Model 230: True Wet-to-Wet Differential Pressure Transducer
Model 230: True Wet-to-Wet Differential Pressure Transducer
Applications:
Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com © Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
Model 230
True Wet-to-Wet Differential Pressure Transducer
0 to 10.0 100 25 Thermal Effects 2 Sensor Cavity Volume 0.27 in3 Positive Port, 0.08 in3 Negative Port
0 to 25.0 350 62.5 (With 1/4”NPT external fittings installed-does not include cavity volume of 1/4”NPT external
Compensated Range °F(°C) +30 to +150 (-1 to +65)
fittings.)
0 to 30.0 350 75
0 to 50.0 350 125
Zero Shift %FS/100°F(%FS/50°C) 2.0 (1.8) Physical Description (3-Valve Manifold Assembly)4
Span Shift %FS/100°F(%FS/50°C) 2.0 (1.8) Manifold Block Brass
0 to 100.0 350 250
Zero shift ±0.004% FS/PSIG line Valves (3)5 V1 for Connection to + port
Line Pressure Effect
pressure V2 for Connection to - port V3 for Equalizing Pressure
Bidirectional Infinite, limited only by output Valve Type 90° On/Off
Resolution
Pressure Range Proof Pressure Proof Pressure noise level (0.02%FS)
PSID High Side PSI Low Side PSI
Static Acceleration Effect 2%FS/g (most sensitive axis) Process Connections 1/4”-18 NPT Internal Thread
0 to ±0.5 50 1.25 Dimensions 7.05”W x 6.25”H x 2.16’’D
Natural Frequency 500 Hz (gaseous media)
0 to ±1.0 50 2.5
Warm-up Shift ±0.1% FS total Weight <2.5 lbs.
0 to ±2.5 100 6.35
Response Time 30 to 50 milliseconds Physical Description (5-Valve Manifold Assembly)6
0 to ±5.0 100 12.5
Long Term Stability 0.5%FS/1 YR Manifold Block Brass
0 to ±10.0 200 25 Maximum Line Pressure 350 PSIG Valve (5) 5
V1 for Connection to ± Port
0 to ±25.0 350 62.5 V2 for Connection to – Port
Environmental Data V3 for Equalizing Pressure
0 to ±50.0 350 125 V4 & V5 for Connection to External
Operating3 Temperature °F (°C) 0 to +175 (-18 to +80) Gauge or Alternate Plumbing Configuration
Storage Temperature °F (°C) -65 to +250 (-54 to +121) Process Connection 1/4”-18 NPT Internal Thread
Vibration 5 g from 5 Hz to 500 Hz Dimensions 7.05”W x 6.25”H x 2.16’’D
Gases or liquids compatible with 360 brass, Copper 122, Acetal plug External Load 0 to 1000 ohms
valves and Nitrile O-rings. Minimum supply voltage (VDC) 9+ 0.02 x (Resistance of receiver plus line).
1
RSS of Non-Linearity, Hysteresis, and Non-Repeatability.
2
Units calibrated at nominal 70˚F. Maximum thermal error computed from this datum.
Maximum supply voltage (VDC) 30+ 0.004 x (Resistance of receiver plus line).
3
Operating temperature limits of the electronics only. Pressure media temperatures may Specifications subject to change without notice.
be considerably higher.
4
Order assembled with the Model 230 (Code 3V) or separately as Option 891.
5
Refer to drawings
6
Order assembled with the Model 230 (Code 5V)
7
Calibrated into a 50K ohm load, operable into a 5000 ohm load or greater.
8
Zero output factory set to within ±25mV (for 5 VDC output) or ±50mV (for 10 VDC
output)
Span (Full Scale) output factory set to ±25 mV (for 5 VDC output ) or ± 50 mV (for 10
VDC output
9
Calibrated at factory with a 24 VDC loop supply voltage and a 250 ohm load.
10
Zero output factory set to within ±0.16mA. Span factory set to within ±0.16 mA
Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com © Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
Model 230
True Wet-to-Wet Differential Pressure Transducer
54.6
2.15
198.3
7.81
142.9
V3 SHUNT VALVES
5.63
SHUT
OFF
VALVES
V1 V2
82.6
3.25
101.6
4.00
54.4
2.14
For differential pressure measurements at high line pressure (350 PSIG max), it is recommended that the pressure sensor be installed with a valve in each line, plus
a shunt valve across the high and low (reference) pressure ports as shown.
Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com © Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
Model 230
True Wet-to-Wet Differential Pressure Transducer
ORDERING INFORMATION
2 3 0 1
Model Range Pressure Fitting Output Bleed Screw Seals Optional
Calibration
2301 = 230 Unidirectional Bidirectional 2F 1/4” NPT (F) 11 4-20 mA Std. B Viton C
Certificate
0.05-5.05
001PD 0 to 1 PSID 0R5PB ±0.5 PSID 3V 3-Valve Manifold 2D Opt. A Buna-N
VDC
0.05-10.05
002PD 0 to 2 PSID 001PB ±1 PSID 5V 5-Valve Manifold 2E
VDC
005PD 0 to 5 PSID 2R5PB ±2.5 PSID
010PD 0 to 10 PSID 005PB ±5 PSID
025PD 0 to 25 PSID 010PB ±10 PSID
030PD 0 to 30 PSID 025PB ±25 PSID
050PD 0 to 50 PSID 050PB ±50 PSID Please contact factory for versions not shown.
100PD 0 to 100 PSID
Ordering Example: 2301005PD2F11B = Model 230 0 to 5 PSID unidirectional, 1/4-18 NPT Ext. fitting, 4 to 20 mA Output, and Viton/Silicone Seals.
2301005PD3V11B = Model 230, 0 to 5 PSID unidirectional, 3-Valve Manifold, 4 to 20 mA, Output, and Viton/Silicone Seals (Assembled w/3- Valve Manifold).
[ 179 ]
7.05
[ 59 ]
102 2.3
4.00
V6 V7
MODEL 230
DIFFERENTIAL PRESSURE
TRANSDUCER
SHUNT VALVE
V3
V4 V5 Low Process/
Commission 198
1/4”NPT 7.8
V1 V2 Connection
149
5.9
4
.156
10 83 8
.38 3.25 .31
For differential pressure measurements at high line pressure (350 PSIG max), it is recommended that the pressure sensor be installed with a valve in each line, plus
a shunt valve across the high and low (reference) pressure ports as shown.
Note: V6 and V7 bleed valves are not required when used with a Setra Model 230. Use the bleed screws on Model 230 to bleed the lines of air.