 3.8 x 106 2.7 x 106
Radiant Output (Watts)3 112 112 250 250
UV Output, <390nm (Watts)3 5.5 11.5 13.0 25.0
IR Output, >770nm (Watts)3 65 62 137 133
Visible Output, 390-770nm (Lumens)3 10500 9600 24000 22000
Color Temperature (Kelvin)4 5600 5050 5600 5050

Beam Geometry5______________________ @ 1 Hour 3.0 3.0 4.5 4.5


(Degrees) @ 100 Hours 4.0 4.0 5.0 5.0
@ 1000 Hours 5.0 5.0 6.0 6.0

Cathode Defocus “D” (inches / mm) .008 / .20 .008 / .20 .008 / .20 .008 / .20
Arc Gap (inches / mm) .090 / 2.29 .090 / 2.29 .090 / 2.29 .090 / 2.29
Window Diameter (inches / mm) 2.00 / 50.8 2.00 / 50.8 2.00 / 50.8 2.00 / 50.8

Focused Output with f/1.0 lens_______ 6mm Aperture 5200 / 48 - 10400 / 96 -


(Lumens / Watts) 3mm Aperture 2300 / 22 - 4600 / 44 -

1
Optimum operating temperature 120-130 (°C). 2 With a PKI approved power supply. 3
Total nominal output from lamp in all directions. 4
Nominal value. 5
Half angle at 10% points.
All specifications subject to change without notice.

Spectral Output

Relative Spectral Distribution


1.0 UV F

0.8

0.6

0.4

0.2

0.0
200 300 400 500 600 700 800 900 1000 1100
Wavelength (nm)

Type D Parabolic Dimensions Type D Parabolic Reflector

NOTES: A. Lamp must not be operated with window facing upwards within 45° of vertical. B. Seal temperature must not exceed 150°C. C. Current/power regulated power supplies and Excelitas lamp housing units are recommended. D. Lamp must be operated
within recommended current and power range. Over powering may lead to arc instability, hard starting and premature aging. E. Hot mirror assembly is available for IR filtering for 300W lamps and below. F. Cermax® lamps are much safer lamps to use than their quartz
xenon arc lamp equivalents. However, caution must be practiced when operating lamps because they are under high pressure, require high voltage, reach temperatures up to 200°C, and their IR and UV radiation can cause skin burns and eye damage. Read hazard sheet
included with each lamp shipment.
Video Projection
Lighting Products

High out put Xenon for excellent color stability and reliability

DC operation for more lumens through smaller apertures Instant

on/off capabilities

One lamp body size for 1,500 - 3,500 lumen projectors Single lamp

design eliminating light coupling design issues Small compact low

wattage lamp for RPTV

Excelitas Technologies is fueling the digital and true white light to produce the accurate and
revolution with its new Cermax® high pressure brilliant colors that consumers expect from the cinema
xenon lamps. These new lamps provide unparallel experience. Xenon provides the added benefit of being
performance ideal for video projection applications environmentally friendly as it is an inert gas and
and are a safe and compact alternative to conventional non-toxic. Cermax® lamps are also designed with
quartz xenon lamps or high pressure mercury lamps revolutionary performance features: shortest arc gap
currently used in the video market. available in the industry, DC operation to reduce any
Cermax® lamps have been innovatively designed to achieve sequencing and flicker issues, and instant on-off with
an unprecedented 4000 hour life rating which, for the first no warm up or cool down time necessary. Lamp
time, makes xenon a viable alternative to high pressure construction is entirely of metal, sapphire and ceramic.
mercury lamps used in RPTV and Home Theatre projector No organic (carbon-based) materials, mercury,
applications. The xenon gas fill provides a balanced D6500 / rare-earth elements, or any other materials with
D5900 color spectrum disposal problems are used in its construction.
14 Video Projection
XE330C-10FH XE330C-13FH XE420C-10FH XE420C-13FH XE550C-10FH XE550C-13FH XE825C-10FH XE825C-13FH
Power (Watts)_________________________ Nominal 330 330 420 420 550 550 825 825
Range 270 - 390 270 - 390 350 - 420 350 - 420 400 - 575 400 - 575 700 - 750 700 - 750
Current (Amps DC)______________________ Nominal 22.8 22.8 25.8 25.8 34.7 34.7 43.0 43.0
Range 21.5- 26.5 21.5- 26.5 24.2- 27.6 24.2- 27.6 32.0- 37.5 32.0- 37.5 40.0- 45.0 40.0- 45.0
Voltage (Volts DC)______________________ Nominal 15.5 15.5 15.8 15.8 15.9 15.9 17.4 17.4
Range 14.5- 16.5 14.5- 16.5 15.0- 16.8 15.0- 16.8 15.0- 17.5 15.0- 17.5 16.0- 18.5 16.0- 18.5
Maximum Top Ceramic Operating Temperature (°C)1 160 160 160 160 160 160 160 160
Minimum Ignition Voltage at lamp (kV)2 35 35 35 35 35 35 38 38
Minimum Ignition Pulse Duration (nS) 100 100 100 100 100 100 100 100

Radiant Output (Watts)3 69 69 83 83 124 124 170 170


UV Output, <390nm (Watts)3 3.0 3.0 3.7 3.7 6.0 6.0 7.6 7.6
IR Output, >770nm (Watts)3 40 40 48 48 72 72 87 87
Visible Output, 390-770nm (Lumens)3 6800 6800 7800 7800 11700 11700 25000 25000
Color Temperature (Kelvin)4 5900 5900 5900 5900 5900 5900 5900 5900
Spot Size at Crossover_____________ @ 50% points4 .067 / 1.7 .067 / 1.7 .067 / 1.7 .067 / 1.7 .067 / 1.7 .067 / 1.7 .10 / 2.5 .10 / 2.5
(inches / mm) @ 10% points4 .220 / 5.6 .220 / 5.6 .220 / 5.6 .220 / 5.6 .220 / 5.6 .220 / 5.6 .24 / 6.1 .25 / 6.4

f-Number 1.0 1.3 1.0 1.3 1.0 1.3 1.0 1.3


Focal Distance “A” (inches / mm) .75 / 19.1 1.10 / 27.9 .75 / 19.1 1.10 / 27.9 .75 / 19.1 1.10 / 27.9 .75 / 19.1 1.10 / 27.9
Cathode Defocus “D” (inches / mm) .008 / .20 .008 / .20 .008 / .20 .008 / .20 .008 / .20 .008 / .20 .008 / .20 .008 / .20
Arc Gap (inches / mm) .038 / .97 .038 / .97 .038 / .97 .038 / .97 .038 / .97 .038 / .97 .038 / .97 .038 / .97
Window Diameter (inches / mm) 1.00 / 25.4 1.00 / 25.4 1.00 / 25.4 1.00 / 25.4 1.00 / 25.4 1.00 / 25.4 1.00 / 25.4 1.00 / 25.4

Focused Output - XGA/SXGA Available upon request

1 2 3 4
Optimum operating temperature 120-130 (°C). With a PKI approved power supply. Total nominal output from lamp in all directions. Nominal value. All specifications subject to change without notice.

Spectral Output

Relative Spectral Distribution


1.0 UV F

0.8

0.6

0.4

0.2

0.0
200 300 400 500 600 700 800 900 1000 1100
Wavelength (nm)

Type C Dimensions Type C Reflector


Video Projection 15

XE1850D-10FH XE1850D-13FH XE2400E-10FH XE2400E-13FH


Power (Watts)_________________________ Nominal 1850 1850 2400 2400
Range 1400 - 1850 1400 - 1850 1800 - 2700 1800 - 2700
Current (Amps DC)______________________ Nominal 65.0 65.0 104.0 104.0
Range 60.0 - 95.0 60.0 - 95.0 75.0 - 110.0 75.0 - 110.0
Voltage (Volts DC)______________________ Nominal 20.0 20.0 23.0 23.0
Range 19.0 - 26.0 19.0 - 26.0 19.0 - 26.0 19.0 - 26.0
Maximum Top Ceramic Operating Temperature (°C)1 160 160 160 160
Minimum Ignition Voltage at lamp (kV)2 28 28 35 35
Minimum Ignition Pulse Duration (nS) 100 100 100 100

Radiant Output (Watts)3 365 365 635 635


UV Output, <390nm (Watts)3 19.5 19.5 31.0 31.0
IR Output, >770nm (Watts)3 200 200 350 350
Visible Output, 390-770nm (Lumens)3 40000 40000 73000 73000
Color Temperature (Kelvin)4 5900 5900 5900 5900
Spot Size at Crossover_____________ @ 50% points4 .15 / 3.8 .16 / 4.1 .19 / 4.8 .19 / 4.8
(inches / mm) @ 10% points4 .40 / 10.2 .53 / 13.5 .65 / 16.5 .65 / 16.5

f-Number 1.0 1.3 1.0 1.3


Focal Distance “A” (inches / mm) .946 / 24.03 1.389 / 35.28 .825 / 20.95 1.885 / 47.88
Cathode Defocus “D” (inches / mm) .008 / .20 .008 / .20 .012 / .30 .012 / .30
Arc Gap (inches / mm) .053 / 1.35 .053 / 1.35 .075 / 1.91 .075 / 1.91
Window Diameter (inches / mm) 1.375 / 34.93 1.375 / 34.93 2.00 / 50.8 2.00 / 50.8

Focused Output - XGA/SXGA Available upon request

1
Optimum operating temperature 120-130 (°C). With a PKI approved power supply. Total nominal output from lamp in all directions. 4 Nominal value. All specifications subject to change without notice.
2 3

Type D Dimensions Type D Reflector

Type E Dimensions Type E Reflector

NOTES: A. Lamp must not be operated with window facing upwards within 45° of vertical. B. Seal temperature must not exceed 150°C. C. Current/power regulated power supplies and Excelitas lamp housing units are recommended. D. Lamp must be operated within
recommended current and power range. Over powering may lead to arc instability, hard starting and premature aging. E. Hot mirror assembly is available for IR filtering for 300W lamps and below. F. Cermax® lamps are much safer lamps to use than their quartz xenon arc lamp
equivalents. However, caution must be practiced when operating lamps because they are under high pressure, require high voltage, reach temperatures up to 200°C, and their IR and UV radiation can cause skin burns and eye damage. Read hazard sheet included with each
lamp shipment.
.

Excelitas Technologies, 44370 Christy Street, Fremont, CA 94538-3180


P: (510) 979-6500, (800) 775-6786, F: (510) 687-1140
www.excelitas.com