S P E C I F I C A T I O N S

Liquid Chromatography

Flexar UV/Vis LC
and UHPLC Detectors

Flexar UV/Vis LC Detector Flexar FX-UV/Vis UHPLC Detector


CDS control environment Chromera and TotalChrom
® ®
Chromera only
Wavelength range 190-700 nm 190-700 nm
Wavelength accuracy ±1 nm ±1 nm
Wavelength precision ± 1 nm ± 1 nm
Bandwidth 5 nm 5 nm
Linearity ≥ 2.5 AU (with 5% deviation) ≥ 2.5 AU (with 5% deviation)
Optics Dual beam Dual beam
Maximum sampling rate 50 pt/sec 100 pt/sec
Sensitivity range 0.0005 to 3.000 AUFS in 0.0001 increments from 0.0005 to 3.000 AUFS in 0.0001 increments from
0.0005 to 0.1, and 0.01 increments above 0.1 AUFS 0.0005 to 0.1, and 0.01 increments above 0.1 AUFS
Noise <7.5 x 10-6 AU, 210-280 nm, 2 sec response time, <7.5 x 10-6 AU, 210-280 nm, 2 sec response time,
standard test cell standard test cell
Drift <1 x 10-4 AU/hr, after warmup <1 x 10-4 AU/hr, after warmup
Light sources Deuterium (190-360 nm) or Tungsten Deuterium (190-360 nm) or Tungsten
(360-700 nm), pre-focused, no adjustment (360-700 nm), pre-focused, no adjustment
required on replacement required on replacement
Flow cell Flow cell sold separately; compatible 2.4 µL, 6 mm pathlength standard; compatible
with PerkinElmer flow cells: 6 mm x 2.4 µL, with PerkinElmer flow cells: 6 mm x 2.4 µL,
8 mm x 12 µL, 10 mm x 15 µL and 2 mm x 1.7 µL (prep) 8 mm x 12 µL, 10 mm x 15 µL and 2 mm x 1.7 µL (prep)

Materials contacting solvent Kel-F, PEEK, quartz Kel-F, PEEK, quartz,


Leak detection Heated thermistor sensor in glass envelope, Heated thermistor sensor in glass envelope,
located in flowcell drip pan located in flowcell drip pan
Dimensions (HxWxD) 16 x 34 x 53 cm (6.5 x 13.5 x 21 in) 16 x 34 x 53 cm (6.5 x 13.5 x 21 in)
Weight 11.4 kg (25.2 lbs) 11.4 kg (25.2 lbs)

PerkinElmer, Inc.
940 Winter Street
Waltham, MA 02451 USA
P: (800) 762-4000 or
(+1) 203-925-4602
www.perkinelmer.com

For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs