BS-120 Chemistry Analyzer Quick Operation Card: Checking Before Startup
BS-120 Chemistry Analyzer Quick Operation Card: Checking Before Startup
2 Power-on
Power on the analyzer in the sequence presented below:
MAIN POWER of the analyzing unit Æ Power of the analyzing unit Æ Monitor of the operation unit Æ
Computer of the operation unit Æ Printer
4 Setting Parameters
Before requesting tests, you must finish the following settings:
9 To set the options regarding the basic parameters of the system, click Setup Æ System.
9 To set the options regarding the hospital and doctor information, click Setup Æ Hospital.
9 To set the options regarding parameters of calibrators, click Calibration Æ Calibrator.
9 To set the options regarding parameters of controls, click QC Æ Control.
9 To set the options regarding test parameters, reference, calibration rule and quality control (QC) rule,
click Parameter Æ Test.
9 To set the options regarding the reagent parameters, click Reagent.
9 To set the options regarding the carryover information among tests, click Parameter Æ Carryover.
9 To set the options regarding the printing parameters, click Setup Æ Print.
5 Preparing Reagents
Load reagent bottles to their assigned positions on the reagent disk, and then open the bottles.
6 Reagent Blank
You can conduct reagent blanks as needed.
To request reagent blanks, click Calibration Æ Calibration Request.
To run reagent blanks, click Start.
To view the reagent blank results, click Calibration Æ Results.
7 Calibration
1
You can conduct calibrations as needed.
Note: You need to run the calibration again when you change the reagent lots, test parameters,
lamp or other analysis conditions.
To request calibrations, click Calibration Æ Calibration Request.
After requesting calibrations, you should load corresponding calibrators to their assigned positions on the
sample disk.
To run calibrations, click Start.
To view the calibration results, click Calibration Æ Results.
If an ISE module is connected, click Maintenance Æ ISE Æ Daily Maint to perform ISE calibration.
8 QC
To request QCs, click QC Request.
After requesting QCs, you should load corresponding controls to their assigned positions on the sample
disk.
To run QCs, click Start.
To view the QC results, click QC Æ Real-time/ Daily QC / Day to Day QC.
9 Samples
To request samples, click Sample Request.
Note: STAT samples are requested in the same way as routine ones except that STAT on the
Sample Request screen should be selected when requesting.
After requesting, you should load corresponding samples to their assigned positions on the sample disk.
To run samples, click Start.
To view the sample results, click Results.
13 Shutdown
After exiting the Windows, switch off the following powers in the presented order below:
Printer Æ Monitor of the operation unit Æ Power of the analyzing unit
Note: The refrigerator still functions after the Power of the analyzing unit is placed to OFF. To shut
down the refrigerator, place the MAIN POWER of the analyzing unit to OFF.