LAB No. 05 Familiarization With Microwave Equipment: Objective
LAB No. 05 Familiarization With Microwave Equipment: Objective
05
Objective:
The objective of this exercise is to familiarize the students with the equipment they will be using
in this laboratory. All equipment and their description are provided below. We have two type of
microwave trainers.
The IT -8001 is a X-band signal generator, this instrument is especially designed for
laboratory to carry on a wide range of research activities in the X-band.
ExtSYNC:
ExtSYNC stands for “External Synchronization”. This is selected when the SSG external
sync from any other source. The ExtSYNC is shown in figure 1.2.
IntAM:
IntAM stands for “Internal Amplitude Modulated Wave”. It is an internal amplitude
modulated wave, modulated using a square wave and pulse wave. This can be controlled using
the knobs shown in figure 1.2.
RF Out:
This is the output section of the SSG; whenever we need a signal from SSG to be
delivered out to any device we use this connector. You can connect the BNC connector as a
frequency source to any RF waveguide adaptor.
2. Power Meter:
Power meter is a dynamic range device used for the measurement of power inside the
microwave waveguides. The power meter is shown below:
Specifications:
1) The power meter needs a 30 minute warm up to function properly and display stable
readings.
2) The working error of this meter is ±5%.
3) The N-connector coaxial cable connects to its input port to be able to measure the power
of the signal. The N-Connector is shown below:
4) The range of the power meter is 0.01uW to 300mW. The range can be varied by the knob
(can be seen in picture below). The results on this meter can be displayed in digital and
analog.
5) Call Factor: The calibration Factor can be adjusted to requirement by using the Cal Factor
toggle provided on the power meter.
3. VSWR Meter:
Specifications:
1) Gives you the ability to adjust dB toggle, BW toggle, Gain and Tune as per your
requirements
2) Adjustment of the meter on normal or 5dB mode can be made through the following
switch
4. The Variable Attenuator:
As the name suggests a variable attenuator is a device that is used to reduce power level
at the input of the microwave component and the attenuation measurements.
There are two types of variable attenuators; they are Rotary Vane Attenuator (not
available in laboratory) and Side Vane Attenuator (available in laboratory). In the Side Vane
Attenuator a plastic fiberglass blade with a resistive coating is used to produce attenuation. The
blade is inserted vertically into the waveguide, parallel to the short side’s walls. The resulting
attenuation depends upon the position of the blade in the waveguide. Attenuation is increased by
the moving the blade towards the center of the waveguide. Attenuation characteristics differ from
one attenuator to the next hence each attenuator has to be individually calibrated.
1. Objective: This shows the target to be achieved after completing the experiment.
2. Introduction: This gives a brief prerequisite knowledge for the required experiment.
3. Reading book: This gives a list of book to strengthen the knowledge of the topic.
4. Apparatus needed: Gives the list of apparatus required to carry out the experiment.
5. Experiment Procedure: This is the guide that must be followed in the process of doing the
practical activities, since it is closely related to the result.
The microwave trainer consists of different components and peripherals equipment required to perform
the experiments.