User's Manual: AXF Magnetic Flowmeter Integral Flowmeter/ Remote Flowtube (Hardware Edition)
User's Manual: AXF Magnetic Flowmeter Integral Flowmeter/ Remote Flowtube (Hardware Edition)
Please use the attached sheets for the pages listed below in IM 01E20D01-01E (9th).
Page 6-36
Table of optional Sepecifications Added the "Dual Seal" to descriptive text of
"CSA Certifications (CF1)".
Page 8-5
8.3 CSA Added the "Process Sealing Certification".
Apr. 1, 2008
Yokogawa Electric Corporation
6. OUTLINE
Table of Optional Specifications (Size 2.5 mm (0.1 in.) to 400 mm (16 in.)) (continued)
: Available –: Not available
Applicable Model
General Explosion proof Submersible Sanitary
Integral Remote Integral Remote Remote Integral Remote
Item Specifications Flowmeter Flowtube Flowmeter Flowtube Flowtube Flowmeter Flowtube Code
G
D
N
E
AXF***G- P
AXF***C- P
AXF***W- P
AXF***H- P
AXF***G- F
AXF***C- F
AXF***H- F
Five-point Calibration A flow test at 0, 25, 50, 75, and 100% of the user-specified span is
in User-specified performed instead of the flow test of the standard 2m/s span and a test
Span certificate (QIC) is submitted. Specify the span (100% flow span) whose
corresponding flow velocity lies between 0.5 to 10 m/s and that is less than
the maximum line capacity. Selectable range of flow rate span is showing
below.
Electrical Electrical
Remote Flowtube Connection Connection
*2: When specifying the optional code BCC or BSC for a PFA or ceramics lining, it is advisable to specify the optional code GA, GC, or GD at the same time
to prevent potential leakage caused by the difference in elasticity between the flowtube and chloroprene gaskets. Refer to description of “Gasket” in the
“Wetted Part Material”.
*3: Special gaskets are inserted between the flowtube and the grounding ring or grounding electrode.
*4: Enhanced dual frequency excitation is not available for models with calibration code C (High Grade Accuracy).
*5: Applicable only for Australia and New Zealand area.
*6: Select optional code G12 or G11 when TIIS Flameproof type with wiring using a flameproof packing adapter. Available only for JIS G1/2 electric connection.
T26-4.EPS
*7: The TIIS flameproof type is only available for AXF***C-P (remote flowtube for combined use with AXFA14).
*8:In the case of Fieldbus communication type, optional codes C1, C2, C3, EM and G11 are unable to select.
6-36 IM 01E20D01-01E
8. EXPLOSION PROTECTED TYPE INSTRUMENT
Intrinsically safe (electrodes) for Class I, Division 1, T5 +85°C (+185°F) –40°C (–40°F)