AT4011E 05 Ical CFO UsersGuide
AT4011E 05 Ical CFO UsersGuide
AT4011E, Rev. 05
User’s Guide
The contents of this document may not be reproduced in any form or communicated to any third party without the
prior written consent of AVL. While every effort is made to ensure its correctness, AVL assumes no responsibility
neither for errors and omissions which may occur in this document nor for damage caused by them.
All mentioned trademarks or registered trademarks are owned by their respective owners.
WARNING
WARNING
Danger of explosion
Due to the ignition of an explosive mixture of ambient air and escaping gas there
is a danger of personal injury and damage to property.
Ensure there is sufficient circulation and ventilation in the room in which the
device is operated.
Monitor ambient air with appropriate measure (LEL sensor).
Do not switch off or remove exhaust and ventilation equipment when the system
is running.
Connect the gas outlet correctly. The connections and the system to be tested
must not have any leaks.
Keep away from open flames or other sources of ignition.
WARNING
Danger of poisoning
Leaking gas leads to the danger of poisoning.
Ensure there is sufficient circulation and ventilation in the room in which the
device is operated.
Monitor ambient air with appropriate measure (CO sensor).
Do not switch off or remove exhaust and ventilation equipment when the system
is running.
Connect gas outlet correctly. The connections and the system to be tested must
not have any leaks.
CAUTION
When handling gases, the applicable regulations and the manufacturer’s mate-
rial safety data sheets for these substances must be heeded and adhered to with
respect to storage, handling, use, and disposal!
Heed applicable regulations for the handling, storage, and transport of
compressed gas bottles. Eating, drinking, smoking or storage of food in rooms
containing chemicals is forbidden.
Always observe the relevant safety instructions for span gas bottles.
The device must not be operated in potentially explosive atmospheres.
Safe operation can only be ensured if the device is installed and commissioned
by qualified personnel in accordance with the instructions given in this manual.
Information
The device must be set up in such a way that it can be easily disconnected from
the mains at any time. If necessary, provide for an emergency stop switch.
NOTICE
Before connecting up and operating the device, read carefully and observe the
supplied user's guide and the user's guides supplied with the additionally used
devices.
Ensure that all connections are made in accordance with the installation instruc-
tions described in this user's guide and the user's guides supplied with the addi-
tionally used devices. Flawless and reliable operation of the device can only be
guaranteed if these requirements are met.
All work on electrical components or lines, as well as device installation, may
only be performed by trained personnel in accordance with the VDE standards
(VDE = the German Association for Electrical, Electronic and Information Tech-
nologies).
NOTICE
Table of Contents
Warnings and Safety Instructions................................................................................................................ 3
3 Installation ....................................................................................................................... 21
3.1 Assembly ................................................................................................................................ 21
4.3 Operation................................................................................................................................ 28
5 Overview .......................................................................................................................... 29
5.1 Connections and Operating Elements.................................................................................... 29
8 Calibration........................................................................................................................ 73
Index.............................................................................................................................................................. 81
1.3 Warranty
The manufacturer assumes no liability for damage and breakdowns that result
from a failure to adhere to the user’s guide.
The user’s guide must be completed by the operator so that it includes operating
instructions based on the relevant national regulations concerning accident
prevention and environmental protection, including information about supervi-
sory and reporting duties for the consideration of operational particularities, e.g.
with respect to work organization, workflows, and personnel deployed.
In addition to the user’s guide and the binding regulations concerning accident
prevention that apply in the place of use, the recognized technical rules for safe
and professional work must also be followed.
Information
The warranty becomes void if the device is not maintained annually by AVL
Services or by a maintenance technician trained by AVL. During such mainte-
nance, the device is checked, repairs are done and a calibration is performed.
Information
Information
WARNING
CAUTION
Notice:
NOTICE
This text indicates situations or operation errors which could result in property
damage or data loss.
Information
Lists:
1. Instructions
2.
• One-step instructions
Unordered series of concepts, items or options
–
–
2 Scope of Delivery
Please understand that not all of the device's versions and options can be
described in every detail in this user's guide. If you need additional information or
you have a special problem that was not explained in sufficient depth in this
user’s guide, please contact your local sales office or representative.
2.1.1 Versions
The following versions are available:
iCAL CFO Light Duty, Low Emission (iCAL CFO LD-LE)
– CO 80 – 560 smlpm
– C 3H 8 50 – 430 smlpm
1 Deflecting plate
2 Nozzle
2.2.2 Trolley
By mounting the iCAL CFO onto a trolley, the device can be used at different
places and easily moved from one place to another. The trolley also offers addi-
tional space for storing any necessary connecting cables or tubing.
Top
Fragile
Keep dry
Center of gravity
Sealed packaging
NOTICE
During transport, do not subject the device to hard blows. Handle carefully, espe-
cially when lifting and lowering. All gas inlets and outlets must be sealed.
3 Installation
Information
3.1 Assembly
WARNING
Danger of explosion
Due to the ignition of an explosive mixture of ambient air and escaping gas there
is a danger of personal injury and damage to property.
Ensure there is sufficient circulation and ventilation in the room in which the
device is operated.
Monitor ambient air with appropriate measure (LEL sensor).
Do not switch off or remove exhaust and ventilation equipment when the system
is running.
Connect the gas outlet correctly. The connections and the system to be tested
must not have any leaks.
Keep away from open flames or other sources of ignition.
NOTICE
• If a 19" cabinet is used: Guarantee sufficient air supply for the ventilation.
• Maintain sufficient distance to warm air ventilators, heaters, and
high-voltage devices.
• Install the device horizontally.
• Install the device in a vibration-free area.
• Only set the unit up in gas-monitored rooms.
• The ambient air must be clean and free of heavy metals.
• Do not operate the device at heavily fluctuating temperatures.
• Protect the device from direct sunlight and contact with water. In particular,
do not place any containers with liquids on top of the device or next to it, as
the containers could leak or topple over causing liquid to enter the device,
which could lead to short circuits.
• Adhere to the ambient conditions specified in the Technical Data.
For an overview of the back panel with gas-related and electrical connections,
see Chapter Connections and Operating Elements on page 29.
NOTICE
When using gas bottle pressure regulators and supply hoses, we recommend
using new ones only and not old ones that have become dirty from prolonged
use.
NOTICE
Flawless operation of the unit can only be guaranteed if the device is used by
qualified personnel and if the instructions specified in this manual are observed
during the system's installation and commissioning.
Handling It is absolutely necessary to adhere strictly to the safety measures for handling
poisonous and flammable gases. In addition, the local safety standards and fire
department standards or other (e.g. plant-internal) regulations must be heeded
and adhered to during operation.
3.2 Commissioning
For commissioning, proceed as follows:
1. Unpack the device.
2. Close any gas connections that are not used (SWAGELOK).
3. Connect the RS232 and/or TCP interface.
4. Check whether your power supply connection delivers the voltage required
for the iCAL CFO.
The voltage is indicated on the back of the device next to the power
connection socket. These voltage levels must match.
5. Make sure that you use gas bottle pressure regulators within their specified
flow ranges.
The iCAL CFO requires a stable inlet pressure within the specified inlet
pressure limits.
If you set flow rates that are too low, please note that most of the gas
bottle pressure regulators are not designed for such low flow rates and
might become unstable. The iCAL CFO does, however, require a
stable inlet pressure, at least for a period of 30 seconds. If slower
changes occur, the device is capable of compensating them accord-
ingly.
If you happen to have this type of gas bottle pressure regulator, you
might be able to tackle the problem by providing an additional bypass
flow downstream of the gas bottle pressure regulator to achieve a
higher flow rate. It might also be helpful to place a disk with a very
small diameter at the outlet of the gas bottle pressure regulator in order
to get the gas bottle pressure regulator back into its control range.
Please ask the manufacturer of your gas bottle pressure regulator what
could be done.
6. Connect the device to the power supply.
7. Set up the serial or network interface.
For more detailed information, please read the sections below and additionally
the sections:
The iCAL CFO is used to check the tightness of the CVS dilution system in emis-
sion testing. A defined volume flow (l/min) of a pure gas (100% C3H8 or CO) is
applied to the CVS system upstream of the mixing point. In the CVS system, this
gas is diluted as defined in the flow settings. Analysis devices are used to
measure the resulting gas concentration of the gas mixture in order to confirm
the tightness of the CVS system.
The injected volume flow is produced by combining critical sonic nozzles with
differently calibrated flow rates. Their level of precision is better than 0.4%. The
pressure of the used gas is controlled by a high-precision pressure regulator with
high long-term stability.
The iCAL CFO is able to be remote-controlled by a host computer via the TCP/IP
interface (alternatively, also serially via RS232 V24) in accordance with the AK
protocol.
.
8
2
7 3
4
6
All system components that come into contact with gas are made of stainless
steel (V4A) or noble metals. The seals are made of Viton and the tubing is made
of PFA.
P-----1 2
P2
To calculate the mass flow rate through a critical orifice that operates in the sonic
range, apply the following equation:
P1
Q m = ------- K gas
T
In this equation, Kgas represents the specific gas factor in relation to N2 and T
the absolute gas temperature.
The mass flow through the critical orifice is independent of inlet pressure (e.g.
the counter-pressure produced by a connected analyzer) as long as the ratio of
absolute inlet pressure P1 to outlet pressure P2 is greater than or equal to 2. This
allows very precise and stable gas flow rates. The cylindrical nozzle body that
contains the critical orifice and the critical orifice itself are made of stainless
steel. The individual nozzles are protected from foreign particles by stainless
steel filters.
pi
Q s = K v -------
ti
But Kv can only be regarded as constant for a small range of pi. Outside of this
range, Kv is a function of pi, which is estimated with a polynomial as follows:
2 N
Kv pi = a0 + a1 pi + a2 pi + + aN pi
The polynomial coefficients are calculated from the calibration curve (see
Chapter Calibration on page 73). The curve is only valid for the media used for
the calibration and is based on a normalized temperature and a normalized pres-
sure (see below). Normally, a third- to fifth-order polynomial is used. It is also
possible to enter a linearization for the pressure and temperature sensors.
A sample j of the standard volume flow is calculated as follows:
Pj
Q sj = K vj P j ---------
Tj
Ts
C t = -------
T cs
P cs
C p = -------
Ps
Ps and Ts are the new standard values. The calibration values are Pcs=1.01325
bar for inlet pressure and Tcs=293.15 K for inlet temperature. This results in the
new standard volume flow Q'sj
Q' sj = Q sj C t C p
To obtain the current volume flow, the following correction factors for tempera-
ture and pressure are used:
Tj
C atj = -------
T cs
P cs
C apj = -------
Pj
The current volume flow Qaj is therefore:
Q aj = Q sj C atj C apj
4.3 Operation
With the help of display and keypad, it does not take long to learn how to operate
the device. Even so, we recommend reading the following chapters very care-
fully, so that you are familiar with all of the operating options.
The device left the factory in a safe and error-free condition. To ensure safe and
correct operation, all requirements relating to packaging, storing, transport and
delivery as well as correct installation must be fully met.
Operating Options The device can be operated using the keypad and the display attached to the
front of the device. To control the AK commands, please see Chapter AK
Commands for Remote Control on page 63.
For further details, see Chapter Display and Operation on page 33.
Designation Function
MV1, 2, 3 Solenoid valves
N1, 2 Critical orifices
P1 Pressure sensor inlet
P2 Pressure sensor outlet
SV1 Safety valve
PV1 Proportional valve
F1 Filter
T1 Temperature sensor
FR1 Flame trap
EV1, 2, 3 External valves (shut-off valve, gas selection
valve and bypass valve)
5 Overview
NOTICE
Correct functioning is only ensured if the inlet pressure indicated in the data
sheet is met.
NOTICE
The temperature in the block containing the critical orifices and the sensors is
regulated internally. For the device to reach rated accuracy, it requires a
heating-up time of around 45 minutes.
For more detailed information on the functions, please refer to Chapters Manual
Operation [F2] on page 38 and AK Commands for Remote Control on page 63.
CAUTION
Information
It is strongly recommended to install an excess flow cut-off valve in the gas feed
line.
Pin Function
1 DCD
2 RxD
3 TxD
4 DTR
5 SG
6 DSR
7 RTS
8 CTS
9 RI
Pin Function
1 +24V
2 GND
3 not assigned
Shut-off valve After switching on the device, a voltage of 24 V is applied at this connector. By
connecting a shut-off valve, a connected gas bottle can be closed automatically.
Socket designation: XG7: SHUT OFF VALVE
Connecting an This makes it possible to control an external valve, e.g. for a SHED application
external 3/2 valve (evaporative determination testing), in such a way that when the integrators are
switched on (see Chapter Control Commands on page 68) the gas can be
injected into the chamber. During the stabilization phase, the gas flows through
the valve into the extraction system. As a result, only so much gas is injected
that can be integrated.
Socket designation: XG8: BYPASS VALVE
Switching between This option allows switching between 2 gas bottles and is enabled by an addi-
C3H8 and CO tional 3/2-way valve. If C3H8 is selected on the device, no voltage is applied to
the output (0 V). If CO is selected, the voltage applied to this output is 24 V. (see
Chapter Parameters [F9] on page 46)
Socket designation: XG9: C3H8/CO SELEC. VALVE
6.1 Display
The LC display is capable of a maximum pixel resolution of 240 x 128 and
displays maximally 16 lines at 30 characters per line in text mode. It has an inte-
grated back-light which can be switched on and off by pressing the key with the
lamp symbol.
8 3
7 4
6 5
Fig. 11 Membrane keypad
1 2 3
8 7 6
7 Function keys that allow you to display other menus or set the states (this
applies generally for all masks, however the displayed function keys are
not always the same):
F1: Standby. All functions can be canceled using this key. See Chapter
Standby [F1] on page 37
F2: For Manual Operation, see Chapter Manual Operation [F2] on
page 38
F3: For Remote Control via AK interface, see Chapter AK Commands for
Remote Control on page 63
F5: Shows further measured values; see Chapter Measuring Values [F5]
on page 37
8 Device error
If there are more than one pending errors, the error texts are shown one
after another in the bottommost row.
By pressing the [Back] key, you are shown the Flow Mask again.
If you press [F5] again, the second Measuring Values mask is shown. This
one shows the pressure at the outlet, the calculated pressure ratio and the
K factor for the nozzle.
If you press [F5] a third time, you are shown the volume integrators.
6.5.1 General
In Manual Operation mode, settings are adjusted and various activities (e.g. leak
checks) are executed. Via the AK interface, queries are still possible, but no
control functions are possible via AK and no settings can be changed.
The Manual Operation mode can be set in the Flow Mask (see Chapter Flow
Mask on page 36) by pressing [F2]. The Main Menu is shown.
1 2
6 5
Fig. 18 Main Menu
By pressing the [Back] key, you are always taken back to the previous mask.
Pressing the [Main] key always displays the Flow Mask.
In the large field in the middle you are shown all parameters that can be
changed. If there are more than one input fields, the required field is
selected using the up/down arrow keys. The active input field is marked by
a black bar.
To change the value in the marked input field, press the Enter key. The input
field starts showing a flashing cursor (horizontal line underneath the first
character). The cursor is moved by using the right/left arrow keys. It is only
possible to change one value at a time (the value at the position of the
cursor). To enter a value, either enter it directly via the keypad or use the
up/down arrow keys to scroll through the values (this way you can also
enter special characters or letters). Either apply your input by pressing
Enter or press the [Back] key if you do not wish to apply the changed value.
In the smaller field below, a short description of the selected value is shown.
Possible values are shown in square brackets.
The [Back] key takes you back to the previous menu and the Main key to
the main view.
1. This is where you can set the injected volume flow for the AVL iCAL CFO
device. The value must be within the specified range. For information pur-
poses, the standard temperature and standard pressure are shown.
2. Press [F1] or [Back] to return to the Main Menu.
1. This is where you can select a span gas (you can choose between two dif-
ferent gases). If you select a gas number, a description of the gas is shown
in the field below. On system start, the gas set last is used.
2. Press [F1] or [Back] to return to the Main Menu.
This is where operations are performed relating to test location change and leak
check.
Move Device [F1] After activating this option, the device will skip the stabilization time and immedi-
ately switch to normal operation mode the next time it is switched on.
Leak Check [F5] This is where the leak check is started for the device.
Now wait until the device's status (status bar) has changed to SLEC. After this
has been accomplished, the valves at the outlet and inlet have to be closed. Any
errors occurring are displayed.
The "F5 Reset to Default“ function is only shown in the System Operator
access level.
The default settings for passwords are:
Standard user no password
Pause 3333
System operator 4444
These passwords should be changed immediately after commissioning (see
Chapter Diverse Settings on page 56).
Standard User [F1] In the Standard User level no changes can be made to the system configuration
or the parameter settings. No password needs to be entered for this access
level. The Standard User level is set automatically after starting the system.
To deactivate the System Operator access level and access to the corre-
sponding functions, either select the Standard User level or restart the system.
Pause [F3] This function puts the system in Pause state. All power consuming devices are
switched off. The password for Pause must be entered.
The password consists of 4 characters; the default setting is "3333“. For safety
reasons, asterisks are shown instead of characters. When the characters are
entered, each one is replaced by an "x". After you have entered the last char-
acter of the password, a mask is shown that allows you to start the pause:
1. To cancel the process, press the [F2] key. You are taken back to the Access
Level mask (see Fig. 26 on page 44).
2. Press F1 to start the pause. The Main Menu is shown; in the upper right
corner the symbol of the access level "Pause" is shown.
If you wish to terminate the pause, press the [F1] key. The device is restarted.
System Operator In the System Operator access level, all control unit functions are available. To
[F4] switch to this access level, the appropriate password has to be entered (see
Fig. 27 on page 44). After entering the correct password, the Main Menu is
shown and the symbol for the System Operator access level is displayed in the
upper right corner. If an incorrect password is entered, the Access Level menu
(see Fig. 26 on page 44) is shown again and the access level remains the same.
The default setting for the system operator password is "4444“.
Reset to Default This function restores the default setting for the passwords. After that, the Stan-
[F5] dard User access level is set automatically.
WARNING
Danger of explosion
Due to explosions caused by escaping gas, there is a risk of injury and damage
to property.
Connect the gas outlet correctly. The connections and the system to be tested
must not have any leaks.
WARNING
Danger of poisoning
Leaking gas leads to the danger of poisoning.
Connect the gas outlet correctly. The connections and the system to be tested
must not have any leaks.
Select:
1 - for Nozzle 1
2 - for Nozzle 2
3 - for the nozzle combination 1 + 2
To cancel the process, press [F1] or the [Back] key to return to the Parameter
Menu or press the [Main] key to show the Flow Mask.
After selecting the nozzle, you are shown the Nozzle Settings mask.
Calibration This function enables you to display and change the coefficients of the calibra-
tion polynomial for the selected nozzle and a specific gas. First, the gas must be
selected:
1. Select the coefficient you wish to change from a0 to a5. The mask shown in
Figure 30 is displayed.
2. To cancel the process, press the [Back] key and return to the gas selection.
3. To delete the current calibration polynomial, press the [F8] key. The polyno-
mial coefficients are set to the initial values a0=a2=a3=a4=a5=0; a1=1 and
the Parameter Menu is shown again.
4. Enter the new polynomial coefficient. The standard values are: a0 = 0.0; a1
= 1.0; a2 = 0.0; a3 = 0.0; a4 = 0.0; a5 = 0.0
5. The [Back] key will take you back to the previous mask and the [Main] key
will take you to the Flow Mask.
Range Mini- With these two functions you can enter the smallest and the greatest value of the
mum/Range measuring range for the selected nozzle.
Maximum
1. Enter the new minimum and/or maximum for the nozzle measuring ranges.
The measuring ranges for your CFO can be found in Chapter Versions on
page 17.
2. The [Back] key will take you back to the previous mask (see Fig. 31 on
page 47) and the [Main] key will take you to the Flow Mask.
Gas Description After selecting the gas description in the mask shown in Fig. 37, the following
mask is displayed:
Sensor Lineariza- Each sensor can be calibrated using a fourth-degree or lower polynomial:
tion
x' = a0 + a1x + a2x2 + a3x3 + a4x4
The standard values are: a0, a2, a3, a4 = 0; a1 = 1 (no correction).
Fig. 42 Menu for setting the polynomial coefficients for sensor linearization
Sensor Maximum Enter the new maximum. The standard value is 600 kPa.
Sampling Interval The value for the sampling interval indicates at which time interval the measured
values are sampled by the sensors. The standard value is 100ms.
Averaging The averaging value indicates the number of individual measured values the
displayed average is calculated from.
In whichever Sensor Settings sub-menu you may be, the [Back] key will always
bring you back to the mask shown in Fig. 41, and the [Main] key will take you
back to the Flow Mask.
Information
The Sampling Interval and Averaging values can affect flow control speed and
should therefore not be changed.
Units This function allows you to change the units with which the measured values are
to be displayed. Please note that it will only change the value output displayed;
these settings have no effect on the entry of parameters via LCD and the AK
interface.
Select the unit you wish to change by using the arrow keys. Press Enter and use
the arrow keys to select the unit you require. Press the Enter key again to apply
your selection.
Standard settings:
After choosing the language, the Flow Mask appears and the Standard User
access level is set.
Update Rate This parameter allows you to set at which time interval the measuring values
mask is updated.
The values on the Nozzle Switching mask indicate which nozzle is switched.
Until the flow rate specified under Nozzle1 is reached, the first nozzle remains
switched. At higher flow rates, yet no higher than the flow rate specified under
Nozzle2, the second nozzle remains switched. For flow rates any higher, both
nozzles are switched.
The value is specified in slpm separately for both gases.
Select the password you wish to change and press the Enter key. A mask will
appear where you can enter a new password.
1. Enter the new password. The password must consist of 4 characters. The
default settings are:
– Pause 3333
– System operator 4444
2. The [Back] key will take you back to the mask shown in Fig. 51 and the
[Main] key will take you to the Flow Mask.
Standardization The Standardization Base panel specifies the standard values for pressure and
Base temperature which are used to calculate the flow rate. There are two standard-
ization bases, the Factory Base and the Conversion Base.
Information
The calibration data for the AVL iCAL CFO are based on the Factory Base. If the
Factory Base values are changed, the device has to be re-calibrated.
By default, the Conversion Base values are the same as the Factory Base
values. If the conversion values are changed, the flow rate is standardized to the
set pressure and temperature values. This makes it easier to use the AVL iCAL
CFO with devices that have a standardization base other than the AVL iCAL
CFO calibration.
1. The discrete field specifies whether the values apply for the Default Base or
the Conversion Base. To toggle between the two bases, select the discrete
field with the arrow keys and set the required standardization base using
the arrow keys. Confirm your choice with the Enter key.
2. Then choose whether the temperature or the pressure value is to be
changed. Press [F1] or [F2] or select the required function using the arrow
keys. The permissible range for temperature and pressure is shown in the
Help line whenever the respective function is selected.
3. Press the Enter key to show the mask for entering the new value. The input
masks for temperature and pressure look similar.
1. Enter the new standard temperature and/or the new standard pressure. The
standard values are 293.15 K for temperature and 101.325 kPa for pres-
sure.
2. The [Back] key will take you back to Fig. 53 and the [Main] key will take
you to the Flow Mask.
Duration of This function enables you to set the Max. Measurement Time and the Flow
Operation Stabilization Time. (The Flow Stabilization Time function is described under
Stabilization Criterion on page 59.)
After selecting a function and pressing the Enter key, a mask is shown where
you can enter the relevant value.
1. The standard value for Max. Measurement Time is 600 s and the standard
value for Flow Stabilization Time is 10 s.
2. The [Back] key takes you back to the previous mask (Fig. 55) and the
[Main] key takes you to the Flow Mask.
Stabilization The Stabilization Criterion specifies the maximum permitted deviation of the
Criterion actual flow rate from the demand flow rate. If this tolerance is exceeded, the
actual flow rate is output with the prefix '#'. The stabilization criterion is used both
for setting the flow rate and during the measurement.
The stabilization criterion has the following effects:
While the new flow rate is being set, the AVL iCAL CFO displays the status
SDST (setting flow rate).
The system waits until the deviation of the actual flow rate from the demand
rate is within the value specified as stabilization criterion. During this period,
the actual flow rate is displayed with the prefix '#'.
Now the stabilization time starts (see parameter Duration of Operation on
page 58). The flow rate is considered stable if, during this period, the devia-
tion of the actual value from the demand value is within the tolerance speci-
fied by the stabilization criterion. The '#' character disappears from the
output as soon as the flow rate is within the tolerance range. The SDST
status, however, indicates that the level setting is not yet complete.
Once the flow rate is stable, the device switches to the SMGA status
(starting measurement).
The stabilization criterion also determines the maximum permissible deviation of
the actual flow rate from the demand value during a measurement. If this toler-
ance is exceeded, the output flow rate value has the prefix '#'.
1. If you activate the "Erase EEPROM" option, all parameter values saved in
the EEPROM are deleted and the standard values are loaded. This also
applies to the calibration data.
2. The [Back] key will take you back to the Diverse Settings menu (Fig. 50)
and the [Main] key takes you to the Flow Mask.
Host Interface This function allows you to change the parameters of the serial and network
interface (TCP/IP, optional) to the host computer.
Fig. 59 Mask for setting the parameters for the host interface
COM [F2]
Fig. 60 Mask for setting the parameters for the serial interface
The serial interface parameters have to be entered in the following order: Baud
Rate, Parity, Data Bits, Stop Bits, Flow Control. The standard setting is:
"19200,N,8,1,0" (19200 baud, no parity check, 8 data bits, 1 stop bit and flow
control inactive)
TCPIP [F3]
Fig. 61 Mask for setting the parameters for the TCP/IP interface
To use the network, it is first necessary to set the TCP/IP address and the mask.
The port required for AK communication is also adjusted here. New settings are
not applied until the device is restarted.
1. The [Back] key will take you back to the Main Menu.
2. Pressing the Enter key or the [Main] key displays the Flow Mask.
Byte Content
1 STX
2 Don’t Care
3..6 Command (e.g. "SREM")
7 .. 7+n-1 Command parameter (e.g. "K0")
7+n ETX
After an AK command, there must always be a blank and the command parame-
ters must also be separated by blanks. Example: AMBE K0 1
After such a command is sent to the control unit, the control unit returns a
response. The response format is always the same, regardless of which type of
command was sent:
Byte Content
1 STX
2 Don’t Care
3..6 Command (e.g. "SREM")
7 Blank
8 Error status
9.. 9+n-1 Response parameter
9+n ETX
The error status code is 0 if there is no error in the system. Every time the
system errors change, the status code increases by 1 (also if the error was elim-
inated and there is still at least one error left in the system). If the status code is
9 and the error status changes with at least one error left in the system, it is reset
to 1.
7.2.6 AIKO Main Volume Integral Value since SINT or since Last AIKO Command
Information
The following table contains a list of all error codes of the control unit. At the
occurrence of an error, the error codes are sent via the AK interface (see ASTF
Error on page 67) and the error text is shown in the bottom line of the display
(see Chapter Flow Mask on page 36)..
1. SREM K0
2. STBY K0
3. SDST K0 1.55
4. SGAS K0 1
5. SMGA K0
6. Wait for status SMGA in ASTZ.
7. Fill bag, wait
8. STBY K0
8 Calibration
Intervals The calibration of the nozzles is valid for 1 year; then (in accordance with ISO
9000) the device has to be inspected and, if necessary, re-calibrated. The cali-
bration results are recorded in the calibration report (see example Fig. 63) and
are kept together with the device.
Information
The current device flow rate is calculated based on pressure, temperature and
the K-Factor (see Chapter Volume Flow Calculation on page 26)
The nozzles' valid flow ranges can be queried using the AK command (ASWP
Nozzle Switching on page 68) or manually via the LCD/keyboard (Nozzle
Switching on page 56).
Example
Nozzle 1: 0.2 - 0.6 lpm
Nozzle 2: 0.4 - 1.3 lpm
Switching point = (0.6 + 0.4 ) lpm / 2 = 0.5 lpm
I.e.
Nozzle 1: 0.2 < flow rate <= 0.5
Nozzle 2: 0.5 < flow rate <= 1.3
Nozzle 1 + 2: Flow rate > 1.3
NOTICE
The device must be sent in for maintenance once a year. Maintenance jobs must
be performed exclusively by AVL personnel. The device contains no installed
parts that need to be replaced by the user.
Provided that there are no malfunctions at the time, the annual maintenance of
the iCAL CFO is carried out according to the following scheme:
Eliminate any existing errors
Perform the jobs in the maintenance list
Electrical testing
Physical testing
For the maintenance itself, a maintenance report is created.
T Testing
C Cleaning
S Setting
Information
NOTICE
Prior to operate the device after a repair or service schedule, at least once a
year, a leak check has to be executed. A harmless gas (e. g. nitrogen) has to be
connected and used instead of the injection gas (C3H8/CO) during this. For infor-
mation how to execute a leak check, see Operations [F3] on page 42
Equipment Return AVL Emission Test Systems takes back its measuring systems provided the AVL
Emission Test Systems return conditions are fulfilled.
The take-back obligation applies to equipment which is operated within the EU.
AVL Emission Test Systems does not bear the costs for the dismantling and the
transport.
Individual components, e. g. IT parts can be sent to AVL Emission Test Systems
after consultation with the Technical Service.
Disposal of our equipment using the municipal collection points is basically not
permitted.
Please bear in mind that the equipment will be dismantled manually and the
disposal personnel must not be endangered in doing so.
Return Conditions Liquids (condensate bottle if present) must be removed and disposed of by the
customer. Service gas lines must be decoupled and the measuring system
prepared for the transport.
The customer is obligated to inform AVL Emission Test Systems about this in
writing before shipment.
Non-EU Countries Take power tools, accessories, and packaging to collection points where they will
be recycled in an environmentally safe manner in accordance with the environ-
mental regulations applicable in the country of operation.
Substances For a list of the substances contained in the devices, please contact AVL.
contained in the
equipment
11 Technical Data
Tab. 10
Tab. 10
Index
A linearization … 52
access level load default setting for passwords … 45
setting … 44
average … 53 M
main key … 33
B main menu … 39
manual operation
back key … 33
general operation … 38
MD … 14, 17
C measuring values … 37
calibration … 47 membrane keypad … 33
Calibration Date … 64
calibration polynomial … 47, 64 N
change of test location … 42
command format … 63 network interface … 30, 62
control commands … 68 nozzle settings … 46
D O
discrete selection fields … 40 operations … 42
display … 33
display settings … 54 P
displayed units … 54 parameter menu … 46
Disposal … 77 password
duration of operation … 58 change … 56
entering … 44
E pause … 44
EEPROM operations … 60
entering access authorization … 44 Q
error query commands … 64
via AK … 71
Ethernet … 30, 62 R
range … 52
F remote control … 63
flow mask … 36 RS232 … 30, 61
flow rate
setting … 41 S
safety messages … 5
G sampling interval … 53
gas selecting span gas … 42
connection … 29 sensor settings … 51
density … 50 serial interface … 30, 61
description … 50 Serial Number … 67
settings … 49 setting commands … 70
standard
H pressure … 57
temperature … 57
HD … 14, 17
host interface … 60, 63 standard user … 44
system operator … 44
I
T
Initial Commissioning … 77
TCP/IP interface … 30, 62
typographic conventions … 14
K
keypad … 33 U
update rate … 55
L
language … 55 V
LD … 14, 17
LD-LE … 14, 17 version … 34
leak check … 43 versions … 62
volume flow
calculation … 26
setting … 31
volume integrators … 38
W
warranty … 13