Flexim F704 PDF
Flexim F704 PDF
UMG70XV3-4EN
FLUXUS G704
FLUXUS G704 A2
FLUXUS G709
Firmware V5.xx
remark:
MS-DOS, Excel, Windows are registered trademarks of Microsoft Corporation.
FLUXUS is a registered trademark of FLEXIM GmbH.
FLEXIM GmbH
Wolfener Strasse 36
12681 Berlin
Germany
FLUXUS can be operated in the language of your choice (see section 9.4).
1 Introduction..........................................................................................................9
1.1 Regarding this Manual...........................................................................................9
1.2 Safety Precautions.................................................................................................9
1.3 Warranty ..............................................................................................................10
3 Handling .............................................................................................................24
3.1 First Inspection ....................................................................................................24
3.2 General Precautions ............................................................................................24
3.3 Cleaning...............................................................................................................24
4 Flowmeter...........................................................................................................25
4.1 Measuring Principle .............................................................................................25
4.2 Description of the Flowmeter ...............................................................................26
4.3 Keyboard .............................................................................................................28
UMG70XV3-4EN 12.01.2009
6.9 Connection of the Serial Interface ....................................................................... 56
6.10 Connection of the Sensor Module (SENSPROM) ............................................... 56
9 Start-up .............................................................................................................. 76
9.1 Switching on ........................................................................................................ 76
9.2 Displays............................................................................................................... 76
9.3 HotCodes ............................................................................................................ 78
9.4 Language Selection ............................................................................................ 79
9.5 Operation State Indication (G709) ...................................................................... 79
9.6 Interruption of the Power Supply ......................................................................... 79
UMG70XV3-4EN 12.01.2009
11 Displaying the Measured Values ................................................................... 100
11.1 Selection of the Physical Quantity and of the Unit of Measurement ................. 100
11.2 Toggling between the Channels........................................................................ 101
11.3 Adjustment of the Display ................................................................................. 102
11.4 Status Line ........................................................................................................ 103
11.5 Transducer Distance ......................................................................................... 104
UMG70XV3-4EN 12.01.2009
16 SuperUser-Mode ............................................................................................. 146
16.1 Activating/Deactivating ...................................................................................... 146
16.2 Transducer Parameters .................................................................................... 146
16.3 Malfunctions in SuperUser Mode ...................................................................... 147
D Certificates....................................................................................................... 235
UMG70XV3-4EN 12.01.2009
1 Introduction
1 Introduction
1.1 Regarding this Manual
This manual has been written for the personnel operating the ultrasonic flowmeter FLUX-
US. It contains important information about the instrument, how to handle it correctly,
how to avoid damages. Always keep this manual at hand.
Get acquainted with the safety rules and the handling precautions. Make sure you have
read and understood this manual before using the instrument.
All reasonable effort has been made to ensure the correctness of the content of this man-
ual. If you however find some erroneous information, please inform us. We will be grate-
ful for any suggestions and comments regarding the concept and your experience work-
ing with the instrument.
This will ensure that we can further develop our products for the benefit of our customers
and in the interest of technological progress. If you have any suggestions about improv-
ing the documentation and particularly this User Manual, please let us know so that we
can consider your comments for future reprints.
The content of this manual is subject to changes without prior notice. All rights reserved.
No part of this manual may be reproduced in any form without FLEXIM's written permis-
sion.
Note! The notes contain important information which help you use the flow-
meter optimally.
UMG70XV3-4EN 12.01.2009 9
1 Introduction
1.3 Warranty
The FLUXUS flowmeter is guaranteed for the term and to the conditions specified in the
sales contract provided the equipment has been used for the purpose for which it has
been designed and operated according to the instructions given in this User Manual.
Misuse of the FLUXUS will immediately revoke any warranty given or implied.
This includes:
• replacement of a component of FLUXUS by a component that was not authorized by
FLEXIM
• unsuitable or insufficient maintenance
• repair of FLUXUS by unauthorized personnel
FLEXIM assumes no responsibility for injury to the customer or third persons proximately
caused by the material owing to defects in the product which were not predictable or for
any indirect damages.
FLUXUS is a very reliable instrument. It is manufactured under strict quality control, us-
ing modern production techniques. If installed as recommended in an appropriate loca-
tion, used cautiously and taken care of conscientiously, no troubles should appear.
If any problem appears which can not be solved with the help of this manual (see chapter
19), contact our sales office giving a precise description of the problem. Specify the type,
serial number and firmware version of the flowmeter.
10 UMG70XV3-4EN 12.01.2009
2 Safety Instructions according to Directive 94/9/EC
Cable gland openings not used have to be properly closed with ap-
propriate plugs. They must be removable only with tools.
UMG70XV3-4EN 12.01.2009 11
2 Safety Instructions according to Directive 94/9/EC
Technical Data
For further technical data see annex A, section Technical Data.
FLUXUS G704 A2
manufacturer FLEXIM GmbH
Wolfener Str. 36
12681 Berlin
marking
II3G Ex nA II T4 Ta -20…+60 °C
II3D Ex tD A22 IP66 T100°C
12 UMG70XV3-4EN 12.01.2009
2 Safety Instructions according to Directive 94/9/EC
The junction box is safe in explosive atmosphere only when all con-
nections have been made correctly according to the instructions in
the user manual.
If other cable glands are used than those supplied by FLEXIM, they
have to be approved for the application area and the cable diameters
have to be observed.
Make sure that all cables fit firmly and tightly in the cable glands, the
cable glands are firmly tightened and the housing has been tightly
screwed.
UMG70XV3-4EN 12.01.2009 13
2 Safety Instructions according to Directive 94/9/EC
Technical Data
For further information see annex A, section Technical Data.
Junction Box JB01S#E3#
FLEXIM GmbH
manufacturer Wolfener Str. 36
12681 Berlin
II2G
0044;
II2D
marking
Ex e mb II (T6)...T4 Ta -40...+(70)80 °C
Ex tD A21 IP67 T 100 °C
IBExU06ATEX1161
certification
IECEx IBE08.0006
14 UMG70XV3-4EN 12.01.2009
2 Safety Instructions according to Directive 94/9/EC
The transducers
**M1N31, **P1N31, **Q1N31, **K1N31, **K1N41,
**G1N31, **G1N41, **K1N33, **K1N43, **G1N33,
**G1N43, **H1N33, **H1N43, **M2N41, **P2N41,
**M1N43, **P1N43, **M2E45, **P2E45, **K1N71,
**K1N81, **G1N71, **G1N81, **K1N73, **K1N83,
**G1N73, **G1N83, **H1N73, **H1N83, **M2N81,
**P2N81, **M1N83, **P1N83, **K1LI1, **G1LI1
**M2LI1, **P2LI1, **M2E85, **P2E85, **Q2E85,
**Q2N81, **Q1N83
can be used in explosive atmosphere in zones 1, 2, 21 and 22.
The transducers listed above can be connected to the flowmeter of
the FLUXUS series.
The flowmeter either has to be located outside of the explosive atmo-
sphere or has to be checked and approved separately for explosion
protection.
All cable connections either have to be located outside of the explo-
sive atmosphere or have to be used with equipment checked and ap-
proved for explosion protection.
The transducers may be used in explosive atmosphere only under
the following conditions:
• the transducers are fixed to a pipe or to the surface of a vessel
• the contact surface of every transducer is directed towards the sur-
face of the pipe or of the vessel on which it is mounted
If an equipotential bonding of the transducers is necessary, the
ground terminal of the transducer housing has to be used.
UMG70XV3-4EN 12.01.2009 15
2 Safety Instructions according to Directive 94/9/EC
The transducers
**K1N41, **G1N41, **K1N43, **G1N43, **H1N43, **M2N41,
**P2N41, **M1N43, **P1N43, **M2E45, **P2E45, **K1N81,
**G1N81, **K1N83, **G1N83, **H1N83, **M2N81, **P2N81,
**M1N83, **P1N83, **K1LI1, **G1LI1, **M2LI1, **P2LI1,
**M2E85, **P2E85, **Q2E85, **Q2N81, **Q1N83
must always be used in the delivered transducer shoe to protect
them against mechanical stress.
If an equipotential bonding of the transducers is necessary, the
ground terminal of the transducer housing and/or of the transducer
shoe has to be used.
16 UMG70XV3-4EN 12.01.2009
2 Safety Instructions according to Directive 94/9/EC
surface
explosion protection ambient
temperature
temperature of transducers temperature
(pipe/vessel)
UMG70XV3-4EN 12.01.2009 17
2 Safety Instructions according to Directive 94/9/EC
Technical Data
For further information see annex A, section Technical Data.
certification IBExU98ATEX1012 X
marking 0044;
II2G Ex q II T6…T3 Ta -40…+180 °C
II2D Ex tD A21 IP65 TX
certification IBExU04ATEX1011 X
18 UMG70XV3-4EN 12.01.2009
2 Safety Instructions according to Directive 94/9/EC
marking 0044;
II2G Ex q II T6…T3 Ta -55…+180 °C
II2D Ex tD A21 IP65 TX
certification IBExU04ATEX1011 X
marking 0044;
II2G Ex q II T6…T3 Ta -40…+140 °C
II2D Ex tD A21 IP65 TX
certification IBExU04ATEX1011 X
UMG70XV3-4EN 12.01.2009 19
2 Safety Instructions according to Directive 94/9/EC
marking 0044;
II2G Ex q II T6…T3 Ta -55…+140 °C
II2D Ex tD A21 IP65 TX
certification IBExU04ATEX1011 X
marking 0044;
II2G Ex q II T6…T2 Ta -45…+225 °C
II2D Ex tD A21 IP65 TX
certification IBExU04ATEX1011 X
20 UMG70XV3-4EN 12.01.2009
2 Safety Instructions according to Directive 94/9/EC
marking II2G
0044 II2D
Ex e q II T6...T3
Ex tD A21 IP65 TX
certification IBExU07ATEX1168 X
IECEx IBE08.0007X
II2G
marking 0044; II2D
Ex e q II T6...T3
Ex tD A21 IP65 TX
certification IBExU07ATEX1168 X
IECEx IBE08.0007X
UMG70XV3-4EN 12.01.2009 21
2 Safety Instructions according to Directive 94/9/EC
II2G
marking 0044; II2D
Ex e q II T6...T3
Ex tD A21 IP65 TX
certification IBExU07ATEX1168 X
IECEx IBE08.0007X
certification IBExU07ATEX1168 X
IECEx IBE08.0007X
22 UMG70XV3-4EN 12.01.2009
2 Safety Instructions according to Directive 94/9/EC
marking 0044; II 2G 2D
Ex q II T6...T3
Ex tD A21 IP68 TX
certification IBExU07ATEX1168 X
IECEx IBE08.0007X
certification IBExU07ATEX1168 X
IECEx IBE08.0007X
UMG70XV3-4EN 12.01.2009 23
3 Handling
3 Handling
3.1 First Inspection
The flowmeter has already been tested thoroughly at the factory. At delivery, proceed to
a visual control to make sure that no damage has occurred during transportation.
Check that the specifications of the flowmeter delivered correspond to the specifications
given on the purchase order.
Type and serial number of the flowmeter are given on the data plate. The transducer type
is printed on the transducers.
3.3 Cleaning
• Clean the flowmeter with a soft cloth. Do not use detergents.
• Remove traces of acoustic coupling compound from the transducers with a soft paper
tissue.
24 UMG70XV3-4EN 12.01.2009
4 Flowmeter
4 Flowmeter
4.1 Measuring Principle
The flow of the medium is measured by ultrasonic signals using the transit time differ-
ence method.
Ultrasonic signals are emitted by a transducer installed on one side of a pipe, reflected
on the opposite side and received by a second transducer. The signals are emitted alter-
natively in and against the flow direction.
As the medium in which the signals propagate is flowing, their transit time in flow direc-
tion is shorter than against the flow direction.
The transit time difference t is measured allowing to determine the average flow velocity
on the propagation path of the ultrasonic signals. A flow profile correction is then per-
formed to obtain the area average of the flow velocity, which is proportional to the volume
flow.
The received ultrasonic signals will be tested for usefulness for the measurement and the
plausibility of the measured values will be evaluated. The complete measuring procedure
will be controlled by the integrated microprocessors. Disturbance signals will be eliminat-
ed.
UMG70XV3-4EN 12.01.2009 25
4 Flowmeter
4 Flowmeter
4.2 Description of the Flowmeter
Attention! The degree of protection of the flowmeter is only guaranteed if the cable
glands have been firmly tightened and the cover of the housing has been
tightly screwed.
keyboard
transducers
power supply
26 UMG70XV3-4EN 12.01.2009
4 Flowmeter
2x 16-digit
LCD display
(backlit) serial interface RS232
A : B :
KL1 KL3 KL5 KL7
T 4 B
P 7 a T 4 B S A 1 S B 1
P 7 b T 4 b
P 6 a T 4 A S A 2 S B 2
P T 1 0 0
P 6 b T 4 a T 4 A
P 5 a S 3 S A 3 S B 3
P 5 b S 4
P 7 + T 3 B S A 4 S B 4
P 7 -
outputs P 6 +
P 6 -
T 3 A
T 3 b
T 3 a
P 5 + T 2 B
P 5 - T 2 b
P 4 + T 2 A
P 4 - T 2 a
P 3 + S 1
P 3 - S 2 T 1 B KL6 A V KL8 B V
P 2 + T 1 B
P 2 - T 1 b
P 1 + T 1 A A V S B V S
P T 1 0 0
P 1 - T 1 a T 1 A
A G N B G N
A R S B R S
A R B R
KL2 KL4
L - L O W 4 B -
power supply L +
V O L T
D C
4 A +
L 1 4 3
1 0 0 ...
N 2 3 0 V 4 2
A C
P E 4 1
UMG70XV3-4EN 12.01.2009 27
4 Flowmeter
4 Flowmeter
4.3 Keyboard
The keyboard has three function keys ENTER, BRK and C and ten numerical keys.
Several keys have double functions. They can be used for data entry and for navigation
through scroll lists.
OON
through scroll lists and for input of digits and letters in input mode.
BRK + C + ENTER RESET: Press these three keys simultaneously to recover from an error. The
reset has the same effect as restarting the flowmeter. Stored data will not be
affected.
O- O+
scroll to the left or right through a scroll list
C delete values. After the value has been deleted, the previous value will be
displayed.
28 UMG70XV3-4EN 12.01.2009
4 Flowmeter
DISP
changes the currently selected character to an "A"
OON
selection of the precedent/next ASCII character
OOFF
% X
ter set. The character changes every second. The scrolling will be stopped by
pressing any other key.
BRK + C INIT (cold start): Most parameters and settings are reset to the factory default
values. Stored data will not be affected.
Keep the two keys pressed while switching on the flowmeter until the main
menu is displayed.
A cold start during operation is executed as follows:
• Press the keys BRK, C and ENTER simultaneously.
• Release only key ENTER. A RESET is executed.
• Keep the keys BRK and C pressed until the main menu is displayed.
UMG70XV3-4EN 12.01.2009 29
5 Selection of the Measuring Point
Avoid locations where deposits are building in the pipe. The ambient temperature at the
measuring point must be within the operating temperature range of the transducers.
Select the location of the flowmeter within cable reach of the measuring point. Make sure
that the ambient temperature at the location is within the operating temperature range of
the flowmeter (see annex A, section Technical Data).
If the measuring point is within an explosive atmosphere, the danger zone and gases
which may be present have to be determined. The transducers and the flowmeter have
to be appropriate for these conditions.
30 UMG70XV3-4EN 12.01.2009
5 Selection of the Measuring Point
UMG70XV3-4EN 12.01.2009 31
5 Selection of the Measuring Point
l l
l
l
32 UMG70XV3-4EN 12.01.2009
5 Selection of the Measuring Point
l l
UMG70XV3-4EN 12.01.2009 33
5 Selection of the Measuring Point
The reflected pipe wall signals can disturb the measurement, especially if:
• the measuring point is close to the reflection point
• the pipe wall and measuring signals are received by the transducer at the same time
disadvantageous
lS
34 UMG70XV3-4EN 12.01.2009
5 Selection of the Measuring Point
disadvantageous
ls
n cP
l s = --- ------ D
2 cF
example:
medium: natural gas
pipe material: stainless steel
cP: 3000 m/s
cF: 400 m/s
number of soundpaths: 2
lS = 7.5 D
The range (7.5 ± 2) D is disadvantageous for the transducer installation.
UMG70XV3-4EN 12.01.2009 35
5 Selection of the Measuring Point
36D
(7.5 - 2) D
3D
20D 20D
Fig. 5.3: Measuring point area with developed flow profile and low influence of noise
Considering flow profile and influence of noise, the measuring point can be selected in
the area 3...(7.5 - 2) D on the right side of pipe segment 2 (with max. distance from the el-
bow). In the example, a distance of 36 D from the elbow was selected.
36 UMG70XV3-4EN 12.01.2009
5 Selection of the Measuring Point
Both demands can not be observed always at the same time. In these cases, the mea-
suring point has to be selected so that the influence of noise is min. and the measuring
point is as far as possible from disturbance sources of the flow profile.
16 D
(7.5 - 2) D
3D
20D 5D
Fig. 5.4: Measuring point area with low influence of noise and
not fully developed flow profile
In the example, there is no measuring point area where both demands are met at the
same time. The measuring point has to be selected as far as possible from the elbow, at
a point where the influence of noise can be neglected: 3...(7.5 - 2) D on the right side of
pipe segment 1. In the example, a distance of 16 D from the elbow was selected.
UMG70XV3-4EN 12.01.2009 37
6 Installation of FLUXUS G704
35
Ø 3.5
25 35
Ø6
200
163
M4
265
280
38 UMG70XV3-4EN 12.01.2009
6 Installation of FLUXUS G704
1 2 3
1 shackle
2 pipe mounting plate
3 instrument mounting
plate
4 nut
5 tension strap
UMG70XV3-4EN 12.01.2009 39
6 Installation of FLUXUS G704
It is recommended to run the cables from the measuring point to the flowmeter before the
connection of the transducers to avoid load on the connection points.
Overview
section page
6.4.1 Transducers (ATEX Zone 1) - Direct Connection..............................................40
6.4.2 Transducers (ATEX Zone 1) - Connection via Junction Box .............................42
Connection of the Extension Cable with the Flowmeter ....................................42
Connection of the Extension Cable with the Junction Box ................................43
Connection of the Transducer Cable with the Junction Box..............................45
6.4.3 Transducers (ATEX Zone 2, FM Div. 2 and without Explosion Protection) -
Direct Connection ..............................................................................................46
6.4.4 Transducers (ATEX Zone 2, FM Div. 2 and without Explosion Protection) -
Connection via Junction Box .............................................................................47
Connection of the Extension Cable with the Flowmeter ....................................47
Connection of the Extension Cable with the Junction Box ................................48
Connection of the Transducer Cable with the Junction Box..............................48
Attention! For good high frequency shielding it is important to assure good con-
tact between the cable shield and the cap nut (and thus the housing).
• Connect the leads to the terminals of the flowmeter (see Fig. 6.3 and Table 6.1).
40 UMG70XV3-4EN 12.01.2009
6 Installation of FLUXUS G704
equipotential bonding
transducers transducers terminal
measuring channel A measuring channel B (FLUXUS G704 A2)
basic
part
20 mm
10 mm
70 mm
cable gland
with double hole sealing
compres-
sion part
cap nut
transducer
cable
transducers
ATEX zone 1
UMG70XV3-4EN 12.01.2009 41
6 Installation of FLUXUS G704
Attention! For good high frequency shielding it is important to assure good con-
tact between the cable shield and the cap nut (and thus the housing).
• Connect the leads to the terminals of the flowmeter (see Fig. 6.4 and Table 6.2).
transducers transducers
measuring channel A basic measuring channel B
part
10 mm
20 mm
70 mm
cable gland
compres-
sion part
cap nut
extension cable
42 UMG70XV3-4EN 12.01.2009
6 Installation of FLUXUS G704
Attention! The outer shield of the extension cable must not have electrical con-
tact to the junction box. The extension cable has to remain insulated
completely up to the shield terminal of the junction box.
• Remove the blind plug from the housing (see Fig. 6.5).
• Open the cable gland of the extension cable. The compression part remains in the cap
nut.
• Push the extension cable through cap nut, compression part and basic part.
• Insert the extension cable in the junction box.
• Prepare the extension cable with the cable gland. Cut the outer shield and brush it
back.
• Pull the extension cable back until the brushed back outer shield is below the shield
terminal (see Fig. 6.6).
• Tighten the gasket ring side of the basic part in the junction box.
• Fix the cable gland by screwing the cap nut on the basic part.
• Fix the extension cable and the outer shield to the shield terminal of the junction box
(see Fig. 6.6).
• Connect the leads of the extension cable to the terminals of the junction box (see Fig.
6.6 and Table 6.3).
UMG70XV3-4EN 12.01.2009 43
6 Installation of FLUXUS G704
basic
part
10 mm
20 mm
20 mm
100 mm
100 mm
75 mm
cable gland
12 mm
compres-
sion part
cap nut
extension cable
transducer
cable
transducers
ATEX zone 1
44 UMG70XV3-4EN 12.01.2009
6 Installation of FLUXUS G704
TV TVS TG TG TRS TR V VS RS R
shield terminal
equipotential
bonding terminal
Fig. 6.6: Connection of extension cable and transducer cable (ATEX zone 1)
UMG70XV3-4EN 12.01.2009 45
6 Installation of FLUXUS G704
SMB connector
transducers:
• ATEX zone 2
• FM Div. 2
• without explosion protection
46 UMG70XV3-4EN 12.01.2009
6 Installation of FLUXUS G704
Attention! For good high frequency shielding it is important to assure good con-
tact between the cable shield and the cap nut (and thus the housing).
• Connect the leads to the terminals of the flowmeter (see Fig. 6.8 and Table 6.6).
transducers transducers
measuring basic measuring
channel A part channel B
10 mm
20 mm
70 mm
cable gland
compres-
sion part
cap nut
extension cable
UMG70XV3-4EN 12.01.2009 47
6 Installation of FLUXUS G704
Attention! The outer shield of the extension cable must not have electrical con-
tact to the junction box. The extension cable has to remain insulated
completely up to the shield terminal.
• Remove the blind plug from the housing (see Fig. 6.9).
• Open the cable gland of the extension cable. The compression part remains in the cap
nut.
• Push the extension cable through cap nut, compression part and basic part.
• Insert the extension cable in the junction box.
• Prepare the extension cable with the cable gland. Cut the outer shield and brush it
back.
• Pull the extension cable back until the brushed back outer shield is below the shield
terminal (see Fig. 6.9).
• Tighten the gasket ring side of the basic part in the junction box.
• Fix the cable gland by screwing the cap nut on the basic part.
• Fix the extension cable and the outer shield to the shield terminal of the junction box
(see Fig. 6.9).
• Connect the leads to the terminals of the junction box (see Fig. 6.9 and Table 6.7).
Table 6.7: Terminal assignment
terminal connection
TV white or marked cable (core)
TVS white or marked cable (inner shield)
TRS brown cable (inner shield)
TR brown cable (core)
48 UMG70XV3-4EN 12.01.2009
6 Installation of FLUXUS G704
TV TVS TG TG TRS TR
XV XR
basic
part
20 mm
100 mm
75 mm
cable gland
12 mm
compres-
sion part
cap nut
extension cable
transducer
cable
UMG70XV3-4EN 12.01.2009 49
6 Installation of FLUXUS G704
AMP-Quick and
SMB connectors
cable gland
transducers:
• ATEX zone 2
• FM Div. 2
• without explosion protection
50 UMG70XV3-4EN 12.01.2009
6 Installation of FLUXUS G704
• Prepare the power cable. Connect the leads with the terminals of the power supply
(see Fig. 6.11 and Table 6.10).
power supply
(KL3)
KL3
UMG70XV3-4EN 12.01.2009 51
6 Installation of FLUXUS G704
outputs
(KL4)
+ (black)
-
Px-
Px-
52 UMG70XV3-4EN 12.01.2009
6 Installation of FLUXUS G704
UH
V
+
-
(black) -
UH
Px-
V Ic = 1...4 mA
(special
+
(black) -
design) Px- -
UH
binary output Px+ UMAX = 24 V
a
(Reed relay) Pxa IMAX = 150 mA
(special
design) b Pxb
Px-
RLOAD is the sum of all ohmic resistances in the circuit (resistance of the conductors, inner resis-
tance of the external power supply, resistance of the amperemeter/voltmeter, etc.)
UMG70XV3-4EN 12.01.2009 53
6 Installation of FLUXUS G704
TxA
Txb
- +
Fig. 6.13: Connection of an external passive current source
Attention! Observe the polarity to avoid damage of the current source. The cur-
rent input could be destroyed by a permanent shorting of both termi-
nals.
54 UMG70XV3-4EN 12.01.2009
6 Installation of FLUXUS G704
If the flowmeter has a passive current input, an external power supply is necessary. The
power supply UH must be able to
• supply sufficient power for the energy requirements of the passive current source
• cover the voltage drop over the measurement resistance (max. 1 V) and
• cover all other voltage drops (e.g. cable resistance) in the circuit.
A current of min. 20 mA has to be provided.
= 11 V + 20 mA · 50 + 20 mA · 2
= 12.04 V
The terminals TxA and TxB are on the upper terminal row (x =1...4)
If the polarity of the current source is inversed, only the sign of the measured current will
change.
TxA TxB
+ -
Fig. 6.14: Connection of an active current source
UMG70XV3-4EN 12.01.2009 55
6 Installation of FLUXUS G704
Note! The serial number of sensor module and transducer have to be iden-
tical. A wrong or incorrectly connected sensor module will result in in-
correct measured values or in measurement failure.
56 UMG70XV3-4EN 12.01.2009
6 Installation of FLUXUS G704
sensor module
(KL1)
UMG70XV3-4EN 12.01.2009 57
7 Installation of FLUXUS G709
It is recommended to run the cables from the measuring point to the flowmeter before the
connection of the transducers to avoid load on the connection points.
Overview
section page
7.2.1 Transducers (ATEX Zone 1) - Connection via Junction Box .............................59
Connection of the Extension Cable with the Flowmeter ....................................59
Connection of the Extension Cable with the Junction Box ................................60
Connection of the Transducer Cable with the Junction Box..............................62
7.2.2 Transducers (ATEX Zone 2, FM Div. 2 and without Explosion Protection -)
Connection via Junction Box .............................................................................63
Connection of the Extension Cable with the Flowmeter ....................................63
Connection of the Extension Cable with the Junction Box ................................64
Connection of the Transducer Cable with the Junction Box..............................64
58 UMG70XV3-4EN 12.01.2009
7 Installation of FLUXUS G709
A : B :
KL1 KL3 KL5 KL7
T 4 B
P 7 a T 4 B S A 1 S B 1
P 7 b T 4 b
P 6 a T 4 A S A 2 S B 2
P T 1 0 0
P 6 b T 4 a T 4 A
P 5 a S 3 S A 3 S B 3
P 5 b S 4
P 7 + T 3 B S A 4 S B 4
P 7 - T 3 b
P 6 + T 3 A
P 6 - T 3 a
P 5 + T 2 B
P 5 - T 2 b
P 4 + T 2 A
P 4 - T 2 a
P 3 + S 1
P 3 - S 2 T 1 B KL6 A V KL8 B V
P 2 + T 1 B
P 2 - T 1 b
P 1 + T 1 A A V S B V S
P T 1 0 0
P 1 - T 1 a T 1 A
A G N B G N
A R S B R S
A R B R
KL2 KL4
L - L O W 4 B -
V O L T
L + D C
4 A +
L 1 4 3
1 0 0 ...
N 2 3 0 V 4 2
A C
P E 4 1
20 mm
shield terminal
60 mm
50 mm
extension cable
UMG70XV3-4EN 12.01.2009 59
7 Installation of FLUXUS G709
Attention! The outer shield of the extension cable must not have electrical con-
tact to the junction box. The extension cable has to remain insulated
completely up to the shield terminal of the junction box.
• Remove the blind plug from the housing (see Fig. 7.2).
• Open the cable gland of the extension cable. The compression part remains in the cap
nut.
• Push the extension cable through cap nut, compression part and basic part.
• Insert the extension cable in the junction box.
• Prepare the extension cable with the cable gland. Cut the outer shield and brush it
back.
• Pull the extension cable back until the brushed back outer shield is below the shield
terminal (see Fig. 7.3).
• Tighten the gasket ring side of the basic part in the junction box.
• Fix the cable gland by screwing the cap nut on the basic part.
• Fix the extension cable and the outer shield to the shield terminal of the junction box
(see Fig. 7.3).
• Connect the leads to the terminals of the junction box (see Fig. 7.3 and Table 7.2).
60 UMG70XV3-4EN 12.01.2009
7 Installation of FLUXUS G709
basic
part
10 mm
20 mm
20 mm
100 mm
100 mm
75 mm
cable gland
12 mm
compres-
sion part
cap nut
extension cable
transducer
cable
transducers
ATEX zone 1
UMG70XV3-4EN 12.01.2009 61
7 Installation of FLUXUS G709
TV TVS TG TG TRS TR V VS RS R
shield terminal
equipotential
bonding terminal
Fig. 7.3: Connection of extension cable and transducer cable (ATEX zone 1)
62 UMG70XV3-4EN 12.01.2009
7 Installation of FLUXUS G709
A : B :
KL1 KL3 KL5 KL7
T 4 B
P 7 a T 4 B S A 1 S B 1
P 7 b T 4 b
P 6 a T 4 A S A 2 S B 2
P T 1 0 0
P 6 b T 4 a T 4 A
P 5 a S 3 S A 3 S B 3
P 5 b S 4
P 7 + T 3 B S A 4 S B 4
P 7 - T 3 b
P 6 + T 3 A
P 6 - T 3 a
P 5 + T 2 B
P 5 - T 2 b
P 4 + T 2 A
P 4 - T 2 a
P 3 + S 1
P 3 - S 2 T 1 B KL6 A V KL8 B V
P 2 + T 1 B
P 2 - T 1 b
P 1 + T 1 A A V S B V S
P T 1 0 0
P 1 - T 1 a T 1 A
A G N B G N
A R S B R S
A R B R
KL2 KL4
L - L O W 4 B -
V O L T
L + D C
4 A +
L 1 4 3
1 0 0 ...
N 2 3 0 V 4 2
A C
P E 4 1
20 mm
shield terminal
60 mm
50 mm
extension cable
UMG70XV3-4EN 12.01.2009 63
7 Installation of FLUXUS G709
Attention! The outer shield of the extension cable must not have electrical con-
tact to the junction box. The extension cable has to remain insulated
completely up to the shield terminal.
• Remove the blind plug from the housing (see Fig. 7.5).
• Open the cable gland of the extension cable. The compression part remains in the cap
nut.
• Push the extension cable through cap nut, compression part and basic part.
• Insert the extension cable in the junction box.
• Prepare the extension cable with the cable gland. Cut the outer shield and brush it
back.
• Pull the extension cable back until the brushed back outer shield is below the shield
terminal (see Fig. 7.5).
• Tighten the gasket ring side of the basic part in the junction box.
• Fix the cable gland by screwing the cap nut on the basic part.
• Fix the extension cable and the outer shield to the shield terminal of the junction box
(see Fig. 7.5).
• Connect the leads to the terminals of the junction box (see Fig. 7.5 and Table 7.5).
64 UMG70XV3-4EN 12.01.2009
7 Installation of FLUXUS G709
TV TVS TG TG TRS TR
XV XR
basic
part
20 mm
100 mm
75 mm
cable gland
12 mm
compres-
sion part
cap nut
extension cable
transducer
cable
UMG70XV3-4EN 12.01.2009 65
7 Installation of FLUXUS G709
Note! Cap nut, compression part and basic part of the cable gland remain on
the cable.
• Push the extension cable through the shield terminal to terminal strip KL6 for measur-
ing channel A and to terminal strip KL8 for measuring channel B (see Fig. 7.6).
• Pull the extension cable back until the brushed back outer shield is below the shield
terminal (see Fig. 7.6).
• Fix the extension cable and the outer shield to the shield terminal.
• Connect the AMP-Quick and the SMB connectors with the sockets of the flowmeter (see
Fig. 7.6 and Table 7.7).
A : B :
T 4 B
P 7 a T 4 B S A 1 S B 1
P 7 b T 4 b
P 6 a T 4 A S A 2 S B 2
P T 1 0 0
P 6 b T 4 a T 4 A
P 5 a S 3 S A 3 S B 3
P 5 b S 4
P 7 + T 3 B S A 4 S B 4
P 7 - T 3 b
P 6 + T 3 A
P 6 - T 3 a
P 5 + T 2 B
P 5 - T 2 b
P 4 + T 2 A
P 4 - T 2 a
P 3 + S 1
P 3 - S 2 T 1 B
B V
P 2 + T 1 B A V
P 2 - T 1 b
P 1 + T 1 A A V S B V S
P T 1 0 0
P 1 - T 1 a T 1 A
A G N B G N
A R S B R S
KL6 A R B R KL8
L - L O W 4 B -
V O L T
L + D C
4 A +
L 1 4 3
1 0 0 ...
N 2 3 0 V 4 2
A C
P E 4 1
AMP-Quick and
SMB connectors
shield terminal
basic
part
outer shield
cable gland
compres-
sion part
cap nut
transducers:
• ATEX zone 2
• FM Div. 2
• without explosion protection
66 UMG70XV3-4EN 12.01.2009
7 Installation of FLUXUS G709
• Connect the leads with the terminals of terminal strip KL2 (see Fig. 7.7 and Table 7.8).
A : B :
KL1 KL3 KL5 KL7
T 4 B
P 7 a T 4 B S A 1 S B 1
P 7 b T 4 b
P 6 a T 4 A S A 2 S B 2
P T 1 0 0
P 6 b T 4 a T 4 A
P 5 a S 3 S A 3 S B 3
P 5 b S 4
P 7 + T 3 B S A 4 S B 4
P 7 - T 3 b
P 6 + T 3 A
P 6 - T 3 a
P 5 + T 2 B
P 5 - T 2 b
P 4 + T 2 A
P 4 - T 2 a
P 3 + S 1
P 3 - S 2 T 1 B KL6 A V KL8 B V
P 2 + T 1 B
P 2 - T 1 b
P 1 + T 1 A A V S B V S
P T 1 0 0
P 1 - T 1 a T 1 A
A G N B G N
A R S B R S
A R B R
KL2 KL4
L - L O W 4 B -
V O L T
L + D C
4 A +
L 1 4 3
1 0 0 ...
N 2 3 0 V 4 2
A C
P E 4 1
power supply
(KL2)
UMG70XV3-4EN 12.01.2009 67
7 Installation of FLUXUS G709
A : B :
KL1 KL3 KL5 KL7
T 4 B
P 7 a T 4 B S A 1 S B 1
P 7 b T 4 b
P 6 a T 4 A S A 2 S B 2
P T 1 0 0
P 6 b T 4 a T 4 A
P 5 a S 3 S A 3 S B 3
P 5 b S 4
P 7 + T 3 B S A 4 S B 4
P 7 - T 3 b
P 6 + T 3 A
P 6 - T 3 a
P 5 + T 2 B
P 5 - T 2 b
P 4 + T 2 A
P 4 - T 2 a
P 3 + S 1
P 3 - S 2 T 1 B KL6 A V KL8 B V
P 2 + T 1 B
P 2 - T 1 b
T 1 A A V S B V S
outputs P 1 +
P T 1 0 0
P 1 - T 1 a T 1 A
A G N B G N
(KL1) A R S B R S
A R B R
KL2 KL4
L - L O W 4 B -
V O L T
L + D C
4 A +
L 1 4 3
1 0 0 ...
N 2 3 0 V 4 2
A C
P E 4 1
+ (black)
-
Px-
Px-
68 UMG70XV3-4EN 12.01.2009
7 Installation of FLUXUS G709
UH
V Ic = 1...4 mA
+
(black) -
Px- -
UH
b Pxb
Px-
RLOAD is the sum of all ohmic resistances in the circuit (resistance of the conductors, inner resis-
tance of the external power supply, resistance of the amperemeter/voltmeter, etc.)
UMG70XV3-4EN 12.01.2009 69
7 Installation of FLUXUS G709
TxA
Txb
- +
Fig. 7.9: Connection of an external passive current source
Attention! Observe the polarity to avoid damage of the current source. The cur-
rent input could be destroyed by a permanent shorting of both termi-
nals.
70 UMG70XV3-4EN 12.01.2009
7 Installation of FLUXUS G709
If the flowmeter has a passive current input, an external power supply is necessary. The
power supply UH must be able to
• supply sufficient power for the energy requirements of the passive current source
• cover the voltage drop over the measurement resistance (max. 1 V) and
• cover all other voltage drops (e.g. cable resistance) in the circuit.
A current of min. 20 mA has to be provided.
= 11 V + 20 mA · 50 + 20 mA · 2
= 12.04 V
The terminals TxA and TxB are on the upper terminal row (x =1...4)
If the polarity of the current source is inversed, only the sign of the measured current will
change.
TxA TxB
+ -
Fig. 7.10: Connection of an active current source
UMG70XV3-4EN 12.01.2009 71
7 Installation of FLUXUS G709
A : B :
KL1 KL3 KL5 KL7
T 4 B
P 7 a T 4 B S A 1 S B 1
P 7 b T 4 b
P 6 a T 4 A S A 2 S B 2
P T 1 0 0
P 6 b T 4 a T 4 A
P 5 a S 3 S A 3 S B 3
P 5 b S 4
P 7 + T 3 B S A 4 S B 4
P 7 - T 3 b
P 6 + T 3 A
P 6 - T 3 a
P 5 + T 2 B
P 5 - T 2 b
P 4 + T 2 A
P 4 - T 2 a
P 3 + S 1
P 3 - S 2 KL6 KL8
serial interface
T 1 B
T 1 B A V B V
P 2 +
P 2 - T 1 b
P 1 + T 1 A A V S B V S
P T 1 0 0
RS485 (option) P 1 - T 1 A
T 1 a
A G N B G N
A R S B R S
(KL4) A R B R
KL2 KL4
L - L O W 4 B -
V O L T
L + D C
4 A +
L 1 4 3
1 0 0 ...
N 2 3 0 V 4 2
A C
P E 4 1
72 UMG70XV3-4EN 12.01.2009
7 Installation of FLUXUS G709
Note! The serial number of sensor module and transducer have to be iden-
tical. A wrong or incorrectly connected sensor module will result in in-
correct measured values or in measurement failure.
Attention! The sensor module and the connector must not be connected at the
same time.
• If the cable with the 4-pole connector is cut, it can be connected directly to the terminal
strip KL5 or KL7 (see Fig. 7.12 and Table 7.13).
Table 7.13: Terminals for the connection of the sensor module
terminal connection
SA1 gray
SA2
SA3 brown
SA4 green
A : B :
KL1 KL3
T 4 B
KL5 KL7 sensor module
P 7 a
P 6 a
P 7 b
T 4 B
T 4 A
T 4 b
S A 1
S A 2
S B 1
S B 2
measuring
P T 1 0 0
P 6 b T 4 a T 4 A
P 5 a
P 5 b
S 3
S 4
S A 3 S B 3
S B 4
channel B
P 7 + T 3 B S A 4
P 6 +
P 7 -
T 3 A
T 3 b
(KL7)
P 5 +
P 6 -
T 2 B
T 3 a X5 X7
P 5 - T 2 b
P 4 + T 2 A
P 4 - T 2 a
P 3 + S 1
P 2 +
P 3 -
T 1 B
S 2 T 1 B
KL6 A V KL8 B V
P 2 - T 1 b B V S
P 1 + T 1 A A V S
sensor module
P T 1 0 0
P 1 - T 1 a T 1 A
A G N B G N
A R S
A R
B R S
B R
measuring
KL2 KL4 channel A
L -
L +
L O W
V O L T
4 B -
4 A +
(KL5)
D C
L 1 4 3
1 0 0 ...
N 2 3 0 V 4 2
A C
P E 4 1
UMG70XV3-4EN 12.01.2009 73
8 Installation of the Temperature Probe (Option)
a) b)
- Apply a film of thermal conductivity paste (no FLEXIM supply) onto the contact sur-
face of the temperature probe.
- Take the spring end of the ball chain and insert the last ball in one of the slots on the
top of the temperature probe.
- Place the chain around the pipe. Pull the chain firmly and insert it in the second slot
of the temperature probe (see Fig. 8.2).
74 UMG70XV3-4EN 12.01.2009
8 Installation of the Temperature Probe (Option)
UMG70XV3-4EN 12.01.2009 75
9 Start-up
9 Start-up
9.1 Switching on
As soon as the flowmeter is connected with the power sup-
FLEXIM FLUXUS ply it will be displayed which transducer has been detected
G70X-XXXXXXXX at which channel.
Afterwards the serial number of the flowmeter will be dis-
played for a short time.
Data can not be entered while the serial number is dis-
played.
After initialization, the main menu is displayed in the se-
>PAR< mea opt sf
lected language. The language of the display can be set
Parameter (see section 9.4).
9.2 Displays
9.2.1 Main Menu
The main menu contains the program branches:
>PAR< mea opt sf
Parameter • PAR (parameters)
• MEA (measuring)
• OPT (output options)
• SF (special functions)
The selected program branch is displayed in capital letters between arrows. The com-
plete name of the selected program branch is displayed in the lower line.
Select a program branch with key O-
and O+
. Press ENTER.
Note! By pressing key BRK, the measurement will be stopped and the main
menu selected.
Note! In this manual, all program entries and keys are indicated in capital
letters. Program entries are indicated in typewriter characters (PA-
RAMETER). The menu items will be separated from the main menu by
a backslash "\".
76 UMG70XV3-4EN 12.01.2009
9 Start-up
1 install medium
The following menu items are in SYSTEM SETTINGS:
• dialogs/menus
• inputs
• measurement
• gas measurement
• outputs
• storing
• serial output
• miscellaneous
• set clock
• libraries
UMG70XV3-4EN 12.01.2009 77
9 Start-up
9.2.4 Navigation
If a vertical arrow is displayed, the menu item contains a scroll list. The current list
item is displayed in the lower line.
OON
Parameter Scroll with key and OOFF to select a list item in the low-
for Channel A: er line. Press ENTER.
In some menu items, there is a horizontal scroll list in the lower line. The selected list
item is displayed in capital letters between arrows.
In some menu items, there is a horizontal scroll list in the upper line. The selected list
item is displayed in capital letters between arrows. The current value of the list item is
displayed in the lower line.
9.3 HotCodes
A HotCode is a key sequence activating some settings:
• language selection (see section 9.4)
• switching on the SuperUser mode (see chapter 16)
• switching on the FastFood mode (see section 12.6)
• manual input of the lower limit for the inner pipe diameter (see section 12.8)
A HotCode can be entered in the main menu after pressing key C.
78 UMG70XV3-4EN 12.01.2009
9 Start-up
Depending on the technical data of the flowmeter, some of the languages might not be
implemented.
When the last digit is entered, the main menu will be displayed in the selected language.
The selected language remains activated after switching the flowmeter off and on again.
Hinweis! After initialization of the flowmeter the default language will be set
(key BRK and C when switching on).
The stopped measurement will be continued. All selected output options are still active.
The measurement is not continued after return of the power supply if a cold start was
performed.
UMG70XV3-4EN 12.01.2009 79
10 Basic Measurement
10 Basic Measurement
The pipe and medium parameters will be entered for the selected measuring point (see
chapter 5). The ranges are limited by the properties of the transducers and of the flowme-
ter.
Note! The parameters will be stored only if the program branch PARAME-
TER is finished completely once.
It is possible to enter the pipe circumference instead of the outer pipe diameter (see sec-
tion 15.2.1).
If the input of the pipe circumference has been activated and 0 (zero) is entered for the
OUTER DIAMETER, the menu item PIPE CIRCUMFER. will be displayed. If the pipe cir-
cumference is not to be entered, press key BRK to return to the main menu and start the
parameter input again.
80 UMG70XV3-4EN 12.01.2009
10 Basic Measurement
Note! The inner pipe diameter (= outer pipe diameter - 2x pipe wall thick-
ness) will be calculated internally. If the value is not within the inner
pipe diameter range of the connected transducers, an error message
will be displayed.
It is possible to change the lower limit of the inner pipe diameter for a
given transducer type (see section 12.8).
The materials to be displayed in the scroll list can be selected (see chapter 14.5).
When the pipe material is selected, the corresponding sound velocity is set automatical-
ly. If OTHER MATERIAL is selected, the sound velocity has to be entered.
Enter the sound velocity of the pipe material. Press EN-
c-Material
TER.
3230.0 m/s
Note! Enter the sound velocity of the material (i.e. longitudinal or transver-
sal velocity) which is nearer to 2500 m/s.
For the sound velocity of some materials see annex C, Table C.1.
The materials to be displayed in the scroll list can be selected (see chapter 14.5).
UMG70XV3-4EN 12.01.2009 81
10 Basic Measurement
For the sound velocity of some materials see annex C, Table C.1.
Enter the thickness of the liner. Press ENTER.
Liner Thickness
3.0 mm
Note! The inner pipe diameter (= outer pipe diameter - 2x pipe wall thick-
ness - 2x liner thickness) will be calculated internally. If the value is
not within the inner pipe diameter range of the connected transduc-
ers, an error message will be displayed.
It is possible to change the lower limit of the inner pipe diameter for a
given transducer type (see section 12.8).
The media to be displayed in the scroll list can be selected (see chapter 14.5).
For the programmed parameters of common media see annex C, Table C.3.
If a medium is selected from the scroll list, the menu item for the input of the medium
temperature is displayed directly (see section 10.2.5).
If OTHER MEDIUM is selected or no data set for the selected medium is stored in the flow-
meter (e.g. natural gas, as the parameters depend on the composition of the natural
gas), the medium parameters have to be entered first:
82 UMG70XV3-4EN 12.01.2009
10 Basic Measurement
10.2.3 Density
The mass flow will be calculated on the base of the density (product of volume flow and
density). This display is indicated only if OTHER MEDIUM has been selected or no data
set for the selected medium is stored in the flowmeter (e.g. natural gas).
Enter the operating density of the medium. Press EN-
Density
TER.
60.00 kg/m3
UMG70XV3-4EN 12.01.2009 83
10 Basic Measurement
Note! If standard transducer parameters are used, FLEXIM can not guaran-
tee for the precision of the measured values. A measurement might
be even impossible.
84 UMG70XV3-4EN 12.01.2009
10 Basic Measurement
Note! A channel can not be activated if the parameters are not valid, e.g. if
the parameters in the program branch PARAMETER of the channel
are not complete.
UMG70XV3-4EN 12.01.2009 85
10 Basic Measurement
reflection mode, number of sound paths = 2 diagonal mode, number of sound paths = 3
A A
!
A < 0
Note! Exact positioning of the transducers is easier for an even number (re-
flection mode) of transit paths than for an odd number (diagonal
mode).
86 UMG70XV3-4EN 12.01.2009
10 Basic Measurement
Attention! Wear protective gloves and protective goggles when mounting the
damping mats.
In case of contact with the eyes:
• rinse immediately with plenty of water and
• see an eye specialist.
In case of contact with the skin:
• wash immediately with plenty of water.
• Fix the part of the damping mat where the protection sheet is removed to the pipe.
protection sheet
damping mat
UMG70XV3-4EN 12.01.2009 87
10 Basic Measurement
• Fix the damping mat to the pipe: remove the protection sheet bit by bit from the damp-
ing mat and attach the damping mat to the pipe at the same time.
transducer
damping mat
88 UMG70XV3-4EN 12.01.2009
10 Basic Measurement
Note! The damping mat adheres strongly to the pipe. Later alignment is not
possible. The damping mat has to be positioned correctly immediate-
ly.
• Depending on the transducer type, another layer of the damping mat is necessary (see
annex A, Technical Data). Repeat the steps for the installation of the damping mat.
• Mount the transducers onto the damping mat. Select the installation instruction corre-
sponding to the supplied transducer mounting fixture (see section 10.7).
• If the measuring point is close to a reflection point (e.g. flange), a damping mat has to
be installed there, if necessary (see Fig. 10.5). Repeat the steps for the installation of
the damping mat.
reflection point
UMG70XV3-4EN 12.01.2009 89
10 Basic Measurement
• Cut the damping mat to size, if necessary (see annex A, section Technical Data).
• Apply the a bead of ultrasound gel along the center line onto the inner side of the
damping mat.
• Spread the ultrasound gel evenly on the surface of the damping mat by means of a
spattle (see Fig. 10.6).
Note! When measuring in diagonal mode, a damping mat will be fixed addi-
tionally to the opposite side of the pipe (see Fig. 10.4).
• Mount the transducers onto the damping mat. Select the installation instruction corre-
sponding to the supplied transducer mounting fixture (see section 10.7).
• If the measuring point is close to a reflection point (e.g. flange), a damping mat has to
be installed there, if necessary (see Fig. 10.5).
• Repeat the steps for the installation of the damping mat.
90 UMG70XV3-4EN 12.01.2009
10 Basic Measurement
UMG70XV3-4EN 12.01.2009 91
10 Basic Measurement
Straps
Note! The clasp must be in contact completely with the pipe to ensure a
good fixation.
• Place the mounting shoe on the pipe. Hold clasp and mounting shoe with one hand
while pushing the tension strap through parts (2) and (1) of the clasp (see Fig. 10.8).
• Pull the tension strap firmly and engage it in the inner hooks of the clasp.
Note! At pipes with large diameters, use tongs, if necessary, to tense the
tension strap.
92 UMG70XV3-4EN 12.01.2009
10 Basic Measurement
• Mount the second transducer in the same way. Adjust the displayed transducer dis-
tance by means of the measuring tape.
• Tighten the screws of the clasps.
• Push the transducers in the mounting shoes. Press the transducer firmly on the pipe.
There should be no air pockets between transducer surface and pipe wall. Tighten the
screw of the mounting shoe firmly.
Note! When the transducers are mounted on a vertical pipe and the flow-
meter is placed lower than the pipe, the cable of the upper transducer
should be fixed to the tension strap by a cable tie to protect the cable
from mechanical strain.
UMG70XV3-4EN 12.01.2009 93
10 Basic Measurement
3
2
• Push the tension strap through the metal spring (3, see Fig. 10.11).
• Bend the end of the tension strap back. Push the other end of the tension strap through
the groove on the top of the mounting shoe.
Place the clasp on the side of the pipe. Turn the tension strap around the pipe (see Fig.
10.12)
Fig. 10.12: Mounted transducer in mounting shoe, fixed with FLEXIM clasp
94 UMG70XV3-4EN 12.01.2009
10 Basic Measurement
Note! The clasp and the metal spring must be in contact completely with
the pipe to ensure a good fixation.
• Place the mounting shoe on the pipe. Hold clasp and mounting shoe with one hand
while pushing the tension strap through parts (2) and (1) of the clasp (see Fig. 10.11).
• Pull the tension strap firmly and engage it in the inner hooks of the clasp.
Note! At pipes with large diameters, use tongs, if necessary, to tense the
tension strap.
• Mount the second transducer in the same way. Adjust the displayed transducer dis-
tance by means of the measuring tape.
• Tighten the screws of the clasps.
• Push the transducers in the mounting shoes. Press the transducer firmly on the pipe.
There should be no air pockets between transducer surface and pipe wall. Tighten the
screw of the mounting shoe firmly.
Note! When the transducers are mounted on a vertical pipe and the flow-
meter is placed lower than the pipe, the cable of the upper transducer
should be fixed to the tension strap by a cable tie to protect the cable
from mechanical strain.
UMG70XV3-4EN 12.01.2009 95
10 Basic Measurement
10 Basic
10.7.3 Fine Measurement
Adjustment of the Transducer Distance
If the displayed transducer distance is adjusted, press EN-
Transd. Distance
TER.
A: 54 mm !
The positioning procedure is started.
A bar graph S= displays the amplitude of the received sig-
S=
nal.
A:<>=54 mm!
If the LED of the measuring channel lights green, the sig-
nal is sufficient for a measurement (only FLUXUS G709).
If the LED of the measuring channel lights red, the signal is
not sufficient for a measurement (only FLUXUS G709).
Shift a transducer slightly in the range of the recommend-
ed transducer distance
• FLUXUS G704: until the bar graph has max. length
(max. 6 squares)
• FLUXUS G709: until the LED of the measuring channel
lights green.
The following can be displayed in the upper line with key
S= DISP
Q=
• transducer distance
trans. 94.0 s • C (signal noise ratio)
Q= If min. one box is displayed, the signal is sufficient for
the measurement. Optimum for the measurement are
three or more boxes.
• bar graph Q= (signal quality), must have max. length
• transit time TRANS. in µs
• bar graph S= (signal amplitude)
If the signal is not sufficient for measurement, Q= UNDEF
will be displayed.
In case of larger deviations, check if the entered parameters are correct or repeat the
measurement at a different location on the pipe.
After the precise positioning of the transducers, the recom-
Transd.Distance?
mended transducer distance is displayed again.
53.9 mm
Enter the actual (precise) transducer distance. Press EN-
TER.
Repeat the steps for all channels on which will be measured. The measurement will be
started automatically afterwards.
96 UMG70XV3-4EN 12.01.2009
10 Basic Measurement
The transducer distance can be displayed during measurement by scrolling with key .
The optimum transducer distance (here: 50.0 mm) will be
L=(50.0) 54.0 mm
displayed in parentheses in the upper line, followed by the
54.5 m3/h entered transducer distance (here: 54.0 mm). The latter
value must correspond to the actually adjusted transducer
distance. Press ENTER to optimize the transducer dis-
tance.
The optimum transducer distance is calculated on the basis of the measured sound
speed. It is therefore a better approximation than the first value which had been calculat-
ed on the basis of the approximate sound speed range entered in the program branch
PARAMETER.
If the difference of optimum and entered transducer distance is less than specified in Ta-
ble 10.1 or Table 10.2, the measurement is consistent and the measured values are val-
id. The measurement can be continued.
If the difference is greater, adjust the transducer distance to the displayed optimum val-
ue. Afterwards, check the signal quality and the signal amplitude bar graph (see section
10.7.3). Press ENTER.
Table 10.1: Standard values for signal optimization of shear wave transducers
transducer frequency difference between optimum and entered transducer distance [mm]
P 8
M 10
K 15
G 20
Table 10.2: Standard values for signal optimization of Lamb wave transducers
transducer frequency difference between optimum and entered transducer distance [mm]
Q -3...+5
P - 6...+10
M -10...+20
K -25...+40
H -35...+60
G -50...+100
UMG70XV3-4EN 12.01.2009 97
10 Basic Measurement
L=(51.1) 50.0 mm
displayed and check the difference between optimum and
54.5 m3/h entered transducer distance. Repeat the steps if neces-
sary.
If an approximate range for the sound speed has been entered in the program branch
PARAMETER and the transducer distance has been optimized afterwards as described in
section 10.7.4, it is recommended to note the sound velocity for the next measurement.
The optimization procedure does not need to be repeated then.
Also note the medium temperature as the sound speed depends on the temperature. The
value can be entered in the program branch PARAMETER or a user defined medium can
be defined for this sound velocity (see section 14.2 and 14.3).
If the standard volume flow is selected as physical quantity during the gas measurement,
the operating volume flow can be displayed, too.
Press key DISP to display the standard volume flow.
A:Volume(norm.)
31.82 m3/h The character * indicates that the displayed value (here:
operational volume flow) is not the selected physical quan-
tity (here: standard volume flow).
A:Volume(oper.)
* 16.22 m3/h
If more than one measuring channel is available/activated, the flowmeter works with an
integrated measuring point multiplexer providing quasi simultaneous measurement on
the different measuring channels.
The flow is measured on one measuring channel for approx. 1 s, then the multiplexer
switches to the next activated channel.
98 UMG70XV3-4EN 12.01.2009
10 Basic Measurement
The measuring time depends on the measuring conditions. If, e.g. the measuring signal
can not be detected immediately, the measurement might be > 1 s.
The outputs and the serial interface continuously get the measuring results of the as-
signed channel. The results will be displayed according to the actually selected output
options (see chapter 11). The default unit of measurement of the volume flow is m3/h. For
the selection of the values to be displayed and for setting of the output options see chap-
ter 11. For further measuring functions see chapter 12.
UMG70XV3-4EN 12.01.2009 99
11 Displaying the Measured Values
Return to the main menu by pressing key BRK. The further menu items of the program
branch OUTPUT OPTIONS are for the activation of the measured value output.
The AutoMux mode is the default display mode. It will be activated automatically after a
cold start.
All Channels
The measured values of all channels (measuring and calculation channels) will be dis-
played. The next active channel is selected after min. 1.5 s.
Note! This mode can only be activated if min. two calculation channels are
activated.
N E X T
Press key % to display the next activated channel. The measured values of the selected
channel will be displayed as configured in the program branch OUTPUT OPTIONS (see
section 11.1).
The display in the upper line can be changed during measurement with key . The dis-
play in the lower line can be changed during measurement with key . DISP
value explanation
S signal amplitude
0 <5%
… …
9 90 %
Note:
values 3 are sufficient for measurement.
Q signal quality
0 <5%
… …
9 90 %
c sound velocity
comparison of the measured and the expected sound velocity of the medium.
The expected sound velocity is calculated on the basis of the medium parame-
ters (medium selected in program branch PARAMETER, temperature
dependency, pressure dependency).
ok, corresponds to the expected value
> 20 % of the expected value
< 20 % of the expected value
? unknown, can not be measured
R flow profile
information about the flow profile based on the Reynold's number
T fully turbulent flow profile
L fully laminar flow profile
the flow is in the transition range between laminar and turbulent flow
? unknown, can not be calculated
F flow velocity
comparison of the measured flow velocity with the flow limits of the system
ok, flow velocity is not in a critical range
L=(51.2) 50.8 mm
the transducer distance during measurement.
54.5 m3/h
The optimum transducer distance (here: 51.2 mm) will be displayed in parentheses in the
upper line, followed by the entered transducer distance (here: 50.8 mm).
The optimum transducer distance might change during measurement (e.g. due to tem-
perature fluctuations).
A deviation from the optimum transducer distance (here: -0.4 mm) will be compensated
internally.
12.2 Totalizers
Heat quantity, total volume or total mass of the medium at the measuring point can be
determined.
There are two totalizers, one for the positive flow direction, one for the negative flow di-
rection.
The unit of measurement used for totalization corresponds to the heat, volume or mass
unit selected for the physical quantity.
The totalizer value consists of max. 11 digits, including max. 3 decimal places.
To activate the totalizers, press key OON
during measure-
A:Volume(oper.)
ment (see Table 12.1).
54.5 m3/h
display of the totalizer for the press key $ once during measurement
Q +
positive flow direction
display of the totalizer for the
negative flow direction press key O-
once during measurement
Note! The totalizers can only be activated for the measuring channel whose
measured values are just displayed.
Note! A keystroke will influence the totalizers only if the totalizer is dis-
played in the upper line.
Without overflow:
• The totalizer value increases to the internal limit of 1038.
• The values will be displayed as exponential numbers (±1.00000E10), if necessary. The
totalizer can only be reset manually to zero.
With overflow:
• The totalizer will be reset to zero automatically as soon as ±9999999999 is reached.
Select SPECIAL FUNCTION\SYSTEM SETTINGS\MEASURING\QUANT. WRAPPING.
Select ON to work with overflow. Select OFF to work with-
Quant. wrapping
out overflow.
off >ON<
The totalizers can be reset manually to zero independently of the selected list item.
Note! The overflow of a totalizer influences all output channels, e.g. data
memory, online output.
The output of the sum of both totalizers (the throughput Q) via an
output will not be valid after the first overflow (wrapping) of one of
the corresponding totalizers.
To signalize the overflow of a totalizer, an alarm output with switching
condition QUANTITY and type HOLD has to be activated.
If CUT-OFF FLOW\SIGN and USER are selected, two values have to be entered:
If CUT-OFF FLOW\ABSOLUT and USER is selected, only one value has to be entered:
Scroll with key in the upper line until the activated mea-
A:Mode=FastFood
suring mode A:MODE=FASTFOOD or A:MODE=TRANSTIME
54.5 m3/h will be displayed.
If the data memory is activated, a new data set will be defined and storing of measured
values will be started. If the FastFood mode is deactivated or if the measurement is inter-
rupted, storing will be stopped.
Note! The values of the current measuring data set will be deleted when
the FastFood mode is deactivated and activated again without inter-
rupting the measurement.
Note! The values of the current measuring data set are kept if the measure-
ment has been interrupted before the FastFood mode is activated
again. A new measuring data set will be created when the next mea-
surement is started.
Additionally to the ultrasonic measuring channels, the flowmeter has two virtual calcula-
tion channels Y and Z. The measured values of the measuring channels A and B will be
summed up by the calculation channels.
The result of the calculation is the measured value of the selected calculation channel.
This measured value is equivalent to the measured values of a measuring channel. All
operations with the measured values of a measuring channel (totalizing, online output,
storing, outputs, etc.) can be done with the values of a calculation channel, too.
Scroll with key and through the scroll list. All mea-
OOFF
Make sure that the physical quantity selected for the calculation channel can be calculat-
ed from the physical quantities of the selected measuring channels. The possible combi-
nations are shown in Table 12.2.
flow velocity
volume flow
volume flow
mass flow
mass flow
flow velocity x x x x x x
volume flow x x x x
mass flow x x x x
example: The difference of the volume flows of the channels A and B has to be deter-
mined. The physical quantity of measuring channel A can be the volume
flow or the mass flow, but not the flow velocity. The physical quantities of
the two measuring channels do not need to be identical (channel A = mass
flow, channel B = volume flow).
Two cut-off flows for each calculation channel can be defined. They will be defined in the
unit of measurement of the physical quantity selected for the calculation channel.
All positive calculated values less than the limit, will be set
+Cut-off Flow
to 0.
1.00 kg/h
Position the transducers for all activated measuring channels. The measurement will be
started automatically afterwards.
Y: A - B N E X T
53.41 m/s Press key % to display the measured values of the vari-
ous channels.
Online Output
The measured values will be transmitted directly during the measurement to a PC (see
section 13.2.3).
Offline Output
The measured values will be stored in the flowmeter and later transmitted to a PC (see
section 13.2.4).
Measuring point number and parameters will be stored together with the measured val-
ues.
13.1.4 Measurement
If OUTPUT OPTIONS\STORE MEAS.DATA is activated
DATA MEMORY
and SPECIAL FUNCTION\SYSTEM SETTINGS\RING-
OVERFLOW! BUFFER is deactivated, this error message will be dis-
played as soon as the data memory is full. Press EN-
TER.
Note! It is recommended to use the RS485 interface for the online output.
The RS232 interface should be used only if the flowmeter does not
have an RS485 interface.
The line \DATA will be transmitted next, followed once by the column titles (see Table
13.1) for the corresponding channel. The measured values will be transmitted after-
wards.
example: \DATA
A:;\*MEASURE; Q_POS; Q_NEG;
B:;\*MEASURE; Q_POS; Q_NEG;
One data line is transmitted per storage interval and per activated measuring channel.
The line "???" will be transmitted if there are no measured values availabe for the stor-
age interval.
example: With a storage rate of 1 s, 10 lines "???" will be transmitted, if the mea-
surement is restarted after an interruption for transducer positioning of
10 s.
Online Output
During online output, columns will be generated for all quantities which appear during
measurement. The columns Q_POS and Q_NEG will be empty if the totalizers are deac-
tivated.
As the totalizers can not be activated for the physical quantity flow velocity, these col-
umns will not be generated.
Offline Output
During offline output, columns will only be generated if at least one measured value is
stored in the data set. The columns Q_POS and Q_NEG will not be generated if the total-
izers are deactivated.
Transmission Parameters
• the flowmeter sends CRLF-terminated ASCII
• max. line length: 255 digits
• RS232: 9600 bits/s, 8 data bits, even parity, 2 stop bits, protocol (RTS/CTS)
• RS485: 9600 bits/s, 8 data bits, even parity, 1 stop bit
The settings for the RS485 interface can be changed in SPECIAL FUNCTION\SYS-
TEM SETTINGS\NETWORK. These displays will not be indicated if an RS485 interface
does not exist.
Select SPECIAL FUNCTION\SYSTEM SET-
SYSTEM settings
TINGS\NETWORK to change the settings for the trans-
Network mission parameters.
Press ENTER to confirm the instrument address in
Device adress:
the network.
0 ADR
Note! The sample mode does not have influence on the interfaces working
continuously (e.g. current output, voltage output).
If AVERAGE is selected, all primary physical quantities will be aver-
aged, i.e. also the measured temperatures if the corresponding mea-
suring channel is activated.
Note! If a mean value could not be calculated over the complete storage in-
terval while AVERAGE was activated, the value will be marked as in-
valid. "???" appears in the ASCII file of the stored measured values
for invalid mean values and the corresponding physical quantity and
"?UNDEF" for invalid temperatures. There will be no indication of
how many momentary measured values a valid mean value consists
of.
Note! The totalizers will be stored only if they are activated and the data
memory is activated.
Storing of a totalizer reduces the total number of measured values to
be stored by approx. two thirds.
Max. 100 measuring data sets will be stored. The number of measuring data sets de-
pends on the total number of measured values stored in the precedent measuring data
sets.
If the data memory is empty and a measurement is started with one physical quantity on
one measuring channel without storing of the totalizer, approx. 100 000 measured values
can be stored. The available data memory can be displayed:
Select SPECIAL FUNCTION\INSTRUM. INFORM.
Special Function
Press ENTER.
Instrum. Inform.
14 Libraries
The internal material database of the flowmeter contains parameters for pipe and lining
materials as well as for media. It can be extended by user defined materials or media.
User defined materials and media will always be displayed in the scroll lists of the pro-
gram branch PARAMETER.
User defined materials and media can be stored in an integrated coefficient memory (us-
er area). The coefficient memory has to be partitioned first (see section 14.1).
The properties of user defined materials or media can be entered as follows:
• as constants without the extended library (see section 14.2)
• as constants or temperature and pressure dependent functions by means of the ex-
tended library (see section 14.3)
The material and media scroll lists displayed in the program branch PARAMETER can be
arranged (see section 14.5). The shorter scroll lists make the work more efficient.
Enter the number of user defined data sets for the heat
Format USER-AREA
flow coefficients. Press ENTER.
Heat-Coeffs: 00
Heat flow coefficients can be entered only if the flowmeter
is equipped with temperature inputs.
Enter the number of user defined data sets for the steam
Format USER-AREA
coefficients. Press ENTER.
Steam-Coeffs: 00
Steam coefficients can be entered only if the flowmeter is
equipped with temperature inputs.
Enter the number of the user defined data sets for the con-
Format USER-AREA
centration coefficients. Press ENTER.
Concentrat.: 00
The input of steam coefficients is reasonable only if the
flowmeter is equipped with temperature inputs.
The occupancy of the coefficient memory is displayed for a
USER-AREA:
few seconds.
52% used
example 1: The number of user defined materials is reduced from 5 to 3. The data sets
#01 to #03 will be kept. The data sets #04 and #05 will be deleted.
example 2: The number of user defined materials will be increased from 5 to 6. All 5
data sets will be kept.
Note! There are 95 ASCII characters available (letters, capital letters, num-
bers, special characters [!? " + - ( ) > < % * etc.]).
A designation can have max. 16 characters. The input of text is de-
scribed in section 4.3.
Material Properties
Enter the sound velocity of the material.
c-Material
1590.0 m/s For the sound velocity of some materials see annex C, Ta-
ble C.1. Press ENTER.
Medium Properties
Enter the min. and max. value of the sound velocity for the
c-Medium MIN
medium in m/s. Press ENTER.
1400.0 m/s
c-Medium MAX
1550.0 m/s
The dependency of certain material properties from the temperature and pressure can be
described
• as constants
• as linear function
• with polynomials of grade 1 to 4
• with customized interpolation functions
Constants or a linear function are sufficient in most cases.
If e.g. the temperature fluctuations at the measuring point are low compared to the tem-
perature dependency of the material properties, the linearization or the complete neglect
of the temperature dependency will not result in a considerable additional measuring er-
ror.
If, however, the process conditions fluctuate strongly and the medium properties depend
strongly on the temperature (e.g. viscosity of hydraulic oil), polynomials or customized in-
terpolation functions should be used. Contact FLEXIM to find the best solution for the
measuring task.
Material Properties
Enter for the material:
• transversal sound velocity (in m/s)
• longitudinal sound velocity (in m/s)
One to five values depending on the selected function have to be entered. Press ENTER
after each input.
If an already defined material is edited, for each property will be requested whether to
edit it. Select YES or NO. Press ENTER. Change the values if necessary.
Select the type of sound wave to be used for the flow mea-
Default soundsp.
surement. Press ENTER.
long. >TRANS.<
A transversal sound wave has to be selected in most cas-
es.
Enter the typical roughness of the material. Press ENTER.
Roughness
0.4 mm
Medium Properties
Enter for the medium:
• longitudinal sound velocity (in m/s)
• kinematic viscosity (in mm2/s)
• density (in kg/m3)
• gas compressibility factor.
One to five values depending on the selected function have to be entered. Press ENTER
after each input.
If an already defined medium is edited, for each property of some of the functions will be
requested whether to edit it. Select YES or NO. Press ENTER. Change the values if nec-
essary.
Select YES to store the entered properties or NO to quit the
Save changes
menu item without storing. Press ENTER.
no >YES<
Note! The heat flow coefficients can also be edited with the programs Flux-
Data and FluxKoef.
Note! The entered coefficients will not be checked. Absurd values can re-
sult in wrong measured values or in permanent system errors.
Edit Medium Select the list item HEAT-FLOW COEFFS. Press ENTER.
Heat-flow-coeffs
Heat-Coeffs for Select the medium for which the heat flow coefficients
BEER have to be entered.
User defined media will be displayed first, followed by the
media of the internal database.
Select index Select an index for storing the heat flow coefficients of the
02(--not used--) selected medium. Press ENTER.
If the coefficient memory is partitioned so that heat flow co-
efficients for two media can be entered, indices 01 and 02
are available.
Heat-flow Coeffs Enter the 10 heat flow coefficients: a0...a4, r0...r4. Press
0.0 a0 ENTER after each input.
Heat-flow Coeffs Select YES to store the heat flow coefficients. Press EN-
Save? no >YES< TER.
Note! The entered coefficients will not be checked. Absurd values can re-
sult in wrong measured values or in permanent system errors.
Note! If the material/medium scroll list is quit with BRK before storing, all
changes will be lost.
Current list= Press ENTER to return to the scroll list MATERIAL LIST.
Other Material
Note! The materials/media are displayed in the order in which they have
been added.
15 Settings
15.1 Time and Date
The flowmeter has a battery buffered clock. Measured values will be automatically stored
with date and time.
15.1.1 Time
Select in SPECIAL FUNCTION\SYSTEM SETTINGS the
SYSTEM settings
list item SET CLOCK. Press ENTER.
Set Clock
15.1.2 Date
After the time has been set, DATE will be displayed.
Select OK to confirm the date or NEW to set the date. Press
DATE 25.01.2007
ENTER.
ok >NEW<
Note! The settings of the menu item DIALOGS/MENUS will be stored at the
end of the dialog. If the menu item is quit before the end of the dialog,
the settings will not be effective.
Note! The pipe circumference is edited temporarily only. When the flowme-
ter switches back to the display of the pipe circumference (internal re-
calculation), slight rounding errors may occur.
Note! All changes will be stored now at the end of the dialog.
Note! The settings of the menu item MEASURING will be stored at the end of
the dialog. If the menu item is quit before the end of the dialog, the
settings will not be effective.
Cut-off Flow
factory >USER<
Note! All changes will be stored now at the end of the dialog.
OOFF
= min. contrast
= medium contrast
OON
= max. contrast
Note! The display will be reset to medium contrast after a cold start.
16 SuperUser-Mode
The SuperUser mode allows experimental work. Features of the SuperUser mode are:
• Defaults will not be observed.
• There are no plausibility checks when parameters are being entered.
• There is no check whether the entered parameters are within the limit determined by
physical laws and technical data.
• The cut-off flow is not active.
• A value for the number of sound paths has to entered.
It is possible to change the lower limit of the inner pipe diameter for a given transducer
type without the activation of the Super User mode.
16.1 Activating/Deactivating
Press key C. Enter HotCode 071049 .
It is displayed that the SuperUser mode is activated. Press
SUPERUSER MODE
ENTER. The main menu will be displayed.
*IS ACTIVE NOW*
Press key C. Enter HotCode 071049 again to deactivate the SuperUser mode.
It is displayed that the SuperUser mode is deactivated.
SUPERUSER MODE
Press ENTER. The main menu will be displayed.
IS PASSIVE NOW
The SuperUser mode will be deactivated by switching off the flowmeter, too.
17 Inputs
External transducers can be connected to the inputs (if available) to measure the follow-
ing physical quantities:
• temperature
• density
• pressure
• kinematic viscosity
• dynamic viscosity
The values of the current, voltage or temperature inputs can be used by all measuring
channels.
An input has to be linked to a measuring channel (see section 17.1 and 17.3) and has to
be activated (see section 17.4) before it can be used for the measurement or for storing
of measured values.
Depending on the configuration of the flowmeter, one or several of the following list items
will be displayed:
Select the list items for T-FLUID/OUTLET, T(3) and T(4) of measuring channel A and
the other activated channels accordingly. Press ENTER after each input.
If necessary, select the temperature probe for input T2, T3 and T4 accordingly.
Select the list items for EXT. INPUT(2)...(4) of measuring channel A and the other
activated channels accordingly.
Note! The total number of measured values that can be stored will be re-
duced if a temperature input is activated.
Attention! Observe the polarity to avoid damage of the current source. The cur-
rent input might be destroyed by a permanent short circuit.
Inputs have to be activated if the measured values are to be displayed, stored and/or out-
put together with the other measured values.
Note! The total number of measured values that can be stored will be re-
duced if an input is activated.
T1 Corr. Offset Enter the correction value for the temperature input.
0.3 C Press ENTER.
18 Outputs
If the flowmeter is equipped with outputs, they have to be installed and activated before
they can be used:
• assigning a measuring channel (source channel) to the output (if the flowmeter has
more than one measuring channel)
• assigning the physical quantity (source item) to be transmitted to the output by the
source channel, and the properties of the signal
• defining the behavior of the output in case no valid measured value is available
• activation of the installed output in the program branch OUTPUT OPTIONS
Note! The configuration of an output will be stored at the end of the dialog.
If the dialog is quit by pressing key BRK, the changes will not be
stored.
The source items and their scroll lists are described in Table 18.1.
Table 18.1: Configuration of the outputs
source item list item output
MEASURED VALUE ACTUAL MEASURE physical quantity selected in the program branch
OUTPUT OPTIONS
FLOW flow independently of the physical quantity select-
ed in the program branch OUTPUT OPTIONS
HEAT FLOW heat flow independently of the physical quantity
selected in the program branch OUTPUT
OPTIONS
QUANTITY Q+ totalizer for the positive flow direction
* ACTUAL MEASURE totalizer for the physical quantity selected in the
program branch OUTPUT OPTIONS
* FLOW flow totalizer
* HEAT FLOW heat flow totalizer
Q- totalizer for the negative flow direction
* ACTUAL MEASURE totalizer for the physical quantity selected in the
program branch OUTPUT OPTIONS
* FLOW flow totalizer
* HEAT FLOW heat flow totalizer
Q sum of the totalizers (positive and negative flow
direction)
* ACTUAL MEASURE totalizer for the physical quantity selected in the
program branch OUTPUT OPTIONS
* FLOW flow totalizer
* HEAT FLOW heat flow totalizer
LIMIT R1 limit message (alarm output R1)
R2 limit message (alarm output R2)
R3 limit message (alarm output R3)
TEMPERATURE Is available only if a temperature input has been
linked to the channel.
T-INLET (T1) supply temperature for the heat flow
measurement
T-OUTLET (T2) return temperature for the heat flow measurement
T(3)=INPUT T3 further temperature input
T(4)=INPUT T4 further temperature input
TV(=T1)-TR(=T2) difference between supply and return
temperature
TV(=T1)-T3 difference between supply temperature and T(3)
TR(=T2)-T3 difference between return temperature and T(3)
TV(=T1)-T4 difference between supply temperature and T(4)
TR(=T2)-T4 difference between return temperature and T(4)
T3-T4 difference between T(3) and T(4)
IMPULS FROM ABS(X) pulse without sign consideration
FROM X > 0 pulse for positive measured values
FROM X < 0 pulse for negative measured values
MISCELLANEOUS SOUNDSPEED FLUID sound velocity of the medium
CONCENTRATION K concentration
SIGNAL signal amplitude of a measuring channel
I1 Output MAX
11.0 mA
example: The volume flow is selected as source item for a current output, the output
range is 4...20 mA, the error value delay is td > 0.
The volume flow can not be measured during the time interval t0...t1 (see
Fig. 18.1). The error value will be output.
3
v[m
v [ m 3/h]/ h ]
? ? ?
tt
tt00 tt11
Error-value I [m A ]
Minimum (4.0mA)
2 0
td
Error-value I [m A ]
Hold last value
2 0
Error-value I [m A ]
Maximum (20.0mA)
2 0
td
Error-value I [m A ]
Other value ...
2 0
error output = 2.00 mA
td
Note! The settings will be stored now at the end of the dialog.
example: The output range 4…20 mA was selected for a current loop, the zero-scale
value was set to 0 m3/h and the full-scale value to 300 m3/h.
If the volume flow is 300 m3/h, a 20 mA signal is transmitted to the output.
If the volume flow is 0 m3/h, a 4 mA signal is transmitted to the output.
Note! The menu item PULSE OUTPUT will be displayed in the program
branch OUTPUT OPTIONS only if a pulse output has been installed.
The max. flow that the pulse output can work with will be displayed now. This value is cal-
culated from the entered pulse value and pulse width.
If the flow exceeds this value, the pulse output will not function properly. In this case,
pulse value and pulse width have to be adapted to the flow conditions. Press ENTER.
Max. 3 alarm outputs R1, R2, R3 per channel operating independently of each other can
be configured. The alarm outputs can be used to output information on the current mea-
surement or to start and stop pumps, motors, etc.
A scroll list will be selected in the upper line with key and . A list item will be se-
O- O+
OON
lected in the lower line with key and .
OOFF
If the switching condition QUANTITY is selected in the scroll list FUNC, the limit of the out-
put has to be defined:
A positive limit will be compared to the totalizer value for the positive flow direction.
A negative limit will be compared to the totalizer value for the negative flow direction.
The comparison will also be made if the totalizer of the other flow direction is displayed.
Note! The unit of measurement of the limit corresponds to the unit of mea-
surement of the selected physical quantity.
If the unit of measurement of the physical quantity is changed, the
limit has to be converted and entered again.
Press three times key C during measurement to set all alarm outputs to the idle state.
Alarm outputs whose switching condition is still met will be activated again after 1 s. This
function is used to reset alarm outputs of type HOLD if the switching conditions is not met
anymore.
By pressing key BRK, the measurement will be stopped and the main menu selected. All
alarm outputs will be de-energized, independently of the programmed idle state.
Alarm outputs with switching condition ERROR will be activated only after several unsuc-
cessful measuring attempts. Therefore, typical short-term disturbances of the measure-
ment (e.g. switching on of a pump) will not activate the alarm.
Alarm outputs with switching condition +- -+ and of type NON-HOLD will be activated
with each change of the flow direction for approx. 1 s (see Fig. 18.2).
Alarm outputs with switching condition +- -+ and of type HOLD will be activated after
the first change of the flow direction. They can be switched back by pressing key C three
times (see Fig. 18.2).
+ flow
D u r c h flu s s + flow
D u r c h flu s s
0 0
- -
type
A l a r NON-HOLD
m ty p : N IC H T H A L T E N D type
A l a r HOLD
m ty p : H A L T E N D
approx.
c a . 1 s 1s resetM ofa n u the
e l l e s alarm
R ü c k s e tz e n
C key
(3x d e s C)A la r m s
The alarm state can be displayed during measurement. This function is activated in SPE-
CIAL FUNCTION\SYSTEM SETTINGS\DIALOGS/MENUS.
Select the menu item SHOW RELAIS STAT. Select ON to
SHOW RELAIS STAT
activate the display of the alarm state.
off >ON<
Scroll during measurement with key until the alarm state is displayed in the upper line.
example: R1 =
3 MIN
+ - - +
QUANTITY
ERROR
19 Troubleshooting
If any problem appears which can not be solved with the help of this manual, contact our
sales office giving a precise description of the problem. Specify the type, serial number
and firmware version of the flowmeter.
Calibration
FLUXUS is a very reliable instrument. It is manufactured under strict quality control, us-
ing modern production techniques. If installed as recommended in an appropriate loca-
tion, used cautiously and taken care of conscientiously, no troubles should appear. The
flowmeter has been calibrated at the factory and usually, a re-calibration of the flowmeter
will not be necessary. A re-calibration is recommended if
• the contact surface of the transducers show visible wear or
• the transducers were used for a prolonged period at a high temperature (several
months >130 °C for normal transducers or > 200 °C for high temperature transducers).
The flowmeter has to be sent to FLEXIM for recalibration under reference conditions.
The flowmeter does not react when key BRK is pressed during measurement
A program code has been defined. Press key C and enter the program code.
The backlight of the display does not light, but all other functions are available.
The backlight is defective. This problem has no influence on the other functions of the
display. Send the flowmeter to FLEXIM for repair.
Date and time are wrong, the measured values will be deleted when the flowmeter
is switched off
The data backup battery has to be replaced. Send the flowmeter to FLEXIM.
Measurements on porous pipe materials (e.g. concrete or cast iron) are only possi-
ble under certain conditions
Contact FLEXIM.
The pipe liner may cause problems during measurement if it is not attached tightly
to the inner pipe wall or consists of acoustically absorbing material
Try measuring on a liner free section of the pipe.
Higher proportions of droplets or solids in the medium scatter and absorb the ul-
trasound and therefore attenuate the measuring signal
A measurements is impossible if the value is 10 %. If the proportion is high, but < 10 %,
a measurement might be possible under certain conditions.
The flow velocity to be measured is outside the measuring range of the flowmeter
The operating volume flow meets the expectations, but the standard volume flow
deviates strongly.
The parameters for the measurement of the standard volume flow are not correctly en-
tered (see section 15.4).
flowmeter
power supply 100...240 V/50...60 Hz or
20...32 V DC
power consumption < 15 W
number of flow mea- 1, option: 2
suring channels
signal damping 0...100 s, adjustable
measuring cycle (1 100...1000 Hz
channel)
data logger
loggable values all physical quantities and totalized values
capacity > 100 000 measured values
communication
interface - process integration: - process integration: - process integration:
option: option: option:
RS485 (Modbus, emit- RS485 (Modbus, emit- RS485 (Modbus, emit-
ter) ter) ter)
- diagnosis: RS232 - diagnosis: RS2321 - diagnosis: RS232
serial data kit (option)
software (all Win- - FluxData: download of measured data, graphical presentation,
dowsTM versions) conversion to other formats (e.g. for ExcelTM)
- FluxKoeff: creating medium data sets
cable RS232 RS2321 RS232
adapter RS232 - USB RS232 - USB1 RS232 - USB
outputs (option)
The outputs are galvanically isolated from the flowmeter.
number on request
current output
range 0/4...20 mA
accuracy 0.1 % of reading ±15 A
active output Rext < 500
passive output Uext = 4...24 V, dependent on Rext, Rext < 1 k
voltage output
range 0...1 V or 0...10 V
accuracy 0...1 V: 0.1 % of reading ±1 mV
0...10 V: 0.1 % of reading ±10 mV
internal resistance Ri = 500
frequency output
range 0...1 kHz or 0...10 kHz
open collector 24 V/4 mA
1 connection of the interface RS232 outside of explosive atmosphere (housing cover open)
70.5
10
Ø 4.5 screw holes for wall mount-
287 ing
265
200
163
M2.5
122.5
129
43.7
222.2
170.5
110
208.25
15
129
Transducers
Shear Wave Transducers (for ATEX zone 1)
technical type GDK1N31 GDM1N31 GDG1N31
order code GSK-NA1TS GSM-NA1TS GSG-NA1TS
transducer frequency MHz 0.5 1 0.2
medium pressure1
min. extended bar 20 20 20
min. bar metal pipe: 30 metal pipe: 30 metal pipe: 30
plastic pipe: 1 plastic pipe: 1 plastic pipe: 1
outer pipe diameter2
min. extended mm 70 30 250
min. recommended mm 80 40 380
max. recommended mm 500 80 810
max. extended mm 720 120 1100
pipe wall thickness
min. mm 5 2.5 14
max. mm - - -
material
housing PEEK with stainless steel PEEK with
stainless steel 316 Ti (1.4571) stainless steel
cap304 cap304
(1.4301) (1.4301)
contact surface PEEK PEEK PEEK
degree of protection IP 65 IP 65 IP 65
according to EN 60529
transducer cable
type 2549 2549 2549
length m 5 4 5
dimensions
length l mm 126.5 60 129.5
width b mm 50 30 50
height h mm 53.5 33.5 64
dimensional drawing
h
h
h
l l l
b
b
b
operating temperature
min. °C -40 -40 -40
max. °C +130 +130 +130
l l
b
operating temperature
min. °C -40 -40
max. °C +130 +130
h
h
l l l
l
b
b
b
operating temperature
min. °C -40 -40 -40 -40
max. °C +100 +100 +100 +100
Shear Wave Transducers (for ATEX zone 1/2 (gas/dust), high temperature)
technical type GDM2E85 GDP2E85
order code GSM-EA1TS GSP-EA1TS
transducer frequency MHz 1 2
medium pressure1
min. extended bar 20 20
min. bar metal pipe: 30 metal pipe: 30
plastic pipe: 1 plastic pipe: 1
outer pipe diameter2
min. extended mm 30 15
min. recommended mm 40 20
max. recommended mm 80 40
max. extended mm 120 60
pipe wall thickness
min. mm 2.5 1.5
max. mm - -
material
housing PI with stain- PI with stain-
less steel less steel
cap304 cap304
(1.4301) (1.4301)
contact surface PI PI
degree of protection IP 56 IP 56
according to EN 60529
transducer cable
type 6111 6111
length m 4 4
dimensions
length l mm 69.5 69.5
width b mm 32.5 32.5
height h mm 61 61
dimensional drawing
h
l l
b
operating temperature
min. °C -30 -30
max. °C +200 +200
h
h
l l l l
b
b
b
operating temperature
min. °C -40 -40 -40 -40
max. °C +130 +130 +130 +130
Shear Wave Transducers (for ATEX zone 2 and without Explosion Protection, IP
68)
technical type GDK1LI8 GDM2LI8 GDP2LI8 GDG1LI8
order code GSK-NA2TS/ GSM-NA2TS/ GSP-NA2TS/ GSG-NA2TS/
IP68 IP68 IP68 IP68
GSK-NNNTS/ GSM-NNNTS/ GSP-NNNTS/ GSG-NNNTS/
IP68 IP68 IP68 IP68
transducer frequency MHz 0.5 1 2 0.2
medium pressure1
min. extended bar 20 20 20 20
min. bar metal pipe: 30 metal pipe: 30 metal pipe: 30 metal pipe: 30
plastic pipe: 1 plastic pipe: 1 plastic pipe: 1 plastic pipe: 1
outer pipe diameter2
min. extended mm 70 30 15 250
min. recommended mm 80 40 20 380
max. recommended mm 500 80 40 810
max. extended mm 720 120 60 1100
pipe wall thickness
min. mm 5 2.5 1.5 14
max. mm - - - -
material
housing PEEK with PEEK with PEEK with PEEK with
stainless steel stainless steel stainless steel stainless steel
cap316Ti cap316Ti cap316Ti cap316Ti
(1.4571) (1.4571) (1.4571) (1.4571)
contact surface PEEK PEEK PEEK PEEK
degree of protection IP 68 IP 68 IP 68 IP 68
according to EN 60529
transducer cable
type 2550 2550 2550 2550
length m 12 12 12 12
dimensions
length l mm 128.5 70 70 128.5
width b mm 50 28 28 50
height h mm 72 42 42 75
dimensional drawing
h
h
h
l l l l
b
b
b
operating temperature
min. °C -40 -40 -40 -40
max. °C +100 +100 +100 +100
h
h
l l l l
b
b
b
operating temperature
min. °C -40 -40 -40 -40
max. °C +130 +130 +130 +130
1
depending on application, typical value for natural gas, N2, compressed air
2
shear wave transducers:
typical values for natural gas, N2, O2, pipe diameters for other gases on request
pipe diameter min. recommended/max. recommended/max. extended: in diagonal mode and for
a flow velocity of 15 m/s
l
b
operating temperature
min. °C -30 -30
max. °C +200 +200
1
depending on application, typical value for natural gas, N2, compressed air
2
shear wave transducers:
typical values for natural gas, N2, O2, pipe diameters for other gases on request
pipe diameter min. recommended/max. recommended/max. extended: in diagonal mode and for
a flow velocity of 15 m/ss
Shear Wave Transducers (high temperature, for ATEX zone 2 or without Explo-
sion Protection)
technical type GDM2E52 GDP2E52 GDQ2E52
order code GSM-EA2TS GSP-EA2TS GSQ-EA2TS
GSM-ENNTS GSP-ENNTS
transducer frequency MHz 1 2 4
medium pressure1
min. extended bar 20 20
min. bar metal pipe: 30 metal pipe: 30
plastic pipe: 1 plastic pipe: 1
outer pipe diameter2
min. extended mm 30 15
min. recommended mm 40 20
max. recommended mm 80 40
max. extended mm 120 60
pipe wall thickness
min. mm 2.5 1.5
max. mm - -
material
housing PI with stain- PI with stain- PI with stain-
less steel less steel less steel
cap304 cap304 cap304
(1.4301) (1.4301) (1.4301)
contact surface PI PI PI
degree of protection IP 65 IP 65 IP 65
according to EN 60529
transducer cable
type 6111 6111 6111
length m 4 4 3
dimensions
length l mm 69.5 69.5 39
width b mm 32.5 32.5 18
height h mm 61 61 24.5
dimensional drawing
h
l l l
b
operating temperature
min. °C -30 -30 -30
max. °C +200 +200 +200
l l l
b
operating temperature
min. °C -40 -40 -40
max. °C +170 +170 +170
l
l
b
operating temperature
min. °C -40 -40 -40
max. °C +170 +170 +170
l l l
b
operating temperature
min. °C -40 -40 -40
max. °C +170 +170 +170
l
l
b
operating temperature
min. °C -40 -40 -40
max. °C +170 +170 +170
Lamb Wave Transducers (without Explosion Protection and with Connection System AS)
l l l
b
operating temperature
min. °C -40 -40 -40
max. °C +170 +170 +170
1 depending on application, typical value for natural gas, N , compressed air
2
2 Lamb wave transducers:
typical values for natural gas, N2, O2, pipe diameters for other gases on request
pipe diameter min. recommended/max. recommended: in reflection mode and for a flow velocity
of 15 m/s
pipe diameter max. extended: in diagonal mode and for a flow velocity of 25 m/s
Lamb Wave Transducers (without Explosion Protection and with Connection Sys-
tem AS)
technical type GRM1NC3 GRQ1NC3 GRP1NC3
order code GLM-NNNAS GLQ-NNNAS GLP-NNNAS
transducer frequency MHz 1 4 2
medium pressure1
min. extended bar - - -
min. bar metal pipe: metal pipe: metal pipe:
10 (> DN 60) 10 (> DN 15) 10 (> DN 35)
5 (< DN 60) 5 (< DN 15) 5 (< DN 35)
outer pipe diameter2
min. extended mm 30 7 15
min. recommended mm 40 10 20
max. recommended mm 90 22 50
max. extended mm 150 35 70
pipe wall thickness
min. mm 2 0.5 1
max. mm 5 1 3
material
housing PPSU with PPSU with PPSU with
stainless steel stainless steel stainless steel
cap304 cap304 cap304
(1.4301) (1.4301) (1.4301)
contact surface PPSU PPSU PPSU
degree of protection IP 65 IP 65 IP 65
according to EN 60529
transducer cable
type 1699 1699 1699
length m 4 3 4
dimensions
length l mm 74 42 74
width b mm 28 18 28
height h mm 42.9 25.5 42.9
dimensional drawing
h
h
h
l l l
b
b
b
operating temperature
min. °C -40 -40 -40
max. °C +170 +170 +170
1 depending on application, typical value for natural gas, N , compressed air
2
2
Lamb wave transducers:
typical values for natural gas, N2, O2, pipe diameters for other gases on request
pipe diameter min. recommended/max. recommended: in reflection mode and for a flow velocity
of 15 m/s
pipe diameter max. extended: in diagonal mode and for a flow velocity of 25 m/s
Junction Box
technical type JB01S4E3M JB02 JB03
dimensions see dimensional draw- see dimensional draw- see dimensional draw-
ing ing ing
fixation wall mounting wall mounting wall mounting
option: option: option:
2 " pipe mounting 2 " pipe mounting 2 " pipe mounting
material
housing stainless steel 316L stainless steel 304 stainless steel 304
(1.4404) (1.4301) (1.4301)
gasket silicone silicone silicone
degree of protec- IP 67 IP 67 IP 67
tion according to
EN 60529
operating temperature
min. °C -40 -40 -40
max. °C +80 +80 +80
explosion protection
ATEX zone 1 2 -
marking 0044 II3G Ex nA II -
II2G Ex e mb II T6...T4
(T6)...T4 Ta -40...+80 °C
Ta -40...+(70) 80 °C
II 2D Ex tD A21 IP 67
T 100 °C
certification IBExU06ATEX1161 - -
type of protection junction box: non sparking -
increased safety
decoupled network:
encapsulation
Dimensions 156
174
wall mounting
holder
Ø
70 9
2
119 163.5
Terminal Assignment
JB01
Transducers
terminal strip KL1
terminal connection
V signal
T V T V S T G T G T R S T R V V S R S R VS shield
K L 2 K L 1 RS shield
R signal
JB02, JB03
Transducers
terminal connection
XV SMB connector
XR SMB connector
T V T V S T G T G T R S T R
K L 2
Extension Cable (Flowmeter)
X V X R
Technical Data
Damping Mats (Option)
Damping mats will be used for the gas measurement to reduce noise influences on the
measurement.
Transducer damping mats will be installed below the transducers.
Pipe damping mats will be installed at reflection points, e.g. flange, welding.
b
D l
A Technical Data
Units of Measurement
1 US gallon = 3.78 l
1 barrel = 42 US gallons = 158.76 l
g a l/ m in
l/ m in
m 3 /h
b b l/ d
l/ s
10 5
10 4
10 5 10 6
0 0
10
10 3
DN 0
80
DN 10 4
0
60
DN 05 0
10 3
DN 0
40 10 5
10 4
DN
0
100 30
DN 250
ND 10 3
0
20
DN
0
100 15
DN
10 3 10 4
1 00
10 DN
8 0
DN 100
6 5
DN
10 50
DN
100 10 3
1 30
DN
10
10 100
0 ,1
1
0 ,1 1 10
flow
S t r ö velocity
m u n g s [m/s]
g e sc h w in d ig k e it (m / s )
D N 0
m 3 / h
b b l/ d
l/ m in
6 0
0
l/ s
D N 5 0
1 0 D N
5
0
4 0
D N 0
3 0 1 0 4
0
1 0 6 D N 2 5 1 0 5
D N 02 0
D N 1 0 3
0
1 0 4 1 5
D N
0
1 0 1 0 3
D N
1 0 5
8 0 1 0 4
D N 6 5
D N 1 0 0
1 0 3 5 0
D N
3 0 1 0 0
1 0 4 D N
1 0 3
1 0
1 0 0
1 0
1 0 3
1 0 0
1
1 0
1
1 0 0 1 0
0 ,1
1
1 1 0 1 0 0
S t r ö velocity
flow m u n g s g [m/s]
e s c h w in d ig k e it (ft/ s )
B Menu Structure
cold start
resistant
cold start
resistant
cold start
resistant
cold start
resistant
cold start
resistant
cold start
resistant
Current Loop
cold start
resistant
Pulse Output
cold start
resistant
Alarm Output
cold start
resistant
SYSTEM SETTINGS
SYSTEM SETTINGS\LIBRARIES
cold start
resistant
cold start
resistant
SYSTEM SETTINGS\DIALOGS/MENUS
cold start
resistant
SYSTEM SETTINGS\MEASURING
cold start
resistant
cold start
resistant
SYSTEM SETTINGS\GAS-MEASURING
cold start
resistant
SYSTEM SETTINGS\STORING
cold start
resistant
SYSTEM SETTINGS\NETWORK
cold start
resistant
SYSTEM SETTINGS\MISCELLANEOUS
INSTRUM. INFORM.
PRINT MEAS.VAL.
cold start
resistant
DELETE MEAS.VAL.
INSTALL MATERIAL
cold start
resistant
cold start
resistant
INSTALL MEDIUM
cold start
resistant
cold start
resistant
C Reference
The following tables provide assistance for the user. The accuracy of the data depends
on the composition, the temperature and the manufacturing process of the material.
FLEXIM does not assume liability for any inaccuracies.
Take into consideration for the measuring task that the sound velocity depends on the
composition and the manufacturing process of the material.
The sound velocity of alloys and cast materials strongly fluctuates. The values serve as
an orientation only.
Autotex is resistant according to DIN 42115, part 2 to acetic acid for a contact time <1 h
without visible damage.
Autotex is not resistant to following chemicals:
• concentrated mineral acids • benzyl alcohol
• concentrated alkaline solutions • methylene chloride
• high pressure steam >100 °C
D Certificates
UMG70XV3-4EN 12.01.2009