Operation Manual: Impedance Audiometer AT235h
Operation Manual: Impedance Audiometer AT235h
Precautions
Notice - Be sure to insert the probe tip in a way which will assure an
airtight fit without causing any harm to the patient. Using a proper and
clean ear tip is mandatory.
Notice - We recommend using a new ear tip for each patient. If the
clinician rinses the ear tips they should be subjected to standard
disinfecting procedure between patients. This includes physically
cleaning the ear tip and use of a recognised disinfectant. Individual
manufacturer's instruction should be followed for use of this
disinfecting agent to provide an appropriated level of cleanliness.
Notice - Be sure to use only stimulation intensities acceptable for the
patient.
Notice - The transducers (headphones, bone conductor, etc.) supplied
with the instrument are calibrated to this instrument - exchange of
transducers requires a re-calibration.
Note!
Within the European Union it is illegal to dispose electric and electronic
waste as unsorted municipal waste. Electric and electronic waste may
contain hazardous substances and therefore has to be collected
separately. Such products will be marked with the crossed-out
wheeled bin shown below. The cooperation of the user is important in
order to ensure a high level of reuse and recycling of electric and
electronic waste. Failing to recycle such waste products in an
appropriate way may endanger the environment and consequently the
health of human beings.
The equivalent volume of the middle ear, also called the compliance,
is easily derived by subtracting the two volume measurements above.
This is done automatically on the MT10 and the result is presented as
"Compliance", measured in ml.
Understanding Tympanograms
General Considerations:
A given curve drawn in a co-ordinate system will always have its
shape dictated by the vertical and horizontal graduations. The
printout of the MT10 complies with the international standards in
this respect, and therefore may not produce tympanogram
shapes directly comparable to other instruments if these do not
meet the standard requirements.
The Peak:
The peak of the tympanogram will horizontally be placed at the
air pressure of the middle ear, as equal pressures on both sides
of the tympanic membrane produces the highest compliance of
the system. A slight deviation of the peak in the direction of the
air pressure sweep may be experienced, due to an inherent
hysteresis of the middle ear and the test equipment. A slower
sweep speed may diminish the offset.
The Height:
The height of the tympanogram from its more or less horizontal
bottom line (measurements made from start pressure) to the top
shows the difference in compliance between stiffened tympanic
membrane and max. compliance. This difference is referred to
as "compliance" and is a measure for the equivalent volume of
the middle ear.
The Shape:
The shape of the tymp curve will change when the stiffness of
the system is changed (e.g. by ossicular chain disruption, otitis
media, etc.), and this is a primary reason for the diagnostic
value of this measurement. However, normal ears show a great
variety of tymp shapes so this should never be taken as the only
basis for making a diagnosis. Furthermore, two different
abnormalities may have opposing effects, resulting in a normal
shape of the tymp curve.
Classification of Tympanograms
Tympanograms can be classified according to compliance (height,
3
measured in ml. or cm ), pressure at compliance maximum
(measured in daPa), rate of compliance change (gradient in %), and
shape. Please refer to the chapter "Examples of Interpretations" in this
manual for illustrations of the classic curve categories, and the names
given to them by Liden and Jerger. On the following pages a more
detailed description of each category is presented.
Peak Pressure
Negative Normal Pressure: Positive Absence of
Pressure: Pressure: Pressure
Peak:
1) Blocked 1) Ossicular bone 1) Early
Eustachia fixation acute 1) Middle ear
n tube 2) Adhesive fixation otitis effusion
2) Serous 3) Ossicular media 2) Open tymp.
otitis discontinuity membrane
media 4) Middle ear tumour 3) Artifact
5) Eardrum
abnormality
Amplitude
Increased Decreased Unchanged
Amplitude: Amplitude: Amplitude:
Shape
........................Slopes....................... Smoothness:
Decreased / Increased slope: Altered smoothness:
Flattened slope:
1) Eardrum 1) Eardrum
1) Serous otitis abnormality abnormality
2) Ossicular fixation 2) Ossicular 2) Ossicular
3) Tumours of middle discontinuity discontinuity
ear 3) Vascular tumours
4) Patulous
Eustachian tube
Noise:
Acoustical signals showing up in the reflex recording, yet
irrelevant to the Stapedius reflex. As the reflex measurements
are based on observing the change in sound intensity of a 226
Hz tone, as explained in "Popular Introduction to Impedance", it
is possible that environmental noise of this frequency entering
the ear will show up as part of the test result. This is a problem
inherent to the measuring method and therefore common to all
normal impedance meters. Heart beat, talking and external
noises are common causes of noise peaks seen on the reflex
curve.
Negative Onset:
It is quite common to see reflexes start out with a small negative
deflecting dip. In ears with stapedial otosclerosis this dip with an
additional dip at the end of the stimulation can be the only
reaction left from the contraction of the Stapedius muscle. Some
tumour ears have been reported to give only the negative onset,
but no further reaction.
Reflex Threshold:
For a given stimulus the lowest level that elicits a detectable
reflex. This is not an absolute measurement as no exhausting
norm exits defining stimuli and related reflex characteristics.
Therefore, differences in test setups and reflex evaluation will
produce somewhat different results. It is not uncommon to
report the reflex threshold as the intensity which produces a 1%
or 2% change in equivalent volume (Test "A" with 2%
sensitivity).
Figure 3:
Air
Touch
Reflex Decay:
Reflex decay is calculated as the reduction of size of the
acoustic reflex during the first 10 seconds of muscular
contraction. Most normal ears will be able to maintain a
Stapedius contraction for 10 seconds or longer for frequencies
below 1000Hz, at a level 10dB above reflex threshold. A high
reflex decay score could indicate VIIIth nerve disorders.
Another way of stating the reflex decay is by finding the number
of seconds it takes before the Stapedius contraction falls to 50%
of its initial maximum.
The fact that the light band of the probe goes off, and the test
starts is not an absolutely certain indication of a fit good enough
to produce valid decay test results. This is due to the prolonged
test time and the high sensitivity of the test.
Also great care should be taken to have the probe kept in a very
fixed position relative to the ear during testing.
Volume : Normal.
Pressure : -100 daPa to +100 daPa.
Ventilation : Present.
Reflex : Present.
Audiogram : No hearing loss.
Volume : Normal.
Pressure : -100 daPa to +100 daPa.
Ventilation : Present.
Reflex : Present or absent.
Audiogram : Sensory neural hearing loss.
Volume : Normal.
Pressure : -100 daPa to +100 daPa.
Ventilation : Present.
Reflex : (Abnormal Decay.)
Audiogram : Sensory neural hearing loss
(May be unilateral).
Volume : Normal.
Pressure : -100 daPa to +100 daPa.
Ventilation : Present.
Reflex : Present.
Audiogram : Normal.
Volume : Normal.
Pressure : -100 daPa to +100 daPa.
Ventilation : Present.
Reflex : Absent.
Audiogram : Conductive loss.
Volume :
Normal.
Pressure :
-100 daPa to +100 daPa.
Ventilation :
Present.
Reflex :
Absent (Present by contra-
lateral stimulation).
Audiogram : Conductive loss.
Volume : Normal.
Pressure : -100 daPa to +100 daPa.
Ventilation : Present.
Reflex : Absent / Present.
Audiogram : Conductive loss.
Volume : Normal.
Pressure : -100 daPa to +100 daPa.
Ventilation : Present.
Reflex : Present.
Audiogram : Normal.
Volume : Normal.
Pressure : Peak not obtainable.
Ventilation : -
Reflex : Absent or elevated (rare).
Audiogram : Mild to moderate conductive
loss.
Volume : Low
Pressure : Normal.
Ventilation : -
Reflex : Absent.
Audiogram : Mild to moderate conductive
loss.
Volume : Normal.
Pressure : -100 daPa to +100 daPa.
Ventilation : Present.
Reflex : Absent or elevated (rare).
Audiogram : Moderate conductive loss.
Volume : Normal.
Pressure : Negative / moderate.
Ventilation : -
Reflex : Absent.
Audiogram : Moderate conductive loss.
Volume : Normal.
Pressure : Negative / negative.
Ventilation : -
Reflex : -
Audiogram : Mild conductive loss.
Volume : Normal.
Pressure : Not obtainable.
Ventilation : -
Reflex : Absent.
Audiogram : Moderate conductive loss.
Volume : Normal.
Pressure : Negative.
Ventilation : Absent or poor
Reflex : Absent or elevated (rare).
Audiogram : Mild to moderate
conductive loss.
Patient Instruction:
Place the patient in a comfortable chair or on an examining
table if necessary. Small children may feel more comfortable
sitting on a parent's or nurse's lap. Show the probe to the patient
and then explain the following:
• That the aim of the test is to test the mobility of the eardrum.
• That the tip of the probe will be inserted into the ear canal,
and that it has to make a perfect seal.
• That a small amount of air will flow through the probe to
move the eardrum; this will produce a sensation equal to
pressing a finger slightly into the ear canal.
• That one or more tones will be heard during the test.
• That no participation is expected from the patient.
• That coughing, talking and swallowing will ruin test results.
Visual Inspection:
Check the external ear canal for wax with an otoscope and remove
excessive wax to prevent probe opening from clogging which will
inhibit testing. Excessive hairs may have to be cut.
Also check for a perforated eardrum as this may give a tymp curve
which may be mistaken for a fluid filled middle ear.
Picture 1:
Pull
Push
The other probe is then pushed into the socket until a distinct
"click" is heard. Note! The squared hole in the probe connection
should face in the same direction as the push-button.
D) Insert the probe tip into the ear canal until you have an
airtight fit. Pull the transducer house upwards again, until the
probe in the ear is hanging by the tubes from the transducer.
Pause
The pause function operated from “Pause” (9) has been built
into AT235h to prevent the instrument from automatically
starting the test as soon as an airtight fit has been obtained. The
pause function is mainly used when using the clinical headset or
shoulder strap on the patient.
Figure 5:
Hair coming out of the ear canal may make an airtight fit difficult
to obtain. Therefore, ideally, excessive should be removed prior
to testing. Also, a clinical ear tip covered with Vaseline may be
helpful. Make sure the ear tip does not have its opening closed
by the wall of the ear canal or clogged by Vaseline or cerumen.
Test Procedure
Before every test is performed it is recommended to delete all stored
patient data from the instrument. This is done by holding down “shift”
(17) while pressing “new subject” (9). After this action the following
message will appear on the screen:
Figure 7:
Press F1 “New” to delete all stored data from the previous patient and
use the keyboard to type in new patient data.
Printout 1: Printout 2:
Printout 1:
Printout 1 contains the following:
• Tympanogram.
• Reflexes (sequence method with individual reflexes starting at 80
dB at 4 different frequencies).
• ETF 1 test (Eustachian Tube Function test).
Printout 2:
Printout 2 contains the following:
• Tympanogram.
Printout 3:
Printout 3 contains the following:
• Tympanogram.
• Reflexes (sequence method with 5 individual reflexes starting at
80 dB at 4 different frequencies).
• Audiogram with stored threshold for right and left ear.
Printout 4:
Printout 4 contains the following:
• ETF 1 test (Eustachian Tube Function test).
or
2) Tympanometry mode
Figure 8:
The probe will have an airtight fit when the indication light on the probe
is green. When the status “Leaking” or “Blocked” is displayed in the
upper right hand corner of the screen under the Status Bar, the
indication light on the probe will be yellow. This indicates a bad fitting
or a blocked ear tip.
On the screen the tymp test can be followed on-line. When the tymp
curve has been drawn, the AT235h will automatically continue with the
selected reflex test “Test A” or “Test B”.
4. Remove the probe from the first ear when the test sequence
has finished. Select the other ear “Right” (20) or “Left” (21)
and move the probe to the second ear and repeat the test
sequence once again.
5. To print out the test result, simply press “Print” (8).
Figure 9:
Figure 10:
Figure 11:
Figure 12 shows the Modify Menu for Reflex Test A – it is similar for
Reflex Test B:
Figure 12:
When the individual reflex for Test A or B has been modified to the
special personal needs, hold down “shift” (10) while pressing the F-key
underneath the highlighted reflex once again.
In the same way all the reflexes for Test A or B can be modified.
Finally, press “Back” (7) to return to the test screen.
The idea is that light indication in the “Ipsi” (18) and “Contra” (19) keys
indicates whether or not the ipsilateral and / or contralateral reflexes
will be performed in a preceding test session.
Tympanometry - only
When the instrument is switched on it automatically powers-up in
2) Tympanometry mode
Figure 13:
The probe will have an airtight fit when the indication light on the probe
is green. When the status “Leaking” or “Blocked” is displayed in the
upper right hand side of the screen the indication light on the probe will
be yellow. This indicates a bad fitting or a blocked ear tip.
Now place the clinical headset over the patient's head or use
the shoulder strap and insert the probe tip into the ear canal.
When an airtight fit has been obtained the tympanometric
test is ready to be performed. To start the test, select
"Pause" (9) once again and the test will be carried out
automatically.
3. Remove the probe from the first ear when the test sequence
has finished. Select the other ear “Right” (20) or “Left” (21)
and move the probe to the second ear and repeat the test
sequence once again.
Extend Function
When being in the Tympanometry mode it is possible to extend
the pressure range to go from +300 daPa to –600 daPa by
pressing F6. When the Extend Function is active “Extend” in
lower right hand corner of the screen is highlighted. The
pressure range is indicated by two small, bold, vertical lines on
the horizontal 0 ml line. To go back to the standard pressure
range press F6 once again.
Figure 14:
The Tympanometry mode for High Probe Tone looks very similar to
the normal Tympanometry mode; however, the following differences
can be seen on the screen:
Figure 16:
In the Reflex Test mode the right indication light only on the
“Tymp/Reflex” key (10) will be active.
Place the clinical headset over the patient's head or use the
shoulder strap and insert the probe tip into the ear canal.
When an airtight fit has been obtained the reflex test is ready
to be performed. To start the test, select "Pause" (9) once
again and the test will be carried out automatically.
3. Remove the probe from the first ear when the test sequence
has finished. Select the other ear “Right” (20) or “Left” (21),
move the probe to the second ear and repeat the test
sequence once again.
Manual Reflexes
With AT235h there are two different ways to perform manual reflexes.
Manual reflexes can be performed subsequently to an automatic test
session or on their own.
When ready, press the “Tone Switch” (22) and the manual reflex will
be performed and appear on the display. The parameters for the
manual reflex can be changed once again if the result is not
satisfactory. When satisfied, press “Store” (16) to store the manual
reflex next to the automatic reflexes in the lower right hand corner of
the screen and AT235h will automatically prepare for performing
another manual reflex.
Figure 18:
Just above the F1 and F2 keys there are two small arrows pointing to
the left and the right hand respectively. By pressing the F1 and F2
keys it is possible to browse between the recorded reflexes.
Figure 19:
Figure 20:
Note: To ensure the right basis for performing a reflex decay test it is
necessary to find the reflex threshold for the frequency to be tested.
When ready, press the “Tone Switch” (22) and the manual reflex
decay will be performed and appear on the screen. The parameters for
the manual reflex can be changed by pressing the “Frequency Decr /
Incr” keys (14 and 15) and the “Intensity Decr / Incr” keys (23 and 24).
Press “Store” (16) to store the manual reflex decay. AT235h
automatically prepares for performing another manual reflex decay.
Figure 21:
When a reflex decay test has been accepted and stored it will be put
into one of the small reflex boxes. A small “D” (see figure 22) will
indicate that a Decay test has been performed.
Figure 22:
To choose the Eustachian Tube Function test press the “ETF” key (11)
Figure 23:
In the Eustachian Tube Function test mode the indication light on the
“ETF” key (11) will be active and at the same time the indication light
on the “Pause” key (9) will start blinking in order to tell the operator
that his/hers action is now needed.
The probe will have an airtight fit, when the indication light on the
probe is green. When the status “Leaking” or “Blocked” is displayed in
To perform the test the operator will have to press the blinking “Pause”
key (9) to make a tympanometric curve.
After the first sweep an information window appears on the screen:
Now press the blinking “Pause” (9) key once again and the test will
make its second sweep. When the second sweep has been carried
out another information window will appear on the screen:
Press the blinking “Pause” (9) key once again as told in the
information window and the test will make its final and third sweep.
or
Figure 24:
When the audiogram for the first test ear has been completed
the test will automatically continue testing the other ear.
In the horizontal bar above the right and the left audiogram it is
possible to follow the test procedure see figure 23. To the left
there is a small circle, which is highlighted whenever a tone is
audible to the patient. Next to the circle there is a small
rectangle, which is highlighted whenever the patient presses
the patient response button.
Figure 25:
Manual Audiometry:
To perform manual audiometry select “Audiometry” (12) and
press “Man” (13). Now, manual audiometry can be performed.
Select frequency by pressing “Frequency Decr / Incr” (14 and
15) and intensity by pressing “Intensity Decr / Incr” (23 and
24). To store thresholds for the audiogram press “Store” (16).
Figure 26 shows the Reflex and Tympanometry Test with the Child
Function activated:
Figure 26:
The train will continue moving until the selected reflex test session
has finished. When the test session has finished the train will
automatically disappear.
The two different probe systems can be seen in the below picture 11:
Picture 11:
Picture 12:
Picture 13:
The cleaning of the acoustic and air pressure channels of the probe
tip must be performed by means of the cleaning wire (nylon wire)
which can be found in the Ear tips Assortment provided with the
AT235.
Picture 14:
After cleaning all the acoustic and air pressure channels of the probe
tip it can be reassembled. Make sure that the Probe Tip is connected
correctly onto the Transducer Housing – a small flange will ensure
correct positioning - before the plastic nut is gently tightened.
Picture 15:
Medical CE-mark
The CE-mark indicates that Interacoustics A/S meets the
requirements of Annex II of the Medical Device Directive
93/42/EEC. TÜV Product Service, Identification No. 0123,
has approved the quality system.
Power, UPS400
Consumption : 15VA
Mains voltage/fuses: 100-240V AC, 50-60Hz, 40VA
Operation Environment:
Temperature : 15 – 35 °C / 65 – 95 °F
Rel. Humidity : 30 – 90%
Printer (Optional)
Type : Thermal printer with recording paper
in rolls.
Printing Time : Depending on the test printed.
Paper Rolls:
Width : 112 (+/- 0.5) mm
Diameter : 45 (+2.0) mm
Basis weight : 58.5 +/- 3g/m2
Thickness : 62 +/- 3 u
Tensile : Min. 3.2 kN/m
Compliance
Range : 0.1 to 6.0 ml (Ear volume: 0.1 to 8.0
ml).
Types
Tympanometry : Automatic, where the start and stop
pressure can be user-programmed
from the setup menu.
Eustachian Tube Function: Williams test (automatic
function).
Indicators
Graphical display : Compliance is indicated as ml and
pressure as daPa. Stimulus level is
indicated as dB Hearing Level.
Memory
Tympanometry : 2 curves per ear (one of them from
the High Probe Tone.
Eustachian Tube Function: 3 curves per ear.
Inputs
Patient Response : Connection for patient response
switch.
Output
Contra Earphone : TDH39, earphone for reflex and
audiometry measurements.
Ipsi Earphone : Probe earphone incorporated in
probe system for reflex
measurements.
Air : Connection for air system to the
probe.
RS232 : Input/output for PC connection. An
external PC can be setup to both
monitor and control the instrument.
The control actions can be followed
on the display and the operation
panel. Online communication can be
selected, where the measured data
will be sent to an external PC.
Keyboard : Connection for external keyboard,
standard PC type.
Attenuator
Range : 0 to 130 dB in 1 or 5 dB steps.
Typical range is –10 to 120 dB HL.
Test Types
Manual Audiometry: Manual control of all functions.
Automatic Audiometry: Auto threshold according to ISO
8253-1 (Patient controlled Hughson
Westlake). The threshold is
determined by the activation of the
patient response.
Manual Reflex : Manual control of all functions.
Reflex Decay : Manually controlled with stimulus
duration of 10 seconds.
Memory
6 ipsilateral and 6 contralateral graphs / curves. Each of them
can hold up to 6 pulses. There is an additional capacity for 6
manual tests.
Additional Parts:
TDH39 audiometric headsetACC226 Carrying Case
EARtone 5A Audiometric Insert Phones
ACC25 carrying case
CIR22 contralateral insert phone
APS2 patient signal
CAT40 calibration unit 0.2-0.5-2.0-5.0 ml
CAT227 calibration unit 0.2-0.5-1-2-3-4-5 ml
IES impedance ear simulator
Connection to PC
The data from AT235h can be transfer to a PC by means of an IFC59
cable (25 pin computer COM-port) or IFC69 (9 pin computer COM-
Port).
The Baud Rate (data transfer speed) of AT235h must selected in the
Common Setup Menu to match that of the used software.
Pull
Push
The other probe is then pushed into the socket until a distinct "click"
is heard. Note! The squared hole in the probe connection should
face in the same direction as the push-button.
2. Place the new paper roll in the paper compartment in such a way
that the paper ascends from the lower part of the paper roll. With
a pair of scissors cut two triangles of paper away. This will make
the feeding of the paper easier – see picture 7.
Picture 7: Picture 8:
4. Gently insert the paper between the lower part of the black
rubber platen and the black plastic – see picture 8.
5. When the paper appears between the upper part of the rubber
platen and the paper cutter, pull out an extra 10 to 15 centimetres
of paper and return the small blue lever to its original position.
Make sure that the paper is aligned correctly – see picture 9 and
10.
6. Now guide the paper through the slot of the cover of the paper
compartment.
Contents of Shipment
When AT235h is delivered as a standard unit the case contains the
following:
1 Instrument AT235
1 Universal Probe System ATP-AT235U
1 Single Contralateral Earphone TDH39
1 External Switch Mode Power Supply UPS400
1 Power Cable
1 Box of 65 Assorted Ear tips BET50
3 Rolls of Recording Paper TRP35
1 Dust Cover PCR-AT235
1 Operation manual
1 CE-manual
Reporting Imperfections
Inspect before connection:
Prior to connecting AT235h to the Power Supply it should once
more be inspected for damage. All of the cabinet and the
accessories should be checked visually for scratches and
missing parts.
Annual calibration:
The AT235h has been designed to provide many years of
reliable service, but annual calibration is recommended due to
possible impact on transducers.
The mains switch must be "on" and the power switch on the
rear panel of the AT235h must also be "on".
The baud rate set on AT235h and the computer program must
always be the same.
Normally, only the indication light of the “Ipsi” key (18) will be
active informing that only the ipsilateral reflex test will be
performed during the subsequent test procedure. Press the
“Contra” key (19) to activate also the contralateral reflex test
for the subsequent.
What is Gradient?
Gradient is explained in the chapter “Tympanometry Setup
Menu” of this manual.
Biswas, Anirban:
Clinical Audiovestibulometry, (Bhalani Medical Book House,
Bombay, India 1995)
Brask, T.:
Extratympanic Manometry in Man. (Scandinavian Audiology,
supp. 7. 1978)
Fiellau-Nikolajsen, Mogens:
Tympanometry and Secretary Otitis Media. (Acta Oto-L. 1983)
Jerger, J.:
Clinical Experience with Impedance Audiometry. (1970)
Katz:
Handbook of Clinical Audiology, Fourth Edition 1994 (Williams
& Wilkins 1985)
Liden, G. et al.:
International Symposium on Impedance Audiometry and
Pediatric Audiology Göteborg 1982. (Scandinavian Audiology
supp. 17. 1983)
Liden, G.:
Audiology (Almqvist & Wiksell. 1985) (Swedish language)
Popelka, G. R. et al.:
Hearing Assessment with the Acoustic Reflex. (Grune &
Stratton 1981)
Acoustic Compliance:
Another term for Acoustic Admittance.
Acoustic Immitance:
Refers collectively to acoustic impedance and / or acoustic
admittance.
Compliance:
1) Ease with which air moves (e.g. influenced by the
eardrum and middle ear mechanism).
2) Often used to indicate the equivalent volume of air in
the middle ear.
Ear Tip: A cuff which is used to seal the probe into the external
auditory canal.
Memo:
memo is a unit for measuring acoustical impedance:
Myringoplasty:
Surgical repair of the eardrum membrane.
Myringotomy:
(tympanotomy) A small incision made in the eardrum
membrane to remove fluid from the middle ear.
Pascal (Pa):
2
A unit of pressure or stress, equal to one Newton per m .
Probe:
A coupling device that is inserted into the external auditory
canal, to connect it to the acoustic immittance meter.
Probe Signal:
An acoustic signal that is emitted into the external auditory
canal by means of a probe. The signal is used to measure
acoustic immittance.
Probe Tip:
The upper part of the probe tip on which the ear tip, a cuff
which is used to seal the probe into the external auditory canal,
is placed.
Stimulus Ear:
The ear to which the reflex activating stimulus is presented in
order to elicit a middle ear muscle reflex. Note: If a bone
vibrator or a loudspeaker is used to deliver an acoustic reflex it
may not be possible to define the stimulus ear.
Toynbee Test:
Test designed to determine the function of the Eustachian tube
in ears with perforated eardrums.
Toynbee's Manoeuvre:
See Valsalvation.
Tympanogram:
A chart of the results of tympanometry - compliance
measurements at the eardrum.
Valsalvation:
Swallowing with the mouth and nose closed to draw air out of
the middle ear. Syn.: Toynbee's manoeuvre.
Valsalva's Manoeuvre:
Blowing forcibly to open Eustachian tube by holding nose and
closing mouth. Named for its originator, Antonio Valsalva.
Sometimes called Valsalva's experiment.
Williams Test:
Test designed to determine the function of the Eustachian tube
in ears with non-perforated eardrums.
Figure 27:
From the main menu it is possible to adjust the contrast of the display
by means of “LCD-“ (F5) and “LCD+” (F6). The instrument will always
remember the last setting.
When holding down “shift” (17) the writing above F5 and F6 changes
to “LED-“ and “(LED+)”. Now the brightness of the LEDs (indication
lights) of the instrument keys can be adjusted. Also here the
instrument will remember the last setting.
Pressing “Test” (F1) will bring you back to the Test Screens. Pressing
“Setup” (F2) will bring you to the Main Setup Menu.
Figure 28:
To leave the individual setup menus press “Back” (7). This key will
always take you back one level. If one or more parameters have
been changed the message on the next page appears on the screen.
Figure 29:
Start Pressure:
The pressure established as the starting point for the pressure
sweep for the tympanometric curve. The Start Pressure runs
from 25 daPa to 300 daPa.
Stop Pressure:
The pressure where the sweep for the tympanometric curve
stops.
Pump Speed:
There are four different pump speed settings:
• Minimum (17 daPa/sec).
• Medium (50 daPa/sec).
• Maximum (>150 daPa/sec).
A slow speed will be more time consuming, but may give more
detailed information. The horizontal displacement of the
tympanometric curve’s peak in the sweeping direction caused
by inherent hysteresis of the system and the middle ear itself
will be smaller with lower speed.
Compensated Mode:
AT235h can be set to either Compensated Mode or Non
Compensated Mode. Compensated Mode is used when you
want the tympanometric curve to be displayed at the floor of
the co-ordinate system of the tympanogram. See figure 30.
Figure 32:
Figure 33:
Figure 34:
Output:
It is possible to select between ipsilateral and contralateral
reflexes as output.
Stimuli Frequency:
For ipsilateral reflexes the stimuli frequency options are the
following:
500, 1000, 2000, 3000 and 4000 Hz. Including the noise
stimuli WN, LP and HP.
Level:
The intensity level for the reflexes goes from 60 dB to
maximum. Maximum depends on the selected frequency and
the selected transducer. In the final line of the table it is
possible to switch on or off the reflexes according to the
number of reflexes needed.
Reflex Methods
There are four different reflex methods to choose from in the Setup
Menu for Reflex Test A and B. In the following these reflex methods
will be described in details:
Figure 35:
Reflex Sensitivity:
In order to fully understand the reflex search procedure of the
following three Reflex Methods it is important to know about
the Reflex Sensitivity idea. Reflex Sensitivity appears under
the Reflex Method line when Screening, Auto or Sequence
Method are selected.
Figure 36: Sensitive (0.03 ml) Figure 37: Normal (0.05 ml)
Figure 38: Robust (0.08 ml) Figure 39: Robust (0.08 ml)
Note: The pass box is not shown in the screen of AT235h but
serves as an example of what sensitivity is and how it is used.
Figure 40:
If the reflex goes through the pass box it is not accepted. The
intensity level automatically increases with 10 dB for another
reflex. This procedure continues until the reflex is above the
selected pass box or the maximum intensity level is reached.
Auto (5 dB steps):
Figure 41 shows the screen for the Auto (5 dB steps) Reflex
Method. The Auto test is used for testing where only one reflex
is needed per frequency, but a reflex of a certain size is
preferred (see “Reflex sensitivity” above). The Auto test then
searches for suitable stimulus intensity before each reflex is
tested.
Figure 41:
5) If the reflex found the second time also does not meet the
pass criteria, a new reflex is recorded at 5dB higher
intensity. Consecutively higher intensities are then used
until a reflex is accepted. (If a pass cannot be obtained,
the reflex recorded at maximum intensity will be displayed
as the test result for the test).
Sequence:
Figure 42 shows the screen for the Sequence Reflex Method:
Figure 42:
Figure 43:
Level Increase:
Selects how many dB the intensity increases between each
single reflex. It is possible to choose between 5, 6, 8, 9, or 10.
Power-up:
Select between “Tymp”, “Tymp and Reflex”, or “High Probe
Tone” as power up. See figure 47:
Figure 47:
Baud rate:
Selects the Baud rate level for the serial communication
RS232. Select between “57600”, “38400” and “38400 MPT10”
Remote Switch:
Selects the function of the Remote Switch of the probe. Select
between “Off”, “L/R”, “Pause” and “L/R or Pause”.
Printer:
There are three different “Printer” settings; “Off”, “Internal” and
“HP PCL L3”:
To enter your own Clinic data press F1 “Edit”. The Clinic data
is edited by pressing F2.
Figure 51:
Keyboard Connected:
It is possible to connect a keyboard to AT235.
− Before the connection to the mains network, be sure that the local
mains voltage corresponds to the voltage labeled on the
instrument. Always disconnect the power cord if the instrument is
opened or by control / replacement of the mains fuses.
− For maximum electrical safety, turn off the power from a mains
powered instrument when it is left unused.
− Do not site the instrument next to a heat source of any kind, and
allow sufficient space around the instrument to ensure proper
ventilation.
Opr. date: 2003-02-24 by: EC Rev. date: 2003-02-24 by: EC Rev. no.: 1
Address
Company: Drejervænget 8
DK-5610 Assens
Address: Denmark
Phone
(+45) 63713555
Phone:
Fax
(+45) 63713522
Fax:
Date :
Contact person:
To ensure instant and effective treatment of returned goods, it is important that this
form is filled in and placed together with the item.
Please note that the goods must be carefully packed, preferably in original packing, in
order to avoid damage during transport. (Packing material may be ordered from
Interacoustics.
1
EC Medical Device Directive rules require immediate report to be sent, if the device by malfunction
deterioration of performance or characteristics and/or by inadequacy in labelling or instructions for use, has
caused or could have caused death or serious deterioration of health to patient or user.
Page 99
Page 100
Drawing of Front Plate
Page 101