URIECA Chemistry 5.35 Module 2: Synthesis of Coordination Compounds and Kinetics
URIECA Chemistry 5.35 Module 2: Synthesis of Coordination Compounds and Kinetics
DEPARTMENT OF CHEMISTRY
This experiment will contribute to improving your skills in the following lab
techniques:
II. Safety
III. Introduction
A. Coordination Chemistry
Angelici, R.J. Synthesis and Techniques in Inorganic Chemistry, 2nd Ed.; University
If six electrons are added such that they are paired up in the t2g set of orbitals (the
energy difference between the eg and t2g orbitals is large with respect to the
energy required to pair up the electrons in the t2g orbitals), the metal is said (in
CFT) to have the low spin d6 configuration, or more accurately, (t2g)6(eg)0.
complex. In general low spin octahedral complexes of Co(III) are relatively slow
to exchange ligands at the metal because the coordination number is relatively
high (six), thereby discouraging formation of a seven-coordinate species for steric
reasons. Also the metal electrons are in orbitals that lie between the axes (dxy,
dyz, dxz), thereby minimizing repulsion between them and the electrons on the six
ligands (bases). In contrast, a Co(II) complex (d7 therefore with electrons in the eg
orbitals) is usually readily substituted. For this reason we begin the synthesis of a
Co(III) species with [Co(H2O)6](NO3)2, a high spin d7 Co(II) species.
C. Kinetic studies
In this experiment you will follow the conversion of one compound into
another by
measuring the visible spectrum of the compound being consumed at a specified
temperature as a function of time. The reaction rate can be measured as the
change in concentration of a reactant (x) per unit of time Δ t. The change in
concentration of a reactant (or alternatively a product) is followed. Therefore
(3.2)
(3.3)
(3.4)
If b = -1/2, then the reaction is inverse 1/2 order in [y], and so forth. The observed
rate law will be consistent with the mechanism of the reaction, i.e., a series of
elementary reactions. (An example of an elementary reaction is a simple collision
between two species.) It is often possible to think of several elementary reactions
(each with its own "absolute" rate constant) that taken together would yield the
observed rate law with the observed rate constant. The units of the rate of a
reaction in solution are mol l-1 s-1 or M s-1. The rate of a reaction generally is
limited by a relatively slow elementary reaction. Relatively fast elementary
reactions that occur before or after a relatively slow reaction are not observable,
i.e., are not part of the observed law. In a given solvent (e.g., water) generally it
cannot be determined to what extent the solvent itself is involved in the reaction
since its concentration is constant. Therefore a rate constant is given for a
specific solvent at a specific temperature. The Nobel Prize in chemistry in 1903
was awarded to Svante Arrhenius, who proposed that the rate constant for a
(3.5)
a. Reading
b. TA Demonstrations
1. Volumetric Techniques
7. Filtration
11. Balances
V. Procedure
molar quantities of each reactant required for the two syntheses. For each
synthesis specify the limiting reagent.
While heating and stirring, concentrate the solution by allowing the temperature
of the
solution (measure it with a thermometer) to reach but go no higher than 75-80°C.
Add a total of 1 g of (NH4)2CO3 in several small portions during the process of
reducing the volume to ~10 mL. (A small final volume ensures that more of the
product will crystallize out.) Let stand undisturbed for 15 minutes, then place the
beaker in an ice bath for 15 minutes. (Disturbing the solution will cause rapid
formation of fine crystals that are not as easy to filter off.)
Filter off the purple-pink crystals using a 30-50 mL medium glass frit and wash
once with 3 mL of ice water, then twice with 3 mL of ethanol. (Helpful hint: turn off
the vacuum to add the wash solution, and then turn the vacuum back on) Dry in
vacuo (if possible) by placing a rubber stopper over the top of the funnel and
pulling a vacuum through the end of the funnel. Release by shutting off the
vacuum, holding the stopper, and removing the tube to the vacuum. Record the
weight, repeating the drying process until the weight no longer changes (Helpful
hint: use two cork rings stacked to support the glass frit on the balance). Store
the product in your desiccators. (Helpful hint: be sure the desiccant is dry i.e.,
free-flowing) Alternatively you may simply let the product air dry until the next
class period. Calculate your theoretical yield (based on the actual quantity of
limiting reagent used) and your percent yield, which should be in the range 35-
65%. Clean the glass frit by running dilute hydrochloric acid followed by water
through the frit.
Question: You can readily identify the peak due to free or uncoordinated
carbonate in the
CaCO3 spectrum. Without consulting the literature, how can you determine which
peak is due to nitrateʼs NO stretch and which are due to carbonateʼs CO
stretches in [Co(NH3)4(CO3)]NO3? Justify you answer.
Partners should combine their products in order to perform the second synthesis.
Cool the solution to about 50°C. Add 20 mL of concentrated HCl slowly while
gently swirling the mixture. Reheat the mixture for 20 to 30 minutes to 75-85°C;
observe the color change. Cool the solution to 5 °C in an ice bath. Collect the fine
purple-red crystals of [Co(NH3)5Cl]Cl2 by filtration using a medium glass frit and
wash twice with 3 mL of ice-cold distilled water, filtering each time. Wash twice
with 3 mL of ethanol and filter. Dry in an oven at 100-120°C for one hour, cool,
weigh, and calculate the percent yield. Transfer your product to a vial and
return the glass frit to the Stockroom. Typical yields vary between 50 and
75%.
Obtain an IR spectrum of the product and compare it with the spectra obtained
above.
Goals
• Become familiar with the Cary 100 spectrometer by recording the UV-VIS
spectrum (λ = 350-700 nm) of each of 4-6 solutions. Determine the λmax
for [Co(NH3)5Cl]Cl2 in 0.5 M nitric acid, HNO3 solution.
Prepare four to six solutions (each person should prepare two solutions) of
known [Co(NH3)5Cl]Cl2 concentration made up in 25 mL volumetric flasks. Weigh
out your samples first on weigh paper to approximate the correct mass. Next,
tare the flask (zero the balance with the flask in place), then carefully add the
weighed quantity of [Co(NH3)5Cl]Cl2 and finally weigh the flask with the cobalt
complex to 0.1 mg. Rinse any compound adhering to the neck of the flask into
the flask and fill to the 25 mL line with 0.5 M HNO3 (Helpful hints: add part of the
acid solution, swirl to mix getting most of the complex in solution, then use a
Pasteur pipet to add the nitric acid solution dropwise until the meniscus rests on
the calibrated line). Concentrations between 2.5 and 10 mM would be most
useful. Why? Be sure to span a range of concentrations. Calculate the mass
required for 2.5 and 10 mM solutions as part of the pre-lab.
Plot absorbance (A ) versus concentration (c). Don't forget to include the (0, 0)
data point in your calculations. Include the least squares line and correlation
coefficient on the plot. If the Lambert-Beer equation is valid over this range of
concentrations, a straight line should be obtained whose slope is equal to εl, or in
fact to ε if l = 1 cm. What is the error in ε? How should ε be reported?
10
Weigh out two samples of [Co(NH3)5Cl]Cl2 accurately (to 0.1 mg) by taring a 25-
mL volumetric flask and then reweighing it plus the sample of [Co(NH3)5Cl]Cl2.
The amounts should be different, for example ~20 mg and ~40 mg. Label both
flasks. Dilute to the 25 mL mark with an aqueous solution of 0.5 M HNO3.
Remove a sample from each flask with a clean pipette and transfer to a glass
cuvette. Select the scanning kinetics icon on the Cary 100 Spectrophotometer
and follow the instructions outlined in Appendix 1, part B. Obtain a baseline
spectrum of the 0.5 M nitric acid solution. Place the cuvettes in the sample holder
of the Cary 100 Spectrometer and equilibrate for 15 minutes at 60ºC.
IMPORTANT NOTE: There will be several teams using each instrument,
requiring coordination of the start time for the kinetics runs. The spectrometer will
record spectra every 20 minutes. Manually record the time and absorbance at
550 nm for each spectrum and record the location of all isosbestic points or
transfer your data files to a USB drive for analysis.
Calculate the expected value for A∝ and construct a table of ln(A-A∝) and time t
for the two solutions. Continue to follow the reactions until the value of A-A∝ is
approximately half the value of Ao-A∝, where Ao is the expected absorption of
[Co(NH3)5Cl]Cl2 alone at 550 nm, which you know from the quantity you weighed
out and ε for [Co(NH3)5Cl]Cl2 550 nm. The time at which A-A∝ is half the value of
Ao-A∝ is the half-life for this first order reaction, i.e., the time required for it to
proceed halfway. After five half-lives the reaction is essentially
finished, but you will have time to follow the reaction for approximately only one
half-life.
Note carefully the presence of one or more isosbestic points. An isosbestic point
is found at some value of λ where the absorbance of the mixture does not
change during the reaction. A "clean" set of isosbestic points is characteristic of a
simple conversion of one absorbing compound into another.
**CAUTION** Please ensure that you use the cuvettes with Teflon stoppers
for your runs ONLY. Cuvettes do not need to be filled more than ~75% full
for accurate readings.
Each partner will carry out a kinetic study at another temperature, one higher
(e.g., 65 °C) and one lower (e.g., 40 °C). It is desirable to have the minimum and
maximum temperatures differ by at least 25 °C, but be aware that a temperature
that is too high can lead to a reaction that is inconveniently fast to measure, while
a temperature that is too low can lead to a reaction that is inconveniently slow to
measure during a given lab period. For these runs monitor the reaction progress
with the spectrophotometer set to 550 nm. Select the kinetics icon (not scanning)
11
IMPORTANT NOTE: The group will have to come to a consensus on the two
temperatures to be studied. One UV-VIS spectrometer will be set to the lower
temperature and the other UV-VIS will be set to the higher temperature. Again
cooperation will be required, as all teams will need to start their kinetic runs
simultaneously. Manually record the time and absorbance at 550 nm for each
data point or transfer your data files to a USB drive for analysis.
1. Plot ln(A-A∝) versus t in order to obtain k in each kinetic run. Express k in s-1.
Include
2. Determine the activation energy for the aquation reaction over the
temperature range you studied by plotting the various rate constants you
versus 1/T. Include the least squares line and correlation coefficient on the plot.
3. Using the value for the activation energy, predict the rate constant for the
aquation reaction at 22 °C. How long you would have to monitor the reaction at
(essentially
C2, where C1 and C2 are constants. (C2 is actually [Co]o, the concentration of
6. You could have chosen to follow the reaction at a different wavelength. What
12
7. Where did you observe isosbestic points in your spectra? What is true in terms
of
absorption by the reactant and the product at these wavelengths?
10. Rationalize why nitrate is a poorer ligand than carbonate, i.e., why
[Co(NH3)4(CO3)]NO3 is not actually [Co(NH3)4(NO3)]CO3.
12. The colors of the cobalt complexes can be assigned to relatively weak ("not
allowed") "dd" transitions. If (t2g)6(eg)0 is the ground state of a given complex, what
is the excited state configuration after absorption of a photon.
13
To launch the software click on the icon Cary WinUV. Follow the column on the left of the
table below for common commands and the details for the particular experiment in the
appropriate column to the right. Note: items not checked off on the screen or not highlighted for
change are not listed below.
2
Desktop/My Computer/Local Disk C:/Varian/CaryWinUV
14
Set-up: Options x Auto Lamp Off x Auto Lamp Off
SBW(nm): 2.0 SBW(nm): 2.0
Energy: 1.00
Source: UV-VIS Source: UV-VIS Source: VIS
Source changeover: Source changeover: Source changeover:
350 nm 350 nm 350 nm
> Overlay data > Individual data
x Show Status x Show Status x Show Status
Display Display Display
Setup: Baseline > Baseline > Baseline Not available
correction correction
x Show Status x Show Status
Display Display
Setup: Accesories 1 Cells: Cells: Cells:
x Use Cell Changer x Use Cell Changer x Use Cell Changer
> Select cells > Select cells > Select cells
x check off cells to be x check off cells to be x check off cells to be
used Cell 1-Cell 6 used Cell 1-Cell 6 used Cell 1-Cell 6
> 6x6 > 6x6 > 6x6
Temperature: Temperature: Temperature:
x Automatic x Automatic x Automatic
Temperature Temperature Temperature
Setting Setting Setting
> Temperature > Temperature > Temperature
Controller Controller Controller
Block 25.0°C Block 60.0°C Block 45.0 OR 75.0°C
Temperature Temperature Temperature
Display: Display: Display:
x Block x Block x Block
x Show Status x Show Status x Show Status
Display Display Display
Setup: Accesories 2 Not Used Not Used Not Used
Setup: Accesories 3 Not Used Not Used Not Used
Setup: Analyze Not Available Not Used Not Used
Setup: Accesories 3 Not Available Not Available Fill in:
Number of Samples
Sample Names
15
16
17
To open the files in Excel use the least restrictive criteria to choose the file.
Explore the various icons immediately above the graph for scaling graphs, using a cursor,
examining multiple plots, selecting which traces to view and the like.
Please be sure the Cary100, temperature controller and computer are shutdown at the end
of the day.
Note: lifting up on the knob in the sample compartment will facilitate removal of the
cells.
18
MIT OpenCourseWare
http://ocw.mit.edu
For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.