Esprit Operation Manual
Esprit Operation Manual
OPERATOR’S MANUAL
REF 580-1000-01P
For Technical Support, contact:
Respironics. Inc. Customer Service
Within the U.S.A. 1-800-345-6443
Outside the U.S.A. 724-387-4000
Facsimile 724-387-5012
[email protected]
Authorized Representative
Respironics Deutschland, Inc.
Gewerbestrasse 17
D-82211 Herrsching Deutschland
+49-8-15-29-30-60
This work is protected under Title 17 of the United States copyright code and is the sole property of Respironics.
No part of this document may be copied or otherwise reproduced, or stored in any electronic information
retrieval system, except as specifically permitted under United States copyright law, without the prior written
consent of Respironics.
Revision Description
G Update to include Esprit 4.10 software enhancements
H Update to include Esprit 4.10 software enhancements
J Update to include Esprit 5.0 software enhancements
K Update to include EST procedure change and CE mark change
L Update to include 7.0 software enhancements and format
change
M Update to include 7.10 software enhancements
N Update to include 8.10 software enhancements
P Addition of EMC Declaration
What’s New
NICO®-Esprit Interface
The NICO-Esprit Interface software allows the Esprit Ventilator and the NICO Monitor to exchange information
via a bidirectional RS-232 serial link. Through this link, the Esprit ventilator supplies the NICO monitor with
breath type and FIO2 information. In turn, the NICO monitor provides the Esprit ventilator with data for trending
including: Cardiac Output Average, O2 Saturation, End Tidal CO2, Volume CO2, Alveolar Tidal Volume, and
Alveolar Minute Volume. See the NICO-Esprit Interface option insert for complete details.
E-Cycle Range
The E-cycle maximum setting has been increased from 45% to 80%. The new range is 10-80%. E-Cycle
defines the percent of peak inspiratory flow that end inspiratory flow needs to drop to in order for inspiration to
end.
Pplat
Plateau Pressure (Pplat) has been added to the Static C & R display and Patient Status windows. Pplat is a
measured value collected during the last saved Static C and R maneuver.
3. Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
5. Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Back Panel Connections & Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Connecting AC Power Cord . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Power On/Off. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Entering Diagnostic Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
User Configuration Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Backup Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
Extended Self Test (EST) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
9. Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
Visual Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
Audible Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
Alarm Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3
Alert Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4
Alarm Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-8
G. Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-1
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-1
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-4
1. Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Please read this manual thoroughly and become familiar with the ventilator's
operation before using it on a patient. For additional information about
accessories or related equipment, such as humidifiers and remote alarm
systems, refer to the appropriate instruction manual prior to operating with the
Esprit Ventilator.
CAUTION: Federal law (USA) restricts this device to sale by or on the order of a
physician.
NOTE: Follow the setup instructions in this manual before placing the Esprit
Ventilator into service. If you have questions, contact Respironics
Customer Service at 1-800-345-6443.
WARNING: A condition that could cause injury to a patient or operator if the operating
instructions in this manual are not followed correctly.
CAUTION: A condition that could cause damage to, or shorten the service life of,
the Esprit Ventilator.
Summary of Warnings
Warnings, Cautions, • Patients on life-support equipment should be visually monitored
and Notes by competent medical personnel, since life-threatening
circumstances may arise that may not activate alarms. The
operator should heed all appropriate alarms and follow the
instructions and warnings in this operator’s manual. Always check
life-support equipment for proper operation before use.
• Do not use in the presence of flammable anesthetics. Possible
explosion hazard.
• One person alone should not attempt to lift the ventilator or
remove it from the shipping carton. At least two people are
required in order to avoid possible personal injury or damage to
the equipment.
• To reduce the chance of contamination or infection, always use an
inspiratory bacteria filter when the Esprit is in operation. Refer to
manufacturer’s instructions and follow institutional infection
control guidelines when replacing the inspiratory filter.
• Do not use anti-static or conductive hoses or conductive patient
tubing.
• The expiratory filter housing may be hot if removed from the
ventilator immediately after use. Wait 15 minutes after turning off
Cautions
• Federal law (USA) restricts this device to sale by or on the order of
a physician.
• Be sure to check all exterior parts of the ventilator. Problems
found during inspection should be corrected and/or reported to
Respironics before using the ventilator.
Notes
• Follow the setup instructions in this manual before placing the
Esprit Ventilator into service. If you have questions, contact
Respironics Customer Service at 1-800-345-6443.
• Respironics recommends that, before using the ventilator for the
first time, the user should wipe the exterior clean and disinfect or
sterilize its components according to the instructions in Chapter
10, “Care and Maintenance” or the component manufacturer’s
instructions.
• Follow institutional infection control guidelines when replacing
inspiratory bacteria filter.
• When adding attachments or other components or subassemblies
to the breathing system, i.e., HME or Humidifier, the user should
ensure that the inspiratory and expiratory resistances (measured at
the patient connection port) do not exceed 6 cmH2O (hPa) at a
flow of 60L/min for adults, 30L/min for pediatrics.
• Follow institutional infection control guidelines when replacing
heated expiratory bacteria filter.
• The Esprit Ventilator should only be connected to an appropriate
medical grade 100% O2 gas source capable of delivering a
regulated 40 to 90 PSIG (276-620 kPa).
• The Esprit Ventilator is shipped with the appropriate gas fittings
and hoses for the intended environment, i.e. DISS (U.S.A. and
Canada), Ohmeda (Germany), NIST (UK), Air Liquide (France),
SIS (Australia).
• All volumes entered into Esprit are assumed to be BTPS (Body
Temperature atmospheric Pressure Saturated (with H2O)) volumes
unless otherwise noted. All volumes reported by Esprit are
reported as BTPS volumes. All pressures are assumed to be
relative to atmospheric pressure unless otherwise noted.
• The Air Inlet Filter houses a reusable foam filter which should be
periodically cleaned by the user. Refer to Chapter 10, “Care and
Maintenance”, for more information on filter changes.
• Unless the Mains Circuit Breaker is turned OFF, electrical power is
applied to the ventilator even though the front panel switch is in
the OFF position. With the Mains Circuit Breaker ON, if the
optional Backup Battery is connected, Esprit will charge the
battery if it requires a charge.
• To disconnect the ventilator from MAINS power, remove the AC
plug from the wall power receptacle. The MAIN switch/circuit
breaker is covered to prevent unintentional ventilator turn off.
• If the operator sets the %O2 setting to 100%, the 100% O2
indicator does not light. The 100% O2 indicator only lights when
the 100% O2 front panel key has been pressed.
• The Esprit Ventilator selects its power source based on the
following prioritization: AC power (if present), external battery,
then backup battery.
• The ventilator may automatically reset certain types of alarm
conditions once the causes of the alarms are corrected. After an
automatic reset, the ventilator will clear the audible alarm, and
will display a Low Urgency Alarm alert in the Alert Message Insert
to inform the operator that an alarm condition existed. When this
situation occurs, use ALARM RESET to clear the visual alarm
indicator.
• If the 100% O2 key is pressed and a 100% O2 gas source is not
available, the Low O2 alarm will be active for the two-minute
100% O2 delivery period.
• Manual breaths are not permitted during the inspiratory phase of a
breath (whether manual or spontaneous). Pressing the MANUAL
BREATH key during these times will not result in the delivery of a
manual breath.
• Some settings buttons appear active despite the fact they are not
being used in the ACTIVE MODE. This is because the setting is
used in Apnea Ventilation or when manual inspiration is pressed.
The operator should always choose a value for an active button
that is appropriate for the patient being ventilated.
• When the active mode is set to NPPV, the HIP Limit Setting will
automatically be adjusted to 10 cmH2O above the current IPAP
setting.
• Pt. Leak only appears on Patient Data block on Settings screen.
• Esprit keeps a distinct set of alarm limits for each ventilation
breath type (VCV, PCV and NPPV).
• Any of the changes made in the screen shown in Figure 8-14, do
not take effect until the operator switches to the new ventilation
breath type (in this case Pressure Control).
• If the EXP HOLD key is held continuously and the exhalation hold
period exceeds 5 seconds, Esprit automatically terminates the
exhalation hold period and begins a new inspiratory period.
Symbols
IEC 417 This symbol indicates the ON condition for part of the
Symbol 5264 equipment. When pressed, the ventilator will operate from the
MAINS voltage if connected or from the Backup Battery if the
battery charge is within operating specifications.
IEC 417 This symbol indicates the OFF condition for part of the
Symbol 5265 equipment.
IEC 878-02 This symbol indicates TYPE B applied part, which indicates
Symbol 02 equipment that provides a particular degree of protection against
electric shock, particularly with regards to allowable leakage
current and of the protective earth connection.
IEC 417 This symbol is located on the rating plate. It indicates the
Symbol 5032 equipment is SUITABLE FOR ALTERNATING CURRENT.
IEC 529 Drip proof.
ISO 7000 This symbol indicates HOT SURFACE warning. It appears next to
Symbol 0535 the heated expiratory filter.
Symbols
Symbols (Continued)
IEC 5436 This symbol indicates the ALARM SILENCE key (Silences alarm
for two minutes.)
IEC 417 This symbol indicates the portion of the circuit breaker that must
Symbol 5008 be pushed in to turn the CIRCUIT BREAKER OFF.
IEC 417 This symbol indicates the portion of the circuit breaker that must
Symbol 5007 be pushed in to turn the CIRCUIT BREAKER ON.
Unpacking The Esprit Ventilator has been carefully packaged to assure safe shipping. In
addition, the packing container has been designed for easy unpacking. Do not
discard packing materials until ventilator installation is complete and proper
machine function has been verified.
Before unpacking Esprit Ventilator, examine the shipping carton(s) for visible
damage. If the shipping carton(s) arrives damaged or if you suspect the
contents are damaged, contact the carrier for an inspection report. If any
damage is evident, Respironics recommends that you photograph the carton(s)
before the shipment is unpacked. Report any damage to the shipping container
or Esprit Ventilator to your local authorized Respironics distributor and to the
carrier.
Save all packing material after removing the ventilator. In the event that the
Esprit Ventilator needs to be repacked and reshipped, use the original packing
material or order replacement material from a Respironics representative.
Esprit Ventilator
Backup Battery
WARNING: One person alone should not attempt to lift the ventilator or remove it from
the shipping carton. At least two people are required in order to avoid
possible personal injury or damage to the equipment.
Getting Started
Unpacking Instructions
Refer to Figure 4-1.
• Tools Required: Box knife
1. Using box knife, cut slit in packaging tape on top of shipping carton.
2. Remove accessories box and optional flex arm box (not shown).
3. Remove top foam insert from inside carton.
4. Roll the plastic shipping bag (not shown) off the ventilator.
5. Gently lift ventilator from the bag and carton.
6. Remove patient circuit package (not shown).
7. Store carton, foam insert, and plastic bag in safe place for possible
future use.
Inspection After unpacking the ventilator, inspect its cabinet exterior for damage,
including cracks and scratches or blemishes. Inspect the front panel for
scratches, chips, abrasions or other deformities.
CAUTION: Be sure to check all exterior parts of the ventilator. Problems found
during inspection should be corrected and/or reported to Respironics
before using the ventilator.
List of Parts and Using the Packing List that accompanies the ventilator, take an inventory of
Accessories the entire shipment before assembling the ventilator. In case of discrepancies,
immediately contact Respironics Customer Service at 1-800-345-6443.
Getting Started
NOTE: Respironics recommends that, before using the ventilator for the first
time, the user should wipe the exterior clean and disinfect or sterilize
its components according to the instructions in Chapter 10, “Care and
Maintenance” or the component manufacturer’s instructions.
Getting Started
Repacking Should the Esprit Ventilator need to be returned to Respironics for servicing, or
shipped elsewhere for any reason, instructions for repacking are listed below.
The following instructions should be followed closely to avoid damage to the
ventilator.
WARNING: One person alone should not attempt to lift the ventilator or remove it from
the shipping carton. At least two people are required in order to avoid
possible personal injury or damage to the equipment.
CAUTION: Always ship the Esprit Ventilator using the original packing material. If
the original material is not available, contact your Respironics
representative to order replacements.
Repacking Instructions
Refer to Figure 4-1.
• Tools Required: Heavy duty packaging tape
1. Open carton so that bottom foam insert is facing up.
2. Place backup battery in the bottom foam insert if it is being shipped.
3. Place center foam insert into box on top of the bottom foam insert.
4. If you are also shipping the flex arm, place it in its box (not shown)
and place the box in the bottom of the center foam insert.
5. Remove all power cords and accessory items from the Esprit
Ventilator.
6. Set the open ventilator shipping bag (not shown) in the box on the
middle foam insert.
7. Gently place Esprit Ventilator into the open bag. Check to ensure that
the ventilator is firmly positioned into bottom foam insert. Close
plastic bag over the ventilator.
8. Replace accessories box (not shown) in the center foam insert beside
the ventilator (if also being shipped).
9. Place top foam insert onto ventilator. Ensure snug fit.
10. Close top flaps of carton and seal with heavy-duty packaging tape.
Getting Started
Inspiratory Bacteria The inspiratory bacteria filter (4) in Figure 4-2, mounts on the gas outlet port
Filter Installation (1) located in the lower right corner on the front of the Esprit Ventilator. If the
optional O2 sensor (2) will be used then it will be connected to the gas outlet
port (1) before the inspiratory bacteria filter is connected. For more
information regarding the optional O2 sensor refer to Chapter 13, “Options and
Accessories”.
5
4
3
2
1
5 4 3 2 1
Flow
Getting Started
Heated Expiratory
Bacteria Filter
Installation
WARNING: The expiratory filter housing may be hot if removed from the ventilator
immediately after use. Wait 15 minutes after turning off ventilator power
before removing the heated expiratory bacteria filter. Exercise caution when
handling the filter housing.
Getting Started
Getting Started
Oxygen Source
Connection
WARNING: All oxygen connections should be carefully inspected to ensure that leaks
are not present. Excessive leaks can result in higher than normal ambient
oxygen concentrations and create a potentially hazardous oxygen-enriched
environment.
WARNING: Care in the routing of the oxygen inlet hose should be exercised to ensure it
is not exposed to mechanisms that could cause damage by cutting or
heating/melting.
CAUTION: The Esprit Ventilator oxygen filter should be replaced annually as a part
of preventive maintenance.
NOTE: The Esprit Ventilator is shipped with the appropriate gas fittings and
hoses for the intended environment, i.e. DISS (U.S.A. and Canada),
Ohmeda (Germany), NIST (UK), Air Liquide (France), SIS (Australia).
Getting Started
Getting Started
Patient Circuit Flex A patient circuit flex arm is provided for use on the Esprit Ventilator. The
Arm Installation patient circuit flex arm may be installed on either the left or right side rail on
the Esprit Ventilator.
Getting Started
Ventilator Positioning The Esprit Ventilator may be mounted on a table top or on an optional
Respironics approved cart.
CAUTION: Table top must be capable of bearing 66 pounds (30 kg) of weight – 86
pounds (39 kg) with Backup Battery – and must be at least 24 inches
(60 cm) deep by 14 inches (35 cm) wide in dimension in order to
accommodate the ventilator.
CAUTION: To prevent damage to ventilator, ensure that the table top surface is
completely flat, dry, and free of dust, dirt, and debris before positioning
the ventilator.
Getting Started
Back Panel
Connections &
Controls
Remote Alarm Parallel PCMCIA
Serial Port Nurse Call Printer Port Card Slot Analog Port
DC Battery
Connector
AC Inlet
Cooling
Fan Humidifier AC
Inlet Circuit Breaker
Circuit Breaker
Potential Cover
Equalization (not shown)
Connection
O2 Inlet Mains Circuit
Filter & Breaker
Water
Trap Humidifier
AC Outlet
O2 Inlet
Connector
Air Inlet Duct
Elapsed & Filter
Time
Meter:
Remote Alarm Nurse Call: Connection source for remote alarm devices.
Parallel Printer Port: For use with the Communications Option. Refer to
Chapter 13, “Options and Accessories”.
PCMCIA Card Slot: (PC Card) For use with the Trending Option. Refer to
Chapter 13, “Options and Accessories”.
Setup
WARNING: The cover plate for the PCMCIA slot at the back of the ventilator must be
replaced after the adaptor and card are installed. This is to protect the
ventilator.
CAUTION: The PCMCIA card should only be removed by trained service personnel
once power to the ventilator is off.
WARNING: AC power is applied to the humidifier from the Esprit humidifier outlet (only
available on 100-120 VAC ventilators). Under no circumstances does the
Esprit Ventilator provide control for the humidifier. To ensure patient safety,
it is important that any humidifier used with the Esprit Ventilator include an
acceptable temperature control and monitoring mechanism, as well as a
temperature display and appropriate alarm capabilities (refer to ISO 8185).
Circuit Breaker Cover: Circuit breaker cover (not shown) protects the
Humidifier AC Circuit Breaker and Mains Circuit Breaker switches.
Air Inlet Duct and Filter: Filters room air drawn into the ventilator internal air
source.
Elapsed Time Meter: Records time when the Power On/Off Switch is on ( ).
O2 Inlet Filter and Water Trap: Filters oxygen entering the ventilator.
Setup
NOTE: All volumes entered into Esprit are assumed to be BTPS (Body
Temperature atmospheric Pressure Saturated (with H2O)) volumes
unless otherwise noted. All volumes reported by Esprit are reported as
BTPS volumes. All pressures are assumed to be relative to atmospheric
pressure unless otherwise noted.
NOTE: The Air Inlet Filter houses a reusable foam filter which should be
periodically cleaned by the user. Refer to Chapter 10, “Care and
Maintenance”, for more information on filter changes.
Connecting AC Power
Cord
2
3
Circuit Breaker
Cover not shown.
CAUTION: Before connecting the Esprit Ventilator to the AC power source, ensure
that the total electrical load does not exceed the ampere rating of the
AC branch circuit, especially when using the ventilator with other
electrical equipment. An AC branch circuit includes all outlets serviced
by a single circuit breaker. If the maximum current drain through a
branch circuit exceeds the circuit breaker’s rating, the branch circuit
will open, causing the ventilator to lose power. For further information,
consult a service technician or a trained biomedical technician.
CAUTION: The Esprit Ventilator is shipped with a power cord that complies with
electrical safety standards. Do not use substitute power cords unless
specifically instructed to do so by an authorized distributor or qualified
personnel. Do not modify the power cord or connect it with electrical
extension cords or outlet adapters.
Setup
Setup
Power On/Off The Esprit Ventilator ON/OFF ( / ) switch is located in the lower portion of
the front panel, and is recessed to avoid inadvertent access.
When the switch is in OFF ( ) position, the ventilator does not provide
mechanical ventilation. If the AC plug is connected and the Mains circuit
breaker (refer to Figure 5-1) is ON ( ) AC power is active and the green Mains
circuit indicator on the front panel, is illuminated.
Power On/
Off Switch
Setup
WARNING: The Esprit front panel LEDs will indicate the power source that is being
used. If the ventilator is plugged in and the MAINS LED is not lit, either the
circuit breaker is off or the wall power outlet is not functioning.
WARNING: The two circuit breakers (MAINS/Humidifier) located on the back of the
ventilator are covered to prevent unintentional ventilator power-off. Do not
use the circuit breaker to power the ventilator on/off. The power switch is
located on the front of the ventilator below the front panel.
NOTE: Unless the Mains Circuit Breaker is turned OFF, electrical power is
applied to the ventilator even though the front panel switch is in the
OFF position. With the Mains Circuit Breaker ON, if the optional Backup
Battery is connected, Esprit will charge the battery if it requires a
charge.
NOTE: To disconnect the ventilator from MAINS power, remove the AC plug
from the wall power receptacle. The MAIN switch/circuit breaker is
covered to prevent unintentional ventilator turn off.
Setup
User Configuration The buttons across the top and bottom bars are used to select the Diagnostic
Screen Mode functions. To set up the machine for the first time for operation as a
ventilator, USER CONFIG should be selected. When this button is pressed the
following screen appears.
Setup
Setup
Window insert
to change the
value of Month
Window insert
to change the
value of Hour
4. After the date and time values have been set, the changes will be
shown on the screen.
5. Press the APPLY DATE and APPLY TIME buttons to activate the
change.
Setup
Apply Date
button
Apply Time
button
Figure 5-9: User Configuration Screen After Date and Time are Set
Altitude
The User Configuration screen in the Diagnostic Mode is used to input the
altitude of the location of the ventilator. To enter Diagnostic Mode, follow the
instructions “Entering Diagnostic Mode” on page 5-6. At Figure 5-5 select
USER CONFIG and the screen in Figure 5-10 appears. Pressing the ALTITUDE
button allows the operator or technician to set the altitude to that of the
present geographical location of the ventilator. This factor ensures a more
accurate tidal volume delivery.
Setting Altitude
To adjust the altitude:
1. Verify the altitude using an altimeter, if available, or estimate the
altitude in feet (or meters) above sea level.
2. Press the ALTITUDE button.
3. When the window insert appears, select either feet or meters. Press
the INCREASE or DECREASE button, or rotate the control knob, to
change the value. The longer the increase and decrease are touched,
the faster the value will change.
Setup
4. Press ACCEPT to accept the changed value. Press the CANCEL button
to leave the value unchanged.
Window insert
to change the
value of Altitude.
Setup
The operator can have Esprit compensate the volumes delivered in volume
controlled, mandatory breaths, with the tubing compliance volume, by
activating the COMPLIANCE button (Figure 5-11). Circuit compliance is
activated when this button has a white background. When the function is
activated the exhaled volumes reported by Esprit, will also be tubing
compliance compensated.
Backup Battery If the backup battery is installed, the operator can have the Esprit confirm the
backup battery is connected each time that the machine powers on. Pressing
the BKUP BATTERY button (Figure 5-12) allows this confirmation feature,
identified by an active button with a white background.
Backup Battery
(confirm at
startup)
button active
From then on when the machine powers on it searches for a backup battery. If
the backup battery is connected to the ventilator the startup is normal. If the
backup battery is not connected then the ventilator displays a message at
startup and a 5002 Diagnostic Code will be logged.
Setup
Press the OK button to clear the message from the screen. You may continue
to use the ventilator without the backup battery or power off the ventilator and
reconnect the backup battery.
WARNING: Always turn the ventilator power OFF before connecting additional
equipment.
To turn this feature off, enter Diagnostic Mode and deactivate the messaging
feature by pressing the Bkup Battery button.
1. Power off the ventilator.
2. Power on the ventilator while holding down the ALARM RESET and
100% O2 keys for approximately 5 seconds.
3. A message appears on the ventilator screen asking the user to ‘Verify
that the patient is disconnected prior to proceeding.’ Press OK to enter
Diagnostic Mode.
4. Once you’ve entered Diagnostic Mode, select USER CONFIG button
which takes you to the User Config screen.
5. Press the BKUP BATTERY button. The background color should return
to blue signifying that this feature has been deactivated.
Extended Self Test Respironics recommends that the user runs Extended Self Test (EST) upon
receipt of the ventilator to ensure that there has been no shipping damage to
(EST)
the system. Respironics also recommends that the user run EST between
patients to verify the overall functional integrity of the ventilator. Refer to
chapter 11 for instructions on running EST.
If EST was run successfully and all configuration information has been
entered, Esprit is ready to be used. Follow the recommended “Preoperational
Procedure” on page 8-29 once a patient has been selected.
To exit the Diagnostics Mode you must turn the Power On/Off switch OFF ( )
then ON ( ).
Setup
WARNING: Always turn the ventilator power OFF before connecting additional
equipment.
Communication
Interface CAUTION: To prevent the risk of excessive leakage due to external equipment being
connected to Esprit via the communication ports, a means for external
separation of the conductive earth paths must be provided.
CAUTION: All equipment used and connected to the Esprit communications ports
(analog, parallel and serial) must comply with the medical electrical
equipment (IEC601-1) or other applicable standards.
The Esprit Ventilator provides three communications interfaces: one serial RS-
232 port, an analog output port, and a parallel port. (The Parallel port is
reserved for use by the Communications Option.)
Parallel Analog
Remote Alarm
AC Power Cord
Serial
Circuit Breaker
(cover not shown)
Humidifier AC
Outlet
Connecting Serial The serial port on Esprit is designed to transmit data on a one device to one
Communications device serial communications channel. In the connection between the two
devices, Esprit assumes the “slave” role and responds to commands
Devices transmitted to it via the serial port by the external “master”. The serial
communications port uses a standard RS-232, null modem, pin configuration.
Esprit assumes the serial communications is set-up for:
• 19,200 bits/second baud rate
• 8 data bits
• no parity bit
• 1 stop bit
Esprit is sent commands that are 4 ASCII characters from the external device
and responds with a fixed format message. The commands and the responses
are specified in A, “RS-232 Communications Protocol”.
Connecting Remote The Esprit Ventilator is equipped with a remote alarm port allowing ventilator
alarm conditions to be sounded at remote locations away from the ventilator.
Alarm Port
Pressing ALARM SILENCE deactivates the remote alarm. The ventilator signals
an alarm using a normally open or normally closed relay contact. The de-
energized state of the relay represents an alarm state (any Medium or High
Priority alarm) and the energized state represents no alarms.
The remote alarm port is a standard ¼ inch, female, phono jack (ring, tip,
sleeve) connector.
Remote alarm
connector & cable
NO Ring
NC Tip
Common Sleeve
The port is configured to work with the Normally Open (NO), Normally Closed
(NC), and Respironics (LifeCare) systems. Each requires specific cabling
identified in Table 6-1:.
WARNING: Use only Respironics approved cables when connecting to the remote alarm
port. Be sure to fully insert the cable into the remote alarm port and into the
remote alarm.
WARNING: When using the Remote Alarm Port be sure to fully test the Remote Alarm
Port and cable by:
• Verifying that annunciated alarms on the Esprit are also
annunciated on the remote alarm.
• Verifying that disconnecting the cable from the Remote Alarm port
results in an alarm notification at the Remote Alarm.
• Verifying that disconnecting the cable from the remote alarm
results in an alarm notification at the Remote Alarm.
CAUTION: To prevent damage to the remote alarm, relay contacts should not
exceed the maximum rating of 125VAC, 0.5 amp or 24VDC, 1 amp.
Connecting Respironics recommends only those humidifiers that comply with ISO 8185,
for use on Esprit.
Humidifier
When connecting a humidifier to the patient circuit, the operator should follow
the set-up procedure supplied by the humidifier manufacturer. The following
steps should be followed to electrically connect the humidifier to Esprit (for
100-120 VAC ventilators). For ventilators using voltage other than 100-120
VAC, the humidifier must be connected to another AC outlet.
1. Ensure that the Esprit Ventilator is properly positioned on a secure
table top, wall mount, or pedestal.
2. Remove the cover of the humidifier AC power outlet on the back of
Esprit (Figure 6-3).
3. Connect the humidifier AC plug to the humidifier AC power outlet
(Figure 6-4).
(Circuit Breaker
Cover not shown)
Humidifier AC
Outlet Cover
(Circuit Breaker
Cover not shown)
Humidifier AC
Plug
WARNING: AC power is applied to the humidifier from the Esprit humidifier outlet (only
available on 100-120 VAC ventilators). Under no circumstances does the
Esprit Ventilator provide control for the humidifier. To ensure patient safety,
it is important that any humidifier used with the Esprit Ventilator include an
acceptable temperature control and monitoring mechanism, as well as a
temperature display and appropriate alarm capabilities (refer to ISO 8185).
Connecting the 1. The humidifier should be connected between the Esprit inspiratory
Patient Circuit bacteria filter (Figure 6-5) and the inspiratory limb of the patient
circuit which leads to the patient wye (Figure 6-6).
2. Follow steps illustrated by Figure 6-5, Figure 6-6 and Figure 6-7.
Figure 6-5: Ventilator Gas Outlet Port to Humidifier Patient Circuit Connection
Connecting the The analog output port adds a second remote alarm output. Pin 12 (fourth pin
Analog Port from the top left is pin 12) signals an unsilenced high or medium urgency
alarm:
• 0 VDC = active alarm
• 1.5 VDC = no alarm or silenced alarm
Analog
Port
(Circuit
breaker cover
not shown)
Introduction This chapter describes the Esprit Ventilator’s breath delivery capabilities. It
includes a system overview and descriptions of the ventilation modes and
available breath types. For descriptions of button settings and general
operating instructions, refer to Chapter 8, “Operating Instructions”.
Ventilator Breath The Esprit Ventilator provides the following ventilation breath types:
Types • Volume Control Ventilation (VCV) – invasive ventilation
• Pressure Control Ventilation (PCV) – invasive ventilation
• Non-Invasive Ventilation (NPPV) – non-invasive ventilation
Operating Theory
Ventilation Modes In Volume Control Ventilation and Pressure Control Ventilation, the operator
Common to VCV and can select between ventilation modes of Assist/Control, SIMV and CPAP.
PCV
Operating Theory
SIMV Logic
Current Next
SIMV State Inputs Ventilator Response SIMV State
Mand Time Trigger Deliver mandatory breath using operator Mand
Window (breath period settings for mandatory breath type; Restart Window
timer expires) breath period
Mand Patient Trigger; Deliver mandatory breath using operator Spont
Window Operator Trigger settings for mandatory breath type Window
Spont Patient Trigger Deliver a spontaneous breath using Spont
Window operator settings for spontaneous breath Window
type
Spont Operator Trigger Deliver a mandatory breath using operator Spont
Window settings for mandatory breath type Window
Spont Time Trigger Restart breath period timer Mand
Window (breath period Window
elapses)
Operating Theory
Operating Theory
Operating Theory
Operating Theory
Emergency Modes of The Esprit ventilator has two emergency modes of ventilation that are entered
in response to certain alarm conditions. They are:
Ventilation
• Apnea Ventilation
• Safety Valve Open
Apnea Ventilation
Apnea ventilation provides an emergency mode of ventilation if the ventilator
does not deliver a breath for an operator set interval of time. The apnea time
can be set between 10 and 60 seconds. Upon entering this mode of
ventilation, Esprit will alarm and immediately start using the Apnea Rate
setting specified by the operator. In PCV and VCV, Esprit will begin delivering
breaths in Assist/Control (A/C), but with the operator set Apnea Rate. In NPPV,
Esprit will deliver only machine controlled breaths either at the operator set
Apnea Rate or in response to patient effort. In Apnea ventilation, the alarms
used are the ones used for machine controlled breaths for the ventilation
breath type (VCV, PCV and NPPV) Esprit was using when Apnea occurred.
Operating Theory
Table 7-2: Parameters Used in Apnea Ventilation, Settings and Alarm Limits
The ventilator will reset out of Apnea Ventilation if the operator presses the
Alarm Reset button, or if the patient triggers two successive breaths.
During SVO:
1. The safety valve is opened,
2. The exhalation valve is opened,
3. The air and oxygen valves remain closed,
4. A high priority alarm is activated,
5. The Safety Valve Open indicator is illuminated,
6. The Normal indicator is turned off.
Esprit Overview
WARNING: Ensure that an alternative means of ventilation (i.e., a resuscitator or similar
device) is available while the Esprit Ventilator is in use on a patient.
WARNING: The Esprit Ventilator complies with the requirements of IEC 601-1-2 (EMC
collateral standard), including the E-field susceptibility requirements at a
level of 10 volts per meter. However, even at this level of immunity, certain
transmitting devices (cellular phones, walkie-talkies, etc.) emit radio
frequencies that could disrupt ventilator operation if operated in a range too
close to the ventilator.
Each has mandatory or machine controlled breaths and each has patient
controlled or spontaneous breaths.
Each ventilation breath type has its own settings that are mutually exclusive
from the other ventilation modes.1
• In VCV, the operator can select either A/C, SIMV, or CPAP mode
• In PCV, the operator can select either A/C, SIMV, or CPAP mode
• In NPPV, the operator can select Spont/T or Spont mode
Alarms are specific to the ventilation breath type. Alarm limits in one
ventilation breath type are mutually exclusive from the alarm limits of the
other ventilation breath types.
Esprit is easy to use because all mode settings and alarm limits are selected
using the same three step process:
1. Select the parameter to be changed by pressing the associated button.
The screen insert shown in Figure 8-1 appears.
1. Set O2 is an exception that can be changed only in the active mode and applies to all
modes.
Operating Instructions
NOTE: All volumes entered into Esprit are assumed to be BTPS (Body
Temperature atmospheric Pressure Saturated (with H2O)) volumes
unless otherwise noted. All volumes reported by Esprit are reported as
BTPS volumes. All pressures are assumed to be relative to atmospheric
pressure unless otherwise noted.
The Front Panel The Esprit Ventilator’s front panel includes the indicators and controls shown
in Figure 8-2.
Operating Instructions
Indicators
Touch
Display
Controls
Operating Instructions
Alarm High: flashes red to indicate that a high priority alarm condition exists.
Vent Inop: illuminated red when active. Indicates that the Esprit Ventilator is
not capable of supporting mechanical ventilation and requires service. During
Vent Inop, the ventilator opens the safety valve to allow the patient to breathe
room air spontaneously. The ventilator also discontinues detection of new
alarm conditions during Vent Inop.
WARNING: Vent Inop is a serious condition, which is indicated by both visual and
audible alarms. If the ventilator is attached to a patient when Vent Inop
occurs, the patient must be supported with another means of life support
ventilation.
Safety Valve: illuminated red when active; indicates that the safety valve is
open and the ventilator is not in operation. This is a high priority alarm
condition. When the Safety Valve indicator is illuminated, the ventilator is not
providing ventilatory support to the patient. The safety valve opens to allow the
patient to breathe spontaneously through the ventilator circuit. The patient
must be capable of creating a spontaneous breath in order to breathe through
the safety valve.
Alarm Silence: illuminated yellow when the audible alarm has been
disabled; is active and stays on for 120 seconds when the ALARM
SILENCE button is pressed. If Alarm Silence is active, and a new alarm
condition occurs, which involves exceeding an active alarm limit, the visual
alarm functions will be active. (Refer to “Alarm Silence” on page 9-2.) ALARM
RESET clears ALARM SILENCE. If a medium or high priority alarm exists after
ALARM RESET clears ALARM SILENCE the audible alarm will begin.
100%O2: illuminated green, and is active only when the 100% O2 front panel
button has been pressed and the ventilator is delivering 100% O2 to the
patient; remains on for the duration of 100% O2 delivery (two minutes).
NOTE: If the operator sets the %O2 setting to 100%, the 100% O2 indicator
does not light. The 100% O2 indicator only lights when the 100% O2
front panel key has been pressed.
Operating Instructions
disabled until the touch screen is unlocked. MANUAL BREATH, 100% O2,
EXP. HOLD, ALARM RESET, and ALARM SILENCE keys are still active keys.
Battery In Use: illuminated yellow when active; indicates that the ventilator is
running on Backup Battery power. The Backup Battery is used when there is no
AC power and no other battery is available.
Battery Charging: illuminated yellow when active; indicates that the optional
backup battery is charging. The battery should not be considered a fully
charged power source when this indicator is illuminated. The Battery Charging
indicator will stay on for the duration of the charging cycle, which can last for
up to ten hours. Once the battery has fully charged, the Battery Charging
indicator will go off.
Battery Low: illuminated flashing red when the backup battery has
approximately 5 minutes power remaining.
WARNING: When the battery low indicator is flashing red, operation of the ventilator
from battery power should be discontinued.
NOTE: The Esprit Ventilator selects its power source based on the following
prioritization: AC power (if present), external battery, then backup
battery.
ALARM SILENCE: disables the audio alarm for 120 seconds. The alarm silence
period ends after the 120 seconds has elapsed. If Alarm Silence is active and
an alarm takes place that involves an operator alarm limit being exceeded, the
Operating Instructions
alarm indicator is illuminated, but Alarm Silence will continue to be active and
the Alarm Silence timer will continue to time out. Alarms that can not be
silenced are listed in Table 9-1: “Alarm Alert Messages” on page 9-4. When
the ALARM SILENCE button is pressed, the Alarm Silence indicator is
illuminated for the duration of the 120 second alarm silence period.
ALARM RESET: Used to clear the visual indicator for auto reset alarms, certain
active alarms (Refer to Chapter 9, “Alarm Reset”.) and reset out of apnea
ventilation back to the active mode of ventilation (Refer to “Apnea Ventilation”
on page 8-24.) Alarm reset also terminates ALARM SILENCE.
NOTE: The ventilator may automatically reset certain types of alarm conditions
once the causes of the alarms are corrected. After an automatic reset,
the ventilator will clear the audible alarm, and will display a Low
Urgency Alarm alert in the Alert Message Insert to inform the operator
that an alarm condition existed. When this situation occurs, use ALARM
RESET to clear the visual alarm indicator.
100%O2: Delivers 100% O2 to the patient for two minutes. Subsequent button
presses will reset the timer to two minutes. The delivery of 100% O2 cannot be
cancelled during the two minute delivery period.
NOTE: If the 100% O2 key is pressed and a 100% O2 gas source is not
available, the Low O2 alarm will be active for the two-minute 100% O2
delivery period.
NOTE: Manual breaths are not permitted during the inspiratory phase of a
breath (whether manual or spontaneous). Pressing the MANUAL
BREATH key during these times will not result in the delivery of a
manual breath.
SCREEN LOCK: used to lock and unlock the graphic display (touch screen).
When the screen lock is activated, all on-screen buttons are disabled. This
function is available to prevent inadvertent setting and display changes via the
touch screen.
Operating Instructions
ADJUST CONTROL KNOB: used in conjunction with the front panel graphical
display and touch screen, to enter operator-selected values for ventilator
settings and alarms.
ON/OFF SWITCH: is located in the lower portion of the front panel, and is
recessed to avoid inadvertent or accidental access. When the switch is in off
( ) position, the ventilator does not provide mechanical ventilation, although
AC power is active and the green Mains indicator is illuminated.
Operating Instructions
Top Bar
Manometer
Bottom Bar
Figure 8-5: Elements common to all operational screens (VCV settings shown)
Figure 8-6 shows the top bar and describes its buttons.
Active Mode: push to display the Active Alarms: push to show current alarm
settings for the active ventilation mode settings for the active ventilation mode and
and breath type. breath type.
Figure 8-7 shows the bottom bar and describes its buttons.
Operating Instructions
VCV Settings: push to display VCV NPPV Settings: push to display NPPV Respiratory Graphics
settings screen, which allows you to settings screen, which allows you to Mechanics button: selects
view and change settings for volume view and change settings for button: selects option if
controlled ventilation noninvasive positive pressure option if installed.
ventilation installed.
PCV Settings: push to display PCV settings Option: for use in the Printer: Available if
screen, which allows you to view and future Communications
change settings for pressure controlled option is installed
ventilation
Buttons in the bottom bar have two states, “selected” and “not selected”. In
the “selected” state, the button has a white background and black letters (see
VCV SETTINGS in Figure 8-7). In the “not-selected” state the button has a
gray background and black letters (see PCV SETTINGS, and NPPV SETTINGS
in Figure 8-7). In the upper bar the buttons MONITOR, ALARM SETTINGS and
PATIENT DATA are also the “selected/not-selected” type.
Operating Instructions
Upper limit of
real time display High pressure
determined by alarm setting -
High Pressure press this button
alarm setting to modify HIP.
Zero pressure
line
Breath
Indicator
Breath Indicator
Breath Symbol Description
Mand Operator or ventilator triggered mandatory breath.
Plateau Inspiratory hold, can be set at the end of the inspiratory phase of a VCV
breath type.
Support Patient triggered spontaneous breath with PSV>0 or IPAP>EPAP.
Operating Instructions
Pressing the HIP indicator on the screen (Figure 8-8) will immediately
HIP allow the operator to modify the high pressure limit for VCV or PCV. In
NPPV, the high pressure limit is automatically set to 10 cmH2O (hPa)
above IPAP. The HIP indicator does not appear near the manometer in NPPV.
Settings Screens The buttons in the middle of the settings and alarm limit screens all have two
states, active and inactive. Active settings have a gray background with black
letters (Figure 8-9). An active setting is currently being used by Esprit to
control ventilation or as an alarm limit. Inactive settings buttons have a gray
background with gray letters. Inactive settings are not currently being used by
Esprit to control ventilation or as an alarm limit. In both states the button can
be pressed and an insert window will appear that will allow the operator to
change the value of the setting (Figure 8-1).
NOTE: Some settings buttons appear active despite the fact they are not being
used in the ACTIVE MODE. This is because the setting is used in Apnea
Ventilation or when manual inspiration is pressed. The operator should
always choose a value for an active button that is appropriate for the
patient being ventilated.
Pressing the ACTIVE MODE button displays the settings screen for the
currently selected ventilation breath type and mode (Chapter 7, “Operating
Theory”, for more information on ventilation modes, breath types, and
controls).
For example, in Figure 8-9 the current breath type is VCV and the mode is A/C.
Because A/C does not allow the ventilator to deliver any spontaneous breaths,
any settings that apply to spontaneous breaths are not active and are grayed
out. In Figure 8-9, PSV is grayed out, indicating that it isn’t active, but you
can still push this button and change the PSV value. The PSV value will
become active only if the SIMV or CPAP mode is selected.
Operating Instructions
Selecting Settings
Follow these steps to adjust ventilator control settings:
1. Push VCV SETTINGS, PCV SETTINGS, or NPPV SETTINGS.
2. Push the button for the control settings you want to select. The
ventilator displays the current value for that parameter in an insert
that allows the operator to adjust the settings value. (Figure 8-10).
3. Press the bar to increase or decrease the setting to the value you want
or use the front panel knob to adjust the value of the setting.
4. Press the screen ACCEPT button or the front panel ACCEPT button to
activate the new setting or CANCEL to leave the setting unchanged.
Operating Instructions
Press ACCEPT to
activate the new
setting
Press CANCEL to
leave the setting
unchanged
When a value is entered, as shown in Figure 8-10, Esprit checks to assure that
the value has been accepted by the operator, is within limits and will not cause
other settings to be out of limits. If the new value causes these limits to be
exceeded a diagnostic message will be displayed. (Refer to “Value Entry
Message” on page 12-4 for more details about the diagnostic messages).
Operating Instructions
Active
ventilation
mode and
breath type
button
displayed
Indicates this
screen is for
VCV alarms
Push
button of
alarm you
want to
adjust
The High Pressure Alarm Setting may be accessed through the Alarms Settings
Screen or through the HIP Indicator adjacent to the manometer in the PCV and
VCV Settings, Alarm Settings, Patient Data and Monitor Screens.
WARNING: For patient safety the HIP Limit Setting should be set as close to the peak
inspiratory pressure as patient conditions allow.
CAUTION: If clinical conditions do not require setting the HIP Limit above 60
cmH2O, it is recommended the setting normally be adjusted to 60
cmH2O or less in order to prolong the operating life of the blower and to
maximize backup battery run time.
NOTE: When the active mode is set to NPPV, the HIP Limit Setting will
automatically be adjusted to 10 cmH2O above the current IPAP setting.
Follow these steps to adjust alarm limits that are not currently active:
1. Depending on the alarm limits you wish to review, push VCV
SETTINGS, PCV SETTINGS, or NPPV SETTINGS button that has a
gray background and black letters (not active).
Operating Instructions
ACTIVATE
button
Figure 8-12: PCV Setting while NPPV is the active breath type
3. Select the REVIEW ALARMS button and Esprit will display the alarm
settings for PCV (Figure 8-14).
4. Notice that all the alarm buttons (Figure 8-14) are grayed out,
indicating that the alarm limits are not currently active. The ventilator
displays the current value for that set of alarm limits. Push the button
for the alarm limit you want to adjust.
NOTE: Esprit keeps a distinct set of alarm limits for each ventilation breath
type (VCV, PCV and NPPV).
5. Press the bar or use the front panel knob to adjust the value (as shown
in Figure 8-10).
6. Press ACCEPT to activate the new alarm limit or CANCEL to leave it
unchanged.
Operating Instructions
Grayed out buttons indicate that the alarm limits are not active
Push button of
alarm setting to
adjust
Figure 8-14: Setting alarm limits that are not currently active
NOTE: Any of the changes made in the screen shown in Figure 8-14, do not
take effect until the operator switches to the new ventilation breath type
(in this case Pressure Control).
Operating Instructions
Operating Instructions
When the settings window insert being displayed is VCV Rate, the calculated
minute volume that results from the rate and tidal volume settings, is
displayed along with the rate value. This is also the case when the tidal volume
setting window insert is displayed. When the setting window insert for PCV
Rate is displayed, the calculated value for the I:E ratio that will result from the
PCV rate value is also displayed.
When the setting window insert for VCV Apnea Rate is displayed, the
calculated minute ventilation that will exist if the ventilator goes into Apnea
Ventilation in VCV is shown. When the setting window insert for PCV Apnea
Rate is displayed, the I:E Ratio shown is the I:E Ratio that will exist if the
ventilator goes into Apnea Ventilation in PCV.
Operating Instructions
Setting to be
changed on the
top.
Setting that is
calculated as
the setting
above changes,
below.
Operating Instructions
Selecting a New Follow these steps to select the ventilation breath type:
Ventilation Breath 1. Press the button for the new breath type you want to select from the
bottom bar (Figure 8-20) [PCV in our example].
Type (VCV, PCV, or
NPPV)
2. The setting screen for the selected breath type appears. All of the
settings are grayed out, indicating that they are not currently active.
However, you can use the grayed out buttons to change the settings as
necessary, (with the exception of Set O2, which can be changed only
in the active mode).
For example, if you press PCV SETTINGS when the current breath type
is NPPV, all the PCV settings are grayed out (Figure 8-21) because
NPPV is currently active.
Operating Instructions
ACTIVATE
button
Figure 8-21: Settings Screen (PCV settings, PCV not currently active)
Figure 8-22: Prompt when Activate is pressed and PCV settings selected
Operating Instructions
Press to activate
new breath type.
Operating Instructions
6. If the operator reviews alarm settings and then activates the new
breath type, Esprit uses the new breath type and displays the alarm
settings of the new breath type (Figure 8-25).
ALARM
Indicates the active SETTINGS
breath type alarms button is
selected
indicating the
alarm settings
shown are for
Alarm limits no the active breath
longer grayed out, type
indicating that the
selected breath type
is active
Operating Instructions
Figure 8-26: Ventilation Mode buttons (A/C and Spont/T active as shown)
2. The ventilator asks you to confirm the mode change (Figure 8-27).
Press YES to confirm, or press NO to leave the mode as is.
3. Once you confirm a mode change, the ventilator changes the mode
selected to active (Figure 8-28).
Figure 8-28: Ventilation Mode buttons (SIMV and Spont shown as active)
Apnea Ventilation Apnea ventilation provides an emergency mode of ventilation if the ventilator
does not deliver a breath for an operator set interval of time. The apnea time
can be set between 10 and 60 seconds. Upon entering this mode of
ventilation, Esprit will immediately start using the Apnea Rate setting
specified by the operator. In PCV and VCV, Esprit will begin delivering breaths
in Assist/Control (A/C), but with the operator set Apnea Rate. In NPPV, Esprit
will deliver only machine controlled breaths either at the operator set Apnea
Rate or in response to patient effort. In Apnea ventilation, the alarms used are
the ones used for machine controlled breaths in the active breath type (VCV,
PCV and NPPV) Esprit was in, when Apnea occurred.
Operating Instructions
Patient Data Screen You can view the patient data screen (Figure 8-29) by pressing the PATIENT
DATA button in the top bar. The format of this screen is the same in VCV, PCV,
and NPPV. Table 8-3: “Patient Data Definitions: Range, Units & Resolution”
summarizes the definitions, display ranges, and resolution for patient data
parameters.
Table 8-3: Patient Data Definitions: Range, Units & Resolution (Sheet 1 of 2)
Operating Instructions
Table 8-3: Patient Data Definitions: Range, Units & Resolution (Sheet 2 of 2)
NOTE: Pt. Leak only appears on Patient Data block on Settings screen.
Monitor Screen The monitor screen (Figure 8-30) is the default screen, and is automatically
displayed if the screen has not been touched for 15 minutes. The screen can
also be viewed by pressing the MONITOR button. The screen is shown in
Figure 8-30. The patient data and settings displayed on this screen depend on
the active breath type (Refer to Table 8-4: “Monitor Screen Settings
Displayed” and Figure 8-30.)
Operating Instructions
Operating Instructions
Pressure
versus time
waveform.
While the EXP HOLD button is held, the exhalation valve and the air and O2
valves will be closed. During this period there is no flow of gas into or out of
the patient circuit from Esprit. The operator can monitor the airway pressure by
watching the EXHALATION PAUSE PRESSURE digital display and by watching
the pressure versus time waveform. When the operator determines the airway
pressure has stabilized, the EXP HOLD button should be released. If the
button is held for 5 seconds, Esprit will automatically terminate the maneuver
and perform the calculation if stability was achieved. Esprit then calculates
Auto-PEEP as shown in the equation below:
• Auto-PEEP = Exhalation Pause Pressure - End Exhalation Pressure
NOTE: If the EXP HOLD key is held continuously and the exhalation hold
period exceeds 5 seconds, Esprit automatically terminates the
exhalation hold period and begins a new inspiratory period.
Operating Instructions
Preoperational The preoperational procedure verifies that the ventilator is ready for use on a
Procedure patient.
NOTE: All components of the patient circuit must not have leaks in order to
pass SST.
Operating Instructions
7. Push USER CONFIG to check time, time format, date and altitude and
set compliance compensation as required (Figure 8-33).
8. Press START SST to begin Short Self Test, and follow the screen
prompts. Use a cap to plug the patient port of the patient circuit wye
at the prompt.
9. During SST, as each test is performed, the ventilator displays test
results (Figure 8-33). Do not proceed until the ventilator completes
SST without failures.
Push USER
CONFIG to set
time, date,
altitude,
compliance
compensation
enable, and time
format
Any failures
detected during
SST are listed
here
WARNING: The operator will be warned if the compliance is 9.0 ml/cmH2O (hPa) or
larger. Patients should not be put on a patient circuit that does not meet this
requirement.
Operating Instructions
Altitude set
Backup Battery
(confirm at startup)
enable/disable
10. Once you have reviewed all the parameters on the User Config screen,
the ventilator is ready for patient use. Turn the ventilator off, then on,
without holding down any front panel keys.
11. Ventilator settings from the previous use are in effect at power up.
Select appropriate settings for the next patient as described in this
chapter.
Alarm Testing A procedure is available if the operator wants to test the operation of alarms.
Respironics recommends following the preoperational procedure. (Refer to C,
Procedure
“Alarm Testing Procedure”.)
Where To Go For Help For clinical or technical support, contact Respironics Customer Service at
1-800-345-6443.
Operating Instructions
Introduction The Esprit ventilator provides an easy-to-use hierarchical alarm system that
includes both visual and audible alarms. When the Esprit ventilator detects an
operating condition that requires attention, it generates an alarm. The alarm
system communicates three levels of urgency and priority:
• High Urgency: Alerts the operator that immediate response is
required. (red flashing indicator)
• Medium Urgency: Alerts the operator that prompt response is
required. (yellow flashing indicator)
• Low Urgency: Alerts the operator to a change in the ventilator status.
(yellow continuous indicator)
In most cases, the alarm will have the following audible and visual
components:
• an indicator is illuminated
• a sequence of tones sounds
• a screen alert window appears with a message in it
Visual Alarms The Esprit ventilator includes alarm and status indicators located above the
front panel to provide a visual summary of active alarm conditions. Each of
these indicators is illuminated by either a red, green or yellow light, which will
flash or remain lit, depending on the alarm condition.
Pressing the ALARM RESET button in the lower left corner of the front panel
clears the visual indicators for active or auto-reset alarms. If the alarm
condition reoccurs, the visual indicator turns on again.
Alarm messages also appear in an alarm alert insert that appears in any screen
whenever there is a low, medium or high urgency alarm active (Figure 9-2).
Alarms
Audible Alarms When an alarm condition exists, the Esprit Ventilator will generate a sequence
of audible tones to alert the operator. The sequence varies according to the
urgency and priority level of the alarm:
• High Urgency: The ventilator emits a repeating sequence of five tones.
• Medium Urgency: The ventilator emits a repeating sequence of three
tones.
• Low Urgency: No audible tone emitted. (Med/Low indicator
illuminates and alarm messages appear in the front panel Alerts
window)
When more than one alarm is active, only the highest urgency level alarm is
audibly enunciated. Procedures for silencing audible alarms are described
below.
Alarm Silence
To silence alarms, perform the following:
• Press the ALARM SILENCE button at the bottom left corner of the
Esprit front panel. The audible tone will cease and the alarm silence
indicator will light and remain lit for two minutes. Specific alarms that
cannot be silenced are listed in Table 9-1: “Alarm Alert Messages”.
When ALARM SILENCE is pressed before the end of the two-minute period,
the two-minute timer is reset, and alarm silence begins anew. Pressing ALARM
SILENCE multiple times does not give multiple two-minute silence periods.
If new alarms occur during alarm silence, all visual alarm enunciation
continues. If the Alerts insert is not present, it will appear. The message for
any alarms that occur during the alarm silence period will appear in the Alerts
insert if there are fewer than three messages currently displayed or the new
message is higher priority than one of the currently displayed messages. The
highest three priority messages will be displayed.
Alarms
CAUTION: The Esprit alarm indicators and the Alerts insert should be monitored
closely during the Alarm Silence period to ensure that unexpected
alarms are noticed.
For the operator to clear an apnea alarm, the ALARM RESET button may be
pressed. Once it has been pressed, the ALARM RESET will cancel an existing
apnea alarm and return ventilation to the active mode. If the condition that
triggered the apnea alarm persists after the ALARM RESET button has been
pressed, the apnea alarm will re-trigger. If the patient triggers two successive
breaths in apnea ventilation, Esprit will automatically reset out of apnea
ventilation.
Alarms
Alarm Enuciation
Volume Control Knob
The volume control can be adjusted between a minimum and maximum to suit
the particular clinical situation.
Alert Messages When the ventilator detects an alarm condition, it presents an Alert insert
below the Patient Data box in the middle of the screen.
While the alarm is active (i.e. visual and audible alarms are still present), the
message in the alert insert appears in bold-faced type. Once the alarm
condition has been corrected, the message automatically switches to regular
face type. Pressing ALARM RESET clears alarm messages.
When the operator presses the ALARM RESET button (described above), the
alarm messages disappear from the alert window.
When the ventilator detects more than three alarm conditions, the three
highest priority alert messages will be displayed.
Alert messages, along with the corresponding alarm descriptions, are listed in
Table 9-1: “Alarm Alert Messages”.
Alarms
Note: For Esprit Ventilators with software earlier than version 7.10,
the High Minute Volume alarm is available in Pressure Control only.
High O2 A high urgency alarm indicates that the monitored O2 concentration
is at least 6% above the set value (%O2) for 30 seconds. Verify
operation of the O2 sensor. Alarm cannot be manually reset.
High Respiratory Rate The ventilator triggers a medium urgency alarm if the total
respiratory rate is greater than the operator-set High Respiratory
Rate limit. The ventilator evaluates this alarm condition at the start
of inspiration, after calculating the total respiratory rate including
the just completed breath. It is elevated to high urgency after one
minute.
High Temperature Internal temperature monitor detects higher than allowed
temperatures inside the enclosure. Alarm cannot be silenced or
manually reset.
Alarms
Note: For Esprit Ventilators with software earlier than version 7.10,
the alarm is elevated to high urgency after one minute, if still
active. In CPAP it elevates after two consecutive breaths.
Low Backup Battery If the Backup Battery is low, a high priority alarm will be triggered.
Immediately connect an AC power source to avoid a loss of power.
Provide AC power and/or replace the Backup Battery. Alarm cannot
be silenced or manually reset. 5 minutes or less of battery run time
remains.
Low EPAP If the Exhalation Positive Airway Pressure is less than the operator
set Low EPAP Pressure limit for one second, the ventilator signals a
medium urgency condition. If the Low EPAP alarm remains active
for one minute, the ventilator elevates the alarm to high urgency.
Note: A Low EPAP setting of zero (0) will disable this alarm.
Note: A Low Minute Volume setting of zero (0) will disable this
alarm.
Low O2 A high urgency alarm indicates that the monitored O2 concentration
is 18% or at least 6% below set value for 30 seconds. Verify
operation of the O2 sensor. Alarm cannot be manually reset.
Low O2 Supply A high urgency alarm indicates that O2 gas supply is below
acceptable levels and the %O2 setting is above 21%. Check O2 gas
connections and inlet filter. Alarm cannot be manually reset.
Alarms
Note: A Low PEEP setting of zero (0) will disable this alarm.
Note: A Low Tidal Volume setting of zero (0) will disable this alarm.
Low Vt Mandatory The ventilator triggers a medium urgency alarm to indicate that the
VCV or PCV mandatory tidal volume is less than the set limit. It is
elevated to high urgency after one minute.
Alarms
Alarm Indicators
Normal
The Normal indicator, located in the upper left-hand corner above the front
panel, remains lit with a steady green light as long as there are no active or
auto reset alarm conditions present.
Alarm High
The Alarm High indicator, to the right of the Normal indicator, visually
indicates a high priority alarm. When a high priority alarm condition exists, this
indicator flashes red and an audible five tone sequence sounds until the
condition is corrected or reset.
WARNING: A high priority, visual and audible alarm indicates a potentially life-
threatening condition and immediate response is required.
Alarm Med/Low
The Alarm Med/Low indicator, to the right of the Alarm High indicator, visually
indicates a medium or low priority alarm. When a medium priority alarm
condition exists, this indicator will flash yellow and an audible three tone
sequence will be heard until the condition is corrected. When a low priority
alarm condition exists, this indicator remains lit with a steady yellow light,
until the alarm is reset.
Vent Inop
The Vent Inop indicator signals the operator that the ventilator is not capable
of supporting ventilation and requires service. During a ventilator inoperative
condition, the ventilator enters a safe state, where the safety valve is opened
and new alarm condition detection is discontinued. If the ventilator is attached
to a patient when this condition is detected, the ventilator must be replaced
immediately.
WARNING: Vent Inop is a serious condition, which is indicated by both visual and
audible alarms. If the ventilator is attached to a patient when Vent Inop
occurs, the patient must be supported with another means of life support
ventilation.
Alarms
Safety Valve
This indicator signals that the Safety Valve is open and the ventilator is not
providing breath support to the patient. It is accompanied by a five-tone
audible alarm sequence and a Safety Valve Open message in the touch screen
display.
The Safety Valve opens automatically whenever the ventilator is not able to
provide breath support to the patient. It allows the patient to spontaneously
breathe room air through the ventilator system.
The Safety Valve Open condition is normal during start-up and restart, and it
automatically turns off when the start-up sequence is complete. If this
condition occurs at any other time, it cannot be reset. Immediately use an
alternative ventilation source and call for service.
WARNING: When the safety valve open indicator is lit, the ventilator does not provide
any ventilatory support to the patient. Immediately use a backup means of
ventilatory support.
Alarm Silence
The indicator , located in the center of the indicators at the top of the front
panel, will illuminate with a steady yellow light whenever the operator presses
the alarm silence control (located at the bottom left of the front panel). The
indicator remains lit for two minutes (120 seconds) after the Alarm Silence
control is pressed. If Alarm Silence is active and an alarm takes place that
involves an operator alarm limit being exceeded, the alarm will be enunciated
visually, but Alarm Silence will continue to be active and the Alarm Silence
timer will continue to time out. Alarms that can not be silenced are listed in
Table 9-1: “Alarm Alert Messages” on page 9-4. If Alarm Reset is hit at any
time during the Alarm Silence period, the Alarm Silence period is terminated.
If any high or medium priority alarm condition exists after pressing Alarm
Reset, an audible alarm sounds.
WARNING: Visually monitor the patient and ventilator during the Alarm Silence period
to ensure that alarms do not go undetected. Allowing alarm conditions to
continue without intervention may result in harm to the patient and/or
ventilator.
Alarms
General Information Procedures for Esprit Ventilator cleaning, sterilizing and periodic maintenance
must be performed to ensure consistent ventilator operation. Institutional
policies may also provide guidelines, procedures, and schedules for cleaning
and sterilizing equipment before, during and after patient use.
General procedures for cleaning and sterilizing the Esprit Ventilator are
described in the following sections. Some ventilator parts must be
disassembled before cleaning and sterilizing. (Refer to “Removing and
Replacing Internal Air Source Inlet Filter” on page 10-9.)
Cleaning When cleaning parts, avoid the use of hard brushes or other instruments likely
to cause surface damage.
CAUTION: Care should be taken when cleaning the Touch Display. (Refer to Figure
8-2 on page 8-3.). A soft moist cloth should be used that does not drip
water and/or soap solution when in contact with the display. After
cleaning and rinsing with a damp cloth, remove all moisture with a dry,
soft cloth. Never allow solutions of any kind to collect on the bottom
bezel of the display. Never use a brush or device that can cause
abrasion to clean the touch display or its bezel, they will cause
irreparable damage.
Sterilization
WARNING: Do not expose expiratory and inspiratory bacteria filters or reusable patient
tubing to ETO gas.
NOTE: Because conditions and practices in health care institutions vary, this
manual can only describe general guidelines. It is the user’s
responsibility to ensure the validity and effectiveness of the methods
used.
Steam Autoclaving
The O2 Sensor Tee (P/N 1001736) and the O2 Sensor coupling (P/N
1002505) may be steam autoclaved.
1. Ensure that the part has been disassembled, cleaned, and partially
reassembled, as applicable.
2. Separately wrap the part in muslin or equivalent paper wrapper.
3. Steam autoclave according to the autoclave manufacturers
instructions. In many institutions it may be routine to place a
biological indicator in the autoclave load as a subsequent test for
sterility.
4. Aseptically store parts until used.
Chemical Disinfecting
The O2 Sensor Tee (P/N 1001736) and the O2 Sensor coupling (P/N
1002505) may be chemically disinfected.
Ventilator Exterior
Wipe clean with a damp cloth and mild detergent. Do not use liquid or aerosol
bactericide. Do not allow moisture to come in contact with the touch panel
screen or to collect between the keypad and front bezel assembly.
CAUTION: DO NOT allow liquid to penetrate the ventilator rear or front panel. DO
NOT attempt to sterilize the ventilator by exposing to ETO gas. DO NOT
steam-autoclave.
CAUTION: Care should be taken when cleaning the Touch Display. (Refer to Figure
8-2 on page 8-3.). A soft moist cloth should be used that does not drip
water and/or soap solution when in contact with the display. After
cleaning and rinsing with a damp cloth, remove all moisture with a dry,
soft cloth. Never allow solutions of any kind to collect on the bottom
bezel of the display. Never use a brush or device that can cause
abrasion to clean the touch display or its bezel, they will cause
irreparable damage.
Visually inspect tubing for nicks, cuts and holes prior to use with the Esprit
Ventilator.
Bacteria Filters Bacteria filters are typically used on the inspiratory and expiratory port
connections on the front of the Esprit Ventilator. Filter locations are illustrated
below.
The inspiratory filter port is a 22mm standard connector. Use the filter
supplied with the ventilator.
The expiratory port includes a heated exhalation compartment. Use the filter
supplied with the ventilator.
1. Disconnect circuit tubing (1) from inspiratory bacteria filter outlet (2).
2. Disconnect filter (3) from Esprit gas outlet (4). If optional O2 sensor is
installed, ensure it is not dislodged. When used, O2 sensor is located
between gas outlet (4) and filter (3).
3. Insert new filter onto Esprit gas outlet (4) with flow direction indicator
(on the filter) pointing away from the ventilator.
4. Reconnect circuit tubing (1) to new filter outlet (2).
WARNING: The expiratory filter housing may be hot if removed from the ventilator
immediately after use. Wait 15 minutes after turning off ventilator power
before removing the heated expiratory bacteria filter. Exercise caution when
handling the filter housing.
NOTE: Because some environments cause a quicker collection of lint and dust
than others, inspect and clean the fan filters more often than every 250
hours if necessary.
2 1
CAUTION: Do not remove any screws from the cooling fan area. Removing screws
from this area will result in damage to internal components.
5
4
3 2
1 2
Figure 10-6: Removing & Replacing Internal Air Source Inlet Filter
Refer to Figure 10-6. Figure 10-6 shows exploded view of the duct (2) and
filter (1), but it does not have to be disassembled.
1. Move the ventilator so that there is easy access to the underside of the
internal air source duct (2).
2. The internal air source filter (1) can be removed manually. The filter
can be reached from under the internal air source duct (2) as
indicated by the arrow.
3. Wash internal air source filter (1) with mild soap and water.
4. Rinse thoroughly then pat dry.
5. Reinstall internal air source filter (1) from underneath the duct (2) in
the same fashion it was removed.
Storage If you need to store the ventilator for fifteen days or more, ensure that the
altitude, temperature and humidity of the storage site fall within the following
ranges:
• Environmental Temperature: -20 to 60°C (-4 to 140° F)
• Relative Humidity: 10 to 100% noncondensing
• Maximum Altitude: 6,560 m (20,000 ft.)
Repairs For technical service or repairs not included in this chapter, refer to the Esprit
Ventilator Service Manual, P/N 580-1000-02 (English language), or contact
Respironics Customer Service at 1-800-345-6443.
WARNING: The patient must be disconnected from the ventilator before entering the
Diagnostic Mode since normal ventilation is suspended.
NOTE: The “Hardware” function and EST in the Diagnostics Mode should only
be run by qualified personnel.
Diagnostics
Entering Diagnostic To enter Esprit’s Diagnostic Mode, hold down the ALARM RESET and 100%
Mode O2 keys for approximately 5 seconds while you turn on the ventilator. A
message appears on the ventilator asking the user to confirm that the patient
is disconnected before entering Diagnostic Mode (Figure 11-1). Press OK to
enter Diagnostic Mode.
Once you’ve entered Diagnostic Mode, you can select any of diagnostic
functions by pressing its button. Change values in Diagnostic Mode the same
way you change ventilator settings.
WARNING: Do not use a ventilator that has failed SST without verifying operational
readiness by other means. Doing so may place a patient at risk.
WARNING: Never initiate SST while the patient is connected to the ventilator. The high
airway pressures generated during SST can injure a patient.
Diagnostics
Extended Self Test EST verifies the overall functional integrity of the ventilator by testing all
critical hardware subsystems and components. Perform EST between patients
(EST)
as part of preventive maintenance, a performance verification, or if the
operational integrity of the ventilator is in question. EST is typically run by
qualified trained personnel.
WARNING: Never initiate EST while the patient is connected to the ventilator. The high
airway pressures and gas flows generated during EST can injure a patient..
WARNING: Do not use a ventilator that has failed EST without verifying operational
readiness by other means. Doing so may place a patient at risk.
Diagnostics
Hardware
The Hardware function allows a trained service technician to operate critical
components separately to help identify a faulty component in the event that
the ventilator fails SST, EST, or performance verification testing. For more
information on the Hardware function, see the Esprit Ventilator Service Manual
(P/N 580-1000-02).
Software
The Software function displays the ventilators serial number, software part
numbers and version numbers and the part numbers of other critical
components.
User Config
The User Config function allows you to:
• Set the date and time when first setting up the ventilator.
• Set the altitude for the location of the ventilator. This setting allows
for more accurate tidal volume delivery.
• Enable or disable the automatic patient circuit compliance
compensation feature. Compliance compensation corrects the
delivered volumes of VCV mandatory breaths for patient circuit
compliance. It also corrects all exhaled volume for patient circuit
compliance volume. The COMPLIANCE button is grey when
compliance compensation is disabled and white when compliance
compensation is enabled.
• Set the time format (AM/PM or 24 hour).
• Enable or disable the confirmation that the backup battery is
connected each time that the machine powers on.
Diagnostic Codes
The Diagnostic Codes function allows you to review the diagnostic log in
ventilator memory.
To view the diagnostic codes, press the DIAGNOSTIC CODE button on the
diagnostic screen. The ventilator displays the following information:
• Number: Diagnostic codes are numbered in reverse order of
occurrence (most recent first).
Diagnostics
Diagnostic codes, associated with EST and SST are only recorded the first time
the failure is encountered.
The log holds the last 20 diagnostic codes. The screen can only display 10
codes at a time. Press NEXT PAGE button to see the next group of codes or
PREV PAGE to view the previous group.
WARNING: If any of the diagnostic codes, other than diagnostic code 1, 3, 2000, 3000,
5000, 5001, 5002, 8003 , or 8004 occur, remove the ventilator from
service and contact trained service personnel.
WARNING: Use of an Esprit ventilator that has not passed SST or EST is against the
strongest recommendation of Respironics.
Diagnostics
Code Description
1 Normal mode startup
3 Diagnostic Startup
Diagnostics
Code Description
3110 Check Valve 3 Leak
3111 O2 Not Connected
3112 O2 Not Disconnected
3113 FiO2 Sensor Sample Out of Range
3114 FiO2 Sensor Average Out of Range
3115 Primary Audio Not Sounding
3116 Backup Audio Not Sounding
3117 Crossover Circuit Fault
3118 Blower Off Switch Failure
3119 Blower DAC Failure
3120 Pressure Relief Valve Cracking Pressure Too High
3121 Pressure Relief Valve Cracking Flow Not Stable
3122 Pressure Relief Valve Cracking Pressure Too Low
3123 O2 Valve Cracking Flow Outside Range
3124 O2 Valve Full Flow Outside Range
3125 Air Flow Sensor/Exh Flow Sensor Disagreement or Inhalation Pressure/Exhalation
Pressure Disagreement
3126 O2 Flow Sensor/Exh Flow Sensor Disagreement
3127 Heated Filter Backpressure Out of Range
3128 Circuit Compliance Out of Range
3129 Pressure Leak Out of Range
3130 Safety Valve Cannot Open
3131 Patient Wye Not Blocked
3132 keyboard Failure
3133 Rotary Knob Failure
3134 Cannot Calibrate Air Flow Sensor
3135 Cannot Calibrate O2 Flow Sensor
3136 Cannot Calibrate Exh Flow Sensor
3137 Verify Failure—Air Flow Sensor Cal
3138 Verify Failure—O2 Flow Sensor Cal
3139 Verify Failure—Exh Flow Sensor Cal
3140 Cannot Erase Flow Sensor Tables
3141 Cannot Open Inh Autozero Solenoid
3142 Cannot Open Exh Autozero Solenoid
3143 Air Step Position for Open—Outside Range
3144 Air Step Position for Midpoint—Outside Range
Diagnostics
Code Description
3145 Air Step Position for Close—Outside Range
3146 O2 Step Position for Open—Outside Range
3147 O2 Step Position for Midpoint—Outside Range
3148 O2 Step Position for Close—Outside Range
3149 Exh Step Position for Open—Outside Range
3150 Exh Step Position for Midpoint—Outside Range
3151 Exh Step Position for Close—Outside Range
3152 Patient Wye Not Unblocked
3153 Touchscreen Failure
3154 LED Indicator Failure
3155 Remote Alarm Not Sounding
Information
The Information function is reserved for future expansion.
Option
The Option function is reserved for future expansion.
Diagnostics
ensure that it is operating within normal ranges. These checks are described
below.
Self-Test Hardware
The self-test hardware involves hardware components that check the integrity
of the software. The hardware components that perform these checks are the
watchdog timer and the bus activity monitor.
• Watchdog Timer: The watchdog timer is a timer that the software must
reset. If the timer times out it causes a reset that restarts Esprit as if
the power switch had been turned on. Normally the software resets the
watchdog timer and it never times out.
• Bus Activity Monitor: The bus activity monitor is a timer similar to the
watchdog timer. The bus activity monitor timer is a timer that only
times out if there has been no activity on the microprocessor bus. If
there is no activity on the bus, it would indicate there is a malfunction
of some kind and the Esprit microprocessor is restarted in the same
way that the watchdog timer does.
A primary objective of POST is to ensure that the watchdog timer and the bus
activity monitor hardware are working and will catch software malfunctions.
These are critical components of the Esprit safety system and therefore POST
tests these components every time it runs.
POST also checks all critical digital hardware, including the processor,
program memory, data memory, and functions of various measurement
systems.
POST also checks pneumatic components that can be safely tested with the
safety valve open.
Built-In Test
When Esprit is operating as a ventilator it is constantly making reasonableness
checks on the operation of the hardware to ensure that failures have not
occurred and that the hardware appears to be operating normally. Also while
Diagnostics
Restart
During normal operation, the ventilator performs background checks to ensure
the integrity of the system. If a problem is detected, the ventilator enters the
“restart” sequence, in which it opens the safety valve while it performs
additional integrity checks. Restart is 10 to 20 seconds, depending upon the
reason for the restart. The ventilator sounds an alarm and displays a visual
message indicating the unit is in a restart sequence. In some cases, a
countdown timer will display the number of seconds remaining until the
ventilator completes the restart cycle. If, during the restart sequence, an
actual problem is confirmed, the ventilator enters the “Vent Inop” state, in
which it activates audible and visual alarms while the safety valve remains
open. Alternatively, if at the end of the restart sequence the ventilator
determines it is safe to continue operation, the unit will return to normal
ventilation.
Breath Types
Breath Types
Volume Controlled Ventilation VCV
Pressure Controlled Ventilation PCV
Non-Invasive Positive Pressure Ventilation NPPV
Apnea Ventilation
Modes
Modes
Assist/Control (A/C) VCV, PCV
SIMV VCV,PCV
CPAP VCV, PCV
Spont/T NPPV
Spont NPPV
Volume Ventilation Resolution is one unit unless otherwise noted (cmH2O is considered
Settings, Ranges & numerically equivalent to hPa).
Setting Range
Respiratory Rate 1 to 80 Bpm
Tidal Volume 50 to 2500 mL
Peak Inspiratory Flow 3 to 140 Lpm (Compliance compensated, actual to 200 Lpm)
PEEP 0 to 35 cmH2O (hPa)
PSV Pressure 0 to 100 cmH2O (hPa)
Technical Specifications
Setting Range
Inspiratory Trigger
(I-trigger)
Pressure Control Resolution is one unit unless otherwise noted (cmH2O is considered
Ventilation Settings, numerically equivalent to hPa).
Setting Range
Apnea Rate 1-80 Bpm
Respiratory Rate 1 to 80 Bpm
PCV Pressure 5 to 100 cmH2O (hPa) (Relative to PEEP)
Inspiratory Time 0.1 to 9.9 seconds (Resolution is 1 second)
PEEP 0 to 35 cmH2O (hPa)
PSV Pressure 0 to 100 cmH2O (hPa) (Relative to PEEP)
Inspiratory Trigger
(I-trigger)
Flow Sensitivity 0.5 to 20 LPM from base flow of 3 LPM above sensitivity (Resolution is
.1 LPM)
Expiratory Trigger
(E-Cycle)
Table 12-4: Pressure Control Ventilation Settings, Ranges & Resolution (Sheet 1 of 2)
Technical Specifications
Setting Range
Rise Time 0.1 to 0.9 seconds
%O2 21% to 100%
Patient Type Adult/Pediatric
Table 12-4: Pressure Control Ventilation Settings, Ranges & Resolution (Sheet 2 of 2)
Non-Invasive Positive Resolution is one unit unless otherwise noted (cmH2O is considered
numerically equivalent to hPa).
Pressure Ventilation
Settings, Ranges &
Resolution Non-Invasive Positive Pressure Ventilation Settings, Ranges & Resolution
Setting Range
Respiratory Rate 1 to 80 Bpm
EPAP 2 to 25 cmH2O (hPa)
IPAP 2 to 35 cmH2O (hPa)
Inspiratory Time 0.1 to 9.9 seconds (Resolution is .1 second)
Rise Time 0.1 to 0.9 seconds
Inspiratory Trigger
(I-tigger)
Flow Sensitivity 0.5 to 20 LPM from base flow of 3 LPM above sensitivity (Resolution is .1
LPM)
Expiratory Trigger
(E-Cycle)
Table 12-5: Non-Invasive Positive Pressure Ventilation Settings, Ranges & Resolution
Apnea Ventilation Considered equivalent to Assist Control with Apnea Rate in VCV and PCV. In
NPPV only mandatory breaths are delivered, triggered either at the Apnea Rate
or by the patient.
Technical Specifications
Value Entry Message If the operator enters a value that is outside the operational limits of the Esprit
Ventilator, the operator will be shown a message window on the screen. Once
the operator acknowledges the message, the Esprit Ventilator returns to the
value modification display without changing the value.
Assist/Control Ventilation
Value Entry Reasonability Checks
Technical Specifications
Non-Invasive Ventilation
Value Entry Reasonability Checks
Patient Data Screen Monitored patient data is displayed when Patient Data is pressed. Patient data
ranges, resolution, units, and accuracy specifications are provided in the table
below. Minute volume data is based on an eight breath average.
Table 12-10: Patient Data Range, Resolution, Units & Accuracy (Sheet 1 of 2)
Technical Specifications
Table 12-10: Patient Data Range, Resolution, Units & Accuracy (Sheet 2 of 2)
a. Due to the variable nature of patient leaks, this parameter is an estimate only.
Technical Specifications
Level Controls
Level Controls
Control Adjustment
Display Brightness Continuous (min to max)
(underneath front panel Keys)
Calculated Values
from Expiratory Hold Calculated Values from Expiratory Hold Maneuver
Maneuver Value Range
End Expiratory Pressure -20 to 120 cmH2O (hPa)
Expiratory Pause Pressure -20 to 120 cmH2O (hPa)
Auto PEEP (calculated range) 0 to 120 cmH2O (hPa)
Interface Ports
Interface Ports
Parallel Printer Port future
RS-232 output and input
Analog Output 0 to 5 VDC full-scale future
Remote Alarm Nurse Call and Remote Alarm Annunciation
Environmental
Specifications Environmental Specifications
Temperature/Humidity Operating 10º to 40ºC (50º to 104ºF)
10 to 95% R.H. (non-condensing)
Storage -20º to 60ºC (-4º to 140ºF)
10 to 100% R.H. (non-condensing)
Atmospheric Pressure Operating 700 to 1060 cmH2O (hPa)
Storage 500 to 1060 cmH2O (hPa)
Technical Specifications
Environmental
Protection Environmental Protection
Batteries Do not dispose of in fire, possible explosion hazard. Do not dispose of in
garbage, recycle lead batteries.
Ventilator Enclosure The system plastic enclosure should not be disposed of in fire, possible
toxic fumes may be generated.
General The system and accessories (bacteria filters, patient tubing, etc.) may be
subject to medical hazardous waste regulations. Consult with local
authorities for proper disposition of the system and accessories at the end
of their useful life.
Alarms
Alarms
Alarm Audible Alarm Volume adjustable, 45 dB(A) to 85 dB(A)
Controls
Alarm Silence 120 seconds
Alarm Reset
Alarm High Inspiratory Pressure 10 to 105 cmH2O (hPa)
Settings
Low Inspiratory Pressure 3 to 105 cmH2O (hPa)
Low PEEP Pressure 0 to 35 cmH2O (hPa)
High Respiratory Rate 0 to 150 Bpm
Low Exhaled Mandatory Tidal Volume 0 mL to 2500 mL
Low Exhaled Spontaneous Tidal Volume 0 mL to 2500 mL
Low Exhaled Minute Volume 0 to 60 Lpm
High Exhaled Minute Volume 0 to 60 Lpm
High Leak (NPPV mode only) 0 to 60 Lpm
Technical Specifications
Alarms (Continued)
Alarm Normal Indicator
Status
High Urgency Alarm (flashing red)
Indicators
Medium/Low Urgency Alarm (flashing yellow/steady yellow)
Alarm Silence
Safety Valve Open
Ventilator Inoperative (steady red)
Screen Locked
Connectors
Connectors
Inspiratory Limb Connector Gas outlet port: 22 mm conical male
Expiratory Limb Connector Gas return port (on expiratory filter):
Proprietary barb fitting
Filters
Filters
Inspiratory See filter instruction sheets for complete specification.
Bacterial Filter Efficiency: 99.999+%
Viral Filtration Efficiency: 99.99+%
Expiratory See filter instruction sheets for complete specification.
Typical efficiency: 99.97% for nominal particle size of 0.3 micro meter (micron)
at 100L/min flow.
Technical Specifications
Measuring and
Display Devices Measuring and Display Devices
Pressure Measurements Type Solid State
Sensing Position Inhalation output port
Exhalation input port
Measurements Patient airway pressure
Volume Measurements Type Hot Film Anemometer flow integrated
Sensing Position O2, air and exhalation
Measurements O2 flow, air flow and exhalation flow
Oxygen Measurements Type Zirconia solid electrolyte
Sensing Position Internal enclosure
Measurements Oxygen concentration (range 0 to 95%)
Type Galvanic cell
Sensing Position Inspiratory limb of VBS
Measurements Delivered O2% (range 0 to 110%)
Settings, Alarms and Type TFT liquid crystal touch screen
Data Display
Leakage Current
Leakage Current
Ventilator Earth Leakage Current 100 to 240VAC; 300µA maximum
Enclosure/Patient Leakage Current 100 to 240VAC; 100µA maximum
Technical Specifications
Compliance and
Approvals Compliance and Approvals
The Esprit Ventilator system complies with the requirements
of the European Directive 93/42/EEC concerning medical
devices and the requirements of Directive 89/336/EEC
relating to electromagnetic compatibility.
IEC 601-1 Classification Protection class 1, type B applied part, drip proof, continuous
operation.
The Esprit Ventilator system IEC 601-1/EN60601-1
complies with these IEC 601-1-2/EN60601-1-2
international and European EN 794-1
Standards.
The Esprit Ventilator system has CSA/NRTL: CSA C22.2 No. 601-1, CSA C22.2 No. 601-2-
been certified by the following. 12, UL2601-1
Power Requirements
Power Requirements
Dimensions and
Weights Dimensions and Weights
Technical Specifications
Electromagnetic
Compatibility
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
Declaration
The Esprit Ventilator is intended for use in the electromagnetic environment specified below.
The user of the Esprit Ventilator should assure that it is used in such an environment.
Technical Specifications
5% UT 5% UT
(>95% dip in UT) (>95% dip in UT)
for 5 sec for 5 sec
Power Frequency Power frequency magnetic fields
(50/60Hz) magnetic should be at levels characteristic
field 3 A/m 3 A/m of a typical location in a typical
hospital environment.
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Technical Specifications
10 Vrms
12
10 Vrms d = ------ P
150kHz to V2
Radiated 80 MHz in
RF
IEC 12
61000-4-3 d = ------ P 80Mhz to 800Mhz
10 V/m
E1
10 V/m
80Mhz to
23
2.5GHz
d = ------ P 800Mhz to 2.5GHz
E1
Technical Specifications
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
a The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are
6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and
40.66 MHz to 40.70 MHz.
b The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in
the frequency range 80 MHz to 2.5 GHz are intended to decrease the likelihood that mobile/
portable communications equipment could cause interference if it is inadvertently brought into
patient areas. For this reason, an additional factor of 10/3 is used in calculating the
recommended separation distance for transmitters in these frequency ranges.
c Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the Esprit ventilator is used exceeds the applicable RF
compliance level above, the Esprit ventilator should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orienting or
relocating the Esprit ventilator.
d Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 1 V/m.
Technical Specifications
NOTE 2 The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765
MHz to 6.795 MHz;13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66
MHz to 40.70 MHz.
NOTE 3 An additional factor of 10/3 is used in calculating the recommended separation distance for
transmitters in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency
range 80 MHz to 2.5 GHz to decrease the likelihood that mobile/portable communications
equipment could cause interference if it is inadvertently brought into patient areas.
NOTE 4 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Technical Specifications
Pneumatic System
Internal Blower
Inspiratory
Module
To Patient
w/SV and PRV
Exhalation Heated
Valve Exhalation
Expiratory Gas Assembly
From Patient
Filter Assy
to Atmosphere
System
Electromechanical
Controls
User Interface
Technical Specifications
Labels
Ventilator Labels
Label Description
On/Off Switch Label
Analog port
AC Power Labels
OR
Technical Specifications
Manufacturer’s Label
Warning Label
Technical Specifications
Introduction This section of the manual is for inserting information associated with
operating and assembly instructions for Esprit options and accessories. When
you order an option or accessory, the instructions for that item are shipped
with the item. Insert those instructions into this section of the manual.
Software Options
Color 1004956
Communications 1010525
Flow-Trak 1019026
Graphics 1003772a
NICO-Esprit Interface 1022488
Respiratory Mechanics 1006600
RS-232 Communications 2 1015725
Trending 1013446
Hardware Accessories
AC Power Cord 1001832a
Backup Battery 1001470a
Battery, external 1001455a
Cable, Vuelink 1006912
Cart, LX200 1001453
Flex Arm Assembly 1003781
Flex Arm Bracket 1002497
Humidifier Bracket, F&P (cart mount) 1002226
Humidifier Bracket, F&P (ventilator mount) 1002901
Humidifier Bracket, Hudson, Concha III (cart mount) 1002227
Humidifier Bracket, Hudson, Concha IV (cart mount) 1002228
O2 Cylinder Brackets 1002241
O2 High Pressure Hose 1001664a
O2 Manifold 1004415
O2 Sensor Kit 1002541
NICO Bracket 1013499
NICO-Esprit, RS-232 serial communications 3’ cable 1018292
Remote Alarm Cable Kit (Normally Open Protocol) 1003741
Reusable Items
Reusable Exhalation Bacteria Filter Omni (Single) 1002970
Reusable Inspiratory Bacteria Filter (Single) 1003847
Reusable Patient Circuit Kit, Adult 1003058
Kit includes:
1 - Reusable Inspiratory Bacteria Filter 1003847
1 - Reusable Exhalation Bacteria Filter 1002970
2 - Water Traps 1003648
2 - Coupling, Straight Silicone 500-1000-43
Reusable Patient Circuit Kit, Pediatric 1003059
Kit includes:
1 - Reusable Inspiratory Bacteria Filter 1003847
1 - Reusable Exhalation Bacteria Filter 1002970
2 - Water Traps 1003648
2 - Coupling, Straight Silicone 500-1000-43
Reusable Vial System, Water Collection Esprit 1021884
Disposable Items
Disposable Exhalation Bacteria Filter (pkg of 12) 1002240
Disposable Inspiratory Bacteria Filter (Single) 1014047
Disposable Inspiratory Bacteria Filter (pkg of 10) 0342077
Disposable Patient Circuit, Adult 1003698a
Kit includes:
1 - Disposable Inspiratory Bacteria Filter 1014047
1 - Disposable Exhalation Bacteria Filter 1002240
2 - Coupling, Straight Silicone 500-1000-43
Disposable Patient Circuit, Pediatric 1003699a
Kit includes:
1 - Disposable Inspiratory Bacteria Filter 1014047
1 - Disposable Exhalation Bacteria Filter 1002240
2 - Water Traps 1003648
2 - Coupling, Straight Silicone 500-1000-43
Disposable Vial System, Water Collection Esprit 1006241
Documentation
Operator’s Manual 580-1000-01a
Quick Reference Guide 1002780a
Service Manual 580-1000-02a
WARNING: The use of accessories, cables and transducers other than those specified
may result in increased EM emissions or decreased immunity of the system.
The optional Oxygen sensor (O2 sensor) may be installed to allow monitoring of
delivered O2 to the inspiratory limb of the patient circuit. The Esprit Ventilator
will accept the MSA MiniOX® O2 sensors, P/N 1001454.
Attaching the sensor The O2 sensor attaches to the gas outlet port located below the front panel on
to the ventilator the ventilator’s lower right corner.
5
4
3
2
1
5 4 3 2 1
Flow
Warranty Respironics warrants the O2 sensor to be free from defects in material and
workmanship for a period of one year from the date of purchase, provided that
the unit is operated under conditions of normal use as described in this
operator’s manual.
The Esprit ventilator will allow for the transmission of data from Esprit via the
RS-232 communications interface. The ventilator receives commands from
the remote device and responds with fixed format records.
RS-232 Configuration The RS-232 communications port will be configured in the following manner
for all communications functions:
• Baud Rate 19,200
• Data Bits 8
• Parity None
• Stop Bits 1
Transmission of Data Unless stated otherwise, all fields will be left justified and six (6) characters in
length. A comma will separate each field. Each data transmission shall be
from the Ventilator
terminated with a carriage return.
Ventilator Report When the ventilator receives VRPT followed by a carriage return, it will respond
by transmitting the information shown in Table A-1: “Ventilation Report”. The
Command and
ventilator responds to the VRPT command by returning a string with a variable
Response (VRPT) length. Fields 2 through 4 define the length of the message. The last character
transmitted is a stop code indicating the end of the message. The second field
indicates the number of characters between the start and stop codes. The third
field indicates the number of fields between the start and stop codes. The
fourth field is the start code, 0x02. The last field in the string is the stop code,
0x03.
REF 580-1000-01P Esprit® Ventilator Operator’s Manual © Respironics, Inc. Appendix A-1
Appendix A
Ventilation Report
Appendix A-2 Esprit® Ventilator Operator’s Manual © Respironics, Inc. REF 580-1000-01P
Appendix A
REF 580-1000-01P Esprit® Ventilator Operator’s Manual © Respironics, Inc. Appendix A-3
Appendix A
Appendix A-4 Esprit® Ventilator Operator’s Manual © Respironics, Inc. REF 580-1000-01P
Appendix A
REF 580-1000-01P Esprit® Ventilator Operator’s Manual © Respironics, Inc. Appendix A-5
Appendix A
Measured Mean 5.6◆◆◆ 0.1 for -20.0 -20.0 -120 cmH2O The MAP value
Airway Pressure to 99.9 (hPa) from the Patient
1 for 100 to
120 Data screen.
Measured End 2.0◆◆◆ 0.1 -20.0-99.9 cmH2O The End Exp. value
Expiratory Pressure (hPa) from the Patient
Data screen.
Measured End 24.0◆◆ 0.1 for -20.0 -20.0 - 130 cmH2O The Plateau value
Inhalation Pressure to 99.9 (hPa) from the Patient
1 for 100 to
130 Data screen.
Measured Minute 5.83◆◆ 0.01 for 0.00 - 60.0 L The Total VE value
Volume 0.00 to 9.99 from the Patient
0.1 for 10.0
to 60.0 Data screen.
Measured 0.0◆◆◆ 0.1 for 0.0 0.0 - 150 BPM The Spont Rate
Spontaneous Breath to 9.9 value from the
1 for 10 to
Rate 150 Patient Data
screen.
Measured Total Breath 12.0◆◆ 0.1 for 0.0 0 - 150 BPM The Total Rate
Rate to 9.9 value from the
1 for 10 to
150 Patient Data
screen.
Measured Rapid 5◆◆◆◆◆ 1 0 - 500 BPM/L The F/Vt value from
Shallow Breathing the Patient Data
Index screen.
Measured I:E Ratio 1:4.1◆ 0.1 for 9.9:1 4.1:1-1:99 N/A The I:E Ratio
to 1:9.9 display from the
1 for 1:10 to
1:99 Patient Data
screen.
Measured Patient 0.0◆◆◆ 0.1 for 0.0 0.0 - 140 LPM The Pt Leak
Leak to 99.9 display.
1 for 100 to
140
Appendix A-6 Esprit® Ventilator Operator’s Manual © Respironics, Inc. REF 580-1000-01P
Appendix A
REF 580-1000-01P Esprit® Ventilator Operator’s Manual © Respironics, Inc. Appendix A-7
Appendix A
MIP Time of Last 11-14-01◆◆13:11 1 minute 01-01-7 N/A MIP Time of Last
Maneuver 0 0:00 to Maneuver display
01-19-38 from MIP/P0.1
3:14 Screen15
character field, 24
hour clock: MM-
DD-YYuuHH:MM
P0.1 -2.3◆◆ 0.1for -99.9 -100 - 200 cmH20 P0.1 display from
to 99.9 MIP/P0.1 display
1 for -100
and 100 to screen
200
P0.1 Time of Last 11-14-01◆◆13:11 1 minute 01-01-70 N/A P0.1 Time of Last
Maneuver 0:00 to Maneuver display
01-1-/38 from MIP/P0.1
3:14 Screen.15
character field, 24
hour clock: MM-
DD-YYwwHH:MM
Static Resistance 5.43◆◆ 0,01 for 0.00 - 400 cmH20 Static Resistance
0.00 to 9.99 /L/Sec display from the
0.1 for 10.0
to 99.9 Static C & R screen
1 for 100 to
400
Static Compliance 19.2◆◆ 0.01 for 0.00 - 350 mL/ Static Compliance
0.00 to 9.99 cmH2O display from the
0.1 for 10.0
to 99.9 Static C & R
1 for 100 to Screen
350
Static C & R Time of 11-14-01◆◆13:11 1 minute 01-01-70 N/A Static C & R Time
Last Maneuver 0:00 to of Last Maneuver
01-19-38 display from Static
3:14 C & R Screen.15
character field, 24
hour clock: MM-
DD-YYwwHH:MM
AutoPeep 1.2◆◆ 0.1 for -9.9 -20 - 120 cmH2O AutoPeep display
to 9.9 from the
1 for -10 to -
20 Mechanics Patient
1 for 10 to Status Screen
120
Appendix A-8 Esprit® Ventilator Operator’s Manual © Respironics, Inc. REF 580-1000-01P
Appendix A
REF 580-1000-01P Esprit® Ventilator Operator’s Manual © Respironics, Inc. Appendix A-9
Appendix A
Appendix A-10 Esprit® Ventilator Operator’s Manual © Respironics, Inc. REF 580-1000-01P
Appendix A
Static C & R Plateau 8.2◆◆◆ 0.1 for -20 -20.0 - 130 cmH2O Pplat display from
End Pressure to 99.9 Static C & R
1 for 100 to
130 Screen
Volume Control When the ventilator receives VCVS followed by a carriage return, it will respond
Ventilation Settings by transmitting the information shown in Table A-2: “Volume Control
Ventilation Settings Report”. The ventilator responds to the VCVS command by
Report (VCVS) returning a string with a variable length. Fields 2 through 4 define the length
of the message. The last character transmitted is a stop code indicating the
end of the message. The second field indicates the number of characters
between the start and stop codes. The third field indicates the number of
fields between the start and stop codes. The fourth field is the start code,
0x02. The last field in the string is the stop code, 0x03.
REF 580-1000-01P Esprit® Ventilator Operator’s Manual © Respironics, Inc. Appendix A-11
Appendix A
Appendix A-12 Esprit® Ventilator Operator’s Manual © Respironics, Inc. REF 580-1000-01P
Appendix A
REF 580-1000-01P Esprit® Ventilator Operator’s Manual © Respironics, Inc. Appendix A-13
Appendix A
Pressure Control When the ventilator receives PCVS followed by a carriage return, it will respond
Ventilation Settings by transmitting the information shown in Table A-3: “Pressure Control
Ventilation Settings Report”. The ventilator responds to the PCVS command by
Report (PCVS) returning a string with a variable length. Fields 2 through 4 define the length
of the message. The last character transmitted is a stop code indicating the
end of the message. The second field indicates the number of characters
between the start and stop codes. The third field indicates the number of
fields between the start and stop codes. The fourth field is the start code,
0x02. The last field in the string is the stop code, 0x03.
Appendix A-14 Esprit® Ventilator Operator’s Manual © Respironics, Inc. REF 580-1000-01P
Appendix A
REF 580-1000-01P Esprit® Ventilator Operator’s Manual © Respironics, Inc. Appendix A-15
Appendix A
Non-Invasive Positive When the ventilator receives NPVS followed by a carriage return, it will respond
Pressure Ventilation by transmitting the information shown in Table A-4: “Non-Invasive Positive
Pressure Ventilation Settings Report”. The ventilator responds to the NPVS
Settings Report command by returning a string with a variable length. Fields 2 through 4
(NPVS) define the length of the message. The last character transmitted is a stop code
indicating the end of the message. The second field indicates the number of
characters between the start and stop codes. The third field indicates the
number of fields between the start and stop codes. The fourth field is the start
code, 0x02. The last field in the string is the stop code, 0x03.
Appendix A-16 Esprit® Ventilator Operator’s Manual © Respironics, Inc. REF 580-1000-01P
Appendix A
REF 580-1000-01P Esprit® Ventilator Operator’s Manual © Respironics, Inc. Appendix A-17
Appendix A
Patient Data Report When the ventilator receives PTDT followed by a carriage return, it will respond
by transmitting the information shown in Table A-5: “Patient Data Report”.
(PTDT)
The ventilator responds to the PTDT command by returning a string with a
variable length. Fields 2 through 4 define the length of the message. The last
character transmitted is a stop code indicating the end of the message. The
second field indicates the number of characters between the start and stop
codes. The third field indicates the number of fields between the start and
stop codes. The fourth field is the start code, 0x02. The last field in the string
is the stop code, 0x03.
Appendix A-18 Esprit® Ventilator Operator’s Manual © Respironics, Inc. REF 580-1000-01P
Appendix A
Measured Minute 5.83◆◆ 0.01 for 0.00 to 0.00 - 60.0 L The Total VE value
Volume 9.99 from the Patient
0.1 for 10.0 to 60.0 Data screen.
Measured 0.0◆◆◆ 0.1 for 0.0 to 9.9 0.0 - 150 BPM The Spont Rate
Spontaneous 1 for 10 to 150 value from the
Breath Rate Patient Data
screen.
Measured Total 12.0◆◆ 0.1 for 0.0 to 9.9 0 - 150 BPM The Total Rate
Breath Rate 1 for 10 to 150 value from the
Patient Data
screen.
Measured Rapid 5◆◆◆◆◆ 1 0 - 500 BPM/L The F/Vt value from
Shallow Breathing the Patient Data
Index screen.
Measured I:E Ratio 1:4.1◆ 0.1 for 9.9:1 to 1:9.9 4.1:1-1:99 N/A The I:E Ratio
1 for 1:10 to 1:99 display from the
Patient Data
screen.
Measured Patient 0.0◆◆◆ 0.1 for 0.0 to 99.9 0.0 - 140 LPM The Pt Leak
Leak 1 for 100 to 140 display.
Measured Percent 56.2◆◆ 0.1 for 0.0 to 99.9 0.0 - 100 % The Pt Trigger
of Breaths 1 for 100 display on the
Triggered by the NPPV Monitor
Patient (NPPV screen.
Spont/T mode only;
otherwise
◆◆◆◆◆◆)
Monitored Oxygen 55.1◆◆ 0,1 for 0.0 to 99.91 for 0.0 - 100 % The % O2 display
Concentration 100 from the Patient
Data Screen.
Ti/Ttot 0.23◆◆ 0.01 0.00 - 1.00 N/A The Ti/Ttot display
from the
Mechanics Patient
Status Screen
REF 580-1000-01P Esprit® Ventilator Operator’s Manual © Respironics, Inc. Appendix A-19
Appendix A
Appendix A-20 Esprit® Ventilator Operator’s Manual © Respironics, Inc. REF 580-1000-01P
Appendix A
Alarm Status Report When the ventilator receives ALRM followed by a carriage return, it will
(ALRM) respond by transmitting the information shown in Table A-6: “Alarm Status
Report”. The ventilator responds to the ALRM command by returning a string
with a variable length. Fields 2 through 4 define the length of the message.
The last character transmitted is a stop code indicating the end of the
message. The second field indicates the number of characters between the
start and stop codes. The third field indicates the number of fields between
the start and stop codes. The fourth field is the start code, 0x02. The last field
in the string is the stop code, 0x03.
REF 580-1000-01P Esprit® Ventilator Operator’s Manual © Respironics, Inc. Appendix A-21
Appendix A
Appendix A-22 Esprit® Ventilator Operator’s Manual © Respironics, Inc. REF 580-1000-01P
Appendix A
REF 580-1000-01P Esprit® Ventilator Operator’s Manual © Respironics, Inc. Appendix A-23
Appendix A
Appendix A-24 Esprit® Ventilator Operator’s Manual © Respironics, Inc. REF 580-1000-01P
Appendix B. Customer Service & Warranty
Customer Service For further information or assistance in operating the Esprit Ventilator, contact
Respironics Customer Service:
Fax: 724-387-5012
email: [email protected]
Exclusions
This warranty does not apply to any unit or individual parts which have been
repaired or altered in any way that, in Respironics’ judgement, affect its ability
or reliability, or which has been subjected to misuse, negligence, abuse, or
accident.
This warranty does not cover damage that may occur in shipment.
Warranty Limits
This warranty takes precedence over all other warranties, expressed or implied.
This warranty also takes precedence over all other obligations or liabilities on
the part of Respironics including, but not limited to, contingent or
consequential damages, such as costs of repairing or replacing other property
which may be damaged as a direct result of Esprit Ventilator operation.
This warranty, and the rights and obligations described herein, is construed
under and governed by the laws of the State of California, U.S.A.
REF 580-1000-01P Esprit® Ventilator Operator’s Manual © Respironics, Inc. Appendix B-1
Appendix B
Options and Warranties are available for various options and accessories. See the specific
Accessories option or accessory in Chapter 13 for complete warranty information.
Appendix B-2 Esprit® Ventilator Operator’s Manual © Respironics, Inc. REF 580-1000-01P
Appendix C. Alarm Testing Procedure
The following procedure is available if the operator wants to test the operation
of the following alarms. Respironics recommends following “Preoperational
Procedure” on page 8-29 before performing the Alarm Testing procedure. It is
assumed that the preoperational procedure has been run before the Alarm
Testing Procedure is followed.
Setup
1. Connect O2 supply to Esprit.
2. Connect optional O2 sensor.
3. Run EST.
4. Attach the test lung to the patient wye. (Use the test lung provided
with your system.)
5. Use the following settings:
• Mode VCV-A/C
• Tidal Vol: 400ml
• High Pressure (HIP): 50cmH20 (hPa)
(or higher if required by the test lung)
• Rate: 10 Bpm
• Low Insp Press: 3 cmH20 (hPa)
• Peak Flow 40 Lpm
• PEEP: 4 cmH20 (hPa)
• Low PEEP: 2 cmH20 (hPa)
• Low Vt Mand: 0 mL
• PSV: 0 cmH20 (hPa)
• Low Vt Spont: 0 Lpm
• I-Trigger: 2 cmH20 (hPa)
• High Rate: 150 Bpm
• O2: 21%
• Low VE: 1 L
• Insp. Hold: 0 sec.
• Apnea: 15 sec.
• Apnea Rate: 20 Bpm
REF 580-1000-01P Esprit® Ventilator Operator’s Manual © Respironics, Inc. Appendix C-1
Appendix C
Appendix C-2 Esprit® Ventilator Operator’s Manual © Respironics, Inc. REF 580-1000-01P
Appendix C
4. Alarm should reset. (If the backup battery is attached, Esprit should
return to AC power. If no battery is attached, Esprit will resume
operating.)
REF 580-1000-01P Esprit® Ventilator Operator’s Manual © Respironics, Inc. Appendix C-3
Appendix C
Appendix C-4 Esprit® Ventilator Operator’s Manual © Respironics, Inc. REF 580-1000-01P
Glossary
Abbreviations 24 Hrs used to describe format for time-of-day when AM or PM does not appear
Aux auxiliary
BTPS Body temperature (98°F, ambient pressure) 100% saturated (with water vapor)
CL lung compliance
f respiratory frequency
f/VT rate versus tidal volume ratio; rapid shallow breathing index
Inop inoperational
L Liter
mL milliliter, or 1/1000 L
mm millimeters
msec millisecond
NO nitric oxide
O2 molecular oxygen
Plateau inspiratory plateau pressure, pressure at end exhalation and no flow from or out of the
ventilator
V volume
•
V flow
Vent ventilator
VT tidal volume
Definitions Airway Pressure the pressure in the patient circuit, measured at the distal end
of the exhalation filter.
Bias flow a continuous flow of gas used during expiratory phase when flow
triggering is active.
Compliance a measure of stiffness for containers that hold gas (i.e. lungs,
patient tubing). The volume required to increase the pressure in the container
by a unit of pressure (i.e., L/cmH2O (hPa)).
E-Cycle The E-Cycle setting on the Esprit determines when the ventilator will
transition from inspiration to expiration in PSV and NPPV modes. E-Cycle
defines the percent of peak inspiratory flow that end inspiratory flow needs to
drop to in order for inspiration to end.
End Expiratory Pressure (End Exp) the airway pressure measured at the end of
exhalation. The display is updated at the end of each exhalation.
Expiratory phase (exhalation) the part of the ventilatory cycle from the
beginning of expiratory flow to the beginning of inspiratory flow.
Inspiratory phase (inspiration) the part of the ventilatory cycle from the
beginning of inspiratory flow to the beginning of expiratory flow. Any inspiratory
pause (plateau) is included in the inspiratory phase.
Mandatory Breath a breath whose inspiratory flow and or pressure is under the
control of the ventilator.
Mean Airway Pressure (MAP) the average over one inspiration/exhalation cycle.
The value displayed is the average of this calculation over one minute. The
display is updated at the end of each exhalation.
Medical gas a gas that has been refined and purified according to
specifications in the United States Pharmacopoeia (USP) intended for human
use in the diagnosis or treatment of disease.
Peak Inhalation Pressure (PIP) the greatest airway pressure during an inspiratory
cycle no matter what the breath type. The pressure is measured at the
exhalation valve and the new data is displayed at the beginning of exhalation.
PCMCIA Card (PC Card) Acronym for Personal Computer Memory Card
International Association, more commonly referred to as a PC Card. This is a
data storage device with an approximate physical size of a credit card, used in
conjunction with the Trending Option.
primary control over the frequency of breathing, the inspiratory time, and the
inspiratory flow.
Rapid Shallow Breathing Index (F/VT) used to evaluate the adequacy of the
patients spontaneous ventilation. It is calculated by as shown below
• f/Vt = (Spont Rate)/(Spont VE)
Resistance The pressure drop across a pneumatic device (i.e. bacteria filter,
patient circuit tubing) for a unit of flow when the volume of the device remains
constant, i.e., (cmH2O (hPa))/mL/sec.
Risetime the time required for a pressure support or pressure controlled breath
to reach its target pressure.
Spontaneous Respiratory Rate (Spont Rate) the average rate of the spontaneous
breaths in the last eight breaths delivered by the ventilator.
•
Spontaneous Volume Exhaled (Spont V E) the exhaled volume that would come
from spontaneous breaths, projected over one minute. The calculation is done
by averaging the spontaneous exhaled tidal volume from the last eight breaths
and projecting what that volume would be if it continued for one minute.
Tidal Volume (Tidal Vol) the volume of patient gas as measured at the
exhalation flow transducer. The display shows an average unless the current
breath differs substantially from the average. When there is a substantial
change, the current breath is displayed. The average for tidal volume is
restarted when the operator changes the tidal volume setting and the machine
is delivering mandatory breaths.
Total Respiratory Rate (Total RR) the total breaths taken, spontaneous breath
rate + mandatory breath rate, from the last eight breaths and projecting what
that rate would be if it continued for one minute.
•
Total Volume Exhaled (Total V E) the total exhaled volume that would come from
all the patient’s breaths, projected over one minute. The calculation is done by
averaging the total exhaled tidal volume from the last eight breaths and
projecting what that volume would be if it continued for one minute. This value
is updated at the end of each exhalation.
U
Using Default Altitude Alarm, 9-7
Using Default Compliance Alarm, 9-7
Using Default Settings Alarm, 9-7
V
Ventilator Inoperative, 12-9
W
Warnings, 2-1
Warranty, 13A-3, B-1