The document provides instructions for navigating and configuring the menus on a RadarRight monitor device. It describes the various menu items for settings like damping, mapping curves, scaled units, range, near blanking, sensor tag, display values, LCD contrast, diagnostics including peak values, measurement status, echo curves, and current output simulation. Configuring the parameters in these menus allows customizing the device's measurements, display, and testing of functions.
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
17 views

Sensor Nivel004

The document provides instructions for navigating and configuring the menus on a RadarRight monitor device. It describes the various menu items for settings like damping, mapping curves, scaled units, range, near blanking, sensor tag, display values, LCD contrast, diagnostics including peak values, measurement status, echo curves, and current output simulation. Configuring the parameters in these menus allows customizing the device's measurements, display, and testing of functions.
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 5

MONITOR RadarRight

Normally when electronic constant is smaller than 1.4, the direct echofromthe mediiim is
low and hard to detect. However by measuring the echo reflected from the base of the vessel,
the height of the médium can be measured. Two parameters are needed to be entered here.
1. Height of empty vessel. 2. True médium height or médium electronic constant, these two
parameters are related, entering either one is accepted. The precisión of parameters will affect
the precisión of the measurement.
Notes: it should be taking cautions while applying the function, Micro DK. When applying
this function the system will decide whether use direct echo or echo from the base to take
the measurement.

1. 4 Damping Pushing < 4 > will enter this menú when the menú item number is 1.3 . L C D
displays asbelow

Damping

OS

Press E , enter editing menú. See parameter edition to leam how to edit the
parameter. To confirm the modification with E , give up with .

1. 5 Mapping curve This menú define the correlation between the measured valué and the current
output. linear or non-linear mapping can be selected in this menú. For the
non-linear correlations, paramenters setting must be done by a computer previously.
Pushing O will select this menú when the menú item number is 1.4. L C D
displays as below

Mapping curve 1.5

Linear ^

Press E , enter editing menú.


Move arrow with < ^ to Select linear or non-linear. Then confirm with E .

1. 6 Scaledunits The unit of the scaled output valué can be set in this menú. Pushing O will
enter this menú when the menú item number is 1.5. L C D displays as below

Scaled u n i t s

height ^
m •

1. 7 Scaling Pushing E to enter the editing menú, see paramenter editing for the valué
editing. Press E to confirm the modification.

24
mm
RadarRígut''
1.8 Range Measure range should be set in order to get accurate result. When menú is 1.8, press
O key to enter Range. LCD displays as below:
Press E key to enter, please see appendix as reference. When finishing editing
parameter, press E key to confirm, or press to cancel.

1.9 Near blanking Near blanking defines the zone immediately forward of the sensor where measurements
are not possible.

Near blanking 1.9

1.10 Sensor tag In the menú Sensor TAG you edit a 11-digit measurement loop designation. The character
set comprises: LettersfromA - Z and Numbers from 0-9.

Sensor t a g 1.10

SENSOR

See Optional item progranmiing to edit the tag ñame.

2 Display This menú is used to set display mode. Pushing < ^ will select this menú when the
main menú item number is 1. LCD displays as below

Basic settings
^ Display
Diagnostics
Service
Info

Push E , you get

2. 1 Display valué Enter display mode set with E . LCD displays as below

D i s p l a y valué

Distance •

25
MONITOR RadarRi
Push E , you can select different display types of the measured valué as shown below

D i s p l a y valué 2.1
shut o f f map percent
• distance scaled
height current
percent

Move arrow to point to the type you want, confirm it with E .

2. 2 LCDcontrast Pushing O will enter this menú when the menú item is 2. 1. LCD diplays as below
adjustment
LCD c o n t r a s t 2.2

adjust?

Adjust the B / W contrast by pressing E , LCD Displays as below

Increase contrast with decrease contrast with < ^ .confírm with E .

3 Diagnostics The running status of the sensor can be provided by the menú Diagnostics, and
furthermore sensor testing can be done.
Pushing O to select this menú when the main menú item number is 2. L C D displays
as below
Basic settings 3
Display
^ Diagnostics
Service
Info

Push E , you get

3. 1 Peak valúes Peak valúes record the maximal and mininoal distance. The records can be cleared
to zero at menú 4.4 .

Peak valúes 3.1

distance-min 0.000 m ( d )
distance-max 2.109 m ( d )

3. 2 Meas. Status Pushing O to display measuring status when the menú item is 3.1 . L C D displays
(measure) as below

Meas s t a t u s 3.2
meas r e l i a b i l i t y : lOdB
sensor s t a t u s : OK

V J
Ib
líví MONITOR RadarRight'
3. 3 Choose curve At this menú, different curves can be selected to be displayed at menú 3. 4, when
(echo curve) the menú item is 3.2 press O you get

Choose c u r v e

echo curve •

Pushing E to select the curve. There are three curves echo curve, False echo curve
and Output trend curve.

Choose curve
^ echo curve
f a l s e echo curve
output t r e n d

3. 4 Curve Pushing <*> will display the selected curve when the menú item is 3.3 . L C D displays
asbelow
Curve zoom
When the curve is displayed, pushing E will enter Curve Zoom function menú.

echo c u r v e 3.4
+10.000
O02. 384

0.01 M (d) 8

Move arrow with O, select menú item for X / Y axis zoom or unzoom. Then Confírm
with E .
For X axis zoom pushing ^ to mark the start position for zoom, then confírm with E .
Pushing ^ again to mark the end position for zoom and confírm with E . The
selected área of the curve will be shown on the whole screen. Exit zoom with ^ .

Echo Curve 3.4


•X-zoom
Y-zoom
unzoom
V y

3. 5 Simulation Simulation is used to simúlate the 4~20mA current output. By current output simulation
the accuraey and linearity of the current output can be checked. And the system testing
can be carried out. Push O to enter Simulation menú when the menú item number is
3.4 . L C D displays as below

Simulation 3.5

start simulation •

\
Pushing E enter Simulation mode selection menú. L C D displays as below

27
N/ll MONITOR RadarRighr"
Simulation
• percent
current
disttince

Pushing O select the Simulation mode. Then confirm with E . The corresponding valué
will be required and then simulation starts.

Three types ofsimulations


Percent: the output current is decided by a percent valué: 100% is relative to 20mA,
0%is relative to 4mA.
Current: the output current is regulated by a current valué.
Distance: the output current is decided by a distance valué. The current output depends
onMin. adjustment (see 1.1), Max adjustment (see 1.2) and Mapping (see 1.5)
4 Service This menú with professional functions can only be used by trained technicans. They
are False echo storage, Reset, sensor settings back up Password setting and etc.

Basic settings
Display
Diagnostics
Service
Info

4. 1 False echo High sockets or vessel installations, e.g. struts or agitators as well as buildup
and weld joints on the vessel walls cause interfering reflections which can impair the
measurement. A false echo storage detects and marks these false echoes, so that they
are no longer taken into account for the level measurement. A false echo memory
should be created with empty vessel so that all potential interfering reflections will
be detected.
Pushing E will enter this menú when the menú item is 4 . L C D displays as below.

F a l s e echo

change?

Then push E , L C D displays as below


F a l s e echo 4.1
• delete
update
créate new
eéit

With O select Update/Create new/Delete a false echo, confírm with E .

F a l s e echo 4.
update/create new

El.OOOm ( d )

When you select update créate /new, you are asked to input a distance valué for the real
echo. Then push E to confirm it and to start the operation. It will take some time to store
the false echo.
28