User’s AXF

Manual Magnetic Flowmeter


Integral Flowmeter/
Remote Flowtube
[Hardware Edition]
Manual Change No. 08-003-E

Please use the attached sheets for the pages listed below in IM 01E20D01-01E (9th).

Page and Item Contents of Correction


Page 6-11
CSA Added the "Process Sealing Certification".

Page 6-36
Table of optional Sepecifications Added the "Dual Seal" to descriptive text of
"CSA Certifications (CF1)".

Page 8-5
8.3 CSA Added the "Process Sealing Certification".

Apr. 1, 2008
Yokogawa Electric Corporation
6. OUTLINE

CSA: (Remote Flowtube)


*AXF002C – AXF400C For CSA C22.2 Series
Applicable Standard: Explosion proof for Class I, Division 1, Groups A, B, C &
For CSA C22.2 Series; D.
C22.2 No 0, C22.2 No 0.4, C22.2 No 0.5, Dust-ignition proof for Class II/III, Division 1, Groups E,
C22.2 No 25, C22.2 No 30, C22.2 No 94, F & G.
C22.2 No 157, C22.2 No 1010.1 Intrinsically safe (electrodes) for Class I, Division 1,
For CSA E79 Series; Groups A, B, C & D.
CAN/CSA-E79-0, CAN/CSA-E79-1, “SEAL ALL CONDUITS WITHIN 50 cm OF THE
CAN/CSA-E79-7, CAN/CSA-E79-11, ENCLOSURE”
CAN/CSA-E79-18 “WHEN INSTALLED IN DIV. 2, SEALS NOT REQUIRED”
Certificate: 1481213
Electrode Circuit Um: 250 Vac/dc
Process Sealing Certification:
Excitation Circuit: 170V max
Dual Seal certified by CSA to the requirements of
Enclosure: Type 4X
ANSI/ISA 12.27.01.
Temperature Code:
No additional sealing required.
Primary seal failure annunciation; Temperature Maximum Process Minimum Process
Deterioration of the flowrate output at nonzero flow Code Temperature Temperature

point. T6 +70°C (+158°F) –40°C (–40°F)


Unstable flowrate output at zero flow point.
T5 +85°C (+185°F) –40°C (–40°F)

(Integral Flowmeter) T4 +120°C (+248°F) –40°C (–40°F)


For CSA C22. 2 Series T3 +150°C (+302°F) –40°C (–40°F)
Explosion proof for Class I, Division 1, Groups A, B, C & T28-1.EPS

D. Ambient Temp.: –40°C to +60°C (–40°F to +140°F)


Dust-ignition proof for Class II/III, Division 1, Groups E,
F & G. For CSA E79 Series
Intrinsically safe (electrodes) for Class I, Division 1, Flameproof for Zone 1, Ex dme[ia] IIC T6...T3
Groups A, B, C & D. Intrinsically safe (electrodes), Ex ia IIC T6...T3
“SEAL ALL CONDUITS WITHIN 50 cm OF THE
Electrode Circuit Um: 250 Vac/dc
ENCLOSURE”
Excitation Circuit: 170V max
“WHEN INSTALLED IN DIV. 2, SEALS NOT REQUIRED”
Enclosure: IP66, IP67
Electrode Circuit Um: 250 Vac/dc Temperature Code:
Maximum power supply voltage: 250 Vac/130 Vdc
Temperature Maximum Process Minimum Process
Excitation Circuit: 140V max Code Temperature Temperature
Enclosure: Type 4X
T6 +70°C (+158°F) –40°C (–40°F)
Temperature Code:
T5 +85°C (+185°F) –40°C (–40°F)
Temperature Maximum Process Minimum Process
Code Temperature Temperature T4 +120°C (+248°F) –40°C (–40°F)

T6 +70°C (+158°F) –40°C (–40°F) T3 +150°C (+302°F) –40°C (–40°F)


T28-1.EPS
T5 +85°C (+185°F) –40°C (–40°F)
Ambient Temp.: –40°C to +60°C (–40°F to +140°F)
T4 +120°C (+248°F) –40°C (–40°F)
T3 +130°C (+266°F) –40°C (–40°F)
T27-1.EPS

Ambient Temp.: –40°C to +60°C (–40°F to +140°F)

For CSA E79 Series


Flameproof for Zone 1, Ex dme[ia] IIC T6...T3
Intrinsically safe (electrodes), Ex ia IIC T6...T3
Electrode Circuit Um: 250 Vac/dc
Maximum power supply voltage: 250 Vac/130 Vdc
Excitation Circuit: 140V max
Enclosure: IP66, IP67
Temperature Code:
Temperature Maximum Process Minimum Process
Code Temperature Temperature

T6 +70°C (+158°F) –40°C (–40°F)


T5 +85°C (+185°F) –40°C (–40°F)
T4 +120°C (+248°F) –40°C (–40°F)
T3 +130°C (+266°F) –40°C (–40°F)
T27-1.EPS

Ambient Temp.: –40°C to +60°C (–40°F to +140°F) IM 01E20D01-01E


6-11
6. OUTLINE

 Table of Optional Specifications (Size 2.5 mm (0.1 in.) to 400 mm (16 in.)) (continued)
: Available –: Not available
Applicable Model
General Explosion proof Submersible Sanitary
Integral Remote Integral Remote Remote Integral Remote
Item Specifications Flowmeter Flowtube Flowmeter Flowtube Flowtube Flowmeter Flowtube Code

G
D

N
E

AXF***G- P

AXF***C- P

AXF***W- P

AXF***H- P
AXF***G- F

AXF***C- F

AXF***H- F
Five-point Calibration A flow test at 0, 25, 50, 75, and 100% of the user-specified span is
in User-specified performed instead of the flow test of the standard 2m/s span and a test
Span certificate (QIC) is submitted. Specify the span (100% flow span) whose
corresponding flow velocity lies between 0.5 to 10 m/s and that is less than
the maximum line capacity. Selectable range of flow rate span is showing
below.

Size : mm Selectable range of flow rate span : m3/h


(in.) (Flow rate span velocity : m/s)
2.5 (0.1) 0.009 (0.5) to 0.05 (2.83)
5 (0.2) 0.036 (0.5) to 0.2 (2.83)
10 (0.4) 0.15 (0.5) to 0.96 (3.40)
15 (0.5) 0.32 (0.5) to 2.8 (4.40)
25 (1) 0.89 (0.5) to 11 (6.22)
       SC
32 (1.25) 1.45 (0.5) to 28.9 (10.00)
40 (1.5) 2.27 (0.5) to 32 (7.07)
50 (2) 3.54 (0.5) to 56 (7.92)
65 (2.5) 5.98 (0.5) to 80 (6.70)
80 (3) 9.05 (0.5) to 126 (6.96)
100 (4) 14.2 (0.5) to 220 (7.78)
125 (5) 22.1 (0.5) to 300 (6.79)
150 (6) 31.9 (0.5) to 380 (5.97)
200 (8) 56.6 (0.5) to 670 (5.92)
250 (10) 88.4 (0.5) to 1000 (5.66)
300 (12) 128 (0.5) to 1200 (4.72)
350 (14) 174 (0.5) to 1200 (3.47)
400 (16) 227 (0.5) to 1350 (2.98)

FM Approval FM Explosion proof


– –   – – – FF1
See “HAZARDOUS AREA CLASSIFICATION”
CENELEC ATEX ATEX Explosion proof
Certification See “HAZARDOUS AREA CLASSIFICATION” – –   – – – KF2
(KEMA Approval)

CSA Certification CSA Explosion proof


Dual Seal – –   – – – CF1
See “HAZARDOUS AREA CLASSIFICATION”
IECEx IECEx Explosion proof
Certification (*5) – –   – – – SF2
See “HAZARDOUS AREA CLASSIFICATION”
TIIS Certification TIIS Flameproof
– –   (*7) – – – JF3
See “HAZARDOUS AREA CLASSIFICATION”
Flameproof packing Two flameproof packing adapters
adapter for TIIS – –   – – – G12
Flameproof Type (*6)
One flameproof packing adapter and one blind plug.
Available for integral flowmeter and only when a four-wire cable is used for – –  (*8) – – – – G11
power input and signal output with a DC power supply.

*1: ⫹90-degree rotation ⫹180-degree rotation ⫺90-degree rotation


Standard
Optional Code RA Optional Code RB Optional Code RC

Electrical Indicator Electrical Indicator


Integral Flowmeter Connection Connection

Electrical Electrical
Remote Flowtube Connection Connection

*2: When specifying the optional code BCC or BSC for a PFA or ceramics lining, it is advisable to specify the optional code GA, GC, or GD at the same time
to prevent potential leakage caused by the difference in elasticity between the flowtube and chloroprene gaskets. Refer to description of “Gasket” in the
“Wetted Part Material”.
*3: Special gaskets are inserted between the flowtube and the grounding ring or grounding electrode.
*4: Enhanced dual frequency excitation is not available for models with calibration code C (High Grade Accuracy).
*5: Applicable only for Australia and New Zealand area.
*6: Select optional code G12 or G11 when TIIS Flameproof type with wiring using a flameproof packing adapter. Available only for JIS G1/2 electric connection.
T26-4.EPS
*7: The TIIS flameproof type is only available for AXF***C-P (remote flowtube for combined use with AXFA14).
*8:In the case of Fieldbus communication type, optional codes C1, C2, C3, EM and G11 are unable to select.

6-36 IM 01E20D01-01E
8. EXPLOSION PROTECTED TYPE INSTRUMENT

HAZARDOUS LOCATIONS For CSA E79 Series


Conduit Sealing Fitting
Magnetic Flowmeter Flameproof for Zone 1, Ex dme [ia] IIC T6...T3
Intrinsically safe (electrodes), Ex ia IIC T6...T3
Electrode Circuit Um: 250 Vac/dc
Maximum power supply voltage: 250 Vac/130 Vdc
18" (457 mm) Max.
F0802.EPS Excitation Circuit: 140V max
Figure 8.2.1 Conduit Wiring Enclosure: IP66, IP67
Temperature Code:
8.3 CSA Temperature
Code
Maximum Process
Temperature
Minimum Process
Temperature

(1) Technical Data T6 +70°C (+158°F) –40°C (–40°F)

*AXF002C – AXF400C T5 +85°C (+185°F) –40°C (–40°F)

Applicable Standard: T4 +120°C (+248°F) –40°C (–40°F)

For CSA C22.2 Series; T3 +130°C (+266°F) –40°C (–40°F)


T0807.EPS
C22.2 No 0, C22.2 No 0.4, C22.2 No 0.5,
Ambient Temp.: –40°C to +60°C (–40°F to +140°F)
C22.2 No 25, C22.2 No 30, C22.2 No 94,
C22.2 No 157, C22.2 No 1010.1
For CSA E79 Series; (Remote Flowtube)
CAN/CSA-E79-0, CAN/CSA-E79-1, For CSA C22.2 Series
CAN/CSA-E79-7, CAN/CSA-E79-11, Explosion proof for Class I, Division 1, Groups A,
CAN/CSA-E79-18 B, C & D.
Certificate: 1481213 Dust-ignition proof for Class II/III, Division 1,
Process Sealing Certification: Groups E, F & G.
Dual Seal certified by CSA to the requirements of Intrinsically safe (electrodes) for Class I, Division 1,
ANSI/ISA 12.27.01. Groups A, B, C & D.
No additional sealing required. “SEAL ALL CONDUITS WITHIN 50 cm OF THE
Primary seal failure annunciation; ENCLOSURE”
Deterioration of the flowrate output at nonzero “WHEN INSTALLED IN DIV. 2, SEALS NOT
flow point. REQUIRED”
Unstable flowrate output at zero flow point.
Electrode Circuit Um: 250 Vac/dc
(Integral Flowmeter) Excitation Circuit: 170V max
For CSA C22. 2 Series Enclosure: Type 4X
Explosion proof for Class I, Division 1, Groups A, Temperature Code:
B, C & D. Temperature Maximum Process Minimum Process
Code Temperature Temperature
Dust-ignition proof for Class II/III, Division 1,
Groups E, F & G. T6 +70°C (+158°F) –40°C (–40°F)

Intrinsically safe (electrodes) for Class I, Division 1, T5 +85°C (+185°F) –40°C (–40°F)

Groups A, B, C & D. T4 +120°C (+248°F) –40°C (–40°F)

“SEAL ALL CONDUITS WITHIN 50 cm OF THE T3 +150°C (+302°F) –40°C (–40°F)


T28-1.EPS
ENCLOSURE”
“WHEN INSTALLED IN DIV. 2, SEALS NOT Ambient Temp.: –40°C to +60°C (–40°F to +140°F)
REQUIRED”
For CSA E79 Series
Electrode Circuit Um: 250 Vac/dc
Maximum power supply voltage: 250 Vac/130 Vdc Flameproof for Zone 1, Ex dme [ia] IIC T6...T3
Excitation Circuit: 140V max Intrinsically safe (electrodes), Ex ia IIC T6...T3
Enclosure: Type 4X Electrode Circuit Um: 250 Vac/dc
Temperature Code: Excitation Circuit: 170V max
Temperature Maximum Process Minimum Process Enclosure: IP66, IP67
Code Temperature Temperature

T6 +70°C (+158°F) –40°C (–40°F)


T5 +85°C (+185°F) –40°C (–40°F)
T4 +120°C (+248°F) –40°C (–40°F)
T3 +130°C (+266°F) –40°C (–40°F)
T27-1.EPS

Ambient Temp.: –40°C to +60°C (–40°F to +140°F) 8-5 IM 01E20D01-01E