0% found this document useful (0 votes)
69 views

G5 Specification Rev.1

The document provides specifications for the SG Healthcare G5 high-performance color ultrasound system. It describes the system's advanced imaging technologies, standard and optional configurations, operating modes, display options, and measurement and reporting features. Key technologies include full digital beamforming, wide dynamic range, and tissue harmonic imaging. The system is intended for a wide range of clinical applications.

Uploaded by

mhd
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
69 views

G5 Specification Rev.1

The document provides specifications for the SG Healthcare G5 high-performance color ultrasound system. It describes the system's advanced imaging technologies, standard and optional configurations, operating modes, display options, and measurement and reporting features. Key technologies include full digital beamforming, wide dynamic range, and tissue harmonic imaging. The system is intended for a wide range of clinical applications.

Uploaded by

mhd
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 14

rev.

0 (27th May 2013)

Specifications for G5 High-Performance Color System


rev.0 (27th May 2013)

General Specification PW Mode

The high performances of the SG Healthcare G5 THI Mode

stem from the advanced ultrasound Doppler Cardiac Package

imaging technologies that include full digital Gynecology Package

beam-former, wide dynamic range, multi-beam Urology Package

processing, etc. Small Part Package

The ergonomic user-friendly design enables user Vascular Package

to customize the system according to the Orthopedic Package

specific application needs, and the graphic exam Compound Imaging

icon assure you familiar with the system in few Trapezoid Imaging

minutes. Dop Auto Trace

IMT Function

2. Advanced Technologies Dicom 3.0

PC platform Internal Workstation

Full Digital Beam Processing

Compound Imaging 4. Optional Functions


Tissue Harmonic Imaging μ-scan
Panoramic Imaging 4D Imaging

Trapezoid Imaging Panoramic Imaging

4D Imaging Anatomic M Mode

Dicom 3.0 CW Mode

Intelligent Upgrade Triplex Mode

Graphic Exam Icon

Remote Maintenance 5. Optional Components


UPS Power

3. Standard Configuration Biopsy Guide

B Mode Color Ink-jet Printer

Multi-B Mode B/W Video Printer

M Mode Foot Switch

DPI Mode

page 2 of 14
rev.0 (27th May 2013)

6. Sweep Width/Angle 9. Operating Modes


Linear Array: Maximum 46mm B-Mode

Convex Array: ≥70° M-Mode

Phased Array: ≥90° THI-Mode

Micro convex Array: ≥135° Color Flow Mode(CFM)

4D probe: 70° Doppler Power Imaging(DPI)

PW Mode

7. Scanning Methods CW Mode(optional)

Electronic Convex Sector Panoramic Imaging(optional)

Electronic Linear 4D Imaging(optional)

Electronic Phased Array Sector Steer M-Mode(optional)

4D Automatic Scanning

Electronic Endocavity 10. Display Modes


B、Dual B、4B

8. Applications B+Color、Dual B+Color

Small Parts(Thyroid, Brest, Testicles and B+DPI、Dual B+DPI

Superficial) M、B+M、Dual B+M

Abdominal(Liver, Spleen, Kidney, Pancreas) B+PW、Dual B+PW

Vascular(Carotid, Peripheral vessel) B+color+PW

Cardiology B+DPI+PW

Obstetrical(Uterus, Appendages, Fetus) B+CW、Dual B+CW(optional)

Gynecological B+color+CW(optional)

Urological B+DPI+CW(optional)

Musculoskeletal B+ Anatomic M(optional)

Craniocerebral Panoramic Imaging(optional)

Emergency(optional) Trapezoid Imaging

ICU(optional) Compound Imaging

Anesthesia (optional) 4D Imaging(optional)

page 3 of 14
rev.0 (27th May 2013)

11. System Menu Setting Storage Mode:depending on the

System Setting frames

General Setting Storage Region:Full Screen, Imaging

Hospital Name Area

Language Setting Save to USB

English Peripheral Equipments

Simple Chinese Clip Format:NTSC、PAL

Spanish(optional) Internet setting

Russian(Optional) IP Adress

French(Optional) Subnet Mask

Italian(Optional) Default Gateway

German(Optional) DNS Service

Turkish(Optional) Local DICOM AE Title

Automatic Response to Freeze Printer Setting

Trackball Sensitive Annotation Setting

Set Time/ Date Annotation Library Setting

Letter form Display Language:Current Language,

Audio Volume English

Display Self-defined Entry

TGC Curve Body Mark Setting

Patient Information Body Mark Library Setting

Clipboard Automatic Open to Freeze Default Display Position

Parameters Setting Measurement Setting

B Mode General Setting

M Mode Unit:Metric, British

Color Mode Cardiac Cycle

DPI Mode UA Result:Last, Minimum, Maximum,

PW Mode Average

CW Mode Define Quick Key(Obstetric, Cardiac)

Storage Calculation Menu Setting

page 4 of 14
rev.0 (27th May 2013)

2D Mode Cardiac

Distance Measurement Method Setting

Trace BSA: Eastern, Western

Ellipse Obstetric Measurement Method

Angle EFW Method

Volume Report Setting

Ratio Report Logo

Vascular Title, Font Size

Obstetric Display Items

Gynecological DICOM Setting

Cardiac DICOM Image Storage

Small Part DICOM Storage Commitment

Urologic Worklist

Orthopedic MPPS

M Mode DICOM Printing

Time System Information

Distance Software Version

Slope Restore to Factory Defaults

Ratio System Upgrade:USB Upgrade, Remote

Heart Rate Network Upgrade

Left Ventricle

Mitral Valve 12. Standard Features


Aortic Valve Frame Rate: Max 800frames/sec or

PW Mode more(configuration dependent)

Flow Velocity Display Gray Scale: 256 Levels

Time Probe Elements:Up to 128

Heart Rate

Manual Trace 13. B Mode


Vascular Focus: Up to 15, Focus span adjustable

Obstetric Line Density: High/ Med

page 5 of 14
rev.0 (27th May 2013)

Dynamic Range: 20-320 (probe dependent) Flow Invert: ON/OFF(optional)

GSC(Gray Scale Curve): 21 steps selectable Left/Right: ON/OFF

Gain: 0-255 adjustable Resolution: 6 kinds

Left and Right Inversion

Up and Down Inversion 15. M Mode


TGC(Time Gain Control): 8 slide controls Chroma: 8 kinds selectable

Depth: Max 33.3cm(probe dependent) Enhancement: 8 kinds selectable

Frequency Range: 5 steps Video Invert: ON/OFF(optional)

Persist: 8 steps selectable Sweep Speed: 2, 4, 6, 8 sec/plane

Chroma: 9 steps selectable Steer M:3 sample lines, Display frame

SEC.WIDTH: adjustable rate(optional)

Biopsy Guide: (optional) Display Format:H1/1、H1/2、H2/1、V1/1、

Biopsy Angle adjustable V1/2

Biopsy Offset adjustable

Power:1%-100% adjustable,5 steps each 16. Spectral Doppler


B Steer Mode(Linear, optional) Chroma: 8 kinds selectable

Spectrum Inversion

14. Color Flow Mode PRF:0.7~5.7kHz(PW)

Frame Rate:≥50 frames/sec PRF:1-48KHz(CW)

Color Area Size and Position: adjustable Baseline Shift:15 steps

Single auto focus while color ROI movement Video Invert: ON/OFF(optional)

Steer Angle:0, ±12, ±16, ±20 degree Filter: 25-750 adjustable

adjustable 2D-Refresh(optional)

Persistence:8 steps selectable Doppler Methods

Frequency Range: 5 steps PW(pulsed wave) Doppler

PRF:0.5-5.7kHz(probe dependent) CW(continuous wave) Doppler

Color Baseline:±15 steps Doppler Envelope: manual trace, auto trace,

Filter: 5-750 adjustable real time trace, freeze trace

Color/Power Map:15 kinds Max Velocity Range:

B Reject:0-255 0.0004-40.9 m/s(PW)

page 6 of 14
rev.0 (27th May 2013)

0.0013-49.1 m/s(CW) 19. Storage of Images and Cine


Dynamic Range: 10 steps selectable Cine loop: 999 frames or more

Frequency Range: 5 steps Cine loop time:99 seconds or more

θ Angle Correction: 0-80° Archived image can be viewed on PC

Sample Volume Size for PW Doppler: 1 -20 Cropboard function: in Freeze Mode

mm Cine play back mode for Dop.

Sweep Speed: 2, 4, 6, 8 sec/plane

Steer Angle: 0、±12、±16、±20,7 kinds 20. DICOM Network Communication


Display Format:H1/1、H1/2、H2/1、V1/1、 Structural Report(SR): ABD、OB、GYN、

V1/2 Cardiac、 Vascular、 Urology、 Small Parts、

PED、MSK、 Nerve、 Orthopedics

17. 4D Mode (optional) DICOM Image Storage Format: RLE,JPEG

Render Mode: Vol、MaxIP DICOM Remote Storage

Auto Rotate (45、90、180、270、360 degrees DICOM ECHO

adjustable ) DICOM Storage Commitment

Zoom Function DICOM Work list

Rotate Function DICOM MPPS

Color Map:4 kinds DICOM Print

Image Quality: high、med、low DICOM Media Storage

4D Stabilization

Save images 21. Data Communication


Print All patient data in this ultrasound unit can be

outputted to Flash Disk and DICOM, in file

18. Storage Media folder of patient, examination mode, image

Hard Disk memory capacity: 320 G and report.

Storage media: USB Drive

USB Drive 22. Pre-set Function


DVD RW Users can optimize the Pre-set by changing

different parameters for different probes and

examination.

page 7 of 14
rev.0 (27th May 2013)

27. Safety Standard


23. Patient Data Management Meet with IEC60601-1, Class I,BF

Patient Registration: Name, ID, Gender, Date international standard

of Birth, Height, Weight, Blood Pressure,

Body Surface Area, Heart Rate, LMP, GA, 28. Environmental Requirements
Doctor’s comment In operation

Retrieve the patient from Worklist Temperature:+10 to +40 degrees C

Patient Data, report, images can be searched, Relative Humidity: 30% to 75%

retrieved, DICOM sending, Send to Report, (non condensing)

Print and so on Atmospheric pressure: 700 to 1060hPa

In Storage/Transportation

25. Annotation and Body Mark Temperature: -20 to +55 degrees C

Body Mark Relative humidity: 20%- 90%

Divided into different applications (non condensing)

User-definable annotation Atmospheric Pressure: 700 to 1060hPa

Annotation can be revised, moved, edited Power Supply

and deleted 100/220 Volts AC, 3.15Amps

Body mark can be moved, while the probe Frequency: 50/60HZ

mark can be moved and rotated

Arrow can be rotated and the size could be 29. Optional Probe
changed Linear Probe(Vascular, Small Parts)

→L741 (5-10 MHZ)

25. Probe Connectors Linear Probe(Vascular, Small Parts)

Active Connectors: 3 connectors →10L1 (6-12 MHZ)

Linear Probe(Intraoperative)

26. Display Screen →10I2 (5-10 MHZ)

15-inch High-Resolution Color LCD monitor Curved Probe(Abdomen, OB/GYN)

→C311 (2-6 MHZ)

Curved Probe(Abdomen, OB/GYN)

→C344 (2-6 MHZ)

page 8 of 14
rev.0 (27th May 2013)

Curved Probe(Abdomen, OB/GYN)

→C354 (2-6 MHZ) On Spectral Doppler

Curved Probe(Abdomen, OB/GYN) →Time

→C611 (4-8 MHZ) →Heart Rate

Micro-curved Probe(Transvaginal) →Velocity

→6V1 (4-8 MHZ) →AcceleraQon

Endocavity Probe →Resistance Index

→EC9-5 (5-9 MHZ) →Pulse Index

Volumetric Probe (4D) →PV (Peak Velocity)

→VC6-2 (2-6 MHZ) →PG (Pressure gradient)

Phased Array Probe(Cardiology) →Auto Trace

→2P1 (2-4 MHZ) →Manual trace

Phased Array Probe(Cardiology) →Mean Flow Velocity

→5P1 (4-7 MHZ) →End Diastolic Velocity

→Pressure Half Time

30. Measurements/Calculations
General Measurements/Calculations Obstetrical/ Gynecological Measurements &

On B-Mode Calculations

→Distance(real time、freeze) B Mode

→Area and circumference (Trace, Ellipse) →GS (GestaQonal Sac diameter)

(real time、freeze) →CRL (Crown Rump Length)

→Volume (L×W×H, Area×L) →BPD (Biparietal Diameter)

→Angle →HC (Head Circumference)

On M-Mode →AC (Abdominal Circumference)

→Velocity →FL (Femur Length)

→Time →CER (Cerebellum)

→Heart rate →OFD (Occipitofrontal Diameter)

→Slope →Fibula (Fibula Length)

→Distance →Foot (Foot Length)

→Time RaQo/Distance RaQo →AA (Abdominal Area)

page 9 of 14
rev.0 (27th May 2013)

→APAD (Anteroposterior Abdominal →EFA (EsQmated Fetal Age)

Diameter) →EDD (EsQmated Date of Delivery)

→HA (Head Area) →EFW (EsQmated Fetal Weight)

→Humerus (Humerus Length) →AUA (Average Ultrasound Age)

→Kidney (Kidney Length) →GA (GestaQonal Age)

→APTD (Anteroposterior Trunk Diameter) PW Mode

→OOD (Outer Orbital Diameter) →Umb A (Umbilical Artery)

→Radius (Radius Length) →MCA (Middle Cerebral Artery)

→TAD (Transverse Abdominal Diameter) →Rt Uterin A (Right Uterine Artery)

→TC (Thoracic Circumference) →Lt Uterin A (LeR Uterine Artery)

→THD (Thoracic Diameter)

→Tibia (Tibia Length) Cardiac measurements

→TTD (Transverse Trunk Diameter) B-Mode

→Ulna (Ulna Length) →LeR Ventricular FucQon Measurement

→Umb VD (Umbilical Vein Diameter) Single Plane Ellipse Method

→NT (Nuchal Translucency) LVALd: Left Ventricular Long-axis

→LV (Lateral Ventricle) Area at end Diastole

→UT L (Uterus Length) LVLd: Left Ventricular Long-axis

→UT H (Uterus Height) Length at end Diastole

→UT W (Uterus Width) LVALs: Left Ventricular Long-axis

→Cx (Cervix) Area at end Systole

→En-T (Endometriosis) LVLs: Left Ventricular Long-axis

→Rt OV L (Right Ovary Length) Length at end Systole

→Rt OV H (Right Ovary Height) Biplane Ellipse Method

→Rt OV W (Right Ovary Width) LVALd: Left Ventricular Long-axis

→Lt OV L (LeR Ovary Length) Area at end Diastole

→Lt OV H (LeR Ovary Height) LVALs: Left Ventricular Long-axis

→Lt OV W (Left Ovary Width) Area at end Systole

→AFI (AmnioQc Fluid Index) LVAMd: Left ventricular short-axis

→Follicle area at end diastole

page 10 of 14
rev.0 (27th May 2013)

LVIDd: Left ventricular short-axis LVLs: Left ventricular long-axis

diameter at end diastole length at end systole

LVAMs: Left ventricular short-axis Cube

area at end systole IVSTd: Interventricular septal

LVIDs: Left ventricular short-axis thickness at end diastole

diameter at end systole LVIDd: Left ventricular short-axis

Bullet diameter at end diastole

LVAMd: Left ventricular short-axis LVPWd: Left ventricular posterior

area at end diastole wall thickness at end

LVAMs: Left ventricular short-axis diastole

area at end systole IVLTs: Interventricular septal

LVLd: Left ventricular long-axis thickness at end systole

length at end diastole LVIDs: Left ventricular short-axis

LVLs: Left ventricular long-axis diameter at end systole

length at end systole LVPWs: Left ventricular posterior

Simpson Method wall thickness at end

LVAMd: Left ventricular short-axis systole

area at end diastole Teichholz

LVAMs: Left ventricular short-axis LVLDd: Left ventricular short-axis

area at end systole diameter at end diastole

LVAPd: Left ventricular short-axis LVIDs: Left ventricular short-axis

area at the level of the diameter at end systole

papillary muscle at end Gibson

diastole LVLDd: Left ventricular short-axis

LVAPs: Left ventricular short-axis diameter at end diastole

area at the level of the LVIDs: Left ventricular short-axis

papillary muscle at end diameter at end systole

systole →Mitral Valve Diam

LVLd: Left ventricularlong-axis length →Lv OuSlow Diam

at end diastole →Pul.Valve Diam

page 11 of 14
rev.0 (27th May 2013)

M-Mode →TEI Index Doppler Measurement

→LeR Ventricular FucQon Measurement

Cube Vascular Measurements Calculations

LVIDd: Left ventricular short-axis →ICA (Internal CaroQd Artery)

diameter at end diastole →ECA (External CaroQd Artery)

LVIDs: Left ventricular short-axis →CCA (Common CaroQd Artery)

diameter at end systole →INT IL (Internal iliac)

LVPWd: Left ventricular posterior →EXT IL (External iliac)

wall thickness at end →ILIAC (Common iliac)

diastole →CFA (Common Femoral Artery)

LVPWs: Left ventricular posterior →PROFUN (Profunda)

wall thickness at end →LT CIR (Lateral Circumflex)

systole →SFA(Superficial Femoral Artery)

Gibson →POP (Popliteal Artery)

LVLDd: Left ventricular short-axis →PTA (Posterior Tibial Artery)

diameter at end diastole →PERON (Personal Artery)

LVIDs: Left ventricular short-axis →ATA (Anterior Tibial Artery)

diameter at end systole →DR PED (Dorsalis Pedis)

Teichholz →%A REDUC (Area reduction percent)

LVLDd: Left ventricular short-axis →%D REDUC (Diameter reducQon percent)

diameter at end diastole →PI (PulsaQlity Index)

LVIDs: Left ventricular short-axis →RI (ResisQve Index)

diameter at end systole →S/D (Systolic/Diastolic RaQo)

→Mitral Valve Measurement →PG((Pressure gradient)

→AorQc Valve Measurement →PV(peak Velocity)

PW-Mode →IMT

→Mitral Valve Measurement

→AorQc Valve Measurement Urological Measurements Calculations

→Tricuspid Valve Measurement →LeR Kidney

→Pulmonary Valve Measurement →Right Kidney

page 12 of 14
rev.0 (27th May 2013)

→Bladder Volume Biophysical Profile (optional)

→Residual Urine Fetal Compare (quadruplets)

Urine Area Insert Picture

Urine Height Comment

Urine Volume →Cardiac funcQon report

→Whole Prostate Volume →Vascular report

→LeR-Seminal Vesicles →Urological report

→Right- Seminal Vesicles →Small Part report

→LeR-Testicle →IMT report

→Right- Testicle

→Trans Zone Volume

Small Part Measurements *The specifications are subject to change without

→L-Thyroid notice.

→R-Thyroid

→Thyroid Isthmus

→L-Superior Parathyroid

→L-Inferior Parathyroid

→R-Superior Parathyroid

→R-Inferior Parathyroid

→TesQs

→Seminal Vesicle

Orthopedics Measurements

→HIP(Hip Joint)

Report functions (Editable)


→Obstetrical /Gynecological report

Obstetrical Curve

Fetal Anatomy

page 13 of 14
SG HEALTHCARE CO.,LTD
753 ITECO, 150 JOJEONG-DAERO, HANAM-SI, GYEONGGI-DO, KOREA 465736

TEL.+82 31 790 0647 FAX.+82 31 790 0604 www.sghealthcare.com

page 14 of 14

You might also like