Manual Tuxon XP PDF

The document provides technical specifications and instructions for connecting and using the TUXON-XP digital indicator. 1. It has functions like numerical calibration, digital filters, automatic zero tracking, and 4 set points. 2. Connection diagrams show how to connect the power supply, load cells via 6-wire or 4-wire, and optional expansion boards. 3. The calibration, parameter setting, I/O definition and serial communication protocols are described.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
111 views97 pages

Manual Tuxon XP PDF

The document provides technical specifications and instructions for connecting and using the TUXON-XP digital indicator. 1. It has functions like numerical calibration, digital filters, automatic zero tracking, and 4 set points. 2. Connection diagrams show how to connect the power supply, load cells via 6-wire or 4-wire, and optional expansion boards. 3. The calibration, parameter setting, I/O definition and serial communication protocols are described.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 97

Model no.

: TUXON - XP

U s e r ’s M a n u a l
CONTENTS
1 General Description .................................................................. 1
1.1 Functions and Characteristics ..................................... 1
1.2 Front Panel ..................................................................... 3
1.3 Technical Specifications ................................................ 5
1.3.3 Digital: .................................................................. 6
2 Installation and Wiring .............................................................. 7
2.1 Connection of Power Supply ....................................... 7
2.2 Connection of Load Cell ............................................... 7
2.2.1 6 wires connection ............................................. 9
2.3 I/O terminals ................................................................. 10
2.4 Optional Expansion Board Output ............................ 11
2.4.1 Analog Output (Optional)................................. 12
2.4.2 Serial Interface RS485 Output ....................... 16
2.4.3 Serial Interface RS232 output (Optional)...... 17
3 Calibration ................................................................................ 18
3.1 Instruction ............................................................. 18
(1)Calibration procedure. ..................................... 18
3.2 Flow Chart of Calibration ............................................ 19
3.3 Millivolt Value Display ......................................... 24
3.4 Calibration with Weights ............................................. 25
3.5 Millivolt Calibration ...................................................... 25
3.6 Calibration Switch for Communication Interface ..... 28
3.7 Explanation for Calibration Parameters ................... 28
4 Working Parameters Setting ................................................. 30
4.1 Flow Chart of Working Parameters Setting ............. 30
4.2 Parameter Setting Method ......................................... 32
4.3 Descriptions of Operation Parameters ..................... 33
4.4 Set point parameters ................................................... 36
5 I/O Definition ............................................................................ 39
5.1 I/O Definition................................................................. 39
5.2 I/O testing ..................................................................... 42
6 Serial Communication ............................................................ 43
6.1 r-Cont............................................................................. 43
6.2 r-SP1 ............................................................................. 45
6.2.1 Parameters Code Chart .................................. 45
6.2.2 Error Code Explanation ................................... 49
6.2.3 Command .......................................................... 49
6.3 tt TOLEDO Protocol .................................................... 63
6.4 Cb920 ............................................................................ 65
6.5 rECont ........................................................................... 67
6.6 rEREAD......................................................................... 69
6.7 Modbus ......................................................................... 70
6.7.1 Modbus communication address ................... 72
7 Password Input and Setting, Reset...................................... 83
7.1 Password Input ............................................................ 83
7.2 Password Setting ......................................................... 84
7.3 Factory Reset ............................................................... 85
8 Display Testing ........................................................................ 88
9 Errors and Alarm Messages .................................................. 89
10 Dimension of Indicator ......................................................... 90
Tuxon-XP Digital Indicator

1 General Description
TUXON - XP weighing indicator is specially designed for weight transmitting in
industrial fields.

This indicator has the features of small volume, plenty communicating commands,
stable performance, easy operation and practicability. It can be widely applied to
concrete and bitumen mixing equipment, metallurgy furnace and converter, chemical
industry and feed, etc. .

1.1 Functions and Characteristics

 Small volume, unique design, easy operation

 Applicable to all kinds of resistance strain gauge bridge load cell

 Front panel numerical calibration

 Multilevel of digital filter

 Automatic zero -tracking

1
Tuxon-XP Digital Indicator

 Automatically zero when powered on

 4 set points

 1 input and 2 outputs

 Serial communication interface:RS232 or RS485

 Calibration via serial interface

 Optional interfaces: Analog output, serial interface

2
Tuxon-XP Digital Indicator

1.2 Front Panel

TUXON - XP Front panel

Main Display: 6 digits, for displaying weight and the information of parameters.

Status Indicator Lamp:

 ZERO: Light on when present weight is within 0±1/4d.

3
Tuxon-XP Digital Indicator

 STAB: Light on when changes of weight values are within the range of motion
detecting during motion detecting time.

 HOLD: Light on when indicator displays the value of D/A output.

Keypad:

: Zero/Esc, Used to exit from current operation or go previous.

: Used to scroll optional values of parameter and to make flashing digit


increase 1 while data inputting.

: Function Selecting Key, To make flashing position move to the right digit
when data inputting.

: Confirming Key. Used to confirm present operation.

4
Tuxon-XP Digital Indicator

1.3 Technical Specifications

1.3.1 Common:

Power supply: DC24V±5%

Working temperature: -10~40℃

Max humidity: 90%R.H without dew

Power consumption: About 10W

Dimension: 110×89×60(mm)

1.3.2 Analog:
Load cell power: DC5V 200mA(MAX) Input impedance: 10MΩ

Zero steady range: 0.00~12mV(Load cell 3mV/V)

Input sensitivity: 0.1uV/d

5
Tuxon-XP Digital Indicator

Input range: 0.00~15mV(Load cell 3mV/V)

Transfer mode: Sigma - Delta

A/D conversion speed: 15, 30, 60, 120, 480, 960 times/sec

Non-linearity: 0.01% F.S

Gain drift: 10PPM/℃

Display Precision: 1/100,000

1.3.3 Digital:
Weight display: 6 digits red high-brightness LED

Minus display: “-”

Overload display: “OFL”

Decimal point: 5 kinds (optional)

Function keys: 4 keys soniferous keypad

6
Tuxon-XP Digital Indicator

2 Installation and Wiring

2.1 Connection of Power Supply

TUXON - XP weighing indicator connects DC24V power supply as follows:

Power supply connection

2.2 Connection of Load Cell

TUXON - XP weighing indicator connects bridge type resistance strain gauge


load cells by 6 wires or 4 wires as follows. When you use 4-wired load cells, you

7
Tuxon-XP Digital Indicator

must bridge the SN+ with EX+ and bridge the SN- with EX-.

The signal definition of each port of the load cell connector is as follows:

EX+: Excitation+ EX-: Excitation- SN+: Sense+ SN-: Sense- SIG+: Signal+ SIG-: Signal-

6 wires EX+ SN+ EX- SN- SIG+ SIG- Shield

4 wires EX+ EX- SIG+ SIG- Shield

8
Tuxon-XP Digital Indicator

2.2.1 6 wires connection

Note:

1. As load cell output sensitive analog signal, please use shield cable to separate
with other cables, especially AC power.

2. 4 wires connection is suitable for short distance and stable temperature or low
precision field, otherwise use 6 wires connection.

3. For more load cells parallel connection, their sensitivity (mV/V) should be same.

9
Tuxon-XP Digital Indicator

2.3 I/O terminals

I/O tolerant definition as follows:

Output Input

OUT1 Stable IN1 Reset


all

OUT2 OFL

Indicator input terminal connection:

10
Tuxon-XP Digital Indicator

Indicator input terminal connection:

2.4 Optional Expansion Board Output

TUXON - XP weighing indicator supports analog output, RS232 or RS485 as

optional output function, please confirm it when place orders.

11
Tuxon-XP Digital Indicator

2.4.1 Analog Output (Optional)

At normal displaying status, press to check the analog output.

The definition of analog output as below:

The definition of analog output:

V+: voltage-output+, V-: voltage-output-

I+: current-output +, I-: current-output -

Analog output two types:

1) Voltage output: 0-5V/0-10V is optional .

2) Electric current output: 4-20mA/0-20mA/0-24mA is optional.

12
Tuxon-XP Digital Indicator

3) User-define function, users can define analog output type and output range.

The analog output has been calibrated before the delivery of the indicators, so
users do not need to make calibration. If analog output is abnormal, users can
calibrate by themselves as follows:(Suggestion: please calibrate under the
instruction of professionals)

Normal display status


Note: only support calibration under current

mode. 4 points must be finished for current

calibration.

SET UP

13
Tuxon-XP Digital Indicator

and
In this interface,the
multimeter measured
Current calibration interface should be 4.000mA.

If not,press

Measured 4.000mA

This interface is current DA If yes,press

D/A,press to adjust the value,


and the current value measured will
vary accordingly.To adjust it when the
measured value to be 4.000mA, then In this interface,the
14 multimeter measured
should be 4.000mA.
press to save and exit.
Tuxon-XP Digital Indicator

Finish the calibration referring to the process of


4.000mA calibration, and the same for the
calibration of 20mA.

This interface is the highest


point calibration of analog
output.

15
Tuxon-XP Digital Indicator

In the interface of highest point calibration,press to enter into analog


display value interface, the display will be 5 digits (initial value is 24000, means

24.000mA),press to input the value measured by the multimeter.

Note: Analog output calibration, highest point must be calibrated.

2.4.2 Serial Interface RS485 Output

Serial Interface RS485 output is optional, please refer to chapter 6.0 for
communication protocol.

RS485 serial interface connection:

16
Tuxon-XP Digital Indicator

A A

Controller 1

Computer
B B
GND GND

.
.
. GND is ground of RS485, it can very
much improve communication
quality via connecting with GND by

Controller n
A
low-resistance wire when there is a lot
B of disturbance in working field.

GND

2.4.3 Serial Interface RS232 output (Optional)

RS232 serial interface connection:

17
Tuxon-XP Digital Indicator

3 Calibration
3.1 Instruction
(1)Calibration procedure must be executed when a TUXON - XP indicator is put in use at
the first time, the preset parameters may no longer meet the user’s needs, and any part of the
weighing system was changed. Position of decimal point, minimum division, maximum
capacity, zero, and gain can be set and confirmed through calibration.

(2)If you want to set only one parameter, please press to save parameter’s value

and then press to exit.

(3)Please see section 3.7 for parameters’ instruction.

(4)Please record each value in the blank table in section 3.4 during calibration for the
emergency use in future.

(5)See chapter 9 for error alarm message that may be displayed during calibration.

18
Tuxon-XP Digital Indicator

3.2 Flow Chart of Calibration


1. Under this status, press (twice), indicator

will display CAL, then press to enter

password input.
Normal Status
2. After password is input, the indicator will display

CALON for one second, then go to next step.

3. Press to select a desired value for


decimal point among 0,

Password input 0.0, 0.00, 0.000 and 0.0000, and then

press to save it and enter next step.

If there’s no need to change the value,

Decimal point
press directly to enter next step.

19
Tuxon-XP Digital Indicator

4. Press to select a desired value for min.


division among 1,2,

5,10,20 and 50, and then to save it and enter next


step.
Min. divistion If there’s no need to change the value, then

press directly to enter next step.


5. Input max. capacity(≤min. division×100000),

press to save it and enter next step. If


Max. capacity there’s no need to change the value, then

press directly to enter next step.

6. Under this status, press to enter zero


calibration.

Display value near the output value in millivolt


Millivolt value display between SIG+/SIG- of load cell.

See section 3.3 for details about this function.


20
Tuxon-XP Digital Indicator

7. Unloaded scale first, when STAB lamp is on,

press to finish zero calibration.

If there’s no need to calibrate zero,


Zero calibration
press directly to enter gain calibration.

21
Tuxon-XP Digital Indicator

8. The process of gain calibration is as follows. If there’s no need to do gain calibration, press

directly to enter serial ports calibration switch setting.

22
Tuxon-XP Digital Indicator

9. Press to enter setting interface,

press to choose the switch position,

press to set password。If don’t need to set

switch position, then press to enter password


setting.

10. See section 7.2 for reference to set password.


If there’s no need to set password,

press directly to go back to normal


status.

23
Tuxon-XP Digital Indicator

3.3 Millivolt Value Display


This function is mainly used for system test, position-error test for weighing
mechanism and linearity test for load cell.
1. System Test
(1) If display data changes with loaded weight changes, it shows that connection
of load cell is correct and weighing mechanism works well.
(2)If display value is OFL (or –OFL), it means that loaded weight on load cells is
too large (or too small). Please unload the weight (or load more), if display value is still
OFL (or –OFL), the possible reasons are as follows:
a. There is something wrong with weighing mechanism, please check and clear.
b. The connection of load cell is incorrect, please check and clear.
c. Load cells may be damaged, please replace.
2. Position-error Test for Weighing Mechanism
Load a same weight on each corner of weighing mechanism and record
displayed millivolt value respectively. If differences among these values are obvious,
please adjust weighing mechanism.
3. Linearity Test for Load Cell
Load same weight for several times, and record displayed value every time. If
one or two values are obviously much larger or smaller than any others, it means that
the linearity of load cell is bad.

24
Tuxon-XP Digital Indicator

*NOTE: You must use to zero display data before weight is


loaded for each time.
3.4 Calibration with Weights
During calibration with weight,please record the zero millivolt value, gain millivolt
value and the loaded weight value in the blank table below. If it is not convenient to load
a weight to calibrate, these values can be used for calibration without weights.

Zero millivolt Gain millivolt Loaded Date Remarks


value(mV) value(mV) Weight
1
2
3
4
5
3.5 Millivolt Calibration
When it is not convenient to load a weight to calibrate,calibration can be done
without weights using recorded data in the table in section 3.4.
However, this method is just used for some emergencies, it will make calibration
result incorrect if load cells, or indicator has been replaced.

25
Tuxon-XP Digital Indicator

26
Tuxon-XP Digital Indicator

27
Tuxon-XP Digital Indicator

3.6 Calibration Switch for Communication Interface


When calibrate the transmitter through serial port ( Rs, SP1 or Modbus), must set to
“ON” status for the calibration switch for communication interface.
3.7 Explanation for Calibration Parameters
Symbol Parameter Types Value of parameter Default
Pt Decimal Point 5 0 0.0 0.00 0.000 0.0000 0
1d Min. Division 6 1 2 5 10 20 50 1
CP Max. Capacity ≤Min. Division×100000 10000

t Millivolt Value

o Zero

C Gain
Switch for
SIOCAL Calibration OFF
via serial interface
PASS Password Setting 000000

28
Tuxon-XP Digital Indicator

Log Table for Calibration Parameters

Parameter Calibrated Value Date Remarks

Decimal Point

Min. Division

Max. Capacity

Load cell sensitivity

Password

29
Tuxon-XP Digital Indicator

4 Working Parameters Setting

4.1 Flow Chart of Working Parameters Setting

Normal Status

If F4.1(Password Switch) is ON, you


SET UP
should input password first.

F1 F1.1 F1.7

1 .Use to select between the same level. Switch from

display F1 to F2, press .

30
Tuxon-XP Digital Indicator

F2 F2.1 F2.7

2 .Press to enter into sub-selection, press to


return back to previous menu.

When it displays F2, press to enter into F2.1.

F3 F3.1 F3.2

F4 F4.1

31
Tuxon-XP Digital Indicator

4.2 Parameter Setting Method


TUXON - XP has 2 kinds of working parameters: Selection type and data type. For selection

type parameters, press to choose. For data type parameter in parameter interface,

press to choose digit position, press to choose value.

Selection:

F1.1 OFF ON F1.1

Enter setting status Change OFF to ON Confirm the change

Data:

F1.4 50 50 60

Enter setting status Move the flashing from 0 to 5 Adjust value


Confirm change
F1.4
32
Tuxon-XP Digital Indicator

4.3 Descriptions of Operation Parameters


Code Default Description
F1 Null The first major term of working parameter.
Switch for Auto-Zeroing when power-on, OFF: disabled ON:
F1.1 OFF
enabled
F1.2 0 Zero-tracking Range(0~9d optional). This parameter is for
automatic calibration, disabled when is set ”0”.
F1.3 1 Motion Detecting Range(1~9d optional)

F1.4 50 Zeroing Range(00%~99% of Maximum capacity)

F1.5 5 Digital filtering parameter: (1-9 as optional)


0: without filtering
9: strongest digital filtering
Stable filter parameter(the second filter based on the first
F1.6 0 filter): (1-9 as optional)
0: without filtering
9: strongest digital filtering
F1.7 0 A/D conversion rate: 120,480,960,15,30,60 as optional

33
Tuxon-XP Digital Indicator

F2 Null The second major term of working parameter.


F2.1 01 Scale no., indicator no.
F2.2 9600 Baud rate of serial port: 1200/2400/4800/9600/19200/38400

F2.3 Cb920 Serial ports communication mode:


Modbus-RTU:MODBUS RTU mode;
r-Cont:SP1 continuous mode;
r-SP1: SP1 command mode;
tt:TOLEDOcontinuous mode;
Cb920:Cb920 continuous mode。
rE-Cont:rE continuous mode;
rE- rEAd:rEcommand mode;
Data format:
F2.4
7-E-1:7 data bits,even parity check,1 stop bit;
7-O-1:7data bits,odd parity check,1 stop bit;
7-E-1 8-E-1:8 data bits,even parity check,1 stop bit;
8-O-1:8 data bits,odd parity check,1 stop bit;
8-n-1:8 data bits,no parity check,1 stop bit;
8-n-2:8 data bits,no parity check,2 stop bit;

34
Tuxon-XP Digital Indicator

HiLo MODBUS dual-byte register storage turn,Hi Lo:High byte


F2.5 in the front, low byte at the back;Lo Hi:Low byte in the front,
high byte at the back
F2.6 nonE Cont mode automatic sending time interval
tt(TOLEDOcontinuous mode)If send the checksum。
F2.7 0
0:not send, 1:send.
The third major term of working parameter.
F3 Null
(For analog output only)
Analog output:
4-20:4-20mA
0-20:0-20mA
0-24:0-24mA
F3.1 0-5 0-5: 0-5V
0-10: 0-10V
I_out:Current customized
V_out:Voltage customized
In customized mode,F3.2-F3.5 parameters available
F3.2 3920 Minimum output
F3.3 4000 Zero point output

35
Tuxon-XP Digital Indicator

F3.4 20000 Maximum capacity output


F3.5 20020 Maximum output
F4 Null The fourth major term of working parameter.
F4.1 OFF Parameters password setting switch.
F4.2 000000 Parameters password setting:Valid when F4.1 is ON
4.4 Set point parameters

Code Default Description

P1-P4 Null The first term of working parameters

PX.1 OFF Change of state if need stable

Change of state minimum duration


PX.2 0.0

PX.3 P1.3=1 Condition of validity:

36
Tuxon-XP Digital Indicator

P2.3=5 0:forbid;

1:<;
P3.3=0
2:<=;

P4.3=0 3:==;

4:>=;

5:>;

6:!=;compare to minimum value

7:_<>_outside the range, need to set 2 edge value

8:=<__>=inside the range, need to set 2 edge value

9:external trigger. If it’s IO,do 1 state change for 1


trigger, if it’s command, then decide according to
valid or invalid command.

37
Tuxon-XP Digital Indicator

PX.4 0 Set value 1(Set value 1 and set value 2, choose the
minimum to compare)

PX.5 0 Set value 2

Set point has 4 major terms which are user defined.

38
Tuxon-XP Digital Indicator

5 I/O Definition

5.1 I/O Definition


Output/Input code table:

Output

Code Definition Description

O0 None No definition

O1 Stable Effective output in stable status.

O2 Overflow Effective output when overflow.

O3 Sp1 Effective output when set point 1 status output.

O4 Sp2 Effective output when set point 2 status output.

O5 Sp3 Effective output when set point 3 status output.

39
Tuxon-XP Digital Indicator

O6 Sp4 Effective output when set point 4 status output.

Input

Code Definition Description

I0 None No definition

I1 Zeroing Effective input for zeroing, pulse input signals

I2 Sp1 If this signal is valid, Sp1 status will be regarded as invalid.


Output valid state when comparision condition turns to
invalid, and be effective again.
I3 Sp2 If this signal is valid, Sp2 status will be regarded as invalid.
Output valid state when comparision condition turns to
invalid, and be effective again.
I4 Sp3 If this signal is valid, Sp3 status will be regarded as invalid.
Output valid state when comparision condition turns to
invalid, and be effective again.

40
Tuxon-XP Digital Indicator

I5 Sp4 If this signal is valid, Sp4 status will be regarded as invalid.


Output valid state when comparision condition turns to
invalid, and be effective again.
I6 Reset all Reset all paramater value when this signal is valid.

41
Tuxon-XP Digital Indicator

5.2 I/O testing


Under weighing status,press (5 times),then

display TESTio,press enter into I/O testing


interface.

Normal Status

Press OUT1 status flash, press OUT2

status flash.

This interface shows:IN1 input valid, OUT1 output valid.

42
Tuxon-XP Digital Indicator

6 Serial Communication
TUXON - XP has RS232 or RS485 as optional to realize communication with
upper computer。Support r-Cont、r-SP1、Modbus(bus)、tt TOLEDO、Cb920、rECont
protocols and rErEAD protocol.
Serial communication terminal please refer to chapter 2.4.2、2.4.3. Baud rate and
communication format setting please refer to F2.2、F2.3 and F2.4.

※Under main display(display weight value), long press to enter into serial
communication checking interface, it will display ‘------’ if no communication, and ‘------’
will flash if there’s communication.

6.1 r-Cont
Indicator will send weighing data to host computer without command.
Data Format:
STX Scale Channel Status Value CRC CR LF
no. no.

Here:

STX —— 1bit,start character 02H

43
Tuxon-XP Digital Indicator

Scale no. —— 2bits,00~99

Status —— 2bits,high byte:40H;low byte definition as follows:

D6 D5 D4 D3 D2 D1 D0

G./N. Zero
Null Null +/- OFL Stable
weight point

0: 0: 0:not
0:+
non/zero normal stable
1 0 0
1:-
1:zero 1:OFL 1:stable

Weight Value —— 6 bits;when weight is+(-)overflow,return to“space space OFL


space”

CRC —— 2 bits,check sum

CR —— 1 bit,0DH

LF —— 1 bit,0AH

44
Tuxon-XP Digital Indicator

For example:

02 30 31 31 40 4120 20 20 37 30 30 32 34 0D 0A

Means: stable,positive data,present weight 700

6.2 r-SP1
Code : ASCII

Operation code supported: W,write;R, read;C,calibrate;O, zero

6.2.1 Parameters Code Chart

Operation
Para. code Para. Name number of character
code

R WT Read current status 8


and weight

R SP Read set point status 4

45
Tuxon-XP Digital Indicator

W DC Write mini. Divistion and 8


max. capacity

R/W PT Decimal point digit 1

R DD Minimum division 2

R CP Maximum capacity 6

R/W AC Auto. Zeroing switch 1

R/W TR Zero tracking range 1

R/W MR Stable range 1

R/W ZR Zeroing range 2

R/W FL digital filtering para. 1

R/W VC steady filtering 1

R/W AD AD sample rate 1

R PO Set point output status 4

46
Tuxon-XP Digital Indicator

R/W P1M~P4M Set point to judge if 1


need stability

R/W P1T~P4T Set point minimum 3


duration time

R/W P1F~P4F Set point comparison 1


condition to judge

R/W P1L~P4L Set point lower edge 6


value to judge

R/W P1H~P4H Set point upper edge 6


value to judge

7: D6D5D4D3D2D1D0;
R AM Absolute millivolt
D6:+;D5-D0: corresponding ASCII for 6 digits
millivolt,Decimal point is fixed to 3 digits
7:D6D5D4D3D2D1D0
R RM
Relative zero point on D6 : +/-;D5-D0:corresponding ASCII for 6
millivolt digits,Decimal point is fixed to 3 digits
C ZY Zero calibration with
weight

47
Tuxon-XP Digital Indicator

C ZN Zero calibration 6
without weight

C GY Gain calibration with 6


weight

C GN Gain calibration 12
without weight

O CZ Zero clearing
command

O P1S~ Setting the Valid when 9,command valid when external


P4S corresponding set trigger and 1 status change for 1 external
point Trigger

O P1C~P4 Clear the


C corresponding set
point

R ID Read ID No. 6

48
Tuxon-XP Digital Indicator

6.2.2 Error Code Explanation

1:CRC check error

2:Operation code error

3:Parameters code error

4:Write data error

5:Operation invalid

6:Channel no. error

Note:Default channel no. of this indicator : 1(31H)

6.2.3 Command

Indicator will send weighing data to host computer after received command.

49
Tuxon-XP Digital Indicator

6.2.3.1 Host computer read present status

Send command:

STX Scale no. Channel No. R WT CRC CR LF

Correct response:

STX Scale Channel R WT Status Value CRC CR LF


no. No.

Wrong response:

STX Scale Channel R WT E Error code CRC CR LF


no. No.

Here:

STX —— 1bit,start character,02H

R——1 bit,52H

50
Tuxon-XP Digital Indicator

WT——2 bit,57H 54H

E——1 bit,45H

Status —— 2bits,high byte: 40H;low byte definition as follows:

D6 D5 D4 D3 D2 D1 D0

G./N.
Null Null +/- Zero point OFL Stable
weight

0:+ 0:non/zero 0:normal 0:not stable


1 0 0
1:- 1:zero 1:OFL 1:stable

Weight Value —— 6 bits;when weight is+(-)overflow,return to“space space OFL


space”

For example:

02 30 31 31 52 57 54 30 31 0D 0A

51
Tuxon-XP Digital Indicator

Correct response:02 30 31 31 52 57 54 40 4130 30 33 37 35 33 33 36 0D 0A


(stablepresent value 3753)

Wrong response: 02 30 31 31 52 57 54 45 31 31 39 0D 0A(CRC check error)

6.2.3.2 Read other parameters

Send command:

STX Scale no. Channel No. R Para. code CRC CR LF

Correct response:

STX Scale no. Channel No. R Para. code Value CRC CR LF

Wrong response:

STX Scale Channel R Para. E Error code CRC CR LF


no. No. code

Here:

52
Tuxon-XP Digital Indicator

Para. Value—— 1bit

Para. code——2 bits,

For example:

02 30 31 31 524D52 3839 0D 0A

Correct response:02 30 31 31 52 4D 52 36 34 33 0D 0A(stable range:6)

Wrong response:02 30 31 31 53 4D 52 45 32 30 39 0D 0A(Operation code error)

6.2.3.3 Write max. Capacity and min. Division

Send command:

STX Scale Channel W DC Division Max. CRC CR LF


no. No. value capacity

Correct response:

53
Tuxon-XP Digital Indicator

STX Scale no. Channel No. W DC O K CRC CR LF

Wrong response:

STX Scale Channel W DC E Error code CRC CR LF


no. No.

Here:

DC——2 bits,44H 43H

O——1 bit,4FH

K——1 bit,4BH

Division value——2 bits,1/2/5/10/20/50

Max. capacity——6 bits

For example:

02 30 31 3157 44 43 30 35 30 31 30 30 30 30 36 30 0D 0A(division value 5, Max


capacity 10000)

54
Tuxon-XP Digital Indicator

Correct response: 02 30 31 31 57 44 43 4F 4B 32 34 0D 0A

Wrong response: 02 30 31 31 57 44 43 45 35 39 32 0D 0A(Operation can’t execute)

6.2.3.4 Write other parameters

Send command:

STX Scale Channel W Para. Para. CRC CR LF


no. No. code value

Correct response:

STX Scale Channel W Para. O K CRC CR LF


no. No. code

Wrong response:

STX Scale Channel W Para. E Error CRC CR LF


no. No. code code

For example:

55
Tuxon-XP Digital Indicator

02 30 31 3157 5A 52 35 30 30 38 0D 0A(Write zeroing range to 50)

Correct response: 02 30 31 31 57 5A 52 4F 4B 36 31 0D 0A

Wrong response: 02 30 31 31 57 5A 53 45 33 32 38 0D 0A(Para. Code error)

6.2.3.5 Calibration Zero

1)Calibrate zero as per current weight(with weight)

Send command:

STX Scale no. Channel No. C ZY CRC CR LF

Correct response:

STX Scale no. Channel No. C ZY O K CRC CR LF

Wrong response:

STX Scale no. Channel No. C ZY E Error code CRC CR LF

56
Tuxon-XP Digital Indicator

Here:

Z——1 bit,5AH

Y——1 bit,59H

For example:

02 30 31 31 435A 59 39 34 0D 0A

Correct response:02 30 31 31 43 5A 59 4F 4B 34 38 0D 0A

Wrong response:02 30 31 34 43 5A 59 45 36 32 30 0D 0A(channel no. error)

2)Input millivolt calibration zero in the chart(without weight)

Send command:

STX Scale no. Channel No. C ZN Zero millivolt value CRC CR LF

Correct response:

57
Tuxon-XP Digital Indicator

STX Scale no. Channel No. C ZN O K CRC CR LF

Wrong response:

STX Scale no. Channel No. C ZN E Error code CRC CR LF

Here:

ZN——2 bits,5AH4EH

Zero millivolt value——6 bits

For example:

02 30 31 31 43 5A 4E 30 31 32 36 31 30 38 31 0D 0A

Correct response:02 30 31 31 43 5A 4E 4F 4B 33 37 0D 0A

Wrong response:02 30 31 31 43 5A 4E 45 34 30 34 0D 0A(Write data error)

58
Tuxon-XP Digital Indicator

6.2.3.6 Gain calibration

1)With weights

Send command:

STX Scale no. Channel No. C GY Weight value CRC CR LF

Correct response:

STX Scale no. Channel No. C GY O K CRC CR LF

Wrong response:

STX Scale no. Channel No. C GY E Error code CRC CR LF

Here:

GY——2 bits,47H 59H

Weight value——6 bits: Write in weight value

59
Tuxon-XP Digital Indicator

For example:

02 30 31 3143 47 59 30 30 30 32 30 30 36 35 0D 0A(Write in: weight value 200)

Correct response: 02 30 31 31 43 47 59 4F 4B 32 39 0D 0A

Wrong response: 02 30 31 35 43 47 59 45 36 30 32 0D 0A(Channel no. error)

2)Without weights

Send command:

STX Scale no. Channel No. C GN Gain millivolt Weight value CRC CR LF

Correct response:

STX Scale no. Channel No. C GN O K CRC CR LF

Wrong response:

STX Scale no. Channel No. C GN E Error code CRC CR LF

Here:

60
Tuxon-XP Digital Indicator

Gain millivolt——6 bits

Weight value——6bits

For example:

02 30 31 3143 47 4E 30 30 31 39 34 30 30 30 30 32 30 30 35 36 0D 0A (Write in:


weight value 200,corresponding gain millivolt 0.194)

Correct response: 02 30 31 31 43 47 4E 4F 4B 31 38 0D 0A

Wrong response: 02 30 31 31 43 48 4E 45 33 38 35 0D 0A(Para. Code error)

6.2.3.7 Zeroing

Send command:

STX Scale no. Channel No. O CZ CRC CR LF

Correct response:

61
Tuxon-XP Digital Indicator

STX Scale no. Channel No. O CZ O K CRC CR LF

Wrong response:

STX Scale no. Channel No. O CZ E Error code CRC CR LF

For example:

02 30 31 31 4F 43 5A 38 34 0D 0A

Correct response: 02 30 31 31 4F 43 5A 4F 4B 33 38 0D 0A

Wrong response: 02 30 31 31 4F 43 5A 45 35 30 36 0D 0A(Operation can’t execuate)

6.2.3.8 CRC computation


All the values in front of the parity bit add together and convert to decimal data,
then convert the last 2 bits to ASCII code (decade in front and the unit at the back).

For example

62
Tuxon-XP Digital Indicator

The following is a frame of data:

0 3 3 3 4 4 5 3 3 0 0
2 0 1 1 F 3 A 8 4 D A

Add 02~5A: 180(Hex),convert to decimal data: 384. We can calculate from this
that the check code is 38,34 for the data frame.

6.3 tt TOLEDO Protocol

When choose “tt” protocol in working parameter F2.3, indicator will send datas in
continuous mode with TOLEDO protocol.

Continuous sending mode format as below:

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18

63
Tuxon-XP Digital Indicator

STX 0D

︸ A B C display weight(6 bits) 6 个 30H Checksum

Here:start character is standard ASII start character 02(STX)

status byte A definition as below:

D0 0 1 0 1 0

D1 1 1 0 0 1

D2 0 0 1 1 0

Decimal point x .x .xx .xxx .xxxx

D3 D4 D6 为 0(not change) D5 is 1(not change)

64
Tuxon-XP Digital Indicator

status byte B definition as below:

D6 D5 D4 D3 D2 D1 D0

Status Stable Overflow symbol G.W.

Is 0 Is 1 Is 1 1-unstable 1-overflow 1-negative Is 0

(not change) (not change) (not change) 0-stable 0-normal 0-positive (not change)

status byte C is reserved, output 20H.

6.4 Cb920

When F2.3=Cb920 in working parameter, indicator will send weighing data


continuously without command under Cb920 protocol.

65
Tuxon-XP Digital Indicator

Data format:

Status , G.W. 0/1 Symbol Display Unit CR LF

Here:

Status —— 2 bits,OL:( 4FH 4CH ) OFL;ST:( 53H 54H )Stable;US:( 55H


53H )unstable

,—— 1 bit,separator 2CH

G.W. —— 2 bits, GS: gross weight 47H 53H

0/1 —— 1 bit,(30H/31H) interleaved transmission

Symbol —— 1 bit, 2BH(+), 2DH(-)

Display —— 7 bits,including decimal point

66
Tuxon-XP Digital Indicator

Unit —— 2 bits,blank space(20H 20H)

CR —— 1 bit,0DH

LF —— 1 bit,0AH

For example: When indicator send the following automatically:

53 54 2C 47 53 31 2B 20 20 31 39 30 2E 31 20 20 0D 0A

Means: Stable、G.W.、Data value is positive、current weight is 190.1

6.5 rECont

Indicator will send weighing data to the upper computer continuously without any
command.

Return data frame format specification:

67
Tuxon-XP Digital Indicator

Status , GS , +/- Display Unit CR LF

2bits 2C 47 53 2C 2B/2D 7bits 6B 67 0D 0A

Here:

Status —— 2 bits,OL(OFL):4FH 4CH; ST(stable):53H 54H; US(unstable):55H 53H

Display value —— 7bits, including decimal point, high bit is blank if no decimal point.

For example: When indicator send the following automatically:

53 54 2C 47 53 2C 2B 30 31 31 2E 31 32 30 6B 67 0D 0A

Means:Stable,Data value is positive,display value is 11.120kg

68
Tuxon-XP Digital Indicator

6.6 rEREAD

Indicator will send weighing data to the upper computer under command.

Data format:

Data R E A D CR LF

explain 52H 45H 41H 44H 0DH 0AH

The return data frame is the same with that of rECont protocol, please refer to rECont.

1) Zero clearing command: ZERO ON<CR><LF> : 5A 45 52 4F 20 4F 4E 0D 0A

Return YES<CR><LF> or NO? <CR><LF>

2) Calibration zeroing command: TARE ON<CR><LF> : 54 41 52 45 20 4F 4E 0D 0A

69
Tuxon-XP Digital Indicator

Return YES<CR><LF> or NO? <CR><LF>

3) Read ID no.: GET ID<CR><LF> : 47 45 54 20 49 44 0D 0A

Return ASCII code with 6 digits ID no.

6.7 Modbus

Indicator uses RTU mode to communicate, every 8-bit byte of the message are
divided into 2pcs of 4-bit hexadecimal characters to transmit at binary code.

Code: Binary

Function code:

Function code Definition Description

70
Tuxon-XP Digital Indicator

03 read the register

06 preset single register

16 preset several registers Command only support preset double registers.

01 read coil
The length unit is bit.
05 write coil

Exception code response

Code Definition Description

02 Illegal data address Data address received from error code is not allowed

03 Illegal data value Data wrote in is not in permissible range

04 When indicator is trying to execuate operation required,


machine fault
unrecoverable error is produced.

71
Tuxon-XP Digital Indicator

07 Unsuccessful Command received can’t be execuated under current


programming condition.
request

6.7.1 Modbus communication address

PLC addr. Display Description


addr.

The following items are only-read register(code 0x03)

40001 0000
Present weight value(4bits including sign characters, the
high bit is in the front)
40002 0001

40003 0002 D15—D14..D5—D4 — D3—— D2 ——D1——D0

All is 0 0:+ 0:non-zero 0:normal 0:unstable

1:- 1: zero 1:OFL 1:stable

72
Tuxon-XP Digital Indicator

40004 0003

…… ……
Reserve(permit to read, reading value is 0 )
40006 0005

The following items are two bytes and are available to read and write

(write code 0x06,read code 0x03)

40007 0006 Zeroing(zeroing when write in non-zero value)

40008 0007 Automatically zeroing when power on(0:OFF;1:ON)

40009 0008 Zero tracking range(0-9d)

40010 0009 Stable range(1-9d)

40011 0010 Zeroing range(0%-99%)

73
Tuxon-XP Digital Indicator

40012 0011 Digit filter parameters(0-9)

40013 0012 Stability filter series(0-9)

40014 0013 AD sample rate:0:15/s 1:30/s 2:60/s 3:120 /s


4:480/s 5:960/s

40015~ 0014~
Reserved
40018 0017

40019 0018 Decimal point place(0:0bit,1:1bit,2:2bits,3:3bits,4:4bits)

40020 0019 Minimum division(1/ 2/ 5/ 10/ 20/ 50)

The following items are available to read and write

(writing code 0x10,read code 0x03)

40021 0020 Maximum capacity(max.capacity≤mini.division×100000)

74
Tuxon-XP Digital Indicator

40022 0021

40023 0022 Zero calibration with weights: write in 1 and calibrate zero
with the current weight.
40024 0023
Read: Absolute millivolt of current load cell

40025 0024 Zero calibration without weights: Write millivolt value at


zero;Write in range(load cell 3mV/V:millivolt value range
40026 0025 within 0.02-12.000mV) millivolt value at zero when read.

40027 0026 Gain calibration with weights. Write weight value(≤max.


capacity). Return millvolt at present weight value when
40028 0027 read

40029 0028 Gain calibration without weights;input gain millivolt(load


cell 3mV/V:0.000 < millivolt<15.000mV - zero
40030 0029 millivolt).Read: millivolt value for gain calibration.

40031 0030
Gain calibration weight without weights. input gain

75
Tuxon-XP Digital Indicator

millivolt(≤max. capacity) Read: weight value for gain


40032 0031
calibration.

40033~40040 0032~0039 reserved

40041 0040 Set point 1 stable or not(0:no;1:yes)

40042 0041 Set point 1 min. duration time(0-999 : 0-99.9sec.)

40043 0042 Set point 1 valid condition

40044~40045 0043~0044 Set point 1 set value 1

40046~40047 0045~0046 Set point 1 set value2

40048 0047 Set point 2 stable or not(0:no;1:yes)

40049 0048 Set point 2 min. duration time(0-999 : 0-99.9sec.)

40050 0049 Set point 2 valid condition

76
Tuxon-XP Digital Indicator

40051~40052 0050~0051 Set point 2 set value1

40053~40054 0052~0053 Set point 2 set value2

40055 0054 Set point 3 stable or not(0:no;1:yes)

40056 0055 Set point 3 min. duration time(0-999 : 0-99.9sec.)

40057 0056 Set point 3 valid condition

40058~40059 0057~0058 Set point 3 set value 1

40060~40061 0059~0060 Set point 3 set value 2

40062 0061 Set point 4 stable or not(0:no;1:yes)

40063 0062 Set point 4 min. duration time(0-999 : 0-99.9sec.)

40064 0063 Set point 4 valid condition

77
Tuxon-XP Digital Indicator

40065~40066 0064~0065 Set point 4 set value 1

40067~40068 0066~0067 Set point 4 set value 2

40069 0068 Output 1 user-defined

40070 0069 Output 2 user-defined

40071 0070 Input 1 user-defined

40072 0071 I/O output value Note: available only when coil address
00016 is valid.
I/O input value
Input write 1 valid,0 invalid.
40073 0072
Read 1 valid,0 invalid

Reserved

40401 0400
Current weight value( 4 bytes with symbolic number,

78
Tuxon-XP Digital Indicator

40402 0401 high digit in front)

40403 0402 D15—D14……D4 0

D3 0:+ 1:-

D2 0:non-zero 1:zero

D1 0:normal 1:OFL

D0 0:Stable 1:Unstable

40404 0403 0

40405 0404
6 digits without symbolic number, user ID no.
40406 0405

The following items are bit read only. (read code: 0x03)

49001 9000 Version no. If display10024,formatXX XXXX,main

79
Tuxon-XP Digital Indicator

version no., hardware no., software


49002 9001 no..So main version no.01,hardware
no. 00,software no. 24

49003 9002
If display 141024, means 24th Oct.,
Develop time
49004 9003 2014

The following items are bit read only. (read code:0 x 0 1)

00001 0000 0:unstable; 1:stable

00002 0001 0:normal;1:OFL

00003 0002 0:non-zero;1:zero

00004 0003 0:+;1:-

00005 0004 Reserved

80
Tuxon-XP Digital Indicator

00006 0005 Reserved

The following item are available to read and write


(read code:0x01,writing code: 0x05)

00007 0006 Automatically zeroing when power on(0:OFF; 1:ON)

00008 0007 Reserved

00009 0008 Reserved

00010 0009 Reset all

00011 0010 Reset calibration

00012 0011 Reset paramaters

00013 0012 Reset I/O

00014 0013 Reserved

81
Tuxon-XP Digital Indicator

00015 0014 Reserved

00016 0015 I/O testing switch

00017 0016 Set point 1 status

00018 0017 Set point 2 status Only read:

00019 0018 Set point 3 status 0:invalid,1:valid

00020 0019 Set point 4 status

00021~ 0020~
Reserved
00032 0031

82
Tuxon-XP Digital Indicator

7 Password Input and Setting, Reset

7.1 Password Input


(1)Indicator calibration and working paraters setting default password: 000000.

(2)User can set password in parameters when F4.1 is“ON”.

(3)When display is “PASS”, need to input correct password to enter parameters.

Note:

(1)When input password, if first time wrong, it will go to the second chance for
password input(display from turn to ).

(2)If second input wrong, it will enter into interface for inputting password the third
time

(Display change from to ).

83
Tuxon-XP Digital Indicator

(3)If Input wrong for three times, main display show “Error4” and self-lock, but user

can operate when power on again.

7.2 Password Setting

(1)User can set password in parameters when F4.1 is“ON”.

(2)User must input same new password twice in setting password,If not same, main

display show“Error”one second and return to PASS again.

84
Tuxon-XP Digital Indicator

7.3 Factory Reset


Note: Factory reset is only for special technicists, which will reset all of
parameters and will maybe cause not working.

85
Tuxon-XP Digital Indicator

Take reset all for example:


In working status,

press 1 time to display


Normal working status SET UP SET UP.

Press and the


PASS same time to display‘PASS’,
then password input.

Password input interface, input


880202 0---- fixed password 880202, then
-- press save password
and enter parameter reset.
86
Tuxon-XP Digital Indicator

1)In F SEt interface, press to reset working


parameters, enter into reset calibration paramters
interface.
F SEt
2)In F SEt interface, press ,not to make working
parameter reset, enter into calibration para. Reset
interface.

F 1)In F CAL interface, press to reset calibration


para., enter into reset all interface.
CAL
2 ) In F CAL interface, press , not to make
calibration para. reset, enter into reset all interface.

1)In F 87
ALL interface, press to make reset all of
F the paramters(including working para., calibration
Tuxon-XP Digital Indicator

8 Display Testing
The following flow chart is to test lights on main-display and status lights.

Normal weighing status

Under parameter interface,

press to switch
SEt UP Press 5 times tESt to“tESt”interface.

Under“tESt”interface,press
8.8.8.8 to make display testing,
.8.8. Nixie tube display and indicator
lights all shine. Now if

press , display will


sparkle.
tESt
After testing is finished,

press to return to “tESt”


88 interface.
Tuxon-XP Digital Indicator

9 Errors and Alarm Messages


Error ①Input error.

②wrong data beyond parameter range.

Error 2 The present weight value is out of zeroing range.

Error 3 Scale platform is not stable when zeroing.

Error 4 Input wrong password more than 3 times.

OFL Weighing value is positive overflow.

-OFL Weighing value is negative overflow.

89
Tuxon-XP Digital Indicator

10 Dimension of Indicator

90
Tuxon-XP Digital Indicator

Dimension of rear panel:92×45(mm)

Dimension of front panel:105×57(mm)

Panel cutout dimension:93×46(mm)

91