General Electric/Electronic Diagrams List of Fuses Cables: Melody Technical Manual
General Electric/Electronic Diagrams List of Fuses Cables: Melody Technical Manual
CHAPTER 5
♦ GENERAL ELECTRIC/ELECTRONIC
DIAGRAMS
♦ LIST OF FUSES
♦ CABLES
MELODY
MELODY/B
92-082-5
CPU mother board
Unit ON/OFF drive
C-ARM up/down motor
Compression motor drive
Magnetic clutch
Collimator light
Rotation brake
Mo/Rh filter
Exposure counter
Calibration:
RV2 compression force
RV1 Inverter power supply control circuit
Replacement:
IC20: version for MELODY
Jumpers:
JP1 short=voltage safety bypass
JP2 short/open = toggle state of stand by lamp
JP3 = short
JP4 short= standard clutch voltage
JP4 open= max clutch voltage
Options:
Deata Plus (no differences)
01-170
CPU board
Calibration:
no calibration is requested
Replacement:
EEPROM contains specific firmware
2 k RAM contains panel settings and CLOCK
512 k RAM contains calibrations
02-172
Potter-bucky driver
Calibration:
No calibration requested
Replacement:
No adjustment
Jumpers:
JP1, 2, 3, 4, 5, 6 open
JP7 closed
92-083-6
Keyboard / control panel
Calibration:
RV1 LCD contrast
Replacement:
short S8 in solder-side to bypass digital inclination display
XTAL= 4.3008 MHz
92-073-4
Filament power supply
Calibration:
refer to procedure
RV1, RV2, RV3, RV4, RV9 (factory adjusted)
Replacement:
anode current calibration
filament safety circuit calibration:
voltmetric RV3, RV5, JP2
amperometric RV6, RV7, JP1
Configuration:
92-073-3 for single anode tubes
JP3 = closed
92-074-3
Rotating anode starter
Calibration:
refer to procedure
Replacement:
safety circuit calibration (RV1, RV2)
voltage selection (220V - 230V - 240V)
Warning:
Circuit ground is floating with respect to other blocks in the
unit having the same symbol.
Be careful making measurements on this PCB, use
only insulated instruments.
92-090-2
IR remote control transmitter
Calibration:
no calibration is requested
Replacement:
no adjustment
92-091-2
IR remote control receiver
Calibration:
no calibration is requested
Replacement:
no adjustment
96-153-0
Mirror motor driver
Calibration:
no calibration is requested
Replacement:
no adjustment
MELODY-MELODY/B FUSES
Numeration is referred to general schematics
F1 = mains power
20A 10×38 WEBER CH10 CEI 32-1-4 IEC239
F2 = mains breaker
0.1A delayed 6.2×30
F3 = c-arm motor3.15 delayed 6.2x30
F4 = toroidal transformer primary TF205 – 3.15A delayed 6.2 X 30
F5 = Power on circuit on PCB 92-082 40mA delayed 5×20; F8 is in series to F10 on
PCB 92-082.
F6 = toroidal transformer secondary TF203 – 10A 10X38 weber CH10 CEI
32-1-4 IEC 239
F7 = toroidal transformer secondary TF203 – 10A 10X38 weber CH10 CEI
32-1-4 IEC 239
F8 = External lamp
500mA delayed 6x30
TF 203 Protected by termoswitch that acts on mains breaker turning off the device
and avoiding self power-on at the termoswitch restoration.
Further protection against transient overload by fuses on secondary
circuits.
CABLES
PCB 01-170
UART2 4800 BAUD
CN15