Normally when electronic constant is smaller than 1.4, the direct echofromthe mediiim is
low and hard to detect. However by measuring the echo reflected from the base of the vessel,
the height of the médium can be measured. Two parameters are needed to be entered here.
1. Height of empty vessel. 2. True médium height or médium electronic constant, these two
parameters are related, entering either one is accepted. The precisión of parameters will affect
the precisión of the measurement.
Notes: it should be taking cautions while applying the function, Micro DK. When applying
this function the system will decide whether use direct echo or echo from the base to take
the measurement.
1. 4 Damping Pushing < 4 > will enter this menú when the menú item number is 1.3 . L C D
displays asbelow
Damping
OS
Press E , enter editing menú. See parameter edition to leam how to edit the
parameter. To confirm the modification with E , give up with .
1. 5 Mapping curve This menú define the correlation between the measured valué and the current
output. linear or non-linear mapping can be selected in this menú. For the
non-linear correlations, paramenters setting must be done by a computer previously.
Pushing O will select this menú when the menú item number is 1.4. L C D
displays as below
Linear ^
1. 6 Scaledunits The unit of the scaled output valué can be set in this menú. Pushing O will
enter this menú when the menú item number is 1.5. L C D displays as below
Scaled u n i t s
height ^
m •
1. 7 Scaling Pushing E to enter the editing menú, see paramenter editing for the valué
editing. Press E to confirm the modification.
24
mm
RadarRígut''
1.8 Range Measure range should be set in order to get accurate result. When menú is 1.8, press
O key to enter Range. LCD displays as below:
Press E key to enter, please see appendix as reference. When finishing editing
parameter, press E key to confirm, or press to cancel.
1.9 Near blanking Near blanking defines the zone immediately forward of the sensor where measurements
are not possible.
1.10 Sensor tag In the menú Sensor TAG you edit a 11-digit measurement loop designation. The character
set comprises: LettersfromA - Z and Numbers from 0-9.
Sensor t a g 1.10
SENSOR
2 Display This menú is used to set display mode. Pushing < ^ will select this menú when the
main menú item number is 1. LCD displays as below
Basic settings
^ Display
Diagnostics
Service
Info
2. 1 Display valué Enter display mode set with E . LCD displays as below
D i s p l a y valué
Distance •
25
MONITOR RadarRi
Push E , you can select different display types of the measured valué as shown below
D i s p l a y valué 2.1
shut o f f map percent
• distance scaled
height current
percent
2. 2 LCDcontrast Pushing O will enter this menú when the menú item is 2. 1. LCD diplays as below
adjustment
LCD c o n t r a s t 2.2
adjust?
3 Diagnostics The running status of the sensor can be provided by the menú Diagnostics, and
furthermore sensor testing can be done.
Pushing O to select this menú when the main menú item number is 2. L C D displays
as below
Basic settings 3
Display
^ Diagnostics
Service
Info
3. 1 Peak valúes Peak valúes record the maximal and mininoal distance. The records can be cleared
to zero at menú 4.4 .
distance-min 0.000 m ( d )
distance-max 2.109 m ( d )
3. 2 Meas. Status Pushing O to display measuring status when the menú item is 3.1 . L C D displays
(measure) as below
Meas s t a t u s 3.2
meas r e l i a b i l i t y : lOdB
sensor s t a t u s : OK
V J
Ib
líví MONITOR RadarRight'
3. 3 Choose curve At this menú, different curves can be selected to be displayed at menú 3. 4, when
(echo curve) the menú item is 3.2 press O you get
Choose c u r v e
echo curve •
Pushing E to select the curve. There are three curves echo curve, False echo curve
and Output trend curve.
Choose curve
^ echo curve
f a l s e echo curve
output t r e n d
3. 4 Curve Pushing <*> will display the selected curve when the menú item is 3.3 . L C D displays
asbelow
Curve zoom
When the curve is displayed, pushing E will enter Curve Zoom function menú.
echo c u r v e 3.4
+10.000
O02. 384
0.01 M (d) 8
Move arrow with O, select menú item for X / Y axis zoom or unzoom. Then Confírm
with E .
For X axis zoom pushing ^ to mark the start position for zoom, then confírm with E .
Pushing ^ again to mark the end position for zoom and confírm with E . The
selected área of the curve will be shown on the whole screen. Exit zoom with ^ .
3. 5 Simulation Simulation is used to simúlate the 4~20mA current output. By current output simulation
the accuraey and linearity of the current output can be checked. And the system testing
can be carried out. Push O to enter Simulation menú when the menú item number is
3.4 . L C D displays as below
Simulation 3.5
start simulation •
\
Pushing E enter Simulation mode selection menú. L C D displays as below
27
N/ll MONITOR RadarRighr"
Simulation
• percent
current
disttince
Pushing O select the Simulation mode. Then confirm with E . The corresponding valué
will be required and then simulation starts.
Basic settings
Display
Diagnostics
Service
Info
4. 1 False echo High sockets or vessel installations, e.g. struts or agitators as well as buildup
and weld joints on the vessel walls cause interfering reflections which can impair the
measurement. A false echo storage detects and marks these false echoes, so that they
are no longer taken into account for the level measurement. A false echo memory
should be created with empty vessel so that all potential interfering reflections will
be detected.
Pushing E will enter this menú when the menú item is 4 . L C D displays as below.
F a l s e echo
change?
F a l s e echo 4.
update/create new
El.OOOm ( d )
When you select update créate /new, you are asked to input a distance valué for the real
echo. Then push E to confirm it and to start the operation. It will take some time to store
the false echo.
28