M850 User Manual
M850 User Manual
Revision History
This manual has a revision number. This revision
number changes whenever the manual is updated due to
software or technical specification change. Contents of this
manual are subject to change without prior notice.
Document No.: J/M850CE-A-008
Revision number: V1.0
Release time: Feb. 2017
Copyright © 2017 Guangdong Biolight Meditech Co., Ltd.
All rights reserved.
I
CE mark
ve Name:
EC Representativ
Shanghai International Holding Corp. GmbH (Europ
pe)
ve Address:
EC Representativ
Eiffestraβe 80, 20537 Hamburg Germany
Statement
Manufacturer holds the copyright of this manuaal, and we
o deal with this manual as confideential files.
are also entitled to
This manual is only
o used for operation, mainten
nance and
service of productt, someone else can not publish thee manual.
This manual contains
c exclusive information pro
otected by
copyright laws an
nd we reserve its copyright. Withou
ut written
approval of manu
ufacturer no parts of this manuall shall be
photocopied, xero
oxed or translated into other languaages.
The contents contained in this manual are subject
s to
amendments without notification.
Manufactureer's Responsibility
Only under the following circumstancces will
II
manufacturer be responsible for the safety, reliability and
performance of the instrument:
All the installation, expansion, readjustment, renovation
or repairs are conducted by the personnel certified by
manufacturer.
The storage condition, operation condition and electrical
status of the instrument conform to the product
specification.
The instrument is used in accordance with the user’s
manual.
About this manual
This manual contains the instructions necessary to
operate the product safely and in accordance with its
function and intended use. Observance of this manual is a
prerequisite for proper product performance and correct
operation and ensures patient and operator safety.
This manual is based on the maximum configuration and
therefore some contents may not apply to your product. If
you have any question, please contact us.
This manual is an integral part of the product. It should
always be kept close to the equipment so that it can be
obtained conveniently when needed.
All illustrations in this manual serve as examples only.
III
They may not necessarily reflect the setup or data displayed
on your product.
Conventions:
Bold Italic text is used in this manual to quote
the referenced chapter or sections.
【】is used to enclose screen texts.
→ is used to indicate operational procedures.
Signs in this manual:
IV
Contents
Chapter 1 General Introduction........................................... 1-1
1.1 Intended Use .............................................................. 1-1
1.2 Main Unit ................................................................... 1-2
1.3 Display Views ............................................................ 1-7
Chapter 2 Safety ..................................................................... 2-1
2.1 Safety Information ..................................................... 2-1
2.2 Explanation of Symbols ............................................ 2-5
Chapter 3 Basic Operations .................................................. 3-1
3.1 Unpacking and Checking .......................................... 3-1
3.2 Getting Started ........................................................... 3-1
3.3 Starting the Monitor .................................................. 3-2
3.4 Screen Brightness Setting .......................................... 3-2
3.5 Auto-Rotate Setting ................................................... 3-3
3.6 Date & Time Setting.................................................. 3-3
3.7 Patient Information Setting ....................................... 3-4
3.8 Demo Mode Setting................................................... 3-4
3.9 Language Setting ....................................................... 3-5
3.10 Checking the Version ................................................ 3-5
3.11 Restoring the Factory Configuration ......................... 3-6
3.12 Shutting off the Monitor ............................................ 3-6
Chapter 4 Alarm .................................................................... 4-1
4.1 Alarm Categories ....................................................... 4-2
V
4.2 Alarm Levels ............................................................. 4-2
4.3 Alarm Indicators ........................................................ 4-4
4.4 Alarm Status Symbol................................................. 4-7
4.5 Alarm Tone Configuration ........................................ 4-7
4.6 Pausing Alarms .......................................................... 4-8
4.7 Adjust the Alarm Volume to Zero ............................. 4-9
4.8 Alarm Reset ............................................................. 4-10
4.9 When an Alarm Occurs ........................................... 4-11
Chapter 5 Measuring ECG ................................................... 5-1
5.1 Introduction ............................................................... 5-1
5.2 Safety Information ..................................................... 5-1
5.3 Monitoring Procedure................................................ 5-3
5.4 Checking Paced Status .............................................. 5-8
5.5 ECG Display ............................................................ 5-10
5.6 Setting ECG ............................................................. 5-10
Chapter 6 Measuring RESP ................................................. 6-1
6.1 Principle of Measuring .............................................. 6-1
6.2 Preparatory Steps of Measurement ........................... 6-1
6.3 Resp Display .............................................................. 6-3
6.4 Resp Setup ................................................................. 6-3
Chapter 7 Measuring SpO2 .................................................. 7-1
7.1 Introduction ............................................................... 7-1
7.2 Safety Information ..................................................... 7-2
VI
7.3 Monitoring Procedure................................................ 7-4
7.4 SpO2 Display............................................................. 7-5
7.5 PR Display ................................................................. 7-6
7.6 SpO2 Setup ................................................................ 7-7
7.7 Desat limit Setup ....................................................... 7-9
Chapter 8 Trend Review ....................................................... 8-1
8.1 Introduction ............................................................... 8-1
8.2 Review Interface ........................................................ 8-1
8.3 Review Setup ............................................................. 8-2
Chapter 9 Battery .................................................................. 9-1
9.1 Introduction ............................................................... 9-1
9.2 Charging the Battery.................................................. 9-2
9.3 Optimizing Battery Performance .............................. 9-2
9.4 Checking the Lithium Battery ................................... 9-3
9.5 Disposing of the Batteries ......................................... 9-4
Chapter 10 Maintenance and Cleaning ............................. 10-1
10.1 Introduction ............................................................. 10-1
10.2 Seasonal Safety Checking ....................................... 10-2
10.3 Cleaning the Monitor............................................... 10-4
10.4 Cleaning ECG Cable ............................................... 10-5
10.5 Cleaning SpO2 Sensor............................................. 10-6
10.6 Disposal ................................................................... 10-6
Chapter 11 Accessories ........................................................ 11-1
VII
Appendix A Product Specifications ......................................... 1
A.1 Safety Specifications ..................................................... 1
A.2 Physical Specifications .................................................. 2
A.3 Environmental Specifications........................................ 2
A.4 Charging Specifications................................................. 2
A.5 Hardware Specifications................................................ 3
A.6 Data Storage .................................................................. 4
A.7 Measurement Specifications.......................................... 5
Appendix B Factory Defaults ................................................ 13
B.1 Alarm Setup ................................................................. 13
B.2 System Setup................................................................ 13
B.3 ECG Setup ................................................................... 14
B.4 SpO2 Setup .................................................................. 14
B.5 Trend Setup .................................................................. 14
Appendix C Alarm Message .................................................. 15
C.1 Physiological Alarm..................................................... 15
C.2 Technical Alarm ........................................................... 16
C.3 Prompt Message........................................................... 17
Appendix D Guidance and Manufacturer’s Declaration of
EMC ......................................................................................... 18
Appendix E Warranty Registration Card ............................ 27
VIII
Patient Monitor User’s Manual
1-1
Patiient Monitor User’s Manual
6 1
5 3
4
1-2
Patient Monitor User’s Manual
1-3
Patiient Monitor User’s Manual
Topside:
1 2
1-4
Patiient Monitor User’s Manual
Downside:
Rightside:
4 5
o computer.
Export data to
1-5
Patient Monitor User’s Manual
Warning:
The equipment connected to monitor shall meets
requirements of the system standard IEC 60601-1. If
in doubt, consult the technical service department or
your local representative.
Operator shall be responsible for safe of system after
monitor connected to computer.
Don’t touch the patient when operating the USB
connector, if not, it will generate risk of electric shock.
4. Shortcut key
Press this button to freeze or unfreeze the ECG waveform
5. Power buttom
Press it about two seconds to turn on when the monitor is
on the condition of shutdown.
Press it about two seconds to turn off when the monitor is
on the condition of working.
Calibration of touch screen
Press shortcut key firstly and press power button and
immediately loose shortcut key, click the center of
appearing point on screen. If the calibration passes, it will
enter the normall interface, if not, a red fork will appear
on screen and continue to calibrate.
1-6
Patient Monitor User’s Manual
8
9
7
2 10
1 11
12
Fig 1-4 Multi-Parameter Display Mode
1-7
Patient Monitor User’s Manual
1-8
Patient Monitor User’s Manual
2 3
1-9
Patient Monitor User’s Manual
current SpO2 value and its upper and lower alarm limits.
3. PR parameter area: The values shown in the area are
current PR value and its upper and lower alarm limits.
1-10
Patient Monitor User’s Manual
1-11
Patient Monitor User’s Manual
1-12
Patient Monitor User’s Manual
Chapter 2 Safety
Warning:
Explosion hazard: Do not use the monitor in the
presence of flammable anesthetics mixture with air,
oxygen, or hydrogen.
When the monitor is in use, there should not be any
great power appliances as high voltage cables, X-ray
machine, ultrasound equipment and electrizer in use
nearby.
Do not open the monitor housings; electric shock
hazard may exist. All servicing and future upgrades
must be carried out by the personnel trained and
authorized by manufacturer only.
Keep the monitor away from dust, vibration,
corrosive substances, explosive materials, high
temperature and moisture.
Do not come into contact with the patient during
defibrillation. Otherwise serious injury or death could
result.
When the monitor is connecting with high-frequency
devices, sensors and cables should avoid touching
2-1
Patient Monitor User’s Manual
2-2
Patient Monitor User’s Manual
Caution:
The monitor does not contain any parts for self-repair
by users. The repair of the instrument must be
conducted by the technical personnel authorized by
manufacturer.
To ensure patient safety, use only parts and
accessories specified in this manual.
When the monitor is connected to AC power, the
battery is in a state of being recharged. When it is
2-3
Patient Monitor User’s Manual
2-4
Patiient Monitor User’s Manual
Type CF
C applied part, defibrillation proteected.
Refer to
o user’s manual.
Alternaating current
Alarm volume
v off
Alarm paused
p
Alarm reset
r
QRS vo
olume off
2-5
Patient Monitor User’s Manual
Date of manufacture
Manufacturer
Serial number
Power button
2-6
Patient Monitor User’s Manual
2-7
Patient Monitor User’s Manual
3-1
Patient Monitor User’s Manual
modules.
——Check for any mechanical damage;
——Check for any incorrect connection of all the external
cables and accessories
Warning:
If the monitor is mechanically damaged, or if it is not
working properly, do not use it for any monitoring
procedure on a patient. Contact your service
personnel.
To avoid explosion hazard, do not use the monitor in
the presence of flammable anesthetics, vapors or
liquids.
3-2
Patient Monitor User’s Manual
After starting up, you need to set date and time of this
monitor. Operations are as follows:
1. Select 【Menu】→【System】to enter the System menu
shown as follows:
2. Select【Use 24-hour format】, it can be set to【24h】or
【12h】.
3. Select 【Date format】, it can be set to【YYYY/MM/DD】、
【MM/DD/YYYY】or【DD/MM/YYYY】.
3-3
Patient Monitor User’s Manual
4. Set the current date and time and select【OK】to confirm it.
3-4
Patient Monitor User’s Manual
3-5
Patient Monitor User’s Manual
3-6
Patient Monitor User’s Manual
3-7
Patient Monitor User’s Manual
Chapter 4 Alarm
Alarm refers to a prompt that is given by the monitor for
medical personnel through visual, audible and other means when
a vital sign appears abnormal or the monitor occurs technical
problem.
Warning:
Setting alarm limits to extreme values that can render
the alarm system useless.
Alarm settings are restored automatically after power
is interrupted for ≤30s, the alarm setting will lose if the
power is interrupted for>30s.
4-1
Patient Monitor User’s Manual
4-2
Patient Monitor User’s Manual
Caution:
The levels of technical alarms are predefined before
the monitor leaves the factory and cannot be changed
by users.
The level of technical alarm can’t be changed by the
user.
4-3
Patient Monitor User’s Manual
4-4
Patient Monitor User’s Manual
Alarm
Audible prompt
level
High “DO-DO-DO------DO-DO,DO-DO-DO------DO-DO”
Medium “DO-DO-DO”
Low “DO-”
4-5
Patient Monitor User’s Manual
Caution:
When multiple alarms of different levels occur at the
same time, the monitor will select the alarm of highest
level give visual and alarm indications.
When multiple alarms occur at the same time, the
alarm message will be displayed in the alarm area in
turn.
4-6
Patiient Monitor User’s Manual
1. The minimum
m alarm volume setting.
Select 【Menu】
】→【System】→【Maintenancce】→enter
the required passworrd, select 【Machine Mainten.】→【Alarm
Setup】→【Min.Alm
m.Vol.】, you can select “0, High, Mid,
M Low”.
2. Alarm volumee setting
Select 【Menu】
】→【System】→【Alarm Volu
ume】, you
can select from X to high. Attention: “X” means: the value
v of the
4-7
Patiient Monitor User’s Manual
ume.
minimum alarm volu
Warning: Auditory
A alarm signal sound pressure
levels,
l which are less than ambiient levels,
can
c impede operator recognition
n of alarm
conditions.
c
4-8
Patient Monitor User’s Manual
Warning:
Potential hazard can exist if different alarm pre-sets
are used for the same of similar equipment in any
single area.
When the alarm sound is adjusted to 0, the monitor
will give no audible alarm tones even if a new alarm
occurs. Therefore the user should be very carefully
4-9
Patiient Monitor User’s Manual
Select 【Menu
u】→【System】→【Alarm reseet】.
Press alarm reset , you can reset the alarm system::
It will exit alaarm pause if it is on the condition of alarm
pause.
It only turns offf audible alarm, the visual is go
oing on for
the existing alarm.
The audible alarm will be restored when a new
n alarm
occurs.
The parameters of physiological alarm keep on flaashing.
The alarm of lead-off/sensor-off turns into a prompt
message.
4-10
Patient Monitor User’s Manual
4-11
Patient Monitor User’s Manual
5.1 Introduction
Before the mechanical contraction, the heart will firstly
produce electrization and biological current, which will be
conducted to body surface through tissue and humors; the current
will present difference in potential in different locations of the
body, forming potential difference ECG, also known as body
surface ECG or regular ECG, is obtained by recording this
changing potential difference to form a dynamic curve. Monitor
measures the changes in the body surface potentials caused by the
heart of the patient, observe the cardioelectric activities, record
the cardioelectric waveforms and calculate the HR through the
multiple electrodes connected to ECG cable.
Warning:
It is imperative to only use the ECG electrodes and
cables provided by manufacturer or specified in this
manual. Users shall use the electrode which has little
polarization voltage and little contact resistance.
5-1
Patient Monitor User’s Manual
5-2
Patient Monitor User’s Manual
5-3
Patient Monitor User’s Manual
Note:
For patients who tremble a lot or patients with
especially weak ECG signals, it might be difficult to
5-4
Patient Monitor User’s Manual
2. Electrode Placement
3-Lead
Take the AHA standard as an example, when conducting
3-lead ECG monitoring, use 3-lead ECG cable. The three
limb-leads of RA, LA and LL as shown in below figure, will be
placed on the relevant locations. This connection can establish
the lead of I, II, III.
5-5
Patient Monitor User’s Manual
LA RA
LL
5-Lead
Take the AHA standard for example, when conducting
7-lead ECG monitoring, use 5-lead ECG cable. The four
limb-leads of RA, LA, RL and LL as shown in below figure, will
be placed on the relevant locations. This connection can
establish the lead of I, II, III, aVR, aVL, aVF; according to
actual needs, chest lead V can be placed on any of the locations
between V1~V6, respectively making one lead of V1~V6
established.
5-6
Patient Monitor User’s Manual
RA LA
RL LL
Warning:
To avoid burning, when the electrotome operation is
performed, the electrodes should be placed near the
middle between ESU grounding pad and electrotome
and the electrotome should be applied as far as
possible from all other electrodes, a distance of at least
15 cm/6 in is recommended.
When using the ESU device, avoid placing the
electrodes near the ESU grounding pad, otherwise,
5-7
Patiient Monitor User’s Manual
Depending on th
he patient to be monitored, you shall
s select
the proper leads. Enter into 【ECG Setup】to sellect【Lead
Type】.
5-8
Patient Monitor User’s Manual
Warning:
For paced patients, you must set 【Paced】to【Yes】.
If it is incorrectly set to【No】, the monitor could
mistake pace pulses for regular QRS complexes and
fail to alarm during asystole.
Some pace pulses can be difficult to reject. When this
happens, the pulses are counted as a QRS complex, and
could result in an incorrect HR and failure to detect
cardiac arrest or some arrhythmias. Keep pacemaker
patients under close observation.
For non-paced patients, you must set【Paced】to【No】,
otherwise, the system cannot detect the arrhythmia
related to Ventricular Premature (including PVCs
count), and will not conduct ST analysis.
5-9
Patient Monitor User’s Manual
5-10
Patiient Monitor User’s Manual
Fig
F 5-2 ECG Setup Interface
5.6.1 Setting HR
R Alarm
5.6.2 Setting HR
R Alarm Limits
5-11
Patient Monitor User’s Manual
5-12
Patient Monitor User’s Manual
5-13
Patient Monitor User’s Manual
1) Plug the 5-lead ECG cable into the ECG socket of the
monitor.
2) Place the various pads of the electrodes onto the body of
patient and connect them to the relevant lead cables. At this
moment, the screen will show Resp waves and the RESP
rate will be calculated.
3) Set the parameters relevant to Resp monitoring.
To measure Resp parameters, it is unnecessary to use other
cables and it is only necessary to use the two RA and LL leads in
6-1
Patient Monitor User’s Manual
the 5-lead ECG cable. So please plug the 5-lead ECG cable into
the ECG socket and refer to 5.3 Monitoring Procedure to place
the RA and LL leads onto the body of patient.
Caution:
In order to get the best Resp waveforms, when
selecting leadⅡfor measuring Resp, it is advised to
place RA and LL electrodes cornerways.
For reducing the influence of rhythmic blood flow on
Resp electrode pickup impedance changes, avoid the
liver area and ventricles of heart in the line between
RA and LL electrodes. This is particularly important
for neonates.
The measurement of RESP is not applicable for
patient with excessive motion, otherwise it may cause
the mistake of RESP alarm.
6-2
Patient Monitor User’s Manual
6
3
6-3
Patiient Monitor User’s Manual
Fig
F 6-2 Resp Setup Interface
6-4
Patient Monitor User’s Manual
6-5
Patient Monitor User’s Manual
7.1 Introduction
The measurement of oxygen saturation of arterial blood
(also known as pulse oxygen saturation, usually shortened as
SpO2) adopts the principles of light spectra and volume tracing.
The LED emits lights with two specific wavelengths, which are
selectively absorbed by oxygenated hemoglobin and
deoxyhemoglobin. The optical receptor measures the changes in
the light intensity after the light passes the capillary network and
estimates the ratio of oxygenated hemoglobin and the total
hemoglobin.
oxygenated hemoglobin
SpO2 % = x 100%
oxyhemoglobin + deoxyhemoglobin
7-1
Patient Monitor User’s Manual
Warning :
Use only SpO2 sensors specified in this manual. Follow
the SpO2 sensor’s instructions for use and adhere to
all warnings and cautions.
When a trend toward patient deoxygenation is
indicated, blood samples should be analyzed by a
laboratory co-oximeter to completely understand the
patient’s conditions.
Do not use the monitor and the SpO2 sensor during
magnetic resonance imaging (MRI). Induced current
could cause burns.
Prolonged continuous monitoring may increase the
risk of unexpected changes in skin characteristics,
such as irritation, reddening, blistering or burns.
Inspect the sensor site every two hours and move the
sensor if the skin quality changes. For neonates, or
patients with poor peripheral blood circulation or
sensitive skin, inspect the sensor site more frequently.
Measurements and pulse signals can be affected by
certain environmental conditions, sensor application
errors, and certain patient conditions. See the
appropriate sections of this manual for specific safety
7-2
Patient Monitor User’s Manual
information.
Check the SpO2 sensor and its package for any sign of
damage before use. Do not use the sensor if any
damage is detected.
When disposing the disposable SpO2 probe or useless
SpO2 probe, please observe all local, state, and federal
regulations that relate to the disposal of this products
or similar products.
Note:
The pleth wave is not equal to the intensity of PR
signal.
The monitor does not provide automatic
self-examination alarm signal and the operator has to
use SpO2 simulator for self-examination.
A functional tester cannot be used to assess the
accuracy of SpO2 of the monitor.
7-3
Patient Monitor User’s Manual
Warning:
Do not use the SpO2 sensor on a limb where the NIBP
cuff is applied. This may result in inaccurate SpO2
reading due to blocked blood flow during cuff
inflation.
Do not conduct SpO2 measurement on the finger
smeared with nail polish, otherwise unreliable
measurement results might be produced.
When using finger sensor, make sure the nail faces to
the light window.
7-4
Patient Monitor User’s Manual
1 5
7-5
Patient Monitor User’s Manual
Waveform Display
7.5 PR Display
1 5
7-6
Patiient Monitor User’s Manual
7-7
Patient Monitor User’s Manual
Click the right of 【Alarm】, you can set alarm level of SpO2
and PR, you can select “Mid, High”
Select the average time for SpO2. The shorter the averaging
time is, the quicker the monitor responds to the change in the
patient’s oxygen saturation level. Click the right of【Avg Time】,
you can select “4s, 8s, 16s”.
Click the right of【QRS Vol.】, you can select “Off, High,
7-8
Patient Monitor User’s Manual
Mid, Low”.
7-9
Patiient Monitor User’s Manual
8.1 Introductiion
8.2 Review In
nterface
8-1
Patiient Monitor User’s Manual
xt】or【Pre.】.
by clicking the【nex
Fig
g 8-2 ID Review Interface
Click the right off 【More】on the top of review intterface, the
drop-down window shown
s as following:
8-2
Patiient Monitor User’s Manual
8-3
Patiient Monitor User’s Manual
Chapter 9 Battery
9.1 Introductiion
A rechargeable and
a maintenance-free battery is deesigned for
Patient Monitor, wh
hich enables continuous working when AC
power off.
9-1
Patient Monitor User’s Manual
9-2
Patient Monitor User’s Manual
9-3
Patient Monitor User’s Manual
9-4
Patient Monitor User’s Manual
10.1 Introduction
Warning:
Be sure to shut down the system and disconnect all
power cables from the outlets before cleaning the
equipment.
For optimal performance, product service should be
performed only by qualified service personnel
10-1
Patient Monitor User’s Manual
Please clean the plug of power cord at least once a year. Too
much dust on plug may cause the fire.
The following safety checks should be performed at least
every 12 months by a qualified person who has adequate training,
knowledge, and practical experience to perform these tests.
The data should be recorded in an equipment log. If the
device is not functioning properly or fails any of the following
tests, the device has to be repaired.
① Inspect the equipment and accessories for mechanical and
10-2
Patient Monitor User’s Manual
functional damage.
② Inspect the safety relevant labels for legibility.
③ Verify that the device functions properly as described in the
instructions for use.
④ Test the earth leakage current according IEC 60601-1 Limit:
NC 500μA, SFC: 1000μA.
⑤ Test the enclosure leakage current according to IEC 60601-1:
Limit: NC 100μA, SFC: 500μA.
⑥ Test the patient leakage current (normal operation) according
IEC 60601-1
Limit: type CF: for a.c.: 10μA, for d.c.: 10μA.
⑦ Test the patient leakage current under single fault condition
according IEC 60601-1
Limit: type CF: for a.c.: 50μA, for d.c.: 50μA.
⑧ Test the patient leakage current Mains voltage on applied part:
According IEC 60601-1:
Limit: type CF: for a.c.: 50uA.
10-3
Patient Monitor User’s Manual
10-4
Patient Monitor User’s Manual
10-5
Patient Monitor User’s Manual
1. The casing of the sensor and light tube can be cleaned with
swab or non-velvet soft cloth dipped with medical alcohol.
2. The sensor cable can be cleaned or sterilized with Hydrogen
Peroxide 3% or isopropyl alcohol 70%.
3. It is forbidden to put the monitor in high-pressure containers
and put the sensor directly in liquid.
10.6 Disposal
10-6
Patient Monitor User’s Manual
Chapter 11 Accessories
Warning:
Use only accessories specified in this manual. Using
other accessories may cause damage to the monitor.
Disposable accessories are designed for single-patient
use only. Reuse of them may cause a risk of
contamination and affect the measurement accuracy.
Check the accessories and their packages for any sign
of damage. Do not use them if any damage is
detected.
Type Model PN
Adult 15-100-0008
ECG Electrode
Pediatric/Neonatal 15-100-0009
11-1
Patient Monitor User’s Manual
Type Model PN
usable,adult 15-100-0013
usable,neonatal 15-100-0015
11-2
Patient Monitor User’s Manual
1
Patient Monitor User’s Manual
Operating: 5℃ to +40℃;
Temperature
Storage: -20℃ to +55℃;
Atmospheric Operating: 860hPa to 1060hPa;
pressure Storage: 500hPa to 1060hPa;
Operating: 15% to 85%(non condensing)
Humidity
Storage: less than 93%(non condensing)
2
Patient Monitor User’s Manual
A.5.1 Display
Size 4.3inch
Resolution 480*272
Touch Resistive touch
Autorotation four direction
Direction
3
Patient Monitor User’s Manual
A.5.4 Buttons
4
Patient Monitor User’s Manual
Patient ID 1~96
ECG
Lead type 3-lead:I,II,III
5-lead:I, II, III, aVR, aVL, aVF, Vx
ECG input ±6mV
rang
Gain 2.5mm/mV, 5mm/mV, 10mm/mV
Input ≥5.0MΩ
impedance
Leakage <10 uA
Current
GMRR ≥89dB
5
Patient Monitor User’s Manual
Frequency 0.5-40Hz
response
Electrode
offset ±500mV d.c.
potential
Protection
against the
effects of
defibrillation
<5s after Defibrillation.
and blocking
after
defibrillation
6
Patient Monitor User’s Manual
Recovery time
of equipment
from
ECG waveform will recover to the baseline in
electrodes
10 s.
polarization
after
defibrillation
Pacemaker With pacemaker detection , meet the
pulse symbol requirement of IEC60601-2-27: 2011
201.12.101.12 , the pace pulse meet the
following requirements ,the pace symbol
(≥2mm) will be shown on screen:
Pulse amplitude:±2mV~±700mV
Pulse width:0.1~2.0ms
Rise time:10~100us
7
Patient Monitor User’s Manual
8
Patient Monitor User’s Manual
HR
Measurement Adult: 15 bpm to 300 bpm
range Pediatric and Neonatal: 15 bpm to 350 bpm
1 bpm
Resolution
Detecting ≥0.20mVpp
sensitivity (II
lead)
Response time <12s
of HR meter
to change in
HR
Response of
HR to QRS Adult: without response
amplitude is
1mVp-p,
width is 10ms
0 bpm to 300 bpm, high/low limit can be
Alarm range
adjusted continuously.
9
Patient Monitor User’s Manual
RESP
Method Impedance
II(RA-LL)optional, II(default)
range
Gain X0.25、X1、X2
adjusted continuously
Resolution 1 rpm
10
Patient Monitor User’s Manual
A.7.3 BLT-SpO2
SpO2
Measurement 0~100%
range
Resolution 1%
70~100%:± 2%
Accuracy
0~69%:unspecified
Alarm Select the high and low alarm limit of
SpO2
Sensor Pulse oximetry sensors contain LEDs
that emit red light at a wavelength of
approximately 660 nm and infrared light
at a wavelength of approximately 905
nm.
The total optical output power of the
sensor LEDs is less than 15 mW.
This information may be useful to
clinicians, such as those performing
photodynamic therapy.
Data update 13s
period
11
Patient Monitor User’s Manual
Anti-electrotome interference
Recovery time of
equipment after 5s
defibrillation
Resisting low With powerful ability of resisting low
perfusion ability perfusion, PR amplitude can reach to
0.2% with value of SpO2 displaying.
Pitch Tone with
PR
Measurement 25 bpm ~250 bpm
range
resolution 1 bpm
accuracy ±1% or ± 1 bpm, whichever is greater
Alarm Select the high and low alarm limit of PR
12
Patient Monitor User’s Manual
13
Patient Monitor User’s Manual
14
Patient Monitor User’s Manual
15
Patient Monitor User’s Manual
16
Patient Monitor User’s Manual
17
Patient Monitor User’s Manual
18
Patient Monitor User’s Manual
19
Patient Monitor User’s Manual
20
Patient Monitor User’s Manual
(50Hz) necessary to
magnetic position the
field monitor further
IEC from sources of
61000-4-8 power frequency
magnetic fields or
to install magnetic
shielding. The
power frequency
magnetic field
should be
measured in the
intended
installation
location to assure
that it is
sufficiently low.
NOTE UT is the a.c. mains voltage prior to application of the test
level.
21
Patient Monitor User’s Manual
22
Patient Monitor User’s Manual
3.5
Radiated RF 3 V/m 3 V/m d P
IEC 80 MHz to E1
61000-4-3 2.5 GHz 80 MHz to 800 MHz
7
d P
E1
800 MHz to 2.5 GHz
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the
transmitter manufacturer
and d is the recommended
separation distance in
metres (m).
Field strengths from fixed
RF transmitters, as
determined by an
electromagnetic site
survey,a should be less
than the compliance level
in each frequency range.b
Interference may occur in
the vicinity of equipment
marked with the
following symbol:
23
Patient Monitor User’s Manual
24
Patient Monitor User’s Manual
25
Patient Monitor User’s Manual
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
26
Patient Monitor User’s Manual
Product Model
Serial No. Contract
Date Installed Warranty
Name
Address
Contact Tele/fax
Person
27
Product name: Patient Monitor
Product type: M850
Manufacturer: Guangdong Biolight Meditech Co., Ltd.
Address: No.2 Innovation First Road, Technical Innovation
Coast, Hi-tech Zone, Zhuhai, P.R. China
Postcode: 519085
PN: 22-067-0007