100% found this document useful (3 votes)
3K views

Prisma VENT

Uploaded by

Michael Levit
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (3 votes)
3K views

Prisma VENT

Uploaded by

Michael Levit
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 72

EN Service and Repair Instructions for devices of type WM 110 TD and

WM 120 TD

prisma VENT30
prisma VENT30-C
prisma VENT40
prisma VENT50
Ventilators
Contents

Contents
1 Introduction 3 7 Repair 36
1.1 About this document ...................................... 3 7.1 General information ...................................... 36
1.2 Navigation in this document ............................ 3 7.2 Open therapy device ..................................... 36
1.3 Warnings in this document .............................. 4 7.3 Close therapy device ..................................... 39
2 Product description 5 7.4 Replacing the valve control board (only
prisma VENT50) ............................................. 41
2.1 Overview ......................................................... 5
7.5 Replacing the internal battery ........................ 44
2.2 Control panel .................................................. 7
7.6 Replacing the pressure build-up valve (only
2.3 Symbols in the display ..................................... 7
prisma VENT50) ............................................. 47
2.4 Description of function .................................... 9
7.7 Replace main board ....................................... 49
2.5 Navigating in the device ................................ 11
7.8 Replacing the knob/encoder .......................... 53
3 Servicing 12 7.9 Replace front wall of housing with display ..... 55
3.1 General information ...................................... 12 7.10 Replace central part of housing ..................... 55
3.2 Intervals ........................................................ 12 7.11 Replace blower .............................................. 58
3.3 Change filters ................................................ 12 7.12 Replace power supply unit and its
3.4 Perform firmware update .............................. 13 connecting cable ........................................... 61
4 Hygiene treatment 14 7.13 Replace rear of housing ................................. 63
4.1 General information ...................................... 14 8 Storage and disposal 64
4.2 Hygiene treatment for therapy device ............ 14 8.1 Storage ......................................................... 64
4.3 Hygiene treatment for humidifier .................. 20 8.2 Disposal ........................................................ 64
5 Final check 21 9 Replacement parts 65
5.1 General information ...................................... 21 9.1 Replacement parts list .................................... 65
5.2 Activating service mode ................................. 21 9.2 Service sets .................................................... 67
5.3 Conduct test ................................................. 24 10 Components and accessories 69
5.4 Perform function check/TSC in line with §11
of the MP BetreibV [German law governing the 11 Auxiliary equipment required 70
owners of medical devices] ............................ 25 11.1 Tools ............................................................. 70
11.2 Test equipment ............................................. 70
6 Faults 31
11.3 Disinfectants ................................................. 70
6.1 Humidifier faults ............................................ 31
6.2 Display messages ........................................... 31
6.3 Fault memory ................................................ 32

WM 68181b 07/2017

2 | EN
1 Introduction

1 Introduction
1.1 About this document
The objective of these Servicing and Repair Instructions is to familiarize
you with the function, technology, servicing and repair of the device.
This will allow you to:
• Eliminate faults
• Perform function checks
• Perform repairs
Observe the information below when doing so.
• Repairs/servicing work may only be carried out by
Löwenstein Medical or by expert trained specialists with a
valid certificate. You are responsible for repairs carried out
yourself and for guaranteeing them!
• Use only genuine Löwenstein Medical replacement parts.
• Comply with accompanying documents (Instructions for
Use).
If you have any questions or feedback, please contact our Technical
Service team:
T: +49 40 54 70 2-100
e-mail: [email protected]

1.2 Navigation in this document


The buttons explained below provide a quick way to navigate through
this document:
• To view the last page you looked at, click the Back button.
Example: If you are on page 30 and the last page you viewed was
page 20, clicking the Back button will take you back to page 20.
• To return to the page you were viewing prior to clicking the Back
button, click the Forward button.
Example: If you were on page 20 and you clicked the Back button
to skip back to page 10, clicking the Forward button would return
you to page 20.
• Click the Contents button to go to the Table of Contents.
• Click the Previous Page button to go to the previous page
• Click the Next Page button to go to the next page.
WM 68181b 07/2017

EN |3
1 Introduction

1.3 Warnings in this document


Warnings indicate information relevant to safety.
Within procedures, you will find warnings in front of a step which
contains a hazard to persons or objects.
There are three levels of warning depending on the degree of hazard:

Danger!
Indicates an unusually significant hazardous situation. If you ignore this
instruction, severe irreversible injuries or death will result.
Warning!
Indicates an unusually significant hazardous situation. If you ignore this
instruction, severe irreversible or fatal injuries may result.
Caution!
Indicates a hazard. If you ignore this instruction, mild or moderate
injuries may result.
Notice!
Indicates a harmful situation. If you ignore this instruction, material
damage may result.

Indicates useful information within procedures.

WM 68181b 07/2017

4 | EN
2 Product description

2 Product description
2.1 Overview
2.1.1 prisma VENT30, prisma VENT30-C,
prisma VENT40

2
1

4
3 5

13 6
14

15 12

11 7
9 8

17
16

10

1 Humidifier connection with cover


2 Control panel with display
3 System interface for connecting modules
4 Handle
5 Release catch
6 Filter compartment with air filter (and optional pollen filter)
7 O2 supply (option)
8 Connection for power supply cable
9 Facility for connecting optional strain relief
10 Breathing tube with connection for breathing mask
11 Latching bores for connecting modules
12 SD card slot
13 Connection for tube heater
14 Device outlet port
15 Power cord
16 SD card
17 O2 connector (option)
WM 68181b 07/2017

EN |5
2 Product description

2.1.2 prisma VENT50

2
1
4

3 5

6
14
15
13
16

12 7
9 8
10
11

18
17

1 Humidifier connection with cover


2 Control panel with display
3 System interface for connecting modules
4 Handle
5 Release catch
6 Filter compartment with air filter (and optional pollen filter)
7 Cooling air opening
8 O2 supply
9 Connection for power supply cable
10 Facility for connecting optional strain relief
11 Breathing tube with connection for breathing mask
12 Latching bores for connecting modules
13 SD card slot
14 Connection for tube heater, valve control tube and pressure
measuring tube
15 Device outlet port
16 Power cord
17 SD card
18 O2 connector
WM 68181b 07/2017

6 | EN
2 Product description

2.2 Control panel

1
10
9
2

8 Enter 3

7 5 4
6
1 Alarm acknowledgment key - mutes an alarm for 2 minutes
2 LIAM key (only present on prisma VENT50)
3 Dial for navigating in the menu
4 Function keys - have different functions
5 Monitor key for switching between different screen views
6 Home key - switches the view back to the start screen
7 Power supply indicator
8 On/off key
9 Humidifier key
10 Program key for selecting preconfigured programs

2.3 Symbols in the display


SYMBOL DESCRIPTION
Device in patient mode. Expert area disabled.

Expert area enabled.

Set tube diameter 15 mm

Set tube diameter 22 mm

Device on standby. The blower is off.

Air filter change required.


Servicing required (only if servicing function is
activated).
Humidifier connected but not active (gray symbol)

Humidifier switched on (green symbol)

Humidifier empty (orange symbol)

Operation with bacteria filter set in the menu

Pulse rate (if pulsoximetry sensor connected)


SpO2 sensor connected
WM 68181b 07/2017

EN |7
2 Product description

SYMBOL DESCRIPTION
prisma2CLOUD module connected

prismaCONNECT module module connected

prisma CHECK module connected


prismaPSG module connected

Network connection present.

SD card inserted.
Indicates respiratory status:
Arrow pointing upward: Inspiration
Arrow pointing downward: Exhalation
S: Spontaneous breath
T: Mandatory breath
Target volume switched on.

AirTrap Control switched on.

5 segments green: Battery capacity above 85 %

4 segments green: Battery capacity above 65 %

3 segments green: Battery capacity above 45 %

2 segments green: Battery capacity above 25 %

1 segment orange: Battery capacity below 25 %

1 segment red: Battery capacity below 10 %

0 segments: Battery capacity below 5 %

Battery fault

Low-priority alarm triggered.


Medium-priority alarm triggered.
High-priority alarm triggered.

Alarm/alarms deactivated.

Acoustic signal for alarm paused for 2 minutes.

Acoustic signal for alarm deactivated.


WM 68181b 07/2017

8 | EN
2 Product description

2.4 Description of function


2.4.1 Therapy devices
Therapy devices of type WM 110 TD and WM 120 TD essentially
consist of the following components:
• Housing
• Blower
• Main board
• Valve control board (only prisma VENT50)
• Pressure build-up valve (only prisma VENT50)
• Pressure drop valve (only prisma VENT50)
• Display
• Control panel and dial (encoder)
• Power supply unit
The electronically-controlled blower draws in ambient air through a
filter, compresses the air and delivers it to the device outlet port at
therapy pressure. From here, air flows through the patient circuit and
the patient/ventilator interface (nasal cannula, tracheostomy, tube) to
the patient. The exhalation system upstream of the patient/ventilator
interface or optionally integrated in the patient/ventilator interface
prevents CO2-enriched exhaled air accumulating in the patient circuit.
Sensors on the main board enable the therapy device to determine and
analyze the pressure and respiratory flow signal at the patient/
ventilator interface and in the patient circuit, as well as in the
respiratory phase switch. The blower accordingly provides the
respiratory volume and the ventilation pressures prescribed by the
physician. The pressure sensors are connected to the device outlet port
via tubes and a Y-piece, allowing pressure to be measured close to the
device.
The flow sensor is connected directly to the flow element via 2 seals.
The air flowing through the flow element is guided through a large
number of fins which forces a laminar flow pattern. A drop in pressure
results from this laminar flow pattern in accordance with the laws of
flow mechanics. Differential pressure is determined by the flow sensor
and flow is calculated with the aid of other status variables such as
absolute pressure, for example. The sensors are calibrated when the
main board is manufactured and do not need to be recalibrated.
Pressure, flow, leakages and volumes can be saved and/or
(synchronously) output to a PSG system in analog form via the
prismaCONNECT module.
The therapy data are stored in the device and on an SD card for
monitoring therapy. In order to save the therapy data even if there is
no power supply in the device, the memory component is supplied
with power by a buffer battery. The buffer battery can be replaced if
required, it is inserted in a battery holder on the main board.
The device is operated by keys and a dial.
The device can be controlled remotely via the prismaTS therapy
software. In the event of a power failure, the settings are retained and
therapy is continued once the power supply is restored.
WM 68181b 07/2017

EN |9
2 Product description

2.4.2 Control PEEP valve (only prisma VENT50)


The PEEP control takes place via a pressure build-up and pressure drop
valve. The valves are controlled by the valve control board.
The patient-related therapy pressure and the PEEP control pressure are
recorded via a pressure sensor on the valve control board.
During the inspiratory phase, the pressure drop valve is closed and the
pressure build-up valve is opened: thereby a secondary flow of the
therapy pressure built up by the blower builds up the control pressure
required to close the PEEP valve via the pressure-side connection of the
central part of housing.
As soon as the required control pressure is reached and the PEEP valve
is closed, the pressure build-up valve is also closed and the pressure is
maintained.
When changing the respiratory phase to expiration, the pressure drop
valve is opened and the PEEP control pressure builds up via the suction
side connection to the central part of housing and the PEEP valve
opens.

2.4.3 Main board


The main board forms the heart of the therapy device. It includes
important components such as the flow sensor and pressure sensor,
memory component and controller, for example. The controller is
programmed with the relevant device firmware which comprises the
complete algorithm for controlling the device.
The main board performs the whole energy management process for
the device.

2.4.4 Display
The display measuring 4.3" is for displaying all information. It is
permanently connected (bonded) to the front wall of the housing.

2.4.5 Blower
The blower is a radial blower with a finely-weighted fan wheel and a
3-phase motor with Hall sensors to detect speed continuously. The
blower is actuated by the main board, with blower speed dependent
on current pressure requirements and current leakage.
WM 68181b 07/2017

10 | EN
2 Product description

2.5 Navigating in the device


RESULT
ACTION WITHIN A MENU
IN THE MENU
ITEM
Press function key
Function is displayed directly in the display via
the key (e.g. System or Back menu).

Turn dial to the left Navigate upward Reduce value


Turn dial to the right Navigate downward Increase value
Press the dial Select menu item Confirm set value
Press Home key
Back to start screen

Press Monitor key


Switches between different screen views.
WM 68181b 07/2017

EN | 11
3 Servicing

3 Servicing
3.1 General information
Servicing, safety checks [Sicherheitstechnische Kontrollen or STKs] and
maintenance measures such as servicing and repairs may only be
performed by the manufacturer or by specialists expressly so
authorized by the manufacturer.

3.2 Intervals
The therapy device is designed for a service life of 6 years.
Once these 6 years have elapsed, a complete final check (see "5 Final
check", page 21) should be performed.
In the event of use in accordance with purpose in line with the
Instructions for Use, the therapy device requires no maintenance.
In the event of use in accordance with purpose in line with the
Instructions for Use, the humidifier requires no maintenance.
If used and cleaned daily, the humidifier can be used for > 6 months.
In Germany, an STK [sicherheitstechnische Kontrolle or safety check] is
a legal requirement under §11 MPBetreibV [the German law governing
the owners of medical devices] and must be performed every 2 years
(see "5 Final check", page 21).

3.3 Change filters


1. Remove air filter WM 29651.
2. If present: Remove pollen filter WM 29389.
3. Put in new pollen filter WM 29389.
4. Put in new air filter WM 29651.

Reset filter change interval


You must set the filter change interval to zero when you have changed
the air filter and the reminder to change the air filter is activated.
1. Switch on therapy device.
2. To call up the Expert area, keep Home key depressed for
> 3 seconds.
3. Press the System function key.
4. Select Device settings.
5. Select Timer Filter: Select On.
6. Select Reset.
WM 68181b 07/2017

7. Confirm entry with Yes.

12 | EN
3 Servicing

8. To exit the Expert area, keep Home key depressed for


> 3 seconds.

Reset service indicator


Applies only when service indicator is activated: You must reset the
service indicator after every service and every hygiene treatment in
order to zero the hours counter for servicing.
1. To call up the Expert area, keep Home key depressed for
> 3 seconds.
2. Press the System function key.
3. Select Device settings.
4. Select Service timer and set the number of months until the next
reminder.
5. Select Reset.
6. Confirm entry with Yes.

3.4 Perform firmware update


Requirement
The therapy device is disconnected from the power supply or
completely switched off during battery operation.
1. Save the current firmware for the therapy device to a separate SD
card. You will find the up-to-date firmware available to download
from the login area of the manufacturer’s Internet site.
2. Insert SD card in the SD card slot.
3. Connect power supply or for battery pack operation, switch to
operating state Standby.

4. Confirm Perform firmware update with the dial.


The device automatically installs the new firmware.
WM 68181b 07/2017

EN | 13
4 Hygiene treatment

4 Hygiene treatment
4.1 General information
• This product may contain disposable items. Disposable items
are intended to be used only once. Therefore, use these items
only once and do not reprocess them. Reprocessing disposable
items may impair the functionality and safety of the product and
lead to unforeseeable reactions due to ageing, embrittlement,
wear, thermal load, the effects of chemical processes etc.
• Wear appropriate safety gear for the disinfecting process.
• Refer to the Instructions for Use for the disinfectant used.
• The Instructions for Use for the device, the components and the
accessories should also be followed.
• If there is evidence of contamination with critical germs (MRSA,
3Mrgn etc.) of the parts inside the device which carry air, the
device must be reprocessed in line with the Keredusy method or
disposed of. Reprocessing in line with the Keredusy method can be
performed up to 5 times (also in the course of regular processing).
Please follow the manufacturer's instructions for this process
(www.medizinservice-sachsen.de). Air filter WM 29651 and the
pollen filter (if present) must also be replaced. Once the Keredusy
method has been used 5 times, components must be replaced in
line with the set for change of patient, WM 15913.

4.2 Hygiene treatment for therapy


device
4.2.1 Hygiene treatment in use
Follow the Instructions for Use for the therapy device.

4.2.2 Hygiene treatment on repair


1. Disinfect external housing and power cord by wiping them down.
2. Clean or replace the following parts in accordance with the
Instructions for Use (depending on condition):
• Breathing tube WM 24445
• Mask with headgear
3. Open therapy device (see "7.2 Open therapy device", page 36).
4. Clean extremely soiled areas.
5. Change filters (see "3.3 Change filters", page 12).
6. Close therapy device (see "7.3 Close therapy device", page 39).
WM 68181b 07/2017

14 | EN
4 Hygiene treatment

4.2.3 Hygiene treatment on change of patient


Risk of infection when the device is used again!
If the device is used by several patients, infections may be transmitted
to the next patient.
 Use of a bacteria filter is obligatory when the device is used for
several patients.

Requirement
The device was used with a bacteria filter.
1. Disinfect the following parts by wiping:
• Housing, handle and device outlet port
• Power cord
2. Replace the following parts:
• Breathing tube WM 24445
• Mask with headgear
• Carrying bag WM 29337
3. Set device to factory settings to clear all previous settings.
4. Clear therapy data.
5. Recommended: Perform a complete final check (see "5 Final
check", page 21).

4.2.4 Enhanced cleaning in the context of hygiene


treatment
All parts replaced here can be found in the set for cleaning WM 15913
or WM 15916 (depending on device type).
1. Open therapy device (see "7.2 Open therapy device", page 36).
2. Remove blower (see "7.11.1 Remove blower", page 58).

Clean rear of housing


1. Remove flow element 50.
50 2. Remove sealing ring for rear of housing 38.
38 3. Remove insulating elements from rear of housing. In doing so,
please note: Insulating element 70 is glued on and needs to be
pulled off carefully.

70
WM 68181b 07/2017

EN | 15
4 Hygiene treatment

4. Cover the power supply unit.


5. Disinfect the rear of the housing by spraying it.

6. Insert new insulating elements.

WM 68181b 07/2017

16 | EN
4 Hygiene treatment

7. Glue on insulating element 70 so that half of it is located on the


insulating element below it and half is located on the housing wall.
70 There is a line on the housing wall to align the insulating element
correctly.
50
8. Insert new sealing ring for rear of housing 38.
9. Insert new flow element 50.
38

Clean central part of housing


1. Remove insulating elements from the central part of the housing.

2. Disinfect the central part of the housing by spraying it.


WM 68181b 07/2017

EN | 17
4 Hygiene treatment

3. Insert new insulating elements.

Clean blower
1. Release decoupling tube 15 and spacer 54 from blower 53.
54

15

53

2. Undo and remove 3 screws 56 on blower 53.


3. Remove blower cap from blower 53.
4. Disinfect the following parts by wiping:
• Mass body
• Blower cable
5. Disinfect the blower cap and the fan wheel by spraying.
53
56

56
WM 68181b 07/2017

18 | EN
4 Hygiene treatment

Material damage if screws are tightened against the blower


cap!
The blower cap may be damaged when screws are tightened.
 Only tighten screws against the blower cap gently.

6. Replace blower cap 55 on blower 53.


Ensure correct positioning as you do so.
7. Secure the blower cap with the 3 new screws 56.
55

53

56

8. Attach the new decoupling tube 15 and new spacer 54 to


54 disinfected blower 53.
Ensure correct positioning as you do so.
9. Fit blower (see "7.11.2 Fit blower", page 59).
15 10. Close therapy device (see "7.3 Close therapy device", page 39).
11. Replace air filter and pollen filter (if present) (see "3.3 Change
filters", page 12).
12. Reset therapy device to factory settings.
13. Check therapy device (see "5 Final check", page 21).

53
WM 68181b 07/2017

EN | 19
4 Hygiene treatment

4.3 Hygiene treatment for humidifier


4.3.1 Hygiene treatment in use
Follow the Instructions for Use for the therapy device.

4.3.2 Hygiene treatment on change of patient


Requirement
The therapy device was previously used without a bacteria filter.
1. Check plastic parts and replace if damaged (e.g. cracked).
2. If plastic parts and heater rod are heavily scaled: Offer new device.
3. Use the set for change of patient, WM 29974, for the hygiene
treatment.
4. Remove the upper part of the humidifier.
5. Remove the humidifier insert.
6. Unscrew the heating element from the lower part of the
humidifier.
7. Release the O-ring from the heating element.
8. Disinfect the heating element and the upper part of the humidifier
by wiping them.
9. Place a new O-ring on the heating element.
10. Screw the heating element into the new lower part of the
humidifier.
11. Insert the new humidifier insert into the upper part of the
humidifier.
12. Close the humidifier.

4.3.3 Hygiene treatment on change of patient and


if used with bacteria filter
Follow the Instructions for Use for the therapy device.

WM 68181b 07/2017

20 | EN
5 Final check

5 Final check
5.1 General information
To simplify the test sequence, the therapy device has a programmed
step-by-step test. This is available in activated Service mode (see "5.4.2
Perform step-by-step test", page 25).
The step-by-step test comprises all the test steps necessary to check the
functionality of the therapy device.
The test sequence is independent of current therapy device settings,
i.e. patient parameters are retained unchanged.
Other service functions are furthermore available in Service mode (see
"5.2 Activating service mode", page 21).
• Perform a final check:
• After every service
• After every repair
• After every hygiene treatment
• In the final step of the step-by-step test, there is the option of
having an electronic test record generated in PDF format. The test
record can be saved on the SD card of the device.
• If you find faults or deviations from specified values during the final
check, do not use the therapy device again until the faults have
been rectified.
• Note that the final check is only possible from firmware version
1.3.0 upwards. Perform a firmware update if there is a lower
version of the firmware on the device (see "3.4 Perform firmware
update", page 13).
• Use the final check to troubleshoot. A summary of possible faults
can be found in the section entitled “Faults” (see "6 Faults", page
31).
• The auxiliary equipment required can be found in the section
entitled “Auxiliary equipment required” (see "11 Auxiliary
equipment required", page 70).

5.2 Activating service mode


The therapy device has a Service mode which can be activated with the
aid of service connector WM 29917 which is available to trained
service engineers.
1. Switch the device off completely.
2. Disconnect the power supply.
3. Connect service connector to the system interface.
4. Connect power supply.
5. Switch device to Standby.
WM 68181b 07/2017

EN | 21
5 Final check

The service menu appears after approximately 10 seconds. The service


menu is only available in English and contains the functions below:

FUNCTION TAB PARAMETER DESCRIPTION SETTING OPTION


Device type Displays the device type activated.
Device
Displays the startup logo activated.
Branding
Indicates whether the device can be
Battery support
supplied by an internal battery
Battery Indicates whether a battery is
Versions, types Battery inserted yes/no
installed inserted.
Device serial Displays the serial number of the Enter serial number once if current
Versions, Types number therapy device. serial number is “0”.
Main board Displays the serial number of the Enter serial number once if current
serial number main board. serial number is “0”.
Blower Index Index of blower design Currently “0”
Mainboard HW Displays the hardware version of the
Automatic
version main board.
Versions Display HW Displays the hardware version of the
Automatic
version display.
FW version Displays the current firmware version. Automatic
For testing the color reproduction of Only for test purposes; selection of
Display test
the display and any pixel faults. different color tones.
Display
Display
Test display brightness
brightness test
Alarm button.
Red LED. LC
Alarm button.
Yellow LED. LC
Alarm LEDs Alarm button. Test alarm LEDs
Red LED. MC
Alarm button.
Yellow LED.
MC
Program
Display/ button LED
Buttons Humidifier
button LED
Power button
LED Tests the backlighting of the
LEDs Home button individual keys
LED
Info button LED
1 button LED
2 button LED
3 button LED
Start rotary
Rotary encoder Tests encoder function
encoder test
Start buttons
Buttons Tests key function
and LEDs test
Beeper test
main controller
Beeper control Beeper test Tests beeper
associated
controller
WM 68181b 07/2017

22 | EN
5 Final check

FUNCTION TAB PARAMETER DESCRIPTION SETTING OPTION


Flow, l/min Displays current flow
Ambient
Displays current ambient pressure
pressure
Mask pressure Displays current mask pressure
Pressure sensor Displays the measured value from
1 pressure sensor 1
Pressure sensor Displays the measured value from
Motor & 2 pressure sensor 2
pressure control For testing flow/pressure as a For test purposes only; motor
Motor speed
function of set blower speed. speed
For testing flow/pressure as a For test purposes only; specified
Mask pressure
function of a certain pressure. pressure
Motor current Tests the current consumption of the
test status blower
Blower using
Resets blower running time.
time, hours
Humidifier Displays the current consumption of
current the heating element.
Humidifier and Humidifier Tests the power supply to the heating For test purposes only; different
Motor, voltage
hose heater element. voltages
humidifier
control Hose heater Indicates whether a heatable hose is
connection connected.
Hose heater For activating hose heating.
Power supply Displays the current operating
voltage voltage of the therapy device.
Humidifier Displays the current supply voltage of
voltage the heating element.
Displays the current supply voltage of
Voltages Battery voltage
the battery.
Goldcap Displays the current voltage of the
voltage Goldcap capacitor.
Displays the current supply voltage of
Motor voltage
the blower.
Magnetic valve Used for manual opening/closing.
Used for infinitely variable opening/
Valves Piezo valve
closing.
(prisma VENT50
Airway Displays the current respiratory tract
only) %-indication
pressure pressure.
Pilot pressure Displays the current control pressure.
Current date/ Displays the current setting of the
Date, time Date, time
time Real Time Clock.
Device
Displays the total operating hours of Current total operating hours if
operating
the therapy device. currently < 10
hours
Next device
Indicates the date of the next service. Reset service indicator
Date, time service
reminders Using time/ Optional service reminder and Activation and setting of the
Device service
service dates specification of the service interval if optional service interval between 1
period, months
service reminder is activated. and 48 months
Next filter Displays the date of the next filter
Reset filter change indicator.
change change.
Filter change Optional activation of the filter Activates the filter change
reminder change indicator. reminder (interval 1 month)
WM 68181b 07/2017

EN | 23
5 Final check

FUNCTION TAB PARAMETER DESCRIPTION SETTING OPTION


SD card test
Storage Tests the SD card Tests card access
status
interfaces
SD card Tests the card reader
PSG UART test
Tests the PSG UART Tests PSG UART communication
status
Second MCU
External UART test Tests the second MCU UART Tests the second MCU UART
interfaces status
External
Tests I2C communication with the Tests I2C communication,
interfaces I2C test status
accessory part module accessory part must be connected
Com. Module Tests communication with accessory
on USB part
Pulsox. module Tests pulsoximetry module
Pulsoximetry module connection
on USB communication
Service files Fault memory Clears fault memory.
Delete therapy
Clears patient usage times.
data
Format SD card Clears the SD card. Clears the SD card.
Factory settings Reset to factory settings.
Copy data on
Clear, copy, Transfer data to SD card. Start transfer.
SD card
format
Import data
Adopt data from SD card. Start transfer.
from SD card
Copy service
protocol on SD Save test record to SD card Start transfer.
card
Battery control Display of battery and charge status
Step by step
Programmed test sequence
test
Device SW Displays the current firmware version
version of the device.
SW version of
Displays the firmware version on the
SW update update file on
SD card.
SD card
Status Indicates the status of the SD card.
SW update Starts a software update

5.3 Conduct test


5.3.1 Check housing
Requirement
Power cord is connected.
1. Check general condition of housing (visual inspection).
2. If the housing is damaged or faulty: Replace housing (see "7.12
Replace power supply unit and its connecting cable", page 61).

5.3.2 Check power cord


1. Check power cord.
Requirement:
WM 68181b 07/2017

• The insulation is OK.


• The cable is undamaged.

24 | EN
5 Final check

• There are no loose contacts.


2. If one of the requirements is not met: Replace power cord.

5.4 Perform function check/TSC in line


with §11 of the MP BetreibV
[German law governing the owners
of medical devices]
5.4.1 Check therapy device
Requirement
The patient has been disconnected from the therapy device.
1. Check therapy device for external damage.
If damaged: Do not use therapy device.
2. Check connectors and cables for external damage.
If damaged: Replace parts.

5.4.2 Perform step-by-step test


1. Activate Service mode: Connect service connector WM 29917 to
the system interface.
2. Connect power cord and supply power.
The Service menu opens after approximately 3 seconds.

3. Select Step by step test field.

4. Step 1: Enter tester name or insert SD card with tester data.


• Turn knob until the Tester name field is highlighted yellow.
• Press knob and enter the name.
• Confirm entry with a tick.
• Select Next.
WM 68181b 07/2017

EN | 25
5 Final check

5. Step 2: Select type of test.


• Select and confirm the desired type of testing.
• Select Next.

Note: The TSC in line with §11 MPBetreibV [German law


governing the owners of medical devices] is only a legal
requirement in Germany and must be carried out every 2 years.

6. Step 3: Visual inspection of external condition of housing and


power cord.
• Check general condition of housing (visual inspection).
• Check power cord.
Requirement:
Housing is in good condition and undamaged.
Insulation of the power cord is in good condition.
The cable is undamaged.
There are no loose contacts.
• Select Next.
7. Step 4: Check date and time.
• Compare date and time to a reference clock.
Requirement:
Date and time are correct.
• Confirm selection Date/time correct? with Yes.
• Select Next.

8. Step 5: Check knob/encoder.


• Select Start rotary encoder test.
• Follow the instructions in the display.
Requirement:
The selection Rotary encoder test passed? is automatically
confirmed with Yes.
• Select Next.

9. Step 6: Check keys and LEDs.


• Select Start buttons and LEDs test.
Requirement:
All the LEDs behind the keys light up. The alarm key lights up
orange.
• Select Yes to confirm the requirement.
• Press all the keys one after another.
Requirement:
The illumination behind each key pressed goes out. The Alarm key
switches from orange to red and starts flashing.
WM 68181b 07/2017

• Confirm that the Alarm key is flashing by selecting Yes.


• Press the Alarm key to stop the flashing.

26 | EN
5 Final check

Requirement:
All the LEDs are off.
• Confirm the requirement with Yes.
The selection Buttons and LED test passed? is automatically
confirmed with Yes.
• Select Next.
10. Step 7: Test alarm.
• Select Start beeper test.
Requirement:
Alarms of varying volume are heard.
• Confirm Did you hear beeper with 4 tones? with Yes.
• Select Next.

11. Step 8: Check the SD card.


• Select Start SD card test.
Requirement:
Display: SD card test status -> Passed
(if SD card present)
or
Display: SD card test status -> SD card not recognized
(if no SD card present)

• If no SD card present:
Insert SD card and select Start SD card test.
Requirement:
Display: SD card test status -> Passed
The selection SD card test passed? is automatically confirmed
with Yes.
• Select Next.
12. Step 9: Check PSG/system interface.
• Select Start PSG test.
Requirement:
Display: PSG UART test status -> Passed
(UART = Universal Asynchronous Receiver Transmitter)
The selection PSG test passed? is automatically confirmed with
Yes.
• Select Next.

13. Step 10: Check ambient pressure sensor.


• Enter current ambient pressure in the Reference ambient
pressure (mBar) field.
Requirement:
The ambient pressure displayed corresponds to current air
pressure. Deviation may not exceed 20 mbar.
The selection Ambient pressure test passed? is automatically
confirmed with Yes.
WM 68181b 07/2017

• Select Next.

EN | 27
5 Final check

14. Step 11: Check specified pressure values.


• Plug pressure measurement adapter onto the device outlet port
and seal it off.
• Connect the lateral connector of the pressure measurement
adapter to a pressure measurement device.
• Select Start pressure test.
• Read measured value off pressure measurement device at
4 mbar and enter it in the Reference pressure for 4 mbar
field.
• Read measured value off pressure measurement device at
15 mbar and enter it in the Reference pressure for 15 mbar
field.

The leak flow may be a maximum of 2 l/min.

• Read measured value off pressure measurement device at


40 mbar and enter it in the Reference pressure for 40 mbar
field.
Requirement:
The pressure values measured are within tolerance. The selection
Pressure test passed? is automatically confirmed with Yes.
• Remove pressure measurement adapter from device outlet port.
• Select Next.
15. Step 12: Check flow values
• If present: Remove adapter or breathing tube from device outlet
port.
• Select Start flow test.
Requirement:
The flow values displayed are within the specified tolerance. The
selection Flow test passed? is automatically confirmed with Yes.
• Select Next.

16. Step 13: PEEP test (only prisma VENT50).


• Close device outlet port with suitable plug.
• Connect pressure measurement port and connection valve
control with suitable tube.
• Select Start PEEP test.
Requirement:
The first three test steps are confirmed with OK.
• Open the device outlet port.
• Remove connecting tube.
• Select Continue test.
Requirement:
The individual test steps are confirmed with OK. The selection
PEEP test passed is automatically confirmed with Yes.
WM 68181b 07/2017

• If the test fails, perform pressure sensor check.


• Select Next.

28 | EN
5 Final check

17. Step 14: Check humidifier interface (heating element).


• Connect the humidifier.
or
Remove side cover and connect individual heating element to
the jack on the therapy device.
• Select Start humidifier test.
Requirement:
The voltage and current consumption displayed are each within
the specified tolerance. The selection Humidifier test passed? is
automatically confirmed with Yes.
• Select Next.
18. Step 15: Check Goldcap voltage.
• The device automatically checks whether the Goldcap voltage is
within the specified tolerance.
Requirement:
The selection Goldcap test passed? is automatically confirmed
with Yes.
• Select Next.

19. Step 16: Check battery (if present)


• The device automatically checks the status of the battery.
Requirement:
Display Battery status OK
The test duration and operating state of the device may vary
depending on the battery level.
• Select Next.

20. Step 17: Check summary of test result.


• The test results of each individual test item are displayed.
Requirement:
All test items have the status OK.
As confirmation that the requirements of §11 MPBetreibV
[German law governing the owners of medical devices] have been
met, the selection The device meets the requirements of
Article 11 of the [German] law governing the owners of
medical devices is automatically confirmed with Yes.
• Select Next.
21. Step 18: Record the test results.
• In order to generate a test record in PDF format, select Start
generating service protocol.
• To copy existing test records to the SD card, select Copy service
protocols on SD card.
WM 68181b 07/2017

22. Disconnect the service connector from the therapy device to exit
the Service menu.

EN | 29
5 Final check

Result The function check is complete.

5.4.3 Check humidifier


Requirement • The patient has been disconnected from the therapy device.
• The therapy device is connected to the power supply.
• The therapy device is on Standby.
1. Check the housing of the humidifier for cracks, damage and severe
soiling.
2. If there are cracks, damage or soiling: Replace housing parts or
humidifier insert.
3. Fill humidifier with water up to the mark.
4. Check whether the humidifier has any leaks.
If the humidifier leaks: Replace damaged parts.
5. Pour out water.
6. Fill humidifier with 200 ml water.
7. Connect humidifier to the therapy device.
8. Switch on humidifier.
9. Set the highest humidifier stage on the therapy device.
10. Check whether the humidifier is heating up.
If the humidifier does not heat up slightly after 10 minutes: Replace
heating element or main board (see "7.7 Replace main board",
page 49).
Result The function check is complete.

WM 68181b 07/2017

30 | EN
6 Faults

6 Faults
If you are unable to eliminate faults immediately with the help of the
table, contact the manufacturer. To avoid exacerbating the damage,
do not continue operating the device.

FAULT CAUSE OF FAULT REMEDY


Check that the power cord is securely connected.
Check function of socket.

Check the power supply with a different device


No running noise, nothing in the No power supply. (e.g. a lamp).
display.
If necessary: Replace power cord.
If necessary: Replace power supply unit (see 7.12,
p. 61).
Fuse faulty. Replace main board (see 7.7, p. 49).
No running noise after display comes Replace main board (see 7.7, p. 49).
Blower not running.
on. Replace blower (see 7.11, p. 58).
Autostart function not activated. Activate autoSTART-STOP function.
Therapy cannot be started by taking a autoSTART function may be
breath. restricted in the case of Contact your specialist dealer.
accessories with a high resistance.
Clean air filter. If necessary: Replace filter (see "4
Air filter dirty.
Therapy device does not reach Hygiene treatment", page 14).
therapy pressure. Adjust headgear so that the mask is tight.
Breathing mask leaking.
If necessary, replace faulty mask.
Keys cannot be operated. Control panel faulty. Replace front wall of housing (see 7.9, p. 55).

6.1 Humidifier faults


FAULT CAUSE REMEDY
Humidifier heating switched off. Adjust humidifier heating.
Humidifier does not warm up.
Humidifier is faulty. Replace humidifier.
Heater element seal is faulty. Replace seal.
Humidifier insert is not inserted correctly. Insert humidifier insert correctly.
Humidifier is leaking.
Humidifier insert is faulty. Replace humidifier insert.
Cracks in lower part of humidifier. Replace lower part of humidifier.
Humidifier switches off. No water present in the humidifier. Fill humidifier with water.

6.2 Display messages


If the message Error (xxx): Please follow the instructions in the
Instructions for Use appears in the display, look in the table (see
"6.3.1 Open/clear fault memory", page 32) for the fault code
displayed. Remedy the fault in accordance with the description.
WM 68181b 07/2017

EN | 31
6 Faults

6.3 Fault memory


The therapy device has an internal fault memory which can store up to
100 fault entries. Once the maximum number of entries has been
reached, the oldest entry is overwritten.
Fault entries are displayed as follows:
date - time - number of fault message - description

6.3.1 Open/clear fault memory


1. Activate Service mode (see 5.2)
2. Select Service files.
3. To delete the error memory: Select Delete all.
4. Confirm the confirmation prompt with Yes.
or
5. Select No to cancel.
The causes of the individual faults can be found in the table below.

FAULT
DESCRIPTION CAUSE REMEDY
NUMBER
101 Sensor fault Main board faulty Replace main board
102 Sensor fault Main board faulty Replace main board
Main board faulty Replace main board
103 Sensor fault
Device outlet port incorrectly assembled
Check device outlet port assembly
104 Sensor fault Main board faulty Replace main board
105 Sensor fault Main board faulty Replace main board
106 Sensor fault Main board faulty Replace main board
107 Sensor fault Main board faulty Replace main board
Set date and time so that the course of
Time not set
therapy is set correctly.
108 Real Time Clock fault
Battery on main board discharged Replace battery on main board
Main board faulty Replace main board
109 Real Time Clock fault Main board faulty Replace main board
110 Real Time Clock fault Main board faulty Replace main board
111 Real Time Clock fault Main board faulty Replace main board
Main board faulty Replace main board
113 Self-test fault
Blower faulty Replace blower
151 Sensor fault Main board faulty Replace main board
152 Sensor fault Main board faulty Replace main board
Main board faulty Replace main board
201 Blower speed fault
Blower faulty Replace blower
Main board faulty Replace main board
202 Blower speed fault
Blower faulty Replace blower
Main board faulty Replace main board
203 Sensor fault Pressure measurement tube kinked/ Check position of pressure measurement
blocked tube/remove blockage, replace tube
Main board faulty Replace main board
205 Voltage supply fault
Power supply unit faulty Replace power supply unit
Replace prismaCONNECT module
206 Module fault prismaCONNECT module module faulty
module
WM 68181b 07/2017

32 | EN
6 Faults

FAULT
DESCRIPTION CAUSE REMEDY
NUMBER
Main board faulty Replace main board
PSG module faulty Replace PSG module
207 Module fault
prisma CHECK faulty Replace prisma CHECK
Connecting cable faulty Replace connecting cable
251 Blower speed fault Blower faulty Replace blower
Check tubing and position of pressure
Tubing for the pressure sensors faulty
252 Sensor fault measurement tubes
Main board faulty Replace main board
Main board faulty Replace main board
253 Sensor fault Pressure measurement tube kinked/ Check position of pressure measurement
blocked tube/remove blockage, replace tube
254 Flow fault Main board faulty Replace main board
Power supply
255 Main board faulty Replace main board
fault
Power supply
256 Main board faulty Replace main board
fault
Power supply
257 Main board faulty Replace main board
fault
Power supply
258 Main board faulty Replace main board
fault
Power supply
259 Main board faulty Replace main board
fault
Humidifier faulty Replace humidifier
260 Humidifier fault
Main board faulty Replace main board
Humidifier faulty Replace humidifier
261 Humidifier fault
Main board faulty Replace main board
500 Controller fault Main board faulty Replace main board
501 Controller fault Main board faulty Replace main board
Send therapy device to the manufacturer
502 Initialization fault Further investigation required
or replace main board
Send therapy device to the manufacturer
504 Initialization fault Further investigation required
or replace main board
Send therapy device to the manufacturer
505 Initialization fault Further investigation required
or replace main board
Send therapy device to the manufacturer
507 Initialization fault Further investigation required
or replace main board
Send therapy device to the manufacturer
508 Initialization fault Further investigation required
or replace main board
Send therapy device to the manufacturer
509 Initialization fault Further investigation required
or replace main board
Send therapy device to the manufacturer
510 Initialization fault Further investigation required
or replace main board
550 Battery fault Battery capacity < 5% Connect the device to the power supply
551 Battery fault Battery capacity < 10% Connect the device to the power supply.
552 Battery fault Battery faulty Replace battery
553 Battery fault Battery not present Fit battery
Third-party battery Fit genuine battery
554 Battery fault
Battery faulty Replace battery
Operate device at an ambient
555 Battery fault Battery temperature on limit value
temperature of 5 °C to 40 °C
Operate device at an ambient
556 Battery fault Battery temperature too high
WM 68181b 07/2017

temperature of 5 °C to 40 °C

EN | 33
6 Faults

FAULT
DESCRIPTION CAUSE REMEDY
NUMBER
Connect the device to the power supply.
557 Battery fault Battery stored for too long If fault is still being displayed after 8 h,
replace battery
558 Battery fault Battery cannot be charged Replace battery
Battery cannot be charged due to high Operate device at an ambient
559 Battery fault
temperature temperature of 5 °C to 40 °C
Battery cannot be charged due to low Operate device at an ambient
560 Battery fault
temperature temperature of 5 °C to 40 °C
561 Battery fault Battery service life at an end Replace battery
Power supply
580 All power supplies have failed Provide power supply
fault
Power supply Press alarm acknowledgment key. The
581 Device in battery mode
fault device is in battery mode.
Power supply Battery on main board Replace battery on main
582
fault discharged board
Power supply
583 Main board faulty Replace main board
fault
Humidifier or cover not connected
Connect humidifier or cover correctly
without leaks
Main board faulty Replace main board
Pressure measurement
701 Pressure measurement tube not
section fault Connect pressure measurement tube
connected
Pressure measurement tube kinked/ Check position of pressure measurement
blocked tube/remove blockage, replace tube
Check position of breathing tube or
Breathing tube kinked
replace breathing tube
Device outlet port blocked Remove blockage at device outlet port
702 Flow section fault
Main board faulty Replace main board
Assembly fault Check flow section assembly
Restrictor faulty Replace restrictor
Remove blockage in intake area or
Intake area blocked
703 Flow section fault replace air filter
Blower faulty Replace blower
Humidifier or cover not connected
Connect humidifier or cover correctly
without leaks
Main board faulty Replace main board
Power supply
751 Pressure measurement tube not
fault Connect pressure measurement tube
connected
Pressure measurement tube kinked/ Check position of pressure measurement
blocked tube/remove blockage, replace tube
Check connection from device to the
752 VBS fault Disconnection for > 2 min. patient/ventilator interface at the patient
via the breathing tube
753 VBS fault Leakage too low Check exhalation system
Fault as a consequence of 701, 203, 751,
754 Remedy primary fault
253, 252
SpO2 alarms active and no SpO2 module Deactivate SpO2/pulse alarms or
793 Module fault
present when ventilation starts reconnect Prisma Check
Further investigation Send therapy device to the manufacturer
801 Controller fault
required or replace main board
Further investigation Send therapy device to the manufacturer
802 Controller fault
required or replace main board
WM 68181b 07/2017

Further investigation Send therapy device to the manufacturer


804 Controller fault
required or replace main board

34 | EN
6 Faults

FAULT
DESCRIPTION CAUSE REMEDY
NUMBER
Further investigation Send therapy device to the manufacturer
805 Controller fault
required or replace main board
Further investigation Send therapy device to the manufacturer
806 Controller fault
required or replace main board
Further investigation Send therapy device to the manufacturer
809 Controller fault
required or replace main board
Further investigation Send therapy device to the manufacturer
810 Controller fault
required or replace main board
Further investigation Send therapy device to the manufacturer
811 Controller fault
required or replace main board
Further investigation Send therapy device to the manufacturer
821 Initialization fault
required or replace main board
Further investigation Send therapy device to the manufacturer
824 Initialization fault
required or replace main board
Further investigation Send therapy device to the manufacturer
851 Controller fault
required or replace main board
Further investigation Send therapy device to the manufacturer
852 Controller fault
required or replace main board
Further investigation Send therapy device to the manufacturer
853 Controller fault
required or replace main board
Further investigation Send therapy device to the manufacturer
854 Controller fault
required or replace main board
Further investigation Send therapy device to the manufacturer
860 Controller fault
required or replace main board
Further investigation Send therapy device to the manufacturer
861 Controller fault
required or replace main board
Further investigation Send therapy device to the manufacturer
862 Controller fault
required or replace main board
If only listed in the service files, no further
Delay when firmware is started
action is required
870 Controller fault
Further investigation Send therapy device to the manufacturer
required or replace main board
Further investigation Send therapy device to the manufacturer
871 Controller fault
required or replace main board
Further investigation Send therapy device to the manufacturer
872 Controller fault
required or replace main board
Further investigation Send therapy device to the manufacturer
880 Controller fault
required or replace main board
Further investigation Send therapy device to the manufacturer
897 Update fault
required or replace main board
Further investigation Send therapy device to the manufacturer
898 Update fault
required or replace main board
Update SD card removed before files Insert update SD card again and restart
899 Update fault
could be copied update
WM 68181b 07/2017

EN | 35
7 Repair

7 Repair
7.1 General information
• Follow the Instructions for Use and the Service and Repair
Instructions when carrying out repairs. Follow the safety
instructions in the Instructions for Use in particular.
• Perform repairs only at an ESD workstation.
• Make sure the workstation is clean.
• Only perform repairs described in these Service and Repair
Instructions.
• Perform a final check after any repair (see "5 Final check", page
21).
• Use only genuine Löwenstein Medical replacement parts.
• The auxiliary equipment required can be found in the section
entitled “Auxiliary equipment required”.
• The item numbers in this section are identical to the item numbers
in the section entitled “Replacement parts”.
• Use a sorting aid for screws and washers.

7.2 Open therapy device

Risk of injury from electric shock!


Opening the therapy device with the power supply connected may lead
to electric shocks.
 Disconnect the power supply before opening.
Disconnect the power cord from the therapy device before opening the
device.

1. Press the catch key and remove cover 7/humidifier from the
therapy device sideways.

7
WM 68181b 07/2017

36 | EN
7 Repair

2. Place the device on the table so that the rear of the housing 1 faces
4 downwards.
29 3. Undo the two screws 29 of front panel 4 on the underside of the
device.
4. Take panel 4 off upwards.
5. If present: Remove internal battery (see "7.5.1 Removing the
internal battery", page 44).

6. Undo the 2 screws 29 of the front of the device 3.

29
3

7. Place the device on its underside and remove the 2 plugs 45 on the
rear of the device.
8. Undo 2 screws 29.

29 45

45
29
WM 68181b 07/2017

EN | 37
7 Repair

9. Swivel the front of the device 3 away upwards and lay the device
3
on its rear 1.

10. Detach connectors X1300, X1500, X302, X1501, X300 and


connection line main board->valve control board 20 (only
prisma VENT50).

20
11.Only prisma VENT50: Detach tube connection for double patient
circuit pressure measurement/PEEP control 21 to the tube
connectors.

20
WM 68181b 07/2017

38 | EN
7 Repair

7.3 Close therapy device


When closing the device, ensure that connecting cable for the battery
main board 14 is attached to the central part of the housing with cable
tie 61 (see "7.10.2 Fit central part of housing", page 56).
1. Place the front of the housing on the table with the display side in
front of the blower box.
2. Connect the connectors X1300, X1500, X302, X1501, X300 and
connection line main board->valve control board 20 (only
prisma VENT50).

20

3. Only prisma VENT50: Connect tube connection for double patient


circuit pressure measurement/PEEP control 20 to the tube
connectors.
Warning:
Pay attention to color coding, do not mix up tubes!
WM 68181b 07/2017

EN | 39
7 Repair

4. Hold the front of housing 3 up against the top edge of the blower
box and join both parts together. In doing so, please note:
3
• The cable guide and guidance for the pressure-side tube must
be correct.
• The sealing ring for the rear of the housing 38 must engage in
the device outlet port.

38

5. Place the device on its underside and insert the 2 screws 29 on the
rear of the device and tighten them up.
6. Insert 2 plugs 45.
29 45

45
29

7. Insert 2 screws 29 on the front of the front wall of the housing 3


and tighten them up.
8. If present: Fit internal battery (see "7.5.2 Installing the internal
battery", page 45).
29

3
WM 68181b 07/2017

40 | EN
7 Repair

9. Insert panel 4 on the front of the device.

4 10. Place the device on its rear 1 and insert and tighten the 2 screws
29 29 on the underside of the device.

11. Insert cover 7/humidifier.

7.4 Replacing the valve control board (only


prisma VENT50)
7.4.1 Removing the valve control board
1. Open therapy device (see "7.2 Open therapy device", page 36).
2. Disconnect plug connection to the pressure drop valve (gray cable)
and to the pressure build-up valve (black cable).
WM 68181b 07/2017

EN | 41
7 Repair

3. Undo 2 screws 29 from the valve control board 28.


29 4. Remove valve control board 28.

5. Remove both pressure measurement tubes from the pressure


sensors.

6. Connection line to valve control board - > main board 20 on the


connection.

20
WM 68181b 07/2017

42 | EN
7 Repair

7.4.2 Installing the valve control board


1. Connect the connector of the connection line for the valve control
board -> main board 20 to the new valve control board 28.

20

2. Connect the tube marked blue to the sensor B400 and the tube
without marking to the sensor B401.

3. Insert valve control board 28. Pay attention to the correct position
of the tubes and connecting cables.
29
Secure with 2 screws 29.
WM 68181b 07/2017

EN | 43
7 Repair

4. Connect the connecting cables from pressure build-up and pressure


drop valve.
5. Close therapy device (see "7.3 Close therapy device", page 39).

7.5 Replacing the internal battery


7.5.1 Removing the internal battery
If a device that was previously operated with an internal battery will be
used in the future without an internal battery, the configuration in the
service menu needs to be changed (Versions,types > SN,types >
Battery installed > select „No“).
1. Place the device on the table so that the rear of the housing 1 faces
4 downwards.
29
2. Undo the two screws 29 of front panel 4 on the underside of the
device.
3. Take panel 4 off upwards.

1
WM 68181b 07/2017

44 | EN
7 Repair

4. Undo 2 screws 29 of the cover for battery compartment 8 and


remove the cover.
29 5. Remove battery and disconnect plug connection.

7.5.2 Installing the internal battery


If a device that has previously been operated without an internal
battery will be retrofitted with a rechargeable battery, the battery is
detected automatically by the device, no manual configuration or
activation is required.

Material damage due to incorrect battery!


Batteries with a serial number below 20,000 do not provide adequate
ESD protection and may damage the device.
Only fit batteries with a serial number > 20,000.

1. Connect the battery to the connection line and insert it into the
battery compartment.
WM 68181b 07/2017

EN | 45
7 Repair

2. Insert battery compartment cover 8 and insert and tighten up the


2 screws 29.

29

3. Insert panel 4 on the front of the device.

4 4. Place the device on its rear 1 and insert and tighten the 2 screws 29
29 on the underside of the device.

WM 68181b 07/2017

46 | EN
7 Repair

7.6 Replacing the pressure build-up valve


(only prisma VENT50)
7.6.1 Removing the pressure build-up valve
1. Open therapy device (see "7.2 Open therapy device", page 36).
2. Removing the central part of housing (see "7.10.1 Remove central
part of housing", page 55).
3. Pull off tubes from pressure build-up valve 62.

62

4. Undo screw 29 from holder of pressure build-up valve 64 and


remove together with the holder.
5. Remove pressure build-up valve 62 from the central part of the
housing.

64

29
62
WM 68181b 07/2017

EN | 47
7 Repair

6. Undo connection line of pressure build-up valve to valve control


board 24 from the pressure build-up valve 62.

24 7. Remove plug 63 from connection 3 to pressure build-up valve 62.

63

62

7.6.2 Installing the pressure build-up valve (only


prisma VENT50)
1. Connect plugs 63 to connection 3 of the pressure build-up
valve 62.

24 2. Connect the connection line of pressure build-up valve to valve


control board 24 to the pressure build-up valve 62.
3. Insert pressure build-up valve 62 to the central part of housing.

63

62

4. Insert holder of pressure build-up valve 64 with the screw and


tighten screw 29.

64

29
WM 68181b 07/2017

48 | EN
7 Repair

5. Connect the tube with the Y-connectors to the lower connection of


the pressure build-up valve 62.
6. Connect the 2nd tube to the upper connection of the pressure
build-up valve 62.
7. Fit central part of housing (see "7.10.2 Fit central part of housing",
page 56).
8. Close therapy device (see "7.3 Close therapy device", page 39).

62

7.7 Replace main board


7.7.1 Remove main board
1. Remove SD card 58 from the SD card slot.
58 2. Open therapy device (see "7.2 Open therapy device", page 36).

3. Remove pressure measuring tube 18 from the pressure sensors.


18 19
4. Undo pressure-side tube 19 from device outlet port 57 and remove.
21
5. Remove double patient circuit for pressure measurement/ PEEP
57 control 21 from the holders (only prisma VENT50).
6. Disconnect plug connection X1401.
WM 68181b 07/2017

EN | 49
7 Repair

7. Release connecting cables X1900, X1901 and X1902.

8. Undo 4 screws 32 from the main board.

9. Lift the main board at the side and release the ribbon cable for
display 16 on the underside of the main board.
10. Remove the main board from the front of the device.

16
WM 68181b 07/2017

50 | EN
7 Repair

7.7.2 Fit main board


1. Connect the ribbon cable for the display 16 on the underside of
the main board.
2. Insert the main board in the front of the housing at an angle.
In doing so, please note: The cover for the system interface must
be inserted.

16
cover for the
system interface

3. Connect connecting cables X1900, X1901 and X1902.

4. Fix the main board in position using the 4 screws 32.


WM 68181b 07/2017

EN | 51
7 Repair

5. Connect connecting cable X401.


18 19
6. Connect pressure-side tube 19 to device outlet port 57.
21 Only prisma VENT50: Insert double patient circuit 21 into both
holders.
57 7. Connect pressure measuring tube 18 to pressure sensors.
8. Close therapy device (see "7.3 Close therapy device", page 39).

If required If the message Code 113 appears on the therapy device after the main
board has been replaced, complete the following steps:
9. Activate Service mode: Connect service connector WM 29917 to
the system interface.
10. Connect the power supply unit to the therapy device and provide
a power supply.
The Service menu opens after approximately 3 seconds.
11. Select the Date, time, reminders field.

12. Select Set other field.


13. Set date and time.
14. Select Set device date/time field.
15. Disconnect the power supply unit from the therapy device.
16. Disconnect the service connector from the therapy device to exit
the Service menu.
17. Connect the power supply unit to the therapy device.

Result The message Code 113 is no longer displayed.


WM 68181b 07/2017

52 | EN
7 Repair

7.7.3 Enter serial number of the device


Note: The serial number can only be set once and cannot be
overwritten.
1. Activate Service mode (see "5.2 Activating service mode", page
21).
2. Connect the power supply unit to the therapy device and provide
a power supply.
The Service menu opens after approximately 3 seconds.
3. Select Versions, types field.
4. Read serial number off label on side of device and enter under
Device serial number.

7.7.4 Adopt configuration and usage times


After the main board has been replaced, the saved configuration and
usage times from the original main board can be adopted.
1. Take the SD card out of the old main board and insert it in the new
main board.
2. Select Clear, copy, format.
3. Select Import compliance&config from SD card.
4. Done appears.

7.8 Replacing the knob/encoder


If you just want to replace the knob of the encoder, simply pull it off
the front of the device and replace it with a new one.

7.8.1 Removing the encoder


1. Open therapy device (see "7.2 Open therapy device", page 36).
2. Remove main board (see "7.7.1 Remove main board", page 49).
3. Pull dial 36 off the front of the device.
4. Undo and remove hexagon nut on the outside of the front of the
device.
5. Take out encoder 35 from the inside.

35

36
WM 68181b 07/2017

EN | 53
7 Repair

7.8.2 Installing the encoder


1. Insert encoder 35 into the front wall of the housing from the
inside.
2. Insert and tighten up hexagon nut on the outside of the front of
the device.
3. Press dial 36 onto the axle.
4. Fit main board (see "7.7.2 Fit main board", page 51).
5. Close therapy device (see "7.3 Close therapy device", page 39).
35

36

WM 68181b 07/2017

54 | EN
7 Repair

7.9 Replace front wall of housing with


display
7.9.1 Remove front wall of housing
1. Open therapy device (see "7.2 Open therapy device", page 36).
2. Remove main board (see "7.7.1 Remove main board", page 49).
3. Remove dial (see "7.8.1 Removing the encoder", page 53).
4. Replace front wall of housing.

7.9.2 Fit front wall of housing


1. Fit dial in new front wall of housing (see "7.7.1 Remove main
board", page 49).
2. Fit main board (see "7.7.2 Fit main board", page 51).
3. Close therapy device (see "7.3 Close therapy device", page 39).

7.10 Replace central part of housing


Starting from SN 10.000, for prisma VENT30/prisma VENT40 the
central part of housing of prisma VENT50 is also used. When
replacing, make sure that the connector stubs for venting the PEEP
control valve and the connector stubs for the pressure drop valve are
closed with the plugs.
The connection line of the battery main board of prisma VENT30/
prisma VENT40 before SN 10.000 is somewhat longer, but it can also
continue to be used with the new central part of housing.

7.10.1 Remove central part of housing


1. Open therapy device (see "7.2 Open therapy device", page 36).

29 2. Undo screw 29 on flow measurement adapter 51.

51 3. Remove flow measurement adapter 51. In the process, ensure that


the blue O-rings do not get stuck in the central part of the housing.
4. Remove humidifier connecting cable 17.

17
WM 68181b 07/2017

EN | 55
7 Repair

5. Only prisma VENT50: Removing the valve control board (see "7.4.1
31 Removing the valve control board", page 41).
6. Undo and remove 3 screws 31.
7. Undo and remove 3 screws 30.
8. Remove the central part of the housing from the rear of the device.
9. Only prisma VENT50: Remove pressure drop valve 66.
10. Only prisma VENT50: Remove pressure build-up valve (see "7.6.1
Removing the pressure build-up valve", page 47).
11. Remove blower (see "7.11.1 Remove blower", page 58).
30
12. Remove connection line of battery main board 14 from the old
central part of housing.

7.10.2 Fit central part of housing


1. Insert the connecting cable for battery for main board 14.
2. Fit blower (see "7.11.2 Fit blower", page 59).
3. Only prisma VENT50: Install pressure build-up valve (see "7.6.2
Installing the pressure build-up valve (only prisma VENT50)", page
48).
4. Only prisma VENT50: Insert pressure drop valve 66.

WM 68181b 07/2017

56 | EN
7 Repair

5. Place central part of housing on the rear of the device and fix in
31 position on rear with 3 screws 31.
6. Fix central part of housing in position at the front with 3 screws 30.

30

7. If connecting cable for the battery main board 14 is not yet


attached with cable tie 61:
• Attach connecting cable for the battery main board 14 using
cable tie 61.
• Kink connecting cable for the battery main board 14 at the level
of the central screwed connection.

14
61
WM 68181b 07/2017

EN | 57
7 Repair

8. Insert humidifier connecting cable 17.


9. Insert flow measurement adapter 51. In the process, ensure that
29 the blue O-rings are correctly positioned on the adapter.

51 10. Fix flow measurement adapter 51 in position with screw 29.


11. Close therapy device (see "7.3 Close therapy device", page 39).

17

7.11 Replace blower


7.11.1 Remove blower
1. Open therapy device (see "7.2 Open therapy device", page 36).
2. Remove central part of housing (see "7.10.1 Remove central part
of housing", page 55).
3. Remove blower with membrane 39 from central part of housing 2.

39

53
WM 68181b 07/2017

58 | EN
7 Repair

4. Remove blower 53 from membrane 39.

53

39

5. Release decoupling tube 15 and spacer 54 from blower 53.

54

15

53

7.11.2 Fit blower


1. If necessary: Attach new decoupling tube 15 and new spacer 54 to
new blower 53.
54 Ensure correct positioning as you do so.

15

53
WM 68181b 07/2017

EN | 59
7 Repair

2. Insert blower 53 in membrane 39.


In doing so, please note: The markings on the membrane must line
up with the screws on the blower.

53

39

3. Insert blower with membrane in the central part of the housing.


4. Insert decoupling tube and cable duct in central part of housing.
In doing so, please note: The decoupling tube may not be kinked
or trapped.
5. Fit central part of housing (see "7.10.2 Fit central part of housing",
page 56).
6. Close therapy device (see "7.3 Close therapy device", page 39).

39

53

WM 68181b 07/2017

60 | EN
7 Repair

7.12 Replace power supply unit and its


connecting cable

Risk of injury from electric shock!


Voltage may still be present in the power supply unit after the power
supply has been disconnected.
 Do not touch intact power supply units for at least 60 seconds after
disconnecting the power supply.
 Do not touch faulty power supply units for at least 12 hours after
disconnecting the power supply.
 Wear suitable safety gear (e.g. insulating gloves).

7.12.1 Remove switched power supply unit


1. Open therapy device (see "7.2 Open therapy device", page 36).
2. Remove central part of housing (see "7.10.1 Remove central part
of housing", page 55).
3. Disconnect power input connector 10.

10

4. Undo screw 32 on switched power supply unit 26.


32
49 5. Remove clamping bar 49 and switched power supply unit 26.
6. Remove EMC protective board 27.

26
WM 68181b 07/2017

EN | 61
7 Repair

2
3 7. Undo and remove 2 screws 29 from power supply input
connector 10 on the rear of the housing 1.
1 8. Pull power supply input connector out of the rear of the housing 1.

29

10

7.12.2 Fit switched power supply unit

1. Insert power supply input connector 10 in the rear of the housing 1


and secure with the 2 screws 29.
1

29

10

2. Insert EMC protective board 27 and new switched power supply


unit 26.
3. Secure new switched power supply unit 26 with screw 32 and
32 clamping bar 49.
49

26
WM 68181b 07/2017

62 | EN
7 Repair

4. Connect power supply input connector 10.


5. Fit central part of housing with blower (see "7.10.2 Fit central part
of housing", page 56).
10 6. Close therapy device (see "7.3 Close therapy device", page 39).

7.13 Replace rear of housing

Starting from SN 10.000, for prisma VENT30/prisma VENT40


the housing rear wall of prisma VENT50 is used. The spare part
assembly includes a plug and a baffle. When used for prisma
VENT30/prisma VENT40, the rear opening in the housing rear
wall must be closed with the plug, when used for
prisma VENT50, the silencer must be used at this point.
prisma VENT30/prisma VENT40 devices with SN < 10.000
receive through the exchange of the housing rear wall a
connection for the direct O2 introduction. In this case, the
existing patient's instructions for use must be replaced with
current instructions for use.

7.13.1 Remove rear of housing


1. Open therapy device (see "7.2 Open therapy device", page 36).
2. Remove central part of housing (see "7.10.1 Remove central part
of housing", page 55).
3. Remove power supply unit (see "7.12.1 Remove switched power
supply unit", page 61).
4. Replace rear of housing.

7.13.2 Fit rear of housing


1. Fit power supply unit in new rear of housing (see "7.12.2 Fit
switched power supply unit", page 62).
2. Fit central part of housing with blower (see "7.10.2 Fit central part
of housing", page 56).
3. Close therapy device (see "7.3 Close therapy device", page 39).
WM 68181b 07/2017

EN | 63
8 Storage and disposal

8 Storage and disposal


8.1 Storage
8.1.1 General information
Store the device under the specified ambient conditions (see section
entitled “Technical data”).

8.1.2 Store therapy device


1. Switch off therapy device.
2. Disconnect the therapy device from the power supply.
3. Clean therapy device, components and accessories.
4. Store therapy device, components and accessories in a dry place.
Result Store the therapy device, the components and the accessories in a dry
place.
If the device has an internal battery which is always supposed to be
ready for use, leave the device connected to the power supply. This
ensures that the battery is always fully charged.
If the device is not connected to the power supply for an extended
period, the battery will discharge. We recommend checking charge
status regularly and recharging with the aid of the device (if necessary).

8.2 Disposal
Do not dispose of the product with domestic waste.
To dispose of properly, contact a licensed, certified
electronics scrap disposal merchant. This address is
available from your Environment Officer or from
your local authority.
The device packaging (cardboard and inserts) can
be disposed of in paper recycling facilities.

WM 68181b 07/2017

64 | EN
9 Replacement parts

9 Replacement parts
9.1 Replacement parts list
The item numbers in this section are identical to the item numbers in
the sections entitled “Hygiene treatment” and “Repair”.

POSITION DESIGNATION ITEM NUMBER


HOUSING PARTS
Set housing rear wall for
1 WM 15919
WM 110 TD/WM 120 TD
Central part of housing,
2 WM 30672
preassembled
Front wall of housing, WM 29332
preassembled for WM 110 TD
3
Front wall of housing, WM 30684
preassembled for WM 120 TD
Panel for WM 110 TD
with Weinmann logo WM 29381WM0
4
with Löwenstein-Medical logo WM 29381
Panel for WM 120 TD WM 30686
5 Handle, fitted WM 29355
6 Release catch WM 29313
7 Cover WM 29358
Cover with seal for battery pack
8 WM 29378
compartment
9 Strain relief for power supply cable WM 29322
CABLES / CONNECTIONS / TUBES
10 Power supply inlet connector WM 29388
11 Power cord WM 24177
12 Connecting cable for tube heating WM 29944
Connecting cable for switched
13 WM 29333
power supply unit for main board
WM 29351
Connecting cable for battery for
main board for WM 110 TD
14
Connecting cable for battery for WM 30639
main board for WM 120 TD

15 Decoupling tube WM 29646


16 Ribbon cable for display WM 29308
17 Connecting cable for humidifier WM 29307
18 Pressure measurement tube WM 29365
19 Tube, pressure side WM 29328
Connection line of main board valve
20 WM 30653
control board for WM 120 TD
Double patient circuit for
21 WM 30693
WM 120 TD
Tube with elbow connector for
22 WM 30698
WM 120 TD
23 Tube valve unit for WM 120 TD WM 30695
Connection line of pressure build-
WM 68181b 07/2017

24 up valve, valve control board for WM 30654


WM 120 TD

EN | 65
9 Replacement parts

POSITION DESIGNATION ITEM NUMBER


PCBS
Main board
prisma VENT30 WM 35118
25 prisma VENT30-C WM 35128
prisma VENT40 WM 35119
prisma VENT50 WM 35209
Switched power supply unit for WM 29352
WM 110 TD
26
Switched power supply unit for WM 30642
WM 120 TD
27 EMC protective board with spacer WM 29517
28 Valve control board WM 35140
SCREWS / NUTS / WASHERS
Screw EJOT DELTA PT 30x14
29 WM 50572
WN 5451
Screw EJOT DELTA PT 40x25
30 WM 50563
WN 5452
Screw EJOT DELTA PT 40x68
31 WM 50557
WN 5452
Screw EJOT DELTA PT 25x8
32 WM 50564
WN 5451
Screw EJOT DELTA PT 25x12
33 WM 50551
WN 5451
34 Washer ISO 7089-4-200HV-A2 WM 50242
ENCODER
35 Encoder WM 29334
36 Dial for encoder WM 29314
Bezel for encoder
prisma VENT30 WM 29341
37 prisma VENT30-C WM 29343
prisma VENT40 WM 29342
prisma VENT50 WM 30647
SEALS
38 Sealing ring for rear of housing WM 29658
39 Membrane WM 29644
40 Insulating element WM 29606
41 Seal for flow element WM 29677
42 Seal for flow sensor WM 29678
43 Seal for system interface WM 29656
44 Seal for SD card slot WM 29648
MISCELLANEOUS
Plug for screwed connection for
45 WM 29362
front wall of housing
46 Cover WM 29371
47 Spacer for insulating element WM 29613
48 Foam for main board WM 27418
Clamping bar for switched power
49 WM 29374
supply unit
50 Flow element WM 29676
51 Flow measurement adapter WM 29321
O-ring for flow measurement
52 WM 29367
adapter
WM 68181b 07/2017

66 | EN
9 Replacement parts

POSITION DESIGNATION ITEM NUMBER


Fan, complete for WM 110 TD WM 29375
53 Blower, complete with NTC for WM 30665
WM 120 TD
54 Spacer for blower WM 29376
55 Blower cap WM 29607
Tallow-drop screw M3x4 for blower
56 WM 53017
cap
Device outlet port for WM 110 TD WM 29329
57
Device outlet port for WM 120 TD WM 30657
58 SD card WM 29794
59 Air filter WM 29651
60 Pollen filter WM 29389
61 Cable tie WM 12479
62 Pressure build-up valve WM 30682
63 Sealing plug WM 29518
Holder for pressure build-up valve
64 WM 30662
for WM 120 TD
Attachment for pressure drop valve
65 WM 30663
for WM 120 TD
66 Pressure drop valve for WM 120 TD WM 30667
67 Baffle for WM 120 TD WM 30661

9.2 Service sets


POSI-
DESIGNATION ITEM NUMBER
TION
SET, SPARE RECHARGEABLE BATTERY WM 15876
SET, CLEANING FOR WM 110 TD WM 15913
SET, CLEANING FOR WM 120 TD WM 15916
Sealing ring for rear of housing, pressure
38 WM 29658
side
19 Tube, pressure side WM 29328
41 Seal for flow element WM 29677
50 Flow element WM 29676
39 Membrane WM 29644
40 Insulating element, membrane WM 29606
Spacer for insulating element,
47 WM 29613
membrane
15 Decoupling tube WM 29646
54 Spacer for blower WM 29376
7 Cover, white WM 29358
51 Flow measurement adapter WM 29321
O-ring for flow measurement adapter
52 WM 29367
(2 off)
Plug for cover, screwed connection for
45 WM 29362
front wall of housing
59 Air filter WM 29651
68 Label, reprocessing WM 75070
Set, insulating elements for rear of housing
69 Insulating element 1 WM 29661
WM 68181b 07/2017

70 Insulating element 2 WM 29984

EN | 67
9 Replacement parts

POSI-
TION DESIGNATION ITEM NUMBER
71 Insulating element 3 WM 29663
72 Insulating element 4 WM 29664
Set, insulating elements for central part of housing
73 Insulating element 6 WM 29666
74 Insulating element 7 WM 29667
75 Insulating element 8 WM 29668
76 Insulating element 9 WM 29669
SET FOR CHANGE OF PATIENT WM 100 TH WM 29974
77 Humidifier insert WM 29683
78 Lower part of humidifier WM 29682
79 O-ring 8-1.5 WM 1145/190
80 Label, set for change of patient WM 77161

WM 68181b 07/2017

68 | EN
10 Components and accessories

10 Components and accessories

DESIGNATION ITEM NUMBER


Breathing tube with 19 mm diameter WM 24445
Breathing tube with 19 mm diameter, can be
WM 24667
autoclaved
Breathing tube with 15 mm diameter WM 29988
Bacteria filter WM 27591
Set, 12 pollen filters WM 29652
prismaBAG advanced WM 29337
prismaCONNECT module, communication module WM 29670
prismaPSG, PSG module WM 29690
prismaAQUA WM 29490
SD card WM 29794
prismaTS, therapy software, complete with USB
WM 93335
data cable
prisma CHECK WM 29390
prisma2CLOUD WM 29620
WM 68181b 07/2017

EN | 69
11 Auxiliary equipment required

11 Auxiliary equipment required


11.1 Tools
• Torx screwdriver size T 20
• Torx screwdriver size T 10
• Torx screwdriver size T 9
• Torx screwdriver size T 8
• ESD workstation
• Socket wrench 11 mm

11.2 Test equipment


• Manual pressure gage
Measuring range 0 - 40 hPa
• Ambient pressure measuring device/barometer
• Prisma therapy software
• ESD workstation
• Pressure measurement adapter (WM 23456)
• Sound insulation (WM 23685)
• Service connector (WM 29917)

11.3 Disinfectants
• terralin® protect
• gigasept FF® (new)
• MIKROZID LIQUID
can be ordered from:
Schülke & Mayr GmbH
Robert-Koch-Str. 2
22851 Norderstedt, Germany
Tel.: +49 40 52 100-0
Fax: +49 40 52 100-318
Internet: www.schuelkemayr.de
WM 68181b 07/2017

70 | EN
WM 68181b 07/2017 EN

Löwenstein Medical
Technology GmbH + Co. KG
Kronsaalsweg 40
22525 Hamburg, Germany
T: +49 40 5 47 02 - 100
F: +49 40 5 47 02 - 476
www.loewensteinmedical.de

You might also like