0% found this document useful (0 votes)
30 views14 pages

Manhi 8033

The document provides instructions for using multiple portable conductivity and TDS meters. It includes detailed specifications, operational instructions, calibration procedures, maintenance instructions, and other information to properly use the devices.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
30 views14 pages

Manhi 8033

The document provides instructions for using multiple portable conductivity and TDS meters. It includes detailed specifications, operational instructions, calibration procedures, maintenance instructions, and other information to properly use the devices.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 14

Untitled-19

1 2

Instruction Manual Dear Customer,


Thank you for choosing a Hanna Product. Please read this instruction
manual carefully before using the instrument. It will provide you with
the necessary information for a correct use of the instrument, as well
HI 8033 - HI 8633 as a more precise idea of its versatility.

HI 8733 - HI 8734 These instruments are in compliance with the


EN 50081-1 and EN 50082-1.
directives

HI 933000
Portable Multi-Range
Conductivity/TDS Meters TABLE OF CONTENTS

PRELIMINARY EXAMINATION ....................................................... 3


mS GENERAL DESCRIPTION .............................................................. 3
FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 8033 ......... 4
FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 8633 ......... 5
FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 8733 ......... 6
HI 8633
ins rum
ns
FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 8734 ......... 7
Conduc
COND.
tivity me
ter FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 933000 ..... 8
ON
TEMP.
OFF OPERATIONAL GUIDE ................................................................. 9
199.9 CALIBRATION ........................................................................... 11
µS 1999
µS CONDUCTIVITY VERSUS TEMPERATURE CHART ........................... 18
19.99
mS 199.9 TDS VERSUS TEMPERATURE CHART .......................................... 19
DETERMINING THE TEMPERATURE COEFFICIENT OF A SOLUTION
mS

(HI 8733) .............................................................................. 20


TEMPE
RATURE PROBE MAINTENANCE .............................................................. 21
BATTERY REPLACEMENT ............................................................ 22
SHOCKPROOF RUBBER BOOTS (for HI 933000 only) ................. 23
ACCESSORIES .......................................................................... 23
WARRANTY .............................................................................. 25
CE DECLARATION OF CONFORMITY ............................................ 26

ISO 9000 Certified


These Instruments are in
http://www.hannainst.com Compliance with the CE Directives Company since 1992
2
Untitled-19

3 4

PRELIMINARY EXAMINATION FUNCTIONAL DESCRIPTION &


SPECIFICATIONS OF HI 8033
Remove the instrument from the packing material and examine it
carefully to make sure that no damage has occurred during shipping.
If there is any noticeable damage, notify your Dealer or the nearest 1
mS

Hanna office immediately.


Each meter is supplied with:
• Conductivity probe with 1m (3.3') cable Conductivity meter
HI 8033

• 9V battery. ppm
CaCO3
salinity
0-19,990
µS
0-199,9
µS
0-1999

Note: Save all packing materials until you are sure that the instru- OFF µS
0-19,990

ment functions correctly. Any damaged or defective item must 2


be returned in its original packing materials together with the
supplied accessories. °C k%
0 10 10 15
20 + 5

3 30
0
5
4
-
50 40 10 15

ins rum n s

1) 3½-digit Liquid Crystal Display


GENERAL DESCRIPTION 2) Rotary switch
3) Manual temperature compensation knob
HI 8033, HI 8633, HI 8733, HI 8734 and HI 933000 are some of 4) Calibration knob
the most complete and versatile portable conductivity/TDS meters Range µS/cm 0.0 to 199.9 / 0 to 1999
ever manufactured. Designed with utmost precision and simplicity, mS/cm 0.00 to 19.99
these meters provide for up to 3 or 4 measurement ranges. The ppm CaCO 3 TDS 0 to 19990
conductivity of a solution depends on the temperature and for this Resolution µS/cm 0.1 / 1
reason measurements are carried out with reference to a standard mS/cm 0.01
temperature of 25°C. If the solution measured has a different tem- ppm CaCO 3 TDS 10
perature than 25°C, compensation must be performed. Accuracy ±1% Full Scale
(@20°C/68°F) excluding probe error
HI 8033, HI 8633 and HI 8734 compensate for temperature manu-
ally. Typical EMC ±2 % Full Scale
Deviation
HI 8733 and HI 933000, with a built-in temperature sensor and
Calibration Manual single setpoint through K% knob
circuitry, automatically compensate for temperature changes.
Temperature Manual from 0 to 50°C (32 to 122°F)
With HI 8733, the temperature coefficient is adjustable from 0 to
Compensation with a ß of 2% per degree °C
2.5% per degree Celsius depending on the type of solution tested.
Probe (included) HI 76301W with 1 m (3.3') screened cable
For all the other meters the temperature coefficient is fixed at 2%.
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type 9 Volt (alkaline)
Life 100 hours of continuous use
Dimensions 185 x 82 x 40 mm (7.3 x 3.2 x1.8")
Weight 355 g (13 oz.)

3 4
Untitled-19

5 6

FUNCTIONAL DESCRIPTION & FUNCTIONAL DESCRIPTION &


SPECIFICATIONS OF HI 8633 SPECIFICATIONS OF HI 8733

mS
1 mS
1

6 5
ins rum n s
HI 8633 Conductivity meter ins rum n s

COND. ON
3 2 HI 8733 ON
TEMP. OFF Conductivity
meter OFF
2
199.9 1999
µS µS 199.9 1999
µS µS
4
19.99 199.9 3
mS mS 19.99 199.9
mS mS

5 0 2.5

4 0.5 2

TEMPERATURE 1 1.5
TEMPERATURE COEFFICIENT

1) 3½-digit Liquid Crystal Display


2) ON/OFF key 1) 3½-digit Liquid Crystal Display
3) Conductivity/temperature selection key 2) ON/OFF key
4) Measurement range selection keys 3) Measurement range selection keys
5) Manual temperature compensation knob 4) Automatic Temperature Compensation coefficient knob
6) Calibration trimmer 5) Calibration trimmer

Range µS/cm 0.0 to 199.9 / 0 to 1999 Range µS/cm 0.0 to 199.9 / 0 to 1999
mS/cm 0.00 to 19.99 / 0.0 to 199.9 mS/cm 0.00 to 19.99 / 0.0 to 199.9
Resolution µS/cm 0.1 / 1 Resolution µS/cm 0.1 / 1
mS/cm 0.01 / 0.1 mS/cm 0.01 / 0.1
Accuracy 1% Full Scale Accuracy 1% Full Scale
(@ 20°C / 68°F) excluding probe error (@ 20°C / 68°F) excluding probe error
Typical EMC ±2 % Full Scale Typical EMC ±2 % Full Scale
Deviation Deviation
Calibration Manual single setpoint through trimmer Calibration Manual single setpoint through trimmer
Temperature Manual from 10 to 40°C (50 to 104°F) Temperature Automatic from 0 to 50°C (32 to 122°F) with
Compensation with a ß of 2% per degree °C Compensation an adjustable ß from 0 to 2.5% per degree °C
Probe (included) HI 76301W with 1 m (3.3') screened cable Probe (included) HI 7633W ATC with 1 m (3.3') screened cable
Environment 0 to 50°C (32 to 122°F); Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing max 95% RH non-condensing
Battery Type 9 Volt (alkaline) Battery Type 9 Volt (alkaline)
Life 100 hours of continuous use Life 100 hours of continuous use
Dimensions 185 x 82 x 45 mm (7.3 x 3.2 x1.8") Dimensions 185 x 82 x 45 mm (7.3 x 3.2 x1.8")
Weight 355 g (13 oz.) Weight 355 g (13 oz.)

5 6
Untitled-19

7 8

FUNCTIONAL DESCRIPTION & FUNCTIONAL DESCRIPTION &


SPECIFICATIONS OF HI 8734 SPECIFICATIONS OF HI 933000

1) 3½-digit Liquid Crystal Display


1) 3½-digit Liquid Crystal Display 2) ON/OFF key
2) ON/OFF key 3) Measurement range selection keys
3) TDS/temperature selection key 4) Probe connector
4) Measurement range selection keys
5) Manual temperature compensation knob
6) Calibration trimmer
Range µS/cm 0.0 to 199.9 / 0 to 1999
Range mg/L 0.0 to 199.9 / 0 to 1999 mS/cm 0.00 to 19.99 / 0.0 to 199.9
g/L 0.00 to 19.99 Resolution µS/cm 0.1 / 1
Resolution mg/L 0.1 / 1 mS/cm 0.01 / 0.1
g/L 0.01 Accuracy 1% Full Scale
Accuracy ±1% Full Scale (@ 20°C / 68°F) excluding probe error
(@ 20°C / 68°F) excluding probe error Typical EMC ±2% Full Scale
Typical EMC ±2 % Full Scale Deviation
Deviation Calibration Manual single setpoint through
Calibration Manual single setpoint through trimmer trimmer in the battery compartment
Temperature Manual from 0 to 50°C (32 to 122°F) Temperature Automatic from 10 to 40°C (50 to 104°F)
Compensation with a ß of 2% per degree °C Compensation with a ß of 2% per degree °C
Probe (included) HI 76301W with 1 m (3.3') screened cable Probe (included) HI 76302W ATC with 1 m (3.3') screened cable
Environment 0 to 50°C (32 to 122°F); Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing max 95% RH non-condensing
Battery Type 9 Volt (alkaline) Battery Type 9 Volt (alkaline)
Life 100 hours of continuous use Life 100 hours of continuous use
Dimensions 185 x 82 x 45 mm (7.3 x 3.2 x1.8") Dimensions 143 x 80 x 38 mm (5.6 x 3.2 x1.5")
Weight 355 g (13 oz.) Weight 360 g (13 oz.)

7 8
Untitled-19

9 10

With HI 8633 and HI 8734 press the COND/TEMP key (for


OPERATIONAL GUIDE HI 8633) or TDS/TEMP (for HI 8734) to display the temperature
and adjust the temperature knob to that of the solution e.g. 20°C.
• Each meter is supplied complete with a 9V battery. Slide off the
battery compartment cover on the back of the meter (see page 22). COND.
TEMP.

Install the battery while paying attention to its polarity.


• Connect the probe to the meter securely by aligning the pins with TDS ºC

the socket and pushing the plug in.


TEMP.

With HI 933000, tighten the threaded ring. • Select the appropriate measurement range.
Note: If the display shows only a "1" on the
far left hand side, the meter is out of
range. Select the next (higher) range.
• For HI 8733:
Adjust the TEMPERATURE COEFFICIENT 0 2.5

knob to 2% to compensate for the tem- 0.5 2

perature effect of average solutions (to


1 1.5

• Make sure that the meter has been calibrated before taking any TEMPERATURE COEFFICIENT

measurements (see page 11 for calibration procedure). determine exact value for a particular so-
• Immerse the conductivity probe into the lution, see page 20).
sample, with the holes on the shaft com- • For HI 8733 and HI 933000:
pletely submerged. Wait for a couple of minutes for the temperature sensor to reach
If possible, use plastic beakers or con- thermal equilibrium with the sample before taking measure-
tainers to minimize any EMC interference. ments.
• Tap the probe lightly on the bottom of the beaker to remove any When the sample's temperature is lower than 20°C or higher
air bubbles which may be trapped inside the PVC sleeve. than 30°C, allow more time for the thermal equilibrium of the
• Turn the instrument on by pressing the ON
system to be achieved.
• After the measurement has been completed, the instrument should
µS
ppm salinity
OFF 0-199,9

ON/OFF key or by setting the rotary switch


CaCO3 0-19,990 µS
0-1999

OFF µS

be switched off and the probe should be cleaned and dried (see
0-19,990

(for HI 8033 only) to the desired mea- ON


OFF

surement range. "Probe Maintenance" on page 21).


• For HI 8033, HI 8633 and HI 8734 only: °C

Take the temperature of the solution with


a ChecktempC or an accurate thermom-
eter.

With HI 8033 set the temperature knob to


the measured value e.g. 20°C.
TEMPERATURE

9 10
Untitled-19

11 12

• All subsequent measurements will be compensated to 25°C (77°F).


CALIBRATION
If you prefer to standardize the temperature compensation to
Accessories needed: 20°C (68°F) rather than 25°C (77°F), leave the °C knob at 18°C
(if the temperature of your solution is 18°C), adjust the trimmer
• Use any calibration solution within the meter's range. The
to read "11.67 mS" (see the conductivity vs. temperature chart at
solution should ideally be close to the samples being mea-
page 18). All subsequent measurements will be compensated to
sured. Use for example HI 7030 or HI 8030, 12880 µS/cm
20°C.
(=12.88 mS/cm) conductivity solution for HI 8033, HI 8633,
HI 8733 and HI 933000 and HI 7032, 1382 mg/L • The calibration is now complete and the instrument is ready for
(=2764 µS/cm) TDS solution for HI 8734 use.
• ChecktempC or an accurate thermometer with 0.1°C resolution The instrument should be re-calibrated at least once a month, or
(not necessary for HI 8733 and HI 933000) when the probe or battery is changed.
• a small screwdriver. Note: For more accurate results, it is advisable to use a calibration
solution close to the range to be measured. See the accessories
PROCEDURE FOR HI 8033 section on page 23 for a wide selection of conductivity solu-
0
703

• Pour sufficient quantity of a conductivity


HI

tions.
calibration solution (e.g. HI 7030/
PROCEDURE FOR HI 8633

0
703
HI 8030) into a beaker to cover the holes

HI
on the probe. If possible, use plastic bea- • Pour sufficient quantity of a conductivity
kers to minimize any EMC interference. calibration solution (e.g. HI 7030/
• Immerse the conductivity probe, making
°C
HI 8030) into a beaker to cover the holes
sure that holes are completely submerged, on the probe. If possible, use plastic bea-
and the ChecktempC in the solution. kers to minimize any EMC interference.
• Immerse the conductivity probe, making
°C

• Wait for a couple of minutes for thermal


equilibrium to be reached. sure that holes are completely submerged,
and the ChecktempC in the solution.
• Tap the probe on the bottom, then shake it while rotating to
make sure no air bubbles remain trapped in the sleeve. • Wait for a couple of minutes for thermal
equilibrium to be reached.
• Record the temperature of the HI 7030/HI 8030 calibration
solution from the ChecktempC (e.g. 18°C). • Tap the probe on the bottom, then shake it while rotating to make
sure no air bubbles remain trapped in the sleeve.
• Turn the °C knob to 18°C. • Record the temperature of the HI 7030/HI 8030 buffer solution
from the ChecktempC (e.g. 18°C).
ppm salinity
µS
0-199,9
• Switch the instrument on by pressing ON/OFF. ON
CaCO3 0-19,990 µS OFF
• Turn the rotary knob to select 19990 µS/cm OFF
0-1999

µS

range.
0-19990

COND.
• Press COND/TEMP to display the temperature.
• Turn the k% calibration knob until the display shows the conduc- TEMP.

tivity reading at 25°C (see the conductivity vs. temperature chart


• Adjust the TEMPERATURE knob to display 18°C.
on page 18). E.g. @ 25°C, 12880 µS/cm = 12.88 mS/cm.
k%
10 15 mS
+ 5
0
5
- TEMPERATURE
10 15

11 12
Untitled-19

13 14

COND.
• Press COND/TEMP again to display conductivity PROCEDURE FOR HI 8733

0
7 03
HI
TEMP.
measurement. • Pour sufficient quantity of a conductivity
calibration solution (e.g. HI 7030/
• Select 19.99 mS/cm range by pressing the 19.99 HI 8030) into a beaker to cover the holes
appropriate range key.
mS
on the probe. If possible, use plastic bea-
kers to minimize any EMC interference.
• Immerse the conductivity probe in the
• Adjust the calibration trimmer (see page 5, #6) on the side of
solution, making sure that holes are com-
the instrument with the calibration screwdriver until the display
pletely submerged.
shows the conductivity reading at 25°C (see the conductivity vs.
temperature on page 18). E.g. @ 25°C, 12880 µS/cm = • Wait for a couple of minutes for thermal
12.88 mS/cm. equilibrium to be reached.
mS
• Tap the probe on the bottom, then shake it while rotating to
make sure no air bubbles remain trapped in the sleeve.
ins rum

• Switch the instrument on by pressing


HI 8633 ns
ON
Conductivity
meter
COND.
TEMP. ON
OFF
mS
ON/OFF. OFF
199.9
µS 1999
µS
19.99
mS 199.9

• Set the temperature coefficient knob to


mS

2% to compensate for the temperature


TEMPERA
TURE
0 2.5

effect of average solutions (to determine


0.5 2

1 1.5

• All subsequent measurements will be compensated to 25°C (77°F). exact value for a particular solution, see
TEMPERATURE COEFFICIENT

If you prefer to standardize the temperature compensation to page 20).


19.99
20°C (68°F) rather than 25°C (77°F), leave the TEMPERATURE
knob at 18°C (if the temperature of the solution is 18°C), adjust • Select 19.99 mS/cm range by pressing mS

the trimmer to read "11.67 mS" (see the conductivity vs. tem- the appropriate range key.
perature chart on page 18). All subsequent measurements will be • Adjust the calibration trimmer (see page 6, #5) on the side of
compensated to 20°C. the instrument with the calibration screwdriver until the display
• The calibration is now complete and the instrument is ready for shows "12.88 mS" i.e. the conductivity reading @ 25°C.
use.
The instrument should be re-calibrated at least once a month, or
mS

when the probe or battery is changed.


Note: For more accurate results, it is advisable to use a calibration mS

solution close to the measurement range . See the accessories


section on page 23 for a wide selection of conductivity solu-
tions.
• All subsequent measurements will be compensated to 25°C (77°F).
If you prefer to standardize the temperature compensation to
20°C (68°F) rather than 25°C (77°F), adjust the trimmer to read
"11.67 mS" (see the conductivity vs. temperature chart on page
18). All subsequent measurements will be compensated to 20°C.
• The calibration is now complete and the instrument is ready for
use.

13 14
Untitled-19

15 16

The instrument should be re-calibrated at least once a month, or • Adjust the calibration trimmer (see page 7,#6) on the side of the
when the probe or battery is changed. instrument with the calibration screwdriver until the display shows
Note: For more accurate results, it is advisable to use a calibration the TDS reading at 25°C (see the TDS vs. temperature chart on
solution close to the range to be measured. See the accessories page 19). E.g. @25°C "1382 mg/L".
section on page 23 for a wide selection of conductivity solu-
tions.
PROCEDURE FOR HI 8734
0
703

• Pour sufficient quantity of a TDS calibra-


HI

tion solution (e.g. HI 7032) into a beaker


TEMPERAT
URE

to cover the holes on the probe. If pos-


sible, use plastic beakers to minimize any • All subsequent measurements will be compensated to 25°C (77°F).
EMC interference. If you prefer to standardize the temperature compensation to
• Immerse the TDS probe, making sure
°C

20°C (68°F) rather than 25°C (77°F), leave the TEMPERATURE


that holes are completely submerged, and knob to 18°C (if the temperature of the solution is 18°C). Adjust
the ChecktempC in the solution. the trimmer to read "1251 mg/L" (see the TDS vs. temperature
• Wait for a couple of minutes for thermal chart on page 19). All subsequent measurements will be compen-
equilibrium to be reached. sated to 20°C.
• Tap the probe on the bottom, then shake it while rotating to make • The calibration is now complete and the instrument is ready for
sure no air bubbles remain trapped in the sleeve. use.
• Record the temperature of the HI 7032 buffer solution from the The instrument should be re-calibrated once a month, or when the
ChecktempC (e.g. 18°C). probe or battery is changed.
• Switch the instrument on by pressing ON Note: For more accurate results, it is advisable to use a calibration
ON/OFF. OFF solution close to the range to be measured. See the accessories
section on page 23 for a wide selection of TDS solutions.

• Press TDS/TEMP to display the temperature. TDS PROCEDURE FOR HI 933000

0
703
TEMP. • Pour sufficient quantity of a conductivity

HI
calibration solution (e.g. HI 7030/
• Adjust the TEMPERATURE knob to display 18°C. HI 8030) into a beaker to cover the holes
on the probe. If possible, use plastic bea-
kers to minimize any EMC interference.
• Immerse the conductivity probe in the
TEMPERATURE solution, making sure that holes are com-
TDS
pletely submerged.
• Press TDS/TEMP again to display the TDS TEMP. • Wait for a couple of minutes for thermal
measurement. equilibrium to be reached.
• Tap the probe on the bottom, then shake it while rotating to make
• Select 1999 mg/L range by pressing the ap- 1999 sure no air bubbles remain trapped in the sleeve.
mg/l
propriate range key.

15 16
Untitled-19

17 18

ON
• Switch the instrument on by pressing ON/OFF.
CONDUCTIVITY VERSUS
OFF

TEMPERATURE CHART
The conductivity of an aqueous solution is the measure of its ability to
• Select 19.99mS/cm range by pressing the 19.99
mS carry an electrical current by means of ionic motion.
appropriate range key.
The conductivity invariably increases with increasing temperature.
It is affected by the type and number of ions in the solution and by
• Adjust the calibration trimmer in the battery compartment with a the viscosity of the solution itself. Both parameters are temperature
screwdriver until the display shows "12.88 mS" i.e. the conductiv- dependent. The dependency of conductivity on temperature is ex-
ity reading @ 25°C. pressed as a relative change per degree Celsius at a particular
temperature, commonly as percent per °C.
For manual temperature compensation, refer to the following chart:

°C °F HI 7030 HI 7031 HI 7033 HI 7034 HI 7035 HI 7039


HI 8030 HI 8031 HI 8033 HI 8034 HI 8035 HI 8039
mS (µS/cm) (µS/cm) (µS/cm) (µS/cm) (µS/cm) (µS/cm)
0 32 7150 776 64 48300 65400 2760
5 41 8220 896 65 53500 74100 3180
9 VOLT BATTERY
10 50 9330 1020 67 59600 83200 3615
15 59 10480 1147 68 65400 92500 4063
16 60.8 10720 1173 70 67200 94400 4155
17 62.6 10950 1199 71 68500 96300 4245
18 64.4 11190 1225 73 69800 98200 4337
19 66.2 11430 1251 74 71300 100200 4429
• All subsequent measurements will be compensated to 25°C (77°F). 20 68 11670 1278 76 72400 102100 4523
21 69.8 11910 1305 78 74000 104000 4617
If you prefer to standardize the temperature compensation to
22 71.6 12150 1332 79 75200 105900 4711
20°C (68°F) rather than 25°C (77°F), adjust the trimmer to read 23 73.4 12390 1359 81 76500 107900 4805
"11.67 mS" (see the conductivity vs. temperature chart on page 24 75.2 12640 1386 82 78300 109800 4902
18). All subsequent measurements will be compensated to 20°C. 25 77 12880 1413 84 80000 111800 5000
26 78.8 13130 1440 86 81300 113800 5096
• The calibration is now complete and the instrument is ready for
27 80.6 13370 1467 87 83000 115700 5190
use. 28 82.4 13620 1494 89 84900 117700 5286
The instrument should be re-calibrated once a month, or when the 29 84.2 13870 1521 90 86300 119700 5383
probe or battery is changed. 30 86 14120 1548 92 88200 121800 5479
31 87.8 14370 1575 94 90000 123900 5575
Note: For more accurate results, it is advisable to use a calibration
solution close to the range to be measured. See the accessories
For instance, the conductivity values of the calibration solutions at
section on page 23 for a wide selection of conductivity solu-
25°C are 12880 µS/cm, 1413 µS/cm or 5000 µS/cm when using
tions. HI 7030, HI 7031 or HI 7039, respectively.
At 20°C, the values are 11670 µS/cm, 1278 µS/cm or
4523 µS/cm, respectively.
With the solutions at 30°C, the values are 14120 µS/cm, 1548 µS/
cm or 5479 µS/cm, respectively.

17 18
Untitled-19

19 20

TDS VERSUS DETERMINING THE TEMPERATURE


TEMPERATURE CHART COEFFICIENT OF A SOLUTION (HI 8733)
The TDS value in aqueous solutions is directly proportional to conduc- Highly acidic, alkaline samples or solutions with high salt content
tivity. The ratio between the two parameters depends on the solution might have a different coefficient than the customary 2% per degree °C.
and usually it is set to a factor of 0.5 (corresponding to a solution of In order to calculate this coefficient follow the procedure below:
CaCO3). This means that 1 µS/cm is equal to 0.5 mg/L (ppm) of TDS. • Immerse the probe of HI 8733 in the sample and adjust the
For manual temperature compensation, refer to the following chart: TEMPERATURE COEFFICIENT knob to 0% (i.e. no compensation).

°C °F HI 7032 HI 7036 0 2.5

mg/L g/L 0.5 2

(ppm) (ppt) 1 1.5


TEMPERATURE COEFFICIENT

0 32 758 6.82
5 41 876 7.88
10 50 999 8.99 • Condition the sample and probe to 25°C and note the conductiv-
15 59 1122 10.10 ity reading, C25 .
16 60.8 1148 10.33 • Condition the sample and probe to a different temperature t°C
17 62.6 1173 10.56 (approximately 10°C different from 25°C) and note the conduc-
18 64.4 1200 10.78 tivity reading Ct.
19 66.2 1224 11.01
• The temperature coefficient ß of the solution is calculated as given
20 68 1251 11.24
21 69.8 1277 11.47
by the following formula:
22 71.6 1303 11.71 (C t - C25)
ß = 100 x
23 73.4 1329 11.94 (t - 25) x C25
24 75.2 1358 12.18
25 77 1382 12.41 The above procedure is suitable for determining the temperature
26 78.8 1408 12.65 coefficient in a laboratory or where the temperature of the solution
27 80.6 1438 12.89 can be controlled.
28 82.4 1461 13.13 If this is not possible (e.g. on-site measurements), the following
29 84.2 1476 13.37 procedure can be used providing the sample temperature varies by at
30 86 1515 13.61 least 5°C or preferably 10°C:
31 87.8 1541 13.85
• Immerse the probe of HI 8733 in the test solution and turn the
For instance, the TDS values of the calibration solutions at 25°C are TEMPERATURE COEFFICIENT knob to 0% (no compensation).
1382 mg/L or 12.41 g/L when using HI 7032 or HI 7036, respec- • Check the conductivity reading and record the value. Make sure
tively. the reading is stable, i.e. no greater variations than ±0.2 mS/
At 20°C, the values are 1251 mg/L or 11.24 g/L, respectively. cm within a minute.
With the solutions at 30°C, the values are 1515 mg/L or 13.61 g/L, • Repeat the procedure when the temperature of the test solution
respectively. has changed by at least 5°C. Wait for the conductivity reading to
stabilize.
• Adjust the TEMPERATURE COEFFICIENT knob until the display
shows the same value as recorded earlier.
• The value indicated by the knob is the temperature coefficient of
the solution.
19 20
Untitled-19

21 22

PROBE MAINTENANCE BATTERY REPLACEMENT


When the battery becomes weak the meters will display "V" or an
Rinse the probe with tap wa- additional blinking decimal point (HI 8733).
ter after every series of
measurements. If a more thor-
ough cleaning is required, V

remove the PVC sleeve and


clean the probe with a cloth or When the low battery indicator appears, the battery has only a few
a non-abrasive detergent. hours left. A low battery will result in unreliable measurements.
When reinserting the sleeve It is recommended that the battery be replaced immediately.
onto the probe, be sure that
Battery replacement must only take place in a non-hazardous area
the sleeve is in the right direc-
using an alkaline 9V battery.
tion with the four holes towards
the cable end. Slide off the battery compartment cover at the rear of the meter and
replace the 9V battery with a new one. Make sure the battery contacts
After cleaning the probe, re-
are tight and secure before replacing the cover.
calibrate the instrument.
The probe body is in PVC. For
this reason it must never come
HI 933000
into close contact with a heat
source. If the probe is exposed
to high temperatures (above 9 VOLT BATTERY

50°C/122°F), the rings might


become loose or detached, re-
sulting in a serious impairment
HI 8033
of the probe. In such cases, HI 8633
SLIDE
the probe has to be replaced. HI 8733 OFF
HI 8734

BATTERY

SLIDE
OFF

21 22
Untitled-19

23 24

CONDUCTIVITY BUFFER SOLUTIONS IN FDA APPROVED


SHOCKPROOF RUBBER BOOTS
BOTTLES
( for H I 933000 only
only)) HI 8030L 12880 µS/cm (µmho/cm), 460 mL
HI 8031L 1413 µS/cm (µmho/cm), 460 mL
These rubber boots are specially
HI 8033L 84 µS/cm (µmho/cm), 460 mL
made to prolong the life of your
HI 8034L 80000 µS/cm (µmho/cm), 460 mL
instrument and to prevent dam-
HI 8035L 111800 µS/cm (µmho/cm), 460 mL
age due to accidental bangs and
HI 8039L 5000 µS/cm (µmho/cm), 460 mL
mS

falls.
They measure 155 x 90 x 45 mm CONDUCTIVITY PROBES
CONDUCTIVITY
METER
ON
OFF

(6.1x3.5x1.8") and are available HI 76301W 4-ring conductivity/TDS probe with 1 m (3.3') screened
in two different colors: 199.9
µS
1999
µS
cable (for HI 8033, HI 8633 and HI 8734)
HI 710007 = blue
19.99 199.9

HI 710008 = orange.
mS mS

19,5mm
HI933000 DIA
0.77"

85mm
3.34"
11.5mm
DIA
0.45"

118.5mm
4.66"

ACCESSORIES HI 76302W 4-ring conductivity probe with built-in temperature


sensor and 1 m (3.3') screened cable (for
CONDUCTIVITY & TDS BUFFER SOLUTIONS HI 933000)
HI 7030L 12880 µS/cm (µmho/cm), 460mL HI 7633W 4-ring conductivity probe with built-in temperature
HI 7030M 12880 µS/cm (µmho/cm), 230mL sensor and 1 m (3.3') screened cable (for HI 8733)
HI 7031L 1413 µS/cm (µmho/cm), 460mL
HI 7031M 1413 µS/cm (µmho/cm), 230mL 19,5mm
0.77"
DIA

HI 7033L 84 µS/cm (µmho/cm), 460 mL 85mm


HI 7033M 84 µS/cm (µmho/cm), 230 mL 3.34"
11.5mm
DIA
HI 7034L 80000 µS/cm (µmho/cm), 460mL
0.45"

HI 7034M 80000 µS/cm (µmho/cm), 230mL


HI 7035L 111800 µS/cm (µmho/cm), 460mL 118.5mm
4.66"

HI 7035M 111800 µS/cm (µmho/cm), 230mL


HI 7039L 5000 µS/cm (µmho/cm), 460mL
HI 7039M 5000 µS/cm (µmho/cm), 230mL OTHER ACCESSORIES
HI 7032L 1382 ppm (mg/L), 460 mL HI 710001 Soft carrying case for HI 8033, HI 8633, HI 8733,
HI 7032M 1382 ppm (mg/L), , 230 mL HI 8734 and HI 933000 measuring 230x100x50 mm
HI 7036L 12.41 ppt (g/L), 460 mL CHECKTEMPC Electronic thermometer (range: -50.0 to 150.0°C)
HI 7036M 12.41 ppt (g/L), 230 mL HI 721313 Rugged carrying case for HI 8733 and HI 8633 measur-
ing 340 x 230 x 90 mm complete with conductivity
calibration solution (HI 7030S, 120 mL) and screwdriver

23 24
Untitled-19

25 26

HI 721314 Rugged carrying case for HI 8734 measuring


340x230x90 mm complete with TDS calibration solu- CE DECLARATION OF CONFORMITY
tion (HI 7032S, 120 mL) and screwdriver
HI 731326 20 small screwdrivers, length 90 mm, for calibration
purposes (except for HI 8033)
HI 710007 Blue shockproof rubber boot for HI 933000
HI 710008 Orange shockproof rubber boot for HI 933000
HI 710009 Blue rubber boot for HI 8033, HI 8633, HI 8733 and
HI 8734
HI 710010 Orange rubber boot for HI 8033, HI 8633, HI 8733
and HI 8734
MANCONDR2 Instruction manual

WARRANTY

All Hanna Instruments meters are warranted for two years against
defects in workmanship and materials when used for their intended
purpose and maintained according to instructions. The electrodes
and the probes are warranted for a period of six months. This
warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased
the instrument. If under warranty, report the model number, date of
purchase, serial number and the nature of the failure. If the repair is
not covered by the warranty, you will be notified of the charges
incurred. If the instrument is to be returned to Hanna Instruments,
first obtain a Returned Goods Authorization number from the Cus-
tomer Service department and then send it with shipping costs prepaid.
When shipping any instrument, make sure it is properly packaged for
Recommendations for Users
complete protection. Before using these products, make sure that they are entirely suitable for the environment in which they are used.

To validate your warranty, fill out and return the enclosed warranty Operation of these instruments in residential area could cause unacceptable interference to radio and TV equipment, requiring the operator to take
all necessary steps to correct interferences.
card within 14 days from the date of purchase. The metal band at the end of the probe is sensitive to electrostatic discharges. Avoid touching this metal band at all times.
During calibration of the instruments, ESD wrist straps should be worn to avoid possible damage to the probe by electrostatic discharge.
All rights are reserved. Reproduction in whole or in part is prohibited Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance.

without the written consent of the copyright owner, To avoid electrical shock, do not use these instruments when voltages at the measurement surface exceed 24VAC or 60VDC.

Hanna Instruments Inc., 584 Park East Drive, Woonsocket, Rhode Use plastic beakers to minimize any EMC interferences.
To avoid damage or burns, do not perform any measurement in microwave ovens.
Island, 02895 , USA.
Hanna Instruments reserves the right to modify the design, construc-
tion and appearance of its products without advance notice.

25 26
Untitled-19

27 28

OTHER PRODUCTS FROM HANNA HANNA LITERATURE

• CALIBRATION AND MAINTENANCE SOLUTIONS


• CHEMICAL TEST KITS
• CHLORINE METERS
• DISSOLVED OXYGEN METERS
• HYGROMETERS
• ION SPECIFIC METERS (Colorimeters)
• MAGNETIC STIRRERS
• Na/NaCl METERS
• pH/ORP/Na ELECTRODES
• pH/ORP METERS LAB RECORDING WATER ANALYSIS
• PROBES (DO, µS/cm, RH, T, TDS)
• PUMPS
• REAGENTS
• SOFTWARE
• THERMOMETERS
• TITRATORS
• TRANSMITTERS
• TURBIDITY METERS
• Wide Range of Accessories
ENVIROCARE GENERAL CATALOG
Most Hanna meters are available in the following formats:
• BENCH-TOP METERS

PRINTED IN ITALY
These and many others catalogs, handbooks and leaflets are avail-
• POCKET-SIZED METERS able from Hanna. To receive your free copy, contact your dealer or the
• PORTABLE METERS nearest Hanna Customer Service Center.
• PRINTING/LOGGING METERS
• PROCESS METERS (Panel and Wall-mounted)
MANCONDR2
• WATERPROOF METERS 06/97

• METERS FOR FOOD INDUSTRY

For additional information, contact your dealer or the nearest Hanna


Customer Service Center. You can also e-mail us at:
[email protected].

h t t p : / / w w w . h a n n a i n s t . c o m

27

You might also like