Julian PDF
Julian PDF
"
Julian
Anaesthetic Workstation
Instructions for Use
Software 2.n
1-36-96
On each page…
the instructions for use
combining text with illustrations. The information is
translated directly into actions to enable the user to 2
1
learn „hands-on“ how to use the workstation.
CO2-Al.
CO2 38 aus/ein
product. 40 CO2
AMV 5.2 8 Alarm
Info
VT O2 40
Fi Fet
36
Liste
Kurven
where there are several steps, numbers refer to the 0.35 0 0.5
Volumeter --s
1
Hal. 0.8 0.6 Konfig.
-- N 2O 58 56
illustration and indicate the sequence. Volumeter starten:
0 5
Bestätigen!
10
Frischgas O2 + N2O
Frischgas
O2 % L/min intern
30 4.00 extern
Screen displays guide the user and confirm the steps 3 Press the »O2 +« button.
O2 flows into the breathing system without
to be followed. anaesthetic gas as long as the button is held
down.
2
Downloaded from www.Manualslib.com manuals search engine
Contents
Contents
For Your Safety and that of Your Patients 5
Intended Use 7
Operating Concept 11
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
Preparation 21
Starting Up 25
Operation 35
Monitoring 55
Care 95
Index 142
3
Downloaded from www.Manualslib.com manuals search engine
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
4
Downloaded from www.Manualslib.com manuals search engine
For Your Safety and that of Your Patients
Maintenance
The apparatus must be inspected and serviced regularly
by trained service personnel at six monthly intervals
(and a record kept).
Repair and general overhaul of the apparatus may only
be carried out by trained service personnel.
We recommend that a service contract be obtained with
DrägerService and that all repairs also be carried out by
them. Only authentic Dräger spare parts may be used for
maintenance.
Observe chapter "Maintenance Intervals".
Accessories
Do not use accessory parts other than those in the
order list.
5
Downloaded from www.Manualslib.com manuals search engine
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
Intended Use
Contents
Intended Use............................................................................................. 9
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
7
Downloaded from www.Manualslib.com manuals search engine
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
8
Downloaded from www.Manualslib.com manuals search engine
Intended Use Intended Use
Intended Use
closed systems with "low flow" and "minimal flow" supervision of qualified medical personnel, so that
techniques (for minimal gas and anaesthetic agent assistance can be provided immediately in the event
consumption). of any malfunctions.
– Inhalation anaesthesia in non-rebreathing systems,
with separate fresh gas outlet for the connection of
e.g. the Bain system or Magill system
with a fresh gas flow of 0.5 to 12 L/min.
Ventilation modes:
– Automatic ventilation (IPPV)
and Explosive anaesthetics, such as ether or cyclo-
pressure-controlled ventilation (PCV). propane, must not be used due to the risk of fire.
– Manual ventilation (MAN).
Mobile radio telephones must not be used within
– Spontaneous ventilation (SPONT).
10 metres of the workstation!
Mobile telephones may interfere with the operation of
electrical and electronic medical equipment.
The following measured values are displayed:
– Peak pressure, mean pressure, plateau pressure and
PEEP Julian must not be used with nuclear spin
tomography (MRT, NMR, NMI)!
– Expiratory minute ventilation
Operation of the apparatus may be impaired.
Tidal volume VT
Breathing rate
Patient compliance
– Inspiratory and expiratory concentration of O2, N2O,
anaesthetic gas and CO2
optional:
– Functional oxygen saturation (SpO2) and pulse rate.
9
Downloaded from www.Manualslib.com manuals search engine
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
10
Downloaded from www.Manualslib.com manuals search engine
Operating Concept
Contents
Operating Concept
Contents
Screen ergonomics.................................................................................12
Selecting / setting ventilation parameters.................................................. 14
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
Screen layout.......................................................................................... 16
Three basic screens for monitoring........................................................... 17
The standard screen.................................................................................17
The data screen....................................................................................... 17
The trend screen...................................................................................... 18
11
Downloaded from www.Manualslib.com manuals search engine
Operating Concept
Screen ergonomics
Operating Concept
Screen ergonomics
Monitoring
All the settings required for
IPPV
– Fresh gas delivery Pleth
SpO2 98 67 alarm
limits
– Ventilation CO2 auto set
38 _ freq
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
etCO2 limits
CO2 Ve
– Monitoring 40
MV 6.0 10 alarm
info
Fi Fet list ..
12...
are entered on the system screen using the appropriate 0
PAW O2 40 36
20 curves
keys and the rotary knob. Hal. 0.8 0.6 config.
The keys are grouped in function fields: 0
N 2O 58 56
Left-hand field
Freshgas O2+N2O
Fresh gas delivery O2 % L/min
Pmax
mbar
VT
mL
Freq.
1/min
TI:TE TIP:TI
%
PEEP
mbar
CO2
etCO2 38 _ freq limits
Ve
1 Left-hand block 40
MV 6.0 10 alarm
info
These function keys are located in the bottom row on the 30 4.00 25 600 10 1:2 10 0
control panel.
Left-hand block for fresh gas setting
Right-hand block for ventilation N 2O AIR
MAN
IPPV PCV
SPONT
1 2
12
Downloaded from www.Manualslib.com manuals search engine
Operating Concept
Screen ergonomics
group of hardkeys. These softkeys are used to set the SpO2 98 67 alarm
limits
CO2
fresh gas delivery parameters and ventilation parameters. etCO2 38 _ freq
auto set
limits
CO2 Ve
1 Left-hand block:
40
MV 6.0 10 alarm
info
Fi Fet list
The keys for setting the O2 concentration and fresh 0
..
12...
PAW O2 40 36
gas flow. 20 curves
Hal. 0.8 0.6 config.
0
2 Right-hand block: N 2O 58 56
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
field.
20 PAW O2 40 36 curves
– to set and confirm the parameters for fresh gas and 30 4.00 25 600 10 1:2 10 0
ventilation modes,
– to set and confirm the monitoring functions. 3
MAN
N 2O AIR IPPV PCV
SPONT
13
Downloaded from www.Manualslib.com manuals search engine
Operating Concept
Screen ergonomics
1
2
20 PAW O2 40 36 curves
Hal. 0.8 0.6 config.
0
N 2O 58 56
Freshgas O2+N2O
Pmax VT Freq. TI:TE TIP:TI PEEP
O2 % L/min mbar mL 1/min % mbar
MAN
N 2O AIR IPPV PCV
SPONT
14
Downloaded from www.Manualslib.com manuals search engine
Operating Concept
Screen ergonomics
IPPV
Pleth
-- alarm
SpO2 98 95 limits
67 120
50
auto set
limits
CO2 42
40 etCO2 38 30
alarm
info
--
MV 6.0 3.0 list
0 --
20 PAW FiO2 29 20 curves
Freshgas O2 + N2O
Pmax VT Freq. TI:TE TIP:TI PEEP
O2 % L/min mbar mL 1/min % mbar
20
PAW FiO2 29 20
curves
FiHal. 0.8 1.0
0.0
config.
0 40
PAW 8
Freshgas O2 + N2O
Pmax VT Freq. TI:TE TIP:TI PEEP
O2 % L/min mbar mL 1/min % mbar
15
Downloaded from www.Manualslib.com manuals search engine
Operating Concept
Screen ergonomics
Screen layout
2
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
20 PAW O2 40 36 curves
Hal. 0.8 0.6 config.
0
N 2O 58 56
Freshgas O2+N2O
Pmax VT Freq. TI:TE TIP:TI PEEP
O2 % L/min mbar mL 1/min % mbar
MAN
N 2O AIR IPPV PCV
SPONT
Screen layout
➀ Status field: ➀ ➁ ➂ ➃
Displays information on the current operating mode IPPV
➁ Graphics field: Pleth
SpO2 98 67 alarm
For curves and bargraphs CO2
limits
auto set
PAW O2 40 36
➃ Numerical value field: 20 curves
16
Downloaded from www.Manualslib.com manuals search engine
Operating Concept
Screen layout
2
1
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
CO2
etCO2 38 _ freq
limits
Ve alarm
40
MV 6.0 10 info
Fi Fet list
0
20 PAW O2 40 36 curves
Freshgas O2+N2O
Pmax VT Freq. TI:TE TIP:TI PEEP
O2 % L/min mbar mL 1/min % mbar
O2 alarm
compliance 65 mL/mbar % 40 33 info
20 Hal. list
MV 6.0 L/min % 0.8 0.6
VT 0.60 L N2O
% 58 56
0
freq 10 1/min 01. Nov. 98 12 : 00 config.
sys. comp. 6.40 mL/mbar 01-10
AW-temp. 34 °C leckage 10 mL/min 8 : 00
Freshgas O2 + N2O
Pmax VT Freq. TI:TE TIP:TI PEEP
O2 % L/min mbar mL 1/min % mbar
17
Downloaded from www.Manualslib.com manuals search engine
Operating Concept
Screen layout
10
O2 40 36 SpO2
pulse
0.8 0.6
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
Iso. full
5
0 trend
0
N 2O 58 56
Select zoom area: Turn ! Zoom in: Press !
Freshgas O2 + N2O
Pmax VT Freq. TI:TE TIP:TI PEEP
O2 % L/min mbar mL 1/min % mbar
18
Downloaded from www.Manualslib.com manuals search engine
Before Using for the First Time
Contents
Contents
19
Downloaded from www.Manualslib.com manuals search engine
Before Using for the First Time
Charging the battery for emergency operation
When Julian is not in use
1
The battery must be charged for 10 hours before using
the Workstation for the first time.
● Plug the mains plug of the Julian workstation into the
mains socket. The mains voltage must correspond to
the voltage specified on the nameplate.
1 The green LED N lights up.
● Leave Julian connected to the mains for 10 hours.
The workstation does not have to be switched on.
The devices connected to auxiliary power sockets will
not be powered by the UPS in the event of a power
failure.
20
Downloaded from www.Manualslib.com manuals search engine
Preparation
Contents
Preparation
Contents
21
Downloaded from www.Manualslib.com manuals search engine
Preparation
Connecting the gas supply
Preparation
supply block. 1 2
The two ports at the front are reserved for the backup N2O AIR O2 VAC
gas cylinders.
2 The vacuum supply block (VAC) is available as option
for secretion aspiration.
Plug the other end of the pressure hoses into the wall
supply points.
22
Downloaded from www.Manualslib.com manuals search engine
Preparation
Connecting the gas supply
Connecting the anaesthetic gas scavenging system
cylinder valves.
2 Screw the gas hoses into the ports on the front of the 1 1
gas supply block.
Switch over immediately to cylinder supply if the central
gas supply fails:
● Open the cylinder valves.
N2O O2
Caution when handling O2 cylinders
● Do not oil or grease the O2 cylinder valves or O2
pressure reducing adapters, and do not handle with
greasy fingers.
Danger of explosion!
● The cylinder valves must be opened / closed by hand.
Do not use tools.
● If a cylinder valve is leaky or difficult to operate, it
must be repaired by an expert.
23
Downloaded from www.Manualslib.com manuals search engine
Preparation
Connecting the power supply
Equipotential bonding
e.g. for intracardial or intracranial operations.
2 Connect one end of the earth cable to one of the
connecting pins on the back of the workstation.
● Connect the other end of the earth cable to the
specified equipotential bonding point, e.g. on the
operating table or ceiling lamp.
24
Downloaded from www.Manualslib.com manuals search engine
Starting Up
Contents
Starting Up
Contents
Checking the workstation against the checklist.................................... 26
Switching on............................................................................................ 26
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
Gas pressures.........................................................................................29
Central gas supply................................................................................... 29
Backup gas cylinders............................................................................... 29
Suction system.........................................................................................30
O2 flush .................................................................................................. 30
O2 emergency metering........................................................................... 30
Preparing Julian for the self-test................................................................ 31
Self-test...................................................................................................31
Electronics............................................................................................... 31
Fresh gas mixer........................................................................................31
Ventilator and breathing system ............................................................... 31
System compliance................................................................................ 32
Leakage................................................................................................... 32
Emergency start......................................................................................33
25
Downloaded from www.Manualslib.com manuals search engine
Starting Up
Checking the workstation against the checklist
Starting Up
1
Version without rotary knob for O2 emergency
metering:
After switching on, Julian delivers an O2 flow of 4 L/min
for manual ventilation.
This O2 flow is switched off automatically by Julian as
soon as the »cancel test« key appears after
acknowledging the checklist.
The opening screen appears with the current software
version after about 10 seconds. Julian now loads its
software and tests its internal memory.
Please wait !
● Press the »cancel test« key. See "Emergency start" Central gas supply O2 pressure, AIR, N2O pressure, 2,7 bar
Back up cylinders O2 pressure > 50 bar, N2O pressure > 30 bar
on page 33 to continue. Suction unit
O2-Flush
functional
audible flow, breathing bag fills
26
Downloaded from www.Manualslib.com manuals search engine
Starting Up
Checking the workstation against the checklist
Water trap
● Remove water trap container by pulling downwards
and empty if necessary.
● Observe the hygiene regulations of the hospital.
● Replace container from below.
27
Downloaded from www.Manualslib.com manuals search engine
Starting Up
Checking the workstation against the checklist
1 2
Breathing system
● Complete and engaged, with both valve discs
inserted, bellows fitted and hoses securely
connected.
● Soda lime renewed, no violet discoloration.
28
Downloaded from www.Manualslib.com manuals search engine
Starting Up
Checking the workstation against the checklist
29
Downloaded from www.Manualslib.com manuals search engine
Starting Up
Checking the workstation against the checklist
1 Switch to I.
2 Set suction pressure with rotary knob »Vac.«.
3 Block the secretion viewing window or kink the
suction hose.
4 Measure the suction pressure on the pressure gauge.
1 Switch to 0. 4
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
1
3 2
O2 flush
D Julian
30
Downloaded from www.Manualslib.com manuals search engine
Starting Up
Checking the workstation against the checklist
Self-test
If all points on the checklist are OK:
● Confirm = press the rotary knob.
The self-test is started.
The self-test is run automatically
and takes 3 to 4 minutes.
Julian carries out the following automatic tests and
actions:
Electronics
– Testing and calibration of gas metering bench
– Testing and calibration of O2 sensor
– Testing of pressure sensor for airway pressure
measurement
– Testing the flow sensor
– Testing the battery
– Activating the default settings for alarm limits,
monitoring parameters and the default settings of
the ventilator and fresh gas settings
Fresh gas mixer
– Testing of gas supply O2, N2O, AIR
– Testing of gas inlet valves
– Calibration of pressures sensors in the compressed
air supply
– Testing of O2 emergency metering
– Testing of fresh gas failure monitoring (optional)
Ventilator and breathing system
– Testing and calibration of PEEP/Pmax valves
– Testing of safety valve
– Testing of MAN/AUTO switch-over valve
– Calibration of flow monitoring valve
– Testing of time monitoring valve
– Testing of bellows
31
Downloaded from www.Manualslib.com manuals search engine
Starting Up
Checking the workstation against the checklist
System compliance
Julian determines the current system compliance.
Depending on the breathing hoses used, the system
compliance is 5 to 6 mL/bar.
Leakage
Julian determines the current leakage of the breathing
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
Display (example):
A tick (✓) indicates that the relevant test point has been
completed successfully.
The clock symbol u indicates the test stage is currently
in progress. Self-test
Malfunctions detected in the self-test and any gas failures
are displayed on the screen. Some malfunctions can be
ventilator ✓
accepted by confirming = pressing the rotary knob, for and
breathing system
example no AIR supply (AIR FAILURE!!!).
Other malfunctions need to be corrected before starting electronics ✓ system 3.40 mL/mbar cancel
compliance test
up the workstation, for example no O2 supply
freshgas ✓ leakage 20 mL/mbar
(O2 FAILURE!!!). mixer
Julian leakage
test
SW-version 2.0
delete
01. Nov. 98
trend
config.
MAN
Operation with SPONT , IPPV or PCV
Freshgas O2+N2O
O2 % L/min
30 4.00
32
Downloaded from www.Manualslib.com manuals search engine
Starting Up
Emergency start
Emergency start
This procedure should only be used when there is no
D Julian
time for the self-test!
1 Switch on the workstation.
Display:
Self-test
39
33
Downloaded from www.Manualslib.com manuals search engine
Starting Up
Emergency start
Julian leakage
test
SW version 2.0
delete
01. Nov. 98
trend
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
config.
MAN
Operation with SPONT , IPPV or PCV
Freshgas O2
the standby screen is switched off and a screen saver 100 2.00
with the Dräger logo is displayed instead.
To start Julian:
Version with rotary knob for O2 emergency metering:
● Turn rotary knob for O2 emergency metering to 0.
34
Downloaded from www.Manualslib.com manuals search engine
Operation
Contents
Operation
Contents
Setting the fresh gas concentrations..................................................... 36
Adjustment ranges................................................................................... 36
Selecting the carrier gas...........................................................................36
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
Aspirating secretion................................................................................ 46
Changing patients...................................................................................47
Leakage test........................................................................................... 48
Ventilating children................................................................................. 49
Connecting the breathing hoses............................................................... 49
End of Operation.....................................................................................54
35
Downloaded from www.Manualslib.com manuals search engine
Operation
Setting the fresh gas concentrations
Operation
O2 % 25 to 100 100
36
Downloaded from www.Manualslib.com manuals search engine
Operation
Setting the fresh gas flow
Selecting the ventilation mode
MAN/SPONT
Manual ventilation
2 Set the lever of the APL pressure-limiting valve to 2
MAN.
MAN
MAN
Spontaneous breathing
4
4 Set the lever of the APL pressure-limiting valve to
SPONT.
SPONT
SPONT
37
Downloaded from www.Manualslib.com manuals search engine
Operation
Selecting the ventilation mode
MAN/SPONT
2
1
Fi Fet list
0
VT O2 40 36 curves
0.35 0 0.5
Volumeter --s
1
Hal. 0.8 0.6 config.
-- 0 5 10
N 2O 58 56
Start volumeter: Confirm !
Freshgas O2 + N2O
fresh gas
O2 % L/min internal
30 4.00 external
O2 flush
D Julian
– For flushing and rapidly filling the breathing system
and breathing bag with O2 while bypassing the
Vapor unit.
3 Press the »O2 +« button.
O2 flows into the breathing system without
anaesthetic gas as long as the button is held down.
38
Downloaded from www.Manualslib.com manuals search engine
Operation
Selecting the ventilation mode
MAN/SPONT
SpO2 ON ––
_
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
ON 92
Pulse H _ ON 120
ON 50
etCO2 ON 50 (mmHg)
_ ON ––
FiCO2 > OFF
MV _ OFF ––
OFF ––
FiO2 _ OFF ––
ON 20
Fi Hal. _ ON 1.5
OFF ––
Fi Iso. _ ON 2.3
OFF ––
Fi Enf. _ ON 3.4
OFF ––
Fi Des. _ ON
OFF
12.0
––
Fi Sev. _ ON 3.4
OFF ––
PAW _ ON 40
OFF ––
Apnoea pressure OFF
Apnoea flow OFF
Apnoea CO2 ON activated after
60 seconds
39
Downloaded from www.Manualslib.com manuals search engine
Operation
Selecting the ventilation mode
IPPV
Display (example):
Standby
[mL]
30 4.00 25 600 12 1:2 10 0
Frequency Freq. 6 to 60 12
[1/min]
Insp/Exp ratio 2 : 1, 1 : 1, 1:2
TI : TE 1 : 2, 1 : 3,
1:4
Insp. pause time : 0 to 50 10
Insp. time TIP : TI [%]
PEEP [mbar] 0 to 20 0
40
Downloaded from www.Manualslib.com manuals search engine
Operation
Selecting the ventilation mode
IPPV
Starting IPPV
Julian
1 Press the »IPPV« hardkey
2 and confirm with the rotary knob.
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
1 2
Fi Fet list
The symbol rotates in the softkey »L/min« to indicate 0
41
Downloaded from www.Manualslib.com manuals search engine
Operation
Selecting the ventilation mode
IPPV
SpO2 ON ––
_ ON 92
Pulse H _ ON 120
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
ON 50
etCO2 _ ON 50 (mmHg)
ON ––
FiCO2 > ON Fixed upper alarm
limit 5 mmHg
MV _ ON ––
ON 3.0
FiO2 _ ON ––
ON 20
Fi Hal. _ ON 1.5
ON ––
Fi Iso. _ ON
ON
2.3
––
Fi Enf. _ ON 3.4
ON ––
Fi Des. _ ON 12.0
ON ––
Fi Sev. _ ON 3.4
ON ––
PAW _ ON 40
ON 8
Apnoea pressure ON Active after
15 seconds
Apnoea flow ON Active after
15 seconds
Apnoea CO2 ON Active after
30 seconds
42
Downloaded from www.Manualslib.com manuals search engine
Operation
Selecting the ventilation mode
PCV
2
Presetting ventilation parameters for PCV
1 Press »PCV« its LED flashes. 3
1
● The five ventilation parameters for PCV appear on the
screen with the default settings.
2 Press the softkey of the relevant ventilation parameter.
3 Set and confirm the ventilation parameter with the
rotary knob.
Standby
Display (example):
Julian leakage
test
SW version 2.0
delete
01. Nov. 98
trend
config.
Freshgas O2+N2O
Pmax Freq. TI:TE Insp.Flow PEEP
O2 % L/min mbar 1/min L/min mbar
30 4.00 15 12 1:2 30 0
–––––––––––
* The default settings can be set specifically for each hospital,
see page 82.
** Can be set by DrägerService.
43
Downloaded from www.Manualslib.com manuals search engine
Operation
Selecting the ventilation mode
PCV
Starting PCV
Julian
1 Press »PCV«.
2 Confirm by pressing the rotary knob.
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
1 2
Fi Fet list
If a change in ventilation parameters is required: 0
20 PAW O2 40 36 curves
● Press the relevant softkey. Set and confirm the Hal. 0.8 0.6 config.
ventilation parameter with the rotary knob. 0
N 2O 58 56
Freshgas O2 + N2O
Pmax Freq. TI:TE Insp. flow PEEP
O2 % L/min mbar 1/min L/min mbar
30 4.00 18 10 1:2 30 0
44
Downloaded from www.Manualslib.com manuals search engine
Operation
Selecting the ventilation mode
PCV
SpO2 ON ––
_ ON 92
Pulse H _ ON 120
ON 50
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
etCO2 _ ON 50 (mmHg)
ON ––
FiCO2 > ON Fixed upper alarm
limit 5 mmHg
MV _ ON ––
ON 3.0
FiO2 _ ON ––
ON 20
Fi Hal. _ ON 1.5
ON ––
Fi Iso. ON 2.3
_ ON ––
Fi Enf. ON 3.4
_ ON ––
Fi Des. _ ON 12.0
ON ––
Fi Sev. _ ON
ON
3.4
––
PAW _ ON 40
ON 8
Apnoea pressure ON Active after
15 seconds
Apnoea flow ON Active after
15 seconds
Apnoea CO2 ON Active after
30 seconds
45
Downloaded from www.Manualslib.com manuals search engine
Operation
Setting the Vapor unit
Aspirating secretion
Aspirating secretion
● Swing the bottles forwards.
D Julian
1 Set the switch to I.
2 Seal the "fingertip" or kink the suction hose and
3 Set the appropriate suction pressure for the patient
with the rotary knob »Vac.« and
4 check on the pressure gauge.
4
Observe the hygiene regulations of the hospital!
1
2 3
After aspirating
● Rinse the suction hose with the water purifier.
1 Set the switch to 0.
46
Downloaded from www.Manualslib.com manuals search engine
Operation
Changing patients
Changing soda lime
Changing patients
To switch Julian to standby:
● Press the Standby key E and confirm with the
rotary knob.
Standby
The functions of the workstation are switched off.
The set alarm limits are cancelled and the default alarm
settings are valid again. Julian leakage
test
SW-version 2.0
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
The default settings are loaded for dosing anaesthetic 01. Nov. 98
delete
trend
gas and for the ventilation parameters.
config.
Freshgas O2+N2O
O2 % L/min
30 4.00
47
Downloaded from www.Manualslib.com manuals search engine
Operation
Leakage test
Leakage test
Must not be performed when a patient is connected to
the workstation!
– When the soda lime has been changed
or
– when the breathing hoses have been changed.
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
Display:
After the test has been successfully completed, the clock
symbol disappears and Julian displays the values for
leakage and system compliance for a few seconds. Leakage test
The results of the leakage test are continually displayed
on the data screen.
System-
compliance
To return to the standby screen: Leakage
Please wait.
48
Downloaded from www.Manualslib.com manuals search engine
Operation
Ventilating children
Ventilating children
For tidal volumes VT of less than 200 mL:
● Use paediatric hoses.
Volumeter function:
● Select »Child« scale, page 67.
49
Downloaded from www.Manualslib.com manuals search engine
Operation
Using non-rebreathing systems
VT O2 40 36 curves
0.35 0
Volumeter --s
0.5 1
Hal. 0.8 0.6 config.
-- 0 5 10
N 2O 58 56
Start volumeter: Confirm !
Freshgas O2 + N2O
fresh gas
O2 % L/min internal
30 4.00 external
50
Downloaded from www.Manualslib.com manuals search engine
Operation
Using non-rebreathing systems
CO2
etCO2 38 _ freq on/off
Ve alarm
● Set the fresh gas flow. The fresh gas supply must be
40
MV -- -- info
at least twice the minute volume in order to prevent Fi Fet list
0
rebreathing. 30 28
O2 curves
0.8 0.8
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
Hal.
config.
N 2O 58 57
Freshgas O2 + N2O
fresh gas
O2 % L/min internal
30 6.00 external
Certain alarms are automatically switched off in order to
avoid artefacts.
The inactive alarm limits are highlighted by the grey
background in the following table.
SpO2 ON ––
_ ON 92
Pulse H _ ON 120
ON 50
etCO2 ON 50 (mmHg)
_ ON ––
FiCO2 > ON Fixed upper alarm
limit 5 mmHg
MV OFF ––
_ OFF ––
FiO2 OFF ––
_ ON 20
Fi Hal. ON 1.5
_ OFF ––
Fi Iso. ON 2.3
_ OFF ––
Fi Enf. ON 3.4
_ OFF ––
Fi Des. ON 12.0
_ OFF ––
Fi Sev. ON 3.4
_ AUS ––
PAW ON 40
_ OFF ––
Apnoea pressure OFF
Apnoea flow OFF
Apnoea CO2 ON
51
Downloaded from www.Manualslib.com manuals search engine
Operation
Using non-rebreathing systems
In the event of a power failure
together with the remaining available battery capacity in Hal. 0.8 0.6 config.
0
percent. N 2O 58 56
Example:
J Freshgas O2+N2O
Pmax VT Freq. TI:TE TIP:TI PEEP
95 % O2 % L/min mbar mL 1/min % mbar
52
Downloaded from www.Manualslib.com manuals search engine
Operation
In the event of a gas failure
In the Event of a Major Hardware Fault
53
Downloaded from www.Manualslib.com manuals search engine
End of Operation
End of Operation
54
Downloaded from www.Manualslib.com manuals search engine
Monitoring
Contents
Monitoring
Contents
Selecting the standard screen................................................................56
Alarms..................................................................................................... 60
Displaying and setting alarm limits............................................................ 60
Adapting limits – AutoSet......................................................................... 62
If an alarm occurs................................................................................... 63
Alarm info................................................................................................64
List display.............................................................................................. 65
Delete list................................................................................................. 65
Selecting curves..................................................................................... 66
Configuring in operation......................................................................... 68
Setting the monitoring functions................................................................69
Setting the pulse tone...............................................................................69
Setting the alarm sound............................................................................69
Selecting the scale................................................................................... 69
Setting the measuring parameters............................................................ 70
Record..................................................................................................... 71
Time.........................................................................................................71
Activating monitor defaults........................................................................71
Selecting alarms..................................................................................... 72
Default..................................................................................................... 72
Alarm mode for the heart-lung machine HLM............................................ 73
Manual calibration...................................................................................75
55
Downloaded from www.Manualslib.com manuals search engine
Monitoring
Selecting the standard screen
Monitoring
or
2 Press the S key several times. 2
1
Fi Fet list
0
data is still reduced in comparison with the "Technical 30 4.00 25 600 10 1:2 10 0
data".
For maximum accuracy the sensors can be calibrated
manually – see page 75.
When measuring CO2, anaesthetic agent and N2O, their
numerical values are shown in grey during the warm-up
phase whenever the workstation is switched on.
The display CAL appears instead of the numerical value
during automatic calibration of the O2 sensor (every
24 hours).
The flow sensor is automatically calibrated as soon as a
CO2 signal is detected by the gas measuring system.
Julian takes account of the relevant gas mixture when
calculating the volumetric concentrations.
56
Downloaded from www.Manualslib.com manuals search engine
Monitoring
Selecting the data screen
with their units of measure. The bar display on the left of O2 alarm
the screen shows the change in airway pressure PAW compliance 65 mL/mbar % 40 33 info
20 Hal. list
over time. The lines cutting across the bar graph mark 6.0 0.8 0.6
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
MV L/min %
the upper and lower alarm limits. To set the alarm limits 0.60 N 2O
VT L
% 58 56
for PAW, see page 60. freq 10 1/min 01. Nov. 98 12 : 00 config.
0
sys. comp. 6.40 mL/mbar 01-10
The system compliance (Sys.-Compl.) and leakage are AW-temp. 34 °C leckage 10 mL/min 8 : 00
57
Downloaded from www.Manualslib.com manuals search engine
Monitoring
Selecting the trend screen
– SpO2/pulse rate
● Press the S key several times until the trend screen
appears.
AMV and compliance trends are scaled in accordance
with the setting in the configuration menu for adults or
children.
AMV trend scale:
0 to 3 L/min for children
0 to 15 L/min for adults
Compliance trend scale:
0 to 20 mL/mbar for children
0 to 100 mL/mbar for adults
____________
* Lung properties can be monitored continuously with the compliance
trend display so that ventilation can be adjusted accordingly.
Patient compliance is not measured in MAN/SPONT ventilation
mode.
58
Downloaded from www.Manualslib.com manuals search engine
Monitoring
Selecting the trend screen
Zoom function
IPPV
PAW CO2
The trend display can be magnified with the zoom 60
20
O2
MV Fi Fet compl.
15
display: 5
Iso. 0.8 0.6 full
0 trend
● Press the rotary knob. 0
N 2O 58 56
A new dashed frame appears after a corresponding Select zoom area: Turn ! Zoom in: Press !
period of operation and can also be enlarged. Freshgas O2 + N2O
Pmax VT Freq. TI:TE TIP:TI PEEP
O2 % L/min mbar mL 1/min % mbar
config.
MAN
Operation with SPONT , IPPV or PCV
Freshgas O2+N2O
O2 % L/min
30 4.00
The system asks you to confirm that the trend is really to
be deleted.
To delete:
Standby
● Press the softkey »delete«.
The stored trend for the anaesthetic gas concentration is
deleted automatically when a new anaesthetic gas is Julian delete
SW-version 2.0
selected. 01. Nov. 98
do not
delete
config.
MAN
Operation with SPONT , IPPV or PCV
Freshgas O2+N2O
O2 % L/min
30 4.00
59
Downloaded from www.Manualslib.com manuals search engine
Monitoring
Alarms
Alarms
Displaying and setting alarm limits
Alarms can be displayed and set from all three basic
screens (standard, data and trend screen) during
operation.
The default alarm limits for the ventilation mode are
activated automatically when changing from standby to a
ventilation mode.
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
20
PAW FiO2 29 20
curves
FiHal. 0.8 1.0
0.0
config.
0 40
PAW 8
Freshgas O2 + N2O
Pmax VT Freq. TI:TE TIP:TI PEEP
O2 % L/min mbar mL 1/min % mbar
42
Example etCO2:
The upper alarm limit (42 mmHg) and the lower alarm etCO2 38 30
limit (30 mmHg) are shown alongside the measured value
(38 mmHg).
A deactivated alarm limit is indicated by two dashes (– –).
60
Downloaded from www.Manualslib.com manuals search engine
Monitoring
Alarms
[mmHg] 0 to 74
MV 0.1 to 39
[L/min] _ 0 to 38.9
FiO2 _ 19 to 100
[Vol.%] 18 to 99
Fi Hal. 0.1 to 7
[Vol.%]
_ 0 to 6.9
Fi Iso. 0.1 to 7
[Vol.%] _ 0 to 6.9
Fi Enf. 0.1 to 7
[Vol.%] _ 0 to 6.9
Fi Des. 0.1 to 21.9
[Vol.%] _ 0 to 21.8
Fi Sev. 0.1 to 9.9
[Vol.%]
_ 0 to 9.8
PAW 5 to 98
[mbar] _ 0 to 35
61
Downloaded from www.Manualslib.com manuals search engine
Monitoring
Alarms
MV VT x f x 1.4 VT x f x 0.6
[L/min] At least 2.0 At least 0.5
62
Downloaded from www.Manualslib.com manuals search engine
Monitoring
If an alarm occurs
If an alarm occurs
Alarm messages are assigned to three priority classes
by Julian, depending on their urgency, and identified by
exclamation marks:
Warning:
1 Red lamp flashes, accompanied by a continuous tone. Julian
Caution: 1
1 Yellow lamp flashes, accompanied by an intermittent
2
tone every 30 seconds.
Advisory:
1 Yellow lamp lights up continuously, accompanied by a
single tone.
63
Downloaded from www.Manualslib.com manuals search engine
Monitoring
CO 2 alarms on/off
Alarm info
20 PAW curves
FIO2 < !!!
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
config.
Active warning: 0
ETCO2 < !!
Freshgas O2 + N2O
Pmax VT Freq. TI:TE TIP:TI PEEP
O2 % L/min mbar mL 1/min % mbar
The display only remains visible as long as the softkey is 50 1.00 25 600 10 1:2 10 0
pressed.
CO2 CO2-al.
In this case, the symbol CO2 g appears alongside the CO2 etCO2 38 _ freq on /off
alarm
measured end-expiratory CO2 concentration. 40
MV 5.2 Ve
8 Info
64
Downloaded from www.Manualslib.com manuals search engine
Monitoring
List display
List display
Record of measured values, alarms and ventilation
modes to facilitate compilation of the anaesthetic record.
The control criteria for entries can be configured, see
"Record" on page 71.
● Press the softkey »list«.
Display (example): IPPV
PAW next page previous page
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
Delete list
The list and trend memory are deleted together!
Only possible in standby!
Standby
In standby:
Julian leakage
test
● Press the softkey »delete trend«. SW-version 2.0
delete
01. Nov. 98
trend
config.
Delete trends and list:
● Press the softkey »delete«. Operation with
MAN
SPONT , IPPV or PCV
Freshgas O2+N2O
O2 % L/min
30 4.00
65
Downloaded from www.Manualslib.com manuals search engine
Monitoring
Selecting curves
Selecting curves
Only possible from the standard screen!
Three curves can be selected and positioned for the
momentary ventilation mode in this menu.
The following curves can be selected:
AGas Curve of anaesthetic agent concentration at
the Y-piece
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
Freshgas O2 + N2O
Pmax VT Freq. TI:TE TIP:TI PEEP
O2 % L/min mbar mL 1/min % mbar
66
Downloaded from www.Manualslib.com manuals search engine
Monitoring
Using the volumeter function
0 1
46 s Hal.
config.
Volumeter (minute volume measurement), preceded
4.1 N 2O 58 56
by the numerical value. 0 5 10
on page 83.
in L/min 4.1 ➁
0 5 10
Scale 0.2 2
for children
Scale 1.0 10
for adults
67
Downloaded from www.Manualslib.com manuals search engine
Monitoring
Configuring in operation
Configuring in operation
The configuration menus are used to select or modify
individual monitoring functions for the measurements in
progress.
The settings entered here are only valid while measu-
rement is in progress and are deleted on switching to
standby mode.
In the standard screen (example) or data screen: IPPV
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
Pleth
● Press the softkey »config.«. SpO2 98 67 alarm
limits
CO2 auto set
CO2
etCO2 38 _ freq
limits
Ve alarm
40
MV 6.0 10 info
Fi Fet list
0
20 PAW O2 40 36 curves
Freshgas O2+N2O
Pmax VT Freq. TI:TE TIP:TI PEEP
O2 % L/min mbar mL 1/min % mbar
Freshgas O2 + N2O
Pmax VT Freq. TI:TE TIP:TI PEEP
O2 % L/min mbar mL 1/min % mbar
68
Downloaded from www.Manualslib.com manuals search engine
Monitoring
Configuring in operation
Setting the monitoring functions
parameters
SpO2 on off
Isoflurane record
20 activate monitor defaults
The anesthetic HLM on off
agent is automati-
cally selected by
gas identification
Freshgas O2 + N2O
Pmax VT Freq. TI:TE TIP:TI PEEP
O2 % L/min mbar mL 1/min % mbar
69
Downloaded from www.Manualslib.com manuals search engine
Monitoring
Configuring in operation
Setting the monitoring functions
40 puls tone 0 1 2 3 4 5 6 7 8 9
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
alarm sound 1 2 3 4 5 6 7 8 9
SpO2-measure. on off Scale adult child
parameter
sample rate 60 200 mL/min
record
20 activate monitor defaults
CO2 unit mmHg kPa vol%
● Select »on« or »off« with the rotary knob and press to 50 1.00 25 600 10 1:2 10 0
confirm.
70
Downloaded from www.Manualslib.com manuals search engine
Monitoring
Configuring in operation
Setting the monitoring functions
Record
This function is used to determine the type of event that
triggers an entry in the record list or a printout on the
connected logging printer.
● Select the line »record« with the rotary knob and
press to confirm.
The »record« menu appears on the left-hand side.
Display (example): IPPV
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
71
Downloaded from www.Manualslib.com manuals search engine
Monitoring
Configuring in operation
Selecting alarms
Selecting alarms
● Turn the rotary knob to select the column »alarms«
and press to confirm.
Default
This function is used to activate the default alarm limits
set in the standby configuration which are effective
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
72
Downloaded from www.Manualslib.com manuals search engine
Monitoring
Configuring in operation
Selecting alarms
0
Fi Fet list
O2 30 30 curves
N 2O 0.0 0.0
Freshgas O2 + AIR
fresh gas
HLM alarm mode can be switched off at any time: O2 % L/min internal
73
Downloaded from www.Manualslib.com manuals search engine
Monitoring
Configuring in operation
Selecting the anaesthetic agent
CO2
etCO2 38 _ freq
limits
Ve alarm
Fi display Fet display Situation 40
MV 6.0 10 info
Fi Fet list
grey light AGAS mixture 0
press to confirm.
● Turn the rotary knob until the required gas appears in
0
the display window.
Press the rotary knob to confirm the selection. Confirm to select Isoflurane.
Freshgas O2 + N2O
Pmax VT Freq. TI:TE TIP:TI PEEP
O2 % L/min mbar mL 1/min % mbar
74
Downloaded from www.Manualslib.com manuals search engine
Monitoring
Configuring in operation
Manual calibration
Manual calibration
Julian calibrates its sensors automatically during IPPV
operation. The message CAL appears instead of the PAW anesth. agent alarms calibration settings
current measured value during a calibration process.
Calibration of the O2 sensor and zero calibration of the 40 selection default O2-Sensor pulse tone 0 1 2 3 4 5 6 7 8 9
21 Vol.% ✓ alarm sound 1 2 3 4 5 6 7 8 9
other gas sensors can be repeated manually, for instance auto man
CO2 on off scale adult child
gas sensor ✓
if automatic calibration was unsuccessful. SpO2 on off
parameters
Isoflurane record
20 activate monitor defaults
The anesthetic HLM on off
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
agent is automati-
cally selected by
● Turn the rotary knob to select the »calibration« gas identification
Freshgas O2 + N2O
Pmax VT Freq. TI:TE TIP:TI PEEP
O2 % L/min mbar mL 1/min % mbar
75
Downloaded from www.Manualslib.com manuals search engine
Monitoring
SpO 2 Measurement
– Patient weight
– Patient mobility
– Possible application site
– Patient perfusion
– Duration of use
Duration Short and long-term Short and long-term Short-term Short and long-term Short and long-term
of use monitoring monitoring monitoring zeitüberwachung monitoring
Sterility1) Sterile packaging Sterile packaging ____ Sterile packaging Sterile packaging
–––––––––––––
1) in unopened, undamaged packaging
76
Downloaded from www.Manualslib.com manuals search engine
Monitoring
SpO 2 Measurement
77
Downloaded from www.Manualslib.com manuals search engine
Monitoring
SpO 2 Measurement
adhere properly.
The strips must not be stretched unduly.
Never use two strips together, as this may lead to venous
pulsation and failure of the pulse signal.
High intrathoracic pressure, pressure on the thorax and
other consecutive impairments of the venous flow can
lead to venous pulsation with failure of the pulse signal.
The pulse signal may fail in the presence of shock, low
blood pressure, severe vasoconstriction, major anaemia,
hypothermia, arterial occlusion proximal to the sensor and
asystolia.
The sensor must be protected from exposure to bright
light (e.g. surgical lamps and direct sunlight), otherwise
the pulse signal may fail or inaccurate results may be
obtained.
The sensor should not be positioned on limbs together
with an arterial catheter, sphygmomanometer cuff or
intravascular venous infusion, otherwise the pulse signal
may fail and measurement become inaccurate.
Measurement accuracy may be reduced in the presence
of significant concentrations of dyshaemoglobins, such
as carboxyhaemoglobin or methaemoglobin.
Intravascular dyes, such as methylene blue, may also
impair measurement accuracy.
Electrocautery can impair the measuring accuracy; the
leads and sensor should therefore be positioned as far
away as possible from the electrocautery and its neutral
electrode.
Sensor performance may be impaired and lead to
inaccurate results if the patient moves violently. The
sensor should be applied to a different site in such cases
in order to reduce the risk of artefacts due to movement.
78
Downloaded from www.Manualslib.com manuals search engine
Monitoring
SpO 2 Measurement
finger. The tip of the finger must touch the end and
the soft padding should rest on the nail and tip of the
finger. The lead should be on top of the finger.
● Ensure that the finger is not compressed or hurt by
the clip.
● Change the application site after not more than
4 hours in order to avoid a build-up of blood pressure
(blocked circulation).
79
Downloaded from www.Manualslib.com manuals search engine
Monitoring
Airway temperature measurement
2
3
80
Downloaded from www.Manualslib.com manuals search engine
Configuring in Standby Mode
Contents
Contents
Setting default values............................................................................. 82
Setting the pulse tone...............................................................................83
Setting the alarm sound............................................................................83
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
Manual calibration...................................................................................92
100 Vol.% O2 calibration..........................................................................92
O2 linearity test........................................................................................ 93
81
Downloaded from www.Manualslib.com manuals search engine
Configuring in Standby Mode
Setting default values
The following settings can be made in the menu: selection O2-sensor ✓ pulse tone 0 1 2 3 4 5 6 7 8 9
auto man 21 vol.% alarm sound 1 2 3 4 5 6 7 8 9
– Default settings gas sensor ✓
scale adult child
parameters
Isoflurane record
– Calibration more
interfaces
The anesthetic
agent is automati- alarm limits
– Anaesth. agent cally selected by curves
gas identification
basic configurations
Fields highlighted by a grey background contain the IPPV default settings
PCV default settings
currently valid settings. fresh gas delivery
***
82
Downloaded from www.Manualslib.com manuals search engine
Configuring in Standby Mode
Setting default values
pulse tone 0 1 2 3 4 5 6 7 8 9
alarm sound 1 2 3 4 5 6 7 8 9
scale adult child
parameters
record
Setting the pulse tone (optional) interfaces
alarm limits
0 = OFF curves
9 = Maximum volume basic configurations
IPPV default settings
● Select the line »pulse tone« by turning the rotary knob PCV default settings
fresh gas delivery
and press to confirm. The cursor automatically jumps
to the last value selected (example 5).
● Set the value with the rotary knob and press to
confirm.
83
Downloaded from www.Manualslib.com manuals search engine
Configuring in Standby Mode
Setting default values
pulse tone 0 1 2 3 4 5 6 7 8 9
SpO2-measure. on off
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
alarm sound 1 2 3 4 5 6 7 8 9
sample rate 60 200 mL/min scale adult child
parameters
CO2 unit mmHg kPa vol% record
interfaces
alarm limits
curves
basic configurations
IPPV default settings
PCV default settings
Switch SpO2 measurement (optional) on/off: fresh gas delivery
84
Downloaded from www.Manualslib.com manuals search engine
Configuring in Standby Mode
Setting default values
Record
This function is used to determine the type of event that
triggers an entry in the record list or a printout on the
connected record printer.
● Select the line »record« with the rotary knob and
press to confirm.
The »record« menu appears on the left-hand side.
Display (example):
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
Standby / Configuration
anesth. agent calibration default settings
_____________
* MEDIBUS: Dräger communications protocol for medical equipment.
85
Downloaded from www.Manualslib.com manuals search engine
Configuring in Standby Mode
Setting default values
Interface
Choice between configuration of COM2 or COM3.
Baud rate
Transmission speed (variable, see Instructions for Use of
the equipment to be connected).
Parity
This display is invariable for MEDIBUS and is shown only
for information.
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
Data bits
This display is invariable for MEDIBUS and is shown only
for information.
Stop bits
This display is invariable for MEDIBUS and is shown only
for information.
86
Downloaded from www.Manualslib.com manuals search engine
Configuring in Standby Mode
Setting default values
etCO2 1 to 75 50 (mmHg)
[mmHg] _ 0 to 74 ––
MV 0.1 to 39 12
[L/min] _ 0 to 38.9 3.0
FiO2 19 to 100 ––
[Vol.%]
_ 18 to 99 20
Fi Hal. 0.1 to 7 1.5
[Vol.%] _ 0 to 6.9 ––
Fi Iso. 0.1 to 7 2.3
[Vol.%] _ 0 to 6.9 ––
Fi Enf. 0.1 to 7 3.4
[Vol.%] _ 0 to 6.9 ––
Fi Des. 0.1 to 21.9 12.0
[Vol.%]
_ 0 to 21.8 ––
Fi Sev. _ 0.1 to 9.9 3.4
[Vol.%] 0 to 9.8 ––
PAW _ 5 to 98 40
[mbar] 0 to 35 8
87
Downloaded from www.Manualslib.com manuals search engine
Configuring in Standby Mode
Setting default values
Configuring curves
Three standard curves can be configured for each of the
ventilation modes: MAN/SPONT, IPPV, PCV and Fresh
gas external.
● Turn the rotary knob to select the line »curves« and
press to confirm.
The selection menu appears in the left-hand field.
Display (example):
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
Standby / Configuration
anesth. agent calibration default settings
● Select the required ventilation mode with the rotary pulse tone 0 1 2 3 4 5 6 7 8 9
alarm sound 1 2 3 4 5 6 7 8 9
knob: default curves Man./Spont. scale adult child
parameters
default curves IPPV
record
default curves PCV interfaces
default curves freshgas external alarm limits
curves
basic configurations
IPPV default settings
PCV default settings
fresh gas delivery
The »Standard IPPV screen« has been selected in the Standby / Configuration
example: anesth. agent calibration default settings
The corresponding position remains unused if »blank« To select curve position: turn and confirm .
is selected.
88
Downloaded from www.Manualslib.com manuals search engine
Configuring in Standby Mode
Setting default values
way.
● Set and confirm the date as described above.
89
Downloaded from www.Manualslib.com manuals search engine
Configuring in Standby Mode
Setting default values
25 600 12 1:2 10 0
15 12 1:2 10 0
● Press the rotary knob.
_____________
* Maximum setting for Pmax: 30 mbar.
90
Downloaded from www.Manualslib.com manuals search engine
Configuring in Standby Mode
Setting default values
O2 % L/min
100 2.00
91
Downloaded from www.Manualslib.com manuals search engine
Configuring in Standby Mode
Manual calibration
Manual calibration
Julian calibrates its sensors automatically during opera- Standby / Configuration
tion. The message CAL appears instead of the current anesth. agent calibration default settings
Other settings can already be made during the calibration gas sensor ✓
process.
more
O2-sensor
100 Vol.%
linearity
100 Vol.% O2 calibration test O2
92
Downloaded from www.Manualslib.com manuals search engine
Configuring in Standby Mode
Manual calibration
Selecting anaesthetic agent
linearity
● Display (example): test O2
O2 21 vol.%
A display in the range between 18 and 24 vol.% is
correct.
If the display is outside this range, the O2 sensor is
defective or spent
(replace O2 sensor, see page 117).
● Turn the rotary knob to select »man« and press to Confirm to select Isoflurane.
confirm.
● Turn the rotary knob to select the line »gas« and press
to confirm.
● Turn the rotary knob until the required gas appears in
the display window.
Press the rotary knob to confirm the selection.
● Check that the manually selected anaesthetic agent is
compatible with the Vapor unit.
93
Downloaded from www.Manualslib.com manuals search engine
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
Care
Contents
Dismantling components........................................................................96
Dismantling sample line............................................................................ 96
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
Assembling........................................................................................... 106
Installing the breathing system................................................................106
Inserting the flow sensor........................................................................ 106
Installing the bellows.............................................................................. 107
Filling and installing the absorber............................................................107
Installing the breathing hoses................................................................. 108
Preparing the suction system..................................................................108
Connecting the measured gas recirculation system................................ 109
Fitting waste gas connector....................................................................110
Sample gas scavenging..........................................................................110
Fitting emergency ventilation bag............................................................111
Connecting the anaesthetic gas scavenging system AGS.......................111
95
Downloaded from www.Manualslib.com manuals search engine
Care
Dismantling components
Care
Dismantling components
Dismantling sample line
● Unscrew the sample line from the Y-piece and water
trap on the back of the unit.
The sample line is not reusable and can be disposed
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
96
Downloaded from www.Manualslib.com manuals search engine
Care
Dismantling components
97
Downloaded from www.Manualslib.com manuals search engine
Care
Dismantling components
3 3
98
Downloaded from www.Manualslib.com manuals search engine
Care
Dismantling components
99
Downloaded from www.Manualslib.com manuals search engine
Care
Dismantling components
1
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
1 1
4 3
100
Downloaded from www.Manualslib.com manuals search engine
Care
Dismantling components
101
Downloaded from www.Manualslib.com manuals search engine
Care
Disinfecting / cleaning / sterilizing
Buraton 10 F
– Compounds containing alkylamines (made by Schülke & Mayr GmbH).
– Compounds containing phenols Note the manufacturer's Instructions for Use.
– Halogen-releasing compounds Do not allow any liquids to enter openings in the
– Strong organic acids apparatus.
– Oxygen-releasing compounds.
Users in the Federal Republic of Germany are advised to
use disinfectants specified in the current list published by
Flow-Sensor
the German Society for Hygiene and Microbiology
(DGHM). The DGHM list also specifies the active agent ● Disinfect for approx. 1 hour in 70 % ethanol solution.
of each disinfectant. Leave the sensor to dry in air for at least 30 minutes,
Users in countries in which the DGHM list is not available otherwise the remaining alcohol can damage the
are advised to use the agents mentioned above. sensor when calibrated.
The flow sensor cannot be autoclaved.
Alcohols and agents containing alcohol must not be
allowed to seep into the sample line! ● The flow sensor can be reused as long as it can be
Alcohol can distort the measurement results. calibrated successfully.
● The flow sensor must be disposed of as infectious
special waste. It can be incinerated with little pollution
Breathing system
at temperatures of more than 800 °C.
All parts of the breathing system, valve discs, bellows,
bellows container, Y-piece, breathing hoses, water
trap containers, breathing bag, parts of the absorber
and Resutator 2000, parts of the secretion aspirator
and parts of the anaesthetic gas scavenging system
● Disinfect in hot liquid disinfectant – in a disinfecting
machine at 93 °C for 10 minutes.
Only use cleaning agent, no alkaline or chlorine-
releasing substances – risk of corrosion!
102
Downloaded from www.Manualslib.com manuals search engine
Care
Disinfecting / cleaning / sterilizing
All the components are listed in the care list for Juian on
pages 104 and 105, together with the corresponding
intervals and appropriate means.
103
Downloaded from www.Manualslib.com manuals search engine
Care
Care list
104
Downloaded from www.Manualslib.com manuals search engine
Care
Care list
The parts should preferably be cleaned and disinfected in a washing machine, otherwise by
immersion.
Julian and its components must not be treated with formaldehyde vapours or ethylene oxide!
How
Disinfection and cleaning4) Sterilization
Washing machine2) Wiping3) Immersion3) Steam, 134 °C,
93 °C 10 minutes 10 minutes
no outside no no
no outside no no
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
no yes no no
yes no possible yes
no outside possible 5) no
yes no possible no
no yes no no
yes no possible yes
yes no possible no
yes no possible no
yes no possible no
yes no possible yes
105
Downloaded from www.Manualslib.com manuals search engine
Care
Assembling
Assembling
Installing the breathing system
1 Place the bottom section on a flat surface.
2 Fit the valve plate on the bottom section. 5
3 Insert both valve discs.
4 Tightly fit the cover. 5 5
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
3
2
3
106
Downloaded from www.Manualslib.com manuals search engine
Care
Assembling
5
5
107
Downloaded from www.Manualslib.com manuals search engine
Care
Assembling
108
Downloaded from www.Manualslib.com manuals search engine
Care
Assembling
51
42
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
109
Downloaded from www.Manualslib.com manuals search engine
Care
Assembling
110
Downloaded from www.Manualslib.com manuals search engine
Care
Assembling
3 4
111
Downloaded from www.Manualslib.com manuals search engine
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
112
Downloaded from www.Manualslib.com manuals search engine
Julian as Wall-mounted Unit
Contents
Contents
Features................................................................................................ 114
Preparation............................................................................................ 114
Anaesthetic gas aspiration......................................................................114
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
113
Downloaded from www.Manualslib.com manuals search engine
Julian as Wall-mounted Unit
earth leads are located on the left and right sides of the
lower half of the unit.
The AGS anaesthetic gas scavenging line is built into the
unit.
The on/off switch and display symbols for anaesthetic
gas aspiration are located in the lower half of the unit.
The bacterial filter for secretion aspiration is located on
the left side of the unit.
The wall-mounted unit must be installed by qualified
personnel only.
We recommend DrägerService.
Preparation
Julian can be swung clear of the wall on its hinges to gain
access to the back panel.
1 Pull the handle to swing the unit about 30° clear of the
wall. D Julian
114
Downloaded from www.Manualslib.com manuals search engine
Maintenance Intervals
Contents
Maintenance Intervals
Contents
115
Downloaded from www.Manualslib.com manuals search engine
Maintenance Intervals
Maintenance Intervals
The workstation and its components must be cleaned and disinfected before every maintenance operation -
even when returning for repair!
(if the sample line is correctly installed and free of any blockage).
Can be disposed of as domestic waste.
O2 sensor Replace when calibration is no longer possible or when
the message FIO2 INOP! is displayed.
Disposal, see page 118.
116
Downloaded from www.Manualslib.com manuals search engine
Maintenance Intervals
Cleaning the cooling air filters
Replacing the water separator
Replacing the O2 sensor
● Replace the cooling air filters after not more than one 1
year. The filters can be disposed of as domestic
waste.
117
Downloaded from www.Manualslib.com manuals search engine
Maintenance Intervals
Disposing of batteries and O2 sensors
Disposing of the used device
118
Downloaded from www.Manualslib.com manuals search engine
Fault – Cause – Remedy
Contents
Contents
119
Downloaded from www.Manualslib.com manuals search engine
Fault – Cause – Remedy
Julian divides the alarm messages into three priority The messages are listed below in alphabetical order.
classes identified by exclamation marks: The list is intended to help identify the cause of an alarm
!!! Warning = Message with top priority message and to remedy the fault rapidly.
!! Caution = Message with middle priority
! Advisory = Message with least priority
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
120
Downloaded from www.Manualslib.com manuals search engine
Fault – Cause – Remedy
121
Downloaded from www.Manualslib.com manuals search engine
Fault – Cause – Remedy
JULIAN COM 2 ! Communication via the RS 232 inter- Check the plug connection on Julian and
face COM 2 has been interrupted. the connected device.
MV < !! Minute volume below the lower alarm Check breathing system.
limit. Check setting of ventilator.
Tube blocked/kinked. Check tube.
Leak in breathing system. Seal breathing system.
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
MV > !! Minute volume exceeds upper alarm Correct tidal volume or respiration rate.
limit.
MIXED AGENT ! The anaesthetic gas has been Check fresh gas setting. Wait a while.
confused. Mixture of two anaesthetic
agent present.
MIXER INOP !!! Fault in the fresh gas mixer system. Patient must immediately be
Automatic switchover to MAN/SPONT. ventilated by hand!
A constant flow of 4 L/min O2 is Call DrägerService.
maintained.
N2O INOP ! N2O gas measurement faulty. Call DrägerService.
N2O SUPPLY !!! N2O supply has failed. Open the N2O backup cylinder.
The central supply connector is not Check connection to central supply.
plugged in or the N2O hose is kinked. Restore N2O supply.
N2O cylinder empty. Connect a full N2O cylinder.
NO FRESHGAS !!! Fault on switching over to the external Call Dräger Service.
fresh gas outlet.
O2 INOP ! O2 sensor worn or defective Replace O2 sensor, page 117
O2 SUPPLY !!! The oxygen supply has failed. Open the O2 backup cylinder.
The central supply connector is not Check connection to central supply.
plugged in or the O2 hose is kinked. Restore O2 supply.
O2 cylinder empty. Connect a full O2 cylinder.
PAW > !!! Airway pressure exceeds upper alarm Check hose system.
limit.
Ventilation hose kinked.
Stenosis.
Pressure limit set too high. Correct pressure limit.
PAW NEGATIVE !!! Inadequate supply of fresh gas. Set adequate fresh gas flow on the
anaesthetic unit.
Mean pressure pmean less than –2 mbar.
Airway pressure PAW less than
–7 mbar.
P MAX? ! PCV pressure limit not reached. Check stroke volume.
Correct pressure limit, inspiratory flow
or inspiration time if necessary.
POWER FAIL ! Power supply failure. Check battery capacity.
Prepare manual ventilation.
Short-circuit in one of the devices Remove the device plug from the
connected to the auxiliary sockets. auxiliary socket. Press the switch on the
back panel.
122
Downloaded from www.Manualslib.com manuals search engine
Fault – Cause – Remedy
PRESSURE EXP > !!! End-expiratory pressure is more than Increase expiration time.
10 mbar above the set PEEP. Check hose system and microbial filter.
Drain water trap.
PULSE SPO2 < !!! Pulse rate has dropped below the set Check patient's condition!
alarm limit.
PULSE SPO2 > !! Pulse rate exceeds upper alarm limit. Check patient's condition!
Correct alarm limit if necessary.
SPO2 < !!! Oxygen saturation is below the set lower Check ventilation.
alarm limit.
Check O2 concentration of the fresh
gas.
SPO2 > !! Oxygen saturation exceeds the upper Check O2 concentration of the fresh
alarm limit. gas flow.
Check ventilation.
SPO2 INOP ! SpO2 measurement faulty. Call DrägerService.
SPO2 PULSE ? !!! No pulse signal detected by SpO2 Check patient's condition!
measurement for approx. 10 seconds.
Check SpO2 sensor!
SPO2 SENSOR? ! SpO2 sensor not plugged in. Check sensor connection.
SPEAKER FAIL ! No alarm tone, loudspeaker faulty. Call DrägerService.
SYSTEM ERROR !!! Internal device error.
Automatic switchover to MAN/SPONT.
Version without rotary knob for O2 Immediately ventilate the patient by
emergency metering: hand.
Julian automatically delivers an O2 flow
of 4 L/min.
Version with rotary knob for O2 Set »Safety O2« rotary knob for O2
emergency metering: emergency metering to desired O2 flow.
Set »Safety O2« rotary knob to desired Range 0 to 12 L/min.
O2 flow. Range 0 to 12 L/min. Immediately ventilate the patient by
This O2 flow flows through the Vapor hand.
unit. Check Vapor unit setting.
Call DrägerService.
VENT INOP !!! The ventilator control unit is performing After about 15 seconds, Julian is
an internal restart. This restart procedure restored to its previous operating mode.
takes about 15 seconds.
WATER TRAP ! The collecting jar of the water trap at Empty the jar.
the back of the unit is full.
The prism of the water trap is misted Wipe the prism dry.
over with condensation.
123
Downloaded from www.Manualslib.com manuals search engine
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
124
Downloaded from www.Manualslib.com manuals search engine
What's what
Contents
What's what
Contents
125
Downloaded from www.Manualslib.com manuals search engine
What's what
What's what
Front view
14
13 1
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
12
2
11
10 5
9
1-36-96 8
126
Downloaded from www.Manualslib.com manuals search engine
What's what
O2 Air N2O
5
Safety-O 2
6 4 2
8
Vac. 10 0 O2
12 +
L/min
4 3 2 1
1 O2 flush
2 O2 emergency metering
3 Secretion aspiration on/off switch
4 Rotary knob for secretion aspiration suction
5 Secretion aspiration pressure gauge
6 Central gas supply O2 pressure gauge
7 Central gas supply AIR pressure gauge
8 Central gas supply N2O pressure gauge
127
Downloaded from www.Manualslib.com manuals search engine
What's what
1 2 3 4 5
11
IPPV 12
Pleth
SpO2 98 67 alarm
13
limits
CO2 auto set
38 _
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
0
Fi Fet list ..
12... 14
20 PAW O2 40 36 curves
Hal. 0.8 0.6 config. 15
0
N 2O 58 56
16
Freshgas O2+N2O
Pmax VT Freq. TI:TE TIP:TI PEEP
O2 % L/min mbar mL 1/min % mbar
MAN
N 2O AIR
SPONT
IPPV PCV 17
18
6 7 8 9 10
1 Status field showing the current ventilation mode 10 Softkeys for setting ventilation
2 Curve field for curves and bar graphs 11 Power supply indicator
3 Alarm field for alarm messages and their priority 12 Alarm indication LEDs
4 Measurement field for numerical values 13 Key for suppressing the acoustic alarm for 2 minutes
5 Softkeys for monitoring functions 14 Key for changing monitoring screen pages
6 Keys for selecting the carrier gas (N2O or AIR) 15 Key for calling up the standard screen
7 Softkeys for setting the fresh gas flow 16 User advisory field
8 Status line for fresh gas 17 Rotary knob for selection, setting and confirmation
9 Keys for selecting the ventilation mode 18 Key for switching to standby mode
128
Downloaded from www.Manualslib.com manuals search engine
What's what
Rear view
11 12 13
1
2
3
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
5
10 6
7
9
MT-490-98
129
Downloaded from www.Manualslib.com manuals search engine
What's what
Interface panel
Analog 1
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
[
Sync.
CAN 2 COM 3 2
SpO2
m
CAN 1 COM 2 3
Temp.
COM 1 4
SA-Bus 5
6 7 8 9 10
130
Downloaded from www.Manualslib.com manuals search engine
Technical Data
Contents
Technical Data
Contents
Ambient conditions...............................................................................132
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
Ventilator...............................................................................................132
Breathing system..................................................................................133
Measuring systems...............................................................................133
Pressure measurement ..........................................................................133
O2 measurement....................................................................................133
Flow measurement................................................................................. 134
Interfaces.............................................................................................. 136
131
Downloaded from www.Manualslib.com manuals search engine
Technical Data
Technical Data
Ambient conditions
In operation:
Temperature 10 to 35 °C
Air pressure 700 to 1060 hPa
Rel. humidity 20 to 80 % (no condensation)
CO2 concentration of ambient air 300 to 800 ppm
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
In storage:
Temperature –20 to 60 °C
O2 sensor max. 50 °C
Air pressure 500 to 1100 hPa
Rel. humidity <98 % (no condensation)
Fresh gas
(electronic mixer)
Settings:
O2 concentration 25 to 100% by vol., at least 25% by vol., 250 mL/min.
± 5% of set value
Fresh gas flow 0 and 0.5 to 12 L/min
± 10% of set value or max. ±100 mL
O2 flush >35 L/min
Ventilator
(electronically controlled, pneumatically
driven bellows in bottle ventilator)
Settings:
Pressure limit Pmax
for IPPV 10 to 70 mbar
±10 % of setting or at least ±2 mbar
for PCV (PEEP+1) to 70 mbar
±10 % of setting or at least ±2 mbar
Tidal volume 50 to 1400 mL
± 15% of the setting for inspiration flows ≥ 10 L/min.
Ventilation frequency 6 to 60 per minute
Tinsp 0.2 to 6.7 seconds
Texp 0.33 to 8 seconds
Insp/Exp. time ratio TI : TE 1 : 4 to 2 : 1 with automatic ventilation
132
Downloaded from www.Manualslib.com manuals search engine
Technical Data
Breathing system
Total gas volume Approx. 4.5 L enclosed gas volume
Compliance with filled absorber container,
without breathing hoses Approx. 4.5 mL/mbar
Absorber volume 1.5 L
Leakage (EN 740) <150 mL/30 mbar
Pressure limiting valve APL
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
Measuring systems
Pressure measurement
(piezoresistive)
Airway pressure –10 to 99 mbar
Resolution 1 mbar
Accuracy Better than ±4 % of measured value or at least 1 mbar,
the larger value applies.
O2 measurement
(fuel cell in sidestream)
Sample rate 60 mL/min or 200 mL/min
Delay time for sampling Less than 1 second
Measuring range 5 to 100 vol%
Resolution 1 vol%
Accuracy Calibration in air:
±3 vol% in the range from 5 to 50 vol%
±5 vol% in the range from 50 to 100 vol%
Calibration with 100 % O2:
±3 vol% in the range from 5 to 100 vol%
Drift Within the ranges specified above
Response time t10...90
at 200 mL/min 500 ms
at 60 mL/min 1s
133
Downloaded from www.Manualslib.com manuals search engine
Technical Data
Flow measurement
(hot wire anemometry)
Tidal volume VT
Range 0.02 to 9.99 L
Resolution 0.01 L
Accuracy (under calibration conditions Better than ±8 % of the measured value or 0.01 L, the larger value
and 1013 hPa) applies
Minute volume MV
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
CO2 measurement
(infrared spectrometry)
Sample rate 60 mL/min or
(selectable) 200 mL/min
Delay time for sampling Less than 1 s
Measuring range 0 to 9.9 kPa (0 to 80 mmHg)
Resolution 0.1 kPa
Accuracy ±0.2 vol% or 5 % of measured value
Response time t10...90
at 200 mL/min 300 ms
at 60 mL/min 1s
Warm-up phase 4 min (accuracy as specified by DIN EN 864)
2 min for first measured values
Zero drift Within the ranges specified above
___________
134
Downloaded from www.Manualslib.com manuals search engine
Technical Data
Accuracy for
0 to 4 vol% Better than ±0.2 vol% absolute
4 to 8.5 vol% Better than 10 % of measured value
Display range for enflurane and sevoflurane 0 to 9.9 vol%
Accuracy for
0 to 4 vol% Better than ±0.2 vol% absolute
4 to 9.9 vol% Better than 10% of measured value
Display range for desflurane 0 to 22 vol%
Accuracy for
0 to 8 vol% Better than ±0.4 vol% absolute
8 to 22 vol% Better than 10 % of measured value
Response time t 10...90
at 200 mL/min 350 ms
at 60 mL/min 1s
Warm-up phase 4 min (accuracy as specified by DIN ISO 11196)
for first measured values
Zero drift Within the ranges specified above
Automatic anaesthetic gas identification The measurement accuracy required by ISO standards
is attained at the latest 4 minutes after switching on the
apparatus. The threshold level for automatic identification
equals 0.15 % by volume.
Anaesthetic gases are identified from a concentration of
0.4 % by volume upwards. If a mixture is present, the gas
with the higher concentration is measured.
135
Downloaded from www.Manualslib.com manuals search engine
Technical Data
Sensors
Type Compatible with Nellcor sensors
Oxisensor, Oxiband and Durasensor
Wavelengths 660 nm (red)
920 nm (infrared)
Acoustic pulse signal A tone is produced for each detected pulse,
tone pitch proportional to oxygen saturation
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
Interfaces
Analog output
Plug connector 9-pin sub-D, electrical isolation 1.5 kV
PAW 0 to 100 mbar; 0 to 5 V =; PIN 3
Flow 0 to 150 L/min; 0 to 5 V=; PIN 6
CO2 0 to 10 kPa; 0 to 5 V=; PIN 9
Earth PIN 1, 4, 7
COM1 record (printer)
Connector 9-pin sub-D, electrical isolation 1.5 kV
Pin assignment 1 shield
2 TxD
3 RxD
5 GND
COM2 (MEDIBUS 1)
Connector 9-pin sub-D, electrical isolation 1.5 kV
Pin assignment 1 shield
2 TxD
3 RxD
5 GND
COM3 (MEDIBUS 2)
Connector 9-pin sub-D, electrical isolation 1.5 kV
Pin assignment 1 shield
2 TxD
3 RxD
5 GND
136
Downloaded from www.Manualslib.com manuals search engine
Technical Data
Operating data
Operating voltage 90 – 265 V~
Power input 400 W typically, max. 2.3 kW
Uninterruptible power supply with fully charged battery: typically 30 min
(auxiliary sockets are not powered!)
Classification II b
according to Directive 93/42/EC
Appendix IX
137
Downloaded from www.Manualslib.com manuals search engine
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
138
Downloaded from www.Manualslib.com manuals search engine
Abbreviations / Symbols
Contents
Abbreviations / Symbols
Contents
Abbreviations........................................................................................ 140
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
139
Downloaded from www.Manualslib.com manuals search engine
Abbreviations
Abbreviations
ç
saturated gas
O2 Oxygen
Conformité Européenne
PAW Airway pressure
Directive 93/42/EC
on Medical Products PCV Pressure Controlled Ventilation
CC-Lock The SpO2 signal is synchronized with the PEAK Peak pressure
ECG signal
PEEP Positive end-expiratory pressure
Compl. Compliance
PLAT Plateau pressure
Des. Desflurane
Pleth Plethysmogram
etCO2 End-expiratory CO2 concentration
Pmax Maximum pressure
Enf. Enflurane
Sev. Sevoflurane
Fet End-expiratory fraction
SpO2 Functional O2 saturation
Fi Inspiratory fraction
Tip:Ti Ratio of inspiratory pause time to
Fi Des Inspiratory desflurane concentration inspiration time
Fi Enf Inspiratory enflurane concentration TI:TE Ratio of inspiration time to expiration time
Fi Hal Inspiratory halothane concentration USV Uninterruptible power supply
Fi Iso Inspiratory isoflurane concentration VAC Vacuum (e.g. for secretion aspiration)
Fi Sev Inspiratory sevoflurane concentration VT Tidal volume
FLOW Expiration flow ZV Piped medical gas supply for O2, N2O,
AIR and vacuum
Fi O2 Inspiratory O2 concentration
Hal. Halothane
HLM Alarm mode for heart-lung machine
INOP Function failure
IPPV Intermittent Positive Pressure Ventilation
Iso. Isoflurane
140
Downloaded from www.Manualslib.com manuals search engine
Symbols
Symbols
Symbol Meaning
ª Pulse rate
u Action in progress
? Repeat calibration
! Advisory message
!! Caution message
141
Downloaded from www.Manualslib.com manuals search engine
Index
Index
Abbreviations ......................................................... 140 COM3 ..................................................................... 85
Absorber, dismantling ............................................... 99 Compliance ...................................................... 32, 133
Absorber, filling and installing ................................. 107 Configuring in operation ........................................... 68
Accessories ................................................................ 5 Configuring in standby .............................................. 82
Advisory ........................................................... 63, 120 Connection ................................................................. 5
Airway temperature measurement ............................. 80 Cooling air filter, cleaning ............................... 116, 117
Alarm ................................................................. 63, 75 Curves ............................................................... 66, 88
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
142
Downloaded from www.Manualslib.com manuals search engine
Index
143
Downloaded from www.Manualslib.com manuals search engine
These Instructions for Use apply only to
Julian
with Serial No.:
If no Serial No. has been filled in by
Dräger these Instructions for Use are
provided for general information only and
are not intended for use with any specific
ç
machine or device.
Directive 93/42/EEC
concerning Medical Devices
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
90 29 439 - GA 5132.005 en
Dräger Medizintechnik GmbH
1st edition - December 1998
Subject to alteration