FAQs On Particle Counting
FAQs On Particle Counting
by Sheesh Gulati
Secretary, ICCCS
Chairman, CCSI
FLOW RATES
• Airborne particle counters are available with
flow rates of 0.1cfm,1.0 cfm, 50 lpm or 100
lpm.
1
24-03-2019
2
24-03-2019
Smoke Studies
(Impact of exhausted air)
Particle Counter Off
Photo shows horizontal
laminar flow across table
Particle Counter On
Photo shows disruption of
laminar flow near the
particle counter exhaust
The greater the flow rate of sampled
air, the greater the amount of
Exhaust Air
3
24-03-2019
FAQs ON PARTICLE
COUNTING
What is the Minimum
Sample Volume to be
collected?
Clause A.4.4: “At each sample location, sample a sufficient volume of air that a
minimum of 20 particles would be detected if the particle concentration for the
largest considered particle were at the class limit for the designated ISO Class”
Vs = 20_ X 1000
Cn,m
Vs is the minimum Sample Volume Per Location (in Litres)
Cn,m is the class limit (particles / M3 ) for the largest considered particle size specified for
the relevant class
20 is the defined number of particles that could be counted if the particle concentration
were at the class limit
20 particles
Volume = X 1000 litres
3520 particles
4
24-03-2019
EU Annex 1 : Monitoring
Section 12:
“The sample sizes taken for monitoring purposes using
automated systems will usually be a function of the
sampling rate of the system used. It is not necessary for
the sample volume to be the same as that used for formal
classification of clean rooms and clean air devices.”
5
24-03-2019
FAQs ON PARTICLE
COUNTING
What is the Maximum
Permissible Sampling
Tube Length?
• Flow rate should be fairly high i.e. 1 cfm not 0.1 cfm.
The higher the flow rate, the lesser the number of larger
particles that will be "lost"
• Sample tubing length should be minimised, although
tubing cannot be totally eliminated.
• In the case of portable particle counters, the tubing
length should not be more than 3 metres i.e. 10 ft
and
6
24-03-2019
7
24-03-2019
Probe With
Curvature of 15
CM Radius
FAQs ON PARTICLE
COUNTING
How Often Should we
Classify an ISO Class 5
cleanroom?
8
24-03-2019
FREQUENCY OF CLASSIFICATION OF
CLEANROOMS
• At-rest and operational classification should performed
periodically based upon risk assessment of the cleanroom
and clean zone operations, typically on an annual basis
• Where the cleanroom and clean zone is equipped with
instrumentation for continuous or frequent monitoring of air
cleanliness (airborne particles, room pressure differentials),
the time intervals between classification may be extended
provided that the results of the monitoring remain within the
specified limits. However, in the pharmaceutical and
related industries formal classification / re-qualification
must be undertaken at least annually
• Frequency of re-testing cleanroom no longer dictated by
Class, to be done annually. Previously it was 6 months for
ISO Class 5
FAQs ON PARTICLE
COUNTING
How Should we
conduct the Recovery
Test?
•Test should be performed when the clean room is in the ‘As Built’
or ‘At Rest’ states, or after major modification to the
cleanroom, or its operation.
•It is not recommended that the 100:1 test be used for ISO 8 & 9
9
24-03-2019
RECOVERY TEST
. Where to test:
•Working level
•At places where particle concentration indicates further
investigation
•At product / process level
Do not test at:
•Places where there is Direct influence of air supplies
•Near air returns
10
24-03-2019
11