Phosphorus Determination Using The Colorimetric Ascorbic Acid Technique
Phosphorus Determination Using The Colorimetric Ascorbic Acid Technique
Spectrophotometer Limitations
The diode array spectrophotometer has 316 diodes that cover the wavelength range
of 190 nm to 820 nm. Each diode generates a voltage output that is proportional to the
number of incident photons. The voltage is then digitized, but the manufacturer of the
instrument in the Cornell Environmental Laboratory, Hewlett-Packard, doesn't report
the resolution of the analog to digital converter. At very low concentrations the
difference between the intensity of light transmitted through the reference and the
intensity of light transmitted through the sample approaches zero. At some low
concentration the difference in light intensity approaches the resolution of the analog
to digital converter. Another source of instrument error is drift in lamp intensity over
time. The lamp intensity is measured when a reference sample is made. The light
intensity recorded by the diodes will vary proportionally to any lamp intensity drift.
The IDL should decrease as the number of diodes used in the analysis increases (as
in Spectral analysis) for the same reason that replicate analysis of samples decreases
the standard deviation. The "Spectral analysis" feature, which can be used to measure
either single or multiple components, uses as much of the spectrum as the user desires
and thus potentially decreases the IDL. Spectral analysis uses general least squares
regression to add multiples of extinction coefficient arrays for each component to
obtain the best curve fit for the sample. The extinction coefficient arrays are obtained
from the slope of the linear regression line for absorbance as a function of
concentration at each wavelength.
Experimental Objectives
1) Measure the concentration of phosphorus in several samples to test the precision
of the ascorbic acid technique.
2) Compare the results obtained using conventional analysis at a single wavelength
with spectral analysis.
3) Analyze the data using spectrophotometer software outside the lab.
4) Analyze multiple samples so that confidence intervals can be calculated.
5) Estimate the method detection limit (MDL).
6) Discuss methods to improve the method detection limit.
7) Discuss method automation.
Experimental Procedures
Standards Preparation Method
1) Use 100 µg P/L stock.
2) Use a digital pipet and prepare 1 mL of each standard.
3) Use E-pure water to dilute the 100 µg P/L stock.
4) Prepare 5 samples such as Cayuga Lake water, tap water, Chem lawn runoff, local
creeks…
Spectrophotometer Method
1) Use Sample Cuvettes. (Make sure to orient all cuvettes with the arrow on the left
because the cuvettes are not symmetrical and have different absorbance when
turned 180°.)
2) Use the reagent blank as the reference sample for all samples.
3) Use units of µg P/L.
4) Label all samples with descriptions that your classmates will understand!
5) Fill in the general description with your initials and a description of the type of
samples
Samples Analysis
1) Measure the reference using a reagent blank. (The reagent blank is also the 0
mg/L standard.)
2) Analyze the reagent blank as a sample and verify that the absorbance deviates less
than 0.004 AU (absorbance units) from zero. If the absorbance deviates more than
0.004 AU reanalyze the reference sample.
3) Analyze 6 standards as standards using the spectrophotometer and save the data as
\\Enviro\enviro\Courses\453\phosphorus\netid_Pstd.
4) Analyze 6 standards as samples using the spectrophotometer and save the data as
\\Enviro\enviro\Courses\453\phosphorus\netid_Pstdsam.
5) Analyze 7 10-µg P/L standards as samples and save the data as
\\Enviro\enviro\Courses\453\phosphorus\netid_10Pstdsam.
6) Analyze 5 samples as samples using the spectrophotometer and save the data as
\\Enviro\enviro\Courses\453\phosphorus\netid_Psam.
Prelab Questions
1) You will be creating 1 mL standards by diluting a stock of 100 µg P/L. Create a
table showing how you will prepare 1 mL of each of the standards.
2) All of the samples are diluted with a small amount of combined reagent. How is
this dilution accounted for when calculating the concentration of samples?
Data Analysis
1) Plot the absorbance spectra of the standards. What is happening in the UV region?
Are there any absorbance peaks?
2) Choose an appropriate wavelength (perhaps an absorbance peak) and use Excel to
create a calibration curve. For the calibration curve, absorbance should be a
function of phosphorus concentration.
3) Use the 10-µg/L standards that were analyzed as samples to evaluate the Method
Detection Limit using single wavelength analysis. Use your Spreadsheet to
calculate the concentration of each of the replicates.
4) Use the 10-µg/L standards that were analyzed as samples to evaluate the Method
Detection Limit using spectral analysis. Report the wavelength range used for
your analysis. You will need to analyze each sample and copy the results into a
spreadsheet. (Note that within a data file you change which sample the software is
analyzing with the sample number control.)
5) Use the method with the lowest MDL to analyze your samples. Create a bar graph
showing the concentration of each of the samples. Report the MDL in the figure
caption.
Spreadsheet requirements
Your spreadsheet must contain all of the analysis requested above as well as the
following capabilities:
1) A well-marked cell containing the analytical wavelength for single wavelength
analysis. Changes to this cell must be reflected in all calculations and graphs.
2) The graph showing the absorbance spectra of the standards must have a vertical
line indicating the analytical wavelength.
3) All of the graphs must be on the same page as the analytical wavelength control
so the effect of changing the wavelength can be easily observed.
4) A separate sheet where you answer the 4 questions.
Hints
If you haven't already learned how to use Vlookup() now is the time!
The row() function returns the number of the row. I found it useful for this analysis!
The slope() and intercept() functions eliminate the need to type equations off of
graphs!
Questions
1) Which analysis technique gave the best results? Explain why. If the analytical
technique didn't significantly affect the MDL, explain why not.
2) What types of errors dominated your ability to measure phosphorus?
3) What method modifications do you propose to improve phosphorus
measurements?
4) Total phosphorus concentration in Cayuga Lake varies between 10 and 50 ppb.
Would the techniques used in this lab be able to measure these phosphorus levels?
References
http://wwwrvares.er.usgs.gov/nawqa/circ-1136/h6.html#PHOS
http://www.epa.gov/glnpo/lmmb/methods/index.html#Volume 3
Standard Methods for the Examination of Water and Wastewater.
Reagents
Table 1. Reagents
A Sulfuric acid solution, 4.9 N: Add
136 mL concentrated H2 SO4 to Description Supplier Catalog
800 mL E-pure water. Cool and number
dilute to 1 L with E-pure water. concentrated Fisher Scientific
H2 SO4
B Ammonium molybdate solution: (NH4 )6 Fisher Scientific
Dissolve 40 g of (NH4 )6 Mo 7 O24 •4H2 O
Mo7 O24 •4H2 O in 900 mL E-pure C6 H8 O6 Fisher Scientific
water and dilute to 1 L. Store at K(SbO)C4 H4 O6 • Fisher Scientific
½H2 O
4°C. sodium lauryl
C Ascorbic acid: Dissolve 9 g of sulfate
ascorbic acid (C 6 H8 O6 ) in 400 mL KH2 PO4 Fisher Scientific
E-pure water and dilute to 500
mL. Store at 4°C. Keep well
stoppered. Prepare fresh monthly Table 2. Equipment list
or as needed.
Description Supplier Catalog #
D Antimony potassium tartrate:
100-1095 µL Fisher Scientific 13-707-5
Dissolve 3.0 g of pipette
K(SbO)C 4 H4 O6 •½H2 O in 800 mL 10-109.5 µL pipette Fisher Scientific 13-707-3
E-pure water and dilute to 1 L. Disposable cuvets Fisher Scientific 14-385-942
Store at 4°C. Cuvet holder Fisher Scientific 14-385-939
UV-Vis Hewlett-Packard 8452A
Combined color reagent: Combine spectrophotometer Company
the following solutions in order,
mixing (but do not entrain air as oxygen oxides the ascorbic acid) after
each addition: (Prepare fresh weekly. Store at 4°C)
Stock A, (4.9 N H2 SO4 ) 50 mL
Stock B, (Ammonium molybdate solution) 15 mL
Stock C, (Ascorbic acid solution) 30 mL
Stock D, (Antimony-tartrate solution) 5 mL
Water diluent solution: Add 4.0 g sodium lauryl sulfate and 5 g NaCl per L of E-pure
water.
Stock phosphorus standard: Dissolve 0.4394 g of Potassium phosphate monobasic
(KH2 PO4 ) (dried at 105°C for one hour) in 900 mL E-pure water. Add 2 mL of
concentrated H2 SO4 and dilute to 1 L. 1.0 mL = 0.100 mg P (100 mg P/L).
Standard phosphorus solution: Dilute 1 mL of stock solution to 1 L. (1.0 mL = 0.1 µg
P) (100 µg P/L).