01.54.107033-1.6 SE-12 Series Electrocardiograph User Manual - EDAN 2 PDF
01.54.107033-1.6 SE-12 Series Electrocardiograph User Manual - EDAN 2 PDF
P/N: 01.54.107033-16
Statement
This manual will help you understand the operation and maintenance of the product better. It is
reminded that the product shall be used strictly complying with this manual. User’s operation
failing to comply with this manual may result in malfunction or accident for which the
manufacturer can not be held liable.
The manufacturer owns the copyrights of this manual. Without prior written consent of the
manufacturer, any materials contained in this manual shall not be photocopied, reproduced or
translated into other languages.
Materials protected by the copyright law, including but not limited to confidential information
such as technical information and patent information are contained in this manual, the user shall
not disclose such information to any irrelevant third party.
The user shall understand that nothing in this manual grants him, expressly or implicitly, any
right or license to use any of the intellectual properties of the manufacturer.
The manufacturer holds the rights to modify, update, and ultimately explain this manual.
Product Information
Product Name: Electrocardiograph
The electrical installation of the relevant room complies with national standards, and
I
The instrument is used in accordance with the instructions for use.
Upon request, the manufacturer may provide, with compensation, necessary circuit diagrams, and
other information to help qualified technician to maintain and repair some parts, which the
manufacturer may define as user serviceable.
WARNING
A WARNING label advises against certain actions or situations that could result in personal
injury or death.
CAUTION
A CAUTION label advises against actions or situations that could damage equipment, produce
inaccurate data, or invalidate a procedure.
NOTE
II
Table of Contents
Chapter 1 Safety Guidance ........................................................................................................... 1
1.1 Intended Use ........................................................................................................................... 1
1.2 Warnings and Cautions ........................................................................................................... 1
1.2.1 Safety Warnings ............................................................................................................... 2
1.2.2 Li-ion Battery Care Warnings .......................................................................................... 5
1.2.3 General Cautions .............................................................................................................. 6
1.2.4 Preparation and Operation Warnings (for SE-12 Express Exercise ECG) ....................... 7
1.2.5 Contraindications (for SE-12 Express Exercise ECG)..................................................... 8
1.3 List of Symbols ....................................................................................................................... 9
III
3.4 Inspection Before Power-On ................................................................................................ 37
3.5 Turning On/Off the Electrocardiograph................................................................................ 38
3.6 Loading/Replacing Recorder Paper ...................................................................................... 39
IV
6.5.2 Rhythm ECG Report ...................................................................................................... 73
6.5.3 Manual ECG Report ....................................................................................................... 74
6.5.4 R-R Analysis ECG Report ............................................................................................. 75
6.5.5 VCG Report ................................................................................................................... 78
6.5.6 Auto ECG Report Printed by the USB Printer ............................................................... 82
6.5.7 Auto ECG Report of Patient with Pacemaker ................................................................ 83
6.5.8 Nehb Lead ECG Report ................................................................................................. 84
V
10.2 Work Mode Setup ............................................................................................................. 103
10.3 Filter Setup ....................................................................................................................... 104
10.4 Record Info Setup ............................................................................................................. 105
10.4.1 Setup 1........................................................................................................................ 105
10.4.2 Setup 2........................................................................................................................ 109
10.4.3 Setup 3........................................................................................................................ 110
10.5 Patient Information Setup ................................................................................................. 111
10.6 Transmission Setup ........................................................................................................... 113
10.7 Lead Setup ........................................................................................................................ 114
10.8 Display&Sound Setup ...................................................................................................... 115
10.9 Date&Time Setup ............................................................................................................. 116
10.10 File Setup ........................................................................................................................ 117
10.11 System Maintenance Setup ............................................................................................. 119
10.12 Other Setup ..................................................................................................................... 120
Chapter 11 Operation Instructions for Exercise ECG (Configurable for SE-12 Express).. 122
11.1 About the Main Screen of Exercise ECG ......................................................................... 122
11.2 Operation Preparations ..................................................................................................... 127
11.3 Exercise Test ..................................................................................................................... 131
11.4 Factory Defaults of Exercise ECG.................................................................................... 132
11.5 System Setup for Exercise ECG ....................................................................................... 134
11.5.1 General Information Setup ......................................................................................... 135
11.5.2 Record Information Setup .......................................................................................... 137
11.5.3 Managing Protocols.................................................................................................... 139
11.6 Exercise ECG Reports ...................................................................................................... 142
11.6.1 12-Lead Report ........................................................................................................... 142
11.6.2 12-Lead Peak Report .................................................................................................. 143
11.6.3 Summary Report ........................................................................................................ 144
11.6.4 ST Scope Report ......................................................................................................... 145
11.6.5 Average Template Report ........................................................................................... 146
11.6.6 Summary Average Template Report ........................................................................... 147
11.6.7 Trend Graph Report .................................................................................................... 148
VI
Chapter 14 Cleaning, Care and Maintenance ......................................................................... 155
14.1 Cleaning ............................................................................................................................ 155
14.1.1 Cleaning the Main Unit and the Patient Cable ........................................................... 155
14.1.2 Cleaning the Reusable Electrodes (for Resting ECG) ............................................... 155
14.1.3 Cleaning the Print Head ............................................................................................. 156
14.1.4 Cleaning the Treadmill ............................................................................................... 156
14.2 Disinfection ...................................................................................................................... 156
14.3 Care and Maintenance ...................................................................................................... 157
14.3.1 Recharge and Replacement of Battery ....................................................................... 157
14.3.2 Recorder Paper ........................................................................................................... 158
14.3.3 Visual inspection ........................................................................................................ 159
14.3.4 Maintenance of the Main Unit and the Patient Cable ................................................ 159
VII
SE-12 Series Electrocardiograph User Manual Safety Guidance
This chapter provides important safety information related to the use of SE-12 series
electrocardiograph.
The intended use of SE-12 series electrocardiograph is to acquire ECG signals from adult and
pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended
to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The
cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart
disease. However, the interpreted ECG with measurements and interpretive statements is offered
to clinicians on an advisory basis only.
WARNING
1. This equipment is not designed for intracardiac use or direct cardiac application.
4. This equipment is intended for use on adult and pediatric patients only.
5. The results given by the equipment should be examined based on the overall clinical
condition of the patient, and they can not substitute for regular checking.
In order to use the electrocardiograph safely and effectively, and avoid possible dangers caused
by improper operation, please read through the user manual and be sure to be familiar with all
functions of the equipment and proper operation procedures before use.
Please pay more attention to the following warning and caution information.
-1-
SE-12 Series Electrocardiograph User Manual Safety Guidance
WARNING
2. Only qualified service engineers can install this equipment, and only service
engineers authorized by the manufacturer can open the shell.
4. SHOCK HAZARD - The power receptacle must be a hospital grade grounded outlet.
Never try to adapt the three-prong plug to fit a two-slot outlet.
5. Make sure that the power is turned off and the power cord is disconnected from the
AC socket before connecting or disconnecting equipment. Otherwise, electrical
shock or other injuries may happen to the patient or operator.
6. If the integrity of the external protective conductor is in doubt, the equipment should
be powered by an internal li-ion rechargeable battery.
7. Do not use this equipment in the presence of high static electricity or high voltage
equipment which may generate sparks.
8. Only the patient cable and other accessories supplied by the manufacturer can be
used. Or else, the performance and electric shock protection can not be guaranteed.
The electrocardiograph has been safely tested with the recommended accessories,
peripherals, and leads, and no hazard is found when the electrocardiograph is
operated with cardiac pacemakers or other stimulators.
9. Make sure that all electrodes are connected to the patient correctly before operation.
10. Ensure that the conductive parts of electrodes and associated connectors, including
neutral electrodes, do not come in contact with earth or any other conducting objects.
11. If reusable electrodes with electrode gel are used during defibrillation, the
electrocardiograph recovery will take more than 10 seconds. The manufacturer
recommends the use of disposable electrodes at all times.
-2-
SE-12 Series Electrocardiograph User Manual Safety Guidance
WARNING
12. Electrodes of dissimilar metals should not be used; otherwise it may cause a high
polarization voltage.
13. The disposable electrodes can only be used for one time.
14. Do not touch the patient, bed, table or the equipment while using the ECG together
with a defibrillator.
15. Do not touch accessible parts of non-medical electrical equipment and the patient
simultaneously.
16. The use of equipment that applies high frequency voltages to the patient (including
electrosurgical equipment and some respiration transducers) is not supported and
may produce undesired results. Disconnect the patient data cable from the
electrocardiograph, or detach the leads from the patient prior to performing any
procedure that uses high frequency surgical equipment.
17. If the wireless AP technology is used, in order to maintain compliance with the FCC
RF exposure guidelines, the wireless AP should be installed and operated with a
minimum distance of 20cm between the radiator and the human body. There should
be no shield in or around the room where the wireless AP is used.
18. Fix attention on the examination to avoid missing important ECG waves.
19. SHOCK HAZARD - Don’t connect non-medical electrical equipment, which has been
supplied as a part of the system, directly to the wall outlet when the non-medical
equipment is intended to be supplied by a multiple portable socket-outlet with an
isolation transformer.
20. SHOCK HAZARD - Don’t connect electrical equipment, which has not been supplied
as a part of the system, to the multiple portable socket-outlet supplying the system.
21. Do not connect any equipment or accessories that are not approved by the
manufacturer or that are not IEC/EN 60601-1-1 approved to the electrocardiograph.
The operation or use of non-approved equipment or accessories with the
electrocardiograph is not tested or supported, and electrocardiograph operation and
safety are not guaranteed.
22. Any non-medical equipment (such as the external printer) is not allowed to be used
within the patient vicinity (1.5m/6ft.).
-3-
SE-12 Series Electrocardiograph User Manual Safety Guidance
WARNING
23. Do not exceed the maximum permitted load when using the multiple portable
socket-outlet(s) to supply the system.
25. Do not use the additional multiple portable socket-outlet or extension cord in the
medical electrical system, unless it’s specified as part of the system by the
manufacturer. And the multiple portable socket-outlets provided with the system shall
only be used for supplying power to equipment which is intended to form part of the
system.
26. Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data
processing equipment and IEC/EN 60601-1 for medical equipment). Furthermore all
configurations shall comply with the valid version of the standard IEC/EN 60601-1-1.
Therefore anybody, who connects additional equipment to the signal input or output
connector to configure a medical system, must make sure that it complies with the
requirements of the valid version of the system standard IEC/EN 60601-1-1. If in
doubt, consult our technical service department or your local distributor.
27. Connecting any accessory (such as external printer) or other device (such as the
computer) to this electrocardiograph makes a medical system. In that case,
additional safety measures should be taken during installation of the system, and the
system shall provide:
b) Outside the patient environment, the level of safety appropriate for non-medical
electrical equipment complying with other IEC or ISO safety standards.
28. All the accessories connected to system must be installed outside the patient vicinity,
if they do not meet the requirements of IEC/EN 60601-1.
29. You should purchase computer, printer, treadmill, ergometer and BP monitor from the
manufacturer. Otherwise, the manufacturer will not be held responsible for the
maintenance of the PC hardware, operating system and other accessories.
30. If multiple instruments are connected to a patient, the sum of the leakage currents
may exceed the limits given in the IEC/EN 60601-1 and may pose a safety hazard.
Consult your service personnel.
-4-
SE-12 Series Electrocardiograph User Manual Safety Guidance
WARNING
31. The potential equalization bar can be connected to that of other equipment when
necessary. Make sure that all the equipment is connected to the potential
equalization terminal.
WARNING
1. Improper operation may cause the internal li-ion battery (hereinafter called battery) to
be hot, ignited or exploded, and it may lead to the decrease of the battery capacity. It
is necessary to read the user manual carefully and pay more attention to warning
messages.
2. Only qualified service engineers authorized by the manufacturer can open the battery
compartment and replace the battery, and batteries of the same model and
specification as manufacturer configuration should be used.
3. DANGER OF EXPLOSION -- Do not reverse the anode and the cathode when
installing the battery.
5. Do not destroy the battery; Do not pierce battery with a sharp object such as a needle;
Do not hit with a hammer, step on or throw or drop to cause strong shock; Do not
disassemble or modify the battery.
6. When leakage or foul smell is found, stop using the battery immediately. If your skin
or cloth comes into contact with the leakage liquid, cleanse it with clean water at once.
If the leakage liquid splashes into your eyes, do not wipe them. Irrigate them with
clean water first and go to see a doctor immediately.
8. Only when the device is off can the battery be installed or removed.
9. Remove the battery from the electrocardiograph when the electrocardiograph isn’t
used for a long time.
10. If the battery is stored alone and not used for a long time, we recommend that the
battery be charged at least once every 6 months to prevent over-discharge.
-5-
SE-12 Series Electrocardiograph User Manual Safety Guidance
CAUTION
1. Avoid liquid splash and excessive temperature. The temperature must be kept
between 5 ºC and 40 ºC during operation, and it should be kept between -20 ºC and
55 ºC during transportation and storage.
2. Do not use the equipment in a dusty environment with bad ventilation or in the
presence of corrosive.
3. Make sure that there is no intense electromagnetic interference source around the
equipment, such as radio transmitters or mobile phones etc. Attention: large medical
electrical equipment such as electrosurgical equipment, radiological equipment and
magnetic resonance imaging equipment etc. is likely to bring electromagnetic
interference.
4. Ruptured fuse must only be replaced with that of the same type and rating as the
original.
5. The device and accessories are to be disposed of according to local regulations after
their useful lives. Alternatively, they can be returned to the dealer or the manufacturer
for recycling or proper disposal. Batteries are hazardous waste. Do NOT dispose of
them together with house-hold garbage. At the end of their lives hand the batteries
over to the applicable collection points for the recycling of waste batteries. For more
detailed information about recycling of this product or battery, please contact your
local Civic Office, or the shop where you purchased the product.
6. Federal (U.S.) law restricts this device to sale by or on the order of a physician.
-6-
SE-12 Series Electrocardiograph User Manual Safety Guidance
Exercise ECG)
WARNING
1. Test the safety stop (mushroom type) and safety stop (cord type) of the treadmill
before using the system.
2. During the exercise test, ensure that there are at least 2 experienced physicians
present. One of them observes the patient and deals with the emergency.
3. Make sure that there is necessary valid first-aid equipment such as defibrillators,
blood-pressure meters etc, and necessary valid medication in the exercise test room.
4. Turn off the system power and disconnect the power cord from the wall outlet after
using the system.
5. Make sure that the power is turned off and the power cord is disconnected from the
AC socket before defibrillation.
6. Keep the four feet of the machine on the ground and make sure that it’s stably
working.
9. The patient undergoing the exercise test should wear suitable clothes and shoes.
10. Keep hands, hair, jewelry, and loose clothing away from moving parts.
11. Don’t let the patient stand on the running belt when starting the treadmill. The patient
should stand on the foot rails and hold the handrails during start-up. Wait until the
running belt is moving before placing feet on the belt.
12. To avoid the static electricity, the patient should not wear loose clothing or clothing
(such as nylon) that easily produces static electricity.
13. Stop exercising immediately when the patient feels uncomfortable or something
abnormal in the operation.
14. Press down the safety stop (mushroom type) or pull out the safety stop (cord type) to
stop the treadmill immediately when an emergency happens.
-7-
SE-12 Series Electrocardiograph User Manual Safety Guidance
Absolute Contraindications:
Relative Contraindications:
-8-
SE-12 Series Electrocardiograph User Manual Safety Guidance
External output
External input
Caution
Potential equalization
USB socket
Net port
Mains supply
Battery indicator
Delete key
Enter key
-9-
SE-12 Series Electrocardiograph User Manual Safety Guidance
Shift key
Fn key
1mV/COPY key
MODE key
SLEEP/WAKE UP key
PRINT/STOP key
Gender key
- 10 -
SE-12 Series Electrocardiograph User Manual Safety Guidance
Recycle
Part Number
Serial Number
Date of Manufacture
Manufacturer
- 11 -
SE-12 Series Electrocardiograph User Manual Introduction
Chapter 2 Introduction
SE-12 series electrocardiograph gathers ECG signals of 12 leads simultaneously. It displays the
operation menu, ECG parameters as well as electrocardiograms.
The 12-channel ECG waves can be viewed on the LCD screen and printed out by using a
high-quality thermal recorder. The sampled ECG data can be saved, transmitted and exported.
The manual, auto, rhythm, R-R analysis or VCG (only configurable for SE-12 Express) mode can
be chosen freely.
For SE-12 Express, the exercise ECG function is optional. When a patient with coronary heart
disease runs, the added heart load will cause myocardium hypotension, and then the ECG will
change abnormally. Therefore, the exercise ECG function of SE-12 Express can be used to
diagnose concealed coronary heart disease and atypical angina pectoris, prescribe the workload
for patients with myocardial infarction before they leave hospital, and assess the effect of the
treatment.
SE-12 series electrocardiograph can be powered by the mains supply or battery.
With a high resolution thermal recorder, a 32-bit processor and a large-capacity memorizer,
SE-12 series electrocardiograph has advanced performance and high reliability. The compact size
makes it suitable for clinic and hospital uses.
SE-12 series electrocardiograph has two models: SE-12 and SE-12 Express.
SE-12 adopts 320×240 dot single color LCD screen; SE-12 Express adopts 800×600 multicolor
LCD screen.
Configuration: main unit, power cord, patient cable, chest electrodes, limb electrodes, disposable
electrodes, alligator clips, thermal recorder paper, fuses, battery.
NOTE: The pictures and windows in this manual are for reference only.
- 12 -
SE-12 Series Electrocardiograph User Manual Introduction
Indicators
LCD Screen
Recorder Keyboard
Indicators
LCD Screen
Recorder
Keyboard
- 13 -
SE-12 Series Electrocardiograph User Manual Introduction
C
D
E
F
Q
P
G
H
O
N M L K J I
C
D
E
F
Q
G
P
H
O
N M L K J I
Figure 2-4 SE-12 Keyboard
- 14 -
SE-12 Series Electrocardiograph User Manual Introduction
Name Explanation
When the main screen is displayed, press this key to display
the main screen of the exercise test.
When the main screen of the exercise test is displayed, press
this key to enter the pretest phase. The length of the pretest
Pretest
phase is not fixed.
When the main screen of the exercise test is displayed and
the exercise test is terminated, press this key to enter the no
testing state.
Press to display the Comment dialog box, and then you can
Comment annotate in the dialog box. The annotations will be displayed
in the summary report.
- 15 -
SE-12 Series Electrocardiograph User Manual Introduction
- 16 -
SE-12 Series Electrocardiograph User Manual Introduction
- 17 -
SE-12 Series Electrocardiograph User Manual Introduction
A B C
A B C
- 18 -
SE-12 Series Electrocardiograph User Manual Introduction
B C D
B C D
Name Explanation
- 19 -
SE-12 Series Electrocardiograph User Manual Introduction
C D
A B E F
C D
G
A B E F
Name Explanation
Connecting to a PC
In the exercise test, only the treadmill/ergometer
B Serial Port 1
recommended by the manufacturer can be
connected.
- 20 -
SE-12 Series Electrocardiograph User Manual Introduction
: Caution
Definitions of corresponding pins:
NOTE: The left side of “/” is European standard, and the right side is American standard.
2) Serial Port 1
WARNING
The isolated intensity of serial port 1 is 1500V AC and the maximum voltage applied
should not exceed +15V DC.
- 21 -
SE-12 Series Electrocardiograph User Manual Introduction
3) Serial Port 2
6
8 2
7 1
5 3
4
Definitions of corresponding pins:
- 22 -
SE-12 Series Electrocardiograph User Manual Introduction
CAUTION
Only the USB equipment recommended by the manufacturer can be connected to the
USB interface.
WARNING
1. Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data
processing equipment and IEC/EN 60601-1 for medical equipment). Furthermore all
configurations shall comply with the valid version of the standard IEC/EN 60601-1-1.
Therefore anybody, who connects additional equipment to the signal input or output
connector to configure a medical system, must make sure that it complies with the
requirements of the valid version of the system standard IEC/EN 60601-1-1. If in
doubt, consult our technical service department or your local distributor.
2. If multiple instruments are connected to a patient, the sum of the leakage currents
may exceed the limits given in the IEC/EN 60601-1 and may pose a safety hazard.
Consult your service personnel.
- 23 -
SE-12 Series Electrocardiograph User Manual Introduction
D
C D
D
C D
- 24 -
SE-12 Series Electrocardiograph User Manual Introduction
Name Explanation
1) Battery Compartment
The rated voltage and the rated capacity of the battery pack are as follows:
WARNING
1. Improper operation may cause the battery to be hot, ignited or exploded, and it may
lead to the decrease of the battery capacity. Therefore, it is necessary to read the
user manual carefully and pay more attention to warning messages.
2. When leakage or foul smell is found, stop using the battery immediately. If your skin
or cloth comes into contact with the leakage liquid, cleanse it with clean water at once.
If the leakage liquid splashes into your eyes, do not wipe them. Irrigate them with
clean water first and go to see a doctor immediately.
3. Only qualified service engineers authorized by the manufacturer can open the battery
compartment and replace the battery, and batteries of the same model and
specification must be used.
4. Only when the device is off can the battery be installed or removed.
NOTE: If the battery has not been used for two months or more, you should recharge it
before using it again.
2) Fuse
There are two fuses of the same specification installed on the bottom of the main unit. The
specification is: T1AL250VP Ø5×20.
WARNING
Ruptured fuses must only be replaced with those of the same type and rating as the
original.
- 25 -
SE-12 Series Electrocardiograph User Manual Introduction
2.7 Features
♦ Supporting AC and DC power supply modes, internal rechargeable li-ion battery with
professional battery powered circuit, battery management and protection systems
♦ Supporting multi-language
♦ Full alphanumeric keyboard (For SE-12 Express, touch screen is available)
♦ ECG signals of 12 leads are gathered and amplified simultaneously, 12-channel waves
are displayed and recorded simultaneously
♦ Correct detection for failure electrodes
♦ Convenient operation of recording by pressing the PRINT/STOP key with high
efficiency
♦ High resolution thermal recorder, recording frequency response ≤150Hz
♦ Supporting external USB printer
♦ Supporting accurate digital filter to decrease the polarization voltage and other
interferences
♦ Supporting rolled and folded paper recorded with high resolution waveforms,
calibration mark, gain, speed and filter
♦ The auto, manual, rhythm, R-R analysis and VCG (only configurable for SE-12 Express)
modes can be chosen freely
♦ Flexible printing formats
♦ Supporting ECG waves displaying with grid
♦ Automatic baseline adjustment for optimal printing
♦ Convenient operation of system setup and file management
♦ Multiple file formats: DAT, PDF and optional formats (SCP/FDA-XML/DICOM)
♦ Measurement function and interpretation function
♦ Supporting bar code reader
♦ ECG data can be transmitted to the PC software through the serial cable, net cable, or
wireless AP (optional)
♦ Real-time transmission to PC ECG
♦ Supporting order function
- 26 -
SE-12 Series Electrocardiograph User Manual Introduction
The following features are only for the exercise test function of SE-12 Express
(optional)
♦ Real-time display and print of 12-lead simultaneous ECG waveforms with average
template
♦ Providing average templates of three rhythm leads in every stage to observe the change
of ST segments between every two stages
♦ Providing saving PDF files to U disk and transmitting PDF files to the server by
implanting FTP protocol.
♦ Providing classical exercise protocols; exercise protocols can be edited and created
♦ Automatically controlling and adjusting the speed and grade of the treadmill or the
power of the ergometer
- 27 -
SE-12 Series Electrocardiograph User Manual Operation Preparations
Before use, the equipment, patient cable and electrodes should be checked. Replace
them if there is any evident defectiveness or aging which may impair the safety or the
performance, and make sure that the equipment is in proper working condition.
The performance and electric shock protection can be guaranteed only if the original
patient cable and electrodes of the manufacturer are used.
The patient cable includes the main cable and lead wires which can be connected to electrodes.
Connecting to the
Lead Wires
Patient Cable Socket
Main Cable
Connecting to
Electrodes
Main Cable
Lead Wires
Screw
Electrode Connector
Connecting to the
Patient Cable Socket
- 28 -
SE-12 Series Electrocardiograph User Manual Operation Preparations
Connect the patient cable to the patient cable socket on the right side of the main unit, and then
secure them with two screws.
ECG)
Align all lead wires of the patient cable to avoid twisting, and connect the lead wires to the
reusable electrodes or the alligator clips. Firmly attach them.
The identifiers and color codes of electrode connectors used comply with IEC/EN requirements.
In order to avoid incorrect connection, the identifiers and color codes are specified in Table 3-1.
Moreover the equivalent codes according to American requirements are given in Table 3-1 too.
Table 3-1 Electrode Connectors and Their Identifiers and Color Codes
European American
Electrode Connectors Identifier Color Code Identifier Color Code
Right arm/Right deltoid R Red RA White
Left arm/Left deltoid L Yellow LA Black
Right leg/Upper leg as
N or RF Black RL Green
close to torso as possible
- 29 -
SE-12 Series Electrocardiograph User Manual Operation Preparations
ECG)
The identifiers and color codes of electrodes used comply with IEC/EN requirements. In order to
avoid incorrect connections, the electrode identifiers and color codes are specified in Table 3-2.
Moreover the equivalent codes according to American requirements are given in Table 3-2 too.
European American
Electrodes Color Code Electrodes Color Code
R Red RA White
L Yellow LA Black
N or RF Black RL Green
F Green LL Red
C1 White/Red V1 Brown/Red
C2 White/Yellow V2 Brown/Yellow
C3 White/Green V3 Brown/Green
C4 White/Brown V4 Brown/Blue
C5 White/Black V5 Brown/Orange
C6 White/Violet V6 Brown/Violet
Before attaching the electrodes, greet the patient and explain the procedure. Explaining the
procedure decreases the patient’s anxiety. Reassure the patient that the procedure is painless.
Privacy is important for relaxation. When possible, prepare the patient in a quiet room or area
where others can’t see the patient. Make sure that the patient is comfortable. The more relaxed
the patient is, the less the ECG will be affected by noise.
- 30 -
SE-12 Series Electrocardiograph User Manual Operation Preparations
Thorough skin preparation is very important. The skin is a poor conductor of electricity and
frequently creates artifacts that distort the ECG signals. By performing methodical skin
preparation, you can greatly reduce the possibility of noise caused by muscle tremor and baseline
drift, ensuring high-quality ECG waves. There is natural resistance on the skin surface due to dry,
dead epidermal cells, oils and dirt.
To Prepare the Skin
Shave hair from electrode sites, if necessary. Excessive hair prevents a good connection.
Wash the area thoroughly with soap and water.
Dry the skin with a gauze pad to increase capillary blood flow to the tissues and to remove the
dead, dry skin cells and oils.
Two kinds of electrode can be used, one is the reusable electrode (including chest electrodes and
limb electrodes), and the other is the disposable electrode.
WARNING
1. Make sure that all electrodes are connected to the patient correctly before operation.
2. Ensure that the conductive parts of electrodes and associated connectors, including
neutral electrodes, do not come in contact with earth or any other conducting objects.
NOTE: The quality of ECG waveform will be affected by the contact resistance between
the patient and the electrode. In order to get a high-quality ECG, the
skin-electrode resistance must be minimized while connecting electrodes.
- 31 -
SE-12 Series Electrocardiograph User Manual Operation Preparations
The electrodes’ positions on the body surface are shown in the following table and figure.
European American
Electrode Placement
Label Label
- 32 -
SE-12 Series Electrocardiograph User Manual Operation Preparations
NEHB Placement
European American
Electrode Placement
Label Label
Nst A1 Attachment point of the second rib to the right sternal edge
R RA Right arm
L LA Left arm
N or RF RL Right leg
F LL Left leg
- 33 -
SE-12 Series Electrocardiograph User Manual Operation Preparations
European American
Electrode Placement
Label Label
C1 V1 Right mid-axillary line on the same horizontal level as C3 and C4
C2 V2 Sternum at the level of C3 and C4
C3 V3 Mid-clavicular line in the fifth intercostals space
C4 V4 Left mid-axilary line on the same horizontal level as C3
C5 V5 Center of spine on the same horizontal level as C3 and C4
C6 V6 Neck, avoid carotid artery and jugular vein
L LA Right arm/Right deltoid
R RA Left arm/Left deltoid
F LL Right leg/Upper leg as close to torso as possible
N RL Left leg/Upper leg as close to torso as possible
- 34 -
SE-12 Series Electrocardiograph User Manual Operation Preparations
European American
Electrode Placement
Label Label
Reed
Clamp
Limb Electrode
Suction Bulb
Chest Electrode
- 35 -
SE-12 Series Electrocardiograph User Manual Operation Preparations
CAUTION
The disposable electrodes can only be used for one time.
- 36 -
SE-12 Series Electrocardiograph User Manual Operation Preparations
In order to avoid safety hazards and get good ECG records, the following inspection procedures
are recommended before operation.
WARNING
1) Environment:
♦ Make sure that there is no electromagnetic interference source around the equipment,
especially large medical electrical equipment such as electrosurgical equipment,
radiological equipment, magnetic resonance imaging equipment etc. Turn off these
devices when necessary.
♦ Keep the examination room warm to avoid muscle tremor voltages in ECG signals
caused by cold.
2) Power Supply:
♦ If the mains supply is used, please check whether the power cord is connected to the
unit well. The grounded three-slot outlet should be used.
♦ When the battery capacity is low, recharge the battery before use.
3) Patient Cable:
♦ Make sure that the patient cable is connected to the unit firmly, and keep it far away
from the power cord.
4) Electrodes:
♦ Make sure that all electrodes are connected to lead wires of the patient cable
correctly.
♦ Ensure that the chest electrodes do not contact with each other.
5) Patient:
♦ The patient should not come into contact with conducting objects such as earth, metal
parts etc.
♦ Ensure that the patient is warm and relaxed, and breathes calmly.
- 37 -
SE-12 Series Electrocardiograph User Manual Operation Preparations
WARNING
1. If the integrity of the external protective conductor is in doubt, the equipment should
be powered by the battery.
The electrocardiograph can be powered by either the mains supply or the battery.
Specifications) before power-on, and then press on the keyboard to turn on the
unit. The mains supply indicator ( ) is lit, and the logo will be displayed on the LCD
screen after self-test.
If the battery is weak when the mains supply is used, it will be recharged automatically
at the same time. Both the mains supply indicator ( ) and the battery recharging
indicator ( ) will be lit.
♦ When operating on battery power
Press on the keyboard to turn on the unit, and then the battery indicator ( ) will
be lit and the battery symbol will be displayed. The logo will be displayed on the LCD
screen after self-test.
Because of the consumption during the storage and transport course, the battery capacity
may not be full. If the symbol and the hint information Battery Weak are displayed,
which means the battery capacity is low, please recharge the battery first.
CAUTION
2. The electrocardiograph cannot print an ECG report when the battery is weak.
- 38 -
SE-12 Series Electrocardiograph User Manual Operation Preparations
Hold down the key to display the hint System is shutting down… on the screen.
Then the device will be off a few seconds later. Remove the plug from the outlet.
♦ When operating on battery power
Hold down the key to display the hint System is shutting down… on the screen.
Then the device will be off a few seconds later.
NOTE:
1. When turning off the device, follow the above sequence strictly, or else there may be
something wrong on the screen.
2. Do not hold down the key when the device displays the hint information System
is shutting down… on the screen.
CAUTION
Make sure that the recorder paper, is installed in the center of the recorder, and the paper
edge is parallel with the casing edge in the direction of advancing paper, in order to avoid
paper deviation or damage to the paper edge.
- 39 -
SE-12 Series Electrocardiograph User Manual Operation Preparations
When the recorder paper runs out or is not loaded, the hint message No Paper will appear on the
screen. Then you should load or replace the recorder paper immediately.
Groove
Movable Movable
Part 1 Part 2
Paper Tray
Exit Edge
(Edge for
tearing paper)
Casing Button
Exit
Paper Roller
Casing
1) Press the casing button downwards and remove the casing to open the recorder.
- 40 -
SE-12 Series Electrocardiograph User Manual Operation Preparations
3) Take off the wrapper of the new folded paper, and then put it in the paper tray.
4) Pull the paper out with the grid side facing the thermal print head, and replace the casing
on the recorder.
to advance the recorder paper to the next black marker; if Paper Marker is
set to No, you can press to advance the paper for 2.5cm. Press
again to stop advancing the paper.
- 41 -
SE-12 Series Electrocardiograph User Manual Operation Preparations
2) Take out the paper roller, and remove remainder paper from the roller if necessary.
3) Take off the wrapper of the new thermal paper roll, and then put the paper roll through
the roller.
4) Place the paper and the roller gently in the recorder with the roller pin clicking into the
groove.
Groove
5) Pull the paper out with the grid side facing the thermal print head, and replace the casing
on the recorder.
- 42 -
SE-12 Series Electrocardiograph User Manual Operation Preparations
When the main screen is displayed, you can press to advance the paper for
- 43 -
SE-12 Series Electrocardiograph User Manual Basic Operation Guidance
The following sections provide an overview of the main operations and functions.
You can operate the electrocardiograph by using the touch screen (optional).
CAUTION
Do not touch the LCD screen with sharp things such as pencils or pens; otherwise, it will
be damaged.
Press F7 below to
open the main screen2.
Press F1, F2, F3, F4, F5 or F6 to select the corresponding menu function.
To select Patient, press the function key F1 below Patient on the main screen1.
To select File, press the function key F1 below File on the main screen2.
For details about the main screen, please refer to Section 4.3.1 “About the Main Screen”.
- 44 -
SE-12 Series Electrocardiograph User Manual Basic Operation Guidance
1. Press F1 below Patient on the main screen1 to open the Patient Information window.
2. Press Tab or Shift + Tab to move the cursor to the Name textbox.
To input patient name, press the letter or numeric keys on the keyboard.
To input the special character in the top right corner of the key, press Fn and a letter key.
For example, press Fn + a to input è.
To input the special character in the top right corner of the key, press Shift and a numeric
key.
For example, press Shift + 3 to input #.
When Caps Lock is set to Off in the Other Setup window, pressing Shift and a letter
key can input a capital letter.
For example, pressing Shift + P can type a capital P.
When Caps Lock is set to On in the Other Setup window, pressing Shift and a letter
key can input a lowercase letter.
For example, pressing Shift + P can type a lowercase p.
3. Press Bksp on the keyboard to erase the typed information.
4. Press Enter to confirm, or press Tab or Shift + Tab to move the cursor to the OK button, and
then press Enter to confirm.
5. Press Esc to cancel the operation, or press Tab or Shift + Tab to move the cursor to the
Cancel button, and then press Enter to cancel the operation.
6. Press Shift+Bksp to refresh all patient information except for the Gender, Age Group,
Exam.Room, Physician and Technician information after you print an ECG report.
- 45 -
SE-12 Series Electrocardiograph User Manual Basic Operation Guidance
Check Box
Textbox
Dropdown
List Box
1. In the Work Mode Setup window, press Tab or Shift + Tab to move the cursor among
different check boxes. Press spacebar to select a check box, and a check mark √ appears in the
box.
2. In the Work Mode Setup window, press Tab or Shift + Tab to move the cursor to a list box
or dropdown. Press the Up or Down arrow to highlight an option.
3. In the Work Mode Setup window, press Tab or Shift + Tab to move the cursor to a textbox.
Enter data in the selected textbox.
4. Press Enter to confirm, or press Tab or Shift + Tab to move the cursor to the OK button, and
then press Enter to confirm.
5. Press Esc to cancel the operation, or press Tab or Shift + Tab to move the cursor to the
Cancel button, and then press Enter to cancel the operation.
For details on configuring the system settings and the order settings, please refer to Chapter 10
“System Setup”, Chapter 11.5 “System Setup for Exercise ECG” (Only Configurable for SE-12
Express) and Section 8.5 “Setting Orders”.
- 46 -
SE-12 Series Electrocardiograph User Manual Basic Operation Guidance
C D E F G H I
B
J
A
Q P O N M L K
B C D E F G H I
A J
Q P O N M L K
Figure 4-2 SE-12 Main Screen1
- 47 -
SE-12 Series Electrocardiograph User Manual Basic Operation Guidance
U T S
Figure 4-3 SE-12 Express Main Screen2
U T S
- 48 -
SE-12 Series Electrocardiograph User Manual Basic Operation Guidance
Name Explanation
Patient Name: within 60 ASCII characters or the equivalent
A Name number of other characters that can be supported by equivalent
memory used by 60 ASCII characters
When ID Mode is set to Manual, the patient ID is within 30
ASCII characters.
When ID Mode is set to Auto, ID can be automatically generated
after you press Shift+Bksp to refresh the patient information. The
patient ID is 0~1999, 999, 999.
B ID
When ID Mode is set to Time, press Shift+Bksp to refresh the
patient information, and then the patient ID can be automatically
generated according to the time when you press the
PRINT/STOP key to print an ECG report. Entering the patient
ID manually is not supported.
Patient Age
C Age The patient age value and the age unit can be set in the Patient
Information window.
D Gender Patient Gender (Male/Female/Cleared)
Including DEMO, Module Error, Overload, Lead Name (When
the leads are off, the lead names will be shown in black on a
E Hint Information 1 yellow background in SE-12 Express, while in SE-12, they will
be shown in black on a white background.) For details, please
refer to Chapter 12 “Hint Information”.
Including No Paper, Paper Error, Battery Weak, Sampling,
Analyzing, Recording, Testing, Learning, Transmitting, Loading
F Hint Information 2 Orders, Detecting, Memory Full, U Disk, USB Printer, Lead
Off, USB Scanner. For details, please refer to Chapter 12 “Hint
Information”.
Manual, Auto, Rhythm, R-R Analysis and VCG (only
configurable for SE-12 Express)
If Print Out is set to Off in the Record Info Setup1 window,
G Work Mode
appears in the auto or rhythm mode. For details, please
refer to Section 10.4.1 “Setup1”.
H Heart Rate Actual Heart Rate
Current system time. For details, please refer to Section 10.9
I Current Time
“Date&Time Setup”.
J Battery Symbol Identify the current capacity of the battery
- 49 -
SE-12 Series Electrocardiograph User Manual Basic Operation Guidance
- 50 -
SE-12 Series Electrocardiograph User Manual Basic Operation Guidance
Select Setup on the main screen1 to display the System Setup screen.
NOTE: If you set the system password in the System Maintenance window, you need
enter the password before opening the System Setup screen. For details, refer
to Section 10.11, “System Maintenance Setup”.
On the System Setup screen, move the cursor on an item, and then press Enter to open the setup
window of the item.
- 51 -
SE-12 Series Electrocardiograph User Manual Basic Operation Guidance
Select Order on the main screen2 to open the Order Manager screen.
You can press Load to load orders and select a loaded order to start an order examination.
Or, you can press Esc to return to main screen1.
I H G F E D C
Name Explanation
For example, 2/200
200 is the total number of orders that can be stored in the
A Order Count electrocardiograph.
2 is the current number of orders stored in the
electrocardiograph.
Orders will be loaded and displayed in the order list.
The order information includes ID, Name, Request No., Order
B Order List
Date, Exam.Room and State.
State includes: √ indicates an order with examination
- 52 -
SE-12 Series Electrocardiograph User Manual Basic Operation Guidance
D Del All Press to delete all the orders from the electrocardiograph.
Once the electrocardiograph is turned on, you can open the Order Manager screen by pressing
Patient if orders without examination exist on the Order Manager screen.
After you press Esc on the Order Manager screen to return to main screen1, you cannot open
the Order Manager screen by pressing Patient until you complete an order examination next
time.
To open the File Manager screen1, select File on the main screen2.
To open the File Manager screen2, select a file on the File Manager screen1, and then
press Select.
To return to the previous screen, select Return.
- 53 -
SE-12 Series Electrocardiograph User Manual Basic Operation Guidance
I H G F E D C
P O N M L K J
- 54 -
SE-12 Series Electrocardiograph User Manual Basic Operation Guidance
Name Explanation
For example, 1/200
200 is the total number of files that can be stored in the
A File Count electrocardiograph.
1 is the current number of files stored in the
electrocardiograph.
Files will be loaded and displayed in the file list.
The file information includes ID, Name, Time, Mode and State.
State includes: No mark
T indicates the file is transferred successfully.
B File List
E indicates the file is exported successfully.
C indicates the file has been edited.
NOTE: A file can be displayed in more than one state at
the same time.
C Return Press to return to the main screen.
D Import Press to import files from the U disk to the electrocardiograph.
E Search Press to open the SearchInfo Setup window.
Press to highlight a file on the File Manager screen1, and then
F Select press Select to select the file and display the File Manager
screen2.
G Del All Press to delete all the files from the electrocardiograph.
Press to export all the files from the electrocardiograph to the
H Export All
U disk.
I Trans All Press to transmit all the files to the PC.
J Return Press to return to the File Manager screen1.
K Preview Press to open the file preview screen.
- 55 -
SE-12 Series Electrocardiograph User Manual Basic Operation Guidance
VCG In the VCG mode, X, Y, Z waves and frontal, horizontal, sagittal planes can
(only configurable be displayed. 10s sampled ECG data can be analyzed, and vector waves,
for SE-12 Express) vector loops, measure information, diagnosis information can be printed.
For details on printing ECG reports in the Auto, Manual, Rhythm, R-R or VCG (only
configurable for SE-12 Express) mode, please refer to Section 6.1 “Printing an ECG Report”.
- 56 -
SE-12 Series Electrocardiograph User Manual Entering Patient Information
2. Select Patient on the main screen1 to open the Patient Information window.
- 57 -
SE-12 Series Electrocardiograph User Manual Entering Patient Information
Pacemaker Select Yes to detect very small pacemaker pulses. However, when
Pacemaker is set to Yes, the system is very sensitive, and should not be
close to equipment emitting high frequency radiation. High frequency
radiation can interfere with pacemaker pulse detection and normal ECG
acquisition.
NOTE: Pacemaker is recommended to be set to No unless it is
known that the majority of the electrocardiograph usage
will be on patients with pacemakers.
NOTE: The total number of supported characters may be fewer if special Latin characters
are entered.
NOTE: To use the order function, the Smart ECG Viewer software of the manufacturer
must be installed in the PC.
Operation procedures are as follows:
1. Connect the electrocardiograph to the PC with Ethernet cable recommended by the
manufacturer.
- 58 -
SE-12 Series Electrocardiograph User Manual Entering Patient Information
1) Select Setup-> Patient Info to open the Patient Information Setup window.
2) Select On from the Order Acquired list box, and then press Enter to confirm.
5. Select Patient on the main screen1 to open the Patient Information window.
Enter the patient ID manually in the ID textbox or connect a bar code reader, press Order, and
then the matched order will be loaded from Smart ECG Viewer software and the order
information will be displayed in the corresponding textboxes.
- 59 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
NOTE:
1. The working mode can not be changed during the printing course. Stop printing
reports before changing the working mode.
2. Within three seconds after returning to the main screen, if you press the PRINT/STOP
key to print an ECG report in the auto quick mode or the manual mode, the recorder
will not respond.
3. If Print Out is set to Off in the Record Info Setup1 window, the ECG report can be
saved and transmitted, but cannot be printed out by pressing the PRINT/STOP key in
the auto and rhythm modes.
4. When the main screen is displayed, pressing the MODE key can select a working mode.
The auto mode is the most common electrocardiograph usage and is applied to normal ECG test.
10s ECG data can be sampled, analyzed, and printed by pressing the PRINT/STOP key.
Operation Method:
1. Set Mode Options, Auto DispStyle, Preview, Sampling Mode, Record Style, Rhythm
Lead1/2/3, or Lead Sequence (Optional)
1) Select AUTO from the Mode Option list in the Work Mode Setup window.
2) Select a style from the Auto DispStyle list in the Work Mode Setup window.
3) Select a mode from the Sampling Mode list in the Work Mode Setup window.
4) Select On or Off from the Preview list box in the Work Mode Setup window.
5) Select a style from the Record Style list in the Record Info Setup window.
6) Select a lead from the Rhythm Lead1/2/3 list in the Lead Setup window.
7) Select a sequence from the Lead Sequence list in the Lead Setup window.
8) Press Enter to confirm.
- 60 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
- 61 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
The rhythm mode contributes to discover the abiogenetic or frequent arrhythmia. The
rhythm-lead ECG waveform of one lead or three leads can be sampled or printed for a long time.
Operation Method:
1. Set Mode Options, Rhythm Style, Rhythm Lead1/2/3 or Lead Sequence(Optional)
1) Select RHYT from the Mode Option list in the Work Mode Setup window.
2) Select a style from the Rhythm Style list in the Work Mode Setup window.
3) Select a lead from the Rhythm Lead1/2/3 list in the Lead Setup window.
4) Select a sequence from the Lead Sequence list in the Lead Setup window.
5) Press Enter to confirm.
2. Press the PRINT/STOP key to begin sampling, the sampling time will be displayed on the
main screen. When the sampling time reaches 60s in the Single Lead style or 20s in the
Three Leads style, it begins to print an ECG report.
- 62 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
The R-R analysis mode contributes to get the related indicators of HRV (Heart Rate Variability).
HRV analysis refers to the variability between durations of the two following heart rates, and it is
a predicted indicator to sudden death caused by acute myocardial infarction and vicious
arrhythmia. For healthy persons, RR Histogram presents open and multi-peak; for patient with
the low variability caused by cardiovascular diseases (such as myocardial infarction or heart
failure), RR Histogram presents a single peak.
Operation Method:
1. To set Mode Options or Rhythm Lead1
1) Select R-R from the Mode Option list in the Work Mode Setup window.
2) Select a lead from the Rhythm Lead1 list in the Lead Setup window.
3) Press Enter to confirm.
2. Press the PRINT/STOP key to begin sampling, the sampling time will be displayed on the
main screen. When the sampling time reaches 180s, it begins to analyze and print an ECG
report.
3. It will stop automatically after a complete R-R analysis report is printed. During the printing
course, pressing the PRINT/STOP key again can stop printing the ECG report.
NOTE: In the R-R analysis mode, you can not set the speed. The constant speed is
25mm/s and the printing speed is 5mm/s, because in the R-R analysis mode, the
ECG wave length is compressed to one fifth of the original wave length.
- 63 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
VCG (Vectorcardiogram) is stereo images to show phase shifts of heart activities. As an advance
and a complement to the cardiogram diagnosis, VCG excels in diagnosis of myocardial infarction,
intraventricular conduction block and ventricular hypertrophy.
NOTE:
1. Auto diagnosis of VCG reports are for reference only. Except the diagnosis to
myocardial infarction, atrial hypertrophy, ventricular hypertrophy, pre-excitation
syndrome, bundle branch block, myocardial injury and coronary heart disease,
cardiograms are diagnosed as normal VCG.
2. VCG reports cannot be saved to the electrocardiograph or be transmitted.
Operation Method:
1. Prepare for the operation.
For details, please refer to Chapter 3 “Operation Preparations”.
2. Set Mode Options, Preview, VCG Record Info, or QRS Gain (Optional)
1) Select VCG from the Mode Option list in the Work Mode Setup window.
2) Select On or Off from the Preview list box in the Work Mode Setup window.
3) Select VCG record information to be printed out in the Record Info Setup3 window.
4) Select a value from the QRS Gain list box in the Record Info Setup3 window.
NOTE: Pressing F2 twice in the Record Info Setup window can display the
Setup3 window.
5) Press Enter to confirm.
3. If Preview is set to On in the Work Mode Setup window, pressing the PRINT/STOP key
can preview an ECG report.
- 64 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
- 65 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
3) Select Measure to display the measure screen which shows the measurement
information.
In the auto, rhythm and VCG (only configurable for SE-12 Express) modes, pressing the
1mV/COPY key can print the ECG report which was printed out last time. Pressing the
PRINT/STOP key can stop printing the ECG report.
You can freeze the ECG waves displayed on the main screen.
Operation Method:
1) Press F3 to set the paper speed, press F4 to set the gain, and press F5 to set the filter on the
main screen1.
2) Select Freeze to display the freezing screen.
- 66 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
Figure 6-2 VCG Freezing Screen (only configurable for SE-12 Express)
NOTE: Within 10 seconds after returning to the main screen, pressing F2 can not
display the freezing screen.
3) Press PRINT/STOP or select Record on the freezing screen to print the current waveform.
- 67 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
Operation Method:
1. Select a file on the File Manager screen1.
2. Press PRINT/STOP to print the file, or press Select to display the File Manager screen2,
and then press PRINT/STOP to print the file.
3. Or, press PRINT/STOP again to stop printing the file.
Operation Method:
1. Select a file on the File Manager screen1, and then press Select to display the File Manager
screen2.
2. Select Preview on the File Manager screen2 to open the preview screen.
3. Press PRINT/STOP or select Record on the preview screen to print the selected file.
4. Or, press PRINT/STOP again to stop printing the file.
NOTE: Rhythm data can not be previewed.
- 68 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
Quick Mode
(a)
- 69 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
(b)
The above figure (a) and (b) show an ECG report in the auto mode. Template is selected, and
Record Style is set to 6×2+1R.
The ECG report includes:
- 70 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
- 71 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
(a)
(b)
- 72 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
The above figure (a) and (b) show an ECG report in the auto mode. Template is selected, and
Record Style is set to 6×2+1R.
The above figure shows an ECG report in the rhythm mode, and Rhythm Style is set to Three
Leads.
- 73 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
The above figure shows an ECG report in the manual mode, and Manual Style is set to 12
channels.
- 74 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
(a)
- 75 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
(b)
- 76 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
The above figure (a) and (b) show an ECG report in the R-R analysis mode.
Figure (a) shows:
Current Date & Current Time
Patient Information
Measuring Time
Gain, Speed, Filter
II (Lead name)
180s compressed ECG wave of Lead II
80 (Heart Rate)
RR Avg Interval (Average RR interval)
RR Max Interval (Maximum RR interval)
RR Min Interval (Minimum RR interval)
Max/Min (Ratio of Maximum RR interval to Minimum RR interval)
SDNN (Standard Deviation of Normal to Normal Intervals)
RMSSD (Root Mean Square Successive Difference)
RR Histogram
RR Trend
Figure (b) shows:
All the RR interval values within the measuring time
- 77 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
(a)
- 78 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
(b)
(c)
- 79 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
The above figure (a), (b) and (c) show an ECG report in the VCG mode. XYZ Wave, Measure
and Analysis are selected, and QRS Gain is set to 20mm/mV.
Item Definition
CW Clockwise
CCW Counter-clockwise
- 80 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
- 81 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
- 82 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
The above figure shows the ECG report of a patient with a pacemaker.
- 83 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports
- 84 -
SE-12 Series Electrocardiograph User Manual Transmitting ECG Data
CAUTION
It is forbidden to connect or disconnect a U disk or a USB printer during the transmission
course.
1. If the PC has RS232 socket, connect the PC RS232 socket to the electrocardiograph RS232
socket with an RS232 cable.
2. If the PC has no RS232 socket, you can connect the PC USB socket to the electrocardiograph
RS232 socket by RS232-USB Assembly.
- 85 -
SE-12 Series Electrocardiograph User Manual Transmitting ECG Data
Cable
- 86 -
SE-12 Series Electrocardiograph User Manual Transmitting ECG Data
- 87 -
SE-12 Series Electrocardiograph User Manual Transmitting ECG Data
In the real-time transmission, the electrocardiograph functions as an ECG sampling box to the PC
ECG software.
- 88 -
SE-12 Series Electrocardiograph User Manual Managing Orders
NOTE: To use the order function, you must install the Smart ECG Viewer software of the
manufacturer.
Before loading orders, please configure the Order Setup screen. For details, please refer to
Section 8.5 “Setting Orders”.
Operation procedures are as follows:
1. Connect the electrocardiograph to the PC installed with the Smart ECG Viewer software by
using an Ethernet cable recommended by the manufacturer.
2. Set Remote IP, Local IP, Gateway and Subnet Mask in the Transmission Setup window.
For details, please refer to Section 10.6 “Transmission Setup”.
3. Select Load on the Order Manager screen to load orders from the Smart ECG Viewer
software, and then a hint will be displayed as follows.
NOTE: If orders are modified on the Smart ECG Viewer software, the corresponding
orders existing on the Order Manager screen will be refreshed after you load
orders from the software.
4. If you select Exam.Room Filter on the Order Setup screen, orders will be filtered after you
press Load.
For details, please refer to Section 8.5 “Setting Orders”.
5. If 200 orders already exist on the Order Manager screen, the following hint pops up after
you press Load.
- 89 -
SE-12 Series Electrocardiograph User Manual Managing Orders
Select an order on the Order Manager screen, and then select Examine or press Enter to return
to main screen1 for starting an examination.
NOTE: If you select Delete After Examination on the Order Setup screen, the order will
be deleted from the Order Manager screen after you examine the selected order.
Otherwise, the order will be marked by √ mark on the Order Manager screen
after you examine the selected order.
Pressing Del All on the Order Manager screen can delete all the orders from the
electrocardiograph.
Or, you can select an order on the Order Manager screen, select Delete to delete the selected
order from the electrocardiograph.
Select Search on the Order Manager screen to display the following window.
Select the search type, such as ID, Name, Request No., Exam. Room, Time, enter the search
information, and then press Enter to confirm. All the orders which meet the requirements will be
searched and displayed on the Order Manager screen.
NOTE: The time mode in the SearchInfo Setup window is the mode you select in the
Date & Time Setup window.
- 90 -
SE-12 Series Electrocardiograph User Manual Managing Orders
Items Default
Condition Default
Sequence Ascending
Item Description
Condition Choose from: Default, ID, Order Date, Request No. or State
Select Default, orders will be displayed in sequence of the time when the
orders are loaded from the Smart ECG Viewer software.
Select ID, Order Date, Request No. or State, orders will be displayed in
sequence of the selected condition on the Order Manager screen.
- 91 -
SE-12 Series Electrocardiograph User Manual Managing Orders
Item Description
Delete After Select this item, the order will be deleted from the Order Manager
Examination screen after the order is examined.
Exam.Room Filter Select this item, a check mark √ appears in the box before Exam.Room
Filter.
You can enter an exact exam.room in the textbox, such as
Electrocardiograph. All the orders which meet the requirements will be
searched and displayed on the Order Manager screen.
Deselect this item or enter nothing in the textbox, all the orders will be
searched and displayed on the Order Manager screen.
- 92 -
SE-12 Series Electrocardiograph User Manual Managing Files
If you want to save the ECG data in the electrocardiograph, you should set Auto Save to to ECG
in the File Setup window. Then the ECG data in the auto or rhythm mode will be saved on the
File Manager screen automatically. For details, please refer to Section 10.10 “File Setup”.
On the File Manager screen, files can be printed, transmitted, exported, displayed, edited
searched or deleted. SE-12 series electrocardiograph can accommodate 200 files.
CAUTION
1. When files are being printed, transmitted, deleted or exported, you can not turn off
the electrocardiograph.
2. Do not cut off the mains supply directly when no battery is installed in the device, or
else, the stored data may be lost.
Make configuration in accordance with Section 7.1 “Transmitting ECG Data” before you transmit
files.
Pressing Trans All on the File Manager screen1 can transmit all the files from the
electrocardiograph.
Or, select Trans on the File Manager screen2, and then press Enter to transmit the selected file
from the electrocardiograph.
NOTE: If you select Delete After Trans. Or Export in the File Setup window, the files
will be deleted from the File Manager screen after they are transferred.
CAUTION
It is forbidden to connect or disconnect a U disk or a USB printer during the transmission
course.
- 93 -
SE-12 Series Electrocardiograph User Manual Managing Files
4. Or select a file on the File Manager screen1, and then press Select to display the File
Manager screen2. Select Export on the File Manager screen2 to export the selected file to
the directory of ECGDATA\ECG-X\Export\Export Date and Time of the U disk.
NOTE:
1. Please insert the U disk recommended by the manufacturer. Please set the format to
FAT or FAT32 when formatting the U disk.
2. X in the directory of ECGDATA\ECG-X\Export\Export Date and Time can be set in
the Device No. textbox in the Transmission Setup window.
3. If you select Delete After Trans. Or Export in the File Setup window, the files will be
deleted from the File Manager screen after they are exported.
Pressing Del All on the File Manager screen1 can delete all the files from the electrocardiograph.
Or, select Delete on the File Manager screen2 to delete the selected file from the
electrocardiograph.
Select Search on the File Manager screen1 to display the following window.
Select the search type, such as ID, Name, Time, enter the search information, and then press
Enter to confirm. All the files which meet the requirements will be searched and displayed on the
File Manager screen.
NOTE: The time mode in the SearchInfo Setup window is the mode you select in the
Date & Time Setup window.
- 94 -
SE-12 Series Electrocardiograph User Manual Managing Files
Press Select on the File Manager screen1 to display the File Manager screen2, and then select
Edit to open the Patient Information window.
NOTE: The Address item can be defined in the User-defined text box in the Patient
Information Setup window. For details, please refer to Section 10.5 “Patient
Information Setup”.
For details on inputting data, please refer to Section 4.1.2, “Entering Data”.
You can press PRINT/STOP on the File Manager screen1/2 or the preview screen to print ECG
reports. For details, please refer to Section 6.4 “Printing a Stored ECG Report”.
1. Select Preview on the File Manager screen2 to open the file preview screen.
2. Select Waveform to display the waveform screen.
- 95 -
SE-12 Series Electrocardiograph User Manual Managing Files
You can press the Left or Right arrow key to switch between the lead groups when
previewing a file on the waveform screen.
3. Select Template to display the template screen.
- 96 -
SE-12 Series Electrocardiograph User Manual Managing Files
4. Select Measure to display the measure screen which shows the measurement information.
- 97 -
SE-12 Series Electrocardiograph User Manual System Setup
Select Setup on the main screen1 to display the System Setup screen.
Default
Items Default
(Only in the U.S.)
Duration
60 min 60 min
(Periodic Sample)
Interval
1 min 1 min
(Periodic Sample)
Filter Setup
Default
Items Default
(Only in the U.S.)
AC filter On On
- 98 -
SE-12 Series Electrocardiograph User Manual System Setup
Default
Items Default
(Only in the U.S.)
Print Out On On
Default
Items Default
(Only in the U.S.)
Measure / Analysis / Time
On On
Scale
Template / Position Marker
/ Minnesota Code / Device Off Off
No.
- 99 -
SE-12 Series Electrocardiograph User Manual System Setup
Default
Items Default
(Only in the U.S.)
Default
Items Default
(Only in the U.S.)
Gender / Pacemaker On On
ID Hint On On
Patient Information
On On
Refreshed
- 100 -
SE-12 Series Electrocardiograph User Manual System Setup
Transmission Setup
Default
Items Default
(Only in the U.S.)
Lead Setup
Default
Items Default
(Only in the U.S.)
Rhythm Lead 1 II II
Rhythm Lead 2 V1 V1
Rhythm Lead 3 V5 V5
Default
Items Default
(Only in the U.S.)
Brightness 10 10
- 101 -
SE-12 Series Electrocardiograph User Manual System Setup
Grid On On
Default
Items Default
(Only in the U.S.)
File Setup
Default
Items Default
(Only in the U.S.)
Maintenance Setup
Default
Items Default
(Only in the U.S.)
Other Setup
Default
Items Default
(Only in the U.S.)
- 102 -
SE-12 Series Electrocardiograph User Manual System Setup
Item Description
Mode Options Select a working mode.
Choose from: MANU, AUTO, RHYT, R-R or VCG (only configurable
for SE-12 Express)
Auto DispStyle Select a style to display the ECG waves of 12 leads in the auto mode.
Choose from: 3×4, 3×4+1R, 3×4+3R, 6×2, 6×2+1R or 12×1
Manual Style Select a style to display and print the ECG waves in the manual mode.
Choose from: 3 channels, 6 channels or 12 channels
Rhythm Style Select a style to print the ECG waves of the appointed rhythm lead.
Choose from: Single Lead or Three Leads
Sampling Mode Choose from: Pre-Sample, Real-time Sample, Triggered Sample or
(Only available in Periodic Sample
the auto mode) Select Pre-Sample, 10s ECG data sampled before pressing the
PRINT/STOP key will be printed out.
NOTE: When Sampling Mode is set to Pre-Sample, if you press
the PRINT/STOP key before the electrocardiograph
samples for 10s, the recorder will not respond.
- 103 -
SE-12 Series Electrocardiograph User Manual System Setup
Item Description
Sampling Mode Select Real-time Sample, 10s ECG data sampled after pressing the
(Only available in PRINT/STOP key will be printed out.
the auto mode) Select Triggered Sample, after pressing the PRINT/STOP key, if
Arrhythmia ECG data, including Asystole, Ventricular
Fibrillation/Ventricular Tachycardia, 5>PVCS>=3, Paired PVCS,
Bigeminy, Trigeminy, R ON T, single PVC and Missed Beat, is detected
during the learning course, the printing will be triggered automatically.
Select Periodic Sample, first you should set Duration and Interval.
NOTE: ID and patient information will not be changed while carrying
out the periodic printing.
Duration In the auto mode, when Sampling Mode is set to Periodic Sample, if
& Interval Interval is set to 2 min, Duration is set to 24 min, after pressing the
PRINT/STOP key, the printing will be performed every two minutes and
come to 12 times.
Preview Enable or disable the preview function in the auto or VCG (only
configurable for SE-12 Express) mode.
Choose from: On or Off
Auto Arrhythmia Choose from: On or Off
Detection Select On, if arrhythmia is detected in the auto mode, a hint will pop up to
ask you whether to print an extra rhythm report after the 12-lead ECG
report.
- 104 -
SE-12 Series Electrocardiograph User Manual System Setup
Item Description
DFT Filter DFT Filter greatly reduces the baseline fluctuations without affecting the
ECG signals. The purpose of this filter is to keep the ECG signals on the
baseline of the printout.
Choose from: 0.05Hz, 0.15Hz, 0.25Hz, 0.32Hz, 0.5Hz or 0.67Hz
The set value is the low limit of the frequency range.
EMG Filter EMG Filter suppresses disturbance caused by strong muscle tremor.
The cutoff frequency can be set to 25Hz, 35Hz or 45Hz.
Select Record Info on the System Setup screen, and then press Enter to open the Record Info
Setup window.
10.4.1 Setup 1
- 105 -
SE-12 Series Electrocardiograph User Manual System Setup
Item Description
Record Style Select a style to print the ECG waves of 12 leads in the auto mode.
Choose from: 3×4, 3×4+1R, 3×4+3R, 6×2, 6×2+1R or 12×1
Record Mode Choose from: Save Paper or Quickly
Select Save Paper, 10s after pressing the PRINT/STOP key on the main
screen, an ECG report is printed in the auto mode. The patient
information, measure information, interpretation and ECG waves are
printed at the same paper.
Select Quickly, pressing the PRINT/STOP key on the main screen to
begin printing an ECG report immediately in the auto mode, the patient
information, measure information, interpretation and ECG waves are
printed on the different pieces of paper.
NOTE:
1. Only when Sampling Mode is set to Real-time Sample,
Quickly is available.
2. When Record Style is set to 3×4, 3×4+1R or 3×4+3R, only
Save Paper is available. When Record Style is set to 12×1,
only Quickly is available
Record Sequence Choose from: Sequential or Synchronous
Select Sequential, the lead group is printed one by one in a certain
sequence. The start time of a lead group is just the end time of the
previous lead group.
- 106 -
SE-12 Series Electrocardiograph User Manual System Setup
Item Description
Record Sequence Select Synchronous, the lead group is printed one by one in a certain
sequence. All leads are printed with the same start time.
Gain You can set the indicated height of 1mV ECG on the paper.
Choose from: 2.5mm/mV, 5mm/mV, 10mm/mV, 20mm/mV or
10/5mm/mV.
10/5mm/mV means that the gain of limb leads is set to 10mm/mV, while
the gain of chest leads is set to 5mm/mV.
WARNING
If the printer used is not the type listed above, additional safety measures (such as
applying an isolation transformer to supply the medical system) should be taken when the
safety of the medical system has not been evaluated. If in doubt, consult our technical
service department or your local distributor.
CAUTION
It is forbidden to connect or disconnect a U disk or a USB printer during the transmission
course.
HP HP Deskjet 1050
2010/1050/2000 HP Deskjet 2000
HP Deskjet 2050
HP Laserjet P2015
HP2035
HP Laserjet P2035
- 107 -
SE-12 Series Electrocardiograph User Manual System Setup
Item Description
Record Device NOTE:
1. During the USB printing course,pressing the PRINT/STOP key
again can not stop printing ECG reports.
2. For details of the ECG report printed by the USB printer, please
refer to 6.5.6, “ECG Reports Printed by the USB Printer”.
3. USB printing is ineffective in the auto periodic sampling mode,
auto 11~24s sampling mode, manual mode and R-R analysis
mode.
4. Make sure that paper is installed in the USB printer before
printing. Error may occur if no paper is loaded in the USB
Printer.
Print Out Choose from: On or Off
(Only available Select On, the ECG report can be printed out by pressing the
in the auto and PRINT/STOP key.
rhythm modes) Select Off, the ECG report can be saved, but cannot be printed out by
pressing the PRINT/STOP key.
Speed Choose from: 5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s or
50mm/s
In the manual mode, select 5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s,
25mm/s or 50mm/s.
Only 25mm/s and 50mm/s are available in the auto, rhythm and VCG
(only configurable for SE-12 Express) modes.
Only 25mm/s is available in the R-R analysis mode.
Paper Marker Paper Marker is used to identify the start point of each page of the
recorder paper.
Choose from: Yes or No
Select Yes if the paper with black markers on the bottom is used, and the
device can identify the start point of each page of the recorder paper while
printing ECG reports.
Sample Time If Record Style is set to 12×1, you can set the time period.
If the time period is set from 11 to 24 seconds, in the auto mode, after
pressing the PRINT/STOP key, the ECG waves of the set time period are
printed.
NOTE: If the time period is set from 11 to 24 seconds, the ECG data
sampled will not be analyzed or stored.
- 108 -
SE-12 Series Electrocardiograph User Manual System Setup
10.4.2 Setup 2
Item Description
Auto Record Info Select the item printed in the ECG reports.
NOTE:
1. Template, Position Marker, Measure, Analysis and
Minnesota Code are available only in the auto mode, and Time
Scale and Device No. are available in all the work modes.
2. To get more information about the above contents, please refer
to Section 6.5.1, “Auto ECG Reports”.
Baseline Choose from: Horizontal, Auto or Off
Adjustment Select Horizontal, the baselines of the lead groups are adjusted
(not available in simultaneously, and the baselines of the leads in the same row are on the
the VCG mode) same line.
Select Auto, the baselines of the lead groups are adjusted respectively.
Select Off, the baselines of the lead groups are adjusted equally in the
ECG reports.
RR Record Info Select RR Interval List, the item will be printed in the ECG reports.
Grid of Report Select Thermal Report or USB Report, the grid will be printed while
printing ECG reports with the thermal recorder or USB printer.
- 109 -
SE-12 Series Electrocardiograph User Manual System Setup
10.4.3 Setup 3
NOTE:
1. Only if the VCG function is activated, can the Record Info Setup3 window appear.
For details on activating the VCG function, please contact the manufacturer or the
local distributor.
2. Pressing F2 twice in the Record Info Setup window can display the Setup3 window.
Item Description
VCG Record Info Select the item printed in the ECG reports.
- 110 -
SE-12 Series Electrocardiograph User Manual System Setup
Item Description
Patient Options Select the item displayed in the Patient Information window.
NOTE:
1. Pacemaker appears in the Patient Information window after it
is selected in the Patient Information Setup window. Set
Pacemaker to Yes in the Patient Information window, and the
Pacemaker information will be displayed on the report printed
out.
2. Pacemaker is recommended to be set to No unless it is known
that the majority of the electrocardiograph usage will be on
patients with pacemakers.
ID Choose from: Auto, Time or Manual
Select Manual, the patient ID is within 30 ASCII characters.
Select Auto, ID can be automatically generated after you press
Shift+Bksp to refresh the patient information. The patient ID is 0~1999,
999, 999.
Select Time, press Shift+Bksp to refresh the patient information, and
then the patient ID can be automatically generated according to the time
when you press the PRINT/STOP key to print an ECG report. Entering
the patient ID manually is not supported.
- 111 -
SE-12 Series Electrocardiograph User Manual System Setup
Item Description
ID Hint Choose from: On or Off
In the auto, rhythm or VCG (only configurable for SE-12 Express) mode,
when ID is set to Manual and ID Hint is set to On, if you do not input
the patient ID before pressing the PRINT/STOP key, a hint will pop up
to remind you to input the patient ID.
Age Choose from: Age, D.O.B or Age Group
Select Age, you can enter the patient age manually in the Patient
Information window.
Select D.O.B, the D.O.B textbox appears and the Age textbox becomes
unavailable in the Patient Information window, you can enter the
birthday of the patient, and the system will calculate the patient age
automatically.
Select Age Group, the Age Group textbox appears in the Patient
Information window and the 0 key (or Age Group key) can be available.
For details, please refer to Section 2.2 “Keyboard and Keys”.
H/W Unit Choose from: cm/kg or inch/lb
BP Unit Choose from: mmHg or kPa
Prompt Choose from: Confirmed By or Unconfirmed
Select Confirmed By, the physician’s name is printed in the ECG reports
if it is input in the Patient Information window.
Select Unconfirmed, Unconfirmed Report is printed in the ECG reports.
- 112 -
SE-12 Series Electrocardiograph User Manual System Setup
Item Description
Device No. Type the Device No., within 7 ASCII characters
Auto Transmission Choose from: On or Off
Select On, ECG data will be transmitted automatically after an ECG
report is printed out in the auto or rhythm mode.
Transmission
Choose from: Ethernet or UART
Mode
FTP Information Enter data in the FTP Path, FTP User Name textboxes.
IP Addresses Set Server IP, Local IP, Gateway, Subnet Mask
For the cross-network transmission,
a) Set the first two sections of the Local IP item to the first two sections
of the IP of the PC.
b) Set the third section of the Local IP item to the network segment of
the electrocardiograph which depends on the configuration of Router.
c) The last section of the Local IP item can be set at random.
For the same network transmission,
a) Set the first three sections of the Local IP item to the first three
sections of the IP of the PC.
b) The last section of the Local IP item can be set at random, but it can’t
be the same as the last section of the IP of the PC.
- 113 -
SE-12 Series Electrocardiograph User Manual System Setup
Select Lead on the System Setup screen, and press Enter to open the Lead Setup window.
Item Description
Rhythm Lead1/2/3 Choose from: І, П, Ш, aVR, aVL, aVF, V1, V2, V3, V4, V5, or V6
In the auto mode:
When Record Style is set to 3×4+1R or 6×2+1R, the rhythm lead
selected in the Rhythm Lead1 list box will be printed in the ECG reports;
When Record Style is set to 3×4+3R, 3 rhythm leads selected
respectively in the Rhythm Lead1/2/3 list box will be printed in the ECG
reports.
In the rhythm mode:
When Rhythm Style is set to Single Lead, 60s wave of the rhythm lead
selected in the Rhythm Lead1 list box will be printed in the ECG reports;
When Rhythm Style is set to Three Leads, 20s waves of three rhythm
leads selected respectively in the Rhythm Lead1/2/3 list box will be
printed in the ECG reports.
In the R-R analysis mode:
The R-R analysis report of the rhythm lead selected in the Rhythm Lead1
list box will be printed.
- 114 -
SE-12 Series Electrocardiograph User Manual System Setup
Item Description
Lead Sequence Choose from: Standard or Cabrera
Lead
Lead group 1 Lead group 2 Lead group 3 Lead group 4
Sequence
Standard І, II, Ш aVR, aVL, aVF V1, V2, V3 V4, V5, V6
Item Description
Brightness Set the brightness within 0~20.
Key Volume Choose from: Low, Medium, High or Off
Select Low, Medium or High, the electrocardiograph gives a short sound
when you press keys on the keyboard.
Select Off, there is no sound.
- 115 -
SE-12 Series Electrocardiograph User Manual System Setup
Item Description
Select Date&Time on the System Setup screen, and press Enter to open the Date&Time Setup
window.
- 116 -
SE-12 Series Electrocardiograph User Manual System Setup
NOTE: Please set DATE&TIME correctly when it’s the first time you use the
electrocardiograph.
Item Description
Date&Time Enter the current date and time displayed on the main screen and in the
ECG reports.
Select File on the System Setup screen, and press Enter to open the File Setup window.
- 117 -
SE-12 Series Electrocardiograph User Manual System Setup
Item Description
- 118 -
SE-12 Series Electrocardiograph User Manual System Setup
Select Maintenance on the System Setup screen, and press Enter to open the System
Maintenance window.
Item Description
Import Settings Press to load the system settings from the U disk to the
electrocardiograph.
Back up Settings Press to back up the system settings to the ECG memory.
Load Backup
Press to load the backup settings from the ECG memory.
Settings
Load Factory
Press to restore the factory settings.
Settings
System Password Type a password that allows you to access the System Setup window.
- 119 -
SE-12 Series Electrocardiograph User Manual System Setup
Item Description
Language Select the language displayed on the main screen and in the ECG reports.
External Input The external input socket is equipped in the electrocardiograph, through
which the electrocardiograph can receive signals from the external
equipment.
Choose from: On or Off
Select On, the electrocardiograph will display the signals which it
receives from external input port.
- 120 -
SE-12 Series Electrocardiograph User Manual System Setup
Item Description
External Output The external output socket is equipped in the electrocardiograph, through
which the electrocardiograph can send rhythm lead signals to the external
equipment.
Choose from: Off, Standard or Triggered
Select Standard, the electrocardiograph sends ECG signals of rhythm
lead 1.
Select Triggered, the electrocardiograph sends pulses with the height of
5V and the width of 45ms, based on the data of rhythm lead 1.
NOTE: If external output function is enabled, you cannot use the
touch screen of SE-12 Express.
- 121 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
D E F G H I J K
C L
B
A M
V N
U T S R Q P O
- 122 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
Z
Figure 11-3 Closeup View of Template
- 123 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
- 124 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
- 125 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
- 126 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
Closeup View of
Y Closeup view of the selected lead.
Waveforms
CAUTION
RS232 port
RS232
NOTE: The position of the RS232 port may be different on different treadmills.
- 127 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
2) Connect the RS232 port of the ergometer to the RS232 port of the electrocardiograph
with an RS232 cable.
NOTE: The position of the RS232 port may be different on different ergometers.
3) Connect the electrocardiograph to the ergometer
RS232 port
4) Connect the RS232 port of the ergometer to the RS232 port of the electrocardiograph
with an RS232 cable.
NOTE: The position of the RS232 in port may be different on different ergometers.
3. Connect the electrocardiograph to the stress BP monitor
- 128 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
4. Connect the power cords, and connect the earth wire (when necessary)
5. Load the recorder paper
6. Turn on the stress BP monitor and treadmill/ergometer
7. Set the electrocardiograph
1) Activate the stress ECG function.
For details, please contact the manufacturer or the local distributor.
2) After turning on the electrocardiograph, press Pretest on the keyboard to open the main
screen of the exercise test.
3) Select Setup on the main screen1 to open the System Setup screen.
4) Configure the System Setup screen.
For details, please refer to Section 13.4 “System Setup for Exercise Test”.
8. Set the treadmill or the ergometer for external control
Hold down the following two keys for 10 seconds, and then you will enter the service
menu of the Valiant.
By default, the Valiant is shipped with the Lode RS232 communication protocol. It
should be set to the Trackmaster RS232 protocol. You can change the communication
protocol by following the procedures below:
a) Select Prot id and select the second RS232 protocol which is the Trackmaster
communication protocol
b) Select Baudrate and set this to 4800 Baud
c) Press the red button to leave the service menu
d) Turn off the Valiant
- 129 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
- 130 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
WARNING
1. During the exercise test, ensure that there are at least 2 experienced physicians
present. One of them observes the patient and deals with the emergency.
2. Remind patients to take care not to fall down from the treadmill.
3. Press down the emergency stop switch of the treadmill before defibrillating to avoid
the hazard to the patient and the operator.
- 131 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
Default
Items Default
(Only in the U.S.)
Filter Setup
Default
Items Default
(Only in the U.S.)
AC filter On On
Recorder Setup
Default
Items Default
(Only in the U.S.)
- 132 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
- 133 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
After turning on the electrocardiograph, press Pretest on the keyboard to open the main screen of
the exercise test, and then select Setup to open the System Setup screen.
On the System Setup screen, move the cursor on an item, and then press Enter to open the setup
window of the item.
NOTE:
1. If you set the system password in the System Maintenance window, you need enter
the password before opening the System Setup screen. For details, refer to Section
10.11, “System Maintenance Setup”.
2. Except for the General Information Setup window, the Record Information Setup
window and the Protocol Manager screen, the changes of the System Setup screen
are synchronous in both Resting and Exercise ECG tests.
3. The setting methods of filter, recorder and lead of the exercise ECG are the same as
those of the resting ECG. For details about the setting methods, please refer to
Section 10.3 “Filter Setup”, Section 10.4.1 “Setup 1”, Section 10.6 “Transmission
Setup” and Section 10.7 “Lead Setup”.
- 134 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
Select General Info on the System Setup screen, and then press Enter to open the General
Information Setup window.
Item Description
- 135 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
Item Description
Normal BP You can set the normal range of the systolic/diastolic blood pressure. If
Range the patient’s blood pressure values exceed the normal BP ranges, the hint
information will be displayed on the main screen.
Max Predicted HR The following formulas are applied to calculate the target heart rate.
and Target HR Max Predicted HR = 220 - Age
Target HR = Max Predicted HR * 85%
NOTE:
1. 220 and 85 are default values, and you can modify them in
different situations.
2. The range of Max Predicted HR you can enter is from 120 to
250.
- 136 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
Select Record Info on the System Setup screen, and then press Enter to open the Record Info
Setup window.
Item Description
Record Style Select the style of 12-lead report printed in the ECG reports.
Choose from: 3×4+1 or 6×2+1
ST Info Choose from: On or Off
Select On to print out the 12-lead ECG report with ST information during
the exercise test.
Manual Report Choose from: Print, Print&Save or Save.
Select Print, Print&Save or Save to print out, print and save or save the
12-lead ECG report after you press the Recall, 12ld or Medians key
during the exercise test.
For details, please refer to Section 2.2 “Keyboard and Keys”.
Pretest Report Choose from: Close, Print, Print&Save or Save.
Select Print, Print&Save or Save to print out, print and save or save the
12-lead ECG report of the pretest phase automatically at the 20s of the
pretest phase.
- 137 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
Item Description
Exercise Report - Choose Close, Print, Print&Save or Save from the Auto Report
Auto Report list box.
Select Print, Print&Save or Save to print out, print and save or save
the 12-lead ECG report of the exercise phase automatically during
the exercise test.
NOTE: If Auto Report is not set to Close, the 12-lead peak
report including the last 10s ECG data of the exercise
phase will be automatically printed out or saved.
Exercise Report - Choose from: Early Stage, Middle Stage or Late Stage
Print Time Select Early Stage, Middle Stage or Late Stage to print out the
12-lead ECG report automatically at the early, middle or late period
of each stage of the exercise phase.
Recovery Report - Input the start time manually to print the first recovery report.
Start Time For example, if you fill 20 in the Start Time textbox, the first
recovery report will be printed out 20s after entering the recovery
phase.
Recovery Report - Set the interval to print the later recovery reports.
Interval For example, if you select 1 min in the Interval pull-down list, the
later recovery reports will be printed out every 1 min.
NOTE: The time range of Interval is 20-60s.
Final Report - Choose from: Close, Print, Print&Save or Save
Summary Report / Select Print, Print&Save or Save to print out, print and save or save
ST Scope Report / the selected final report after you press the PRINT/STOP key when
Trend Graph Report / the exercise test is terminated.
Sum Template Report NOTE: During the printing course, you can press the
PRINT/STOP key to stop printing the current report.
Final Report - Choose from: On or Off
Edit Conclusion Select On, a hint pops up you to edit conclusion before the Summary
Report is printed out.
- 138 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
Select Protocol on the System Setup screen, and press Enter to open the Protocol Manager
(treadmill or ergometer) screen.
- 139 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
1. Adding Protocols
Select Add on the Protocol Manager screen to display the following window.
Enter the new protocol name in the Protocol Name textbox, and then enter the stage information
of every stage, including the time, speed and grade. After that, press Enter to confirm.
NOTE:
1) When Time of a stage in the exercise phase is set to 00, this stage and its following
stages in the exercise phase will not be carried out.
2) “1/3” in the Add Protocol window stands for “the current page/ the total pages”.
3) Pressing Shift + Left/Right can turn pages in the on the Add Protocol window.
2. Editing Protocols
Select a protocol on the Protocol Manager screen by pressing the Up or Down arrow, and select
Edit to display the Edit Protocol dialog box.
- 140 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
Edit the protocol name or the stage information of every stage, including the time, speed and
grade. After that, press Enter to confirm.
NOTE: For the Bruce or Modified Bruce protocol, only the stage information of the pretest
and recovery phases can be edited.
Select a protocol on the Protocol Manager screen by pressing the Up or Down arrow, and press
Select to set the selected protocol as the default protocol.
4. Deleting Protocols
Pressing Del All on the Protocol Manager screen can delete all the protocols from the
electrocardiograph.
Or, you select a protocol on the Protocol Manager screen, select Delete, and then press Enter to
delete the selected protocol from the electrocardiograph.
NOTE: Only the customized treadmill protocol can be deleted.
5. Restoring Protocols
Pressing Restore on the Protocol Manager screen can restore the factory settings.
6. Return
Pressing Return on the Protocol Manager screen can return to the System Setup screen for the
exercise test.
- 141 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
The above figure shows a 12-lead ECG report in Exercise Stage 2 during the exercise test.
The 12-Lead report includes:
6×2+1R ECG waves
Current Date and Time
Patient Information: Name, ID, Gender, Age
Stage Information
Measure Information
Average Template
Bottom Information
- 142 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
The above figure shows a peak report during the exercise test.
- 143 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
The above figure shows the summary report during the exercise test.
The summary report includes:
Current Date and Time
Patient Information: Name, ID, Gender, Age, height, weight, Room No.
Stage and measuring information at each phase
Comment
Measure Information
Trend graphs of HR and workload
Conclusion
Bottom Information
- 144 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
The above figure shows the ST scope report during the exercise test.
The ST scope report includes:
Patient Information: Name, ID, Gender, Age
Information of Max ST Ascend
Information of Max ST Descend
- 145 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
The above figure shows the average template waves in Exercise Stage 2 during the exercise test.
- 146 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
The above figure shows the average template waves at each phase
- 147 -
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG
The above figure shows the trend graph report during the exercise test.
- 148 -
SE-12 Series Electrocardiograph User Manual Hint Information
Hint information and the corresponding causes provided by the electrocardiograph are listed in
Table 12-1.
Sampling/Analyzing/
ECG signals are being sampled / analyzed / recorded.
Recording
The self-study process of arrhythmia arithmetic in the Trigger
Learning
Sample mode
The examining process of arrhythmia data in the Trigger Sample
Detecting
mode
ECG data is being transmitted from the electrocardiograph to the
Transmitting
PC through the net or serial cable in the auto or rhythm mode.
Memory Full The amount of files on the File Manager screen reaches 200.
Module Error There is something wrong with the signal sample module.
U Disk / USB Printer A U disk, a USB printer or a bar code reader is connected to the
/ USB Scanner USB interface.
- 149 -
SE-12 Series Electrocardiograph User Manual Hint Information
Module Error There is something wrong with the signal sample module.
Lead X off Electrodes fall off the patient or the patient cable falls off the unit.
The safety stop (mushroom type) is revolved and the safety stop
Lanyard Off
(cord type) is pulled out.
The transducer and the main control board lines are not connected
Trans.Comm Err
well.
Transducer Err The transducer doesn’t work well or is not connected well.
Incline Err The incline motor doesn’t work well or is not connected well.
No Speed Signal The speed sensor doesn’t work well or is not connected well.
- 150 -
SE-12 Series Electrocardiograph User Manual FAQ
Chapter 13 FAQ
1. Operating Problems
Q1: I was trying to select a file from the file list on the File Manage screen, but the file was in
the middle of the long list. Is there any way to make the selection faster?
A1: Actually, the system provides a method for fast moving: pressing Shift + Up or Down arrow
can move the cursor up or down in the file list very fast.
Q2: I was just about to input the age when I suddenly realized that I had entered the Name
textbox unintentionally, can I just go back without pressing Tab for a whole circle?
A2: As a matter of fact, the system does take such unintentionalities into consideration by
providing Shift + Tab as the way back, as the Microsoft Windows operating system does.
Q3: I want to save the ECG data without printing, could it be possible?
A3: Yes, you can set Print Out to Off in the Record Info Setup1 window. Or, in the auto or
rhythm mode, you can directly press Shift + PRINT/STOP to enable or disable the print out
function. The ECG data will be collected and saved without printing. In the same way, if the
transmission settings are configured, the ECG data could be transmitted to the PC without
printing.
Q4: The screen of SE-12 series electrocardiograph is too shiny. Could it be possible to weaken
the brightness of the screen?
A4: There is a setup item named brightness in the Display & Sound Setup window, you can
press the Left or Right arrow to change the value, which would lead to the change of the
brightness of the screen, please refer to Section 10.8, “Display & Sound Setup”.
Q5: I want to input the patients’ phone number in the Patient Information window, but there is
no such item. Can I add it manually?
A5: Yes, there is a user-defined item for entering patient information. It works in this way: first
input the name of the item in the User-defined textbox in the Patient Information Setup
window, e.g. Tel. Then return to the main screen1, and open the Patient Information window,
the Tel item will be displayed in this window. Now it’s possible to input the phone number of
the patient in the Tel textbox. For details, please refer to Section 10.5 “Patient Information
Setup” and Section 4.1.2 “Entering Data”.
- 151 -
SE-12 Series Electrocardiograph User Manual FAQ
Q6: Memory Full is displayed on the main screen; Or, the hint Memory full! Replace the earliest
file? pops up every time when I save an ECG report to the electrocardiograph. What am I
supposed to do?
A6: Memory Full is used to remind you that the amount of stored file reaches 200.
The display of the pop-up hint Memory full! Replace the earliest file? is related to the
settings of the File Setup window.
Select Off from the Replace When Memory Full list box, when the amount of stored files
reaches 200 and you save an ECG report to the electrocardiograph, the hint Memory full!
Replace the earliest file? pops up.
Select On from the Replace When Memory Full list box, when the amount of stored files
reaches 200 and you save an ECG report to the electrocardiograph, the hint Memory full!
Replace the earliest file? does not pop up.
You can deal with the hint as follows:
1) You can just delete several stored files from the electrocardiograph to ensure the amount
of stored file not to reach 200.
2) When Memory Full is displayed on the main screen, you can set Auto Save to To U
Disk to save the added ECG reports. However, the amount of stored files in the
electrocardiograph still reaches 200.
2. Printing Problems
Q2: The hint Paper Error is displayed on the screen, what should I do?
A2: It might be the result of unsuccessful detection of the black markers, first open the recorder
casing so as to clear the error information, and then check whether the black marker is on the
bottom of the paper. Reload the paper in the paper tray. If it doesn’t work, change the paper.
If the problem still exists, please contact the manufacturer or the local distributor for further
disposal.
Q3: The hint No Paper is displayed on the screen, what should I do?
A3: Check whether the paper runs out, or the black marker is just facing the black marker
detection window on the thermal printing head, as the following figure shows.
- 152 -
SE-12 Series Electrocardiograph User Manual FAQ
Reload the paper in the paper tray, close the recorder casing firmly. If the problem still exists,
please contact the manufacturer or the local distributor for further disposal.
Q4: I want to print the hospital name in the report, but I can’t find the place to enter it, where is
it?
A4: Please open the Other Setup window, and move the cursor to the Institution textbox, and
then input the hospital name. The content you input in this textbox will be printed in the
report. For details, please refer to Section 10.12, “Other Setup”.
Q5: I pressed the PRINT/STOP key, but the ECG didn’t start printing, what’s wrong with it?
A5: The system will not respond to the PRINT/STOP key during the first 3s after you return to
the main screen. Therefore, you have to wait for a few seconds, and then you are able to start
the printing by pressing the PRINT/STOP key.
If you wait for a few seconds, but you still unable to start the printing by pressing the
PRINT/STOP key, please check whether there is any error information displayed on the
screen.
If the hint No Paper or Paper Error is displayed on the screen, please deal with it according
to the above-mentioned measures.
If the hint Transmitting… is displayed on the screen, which means that the ECG is
transmitting the data to the PC, please wait a few seconds. You can start the printing after the
data is transmitted.
If the problem still exists, please contact the manufacturer or the local distributor for further
disposal.
Q6: I set the filter, speed and gain on the main screen1, but these settings were changed after
printing.
A6: The filter, speed and gain which are set on the main screen1 will not be saved, and they are
changed when you exit the main screen1 or after printing. If you want to save these settings,
please set them in the Record Info Setup window and the Filter Setup window.
- 153 -
SE-12 Series Electrocardiograph User Manual FAQ
3. Transmitting Problems
Q1: The ECG doesn’t respond to any keys after a long time of transmission. It transmits nothing
for there is no new data appearing on the screen of the PC software. What should I do?
A1: Some error may occur during the transmission course, for example, the connection between
the ECG and the net cable may loosen. In this case, please connect the net cable well. If it
doesn’t work, please restart the ECG.
If the problem still exists, please contact the manufacturer or the local distributor for further
disposal.
4. Main Unit Problems
Q1: After power-on, the ECG stays on the logo screen and doesn’t open the main screen. I have
restarted the machine several times, but there is no better change.
A1: The reason for this problem might be: there is a key pressed down, without springing up.
Find that key, and make it spring up, the problem should be solved.
Q2: I was doing the examination when the machine suddenly gave out a sound and displayed the
hint Lead Off. What should I do?
A2: The corresponding electrodes are not connected well. Please find out which lead is off by
checking the Lead Name area on the main screen (please refer to Section 4.3.1, “About the
Main Screen”). The lead whose name is highlighted is off. Please check whether the
corresponding electrode of the lead is connected to the patient skin well, and then make sure
that the patient cable socket is connected to the patient cable firmly.
If none of the above-mentioned measures takes effect, please contact the manufacturer or the
local distributor for further disposal.
Q2: I press the Stop tmill key to stop the test temporarily during the exercise test, but in a while,
the treadmill cannot be started after I press the Start tmill key. What should I do?
A1: Press the Stop tmill key, 1 minute later press the Start tmill key to start the treadmill.
- 154 -
SE-12 Series Electrocardiograph User Manual Cleaning, Care and Maintenance
CAUTION
1. Turn off the power before cleaning and disinfection. The mains supply must be
switched off if it is used.
14.1 Cleaning
CAUTION
Any remainder of detergent should be removed from the unit and the patient cable after
cleaning.
1. Remove the remainder gel from the electrodes with a clean soft cloth first.
2. Take suction bulbs and metal cups of chest electrodes apart, and take clamps and metal parts
of limb electrodes apart.
3. Clean them in warm water and make sure there is no remainder gel.
4. Dry the electrodes with a clean dry cloth or air dry naturally.
- 155 -
SE-12 Series Electrocardiograph User Manual Cleaning, Care and Maintenance
CAUTION
Do not clean the unit and accessories with abrasive fabric and avoid scratching the
electrodes.
Dirty and soiled thermal print head will deteriorate the printing definition. So it should be cleaned
at least once a month regularly.
CAUTION
Prevent the detergent from seeping into the main unit while cleaning. Do not immerse the
unit or the patient cable into liquid under any circumstances.
For details on cleaning the treadmill, please refer to Treadmill User Manual.
14.2 Disinfection
Disinfection of the main unit is not necessary need in daily maintenance, it is only necessary in
operating room. In that case, please use hospital standard disinfectant.
NOTE: Clean and disinfect the chest and limb electrodes after each use.
CAUTION
1. Do not use high-temperature, high-pressure vapour or ionizing radiation as
disinfection methods.
- 156 -
SE-12 Series Electrocardiograph User Manual Cleaning, Care and Maintenance
CAUTION
Operate the cardiograph, charge the battery, and store the battery at a temperature of
40℃ (104ºF) or lower. Exposure to higher temperature may reduce battery life, damage
the battery, and degrade overall cardiograph performance.
1) Capacity Identification
The battery capacity can be identified according to the battery symbol in the top right corner
of the LCD screen.
2) Recharge
SE-12 series electrocardiograph is equipped with the recharge control circuit together with
the battery. When the unit is connected to the mains supply, the battery will be recharged
automatically. Then the battery recharging indicator ( ) and the mains supply indicator
( ) will be lit at the same time. During the recharging course, the symbol flashes in
the top right corner of the LCD screen. After the battery is fully recharged, the symbol stops
flashing, and the battery recharging indicator ( ) is black.
Because of the capacity consumption during the storage and transport course, the battery
capacity is not full when it is used for the first time. Battery recharge should be considered
before the first use.
NOTE: The battery will automatically stop charging if you print an ECG report.
CAUTION
Repeated undercharging of the battery will damage the battery and reduce battery life.
- 157 -
SE-12 Series Electrocardiograph User Manual Cleaning, Care and Maintenance
3) Replacement
When the useful life of the battery is over, or foul smell and leakage are found, please
contact the manufacturer or the local distributor for replacement.
WARNING
1. Only qualified service engineers authorized by the manufacturer can open the battery
compartment and replace the battery, and the battery of the same model and
specification provided by the manufacturer must be used.
2. Danger of explosion -- Do not reverse the anode and the cathode when installing the
battery.
3. When the battery’s useful life is over, contact the manufacturer or the local distributor
for disposal or dispose of the battery according to local regulations.
4. Remove the battery from the electrocardiograph when the electrocardiograph isn’t
used for a long time.
5. If the battery is stored alone and not used for a long time, we recommend that the
battery be charged at least once every 6 months to prevent over-discharge.
CAUTION
If the battery has been fully charged and requires recharging after printing only a few
ECGs, consider replacement.
NOTE: Recorder paper provided by the manufacturer should be used. Other paper may
shorten the life of the thermal print head. The deteriorated print head may lead to
illegible ECG reports and block the advance of the paper.
Storage Requirements:
Recorder paper should be stored in a dry, dark and cool area, avoiding excessive
temperature, humidity and sunshine.
♦ Do not put the recorder paper under fluorescence for a long time.
♦ Make sure that there is no polyvinyl chloride or other chemicals in the storage
environment, which will lead to color change of the paper.
♦ Do not overlap the recorder paper for a long time, or else the ECG reports may
trans-print each other.
- 158 -
SE-12 Series Electrocardiograph User Manual Cleaning, Care and Maintenance
Perform a visual inspection of all equipment and peripheral devices daily. If you notice any items
that need repair, contact a qualified service engineer to make the repairs.
Check the case and display screen for cracks or other damage.
Regularly inspect all plugs, cords, cables, and connectors for fraying or other damage.
Verify that all cords and connectors are securely seated.
Inspect keys and controls for proper operation.
CAUTION
Besides the maintenance requirements recommended in this manual, comply with local
regulations on maintenance and measurement.
The following safety checks should be performed at least every 12 months by a qualified person
who has adequate training, knowledge, and practical experience to perform these tests.
a) Inspect the equipment and accessories for mechanical and functional damage.
b) Inspect the safety related labels for legibility.
c) Inspect the fuse to verify compliance with the rated current and circuit-breaking
characteristics.
d) Verify that the device functions properly as described in the instructions for use.
e) Test the protection earth resistance according to IEC/EN 60601-1: Limit: 0.1 ohm.
f) Test the earth leakage current according to IEC/EN 60601-1: Limit: NC 500μA, SFC
1000μA.
g) Test the enclosure leakage current according to IEC/EN 60601-1: Limit: NC 100μA,
SFC 500μA.
h) Test the patient leakage current according to IEC/EN 60601-1: Limit: NC a.c. 10μA, d.c.
10μA; SFC a.c. 50μA, d.c. 50μA.
i) Test the patient auxiliary current according to IEC/EN 60601-1: Limit: NC a.c. 10μA,
d.c. 10μA; SFC a.c. 50μA, d.c. 50μA.
j) Test the patient leakage current under single fault condition with mains voltage on the
applied part according to IEC/EN 60601-1: Limit: 50μA (CF).
The leakage current should never exceed the limit. The data should be recorded in an equipment
log. If the device is not functioning properly or fails any of the above tests, the device has to be
repaired.
- 159 -
SE-12 Series Electrocardiograph User Manual Cleaning, Care and Maintenance
WARNING
Failure on the part of the responsible individual hospital or institution employing this
equipment to implement a satisfactory maintenance schedule may cause undue
equipment failures and possible health hazards.
1) Main Unit
♦ Avoid excessive temperature, sunshine, humidity and dirt.
♦ Put the dustproof coat on the main unit after use and prevent shaking it violently when
moving it to another place.
♦ Prevent any liquid from seeping into the equipment; otherwise the safety and the
performance of the electrocardiograph can not be guaranteed.
2) Patient Cable
♦ Integrity of the patient cable, including the main cable and lead wires, should be
checked regularly. Make sure that it is conductible.
♦ Do not drag or twist the patient cable with excessive stress while using it. Hold the
connector plug instead of the cable when connecting or disconnecting the patient cable.
♦ Align the patient cable to avoid twisting, knotting or crooking in a closed angle while
using it.
♦ Store the lead wires in a big wheel to prevent any people from stumbling.
♦ Once damage or aging of the patient cable is found, replace it with a new one
immediately.
3) Reusable Electrodes
♦ Electrodes must be cleansed after use and make sure there is no remainder gel on them.
♦ Keep suction bulbs of chest electrodes away from sunshine and excessive temperature.
♦ After long-term use, the surfaces of electrodes will be oxidized because of erosion and
other causes. By this time, electrodes should be replaced to achieve high-quality ECG
records.
CAUTION
The device and accessories are to be disposed of according to local regulations after
their useful lives. Alternatively, they can be returned to the dealer or the manufacturer for
recycling or proper disposal.
- 160 -
SE-12 Series Electrocardiograph User Manual Accessories
Chapter 15 Accessories
WARNING
Only the patient cable and other accessories supplied by the manufacturer can be used.
Or else, the performance and electric shock protection can not be guaranteed.
- 161 -
SE-12 Series Electrocardiograph User Manual Accessories
Fuse 11.21.64073
03.24.38952
Smart ECG Viewer Software
03.24.38953
U Disk 11.18.78105
Wireless AP 11.17.047338
SE-12 series electrocardiograph and accessories are available by contacting the manufacturer or
your local distributor.
NOTE: The chest electrodes, limb electrodes, pediatric chest electrodes and pediatric
limb electrodes are not available in the U.S.
- 162 -
SE-12 Series Electrocardiograph User Manual After-Sales Service
If you have any question about maintenance, technical specifications or malfunctions of devices,
contact your local distributor.
- 163 -
SE-12 Series Electrocardiograph User Manual Technical Specifications
IEC 60601-1:1988+A1:1991+A2:1995
EN 60601-1:1990+A1:1993+A2:1995
IEC60601-1-2:2001+A1:2004
Comply with: EN 60601-1-2:2001+A1:2006
IEC/EN60601-2-25
ANSI/AAMI EC11
IEC/EN 60601-2-51
Degree of safety of
Equipment not suitable for use in the presence of flammable
application in the presence
gas
of flammable gas:
- 164 -
SE-12 Series Electrocardiograph User Manual Technical Specifications
25%~93% 25%~80%
Relative Humidity:
Non-Condensing Non-Condensing
SE-12: approx. 5.0kg (11.0 lbs); SE-12 Express: approx. 6.5kg (14.3 lbs)
Weight
(Excluding recorder paper and battery)
- 165 -
SE-12 Series Electrocardiograph User Manual Technical Specifications
Recording
HR Recognition
Accuracy: ±1 BPM
ECG Unit
- 166 -
SE-12 Series Electrocardiograph User Manual Technical Specifications
AC Filter: On/Off
LOWPASS Filter:150Hz/100Hz/75Hz
CMRR ≥115dB
Pacemaker Detection
Amplitude ±2 to ±700 mV
NOTE:
1. Test the accuracy of input signal reproduction according to the methods described in
clause 4.2.7.2 in ANSI/AAMI EC11: 1991/(R) 2001/(R) 2007, and the result complies
with the clause 3.2.7.2 in ANSI/AAMI EC11: 1991/(R) 2001/(R) 2007.
2. Overall System Error is tested using the method described in AAMI EC11 3.2.7.1.
Overall System Error is not more than ±5%.
3. Frequency Response is tested using the method described in AAMI EC11 3.2.7.2
methods A and D.
- 167 -
SE-12 Series Electrocardiograph User Manual EMC Information
RF emission
Class A
CISPR 11
The 12-channel electrocardiograph is
Harmonic
suitable for use in all establishments, other
emissions Class A
than domestic and those directly connected
IEC/EN 61000-3-2
to the public low-voltage power supply
Voltage network that supplies buildings used for
fluctuations/ flicker domestic purposes.
Complies
emissions
IEC/EN 61000-3-3
- 168 -
SE-12 Series Electrocardiograph User Manual EMC Information
- 169 -
SE-12 Series Electrocardiograph User Manual EMC Information
NOTE UT is the a.c. mains voltage prior to application of the test level.
- 170 -
SE-12 Series Electrocardiograph User Manual EMC Information
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the 12-channel
electrocardiograph is used exceeds the applicable RF compliance level above, the
12-channel electrocardiograph should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the 12-channel electrocardiograph.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
- 171 -
SE-12 Series Electrocardiograph User Manual EMC Information
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 : These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
- 172 -
SE-12 Series Electrocardiograph User Manual Abbreviation
Appendix 3 Abbreviation
Abbreviation Statement
BP Blood Pressure
ECG Electrocardiogram/Electrocardiograph
HR Heart Rate
LA Left Arm
LL Left Leg
RA Right Arm
RL Right Leg
ID Identification
AC Alternating Current
NC Normal Condition
- 173 -
- 174 -