Cobas 6000
Cobas 6000
1. Purpose:
The purpose of this procedure is to provide the instructions of the user manual
Cobas 6000.
2. Definition:
c 501 (ISE): ISE unit of the c 501 module
c 501 (P): photometric unit of the c 501 module
CLAS 2 : Clinical Laboratory Automation System 2
CLIA: Clinical Laboratory Improvement Amendments
DIL: diluent
ECL: electrochemiluminescence
EMC: electromagnetic compatibility
HCFA: Health Care Financing Administration
IEC: International Electrical Commission
IS: internal standard (ISE module)
ISE: ion selective electrode
KVA: kilovolt-Ampere. Unit for expressing rating of AC electrical machinery.
LDL: lower detection limit see analytical sensitivity
LIS: laboratory information system
LLD: liquid level detection
MBC: matrix barcode
MSDS: material safety data sheet
n/a: not applicable
NCCLS:n National Committee for Clinical Laboratory Standards
PC/CC: ProCell M (ProCell) /CleanCell M (CleanCell)
PSM: process system manager (software)
REF: reference solution for ISE module
UL: Underwriters Laboratories Inc.
3. Responsibilities:
Chemistry section Technicians
Laboratory Physician
4. Policy:
Cobas® 6000 analyzer series
5. Procedure:
The cobas 6000 analyzer series is a fully automated, random-access, software-
controlled system for immunoassay and photometric analysis intended for qualitative
and quantitative in vitro determinations using a wide variety of tests. The cobas 6000
analyzer series is a powerful tool for complete diagnostic laboratory automation. It is
optimized for high throughput workloads using an immunoassay analysis module (e 601
module).
Each sample category has its own type of sample rack. Each sample rack
holds a maximum of five samples. To load sample racks, they are either
inserted at the STAT port or placed on a tray and the tray is placed into
the rack loader.
Sample racks are color-coded and equipped with a barcode label and a label with a
unique rack number. Each rack is identified and registered when its barcode label is
scanned by the barcode reader of the rack sampler unit.
The following table explains the different types of racks, their color codes and
numbers:
The table shows that each type of sample rack has its specific range of rack numbers.
The software allows subdividing the range of rack numbers of routine, rerun, and
STAT racks according to sample types (such as Ser/Pl, Urine, CSF…). This
assignment of sample types to rack numbers is displayed in the Rack Assignment
area on Utility > System.
Make sure the rack barcode label is on the right side, when inserting a rack into the
STAT port or when placing a rack into the rack loader. The barcode label can be read
only when racks are supplied in the correct orientation.
Test selections for routine samples are downloaded from the Host.
However, these test selections can manually be changed—regardless
of the mode (Stand By, Stop, Operation, or Sample Stop) the
analyzer is in.
5.2.1.2. Loading routine samples
To load patient samples
1. Prepare sample racks with the correct color and rack
numbers according to sample types (such as Ser/Pl,
Urine, CSF…).
For all routine samples use gray racks.
To verify the correct rack numbers for each sample type
refer to the Rack
Assignment area on the Utility > System screen.
To start processing
1. Choose Start (global button).
Test selection
Test selection for STAT samples is the same as for routine samples. That is,
test selections for samples are downloaded from the Host in real time but can
still be changed from the analyzer’s control unit.
3. Place the STAT rack in the STAT port carefully observing its
correct orientation:
The open slots in the racks must point to the rear of the rack sampler unit.
The side of the STAT rack with the barcode label must point to the
analyzer.
As results are generated on the instrument, they are saved in the database
located on the internal hard disk of the control unit. When all results for all
the tests requested for a sample are available, the analyzer sends them to
the Host where they can be validated.
To view or edit results on the analyzer, use the Workplace > Data Review
screen; all test results in the database are displayed here.
5.3. Reagents:
5 Choose either Unload (to send the cassette to the cassette loading port)
or Dump.Choose Execute.
5.4. Calibration
Calibration is the process that establishes a relation between measurement value
and the corresponding results.
This process may comprise the complete setting up of a new calibration curve or it
is only the update of one or two parameters of an existing calibration curve. In
either case the term calibration is appropriate.
5.4.1. Overview
The calibration menu contains three screens: Calibration > Status,
Calibration > Calibrator, and Calibration > Install.
This screen shows detailed information about the calibration status of the tests on the
modules selected in the Module list box. This screen is used to select tests for start up,
timeout, changeover or manual calibrations.
The graph shows calibration results for the selected photometric test. It displays
the measured absorbance values for both the Std (1) calibrator and the
calibrator with the maximum concentration Std (2)-Std (6) on one plot.
The left and right y-axes are scaled independently:
o The left axis refers to the Std (1) calibrator represented by .
o The right axis refers to the maximum calibrator Std (2)-Std (6) represented
by
5.8.1. QC concept
Regularly perform QC measurements to continually monitor the
instrument performance. After measurement of QC samples, data can
both be transferred to and evaluated on a Host or they can be
evaluated on the system.
The QC > Status screen displays all installed controls with the
corresponding assigned and activated tests for each module and
measuring cell.
The test display can be sorted by module, test name (alphabetical order),
control name (alphabetical order), QC test selection, or QC cause. Tests
can be selected with the Select/Deselect button. Selected tests are
highlighted in green. A control test selection is made for the subsequent
control run. The control test selections of the status list are reset after the
control is measured.
An asterisk (*) in the first column of the QC > Status screen denotes which
control is used as automatic QC.
On the QC > Individual screen, choose Chart to open the Individual QC Chart window.
This window displays all non-accumulated QC results for the selected test in a
Levey-Jennings chart. A Levey-Jennings chart provides a visual method of
monitoring trends in QC results. The result values are scaled to the standard
deviation (SD) of the respective test.
Up to three controls for one test can be selected (Select button) and displayed on
one chart.
5.8.5. Excluding an individual QC point or adding a
comment
Individual QC points on a QC chart can be excluded from QC
calculations if required. It is also possible to annotate individual
QC points.
1 Choose QC > Individual > Chart.
2 Select the QC point to be excluded from QC calculations.
Choose Utility > System > QC Setting to display the QC Setting window.
Accumulation Method
Select one of the two accumulation methods, Mean-R or X-R, from this box.
QC Calculation
If the accumulation method is Mean-R, select one of the two calculation
methods, DayToDay or Overall, from this box to specify how mean and SD
values are calculated in the cumulative QC.
This box is not available if X-R is selected in the Accumulation Method box.
5.8. Maintenance
2 Select the Shut Down option, choose Execute and confirm the
shutdown.
3 Wait until the computer power supply turns off. Then, switch off the
power switch of the printer and monitor.
4 Switch off the operation power switch on the left of the rack sampler
unit.
6 For maintenance tracking, enter the frequency in the text box and select the time
period (Hour, Day and Month) from the Period list box.
7 Select a timer function from the Timer list box:
Realtime: The actual time
Power On: Cumulative power-on time of the analyzer (excluding sleep time)
Operation: Cumulative operation time
Power on time and operation time are both listed on the Maintenance Report
(Print > Utility > Maintenance Report). If a frequency period is not required,
enter 0 in the field to cancel.
8 Enter the warning level (%) in the Warning Level text box. Example: If you define
a period of 10 hours in real time and a warning level of 90%, the analyzer will
remind you to perform the maintenance item after 9 hours.
9 Select the Comment check box to display the Comment field on the appropriate
maintenance item window and to print the comment on the Maintenance Report.
10 Choose Update to add the item to the Maintenance Items list.
11 Repeat steps 3 to 10 for all maintenance items for the maintenance type.
12 Choose OK to save the collection and its settings and close the Edit
Maintenance Type window.
The following tables display the maintenance items and checks and indicate in which
operation mode they can be performed.
5.9. Finalization
6. Appendices:
Cobas 6000 user manual
7. Signature Block:
Revised On