Automated Cell Counts HB Estimation Methods
Automated Cell Counts HB Estimation Methods
Automated blood cell counters have come to stay in a modern Haematology laboratory because
they provide data with increased reliability, precision and accuracy.
The Automated blood cell counters available today are based on two principles
A-Electrical Impedance
B-Light Scatter
Electrical Impedence
Cells passing through an aperture through which a current is flowing cause changes in electrical
resistance that are counted as voltage pulses. This principle is illustrated in Figure 5.4.
Fig. 5.4: Schematic diagram of particle counter in which changes in electrical resistance are
counted as voltage pulses. CS = cell suspension,
GC = glass cylinder, A = aperture, E1 and E2 = platinum electrodes, V = valve, M =
mercury column, EC1 and EC2 = electrical contacts VP = vacuum pump.
Each cell that passes through the aperture displaces an equal volume of conductive fluid,
increasing the electrical resistance and creating the voltage pulse, because its resistance is much
greater than that of the conductive solution. The pulses, which are proportional in height to the
volume of the cells are counted. This is the “Coulter principle”.
In the simplest system, the counting mechanism is started when the mercury contacts EC1 and
stopped when it contacts EC2; during this time the cells are counted in a volume of suspension
exactly equal to the volume of glass tubing between contact wires EC1 and EC2. If two or more
cells enter the aperture simultaneously, they will be counted as one pulse; this produces a
coincidence error that must be corrected. The size of the coincidence error can be diminished by
decreasing the concentration of cells and decreasing the size of the aperture.
Light Scatter
In the electro-optical analysers (Fig. 5.5), a light sensitive detector measures light scattering.
Light is focused on the flow cell. Only light scattered by a cell reaches the photomultiplier tube
(PMT), which converts it to an electrical pulse. The size of the pulse detected is proportional to
the size of the particle.
Fig. 5.5: Schematic diagram of the electro-optical cell counter. Light is focused on the flow
cell. Only light scattered by a cell reaches the photomultiplier tube (PMT), which converts
it to an electrical pulse.
The erythrocyte / platelet channel determines RBC and Platelet counts by the analysis of light
scattering measurements obtained as the diluted cells pass singly through a helium-neon laser
beam. The RBCs are counted and sized by both high-angle and low-angle light scattering
measurements and the MCV is thus determined.
Within the hemoglobin channel, a portion of EDTA anticoagulated blood is mixed with the
hemoglobin diluent. The RBCs are lysed and free hemoglobin is converted to
cyanmethemoglobin. The concentration of hemoglobin is determined photometrically at 546 nm.
The leucocyte count and five – part differential count are obtained using two different methods
and two separate channels, the peroxidase channel and the basophil / lobularity channel. In the
peroxidase channel, neutrophils, monocytes, and eosinophils are identified by the degree of
peroxidase positivity and the amount of forward light scatter. Lymphocytes and the large
unstained cells (LUCs) are identified by the amount of forward light scatter and the fact that they
remain unstained by this peroxidase cytochemical staining method. RBCs are removed before
peroxidase staining by lytic action. The amount of forward scatter and degree of positivity are
detected as the cells pass through a tungsten - based flow cell. Within the basophil / lobularity
channel, a fourth portion of EDTA – anticoagulated blood is mixed with basophil diluent. The
basophil diluent lyses RBCs and platelets and strips all WBCs except basophils of their
cytoplasm. This dilution is measured by the helium – neon laser flow cell. Basophils will have
large, low angle scatter and the remaining cell nuclei will be classified as mononuclear or
polymorphonuclear based on their high angle scatter.
BLOOD FILM MORPHOLOGICAL EXAMINATION
Erythrocyte Morphology:
(1) Study the erythrocytes and report any evidence of rouleaux formation or
signs of immaturity.
(2) Report the erythrocyte morphology with reference to size, shape,
staining characteristics, and inclusions. Report the degree of the specific morphological
characteristic (for example, moderate hypochromia).
(3) If nucleated erythrocytes are found (usually these are megaloblasts),
report the number per 100 leukocytes counted.
Normal Haemoglobins
Abnormal Haemoglobins
These are formed from the abnormal polypepetide chain synthesis of the haemoglobin molecular
structure. They result from either switches, deletions or spontaneous mutations of the amino acid
sequences making up the polypeptide chain. Consequently, they result to abnormal hemoglobin
structures and hence a deranged haemoglobin molecule, which is impaired in its function, a term
to describe them is haemoglobinopathies. The type of Hb resulting from this complications
include:
Hb S, Hb Barts, Hb Heinz, Hb Thellasaemia
a) Calorimetric Methods
These are based on measuring the color of haemoglobin or a derivative of haemoglobin in blood.
The estimation is based on Beer’s law i.e. the optical density (OD) of a colored solution is
directly proportional to the concentration of the colored material in the solution.
iii-Photoelectric method
1) Cyanmethemoglobin method
2) Oxyhemoglobin method
Principle
Haemoglobin is converted to acid hematin by N/10 HCl, the resulting brown color is
compared with standard brown glass.
Instruments
Sahli’s Hemoglobinometer – the main parts of which are a graduated glass tube, color
comparators, glass stirrer and Sahli’s pipette to measure 20 cu mm of blood.
The tubes commonly used are square with graduations in percent on one side and grams
per 100 ml on the other.
The color comparators are made of brown colored glass
Requirements
Procedure
Sources of Error
Technical errors – Most of the errors described under Cyanmeth Hb method discussed later
also occur in this method.
o E.g. Improper mixing of blood, errors in pipetting, tissue fluid contaminating
capillary blood.
Visual errors – Taking the reading is very subjective.
Quality of the color comparators can affect the reading
Insufficient time allowed for the conversion of Hb to acid hematin.
Other forms of hemoglobin: Carboxyhaemoglobin, methaemoglobin and
sulphhemoglobin are not converted to acid hematin.
Time delay – The brown colour of acid hematin is not stable, so undue delay in reading
the test result is not allowed.
In this method, the Hb is converted to alkali hematin by the addition of a strong alkali (N/10
NaOH).
iii-Photoelectric method
1- Cyanmethaemoglobin Method
This is the prefered and the most accurate method for determining the maemoglobin
concentration. It is the standard method used in most of the countries.
In this method, Blood is diluted in a solution of potassium ferricyanide and potassium cyanide.
The ferricyanide oxidizes haemoglobin to hemiglobin (Hi,) or methaemoglobin. Potassium
Cyanide provides cyanide ions (CN–) to form hemiglobincyanide (HiCN) or Cyanmethaemo-
globin. The absorbance of the solution is then measured in a spectrophotometer at a wavelength
of 540 nm.
Haemoglobin being the largest single constituent, affects the specific gravity of blood more than
other substances. Serum proteins are the next heaviest constituents of blood. It is assumed that
(which is not always true) the level of serum proteins and other smaller constituents remain the
same, so any change in the specific gravity of blood is mainly due to change in concentration of
haemoglobin.
Allow a drop of blood to fall into a series of solutions of copper sulphate of varying specific
gravity and note the behaviour of the drop and estimate the specific gravity of blood. If the drop
sinks to the bottom, its specific gravity is more than that of the copper sulphate solution. If the
drop rises after its initial fall then its specific gravity is less than that of the copper sulphate
solution.
The principle is based on the fact that each molecule of hemoglobin contains 4 atoms of iron or
0.347 grams of iron per 100 grams of hemoglobin. The iron present is detached from the
hemoglobin and measured. The hemoglobin is calculated by using the formula.
Hemoglobin (gm/dl) =
Though this is very accurate, it is a complex and very time-consuming test. So it is not done
as a routine procedure. It is used only as a reference method i.e. a method used to check the
accuracy of the other methods.
This is a reference method as it is very accurate, however it is not used for routine laboratory
work. The principle is based on the fact that one molecule of O2 binds to each iron atom. So one
molecule of haemoglobin binds 4 molecules of oxygen. Thus oxygen combining capacity thus
indirectly measures the amount of Hb.
It is estimated that 1 gram of haemoglobin binds about 1.34 ml of oxygen. From this the
haemoglobin concentration is calculated by using the following formula
Hb in gm/dl =
1-Improper sampling
2-Method used
Sahli’s method has an error of 5% to 20%, CUSO4 method is inaccurate while the
cyanmethhemoglobin method is the most accurate.
3-Equipment / Apparatus
The pipette, cuvette and photometer need proper calibration and appropriate checking
periodically to eliminate the errors due to them.
4-Human errors
Human errors can be reduced by good training, understanding the clinical significance of the test
and the necessity for a dependable method, adherence to oral and written instructions and
familiarity with the equipment and with sources of errors. Automated instruments are widely
used now-a-days and this eliminates most of these errors.
Quality Control
The important aspect of quality control is to identify those steps in which the likelihood of error
is high and to consider ways to minimize that likelihood. Some of the measures followed are:
1. Duplicating samples.
2. Haemolysate of known value are run with batches of tests.
3. Haemoglobin values are compared with other values.
Normal Values
The normal value depends on the age and sex of the individual