Supplement 10 TS Calibration ECSPP ECBS
Supplement 10 TS Calibration ECSPP ECBS
598
Supplement 10
WHO Vaccine
Annex 9: Model guidance for the storage and transport of time and
temperature–sensitive pharmaceutical products
August 2014
`
Health Organization be liable for damages arising from its use. The named authors alone are responsible for the views
expressed in this publication.
`
Acknowledgments
The author of this document is Claude Hafner of Berlinger & Co. AG., Switzerland.
Technical Supplement: Checking the accuracy of temperature control and monitoring devices 1
Contents
Acknowledgments ................................................................................................................................ 1
Contents ................................................................................................................................................... 2
Abbreviations ........................................................................................................................................ 3
Glossary ................................................................................................................................................... 4
1. Introduction .................................................................................................................................. 5
1.1 Requirements ...................................................................................................................................... 5
1.2 Objectives.............................................................................................................................................. 6
1.3 Target readership .............................................................................................................................. 6
2. Guidance ......................................................................................................................................... 7
2.1 Associated materials and equipment ........................................................................................ 7
2.2 Procedure ............................................................................................................................................. 8
2.2.1 Prerequisites ............................................................................................................................................ 8
2.2.2 Establishing the ice-point bath (excerpt from ASTM E563-11) ...................................... 8
2.2.3 Placing the device in the bath.......................................................................................................... 9
2.2.4 Carrying out the accuracy check, step-by-step ..................................................................... 10
2.2.5 Maintaining the bath temperature ............................................................................................ 10
2.2.6 Actions to take following the test ............................................................................................... 11
References ............................................................................................................................................12
Annex 1 - Generic temperature accuracy check form ...........................................................13
Revision history ..................................................................................................................................14
Technical Supplement: Checking the accuracy of temperature control and monitoring devices 2
Abbreviations
DUT Device Under Test
SOP Standard Operating Procedure
TTSPP Time and Temperature-Sensitive Pharmaceutical Product
Technical Supplement: Checking the accuracy of temperature control and monitoring devices 3
Glossary
Standard Operating Procedure (SOP): A set of instructions having the force of a
directive, covering those features of operations that lend themselves to a definite or
standardized procedure without loss of effectiveness. Standard operating policies and
procedures can be effective catalysts to drive performance improvement and improve
organizational results.
Temperature control device: A device which actively controls the operation of cooling
plant used to store or transport TTSPPs.
Temperature monitoring device: A device which monitors the temperature of spaces
used to store or transport TTSPPs.
Thermal time constant: The most common definitions of the thermal reaction time are
the so-called "Tau" (τ, the 19th letter of the Greek alphabet) and "T90". "Tau" stands for
the time a device needs to adapt to 63% of the end value of a temperature change whereas
"T90" represents the time to adapt to 90% of the change. "T90" approximately equals to
2.5 * "Tau". These constants are commonly evaluated by experiment on the test device
under well-defined conditions as described in EN12830.
Time and temperature sensitive pharmaceutical product (TTSPP): Any
pharmaceutical good or product which, when not stored or transported within pre-
defined environmental conditions and/or within pre-defined time limits, is degraded to
the extent that it no longer performs as originally intended.
Triple point: The temperature and pressure at which a substance can exist in equilibrium
in the liquid, solid, and gaseous states. The triple point of pure water is at 0.01 degrees
Celsius and 4.58 millimetres of mercury and is used to calibrate thermometers.
Technical Supplement: Checking the accuracy of temperature control and monitoring devices 4
1. Introduction
This technical supplement has been written to amplify the recommendations given in
Section 4.10 of WHO Technical Report Series No. 961, 2011, Annex 9: Model guidance for
the storage and transport of time- and temperature-sensitive pharmaceutical products1. It
describes a way to check the accuracy of temperature monitoring and temperature control
devices using the “ice-water” procedure, and it outlines the limitations of this approach.
This method can be used in situations where it is not possible to carry out a full three
point calibration using the services of a of a nationally or internationally accredited
calibration laboratory.
1.1 Requirements
As a general rule, temperature measurement and control devices must periodically be
calibrated in order to prove their accuracy over the full operating temperature range and
according to the device’s data sheet definitions. Proven accuracy is mandatory because
inaccurate readings can lead to a false sense of security and place TTSPPs at risk.
Some devices are covered by calibration certificates from the device manufacturer. These
certificates are valid for a defined period of time and the associated devices may be used
throughout this period without additional calibration. An example of this is a single use
temperature monitoring device which is designed to be discarded at the end of a journey,
or when the battery powering the device expires. In order to ensure conformity, the
manufacturer of such products must supply a calibration certificate with the device.
However, there are circumstances under which proper device calibration or re-calibration
is needed. The list below is not comprehensive but illustrates some of these
circumstances:
A calibration certificate is not available because it has been lost.
The device is used for longer than the period covered by the calibration certificate,
either at the user’s risk or with the approval of the manufacturer.
The device was used or treated beyond the manufacturer's data sheet limitations
(e.g. excessive temperature, shock, etc.).
The battery powering the device was replaced.
The device's measurements are suspect.
The device manufacturer specifies that a calibration procedure should be carried
out at regular intervals.
Regulatory bodies require regular proof of calibration - e.g. at 12 month intervals –
and proof of calibration cannot be provided by the manufacturer’s certificate.
1 http://apps.who.int/medicinedocs/documents/s18683en/s18683en.pdf
Technical Supplement: Checking the accuracy of temperature control and monitoring devices 5
1.2 Objectives
Wherever possible, calibration should be carried out in accordance with the device
manufacturer’s instructions, or by following a device-specific SOP. Ideally, a full three-
point calibration should be carried out by a nationally or internationally accredited
calibration laboratory with proven accuracy standards and appropriate equipment.
However, there are many circumstances where accredited calibration is not possible
because no suitable laboratory is available. The simple and accurate method described in
this Technical Supplement can be used to prove the device’s functionality and accuracy at
one single point of temperature using the so-called "ice-water procedure".
Technical Supplement: Checking the accuracy of temperature control and monitoring devices 6
2. Guidance
The method described below is relatively simple to carry out but requires close attention
to detail. The accuracy of the results are also dependent upon the use of a high quality
reference thermometer with a valid calibration certificate.
Technical Supplement: Checking the accuracy of temperature control and monitoring devices 7
protection class rating because the protection might have been damaged – for
example by dropping or battery replacement. Where the DUT has an external
sensor, the sensor may be immersed directly in the bath, provided it has an IP7 or
IP8 protection class rating. Otherwise it should be sealed in a pouch as described
above, with the pouch tied tightly around the lead above the level of the ice-water.
2.2 Procedure
The ice-water bath provides an accurate reference temperature at 0.0°C if the melting ice -
water mixture is properly set up, handled and maintained. An accurate temperature is
achieved by this method because an ice-water mixture in a container which is open to the
atmosphere will stabilize at its own “triple point”. At this point all three aggregate states of
water coexist: liquid, solid and gaseous. For more physical details refer ASTM E563-11.
2.2.1 Prerequisites
a. Only place clean equipment and distilled water inside the container. Use clean
latex gloves to handle ice and equipment.
b. Although the temperature of the ice-water mixture stabilizes itself, its temperature
must still be monitored before and during the preocedure using the calibrated
reference thermometer. This is a mandatory requirement in order to prove the
functionality and stability of the ice-water bath.
c. During extended testing periods enough ice must be added to the bath to maintain
the water–ice equilibrium temperature and to prevent a possible temperature rise
caused by excessive ice melting.
d. Note that only DUTs which display a temperature reading can be checked by this
method because the temperature indicated by the DUT must be compared to the
ice-water bath temperature. Devices without a display need additional equipment
to capture an immediate measurement read-out so that this can be compared with
the reference temperature reading. This additional equipment is manufacturer-
specific.
Technical Supplement: Checking the accuracy of temperature control and monitoring devices 8
e. Alternately add shaved ice and chilled water to the vessel, using just enough water
to saturate the ice but not enough to float it. As the vessel fills, compress the ice-
water mixture to force out excess water. The objective is to surround each particle
of ice with water, filling all voids, but to keep the ice particles as close together as
possible. Continue adding ice and water and compressing until the vessel is filled
to the required level. Decant or siphon off excess water.
f. Use the reference thermometer throughout the entire test period to confirm that a
stable temperature is maintained.
g. Cover the ice-point bath to protect it during the test period. Use an opaque and
thermally insulating cover or stopper that is suitable for the application. This
reduces heat transfer to the ambient environment through the surface of the bath.
Allow the bath and vessel to equilibrate for at least 30 minutes before using.
Technical Supplement: Checking the accuracy of temperature control and monitoring devices 9
Figure 2 – Ice-water bath arrangement
Technical Supplement: Checking the accuracy of temperature control and monitoring devices 10
the ice slush column always extends to at least 30mm below the lowest point of the
test object.
b. In order to sustain the ice point over prolonged periods, the ice-point bath can be
immersed in another larger insulated bath that is kept near to 0.0°C.
3 For example, if the reference instrument has an accuracy of ±0.5°C and the DUT deviates 0.3°C
from the reference instrument reading, this is a pass.
4 For example, if the reference instrument has an accuracy of ±0.5°C and the DUT deviates 0.6°C
Technical Supplement: Checking the accuracy of temperature control and monitoring devices 11
References
ASTM E563-11 Standard Practice for Preparation and Use of an Ice-Point Bath as a
Reference Temperature
EN 12830:1999 Temperature recorders for the transport, storage and distribution of
chilled, frozen, deep-frozen/quick-frozen food and ice cream. Tests, performance and
suitability.
IEC 60529: Consolidated Edition 2.1 (incl. am1): Degrees of protection provided by
enclosures (IP Code).
WHO. EVM-SOP-E2-2. Checking the accuracy of temperature monitoring devices.
Available online at:
http://www.who.int/immunization/programmes_systems/supply_chain/evm/en
/index2.html
WHO Technical Report Series No. 961, 2011, Annex 9: Model guidance for the
storage and transport of time- and temperature-sensitive pharmaceutical
http://apps.who.int/medicinedocs/documents/s18683en/s18683en.pdf
Technical Supplement: Checking the accuracy of temperature control and monitoring devices 12
Annex 1 - Generic temperature accuracy check form
Store location: …………………………………………………………………………......
Form start date: …………………………………………………………………………....
Form finish date: .…………………………………………………………………………..
Date Device Device Reference Check Reference Difference from Initials Comments Verified
description location thermometer method temperature reference by/ date
12 Dec 2013 DUT Cold room # 1 XYZ 500 Water-ice 0.0°C +0.2°C UK checked - OK SJ 15
#12341234 bath Dec2013
Technical Supplement: Checking the accuracy of temperature control and monitoring devices 13
Revision history
Date Change summary Reason for change Approved
Technical Supplement: Checking the accuracy of temperature control and monitoring devices 14