0% found this document useful (0 votes)
141 views3 pages

Specs For Basic USG

This document provides specifications for a black and white ultrasound machine, including technical requirements, accessories, warranty, and documentation needs. The machine must be a general purpose, fully digital ultrasound system with multiple imaging modes and a minimum 12" LCD monitor. It requires a battery backup, two probe ports, image storage for 200 patients, and DICOM connectivity. The specifications also list the needed system configuration, probes, printer, and service/operating manuals.

Uploaded by

UMHT BIO-Medical
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as XLSX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
141 views3 pages

Specs For Basic USG

This document provides specifications for a black and white ultrasound machine, including technical requirements, accessories, warranty, and documentation needs. The machine must be a general purpose, fully digital ultrasound system with multiple imaging modes and a minimum 12" LCD monitor. It requires a battery backup, two probe ports, image storage for 200 patients, and DICOM connectivity. The specifications also list the needed system configuration, probes, printer, and service/operating manuals.

Uploaded by

UMHT BIO-Medical
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as XLSX, PDF, TXT or read online on Scribd
You are on page 1/ 3

Specification for Black and White Ultrasound Machine

SN Purchaser's Specification Bidder's Remarks


B/W USG Machine
Manufacturer
Brand
Type/Model
Country of origin
1 Description of function
1.1
A general purpose fully digital B/W Ultrasound imaging system.
2 Operation requirements
2.1 Should operate on 220VAC and shall have battery backup system
for emergency during mains failure.
2.2 Movement should be easier.
3 Technical Specification
3.1
Latest technology and all digital beam former general purpose
standalone ultrasound machine with integrated light weight
mobile cart.
3.2 Main applications: OB/GYN, abdominal, peripheral vessels and
small parts.
3.3 The system shall have at least 12” or higher flat LCD monitor with
tilt & swivel facilities.
3.4 Shall have B-mode, M-mode, B/M mode, 2B mode & 2D mode.
3.5
The system must have at least Two active probe ports for easy use
and convenient operation.
3.6 256 Grey shades for sharp contrast resolutions.
3.7 Controls for depth, gain compensation, body markers with
transducer position.
3.8 Shall have real time continuous dynamic focus.
3.9 Shall have facility for image zoom, freeze, text annotation.
3.1 The system shall have extensive calculation software package for
Ob/Gyn and general imaging.
3.11
The system must have provision for measurement and calculation
of distance, area, volume, heart rate and circumference on the
image.
3.12 The system shall have Tissue Harmonic Imaging.
3.13 Near and far gain adjustable.
3.14 Contrast, adjustable.
3.15 Focus: auto adjustable.
3.16 Shall have an alpha-numeric keyboard with easy access scans controls
and track ball and status display.
3.17 Cine memory of 250 frames for cine loop playback.
3.18 Frame rate: not less than 50fps.
3.19 Display depth: minimum 28-30cm.
3.2 Dynamic range, selectable up to approximately 165dB.
3.21 Image storage: Minimum 200 patient’s images on main unit.
3.22 System shall be DICOM ready and capable of being interfaced with
HIS/RIS/PACS.
3.23 Facility for future upgradeability.
3.24 Probe: 2 to 5 MHz convex probe for Obs. /Gyn. and abdominal
application is to be supplied.
4 System Configuration
4.1
System shall come with main unit, 1 probe, 1 unit of black and
white video thermal printer and Ultrasound gel warmer 1 unit.
5 Accessories, spares and consumables
5.1
All standard accessories and parts required to operate the
equipment, including all standard tools and lubrication materials,
to be included in the offer.
6 Standards and safety requirements
6.1 Must Submit CE/ FDA approved certificate.
6.2 Must submit electrical standard and safety reports.
7 User Training
7.1
User training should be provided on the site including how to
maintain the equipment
8 Warranty
8.1
Comprehesive warranty for 2 years after the date of acceptance.
8 Installation and commissioning
8.1
The bidder must arrange for the equipment to be installed and
commissioned by certified or qualified personnel.
9 Documentation
9.1 Operating manual in English.
9.2 Service manual in English.
9.3 List of important spare parts and accessories with their part
number and costing.
9.4 Certificate of calibration and inspection from factory.
dder's Remarks