0% found this document useful (0 votes)
746 views1 page

AU Routine Cleaning Solutions

This document provides instructions for preparing various cleaning solutions used in an AU routine cleaning process. It lists the bottle names, solutions, and preparation instructions for combining specific amounts of wash solution, bleach, hydrochloric acid, or sodium hydroxide with deionized water and transferring to labeled bottles. Special instructions are given for cleaning sample pots and using different solutions for laboratories with higher sample volumes or dialysis patients.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
746 views1 page

AU Routine Cleaning Solutions

This document provides instructions for preparing various cleaning solutions used in an AU routine cleaning process. It lists the bottle names, solutions, and preparation instructions for combining specific amounts of wash solution, bleach, hydrochloric acid, or sodium hydroxide with deionized water and transferring to labeled bottles. Special instructions are given for cleaning sample pots and using different solutions for laboratories with higher sample volumes or dialysis patients.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

AU routine cleaning solutions 

Bottle  Solution  Preparation Instructions 


Name/Use 
    Combine: 
Det‐1  2% Wash  2mL of Wash Solution + 98mL of DI H20 mix and  
transfer to the 60mL Det‐1 bottle 
Solution 
    Combine: 
Det‐2  2% Wash  2mL of Wash Solution + 98mL of DI H20 mix and  
transfer to the 60mL Det‐2 bottle 
Solution* 
Sample Probe    Combine: 
Wash Solution  2% Wash  2mL of Wash Solution + 98mL of DI H20 mix and  
W1 (AU480 only)  transfer to the 5mL Sample Probe Wash tube  
Solution 
     
CLN‐1  100%  Transfer undiluted Wash Concentrate to the 60mL CLN‐1 bottle 
 
Wash 
Solution 
    Combine: 
CLN‐2  20% Wash  20mL of Wash Solution + 80mL of DI H20 mix and  
transfer to the 60mL CLN‐2 bottle 
Solution 
    Combine: 
W2  10% Bleach   20mL of 5.25% Bleach +180mL of DI H20 mix and transfer to the 
60mL W2 bottle(s) (and 5mL tube ‐AU480 only) specifically 
designated for bleach. 
     
W2  1 Normal  Transfer 1 Normal HCl to the 60mL W2 bottle(s) (and 5mL tube – 
AU480 only) specifically designated for HCl. 
HCl 
    When processing HbA1c reagent (part #B00389), prior to 
W2  0.1 M  weekly maintenance a mandatory additional W2 cycle using 
0.1M NaOH is required. Transfer 0.1M NaOH to the 60mL W2 
NaOH  bottle(s) specifically designated for NaOH. (0.1 M NaOH part # 
B44081) 
Sample Pot    Combine: 
1% Wash  1mL of Wash Solution + 99mL of DI H20 mix and  
cleaning  use for Sample Pot cleaning 
solution  Solution   
*For laboratories with higher than average volumes of samples or dialysis patients, substitute 2% Wash 
Solution with 20% Bleach. Combine 10mL of Bleach + 40mL of DI H20 mix and transfer to the 60mL Det‐
2 bottle. Use Sodium Hypochlorite (part # A32319)  

AU routine cleaning solutions    Page 1 of 1 