Service Manual: MODEL FG-8110 Sweep/Function Generator
Service Manual: MODEL FG-8110 Sweep/Function Generator
MODEL FG-8110
SWEEP/FUNCTION GENERATOR
CONTENTS
2. ADJUSTMENT PROCEDURE
3. TROUBLESHOOTING PROCEDURE
4. PARTS LIST
5. CIRCUIT DIAGRAM
2
1. PERFORMANCE CHECK PROCEDURE
1-1. General
This section contains the Procedures required to check and maintain specified instrument
performance. The adjustments should be performed at an ambient temperature of
23°C ±5°C and a relative humidity of less than 60%. It allows the instrument to stabilize at
this environment for a minimum of 30 minutes.
ACCURACY
+/- 5 % OF SETTING
EQUIPMENT
FREQUENCY COUNTER
BNC CABLE
FEEDTHROUGH TERMINATION 50 Ω (if necessary)
PROCEDURE
2. Set 8210 as follows;
3
- Function ; SINE WAVE ( )
- ATT ; 0 dB(PULL OUT)
- AMPLITUDE ; CENTER
- COUNT ; INT(PULL OUT)
- DC OFFSET ; OFF(PUSH IN)
- SYM ; OFF(PUSH IN)
- SWEEP ; OFF(PUSH IN)
2. Connect the frequency counter to output terminal.
3. Set the frequency counter to frequency measurement.
4. Set the output frequency of this product by FREQUENCY RANGE and DIAL and verify
counter frequency reading as follows;
** For reading less than 200 mHz, you can read by period value.
200mHz ; 5.000 sec
100 mHz ; 10.000 sec
ACCURACY
Lower than 100mVp-p(IN TO 50Ω) at MIN.
Higher than 10Vp-p(IN TO 50Ω) at MAX.
4
EQUIPMENT
Oscilloscope
AC Millivoltmeter
BNC CABLE
FEEDTHROUGH TERMINATION 50 Ω
PROCEDURE
1. Set this product as follows;
- Function ; SINE WAVE ( )
- ATT ; 0 dB(PULL OUT)
- AMPLITUDE ; CENTER
- COUNT ; INT(PULL OUT)
- DC OFFSET ; OFF(PUSH IN)
- SYM ; OFF(PUSH IN)
- SWEEP ; OFF(PUSH IN)
2. Connect the oscilloscope through FEEDTHROUGH TERMINATION 50 Ω
to output terminal.
3. Set Amplitude Control to Min. position and check the output level is lower than
100mVp-p(less 5div in 20mV/div).
4. Set Amplitude Control to Max. position and check the output level is higher than
10Vp-p(higher than 4div in 5V/div).
5. Set Amplitude Control to Min. position and connect AC Millivoltmeter to output
terminal.
6. Set the Range of AC millivoltmeter to 3V range and adjust Amplitude control for
0dB reading on AC millivoltmeter.
7. Push in “20dB ATT switch” and check the reading of AC Millivoltmeter is -20 dB.
EQUIPMENT
Distortion analyzer
BNC CABLE
FEEDTHROUGH TERMINATION 50 Ω (if necessary)
PROCEDURE
1. Set this product as follows;
- Function ; SINE WAVE ( )
- ATT ; 0 dB(PULL OUT)
5
- AMPLITUDE ; CENTER
- COUNT ; INT(PULL OUT)
- DC OFFSET ; OFF(PUSH IN)
- SYM ; OFF(PUSH IN)
- SWEEP ; OFF(PUSH IN)
2. Connect the distortion analyzer to output terminal.
3. Check the distortions of output signal are less than 1% from 10Hz to 100kHz.
EQUIPMENT
Oscilloscope
BNC CABLE
FEEDTHROUGH TERMINATION 50 Ω (if necessary)
PROCEDURE
1. Set this product as follows;
- Function ; TRIANGLE WAVE ( )
- ATT ; 0 dB(PULL OUT)
- AMPLITUDE ; CENTER
- COUNT ; INT(PULL OUT)
- DC OFFSET ; OFF(PUSH IN)
- SYM ; OFF(PUSH IN)
- SWEEP ; OFF(PUSH IN)
2. Connect the oscilloscope to output terminal.
3. Check the linearity of Triangle wave from 10Hz to 100kHz.
4. Change the function to “SQUARE WAVE( )” and check the rise and fall time at
10MHz Max. output level.
EQUIPMENT
Oscilloscope
BNC CABLE
6
PROCEDURE
1. Set this product as follows;
- FREQ. RANGE ; 10k
- FREQ. DIAL ; 10kHz
- COUNT ; INT(PULL OUT)
- DC OFFSET ; OFF(PUSH IN)
- SYM ; OFF(PUSH IN)
- SWEEP ; OFF(PUSH IN)
3. Set the oscilloscope to 1V/div, 5uS/div and connect to TTL output terminal.
4. Check the rise and fall time of TTL SQUARE WAVE on CRT( 5div ).
EQUIPMENT
Oscilloscope
BNC CABLE
PROCEDURE
1. Set this product as follows;
- Function ; SQUARE ( )
- ATT ; 0 dB(PULL OUT)
- FREQ. RANGE ; 10k
- FREQ. DIAL ; 10kHz
- AMPLITUDE ; CENTER
7
EQUIPMENT
Oscilloscope
BNC CABLE
FEEDTHROUGH TERMINATION 50 Ω (if necessary)
PROCEDURE
1. Set this product as follows;
- Function ; SQUARE ( )
- ATT ; 0 dB(PULL OUT)
- AMPLITUDE ; CENTER
- COUNT ; INT(PULL OUT)
- DC OFFSET ; OFF(PUSH IN)
- SYM ; OFF(PUSH IN)
- SWEEP ; OFF(PUSH IN)
2. Connect the oscilloscope to output terminal.
3. Set the output frequency to 10kHz.
4. Pull out SYM Control and adjust to Min. position. And check the symmetry(duty
ratio) is “lower than 1:10”.
5. Adjust SYM Control to Max. position and check the symmetry(duty ratio) is”higher
than 10:1”.
EQUIPMENT
Oscilloscope
Frequency Counter
BNC CABLE
FEEDTHROUGH TERMINATION 50 Ω (if necessary)
PROCEDURE
1. Set this product as follows;
- Function ; SINE WAVE ( )
- ATT ; 0 dB(PULL OUT)
- AMPLITUDE ; CENTER
- COUNT ; INT(PULL OUT)
- DC OFFSET ; OFF(PUSH IN)
- SYM ; OFF(PUSH IN)
2. Connect the oscilloscope and Frequency Counter to output terminal.
8
3. Set Freq. Range to 10k and adjust Freq. Dial to 1.00 kHz (Start Frequency).
4. Pull out SWIDTH Control and adjust to max. position ( 100kHz, Stop frequency).
5. Adjust Sweep Rate Control for desired rate and check sweep operating.
EQUIPMENT
Power supply
BNC CABLE
PROCEDURE
1. Set this product as follows;
- Function ; SINE WAVE ( )
- Frequency Range ; 10K
- Frequency Control ; Max.
- ATT ; 0 dB(PULL OUT)
- AMPLITUDE ; CENTER
- COUNT ; INT(PULL OUT)
- DC OFFSET ; OFF(PUSH IN)
- SYM ; OFF(PUSH IN)
2. Set Power Supply to “0V” and connect to “VCF IN” Terminal.
3. Check the output frequency is “1.00kHz ± 0.2kHz”.
4. Increase the output voltage of Power Supply up to +10V and check the output
frequency as following table.
VCF Voltage Output Frequency Tolerance
0 1.000 kHz ± 0.2kHz
2.000 V 20.00 kHz ± 4 kHz
4.000 V 40.00 kHz ± 8 kHz
6.000 V 60.00 kHz ± 12 kHz
8.000 V 80.00 kHz ± 16 kHz
10.00 V 100.00 kHz ± 20 kHz
EQUIPMENT
9
Signal generator
PROCEDURE
1. Set the COUNT to EXT(PUSH IN).
2. Connect the signal generator to EXT COUNT IN terminal.
3. Set the signal generator to 100mVrms, 200mHz and check the reading of 8210.
4. Vary the output frequency from 200mHz to 110MHz and check the 8210 displays
correct frequencies.
5. Push in LPF(Low Pass Filter) and check the LPF is working (-3dB +/-0.2dB point at
approx. 120kHz).
< Check procedure >
a. Apply 120KHz signal with minimum input level into EXT COUNT IN.
b. Push in LPF switch and check gate time(Frequency display) is stopped.
c. Increase input level and check the gate time(Frequency display) is starting to
and check the input level is increased about 3dB(+/-0.2dB) from “minimum input
level”.
2. ADJUSTMENT PROCEDURE
10
2-4. PRE-ADJUSTMENTS
EQUIPMENT
Oscilloscope
BNC CABLE
FEEDTHROUGH TERMINATION 50 Ω (if necessary)
PROCEDURE
2-4-1. TRIANGLE WAVE OFFSET
1. Set this product as follows;
- Function ; TRIANGLE WAVE ( )
- ATT ; 20 dB(PUSH IN)
- AMPLITUDE ; MIN.
- FREQUENCY ; 10kHz
- COUNT ; INT(PULL OUT)
- DC OFFSET ; OFF(PUSH IN)
- SYM ; OFF(PUSH IN)
2. Set the oscilloscope to 0.5V/DIV, 20uS/DIV , DC Coupling and set the vertical
position to horizontal graticule center.
3. Connect the oscilloscope to output terminal.
4. Set “AMPLITUDE” control for 2DIV on CRT and adjust SFR15(OFFSET ADJUSTER) for “0
OFFSET of waveform”(The waveform should be in center. +/- 0.1DIV from horizontal
graticule center).
11
2-4-3.SINE WAVE LEVEL Adjust.
12
2-4-5. AMPLITUDE CONTROL AND ATTENUATOR CHECK
EQUIPMENT
Oscilloscope
AC Millivoltmeter
BNC CABLE
FEEDTHROUGH TERMINATION 50 Ω
PROCEDURE
1. Set this product as follows ;
- Function ; SINE WAVE ( )
- ATT ; 0 dB(PULL OUT))
- AMPLITUDE ; CENTER
- COUNT ; INT(PULL OUT)
- DC OFFSET ; OFF(PUSH IN)
- SYM ; OFF(PUSH IN)
- SWEEP ; OFF(PUSH IN)
2. Connect the oscilloscope through FEEDTHROUGH TERMINATION 50 Ω
to output terminal.
3. Set Amplitude Control to Min. position and check the output level is lower than
1Vp-p(less 5div in 200mV/div).
4. Set Amplitude Control to Max. position and check the output level is higher than
10Vp-p(higher than 4div in 5V/div).
5. Set Amplitude Control to Min. position and connect AC Millivoltmeter to output
terminal.
6. Set the Range of AC millivoltmeter to 3V range and adjust Amplitude control for
0dB reading on AC millivoltmeter.
7. Push in “20dB ATT switch” and check the reading of AC Millivoltmeter is -20 dB.
8. Check the amplitude of TRIANGLE & SQUARE WAVE same as paragraph 3 to 7.
PROCEDURE
1. Set this product as follows;
- Function ; TRIANGLE WAVE ( )
- ATT ; 20 dB(PULL OUT)
- AMPLITUDE ; CENTER
- COUNT ; INT(PULL OUT)
13
- DC OFFSET ; OFF(PUSH IN)
- SYM ; OFF(PUSH IN)
- SWEEP ; OFF(PUSH IN)
2. Connect the Frequency Counter to Output Terminal.
2. Set Frequency Range to “1M” and Frequency Control(Dial) to MAX. Position.
3. Adjust SFR1 for 10.500 MHz frequency reading on counter.
4. Change Frequency Range to “100k” and adjust SFR2 for 1.0500 MHz frequency
reading on counter.
5. Change Frequency Range to “10k” and adjust SFR3 for 105.00 kHz frequency
reading on counter.
6. Change Frequency Range to “1k” and adjust SFR4 for 10.50 kHz frequency
reading on counter.
7. Change Frequency Range to “100” and adjust SFR5 for 1.050 kHz frequency
reading on counter.
8. Change the frequency range to 10 and 1, and check the frequency reading
in each range. (10 : 105.00 Hz , 1 : 10.500Hz)
9. After adjustment, check the frequency Control(Dial) at MIN. and MAX.
as following table.
PROCEDURE
1. Set this product as follows;
14
- Function ; SQUARE ( )
- ATT ; 0 dB(PULL OUT)
- AMPLITUDE ; CENTER
- COUNT ; INT(PULL OUT)
- DC OFFSET ; OFF(PUSH IN)
- SYM ; OFF(PUSH IN)
- SWEEP ; OFF(PUSH IN)
2. Connect the oscilloscope to output terminal.
3. Set the output frequency to 100kHz.
4. Pull out SYM Control and adjust to Min. position. And adjust SFR6 the
symmetry(duty ratio) is “lower than 1:10”.
5. Adjust SYM Control to Max. position and adjust SFR7 the symmetry(duty ratio)
is”higher than 10:1”.
EQUIPMENT
10MHz Standard Signal Source
Signal Generator up to 110MHz
BNC CABLE
FEEDTHROUGH TERMINATION 50 Ω (if necessary)
PROCEDURE
1. Set this product as follows;
- COUNT ; EXT(PUSH IN)
2. Connect 10MHz Standard Signal to the EXT COUNT IN terminal and adjust TC1 for
10.0000 MHz reading on Counter Display.
3. Set Signal generator to 110MHz, 100mVrms and connect the EXT COUNT IN terminal.
4. Adjust SFR16 (Shield Box IN) for 110.0000 MHz reading on Counter Display.
Reduce the output level of Signal Generator Slowly and adjust SFR16 for Minimum
input sensitivity(as possible as less than 100mVrms).
5. Change the out frequency of Signal Generator from 110MHz to 200mHz and check
the minimum input voltage is less than 100mVrms.
6. Push in LPF(Low Pass Filter) and check the LPF is working (-3dB +/-0.2dB point
at approx. 120kHz).
< Check procedure >
a. Apply 120KHz signal with minimum input level into EXT COUNT IN.
b. Push in LPF switch and check gate time(Frequency display) is stopped.
c. Increase input level and check the gate time(Frequency display) is starting to
and check the input level is increased about 3dB(+/-0.2dB) from “minimum input
level”.
15
3. Troubleshooting Procedure
3-1-1. Check the function selector and switch setting. Incorrect switch settings can
give a false indication of instrument malfunction. If there is any question
about the correct function or operation, refer to OPERATOR'S MANUAL.
3-1-2. Check associated equipments before proceeding, ensure that any equipments used
with this product is operating correctly and verify that out and input signals
are properly connected and that the interconnecting cables are not defective.
3-2-1. NO Display
1. Check the “FUSE” and if it is broken, change to new one with same rating.
2. Check the output voltage of power transformer.
-. 17 to 18 Vac between pin 1 and 2 of CN14, and between pin 2 and 3 of CN14
on main board.
-. 8 to 10 Vac between pin 1 and 2 of CN4,CN24, and between pin 2 and 3 of
CN4,CN24 on main board.
3. Check the DC Power Voltage.
-. +5VDC at U6 and U15 output(U6,15 Assembled at rear panel)
4. Check the frequency at pin 19 of U12. It should be “10MHz”.
5. Check the connector cable from CN12’,CN13’(main board) to display board.
16
Q1,Q2,Q3,Q4,Q5,Q6,D1,D2,D3,D5,Q7).
5. Check the Range Selector Switch operation and Frequency Control Volume and Cable
connection from VR1 to CN2(Main Board).
6. Check the “Amplitude VR4” and Cable connection from VR to CN5 on main board.
7. Check the waveform at R46. The Triangle Wave should be appeared.
8. If not, check the operation of Differential Amplifier(Q12,Q13,Q14,Q15,Q16).
9. Check the Attenuator switch(20 dB) and cable connection from CN7 to Output terminal.
17
3-2-7. Frequency Range and Frequency Dial
1. Check the cable connection from CN2(Main Board) to Frequency Control Volume.
2. Check the operation of Range selector switch(Key Switch operation).
3-2-8. Attenuator(20dB)
1. Check the operation of each Attenuator switch.
2. If not working, check the R146,147,148
18
4. PARTS LIST
4-1.. Main PCB Parts List
19
Part Name Specification q"ty Referance.
20
Part Name Specification q"ty Referance.
21
4-2.. Display PCB Parts List
Part Name Specification Q'ty Referance
NUT(VR) 6.00 VR
22
4-5. Rear Panel Assembly Parts.
BRACKET, POWER
23
4-6. Others.
SHIELD CASE 8210 TOP 0.6t 210x202(156)x35 1.00 MOUNT POLE- TOP
24
4-7. Packing
5. CIRCUIT DIAGRAM
- Refer to Attached File.
25
5 4 3 2 1
R83
10K
COUNTER
+5V(A)
D D
06-R150#1
+
C40
47/16V C39 R87 C48 + C47
+5VD 103 510 47/16V 103
C44 R95
470P(C/C) 330 R99
330
R85 Q20
BNC4 100K R88 U8A 2N3906 R102
1/16
EXT COUNT IN Q18 150 U8B R92 MC10116 U8C 22
1
2N5486 C9 MC10116 150 2 MC10116
1 3 47/16V 7 4 15
10 13
C37
+
5
474/250V
2
2
9 3 12
SW5 C38 6 14 Q21
EXT/INT SW 130P C41 2N3906
11
130P
8
Q19 R93 R94 R96 R97 R98
C D10 D11 2N3906 470 470 470 470 330 R103 C
3
1N4148 1N4148 R100 R101 120
R86 R89 470 470
510 22K
SFR16 2
10K
R90 R91
+ 2.2K 2.2K
C43
1
47/16V
15 A QA 3
R108 D16 C91 + C46 C45 1 2
470 1N4148 103 47/16 103 B QB
10 C QC 6
9 D QD 7
RCO 13
14 CLK
+5VD 4
C90 D17 G
5 D/U
103 1N4148 11
SW6 LOAD
MX/MN 12
L.P.F
C12 U18
103 74LS191
B 06-U9#3 B
06-U9#5
06-U9#4
A
Date 2002/04/13 DAGATRON A
5 4 3 2 1
5 4 3 2 1
/Sa mHZ D1
/Sb HZ D2
D D
/Sg
/Sa
/Sb
/Sa
/Sb
/Sg
/Sa
/Sb
/Sa
/Sb
/Sg
/Sa
/Sb
/Sa
/Sb
/Sg
/Sa
/Sb
/F3
/F4
/F5
/F6
/F7
/F8
/F9
/Sf
/Sf
/Sf
/Sf
/Sf
/Sf
/Sf
/Sc KHZ D3
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
1
2
3
4
5
/Sd MHZ D4
FND1 FND2 FND3 FND4
SEGMENT SEGMENT SEGMENT SEGMENT /Se G.T D5
+5V(D)
8.8. 8.8. 8.8. 8.
18
17
16
15
14
13
12
11
10
18
17
16
15
14
13
12
11
10
18
17
16
15
14
13
12
11
10
10
C49 C50 C51 C52 C56 C53
9
8
7
6
104 104 104 104 104 104
/Se
/Sd
/Sh
/Se
/Sd
/Sg
/Sh
/Se
/Sd
/Sh
/Se
/Sd
/Sg
/Sh
/Se
/Sd
/Sh
/Se
/Sd
/Sg
/Sh
/Se
/Sd
/Sh
/Se
/Sc
/Sc
/Sc
/Sc
/Sc
/Sc
/Sc
/F3
/F4
/F5
/F6
/F7
/F8
/F9
U9
PAL16R4 J13
8
7
6
5
4
3
2
1
C 12 F0 I0 1 CN13 C
13 F1 I1 2
14 3 J13'
F2 I2 05-U18#14
15 4 CN13'
F3 I3 05-U18#7
1
2
3
4
5
6
7
8
16 F4 I4 5 05-U18#6
17 6 U13 J3 J3'
F5 I5 TB62783 CN3 CN3'
1
2
3
4
5
6
7
8
18 F6 I6 7
19 8 1 18 1 /Sa
F7 I7 I0 O1 1 1
/CO0
/CO1
/CO2
/CO3
/CO4
/CO5
/CO6
/CO7
9 2 17 2 /Sb
I8 I1 O2 2 2 /Sc
I9 11 3 I2 O3 16 3 3 3
4 15 4 /Sd
I3 O4 4 4 /Se
5 I4 O5 14 5 5 5
/CO0
/CO1
/CO2
/CO3
/CO4
/CO5
/CO6
/CO7
6 13 6 /Sf
I5 O6 6 6 /Sg
7 I6 O7 12 7 7 7
U12 8 11 8 /Sh Q22 Q23 Q24 Q25 Q26 Q27 Q28 Q32
89C52 I7 O8 8 8 A1015 A1015 A1015 A1015 A1015 A1015 A1015 A1015
U10A
74LS393 16 WR /a
1 A QA 3 17 RD P1.0 1
4 13 2 /b R105 R106 R107 R108 R109 R110 R111 R112
QB INT1 P1.1 /c 10 10 10 10 10 10 10 10
QC 5 P1.2 3
2 6 21 4 /d
CLR QD P2.0 P1.3
/CM0
/CM1
/CM2
/CM3
/CM4
/CM5
/CM6
/CM7
22 5 /e
U10B P2.1 P1.4 /f
23 P2.2 P1.5 6
B 74LS393 24 P2.3 P1.6 7 /g
/h U14
B
13 A QA 11 25 P2.4 P1.7 8
10 26 74LS138
QB P2.5 /CM0 +5V(D)
QC 9 27 P2.6 RXD 10 1 A Y0 15
12 8 15 11 2 14 /CM1
CLR QD T1 TXD B Y1 /CM2
INT0 12 3 C Y2 13
U11A 28 12 /CM3 R149
74LS393 P2.7 +5V(D) +5V(D) Y3 /CM4 10
14 T0 Y4 11
1 3 6 10 /CM5
A QA G1 Y5 /CM6
QB 4 39 P0.0 EA/VP 31 4 G2A Y6 9
5 38 9 5 7 /CM7
QC P0.1 RESET + C55 G2B Y7 Q30
2 CLR QD 6 37 P0.2 X1 19 05-SW5#5
36 10/16V C1815
U11B P0.3 Y1 R150
35 P0.4
74LS393 34 10MHZ 1K
P0.5
ALE/P
PSEN
13 A QA 11 33 P0.6
QB 10 32 P0.7 X2 18
9 R104
QC 8.2K
12 8
29
30
CLR QD T1
C64 C57 20P
30P 56P
Title 8210 - 06 Drawing Design Checked Approved
5 4 3 2 1
5 4 3 2 1
D8
1N4148
J8
3
SW1 CN8 R70 +15V +15V
SWEEP WIDTH 1K
1
3 2
Sweep Width VR5 2 1 3
1
10KB 8
D 2 D
SFR8
1
2.2KB
-15V
+15V
C36
3
R73 0.22uF(MYLER)
7.5K R75 C31 + C30
3
J9 5.1K 0.01(C/C) 22/16V R11
CN9 R71 2.7K
39K R74
VR6 1 5.1K R76
Sweep Rate 2 2 9 -
02-U30A#12
4
10KB 8 13 - 5.1K
3 Q17
10 + 14 5 +
U2C 12 + 2N3906 7 Q2
GL324 U2D 2N3904
1
6 -
R72 GL324 U2B
33 GL324
11
D9 R77
1N4148 10K
+
C33 C32
C 0.01(C/C) 22/16V C
-15V
BNC1 J1 R300 R2 R4
VCF IN CN1 470 1K 220
1 1
3
2
C304 R1 R9
2
R3
1K
1
5 +
7 2 -
3
J2 6 1 Q1
-
CN2 U1B 3 + 2N3906
GL324 U1A
VR1 1 GL324
Frequency 10KB
2 2 R8 R10
3 1K 2.2K
R114
1
B 51
SFR1 +21V
B
2
10KB
3
3 1
1
2
3
SFR3 10 + 1
2
10KB 8
3 1 9 - 02-U30B#3
U1C 2 VR2
+15V GL324 50KB
11
SFR4
2
10KB
1
3 1
R6
1K SFR5
2
10KB 3
3 1 1
A R7
C2
0.22uF 22/16
8 Title 8210 - 01 Drawing Design Checked Approved A
+
7.5K C37 SW2
Symm
Model FREQ/VCF IN CIRCUIT
-5V
Size A3 Rev 02
Sheet 1 of 6 Scale 2:1
02-R19#1 +15V
02-U4A#1
02-R32#1
C76
R126 0.1(MON)
D +5V R46 C19 100 R138 D
2K 0.1(MON) R124 C67 56.2F R140 D5
3
1K 0.1(MON) 10(1/2W) 1N4148
R117 U17
R69 100KF CA3096 C78
10 2 1 16 R125 0.1(MON)
Q1B SUB 1K R127 C75
2 Q1E
SFR15 R118 3 15 825F 0.1(MON) + C77
100K 220 Q1C Q5C 100/25V
Q5B 14
1
4 Q2E Q5E 13
C35 C34 5 R139
22/16V + 0.01 SW1 SW2 SW3 Q2B 21.5F
6 Q2C Q4C 12
SINE TRIANGLE SQUARE 11 R128 Q14
Q4B 150 2N3563
7 Q3E Q4E 10
R119 8 Q15
3.32KF Q3B 2N3906
Q3C 9
R120 R121 Q16
42.2F 576 2N2219
R147
2
240
3 1 D6
C17 C66 C71 1N4148
SFR10 0.1(MON) 150P R132 0.1(MON) R146 R148
1KB 4.99KF 62 62
C R37 C
1KF C68
R33 C87 0.1(MON)
2K 22P R135 R141
Q10 10(1/2W) 10(1/2W)
C3198 R129
24.9KF R144
100(1/2W)
R41 BNC3
270 R115 R134 R136 6 1 OUTPUT
R42 249 1.5KF 2.2KF
1
R40 220 5 2 1
33K
Q8 Q9 2 Q11 4 3
2
C1815 C3198 C3198 R122
2
1.1K SW3
3
1
3
2
2 2 0.1(MON) D7
3 1 R43 1 R123 1N4148
R34 R42 1K C18 1.21K Q29
330 SFR11 1K 0.1(MON) 2N2905
100B R116 R47 R160 Q13
1
B R35 R38
VR3
AMP 500B
10K 22 1K
Q12
2N3904 B
3
CN6 100/25V
+
SFR13 SW4 3 C79
1KB DC OFFSET 2
R36 R137 R136
1
2 1 2 1
1K 21.5F 144F R74
R131 C70 10(1/2W)
R161 100 0.1(MON)
VR4 1K C81
1
5K 0.1(MON)
R44 -15V
C14 + 10 C15
22/10V 0.01uF(CER)
A
Title
8210 - 03 Drawing Design Checked Approved
A
Model AMPLFIER CIRCUIT
-5V
Size A3 Rev 02
Sheet 3 of 6 Scale 2:1
5 4 3 2 1
5 4 3 2 1
+21V
+21V
R68
D 10/0.5W + C20 D
2200/35V
+15V
DB1
+15V
1
J14' J14 DB06M
CN14' CN14 U5 U6 +5V
7815 7805
1 1
2 2 2 - + 4 1 VIN VOUT 3 1 VIN VOUT 3 +5V
GND
GND
3 3
+ C21 + C22 + C23
1000/35V 22/16V 22/16V
2
U7
7915
2 VIN VOUT 3 -15V
GND
C25
+ +
C24 22/16V
C 1000/35V -15V C
1
R66
560/1W
-5V
ZD1 +
5.1V C27
47/35V -5V
+5V(D)
T2
TRANSFORMER POWER
D12 U15
11 1 J4' J4 1N4003 7805
2 CN4' CN4 1 3
VIN VOUT +5V(D)
SW8 ST1 J16' J16 3
GND
F1 F2 J15' J15 POWER S/W SELECT CN16' CN16 1 1
2 2
B NOISE FILTER FUSE CN15' CN15
1 1 4 3 3
C59
104
+ C58
2200/16V
+ C60
470/16V
C61
104
B
1 1 1 1 3 2 2 5
2
3 2 2 3 3 6
2 2 4 D13
3 3 1N4003
7
8
10 9
FAN
D20 U20 J17 DC FAN
C306 C307 J5' J5 1N4003 7805 CN17
104/1KV 104/1KV CN5' CN5 1 3 1 2
VIN VOUT 1
GND
1 1 2
2 2 C201 +
3 3 + C200 470/16V C203
C202 470/16V 104
2
D22 104
1N4003
5 4 3 2 1
5 4 3 2 1
+15V
D D
R12 R13 +5V
1.2KF 120KF L2
INDUCTOR AUDIO
L1
INDUCTOR AUDIO
1
R14 Q3
13 - 10K 5771 3 Q5
14 2N5486
12 + C3
01-R11#1
U30D 33P C101
2
072 R23 10n(C/C)
C4 11KF
14
680P(BOX) D3
C85 1N4148 4 BNC2
472 C5 6 R27 J10 R170 OUTPUT
C 6.8n(BOX) 5 33 CN10 33 C
R24
14
1 1
6
8
D1-1 C6 C8 430F U4B
HSMS2825 68n(BOX) D4 0.01(C/C) 74LS00 R28 C13 2
S
2G
VCC+
1N4148 22 150P TTL SYNC OUT
2
11 2B 2Y 9
12 2A
Q7 C10
2N3563 130P
C86 2 R30 R29
472 1B 4.7K 4.7K
1 1A 1Y 4
R001 C84 U4
VCC-
Q4 ??? C7 30P(C/C) SN75107
1G
3 3563 0.68n(BOX)
01-CN3#1 +
1 R20 R21 1
13
1
1K 1KF
5
2 - 3
U30A R15 3 Q6 2
072 10K 2N5486
L3 U4A
INDUCTOR AUDIO 74LS00
2
R31
240
B R24
R25
11KF R32
B
430 -5V 2.2K
R18 R19 C9
10 1.5K 15P(C/C)
R17
R16 120KF
1.2KF
03-SW3#6
03-SW3#2
03-R46#1
Shield Box
A
Title
8210 - 02 Drawing Design Checked Approved
A
Model OSCILLATOR/TTL CIRCUIT
Size A3 Rev 02
Sheet 2 of 6 Scale 2:1
5 4 3 2 1