Sulphur Compounds TD GC MS
Sulphur Compounds TD GC MS
Tel: +44 (0) 1443 230 935 Fax: +44 (0) 1443 231531
http://www.markes.com
Sulphur compounds are probably best associated with their unpleasant, pungent odours, noticeable even
at low concentrations. These compounds, especially when in contact with metals, become increasingly
sensitive to high temperatures. However, despite the limitations this property brings to analysis, the
detection of such compounds is of particular importance to a wide range of industries such as flavour and
fragrance testing, the purification of water and food studies.
The following method was developed to analyse a standard sulphur solution obtained from Chemservice
and consisting of methylmercaptan, methyl sulphide, acetaldehyde, dimethyl-disulphide and styrene (1%
in methanol). The standard solution was introduced onto a Silcosteel® coated stainless steel thermal
desorption tube containing a 40 mm bed of Tenax™ backed up by a 15 mm bed of UniCarb™. Four
different volumes of sample were injected (5 ng, 0.5 µl, 1 µl and 2 µl) respectively, in a flow of inert gas
(helium) at a rate of 50 ml/min and using the calibration standard solution loading rig from Markes
International. The samples were then desorbed using the UNITY™ thermal desorber from Markes
International linked to an Agilent 6890 GC and 5973 MS. The analytical conditions are given below.
The analytical conditions and results obtained from a similar application carried out in the field and using
a UNITY - Air server system from Markes International are also given. On this occasion the calibration
gas used consisted of hydrogen sulphide, methylmercaptan, ethylmercaptan, dimethyl sulphide, carbon
disulphide and dimethyl-disulphide with a concentration of approximately 150-400 ppb. The standard was
sampled for 3 minutes at a rate of 30 ml/min resulting in a total sample volume of 90 ml. Since the
sampling was carried out on-line, no sorbent tube was required - the standard was sampled directly onto
the focusing trap of UNITY.
Analytical Conditions
Results
Dimethyl disulfide
Dimethyl disulfide
Dimethyl disulfide
February 2003
Styrene
Styrene
Styrene
standard solution
standard solution
standard solution
Page 2 of 4
Figure 2. 1 µl sample of
Figure 3. 2 µl sample of
Figure 1. 0.5 µl sample of
TDTS 32 February 2003 Page 3 of 4
Figure 5. 2 ng sample of
standard solution. This
equates to approximately 2
ppb per litre of air
Results
carbon disulphide
dimethyl-disulphide
hydrogen sulphide
ethyl mercaptan
UNITY-Air Server-GC.
methyl mercaptan
Summary
Both sets of experiments demonstrated the ability of UNITY to successfully thermally desorb a mixture of
sulphur compounds, despite their sensitivity to heat and reaction with metals. This clearly confirms the
inertness of the flow path in UNITY. From Figure 4 it is evident that there is a linear relationship between
the amount of sample introduced to the tube and the peak area produced. Again this emphasizes the
inertness of the UNITY and UNITY-Air Server flow path since all the sample was desorbed and
subsequently passed onto the GCMS. From the low level standards (Figure 5), it also appears that
quantitation limits in the order of 1 ppb can be readily achieved from as little as 1 L of air.
Sulphur compounds can therefore be analysed successfully using an on-line sampling method via the
UNITY-Air Server system, or the more conventional passive/diffusive method via sorbent tubes and a
UNITY thermal desorber.
Trademarks
UNITY™, Air Server™, UniCarb™ are trademarks of Markes International Ltd., UK
Tenax™ is a trademark of Buchem B.V. The Netherlands
Carboxen 1000™ is a trademark of Supelco Inc., USA
Silcosteel™ is a trademark of RestekInc., USA