Schematic - CPAP - Iss06

The document provides a schematic and characteristics for the CPAP MKII system. It includes an O2 and air supply inlet with an O2 sensor, safety PEEP valve set to 20 cmH2O, mask connected via a smooth internal bore flexible hose to a bacterial/viral filter and PEEP valve set to 10 cmH2O. The system requires bacterial/viral filters and has specifications for maximum back pressures on the CPAP unit and for patients.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
51 views

Schematic - CPAP - Iss06

The document provides a schematic and characteristics for the CPAP MKII system. It includes an O2 and air supply inlet with an O2 sensor, safety PEEP valve set to 20 cmH2O, mask connected via a smooth internal bore flexible hose to a bacterial/viral filter and PEEP valve set to 10 cmH2O. The system requires bacterial/viral filters and has specifications for maximum back pressures on the CPAP unit and for patients.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

CPAP MKII

System Schematic and Characteristics


Iss 06

Safety PEEP
20 CmH2O
O2 Air % O2
Z
Z PEEP
Supply Inlet Sensor Flow Flag
MASK
(10L/min)

Z
Z
Smooth
Bacterial internal bore Patient Bacterial
Flow & Viral flexible hose PEEP & Viral
Generator Filter (Ø19mm 10 CmH2O Filter
minimum)

1
System Characteristics
In order for the CPAP System to deliver performance and patient comfort, the following
components characteristics are required:

Bacterial & Viral Filter Peep Valve (10 CmH2O)

CPAP Max Back Pressure Patient Max Back Pressure