Hygiena Test Device Yeast Detection Purpose: ATP Sanitation Monitoring
Hygiena Test Device Yeast Detection Purpose: ATP Sanitation Monitoring
Technical Document
Methods
All yeast were grown in PDB (Potato Dextrose Broth) at 37 °C and 30 °C for 72 hours and then sub-cultured onto PDA
(Potato Dextrose agar) to check purity. The yeasts were grown again in PDB before being diluted into MRD (Maximum
Recovery Diluent). There was no need to wash the yeast in saline prior to measurement as it is more practical to
measure the background ATP (non-yeast ATP) rather than try to remove it and stress the yeast in the process. Yeasts
in the field would never be washed of background (free) ATP, so this method is most practical. Background ATP (non-
yeast ATP) was detected using a proprietary internal quality control test device which detects non-microbial ATP.
From each dilution 10µL was added directly to the swab bud or 100µL added to the dipper. Each device was activated
and read in the EnSURE Monitoring System. Five replicates were run from each dilution. Reference plates were also
done using the method of Miles and Mizra by inoculating each dilution as 10µL spots onto PDA to exactly estimate the
count per 10µL and also to estimate from the M&M method the CV% of the inoculation method independently of the
RLU results.
REVA 032015
www.hygiena.com
ATP Sanitation Monitoring
Technical Document
Results
Table 1: Average Saccharomyces cerevisiae detection in Hygiena test devices (RLU) (n=5)
SuperSnap UltraSnap AquaSnap Non-yeast ATP
Neat 9235 7969 9888 245
-1 7966 5312 9790 42
-2 2486 765 7648 8
-3 183 77 1116 1
-4 28 23 246 0
-5 0 0 12 0
Neat CV% 1 9 1 17
-1 CV% 3 8 2 11
-2 CV% 9 6 4 11
-3 CV% 9 10 13 149
-4 CV% 8 21 7 NA
-5 CV% NA NA 18 NA
REVA 032015
www.hygiena.com
ATP Sanitation Monitoring
Technical Document
Table 4: Average Candida albicans detection in Hygiena test devices (RLU) (n=5)
SuperSnap UltraSnap AquaSnap Non-yeast ATP
Neat 9083 8068 9891 257
-1 7664 6864 9374 74
-2 2838 996 8265 8
-3 290 95 1146 0
-4 91 30 133 0
-5 1 5 9 0
Neat CV% 2 3 1 3
-1 CV% 2 2 4 20
-2 CV% 10 4 3 10
-3 CV% 8 5 13 NA
-4 CV% 6 11 15 NA
-5 CV% 37 16 8 NA
REVA 032015
www.hygiena.com
ATP Sanitation Monitoring
Technical Document
Table 7: Average Zygosaccharomyces bailii detection in Hygiena test devices (RLU) (n=5)
SuperSnap UltraSnap AquaSnap Non-yeast ATP
Neat 6928 7892 9917 659
-1 1718 2993 9400 174
-2 243 263 4023 19
-3 34 27 480 1
-4 4 5 47 0
-5 0 0 5 0
Neat CV% 5 2 1 7
-1 CV% 7 4 3 10
-2 CV% 8 11 5 14
-3 CV% 7 19 14 91
-4 CV% 15 32 10 NA
-5 CV% NA NA 9 NA
REVA 032015
www.hygiena.com
ATP Sanitation Monitoring
Technical Document
Table 10: Average Candida parasilosis detection in Hygiena test devices (RLU) (n=5)
SuperSnap UltraSnap AquaSnap Non-yeast ATP
Neat 8874 8351 9999 129
-1 8122 4875 9999 25
-2 3124 478 8114 7
-3 284 58 897 1
-4 33 12 142 0
-5 0 0 8 0
Neat CV% 2 11 2 11
-1 CV% 4 5 5 9
-2 CV% 4 6 5 14
-3 CV% 8 8 12 121
-4 CV% 9 15 9 NA
-5 CV% NA NA 14 NA
REVA 032015
www.hygiena.com
ATP Sanitation Monitoring
Technical Document
Table 13: Average Rhodotorula rubra detection in Hygiena test devices (RLU) (n=5)
SuperSnap UltraSnap AquaSnap Non-yeast ATP
Neat 8712 7894 8989 88
-1 5841 5111 8478 11
-2 2811 1421 7133 4
-3 198 88 1011 0
-4 21 4 114 0
-5 0 0 5 0
Neat CV% 2 1 1 1
-1 CV% 4 4 4 8
-2 CV% 4 4 6 10
-3 CV% 8 5 10 88
-4 CV% 11 12 19 NA
-5 CV% NA NA 14 NA
Conclusion
In general, SuperSnap and UltraSnap Surfaces ATP test devices detect approximately <50 CFU/swab yeasts.
AquaSnap Total Water ATP test devices detect approximately <10 CFU/mL yeasts. Variation of results at these low
levels was minimal with CV% ranges as follows: SuperSnap (6-11%), UltraSnap (12-21%), and AquaSnap (8-18%).
REVA 032015
www.hygiena.com