Service Manual: FOR Diagnostic Ultrasound System
Service Manual: FOR Diagnostic Ultrasound System
FOR
DIAGNOSTIC ULTRASOUND SYSTEM
MODEL Qbit7
We reserve the right to make changes to this manual without prior notice.
Qbit7 service manual
Regulatory Requirement
This manual is a reference for the Qbit7. Please verify that you are using the latest revision of
this document. If you need the latest revision, contact your distributor.
1 / 52
Qbit7 service manual
4.1 Notices.....................................................................................................20
CHAPTER6 OTHERS.................................................................................... 35
6.2 Network....................................................................................................44
2 / 52
Qbit7 service manual
Reversion History
3 / 52
Qbit7 service manual
1.1 Introduction
This guide covers the preparation, use, and maintenance of the ultrasound system, transducers,
and accessories. Refer to the manufacturers’ instructions for specific information about
peripherals. Please read the information in this guide before providing service to Qbit7 color
ultrasound system. It applies to the ultrasound system, and transducers.
CHISON Service Dept. provides efficient service solutions to customers timely, including
technical support, spare parts support, field service, upgrade, warranty extension, and on
demand service.
The engineer must have ultrasound training before provide service, CHISON can provide
varieties of training, including operation, application maintenance, and troubleshooting.
If the problem is not still solved there, please contact with CHISON Service Dept. at the
following numbers or addresses:
Fax: 0086-510-85310726
E-mail: [email protected]
2 / 52
Qbit7 service manual
NOTE:
Important information
2. The warranty does not cover the following items, even during the warranty period:
a) Damage or loss due to misuse or abuse with system and probes, for
example, drop the probe, the liquid or the metal part fall into the system.
c) Damage or loss caused by failure to meet the specified conditions for this
system, such as inadequate power supply, improper installation or
environmental conditions.
e) Damage or loss due to use the system outside the region where the
system was originally sold.
f) Damage or loss involving the system purchased from a source other than
CHISON or its authorized agents.
4. During operate the system, if user has any doubt, difficulty or any unclear, please
contact CHISON's authorized service engineer immediately. Please describe the
situation clearly to solve the question in time. Before solve the question, please
don’t operate the system.
5. This system shall not be used by persons other than fully qualified and certified
medical personnel.
6. The device can only be sold to qualified medical institutions or doctors. The users
shall fully understand and master the devices before operating. The users shall
have got the qualification, and shall comply with the local laws and regulations,
the local religion and customs, etc.
8. The purpose of this system is to provide physicians with data for clinical diagnosis.
It is the physician’s responsibility for diagnostic procedures. CHISON shall not be
liable for the results of diagnostic procedures
9. This manual contains warnings regarding foreseeable potential dangers, but user
shall always be alert to dangers other than those indicated as well. CHISON shall
not be liable for damage or loss that results from negligence or from ignoring the
precautions and operating instructions described in this operation manual.
3 / 52
Qbit7 service manual
10. Due to negligence not following operation manual, CHISON shall not be liable for
the results.
11. Each time before and after ultrasound examination, please check the probe
surface, probe cable and sheath whether they are abnormal, such as cracking,
peeling and deformation. Also check whether the lens is strongly fixed. Abnormal
probes may cause electric shock and injure the patient. Once any abnormal, user
must stop using and contact CHISON's authorized service engineer.
12. If the probe is dropped or scratched by hard part, please stop using the probe
immediately. And contact CHISON's authorized service engineer to make sure
the safety and effectiveness is in good condition before use.
13. If there is any liquid or metal to enter to the system, please power off the system
and stop using it immediately. Please first contact CHISON’s authorized service
engineer to make sure it’s safe before restart using it.
14. Please don't use solvents (such as paint thinner, Benzene, or alcohol) or abrasive
cleansers for cleaning the system (including monitor and probes, etc.). It may
corrode the system and probes.
15. While the system or probe is over life time, please refer to operation manual
section 9.5
16. Important data must be backed up on external memory media. CHISON shall not
be liable for loss of data stored in the memory of this system caused by operator
error or accidents.
17. LCD display screen may have some dark or light dots, it is normal for the LCD. It
does not mean that LCD screen is defective.
18. Please put this service manual with the system to ensure operator and manager
can reach it at any time.
19. Please start to disassemble the system in 2 minutes after the system is OFF.
20. Please pay the attention to the invert part, that there is high voltage on it.
21. While open the metal box, please note the sharp edge of metal part to avoid
being cut the finger.
4 / 52
Qbit7 service manual
module.
5 / 52
Qbit7 service manual
2.1.1.1
Remove the 4 screws from the back cover and remove the back panel.
2.1.1.2
Remove the 8 screws and pull out the cable.
6 / 52
Qbit7 service manual
2.1.1.3
Pull out the mainframe.
2.1.1.4
Remove the battery from the mainframe.
7 / 52
Qbit7 service manual
2.1.1.5
Remove the hard disk from the mainframe.
2.1.1.5
Remove the 11 screws from the cover to disassemble the ultrasound module.
8 / 52
Qbit7 service manual
2.1.1.6
The following figure is a feature of the main board and signal box board connections. Pull
out cables in the interface and remove the screw marked in the picture.
2.1.1.6
Remove 7 screws from the transducer board, main board, power board by turns.
2.1.1.7
The fan is the last one need to be removed.
9 / 52
Qbit7 service manual
2.1.2.1
Disassemble the PC module, you need to remove the 11 screws marked in the picture.
2.1.2.2
The following picture is a comparison between the dismantling before and after of the PC
part.
2.1.2.3
Remove the 8 screws to remove the fans.
10 / 52
Qbit7 service manual
2.1.2.4
Remove the 7 screws from the PC board to disassemble it.
2.1.2.5
Remove the 4 screws to disassemble the charger adapter.
11 / 52
Qbit7 service manual
2.1.3.1
Remove the 4 screws to disassemble the back panel.
2.1.3.2
Remove the 4 screws to disassemble the box of the back panel.
12 / 52
Qbit7 service manual
2.1.3.3
Remove the 4 screws to disassemble the box of the back panel AC-FILTER.
2.1.3.4
The detail of the AC-FILTER.
13 / 52
Qbit7 service manual
2.2.1.1
The following picture is the screen module.
2.2.1.2
14 / 52
Qbit7 service manual
2.2.1.3
Remove the 7screws and pull out the VGA cable and power cable.
2.2.1.4
Remove the 12 screws to remove the back cover.
15 / 52
Qbit7 service manual
2.3.1.1
Remove the 13 screws under the keyboard and 3 screws on the support arm.
2.3.1.2
Remove the 3 screws to pull out the speaker cables.
16 / 52
Qbit7 service manual
2.3.1.3
Remove the 3 screws to pull out the speaker cables.
17 / 52
Qbit7 service manual
1) POWER board: include the DC-DC module, to generate the power supply to
ultrasound part; Motor board is fixed on this board if the main system has 4D function;
18 / 52
Qbit7 service manual
19 / 52
Qbit7 service manual
4.1 Notices
4.1.1
The following environmental conditions are within system tolerances for operation:
4.1.2
It is necessary to put the machine on a flat place during operating. As a result of the poor
4.1.3
Keep the room clean and tidy, because the dust will accumulate rapidly on the air filter and
cause the system to overheat. Recommend that the customer to ensure that the dust in the
4.1.4
If any breakdown is suspected, please contact the local CHISON’s Authorized Service
Representative at the first time. Don’t disassemble or repair privately, resulting in unnecessary
loss.
20 / 52
Qbit7 service manual
installation environment.
Incorrect operation
The screen is
on, the inside fan doesn't work.
working.
monitor is ok.
21 / 52
Qbit7 service manual
range.
in operation.
Train the customer doesn't move the
automatically.
2002(example) year,
The battery on
sometimes even you adjust it
The system
SYSC-UNIT is out of Replace the battery on the SYSC-UNIT.
correctly, it will go back soon.
time is wrong.
power.
Sometimes the system time is
incorrect.
22 / 52
Qbit7 service manual
switch directly.
23 / 52
Qbit7 service manual
PANEL.
of progress bar.
device or Insert Boot Media in SATA cable 2) Replace the SATA cable
error key_”
The system shows "error report", 2) Contact the service engineer and
and asks for loading the license System then install the license
24 / 52
Qbit7 service manual
The system system shows notice "Don't ....." 1. Upgrade the software after "one key
can't recovery"
DSP,DBF,USB items are blank in Software/Probe
recognize 2. Replace the probe.
system information interface.
the probe 3. Replace the MAIN-UNIT.
properly.
The OP
PANEL Most of the buttons on the OP 1) Restart the system and check it
properly
OP PANEL cable
2) Reconnect the OP PANEL cable.
The system shows " Keyboard
3) Replace the OP PANEL cable.
error ".
4) Replace the PORT board.
25 / 52
Qbit7 service manual
Abnormal
"Zi Zi" sound HDD Replace HDD
sound in the
"Wu Wu" sound SYSC-UNIT Replace SYSC-UNIT
system
26 / 52
Qbit7 service manual
Abnormal
There are some bright lines in
display in
scanning area, no response with
scanning
all the probe scanning.
area MAIN-UNIT Replace the MAIN-UNIT.
Find different probes lacks of
regular.
27 / 52
Qbit7 service manual
setting, the image size changes to incorrect operation One Key Recovery
resolution.
signal from extra VGA LCD-UNIT 1) Check the connection of VGA cable.
VGA
is OK.
2) Replace VGA cable
cable/LCD-UNIT
The system shows notice “no 3) Replace the LCD-UNIT
signal”.
28 / 52
Qbit7 service manual
The printer
No signal from BNC port.
doesn't
Connection/SYSC-U 1. Connect well all the cables.
work. The printer can't be found by the
29 / 52
Qbit7 service manual
1) The upgrade file should be at the root of the USB flash disk.
3) When upgrade SW or FW, do not remove the USB flash disk or turn off power. It will
cause system damaged.
Before upgrade, you can backup the patient and examination data for backup, to avoid the
damage of hard disk. The steps for backup is as below:
30 / 52
Qbit7 service manual
3. Select the defined patient or select all, press backup to the destination. Press OK and
to start he backup.
4. Recover from the external memory. Choose the target source, and then it will display
the patient data in the list as below picture. Choose the data and press recover to recover
Fig.128
31 / 52
Qbit7 service manual
2) Press key on the keyboard, choose the "system setting", press "SW Upgrade".
32 / 52
Qbit7 service manual
2) Press key on the keyboard, choose the "system setting", press "FW Upgrade".
33 / 52
Qbit7 service manual
1) Software
Update SW OK
Fail
OK
Check the update file
Finish upgrading.
and Update SW again.
Fail
OK
Please replace the hard
disk directly.
2) Firmware
Update FW OK
Fail
Fail
MAIN-UNIT.
34 / 52
Qbit7 service manual
Chapter6 Others
6.1.1.1
Press the “setup” key; enter "system" interface, you will see the "Display Setting" bottom is
dark.
6.1.1.2
Set the video print option to choose different methods for video print. Choose the “General”
setting, and select Keyboard submenu. Choose “video print” under print key: print1,
choose “video enable” and “VGA enable” under video config.
35 / 52
Qbit7 service manual
“Picture and info” means print the ultrasound image with patient information.
6.1.1.3
Connect the cables between Video printer and the System as the following picture.
Need to restart the system after connect the cables between Video printer and the
System.
36 / 52
Qbit7 service manual
6.1.2.1
Place the printer smoothly.
6.1.2.2
Connect the printer to the system.
6.1.2.3
Set the print manager.
6.1.2.4
Choose “PC print” in system setting, and in the “PC print selection”, chooses “print the
6.1.3.1
Press key to enter into the System Settings., and then click the System page .Click
37 / 52
Qbit7 service manual
6.1.3.2
Click “Add” to add a printer, the system will pop up the “Add Printer Wizard”, follow the
6.1.3.3
Select “Local printer attached to this computer”, click “Next” to continue.
38 / 52
Qbit7 service manual
6.1.3.4
Select “Use the following port: USB001”, click “Next” to continue.
6.1.3.5
Plug the USB which contains the driver of the printer, then click “Have Disk” to continue.
39 / 52
Qbit7 service manual
6.1.3.6
Click “Browse” to locate the driver.
6.1.3.7
Select the driver, click “Open” to continue.
40 / 52
Qbit7 service manual
6.1.3.8
Click “OK” to confirm.
6.1.3.9
Click “Next “to continue.
41 / 52
Qbit7 service manual
6.1.3.10
Input printer name, select “Yes” to print a test page”, click “Next” to continue.
6.1.3.11
Click “Finish” to install the driver
42 / 52
Qbit7 service manual
6.1.3.12
Click “Continue anyway” to continue, the system will install the driver automatically.
6.1.3.13
Before using the printer, please check the printer configuration. Select the printer and then
click “Properties” and “Preferences” to adjust the printing Properties and Preferences.
43 / 52
Qbit7 service manual
6.2 Network
If you want to save the documents and files to the PC directly, you can set the network.
With this feature , To use the network storage function, the user should connect the target
computer to the network first, and build a new folder on the computer, right-click it to open
the attribute, select shared page, set to share this folder on the network and allow network
6.2.1.1
Create a folder on the computer, named as “xxxx”, for example “denny” , Right-click the
folder and choose its “Properties”, then click “sharing” => “Advanced Sharing”
44 / 52
Qbit7 service manual
6.2.1.2
Click the Permissions bottom, choose everyone as the user and allow: Full Control、
Change、Read。
6.2.1.3
Set computer network properties, click the “internet”.
45 / 52
Qbit7 service manual
6.2.1.4
Click the “properties”, get into “Network connection Status”
6.2.1.4
Double click the “TCP/IPv4”, get into “Internet Protocol Version4 Properties”
46 / 52
Qbit7 service manual
6.2.1.5
Get into “Internet Protocol Version4 Properties” interface, choose “use the following IP
6.2.2.1
Press key get into the System Setting interface. Click Network, Setting, then
plug the network cable into the LAN port between the ultrasound main unit and computer.
47 / 52
Qbit7 service manual
IP address: Input the IP address of the system manually, and make sure the IP address of
the network printer is in the same network segment. Subnet Mask: Input the subnet mask
manually. Default Gateway: Input the gateway manually. DNS server: Input DNS server
system is not connected to the network. Network Information: Display the current
network information. Configuration information: Click this button to display the network
connected, enter the IP address of the test, click the test, you can test the network the
connection status between the IP of current system and the IP that needs to be tested.
6.2.2.2
When fill out all the data, click “Test” to check the network info.
48 / 52
Qbit7 service manual
6.2.2.3
Click Network Storage interface, input “Shared Dir” and “IP address”, If Ping
Shared Dir : the folder in your computer which is shared. IP address: computer IP. User
49 / 52