List of Spec

The document provides a list of equipment for a biomedical workshop including multi-meters, clamp meters, soldering stations, de-soldering stations, voltage/current sources, RCL meters, and other electrical tools. The list includes 9 items with descriptions of each item detailing specifications such as voltage and current ranges, accuracy levels, and resolution. The estimated budgets are also provided for each quantity listed.

Uploaded by

Flame St
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
85 views

List of Spec

The document provides a list of equipment for a biomedical workshop including multi-meters, clamp meters, soldering stations, de-soldering stations, voltage/current sources, RCL meters, and other electrical tools. The list includes 9 items with descriptions of each item detailing specifications such as voltage and current ranges, accuracy levels, and resolution. The estimated budgets are also provided for each quantity listed.

Uploaded by

Flame St
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
You are on page 1/ 17

EQUIPMENT LIST FOR BIOMEDICAL WORKSHOP, FLOW METERS AND AUTOCLAVE

# Name of Description Unit Quantity Est. Remark


Equipment Budget

1. True rms Voltage DC Accuracy1 ±(0.09% + 2) Pcs 4


multi-meter
Max. resolution 0.1 mV

Maximum 1000 V

Voltage AC Accuracy1 ±(1.0% + 3)

Max. resolution 0.1 mV

Maximum 1000 V

Current DC Accuracy1 ±(1.0% + 3)

Max. resolution 0.01 mA

Maximum 10 A

Current AC Accuracy1 ±(1.5% + 3)

Max. resolution 0.01 mA

Maximum 10 A

Resistance Accuracy1 ±(0.9% + 1)

Max. resolution 0.1 Ω

Maximum 50 MΩ
1
Capacitance Accuracy ±(1.2% + 2)

Max. resolution 1 nF

Maximum 10,000 µF

Agency approvals UL, CSA, TÜV listed


and VDE

Battery life Alkaline ~200 hours


typical
2. True rms AC current Range 40.00 A / 4
clamp meter 400.0 A
Accuracy 2% ± 5 digits
(45 Hz to 65
Hz)
2.5% ± 5
digits (65 Hz
to 400 Hz)
DC current Range 40.00 A /
400.0 A
Accuracy 2% ± 5 digits
AC voltage Range 600.0 V
Accuracy 1.5% ± 5
digits
DC voltage Range 600.0 V
Accuracy 1.0% ± 5
digits
Resistance Range 400.0 Ω /
4000 Ω /
40.00 kΩ
Accuracy 1.0% ± 5
digits
Continuity ≤ 30 Ω
Capacitance 0 to 100.0 μF /
100μF to 1000
μF

Frequency 5.0 Hz to
500.0 Hz
AC response
True-RMS
Backlight Yes
Min/Max Yes
Category rating CAT III 600
V
CAT IV 300
V
3. Electronics Complete kits for maintenance purpuse 2
tool box,
complete
4. Electrician tool Allen key set of ten set 2
box, complete
Combination pliers

Long nose pliers

Philips screw drivers: #0, #1, #2, , #3, #4

Screw drivers flat : Stubby, 1/4”x4”,5/16”x6”,3/8”x8”,


3/8”x10”

Miniature long nose pliers N-25

Sniper with stripper cut N-33


Cutter with blade

Center punch set of 3

Open end wrench: 4x5, 6x7, 8x9, 10x11, 12x13, 14x15 mm

Wire stripper

Electrical terminal press pliers

Adjustable driver

Reversible flat and Phillips

Miniature snipper

End-snipper

Electric knife

Hacksaw small size with a pair of blade

Nut drivers: 4, 5, 5.5, 6, 7, 8 , 9,10, 11, 12, 13mm

Forceps Normal

Scissors

Heat radiation clip

File set of 5

Hammer

Soldering aid set of 3


De-soldering bridge

Core driver

Solder puller

Soldering cleaner

Pin vice

Forceps reverse action

Measuring tape 2m

Adjustable wrench

Electric tester

Soldering iron with box

Cable stripper
 100 Watt Solder and Disoldering Station, with internal
5. Soldering vacuum pump. set 2
station
 ESD Safe and Spike Free circuitry zero voltage
protection
 SMD Tweezers ,Iron Working Option mandatory both
at the same time
 Temperature range: 177 to 454(C
 Temperature accuracy calibration range: (5(C
 Temperature stability: (6(C
 With temperature lock out
 With accessory soldering tips: 1.6mm, 0.48mm and
0.8mm
 Soldering iron with pre-loaded standard tip
 Disoldering iron with pre-loaded standard tip
 2 iron holders
 Maintenance kit
 426DLX Fume extractor
 AC power cord and plug
 Operation manual

 Soldering pencil: 50 Watts


 Input voltage: 220V, 50Hz
5. Core Solder  8.0 oz set 2

6 ESR meter Analyzer type: ESR and Capacitance.

Component types: Capacitors (>1uF).

ESR range: 0.00 Ohms to 40.0 Ohms.

ESR resolution: From 0.01 Ohms.

In-circuit use: ESR only.

Capacitance range: 1uF to 22000uF.

Battery type: 1.5V Alkaline AAA Cell (supplied)


6. Soldering iron  50W,220v set 2
 Electronically controlled zero crossover
7. De-soldering Pcs 2
station  Temperature range: 260(C to 540(C
 Closed loop electronic temperature control
 With isolated output voltage
 With fully grounded de-soldering iron
 With replacement de-soldering irons
 Power supply: 220V, 50Hz
8. DC voltage Pcs 2
current source  Voltage source up to ±32 V and current source

up to ±200 mA

 51/2-digit, 120000-count output resolution


 Adjustable current output: 0-10A
 Load regulation: (0.01%+3mV; (0.02%+3mA
 Line regulation: (0.01%+3mV; (0.02%+3mA
 Ripple: (0.5mVrms; 3mArms
 With digital display
 Power: 220V, Adjustable voltage output: 0-35V,
50Hz

Accuracy:
±0.016% of setting + 240 μV (at 10 V range for one
year)
±0.03% of setting + 5 μA (at 100 mA range for one
year)
Stability:
±0.001% of setting 20 μV (at 10 V range for one day)
±0.004% of setting + 3 μA (at 100 mA range for one
day)
Resolution:
100 nV (VDC, 10 mV range)
10 nA (1 mA range)
Noise:
100 μVp-p (10 V range, DC to 10 kHz)
3 μAp-p (100 mA range, DC to 10 kHz)

 Frequency range: 12Hz - 100KHz


9. RCL meter Pcs 2
(digital)  Accuracy: 0.02%
 Test voltages: 5mV - 1.275V

 Automatic Zero Trimming


 Connectors: 2 mm sockets
 Resistance (Rp, Rs, Z): 0.000Ω to 200 MΩ
 Capacitance (Cp, Cs): 0.0 pF to 100 mF
 Inductance (Lp, Ls): 0.0 mH to 32 kH
 Q and D factor: 0.002 to 500
 Phase angle F: -90.0 to +90.0 deg
Mandatory Components
 Two test pods
 Kelvin Cable
 User Manual
 Operating Card
 Instruction manual
 Mains Cable
Wire Test Adapter
Wire Test Adapter SMD

10. Dust blower  Continuous rating input 600w Pcs 3


 Max sealed suction 5.7kpa
 220Vac
11. Electronics  Standard for work shop facility Pcs 2
workbench
12. Electrical  Standard for work shop facility Pcs 2
workbench
13 Bench light Long reach bench light Pcs 3

Steel construction

220V, 60 watt incandescent lamp

14. Digital Range: -100.0 to 199.9(C Pcs 2


infrared
thermometer Accuracy: (0.2(C

Battery powered

Resolution: 0.1(C

With cable and penetrating type probe

15. Venire caliper 20cm with inches and mm measurement capability Pcs 2

16. Tap and die set 2.5mm- 18mm diameter set 2


17. Mechanics tool Hexagon socket heads ¼ -( drive: 2, 2.5, 3, 3.5, 4, 4.5, 5, set 2
box complete
5.5, 6, 7, 8, 9, 10 up to 20mm

Ratchet wrench, jointed ( 3/8

Tommy bar 150mm ( 3/8

Spark plug socket 21m ( 3/8

Reducing adapter 3/8 – ¼

Screw driver handle ( ¼

Adjustable pliers, length 240mm

Combination pliers, 160mm

Screw driver 100mm , 1.0 x 6

Screw driver 40mm , 1.0 x 6

Allen keys on ring 1.5, 2, 2.5, 3, 3.5, 4, 4.5, 5, 5.5, 6, 7, 8,


9, 10mm

Combination spanners 4, 5, 6, 8, 10, 11, 12, 13, 14 up to


30mm

18. Advanced Pcs 2


semiconductor
analyzer
Analyzer type: Semiconductor components.

Component Automatic.
detection:

Pinout detection: Automatic, connect any way round.

Display: Graphic LCD.

Interface type: USB for optional PC connection.

PC functions: Curve tracing.

Software: Included on USB drive.

Battery: Single AAA cell (supplied)

19 Electronic Pcs 3
data book
24. Portable capacity 8mm, with chuck, key and cable, suitable for an Pcs 2
electric drill
electric supply of 220v,AC single phase 50hz

25. Goggle clear For all around the eye protection and ventilation slots Pcs 10
glass
26. Glove ,rubber heavy duty, suitable for electrical work pack 10
TECHNIAL SPECIFICATION FOR BS FLOW METER WITH HUMIDIFIER

Mat type Flowmeter with humidifier

Gas type Oxygen

Probe type British standard

Adapter type British type

Flow rate 0-15lpm

Input pressure 0.4 MPa

Inlet thread 1/4NPT-18-RH

outlet thread 8mm

Humidifier bottle 200ml


capacity

certification CE, ISO13485

quantity 100 pieces


TECHNICAL SPECIFICATIONS: HIGH PRESSURE AUTOCLAVE

1- Horizontal Rectangular High Pressure High Vacuum steam Sterilizer.


2- Chamber size shall be minimum of 2 feet width x 2 feet height x 4 feet length.
3- Should be operated on 380-440V, 3 phase with neutral, AC power supply.
4- Chamber volume shall be minimum of 500 liters.
5- Steam generator volume shall be minimum of 60 liters and should be fitted with suitable
Electrical heater load of 18KW, to produce steam to sterilizer.
6- The sterilizer shall have single door (Hinge type) with radial locking using shooting bolts
Having high pressure locking safety facility and made of good quality stainless steel 304
Quality. Should provide heat resistant SILICON door gasket withstand up to 140 Deg. C.
7- The sterilizer shall have to draw the water, automatically, when needed in the inbuilt boiler.
8- Thickness of chamber of sterilizer shall be minimum 6mm and to be made of stainless steel
316 quality is capable to withstand the negative pressure (vacuum) of 24-26’’Hg created by the
Vacuum pump.
9- Thickness of door of sterilizer shall be minimum of 12mm and to be made of stainless steel
316 quality.
10- Sterilizer jacket shall be made of Boiler quality steel plate with a material thickness of 6-8mm.
11- Glass wool insulation thickness shall be 50mm. Insulation cover shall be made of good quality
Stainless steel 304 quality.
12- All connecting pipes shall be made of good quality stainless steel.
13- Stand shall be made of Mild steel with anticorrosion paint.
14- The unit shall be fitted with suitable water ring Vacuum pump, motor capacity: 3 HP, to create
High vacuum of 26”hg for efficient drying and sterilization of loads. The sound level of vacuum
Pump shall be <80DP and no vibration. Vacuum pump. Fitted with suitable stainless steel
Condenser, piping’s and pipes to enable efficient drying with the use of the Vacuum pump.
15- Pressure gauge range shall be 0 to 6 Kg. /cm2.
16- Compound gauge range shall be – 1 to 6 Kg/cm2.
17- Safety valve range shall be 0.3 to 3.5kg/cm2.
18- The unit shall be capable of being stored continuously in ambient temperature of 0-50 Deg C
And relative humidity of 15-90%.
19- Suitable validation port shall be provided.
20- Working temperature of sterilizer is 121-134 Deg. C and the corresponding pressure is 1.2-2.1
Kg/cm2.
21- Safety features of sterilizers: Door locking facility, Low water protection system, Pressure cut
Off facility and all other necessary safety features.
22- Equipment shall have no sharp edges, will be securely mounted and would provide adequate
Protection against moving and electrically energized parts.
23- Controls (e.g. switches, knobs) shall be visible and clearly identified.
24- Labels and markings shall be clear and visible.
.
25- Equipment shall be simple to use, operate and maintain (User friendly). It shall be designed for
Easy access to serviceable parts.
26- AUTOMATIC OPERATION WITH PRINTER:
(A) The sterilizer shall be fitted with suitable PLC (Microprocessor) for fully automatic cycle
Operation instead of manual operating valve with following Features:
(i) PLC based microprocessor which is incorporated with the sterilizer.
(ii) Digital displays of Chamber Pressure, temperature, cycle no., Batch no., Time &
Date, alarm indicator, Low water indicator.
(iii) Provision of ‘error code analysis’ inbuilt, Leak test, Bowie & Dick and Standard
Process.
(B) Printer: Printer that shall automatically and continuously monitor and record dates, time
Of day, load, identification no. and operating parameters.
(C) The system shall be designed, primarily, for carrying out the following:
 Leak test cycle
 Bowie & Dick Process
 Standard Process
 High pressure and High Vacuum Process – vacuum holding.
27- STANDARDS & SAFETY:
(A) The unit should be manufactured as per IS specifications Mark ISI: 3829 and also should bear the
Certification.
(B) Electrical safety shall conforms to standards for electrical safety IEC-60601-2-25 Safety of
Electrocardiograms (OR EQUIVALENT BIS Standard).
(C) Electrical safety shall conforms to standards for electrical safety IEC-60601/IS-12450
(D) Equipment performance should not be affected by electromagnetic interference radiated or
Conducted through power lines from another device.
(E) Necessary operational training / day-to-day maintenance training shall be imparted to our staff
After commissioning of the equipment at site.
28- AFTER SALES SERVICE:
(A) After-sales-service/ maintenance shall be provided from your factory trained engineer.
(B) Response time from time of lodging the complaint shall be 24-36 hours and total uptime in a
Year shall be not less than 97% including PPM.
(C) Back to back warranty to be taken by the supplier from the principal to supply spares for a
Minimum period 10 year.
(D) Should have service facility. The service provider should have the necessary equipment’s
Recommended by the manufacturer to carry out preventive maintenance test as per guidelines
Provided in the service/maintenance manual.
29- SPARES AND ACESSORIES TO BE PROVIDED ALONG WITH EQUIPMENT:
(A) Spare fuses 10nos.shall be provided, if fuses are used.
(B) Spare water reading glass and washers shall be provided
(C) One number single water still (distillation) with following features: Electrically operated water
Still capable of providing 10 liter / hour pyrogen free distilled water, made of stainless steel,
Having immersion water heater (heating coils) with low water protection (low water cut off
Device).
(D) 60 liter plastic tank to store distilled water, fitted with two outlet taps to withdraw water.
30- DOCUMENTATION – SHALL BE PROVIDED AT THE TIME OF DELIVERY:
(A) Operator’s / instruction/ user manual in English should be provided.
(B) Installation qualification
(C) Operation qualification
(D) Design qualification
(E) Performance qualification
(F) Hydraulic test certificate
(G) Material test certificate
(H) Gauge calibration certificate
(I) Master gauge calibration certificate
(J) Warranty certificate
31- Vendor has to support the specifications with manufacturer’s brochure failing
Which offer may be rejected? Vendor has to demonstrate the equipment at
Hyderabad, within specified time limit, if asked for; failing which offer will
Be rejected.
32- Installation (Erection & commissioning) and training to be provided by vendor.