SmartLine

Technical Information

STG800 SmartLine Gauge Pressure


Specification 34-ST-03-83, March 2018

Introduction

Part of the SmartLine® family of products, the STG800 and


STG80L are high performance gauge pressure transmitter
featuring piezoresistive sensor technology combining
pressure sensing with on chip temperature compensation
capabilities providing high accuracy, stability and
performance over a wide range of application pressures and
temperatures. The SmartLine family is also fully tested and
compliant with Experion® PKS providing the highest level of
compatibility assurance and integration capabilities.
SmartLine easily meets the most demanding application
needs for pressure measurement applications.

Best in Class Features:

o Accuracy up to 0.055 % of calibrated span & 0.025% Opt.

o Stability up to 0.015% of URL per year for ten years

o Automatic temperature compensation


Figure 1 – STG800 Gauge Pressure Transmitters feature
o Rangeability up to 100:1 field-proven piezoresistive sensor technology
o Response times as fast as 80ms
Communications/Output Options:
o Multiple local display capabilities
o 4-20mA dc
o External zero, span, & configuration capability o Honeywell Digitally Enhanced (DE)

o Polarity insensitive electrical connections o HART ® (version 7.0)


o FOUNDATION™ Fieldbus
o Comprehensive on-board diagnostic capabilities
All transmitters are available with the above listed
o Integral Dual Seal design for safety based on communications protocols.
ANSI/NFPA 70-202 and ANSI/ISA 12.27.0 Span & Range Limits:
o Full compliance to SIL 2/3 requirements as a standard. URL/Max
LRL Min Turn
Model Span
o Modular design characteristics psi (bar) Span down
psi (bar)
o Available with 15 year warranty STG830/STG83L 50 (3.5) -14.7 (-1.0) 0.5 (.35) 100:1
STG840/STG84L 500 (35) -14.7 (-1.0) 5 (.35) 100:1
o Plugged Impulse Line Detection Option
STG870/STG87L 3000 (210) -14.7 (-1.0) 30 (2.1) 100:1
o Dual/Triple Calibration Option (HART & Fieldbus Only) STG88L 6000 (420) -14.7 (-1.0) 60 (4.2) 100:1
STG89L 10000 (690) -14.7 (-1.0) 100 (6.9) 100:1
2 STG800 Smart Pressure Transmitter

Description Configuration Tools

The SmartLine family of gauge pressure, differential Integral Three Button Configuration Option
pressure, and absolute pressure transmitters is designed
Suitable for all electrical and environmental requirements,
around a high performance piezo-resistive sensor. This one
SmartLine offer the ability to configure the transmitter and
sensor actually integrates multiple sensors linking process
display via three externally accessible buttons when either
pressure measurement with on-board static pressure (DP
display option is selected. Zero/span capabilities are also
Models) and temperature compensation measurements
optionally available via these buttons with or without selection
resulting in the best total performance available. This level of
of a display option.
performance allows the ST 800 to replace virtually any
competitive transmitter available today. Hand Held Configuration

Unique Indication/Display Options SmartLine transmitters feature two-way communication and


configuration capability between the operator and the
The ST 800 modular design accommodates a basic
transmitter. This is accomplished via Honeywell’s field-rated
alphanumeric LCD display or a unique advanced graphics
Multiple Communication Configurator (MCT404).
LCD display with many unparalleled features.
The MCT404 is capable of field configuring DE and HART
Basic Alphanumeric LCD Display Features Devices and can also be ordered for use in intrinsically safe
environments. All Honeywell transmitters are designed and
o Modular (may be added or removed in the field)
tested for compliance with the offered communication protocols
o 0, 90,180, & 270 degree position adjustments and are designed to operate with any properly validated hand
o Configurable (HART only) and standard (Pa, KPa, MPa, held configuration device.
KGcm2, Torr, ATM, inH2O, mH2O, bar, mbar, inH2O,
Personal Computer Configuration
inHG, FTH2O, mmH2O, mm HG, & psi) measurement
units Honeywell’s SCT 3000 Configuration Toolkit provides an easy
o 2 Lines 16 Characters (4.13H x 1.83W mm) way to configure Digitally Enhanced (DE) instruments using a
personal computer as the configuration interface. Field Device
o Square root output indication (√)
Manager (FDM) Software and FDM Express are also available
Advanced Graphics LCD Display Features for managing HART & Fieldbus device configurations.

o Modular (may be added or removed in the field) System Integration


o 0, 90, 180, & 270 degree position adjustments
o SmartLine communications protocols all meet the most
o Standard and custom measurement units available. current published standards for HART/DE/Fieldbus.
o Up to eight display screens with 3 formats are possible
o Integration with Honeywell’s Experion PKS offers the
(Large PV with Bar Graph or PV with Trend Graph) following unique advantages.
o Configurable screen rotation timing (1 to 30 sec)
o Transmitter messaging
o Display Square Root capabilities may be set separately
from the 4-20mA dc output signal o Maintenance mode indication

o Unique “Health Watch” indication provides instant o Tamper reporting


visibility of diagnostics
o FDM Plant Area Views with Health summaries
o Multiple language capability. (EN, DE, FR, IT, ES, RU,
TR, CN, JP) o All ST 800 units are Experion tested to provide the
highest level of compatibility assurance
Diagnostics

SmartLine transmitters all offer digitally accessible


diagnostics which aid in providing advanced warning of
possible failure events minimizing unplanned shutdowns,
providing lower overall operational costs
STG800 Smart Pressure Transmitter 3

Plugged Impulse Line Detection:


Modular Design STG800 models are offered with a PILD option which
To help contain maintenance & inventory costs, all ST 800 provides indication of a plugged impulse line or process
transmitters are modular in design supporting the user’s ability connection. When used in conjunction with a basic or
to replace meter bodies, add indicators or change electronic advanced display, a non-critical diagnostic indication appears
modules without affecting overall performance or approval body on the integral display. For units without an integral display,
certifications. Each meter body is uniquely characterized to an indication can be seen via the host or hand held device
provide in-tolerance performance over a wide range of when HART Protocol is utilized.
application variations in temperature and pressure and due to
the Honeywell advanced interface, electronic modules may be
swapped with any electronics module without losing in- Dual/Triple Calibration:
tolerance performance characteristics. STG800 models are optionally offered with multiple
calibrations. In lieu of a standard factory calibration, units
Modular Features can be supplied with 1, 2, or 3 customer specified
o Meter body replacement calibrations. These calibrations are stored in the meter body
o Exchange/replace electronics/comms modules* and provide users with factory calibrated performance at up
o Add or remove integral indicators* to three different calibrated ranges. This increases
o Add or remove lightning protection (terminal connection)* application flexibility without requiring any costly recalibration
* Field replaceable in all electrical environments (including IS) or additional inventory.
except flameproof without violating agency approvals.
With no performance effects, Honeywell’s unique modularity
results in lower inventory needs and lower overall operating
costs.
4 STG800 Smart Pressure Transmitter

Performance Specifications
Reference Accuracy: (conformance to +/-3 Sigma)
Reference
1,2
Maximum Accuracy
Turndown Stability (% Span)
Model URL LRL Min Span Ratio (% URL/Year) Std / Opt.
STG830 50 psi (3.5 bar) -14.7 psi (-1.0 bar) 0.5 psi (.035 bar)
STG83L 50 psi (3.5 bar) -14.7 psi (-1.0 bar) 0.5 psi (.035 bar)
STG840 500 psi (35 bar) -14.7 psi (-1.0 bar) 5 psi (.35 bar)
0.055 /
STG84L 500 psi (35 bar) -14.7 psi (-1.0 bar) 5 psi (.35 bar)
0.025%
STG870 3000 psi (210 bar) -14.7 psi (-1.0 bar) 30 psi (2.1 bar) 100:1 0.010
STG87L 3000 psi (210 bar) -14.7 psi (-1.0 bar) 30 psi (2.1 bar)
STG88L 6000 psi (420 bar) -14.7 psi (-1.0 bar) 60 psi (4.2 bar)
0.055 /
STG89L 10000 psi (690 bar) -14.7 psi (-1.0 bar) 100 psi (6.9 bar) 0.04%
Zero and span may be set anywhere within the listed (URL/LRL) range limits

Accuracy at Specified Span and Temperature: (Conformance to +/-3 Sigma)


Combined Zero & Span
Accuracy1,2,3 Temperature Effect
(% of Span)
(% Span/50o F)

For C
Turndowns A B D E
psi (bar)
Model URL Greater Then
STG830 50 psi (3.5 bar) 50:1 1 (0.07) 0.003
0.030
Standard Accuracy

STG83L 50 psi (3.5 bar) 16:1 3 (0.21) 0.006


STG840 500 psi (35 bar) 35:1 14.5 (1.0) 0.004
STG84L 500 psi (35 bar) 35:1 14.5 (1.0) 0.007
0.015 0.04
STG870 3000 psi (210 bar) 10:1 300 (20.7) 0.005
0.025
STG87L 3000 psi (210 bar) 10:1 300 (20.7)
STG88L 6000 psi (420 bar) 12:1 500 (34.5) 0.010
STG89L 10000 psi (690 bar) 10:1 1000 (69)
0.015 0.01 0.003
High Accuracy Option

STG830 50 psi (3.5 bar) 50:1 1 (0.07) 0.030


STG83L 50 psi (3.5 bar) 16:1 0.010 0.015 3 (0.21) 0.030 0.006
STG840 500 psi (35 bar) 35:1 14.5 (1.0) 0.004
STG84L 500 psi (35 bar) 35:1 14.5 (1.0) 0.007
STG870 3000 psi (210 bar) 10:1 0.01 300 (20.7) 0.005
0.015 0.025
STG87L 3000 psi (210 bar) 10:1 300 (20.7)
STG88L 6000 psi (420 bar) 12:1 500 (34.5) 0.001
STG89L 10000 psi (690 bar) 10:1 0.025 1000 (69)
Turn Down Effect Temp Effect
  C    URL  
 A  B   D E  
  Span    Span  
% Span % Span per 28oC (50oF)
Total Performance (% of Span):

Total Performance Calculation: = +/- √ (Accuracy)2 + (Temperature Effect)2


Standard Accuracy Total Performance Examples (for comparison): @ 5:1 Turndown, +/-50 oF (28oC) shift
STG830 @ 10 psi 0.071% of span STG840 @ 100 psi 0.071% of span STG870 @ 600 psi 0.074 % of span
STG83L @ 10 psi 0.081% of span STG84L @ 100 psi 0.081% of span STG87L @ 600 psi 0.093% of span
STG88L @ 1200 psi 0.093% of span STG89L @ 2000 psi 0.093% of span
Typical Calibration Frequency:
Calibration verification is recommended every four (4) years
Notes:
1. Terminal Based Accuracy - Includes combined effects of linearity, hysteresis, and repeatability. Analog output adds 0 .005% of span.
2. For zero based spans and reference conditions of: 25 oC (77oF) for LRV>= 0 psia, 10 to 55% RH, and 316 Stainless Steel barrier diaphragm.
3. STG830 and STG83L, for LRV<0 and / or URV<0, B = 1.25 and E = 0.25.
STG800 Smart Pressure Transmitter 5

Operating Conditions – All Models


Reference Rated Condition Operative Limits Transportation and
Parameter
Condition Storage
C F C F C F C F
Ambient Temperature1 25±1 77±2 -40 to 85 -40 to 185 -40 to 85 -40 to 185 -55 to 120 -67 to 248
Meter Body Temperature2 25±1 77±2 -40 to 1101 -40 to 2301 -40 to 125 -40 to 257 -55 to 120 -67 to 248
Humidity %RH 10 to 55 0 to 100 0 to 100 0 to 100
Vac. Region – Min. Pressure
mmHg absolute Atmospheric 25 2 (short term) 3
inH2O absolute Atmospheric 13 1 (short term) 3

Supply Voltage 10.8 to 42.4 Vdc at terminals


Load Resistance 0 to 1,440 ohms (as shown in Figure 2)
Maximum Allowable
Working Pressure
(MAWP)4, 5 STG830: 50 psi (3.5 bar) STG83L: 50 psi (3.5 bar)
(Maximum allowable working
STG840: 500 psi (35 bar) STG84L: 500 psi (35 bar)
pressure is the maximum pressure STG870: 3000 psi (210 bar ) STG87L: 3000 psi (210 bar)
allowed for the normal transmitter STG88L: 6000 psi (420 bar)
operation. MAWP depends on
Approval agencies and transmitter STG89L: 10000 psi (690 bar)
material of construction. MAWP are
same as URL for gauge and
absolute pressure transmitters)

Over pressure STG830: 75 psi (5.25 bar) STG83L: 75 psi (5.25 bar)
These are the pressure limits
the transmitters can STG840: 1500 psi (103 bar) STG84L: 1500 psi (103 bar)
withstand without damage STG870: 4500 psi (310 bar) STG87L: 4500 psi (310 bar)
STG88L: 9000 psi (620 bar)
STG89L: 15000 psi (1035 bar)
1
LCD Display operating temperature -20C to +70C . Storage temperature -30C to 80C.
2
Silicone 704 minimum temperature rating is 0oC (32oF). NEOBEE M-20 minimum temperature rating is -15oC (5oF)
NEOBEE is a registered trademark of Stepan Company
3
. Short term equals 2 hours at 70C (158F)
4.
Units can withstand overpressure of 1.5 x MAWP without damage
5
Consult factory for MAWP of ST 800 transmitters with CRN approval
6 STG800 Smart Pressure Transmitter

Figure 2 - Supply voltage and loop resistance chart & calculations

Performance Under Rated Conditions – All Models


Parameter Description
Analog Output Two-wire, 4 to 20 mA (HART & DE Transmitters only)
Digital Communications: Honeywell DE, HART 7 protocol or FOUNDATION Fieldbus ITK 6.0.1 compliant
All transmitters, irrespective of protocol have polarity insensitive connection.
HART & DE Output Failure Modes Honeywell Standard: NAMUR NE 43 Compliance:
(NAMUR for DE Units requires Normal Limits: 3.8 – 20.8 mA 3.8 – 20.5 mA
selecting display and configuration Failure Mode: ≤ 3.6 mA and ≥ 21.0 mA ≤ 3.6 mA and ≥ 21.0 mA
buttons or factory configuration)
Supply Voltage Effect 0.005% span per volt.

Transmitter Turn on Time HART or DE: 2.5 sec


(includes power up & test algorithms)
Foundation Fieldbus: Host dependent
Response Time DE/HART Protocol FOUNDATION Fieldbus
(delay + time constant)
80ms 150ms (Host Dependent)
Damping Time Constant HART: Adjustable from 0 to 32 seconds in 0.1 increments. Default Value: 0.5 seconds
DE: Discrete values 0, 0.16, 0.32, 0.48, 1, 2, 4, 8, 16, 32 seconds. Default Value: 0.48 seconds
Vibration Effect: Less than +/- 0.1% of URL w/o damping
Per IEC60770-1 field or pipeline, high vibration level (10-2000Hz: 0.21 displacement/3g max
acceleration)
Electromagnetic Compatibility Complies with EMC directive 2004/108/EC per IEC 61326-3-1 (HART/ DE Units) IEC 61326-1 (FF
Units)
Lightning Protection Option Leakage Current: 10uA max @ 42.4VDC 93C
Impulse rating:
8/20uS 5000A (>10 strikes) 10000A (1 strike min.)
10/1000uS 200A (> 300 strikes)
STG800 Smart Pressure Transmitter 7

Materials Specifications (see model selection guide for availability/restrictions with various models)
Parameter Description
Barrier Diaphragms Material STG800: 316L SS, Hastelloy® C-2762, Monel® 400 3, Tantalum, Gold-plated 316L SS, Gold-plated
Hastelloy® C-276, Gold-plated Monel® 400
STG80L: 316L SS, Hastelloy C-276
Process Head Material STG800: Carbon Steel (Zinc Plated), 316 SS4, Hastelloy® C-2766, Monel® 400 7
STG80L: 316L SS, Hastelloy C-2766
Vent/Drain Valves & Plugs 1 STG800:316 SS4, Hastelloy C-2762, Monel 400 7

STG80L: N/A
Head Gaskets STG800: Glass-filled PTFE standard. Viton® and graphite are optional. STG80L: N/A
STG800: Carbon Steel (Zinc plated) standard. Options include 316 SS, NACE A286 SS bolts and
Meter Body Bolting
nuts or NACE A286 SS bolts and 304 SS nuts STG80L: N/A
Carbon Steel (Zinc-plated) or 304 Stainless Steel or 316SS angle bracket or Carbon Steel or
Mounting Bracket
304SS or 316SS flat bracket available .with 2” pipe bracket. See Figure 3
Fill Fluid Silicone 200, CTFE, NEOBEE M-20 or Silicone 704
Pure Polyester Powder Coated Low Copper (<0.4%)-Aluminum. Meets NEMA 4X, IP66, IP67 and
Electronic Housing
NEMA 7 (explosion proof). All stainless steel housing is optional.
Process Connections STG800: ½ -inch NPT(female), DIN 19213 (standard)
STG80L: ½ -inch NPT(female), ½ -inch NPT male, 9/16 Aminco, DIN19213 (except STG89L),
G ½ -B Male threaded
Wiring Accepts up to 16 AWG (1.5 mm diameter).
Dimensions See Figure 4.
Net Weight STG800: 8.3 pounds (3.8 Kg). STG80L: 3.6 pounds (1.6 Kg) with Aluminum Housing
1 2
Vent/Drains are sealed with Teflon® Hastelloy® C-276 or UNS N10276
3 4
Monel® 400 or UNS N04400 Supplied as 316 SS or as Grade CF8M, the casting equivalent of 316 SS.
5
Carbon Steel heads are zinc-plated and not recommended for water service due to hydrogen migration. For that service, use 316 stainless steel wetted
Process Heads.
6
Hastelloy® C-276 or UNS N10276. Supplied as indicated or as Grade CW12MW, the casting equivalent of Hastelloy® C-276
7
Monel® 400 or UNS N04400. Supplied as indicated or as Grade M30C, the casting equivalent of Monel® 400
8 STG800 Smart Pressure Transmitter

Communications Protocol & Diagnostics

HART Protocol
Compliance Testing:
Version: o Tested according to ITK 6.0.1
HART 7 Software Download
Power Supply Utilizes Class-3 of the Common Software Download
Voltage: 10.8 to 42.4Vdc at terminals procedure as per FF-883 which allows the field devices of any
Load: Maximum 1440 ohms See figure 2 manufacturer to receive software upgrades from any host.
Minimum Load: 0 ohms. (For handheld communications a Honeywell Digitally Enhanced (DE)
minimum load of 250 ohms is required) DE is a Honeywell proprietary protocol which provides
Foundation Fieldbus (FF) digital communications between Honeywell DE enabled
Power Supply Requirements field devices and Hosts.
Voltage: 9.0 to 32.0Vdc at terminals Power Supply
Steady State Current: 17.6mAdc Voltage: 10.8 to 42.4Vdc at terminals
Software Download Current: 27.4mAdc Load: Maximum 1440 ohms See figure 2
Available Function Blocks
Standard Diagnostics
Block Type Quantity Execution Time ST 800 top level diagnostics are reported as either critical
Resource 1 n/a or non-critical and readable via the DD/DTM tools or
integral display as shown
Transducer 1 n/a

Diagnostic 1 n/a

Analog Input 1* 30 ms

PID w/Autotune 1 45 ms

Integrator 1 30 ms

Signal Char (SC) 1 30 ms

LCD Display 1 n/a

Flow Block 1 30 ms

Input Selector 1 30 ms

Arithmetic 1 30 ms
* AI block may have two (2) additional instantiations.
All available function blocks adhere to FOUNDATION
Fieldbus standards. PID blocks support ideal & robust PID
algorithms with full implementation of Auto-tuning.
Link Active Scheduler
Transmitters can perform as a backup Link Active
Scheduler and take over when the host is disconnected.
Acting as a LAS, the device ensures scheduled data
transfers typically used for the regular, cyclic transfer of
control loop data between devices on the Fieldbus.
Number of Devices/Segment
Entity IS model: 6 devices/segment

Schedule Entries Other Certification Options


18 maximum schedule entries Materials
o NACE MRO175, MRO103, ISO15156
Number of VCR’s: 24 max
Approval Certifications:

COMM. FIELD AMBIENT TEMP


AGENCY TYPE OF PROTECTION
OPTION PARAMETERS (Ta)

Explosionproof:
Class I, Division 1, Groups A, B, C, D;
Dust Ignition Proof:
Class II, III, Division 1, Groups E, F, G;
T5: -50 ºC to 85ºC
T6..T5 All Note 1
T6: -50 ºC to 65ºC
Class l, Zone 0/1, AEx db IIC T6..T5
Ga/Gb
Class ll, Zone 21, AEx tb IIIC T95o Db
Intrinsically Safe: 4-20 mA / DE/
Class I, II, III, Division 1, Groups A, B, C, Note 2a -50 ºC to 70ºC
HART
FM ApprovalsTM D, E, F, G: T4
USA
Class l, Zone 0, AEx ia IIC T4 Ga
Foundation
Note 2b -50 ºC to 70ºC
Fieldbus
FISCO Field Device (Only for FF Option)
Ex ia IIC T4 Ga; Ex ic IIC T4 Gc
Nonincendive:
Class I, Division 2, Groups A, B, C, D 4-20 mA / DE/
locations, T4 HART/ Foundation Note 1 -50 ºC to 85ºC
Fieldbus
Class l, Zone 2, AEx nA IIC T4 Gc
Enclosure: Type 4X/ IP66/ IP67 All All -
Explosion Proof:
Class I, Division 1, Groups A, B, C, D;
Dust Ignition Proof:
Class II, III, Division 1, Groups E, F, G;
T6..T5
T5: -50 ºC to 85ºC
All Note 1
T6: -50 ºC to 65ºC
Class I Zone 1 AEx db IIC T6..T5 Ga/Gb
Ex db IIC T6..T5 Ga/Gb

Zone 22 AEx tb IIIC T95o Db


Canadian Ex tb IIIC T95o Db
Standards Intrinsically Safe:
Association 4-20 mA / DE/
(CSA) Class I, II, III, Division 1, Groups A, B, C, Note 2a -50 ºC to 70ºC
D, E, F, G; T4 HART
USA and
Canada
Class I Zone 0 AEx ia IIC T4 Ga
Ex ia IIC T4 Ga
Foundation
Note 2b -50 ºC to 70ºC
Fieldbus
FISCO Field Device (Only for FF Option)
Ex ia IIC T4 Ga; Ex ic IIC T4 Gc
Nonincendive:
Class I, Division 2, Groups A, B, C, D; T4 4-20 mA / DE/
HART/ Foundation Note 1 -50 ºC to 85ºC
Class I Zone 0 AEx nA IIC T4 Gc Fieldbus
Ex nA IIC T4 Gc
Enclosure: Type 4X/ IP66/ IP67 All All -
Approval Certifications: (Continued)
Flameproof:
T5: -50 ºC to 85ºC
II 1/2 G Ex db IIC T6..T5 Ga/Gb All Note 1
T6: -50 ºC to 65ºC
II 2 D Ex tb IIIC T95o Db
Intrinsically Safe: 4-20 mA / DE/
Note 2a -50 ºC to 70ºC
II 1 G Ex ia IIC T4 Ga HART

ATEX FISCO Field Device (Only for FF Option) Foundation -50 ºC to 70ºC
Note 2b
Fieldbus
Ex ia IIC T4 Ga; Ex ic IIC T4 Gc
4-20 mA / DE/
Nonincendive: HART/
Foundation Note 1 -50 ºC to 85ºC
II 3 G Ex nA IIC T4 Gc
Fieldbus
Enclosure: IP66/ IP67 All All -
Flameproof :
T5: -50 ºC to 85ºC
Ex db IIC T6..T5 Ga/Gb All Note 1
T6: -50 ºC to 65ºC
Ex tb IIIC Db T 95oC Db
Intrinsically Safe: 4-20 mA / DE/
Note 2a -50 ºC to 70ºC
Ex ia IIC T4 Ga HART
IECEx
World FISCO Field Device (Only for FF Option) Foundation
Note 2b -50 ºC to 70ºC
Fieldbus
Ex ia IIC T4 Ga; Ex ic IIC T4 Gc
Nonincendive: 4-20 mA / DE/
HART/
Ex nA IIC T4 Gc Foundation Note 1 -50 ºC to 85ºC
Fieldbus
Enclosure: IP66/ IP67 All All -
Flameproof :
Ex d IIC Ga/Gb T4 All Note 1 -50 ºC to 85ºC
Ex tb IIIC Db T 95oC
Intrinsically Safe: 4-20 mA / DE/
Note 2a -50 ºC to 70ºC
Ex ia IIC Ga T4 HART
SAEx
South Africa FISCO Field Device (Only for FF Option) Foundation
Note 2b -50 ºC to 70ºC
Fieldbus
Ex ia IIC T4 Ga; Ex ic IIC T4 Gc
Nonincendive: 4-20 mA / DE/
HART/
Ex nA IIC Gc T4 Foundation Note 1 -50 ºC to 85ºC
Fieldbus

Enclosure: IP66/ IP67 All All -


Flameproof:
Ex db IIC T6..T5 Ga/Gb All Note 1 50 ºC to 85ºC
Ex tb IIIC T 95oC Db
Intrinsically Safe: 4-20 mA / DE/
Note 2a 50 ºC to 70ºC
INMETRO Ex ia IIC T4 Ga HART

Brazil FISCO Field Device (Only for FF Option) Foundation


Fieldbus
Note 2b 50 ºC to 70ºC
Ex ia IIC T4 Ga; Ex ic IIC T4 Gc
Nonincendive: 4-20 mA / DE/
HART/
Ex nA IIC T4 Gc Foundation Note 1 -50 ºC to 85ºC
Fieldbus

Enclosure : IP 66/67 All All -


Approval Certifications: (Continued)
Flameproof:
Ex d IIC Ga/Gb T4 All Note 1 -50 ºC to 85ºC
Ex tb IIIC Db T 85oC
Intrinsically Safe: 4-20 mA / DE/
Note 2a -50 ºC to 70ºC
Ex ia IIC Ga T4 HART
NEPSI
China FISCO Field Device (Only for FF Option) Foundation
Note 2b -50 ºC to 70ºC
Fieldbus
Ex ia IIC T4
Nonincendive: 4-20 mA / DE/
Ex nA IIC Gc T4 HART/ Foundation Note 1 -50 ºC to 85ºC
Fieldbus

Enclosure : IP 66/67 All All -


Flameproof:
1 Ex d IIC Ga/Gb T4 All Note 1 -50 ºC to 85ºC
Ex tb IIIC Db T 85oC
EAC
Russia, Belarus Intrinsically Safe: 4-20 mA / DE/
Note 2a -50 ºC to 70ºC
and 0 Ex ia IIC Ga T4 HART
Kazakhstan
FISCO Field Device (Only for FF Option) Foundation
Note 2b -50 ºC to 70ºC
Fieldbus
Ex ia IIC T4
Enclosure : IP 66/67 All All
T6: Ta= -50 ºC to
Flameproof :
65ºC
Ex d IIC T6..T5 All Note 1
T5: Ta= -50 ºC to
Ex tD T 95oC
85ºC
KOSHA
4-20 mA / DE/
Korea Note 2a Ta= -50 ºC to 70ºC
Intrinsically Safe: HART
Ex ia IIC T4 Foundation Ta= -50 ºC to 70ºC
Note 2b and 2c
Fieldbus

Enclosure: IP66/ IP67 All All -

Notes:
1. Operating Parameters:
Voltage= 11 to 42 V DC Current= 4-20 mA Normal
= 10 to 30 V (FF) = 30 mA (FF)
2. Intrinsically Safe Entity Parameters
a. Analog/ DE/ HART Entity Values:
Vmax= Ui = 30V Imax= Ii= 105mA Ci = 4.2nF Li =984 uH Pi =0.9W
Transmitter with Terminal Block Revision E or Later
Vmax= Ui = 30V Imax= Ii= 225mA Ci = 4.2nF Li = 0 Pi =0.9W
Note : Transmitter with Terminal Block Revision E or later
The revision is on the label that is on the module. There will be two lines of text on the label:
 First is the Module Part #: 50049839-001 or 50049839-002
 Second line has the supplier information, along with the REVISION:
XXXXXXX-EXXXX, THE “X” is production related, THE POSITION of the “E” IS THE REVISION.

b. Foundation Fieldbus- Entity Values


Vmax= Ui = 30V Imax= Ii= 180mA Ci = 0nF Li = 984 uH Pi =1W
Transmitter with Terminal Block Revision F or Later )
Vmax= Ui = 30V Imax= Ii= 225mA Ci =0nF Li = 0 Pi =1 W

FISCO Field Device Imax= Ii= 380 mA Ci = 0nF Li = 0 Pi =5.32 W


Vmax= Ui = 17.5V
Note : Transmitter with Terminal Block Revision F or later
The revision is on the label that is on the module. There will be two lines of text on the label:

 First is the Module Part #: 50049839-003 or 50049839-004


 Second line has the supplier information, along with the REVISION:
XXXXXXX-EXXXX, THE “X” is production related, THE POSITION of the “E” IS THE REVISION.

Approval Certifications: (Continued)

This certificate defines the certifications covered for the ST 800 Pressure Transmitter family of
products, including the SMV 800 Smart Multivariable Transmitter. It represents the compilation
of the five certificates Honeywell currently has covering the certification of these products into
marine applications.

For ST 800 Smart Pressure Transmitter and SMV800 Smart Multivarible Transmitter

American Bureau of Shipping (ABS) - 2009 Steel Vessel Rules 1-1-4/3.7, 4-6-2/5.15, 4-8-3/13
& 13.5, 4-8-4/27.5.1, 4-9-7/13. Certificate number: 04-HS417416-PDA

Marine Certificates
Bureau Veritas (BV) - Product Code: 389:1H. Certificate number: 12660/B0 BV

Det Norske Veritas (DNV) - Location Classes: Temperature D, Humidity B, Vibration A, EMC B,
Enclosure C. For salt spray exposure; enclosure of 316 SST or 2-part epoxy protection with 316
SST bolts to be applied. Certificate number: A-11476

Korean Register of Shipping (KR) - Certificate number: LOX17743-AE001

Lloyd's Register (LR) - Certificate number: 02/60001(E1) & (E2)

SIL 2/3 Certification IEC 61508 SIL 2 for non-redundant use and SIL 3 for redundant use according to EXIDA and
TÜV Nord Sys Tec GmbH & Co. KG under the following standards: IEC61508-1: 2010; IEC
61508-2: 2010; IEC61508-3: 2010.

MEASUREMENT Certificate Issued by NMI Certin B.V.


Mechanical Class: M3 Electromagnetic Environment: E3
INTRUMENTS
Ambient Temperature Range: -25 oC to + 55 oC
DIRECTIVE (MID)
2004/ 22/ EC
Unit Custom Calibration
STD820 0 to 1000 mBar
STD830 0 to 7 Bar
STA84L 0 to 35 Bar A
STG84L 0 to 35 Bar
STD870 0 to 100 Bar
STA87L 0 to 100 Bar A
STG87L 0 to 100 Bar
Mounting & Dimensional Drawings)

millimeters
Reference Dimensions: inches

Mounting Configurations: (Dual head design)

Mounting Configurations (Inline Designs)


millimeters
Reference Dimensions: inches

Dimensions: (Dual head design)

Figure 4 – Typical mounting dimensions of STG840 & STG870 for reference


millimeters
Reference Dimensions: inches

Dimension (Inline Design)

Figure 5 – Typical mounting dimensions of STG84L, STG87L, STG88L, & STG89L for reference
Model Selection Guides are subject to change and are inserted into the specifications as guidance only.
Prior to specifying or ordering a model check for the latest revision Model Selection Guides which are published at:
www.honeywellprocess.com/en-US/pages/default.aspx

Model Selection Guide

Model STG800
Gauge Pressure Transmitters
Model Selection Guide
34-ST-16-83 Issue 18
Instructions: Make selections from all Tables using column below the proper arrow. Asterisk indicates availability. Letter (a)
refers to restrictions highlighted in the restrictions table. Tables delimited w ith dashes.

KEY NUMBER URL/Max Span LRL Min Span Units Selection Availability
50 (3.5) -14.7 (-1.0) 0.5 (.035) psi (bar) STG830
Gauge
500 (35) -14.7 (-1.0) 5 (.35) psi (bar) STG840
Dual Head
3000 (210) -14.7 (-1.0) 30 (2.1) psi (bar) STG870
50 (3.5) -14.7 (-1.0) 0.5 (.035) psi (bar) STG83L
500 (35) -14.7 (-1.0) 5 (.35) psi (bar) STG84L
Gauge -14.7 (-1.0)
3000 (210) 30(2.1) psi (bar) STG87L
In-Line
6000 (420) -14.7 (-1.0) 60 (4.2) psi (bar) STG88L
10000 (690) -14.7 (-1.0) 100 (6.9) psi (bar) STG89L

TABLE I METER BODY SELECTIONS


P ro c e s s H e a d/ R e f e re nc e H e a d M a t e ria l 1 b Barrier Diaphragm Material
316L SS A_ _ _ _ _ * *
®
Hastelloy C - 276 B_____ * *
®
Monel 400 C_____ * *
Plated Carbon Steel / Plated
Tantalum D_____ a a
Carbon Steel
Gold Plated 316L SS 1_____ * *
Gold Plated Hastelloy C-276 2_____ * *
Gold Plated Monel 400 3_____ * *
a. Process
316L SS E_____ * * * * * * *
Head &
Hastelloy C - 276 F_____ * * * * * * *
Diaphragm
Monel 400 G_____ * *
Materials 316 Stainless Steel / 316
Tantalum H_____ a a
Stainless Steel
Gold Plated 316L SS 4_____ * *
Gold Plated Hastelloy C-276 5_____ * *
Gold Plated Monel 400 6_____ * *
Hastelloy C - 276 J_____ * * * * * * *
Hastelloy C - 276 / 316
Tantalum K_____ a a
Stainless Steel
Gold Plated Hastelloy C-276 7_____ * *
Monel 400 / 316 Monel 400 L_____ a a
Stainless Steel Gold Plated Monel 400 8_____ a a
Silicone Oil 200 _ 1_ _ _ _ * * * * * * *
Fluorinated Oil CTFE _2____ * * * * * * *
b. Fill Fluid Silicone Oil 704 _3____ * * * * * * *
NEOBEE® M-20 _4____ * * * * * * *
Size/Type Material
9/16" Aminco Same as Process Head _ _ A_ _ _ * * * * *
c. Process 1a
1/2" NPT (female) Same as Process Head __G___ * * * * * * *
Connection
1/2" NPT (male) Same as Process Head __H___ * * * * *
DIN 19213 (1/4" female NPT) Same as Process Head __D___ * * * * * *
G 1/2 B Threaded Fitting Same as Process Head __B___ * * * * *
None ___0__ * * * * *
Carbon Steel ___C__ * *
316 SS ___S__ * *
d. Bolt/Nuts Grade 660 (NACE A286) with NACE 304 SS Nuts ___N__ * *
Materials Grade 660 (NACE A286) Bolts & Nuts ___K__ p p
Monel K500 _ _ _ M_ _ p p
Super Duplex ___D__ p p
B7M ___B__ * *
Head Type Vent Type Location Vent Material
None None None None ____0_ * * * * *
Single Ended None None None ____1_ * *
1
e. Vent/Drain Single Ended Standard Vent Side Matches Head Material ____2_ * *
Type/Location Single Ended Center Vent Side Stainless Steel Only ____3_ t t
1
Dual Ended Standard Vent End Matches Head Material ____4_ * *
Dual Ended Center Vent End Stainless Steel only ____5_ t t
1
Dual Ended Std Vent/Plug Side/End Matches Head Material ____6_ * *
None _____0 * * * * *
®
f. Gasket Teflon or PTFE (Glass Filled) _____A * *
®
Materials Viton _____B * *
Graphite _____C * *
1
Except Carbon Steel Heads shall use 316SS Vent/Drain & Plugs and or 1/2" adapters
1a
STG830,840,870 supplied via 1/2" flange adapter same material as process head except carbon steel shall use 316 SS
1b
Reference head available w ith Dual Head Gage models only. In-Line Gage models are supplied w ith Process Head only.
STG89L
STG88L
STG87L
STG84L
STG83L
STG870
STG830, STG840

TABLE II METER BODY & CONNECTION ORIENTATION


2
Standard High Side Left, Low Side Right / Std Head Orientation 1 * * * * * * *
Head/Connect
Reversed Low Side Left, High Side Right2 / Std Head Orientation 2 * *
Orientation 2 0
90/Standard High Side Left, Low Side Right / 90 Head Rotation 3 h h

TABLE III AGENCY APPROVALS


No Approvals Required 0 * * * * * * *
<FM> Explosion proof, Intrinsically Safe, Non-incendive, & Dustproof A * * * * * * *
CSA Explosion proof, Intrinsically Safe, Non-incendive, & Dustproof B * * * * * * p
ATEX Explosion proof, Intrinsically Safe & Non-incendive C * * * * * * *
IECEx Explosion proof, Intrinsically Safe & Non-incendive D * * * * * * *
Approvals SAEx/CCoE Explosion proof, Intrinsically Safe & Non-incendive E * * * * * * *
INMETRO Explosion proof, Intrinsically Safe & Non-incendive F * * * * * * *
NEPSI Explosion proof, Intrinsically Safe & Non-incendive G * * * * * * *
KOSHA Explosion proof, Intrinsically Safe & Non-incendive H * * * * * * *
EAC Customs Union(Russia,Belarus,Kazakhstan)Ex Approval, Flame proof, Intrinsically Safe I * * * * * * *

TABLE IV TRANSMITTER ELECTRONICS SELECTIONS


Material Connection Lightning Protection
Polyester Powder Coated Aluminum 1/2 NPT None A_ _ * * * * * * *
Polyester Powder Coated Aluminum M20 None B__ * * * * * * *
a. Electronic
Polyester Powder Coated Aluminum 1/2 NPT Yes C__ * * * * * * *
Housing
Polyester Powder Coated Aluminum M20 Yes D__ * * * * * * *
Material &
316 Stainless Steel (Grade CF8M) 1/2 NPT None E__ * * * * * * *
Connection Type
316 Stainless Steel (Grade CF8M) M20 None F__ * * * * * * *
316 Stainless Steel (Grade CF8M) 1/2 NPT Yes G__ * * * * * * *
316 Stainless Steel (Grade CF8M) M20 Yes H__ * * * * * * *
Analog Output Digital Protocol
b. Output/ 4-20mA dc HART Protocol _H_ * * * * * * *
Protocol 4-20mA dc DE Protocol _D_ u u u u u u u
none Foundation Fieldbus _F_ * * * * * * *
Indicator Ext Zero, Span & Config Buttons Languages
None None None __0 * * * * * * *
None Yes (Zero/Span Only) None __A f f f f f f f
c. Customer Basic None EN __B * * * * * * *
Interface Basic Yes EN __C * * * * * * *
Selections Advanced None EN, GR, FR, IT, SP, RU, TU __D * * * * * * *
Advanced Yes EN, GR, FR, IT, SP, RU, TU __E * * * * * * *
Advanced None EN, CH, JP __H * * * * * * *
Advanced Yes EN, CH, JP __J * * * * * * *

TABLE V CONFIGURATION SELECTIONS


Diagnostics
a. Application
Standard Diagnostics 1__ * * * * * * *
Software
Advanced Diagnostics (Above with Plugged Impulse Detection PILD) 2__ * * * * * * *
Write Protect Fail Mode High & Low Output Lim its 3
Disabled High> 21.0mAdc Honeywell Std (3.8 - 20.8 mAdc) _1_ f f f f f f f
b. Output Limit, Disabled Low< 3.6mAdc Honeywell Std (3.8 - 20.8 mAdc) _2_ f f f f f f f
Failsafe & Write Enabled High> 21.0mAdc Honeywell Std (3.8 - 20.8 mAdc) _3_ f f f f f f f
Protect Settings Enabled Low< 3.6mAdc Honeywell Std (3.8 - 20.8 mAdc) _4_ f f f f f f f
Enabled N/A N/A Fieldbus or Profibus _5_ g g g g g g g
Disabled N/A N/A Fieldbus or Profibus _6_ g g g g g g g
General Configuration
c. General Factory Standard __S * * * * * * *
Configuration Custom Configuration (Unit Data Required from customer) __C * * * * * * *
2
Left side/Right side as view ed from the customer connection perspective
3
NAMUR Output Limits are configurable by customer
STG89L
STG88L
STG87L
STG84L
STG83L
STG870
TABLE VI CALIBRATION & ACCURACY SELECTIONS STG830, STG840
Accuracy Calibrated Range Calibration Qty
Standard Factory Standard Single Calibration A * * * * * * *
Standard Custom (Unit Data Required) Single Calibration B * * * * * * *
Standard Custom (Unit Data Required) Dual Calibration C * * * * * * *
a. Accuracy and
Standard Custom (Unit Data Required) Triple Calibration D * * * * * * *
Calibration
High Accuracy Factory Standard Single Calibration E s s s s s s s
High Accuracy Custom (Unit Data Required) Single Calibration F s s s s s s s
High Accuracy Custom (Unit Data Required) Dual Calibration G s s s s s s s
High Accuracy Custom (Unit Data Required) Triple Calibration H s s s s s s s

TABLE VII ACCESSORY SELECTIONS


Bracket Type Material
None None 0___ * * * * * * *
Angle Bracket Carbon Steel 1___ * * * * * * *
Angle Bracket 304 SS 2___ * * * * * * *
Angle Bracket 316 SS 3___ * * * * * * *
a. Mounting Marine Approved Bracket Carbon Steel 8___ * *
Bracket Marine Approved Bracket (In-Line) Carbon Steel 9___ * * * * *
Marine Approved Bracket 304 SS 4___ * *
Marine Approved Bracket (In-Line) 304 SS A_ _ _ * * * * *
Flat Bracket Carbon Steel 5___ * * * * * * *
Flat Bracket 304 SS 6___ * * * * * * *
Flat Bracket 316 SS 7___ * * * * * * *
Custom er Tag Type
b. Customer No customer tag _0__ * * * * * * *
Tag One Wired Stainless Steel Tag (Up to 4 lines 26 char/line) _1__ * * * * * * *
Two Wired Stainless Steel Tag (Up to 4 lines 26 char/line) _2__ * * * * * * *
Unassem bled Conduit Plugs & Adapters
No Conduit Plugs or Adapters Required _ _ A0 * * * * * * *
c. Unassembled
1/2 NPT Male to 3/4 NPT Female 316 SS Certified Conduit Adapter _ _ A2 n n n n n n n
Conduit
1/2 NPT 316 SS Certified Conduit Plug _ _ A6 n n n n n n n
Plugs &
M20 316 SS Certified Conduit Plug _ _ A7 m m m m m m m
Adapters
Minifast® 4 pin (1/2 NPT) (not suitable for X-Proof applications) _ _ A8 n n n n n n n
Minifast® 4 pin (M20) (not suitable for X-Proof applications) _ _ A9 m m m m m m m

TABLE VIII OTHER Certifications & Options: (String in sequence com m a delim ited (XX, XX, XX,….)
No additional options 00 * * * * * * *
Low Temperature Rating ( -50 deg C min. ambient operative temperature limit) LT w w w w w w w
NACE MR0175; MR0103; ISO15156 (FC33338) Process wetted parts only FG * * * * * * *
b
NACE MR0175; MR0103; ISO15156 (FC33339) Process wetted and non-wetted parts F7 c c c c c c c
Marine (DNV, ABS, BV, KR, LR) MT d d d d d d d
EN10204 Type 3.1 Material Traceability (FC33341) FX * * *
* * * *
MID Approved Transmitter - Contact Tech Support for specific MID approved ranges MD v v
Certificate of Conformance (F3391) F3 * * * * * * *
b
Calibration Test Report & Certificate of Conformance (F3399) F1 * * * * * * *
Certifications &
Certificate of Origin (F0195) F5 * * * * * * *
Warranty
FMEDA (SIL 2/3) Certification (FC33337) FE j j j j j j j
Over-Pressure Leak Test Certificate (1.5X MAWP) (F3392) TP * * * * * * *
Cert Clean for O2 or CL 2 service per ASTM G93 OX e e e e e e e
PMI Certification 1 PM * * * * * * *
Extended Warranty Additional 1 year 01 * * * * * * *
Extended Warranty Additional 2 years 02 * * * * * * *
Extended Warranty Additional 3 years 03 * * * * * * * b
Extended Warranty Additional 4 years 04 * * * * * * *
Extended Warranty Additional 15 years 15 * * * * * * *

TABLE IX Manufacturing Specials


Factory Factory Identification 0000 * * * * * * *
RESTRICTIONS
Restriction Available Only with Not Available with
Letter Table Selection(s) Table Selection(s)
a VIII FG, F7
c Id _ _ _ 0,N,K,D,B _ _ _ Ia D,H,K,L,8 _ _ _ _ _ _
d IV a C,D,G,H _ _ VIIa 1,2,3,5,6,7 _ _ _
e Ib _2_____
f IV b _ F_
g IVb _ H, D _
Ie _ _ _ _ 4, 5, 6 _ _
h
Vlla 1,2,3,4,5,6,7,8 _ _ _
j IV b H Vb _ 1,2,6_
m IV a B,D, F, H _ _
n IV a A,C, E, G _ _
p III B- No CRN number available
s Ia A,E _ _ _ _ _ _
t Ia J, K, 7, L, 8
Va 2__
u
VIa C,D,G,H
v IV a C,D,G,H _ _ IVb _D,F_
w Ib _1_____ VIII FE
b Select Only one option from this group

1
The PM option is available on all Smartline Pressure Transmitter process wetted parts such as process heads, flanges, bushings and vent plugs except plated carbon
steel process heads and flanges. PM option information is also available on diaphragms except Gold plated and STG and STA in-line construction pressure
transmitters.

FIELD INSTALLABLE ACCESSORY KITS


Description Kit Number
Integrally Mounted Basic Indicator Kit (Compatible with all Electronic Modules) 50049911-501
Integrally Mounted Advanced Indicator Kit (compatible with all Electronic Modules) 50049846-501
Terminal Strip w/o Lightening Protection for HART or DE Modules 50075472-531
Terminal Strip w/Lightning Protection Kit for HART or DE Modules 50075472-532
Terminal Strip w/o Lightening Protection FFB/Profib us Module 50075472-533
Terminal Strip w/Lightning Protection Kit for FFB/Profib us Module 50075472-534
HART Electronics Module 50049849-501
HART Electronics Module w/connection for external configuration buttons 50049849-502
DE Electronics Module 50049849-503
DE Electronics Module w/connection for external configuration buttons 50049849-504
FFB Electronics Module Kit 50049849-507
FFB Electronics Module w/connection for external configuration buttons 50049849-508
Sales and Service
For application assistance, current specifications, pricing, or name of the nearest Authorized Distributor, contact one
of the offices below.

ASIA PACIFIC EMEA AMERICA’S


Honeywell Process Solutions, Honeywell Process Solutions, Honeywell Process Solutions,
(TAC) hfs-tac- Phone: + 80012026455 or Phone: (TAC) 1-800-423-9883 or
[email protected] +44 (0)1344 656000 215/641-3610
(Sales) 1-800-343-0228
Australia Email: (Sales)
Honeywell Limited [email protected] Email: (Sales)
Phone: +(61) 7-3846 1255
FAX: +(61) 7-3840 6481 or [email protected]
Toll Free 1300-36-39-36 (TAC) or
Toll Free Fax: [email protected] (TAC)
1300-36-04-70 [email protected]
China – PRC - Shanghai
Honeywell China Inc.
Phone: (86-21) 5257-4568
Fax: (86-21) 6237-2826

Singapore
Honeywell Pte Ltd.
Phone: +(65) 6580 3278
Fax: +(65) 6445-3033

South Korea
Honeywell Korea Co Ltd
Phone: +(822) 799 6114
Fax: +(822) 792 9015

Specifications are subject to change without notice.

For more information


To learn more about SmartLine Transmitters,
visit www.honeywellprocess.com
Or contact your Honeywell Account Manager

Process Solutions
Honeywell
1250 W Sam Houston Pkwy S
Houston, TX 77042

Honeywell Control Systems Ltd


Honeywell House, Skimped Hill Lane
Bracknell, England, RG12 1EB

Shanghai City Centre, 100 Jungi Road


Shanghai, China 20061
34-ST-03-83
March 2018
www.honeywellprocess.com 2018 Honeywell International Inc.