Manual Tuxon XP PDF
Manual Tuxon XP PDF
: TUXON - XP
U s e r ’s M a n u a l
CONTENTS
1 General Description .................................................................. 1
1.1 Functions and Characteristics ..................................... 1
1.2 Front Panel ..................................................................... 3
1.3 Technical Specifications ................................................ 5
1.3.3 Digital: .................................................................. 6
2 Installation and Wiring .............................................................. 7
2.1 Connection of Power Supply ....................................... 7
2.2 Connection of Load Cell ............................................... 7
2.2.1 6 wires connection ............................................. 9
2.3 I/O terminals ................................................................. 10
2.4 Optional Expansion Board Output ............................ 11
2.4.1 Analog Output (Optional)................................. 12
2.4.2 Serial Interface RS485 Output ....................... 16
2.4.3 Serial Interface RS232 output (Optional)...... 17
3 Calibration ................................................................................ 18
3.1 Instruction ............................................................. 18
(1)Calibration procedure. ..................................... 18
3.2 Flow Chart of Calibration ............................................ 19
3.3 Millivolt Value Display ......................................... 24
3.4 Calibration with Weights ............................................. 25
3.5 Millivolt Calibration ...................................................... 25
3.6 Calibration Switch for Communication Interface ..... 28
3.7 Explanation for Calibration Parameters ................... 28
4 Working Parameters Setting ................................................. 30
4.1 Flow Chart of Working Parameters Setting ............. 30
4.2 Parameter Setting Method ......................................... 32
4.3 Descriptions of Operation Parameters ..................... 33
4.4 Set point parameters ................................................... 36
5 I/O Definition ............................................................................ 39
5.1 I/O Definition................................................................. 39
5.2 I/O testing ..................................................................... 42
6 Serial Communication ............................................................ 43
6.1 r-Cont............................................................................. 43
6.2 r-SP1 ............................................................................. 45
6.2.1 Parameters Code Chart .................................. 45
6.2.2 Error Code Explanation ................................... 49
6.2.3 Command .......................................................... 49
6.3 tt TOLEDO Protocol .................................................... 63
6.4 Cb920 ............................................................................ 65
6.5 rECont ........................................................................... 67
6.6 rEREAD......................................................................... 69
6.7 Modbus ......................................................................... 70
6.7.1 Modbus communication address ................... 72
7 Password Input and Setting, Reset...................................... 83
7.1 Password Input ............................................................ 83
7.2 Password Setting ......................................................... 84
7.3 Factory Reset ............................................................... 85
8 Display Testing ........................................................................ 88
9 Errors and Alarm Messages .................................................. 89
10 Dimension of Indicator ......................................................... 90
Tuxon-XP Digital Indicator
1 General Description
TUXON - XP weighing indicator is specially designed for weight transmitting in
industrial fields.
This indicator has the features of small volume, plenty communicating commands,
stable performance, easy operation and practicability. It can be widely applied to
concrete and bitumen mixing equipment, metallurgy furnace and converter, chemical
industry and feed, etc. .
1
Tuxon-XP Digital Indicator
4 set points
2
Tuxon-XP Digital Indicator
Main Display: 6 digits, for displaying weight and the information of parameters.
3
Tuxon-XP Digital Indicator
STAB: Light on when changes of weight values are within the range of motion
detecting during motion detecting time.
Keypad:
: Function Selecting Key, To make flashing position move to the right digit
when data inputting.
4
Tuxon-XP Digital Indicator
1.3.1 Common:
Dimension: 110×89×60(mm)
1.3.2 Analog:
Load cell power: DC5V 200mA(MAX) Input impedance: 10MΩ
5
Tuxon-XP Digital Indicator
A/D conversion speed: 15, 30, 60, 120, 480, 960 times/sec
1.3.3 Digital:
Weight display: 6 digits red high-brightness LED
6
Tuxon-XP Digital Indicator
7
Tuxon-XP Digital Indicator
must bridge the SN+ with EX+ and bridge the SN- with EX-.
The signal definition of each port of the load cell connector is as follows:
EX+: Excitation+ EX-: Excitation- SN+: Sense+ SN-: Sense- SIG+: Signal+ SIG-: Signal-
8
Tuxon-XP Digital Indicator
Note:
1. As load cell output sensitive analog signal, please use shield cable to separate
with other cables, especially AC power.
2. 4 wires connection is suitable for short distance and stable temperature or low
precision field, otherwise use 6 wires connection.
3. For more load cells parallel connection, their sensitivity (mV/V) should be same.
9
Tuxon-XP Digital Indicator
Output Input
OUT2 OFL
10
Tuxon-XP Digital Indicator
11
Tuxon-XP Digital Indicator
12
Tuxon-XP Digital Indicator
3) User-define function, users can define analog output type and output range.
The analog output has been calibrated before the delivery of the indicators, so
users do not need to make calibration. If analog output is abnormal, users can
calibrate by themselves as follows:(Suggestion: please calibrate under the
instruction of professionals)
calibration.
SET UP
13
Tuxon-XP Digital Indicator
and
In this interface,the
multimeter measured
Current calibration interface should be 4.000mA.
If not,press
Measured 4.000mA
15
Tuxon-XP Digital Indicator
Serial Interface RS485 output is optional, please refer to chapter 6.0 for
communication protocol.
16
Tuxon-XP Digital Indicator
A A
Controller 1
Computer
B B
GND GND
.
.
. GND is ground of RS485, it can very
much improve communication
quality via connecting with GND by
Controller n
A
low-resistance wire when there is a lot
B of disturbance in working field.
GND
17
Tuxon-XP Digital Indicator
3 Calibration
3.1 Instruction
(1)Calibration procedure must be executed when a TUXON - XP indicator is put in use at
the first time, the preset parameters may no longer meet the user’s needs, and any part of the
weighing system was changed. Position of decimal point, minimum division, maximum
capacity, zero, and gain can be set and confirmed through calibration.
(2)If you want to set only one parameter, please press to save parameter’s value
(4)Please record each value in the blank table in section 3.4 during calibration for the
emergency use in future.
(5)See chapter 9 for error alarm message that may be displayed during calibration.
18
Tuxon-XP Digital Indicator
password input.
Normal Status
2. After password is input, the indicator will display
Decimal point
press directly to enter next step.
19
Tuxon-XP Digital Indicator
21
Tuxon-XP Digital Indicator
8. The process of gain calibration is as follows. If there’s no need to do gain calibration, press
22
Tuxon-XP Digital Indicator
23
Tuxon-XP Digital Indicator
24
Tuxon-XP Digital Indicator
25
Tuxon-XP Digital Indicator
26
Tuxon-XP Digital Indicator
27
Tuxon-XP Digital Indicator
t Millivolt Value
o Zero
C Gain
Switch for
SIOCAL Calibration OFF
via serial interface
PASS Password Setting 000000
28
Tuxon-XP Digital Indicator
Decimal Point
Min. Division
Max. Capacity
Password
29
Tuxon-XP Digital Indicator
Normal Status
F1 F1.1 F1.7
30
Tuxon-XP Digital Indicator
F2 F2.1 F2.7
F3 F3.1 F3.2
F4 F4.1
31
Tuxon-XP Digital Indicator
type parameters, press to choose. For data type parameter in parameter interface,
Selection:
Data:
F1.4 50 50 60
33
Tuxon-XP Digital Indicator
34
Tuxon-XP Digital Indicator
35
Tuxon-XP Digital Indicator
36
Tuxon-XP Digital Indicator
P2.3=5 0:forbid;
1:<;
P3.3=0
2:<=;
P4.3=0 3:==;
4:>=;
5:>;
37
Tuxon-XP Digital Indicator
PX.4 0 Set value 1(Set value 1 and set value 2, choose the
minimum to compare)
38
Tuxon-XP Digital Indicator
5 I/O Definition
Output
O0 None No definition
39
Tuxon-XP Digital Indicator
Input
I0 None No definition
40
Tuxon-XP Digital Indicator
41
Tuxon-XP Digital Indicator
Normal Status
status flash.
42
Tuxon-XP Digital Indicator
6 Serial Communication
TUXON - XP has RS232 or RS485 as optional to realize communication with
upper computer。Support r-Cont、r-SP1、Modbus(bus)、tt TOLEDO、Cb920、rECont
protocols and rErEAD protocol.
Serial communication terminal please refer to chapter 2.4.2、2.4.3. Baud rate and
communication format setting please refer to F2.2、F2.3 and F2.4.
※Under main display(display weight value), long press to enter into serial
communication checking interface, it will display ‘------’ if no communication, and ‘------’
will flash if there’s communication.
6.1 r-Cont
Indicator will send weighing data to host computer without command.
Data Format:
STX Scale Channel Status Value CRC CR LF
no. no.
Here:
43
Tuxon-XP Digital Indicator
D6 D5 D4 D3 D2 D1 D0
G./N. Zero
Null Null +/- OFL Stable
weight point
0: 0: 0:not
0:+
non/zero normal stable
1 0 0
1:-
1:zero 1:OFL 1:stable
CR —— 1 bit,0DH
LF —— 1 bit,0AH
44
Tuxon-XP Digital Indicator
For example:
02 30 31 31 40 4120 20 20 37 30 30 32 34 0D 0A
6.2 r-SP1
Code : ASCII
Operation
Para. code Para. Name number of character
code
45
Tuxon-XP Digital Indicator
R DD Minimum division 2
R CP Maximum capacity 6
46
Tuxon-XP Digital Indicator
7: D6D5D4D3D2D1D0;
R AM Absolute millivolt
D6:+;D5-D0: corresponding ASCII for 6 digits
millivolt,Decimal point is fixed to 3 digits
7:D6D5D4D3D2D1D0
R RM
Relative zero point on D6 : +/-;D5-D0:corresponding ASCII for 6
millivolt digits,Decimal point is fixed to 3 digits
C ZY Zero calibration with
weight
47
Tuxon-XP Digital Indicator
C ZN Zero calibration 6
without weight
C GN Gain calibration 12
without weight
O CZ Zero clearing
command
R ID Read ID No. 6
48
Tuxon-XP Digital Indicator
5:Operation invalid
6.2.3 Command
Indicator will send weighing data to host computer after received command.
49
Tuxon-XP Digital Indicator
Send command:
Correct response:
Wrong response:
Here:
R——1 bit,52H
50
Tuxon-XP Digital Indicator
E——1 bit,45H
D6 D5 D4 D3 D2 D1 D0
G./N.
Null Null +/- Zero point OFL Stable
weight
For example:
02 30 31 31 52 57 54 30 31 0D 0A
51
Tuxon-XP Digital Indicator
Send command:
Correct response:
Wrong response:
Here:
52
Tuxon-XP Digital Indicator
For example:
02 30 31 31 524D52 3839 0D 0A
Send command:
Correct response:
53
Tuxon-XP Digital Indicator
Wrong response:
Here:
O——1 bit,4FH
K——1 bit,4BH
For example:
54
Tuxon-XP Digital Indicator
Correct response: 02 30 31 31 57 44 43 4F 4B 32 34 0D 0A
Send command:
Correct response:
Wrong response:
For example:
55
Tuxon-XP Digital Indicator
Correct response: 02 30 31 31 57 5A 52 4F 4B 36 31 0D 0A
Send command:
Correct response:
Wrong response:
56
Tuxon-XP Digital Indicator
Here:
Z——1 bit,5AH
Y——1 bit,59H
For example:
02 30 31 31 435A 59 39 34 0D 0A
Correct response:02 30 31 31 43 5A 59 4F 4B 34 38 0D 0A
Send command:
Correct response:
57
Tuxon-XP Digital Indicator
Wrong response:
Here:
ZN——2 bits,5AH4EH
For example:
02 30 31 31 43 5A 4E 30 31 32 36 31 30 38 31 0D 0A
Correct response:02 30 31 31 43 5A 4E 4F 4B 33 37 0D 0A
58
Tuxon-XP Digital Indicator
1)With weights
Send command:
Correct response:
Wrong response:
Here:
59
Tuxon-XP Digital Indicator
For example:
Correct response: 02 30 31 31 43 47 59 4F 4B 32 39 0D 0A
2)Without weights
Send command:
STX Scale no. Channel No. C GN Gain millivolt Weight value CRC CR LF
Correct response:
Wrong response:
Here:
60
Tuxon-XP Digital Indicator
Weight value——6bits
For example:
Correct response: 02 30 31 31 43 47 4E 4F 4B 31 38 0D 0A
6.2.3.7 Zeroing
Send command:
Correct response:
61
Tuxon-XP Digital Indicator
Wrong response:
For example:
02 30 31 31 4F 43 5A 38 34 0D 0A
Correct response: 02 30 31 31 4F 43 5A 4F 4B 33 38 0D 0A
For example
62
Tuxon-XP Digital Indicator
0 3 3 3 4 4 5 3 3 0 0
2 0 1 1 F 3 A 8 4 D A
Add 02~5A: 180(Hex),convert to decimal data: 384. We can calculate from this
that the check code is 38,34 for the data frame.
When choose “tt” protocol in working parameter F2.3, indicator will send datas in
continuous mode with TOLEDO protocol.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
63
Tuxon-XP Digital Indicator
STX 0D
D0 0 1 0 1 0
D1 1 1 0 0 1
D2 0 0 1 1 0
64
Tuxon-XP Digital Indicator
D6 D5 D4 D3 D2 D1 D0
(not change) (not change) (not change) 0-stable 0-normal 0-positive (not change)
6.4 Cb920
65
Tuxon-XP Digital Indicator
Data format:
Here:
66
Tuxon-XP Digital Indicator
CR —— 1 bit,0DH
LF —— 1 bit,0AH
53 54 2C 47 53 31 2B 20 20 31 39 30 2E 31 20 20 0D 0A
6.5 rECont
Indicator will send weighing data to the upper computer continuously without any
command.
67
Tuxon-XP Digital Indicator
Here:
Display value —— 7bits, including decimal point, high bit is blank if no decimal point.
53 54 2C 47 53 2C 2B 30 31 31 2E 31 32 30 6B 67 0D 0A
68
Tuxon-XP Digital Indicator
6.6 rEREAD
Indicator will send weighing data to the upper computer under command.
Data format:
Data R E A D CR LF
The return data frame is the same with that of rECont protocol, please refer to rECont.
69
Tuxon-XP Digital Indicator
6.7 Modbus
Indicator uses RTU mode to communicate, every 8-bit byte of the message are
divided into 2pcs of 4-bit hexadecimal characters to transmit at binary code.
Code: Binary
Function code:
70
Tuxon-XP Digital Indicator
01 read coil
The length unit is bit.
05 write coil
02 Illegal data address Data address received from error code is not allowed
71
Tuxon-XP Digital Indicator
40001 0000
Present weight value(4bits including sign characters, the
high bit is in the front)
40002 0001
72
Tuxon-XP Digital Indicator
40004 0003
…… ……
Reserve(permit to read, reading value is 0 )
40006 0005
The following items are two bytes and are available to read and write
73
Tuxon-XP Digital Indicator
40015~ 0014~
Reserved
40018 0017
74
Tuxon-XP Digital Indicator
40022 0021
40023 0022 Zero calibration with weights: write in 1 and calibrate zero
with the current weight.
40024 0023
Read: Absolute millivolt of current load cell
40031 0030
Gain calibration weight without weights. input gain
75
Tuxon-XP Digital Indicator
76
Tuxon-XP Digital Indicator
77
Tuxon-XP Digital Indicator
40072 0071 I/O output value Note: available only when coil address
00016 is valid.
I/O input value
Input write 1 valid,0 invalid.
40073 0072
Read 1 valid,0 invalid
Reserved
40401 0400
Current weight value( 4 bytes with symbolic number,
78
Tuxon-XP Digital Indicator
D3 0:+ 1:-
D2 0:non-zero 1:zero
D1 0:normal 1:OFL
D0 0:Stable 1:Unstable
40404 0403 0
40405 0404
6 digits without symbolic number, user ID no.
40406 0405
The following items are bit read only. (read code: 0x03)
79
Tuxon-XP Digital Indicator
49003 9002
If display 141024, means 24th Oct.,
Develop time
49004 9003 2014
80
Tuxon-XP Digital Indicator
81
Tuxon-XP Digital Indicator
00021~ 0020~
Reserved
00032 0031
82
Tuxon-XP Digital Indicator
Note:
(1)When input password, if first time wrong, it will go to the second chance for
password input(display from turn to ).
(2)If second input wrong, it will enter into interface for inputting password the third
time
83
Tuxon-XP Digital Indicator
(3)If Input wrong for three times, main display show “Error4” and self-lock, but user
(2)User must input same new password twice in setting password,If not same, main
84
Tuxon-XP Digital Indicator
85
Tuxon-XP Digital Indicator
1)In F 87
ALL interface, press to make reset all of
F the paramters(including working para., calibration
Tuxon-XP Digital Indicator
8 Display Testing
The following flow chart is to test lights on main-display and status lights.
press to switch
SEt UP Press 5 times tESt to“tESt”interface.
Under“tESt”interface,press
8.8.8.8 to make display testing,
.8.8. Nixie tube display and indicator
lights all shine. Now if
89
Tuxon-XP Digital Indicator
10 Dimension of Indicator
90
Tuxon-XP Digital Indicator
91