Cleaning and Maintenance of FCI Instruments Final
Cleaning and Maintenance of FCI Instruments Final
Generalities
- This document is intended to provide instructions for handling, decontaminating, cleaning and sterilizing FCI reusable surgical instruments,
unless different instructions are provided with the instrument.
- The equipment, personnel, cleaning/decontaminating agents and procedures all participate in the efficiency of the process. It remains however
the user’s responsibility to make sure the process has produced the expected results. This requires a validation and routine surveillance of the
procedures. All the cleaning and sterilization processes need to be validated. However, we can only provide general advice for adequate
cleaning and sterilization of reusable instruments.
Warning
- Unless otherwise indicated, instruments are delivered non-sterile by FCI and should not be used without being cleaned, disinfected and
sterilized.
- New instruments must be cleaned and sterilized before initial use.
- Instruments should not be soaked in sodium hypochlorite or any kind of solution containing chlorine or chloride, as they promote corrosion.
- Only use cleaning and decontaminating solutions adequate for use on surgical instruments. Comply with the soaking time, temperature, water
quality and concentration recommended in the manufacturers’ instruction for use
- Never use metallic brushes or scouring pads
- Instruments made of different metal or metallic alloys should be processed separately to prevent corrosion by contact.
Instructions
Point of use
- Instruments should be cleaned immediately after use, to prevent blood or any other organic substance from drying on surface or in cavities.
- After each use, you should:
1- Remove all contaminants (antiseptic products, biological material ...) by wiping the instruments with a lint-free cloth,
2- Rinse the reusable instruments carefully with demineralized or distilled water,
3- Flush all lumen instruments (such as cannulas, suction pipes, hand pieces ...) several times with demineralized or distilled water to remove
any particles that could prevent the liquid from entering the instruments.
1- When needed, pull apart cannulated instruments and instruments with removable nozzle (mainly microsurgical instruments).
2- Prepare the cleaning solution as recommended by the supplier. It is possible to use a combined solution of decontamination and cleaning
product. Comply with the instructions for use.
3- In the decontamination bath, instruments should be soaked in open position for the length of time recommended by the supplier.
4- Flush lumen instruments (cannulas, suction pipes, hand pieces, removable nozzle …) several times with the cleaning solution.
5- Brush the instruments with a soft brush while maintaining them within the cleaning solution. Pay special attention to rough surfaces, lumens,
joints, blind holes and moving parts.
6- Use a swab of the correct diameter and length for cleaning hollow instruments (such as cannulas, suction pipes, hand pieces, instruments with a
blind hole ...).
7- Rinse instruments with demineralized or distilled water until full removal of the cleaning solution.
8- Lumen instruments (such as cannulas, suction pipes, hand pieces, instruments with a blind hole ...) and instruments with moving pieces should
be flushed / rinsed with demineralized or distilled water until full removal of the cleaning solution. Then, flush /rinse the instruments with 100%
alcohol to remove any trace of water and flush/dry them with air.
9- Cleaning cycle in ultrasonic tank (if possible): manual cleaning cycle can then be completed by an ultrasonic cleaning cycle of 5 to 10 minutes
with a fresh cleaning solution. At the end of the cycle, rinsing steps « 7 »and « 8 » must be repeated.
Refer to section « Ultrasonic Cleaning Procedure» for ultrasonic cleaning tank instructions for use.
10- Cleaning and rinsing operations should be repeated until full removal of any visible soiling.
11- Carefully dry the instruments with a medical fabric, hot air or compressed micro-filtered air.
2138 311 86/A – September 2011
Maintenance
- The instruments joints, hinges and moving pieces should be regularly lubricated (in particular after an ultrasonic bath) with a lubricant meant
for medical grade surgical instruments, to reduce friction and wear.
- Complying with the lubricant supplier’s instructions is highly recommended. The lubricant used must be compatible with the chosen
sterilization method.
Packing
- If instruments are sterilized individually, the use of specialized bags is recommended. The bags should be big enough for the instruments to be
placed in an open position. Instruments should NEVER be sterilized in a closed position.
- If the instruments are sterilized together or by batch, the use of the appropriate size of sterilization racks with silicone pads is recommended.
Instruments should not come into contact with one another. The use of protection nozzles made of soft silicone tubing, of an appropriate size
and thickness is recommended. Do not use protection nozzles made of rubber or plastic, as they could melt during autoclaving and damage the
instruments.
Sterilization Cycles
FCI instruments may be sterilized using the following methods:
Ethylene Oxide:
Gas concentration: 850±50mg/l Temperature: 37°C – 47°C
Exposure time: 3- 4 hours Relative humidity: 70% HR minimum
2138 311 86/A – September 2011
Stan Autoclaving:
Type of autoclave Gravity displacement Prevacuum
Configuration Wrapped Wrapped
Temperature 121°C to 123°C 132°C to 135°C
250°F to 253°F 270°F to 275°F
Exposure Time 15 to 30 minutes 3 to 4 minutes
“Flash” Autoclaving:
Type of autoclave Gravity displacement Prevacuum
Configuration Unwrapped Unwrapped
Temperature 132°C / 270°F 132°C / 270°F
Exposure Time 3 minutes 3 minutes
Notes:
- Sterilization cycles described above correspond to standard procedures and should produce a sterile device. Because of differences
between sterilization equipments and in the microbial charge of devices when used clinically, FCI cannot recommend specific cycle
parameters. It is each user’s responsibility to control and validate that the sterilization cycle will provide the right level of guarantee of an
appropriate sterilization of the instruments.
- FCI does not recommend “flash” autoclaving for reusable instruments. “Flash” autoclaving should only be reserved for emergency
reprocessing, i.e. in case of a problem of asepsis.