Manual For HIFU PDF
Manual For HIFU PDF
1.Introduction
1.1Purpose
This Instructions for Use manual provides a description of System components, its controls and
displays, instructions for its operation, and other equipment information important to be user.
1.2 Conventions
Note: Notes designate information of special interest.
Caution: Cautions alert the user to precautionary steps necessary to properly operate the system.
Failure to observe these cautions may void the warranty. All procedures are broken down by
numbered steps. Steps must be completed in the sequence they are presented. Bulleted lists
indicate general information about a particular function or procedure. They do not imply a
sequential procedure. Control names are spelled as they are on the system, and they appear in
Bold text.
2. Parameter
Machine size 40*38*25cm
Frequency 50Hz/60Hz
Power output 0.2J-2.0J
Handpiece Standard: Face lifting(1.5mm,3.0mm,4.5mm);
Optional: Body slimming(8.0mm,13.0mm);
Optional: Vagina tightening(3.0mm,4.5mm)
Voltage AC110V-240V
Machine weight about 10kg
3. Accessories picture
4. Instrument Description
4.1 Host Control
Vaginal Handle
HIFU Handle
Vents
4.3 Cartridges
Cartridges Description
Item description
① Top marker Cartridges Transducers
② Treatment instructions Maximum processing line length and Center mark
5. Introduction to Operation
5.1 Click on the interface to enter the next interface(Choose treatment handles)
Remaining
Cartridges shots/Total
identification
Spacing adjustment
Length Power
Return to main
interface
Spacing
adjustment
Length Power
Return to Main
interface
6. Treatment guidelines
Note: Operation Please follow the instruction
Note:In order to avoid broke the cartridges , please clean only with the above cleaning method,
do not use acetone or other solvent.
Note:Please check the date of the cartridge before use, do not use the expired probe.
8.Taboo crowd
1. Pregnant
2. Heart disease, diabetes, hyperthyroidism patients
3. Patients with advanced cancer
4. Surgical wounds are not healed
5. Injection of botulinum toxin within one year
6. Have unrealistic effect on the patients
9.Safety matters
9.1 Handling safe
The operating system memory precision instrument,Users do not try to move the instrument at a
long distance. The original shock-absorbing package must be used during transportation, or it
may lead to system disorders, the energy output will be affected.
1.Instrument overview
The Vagina Tightening HIFU System uses an noninvasive ultrasonic focusing technique to focus
on the mucous membrane fibrous layer and muscle layer directly. Using ultrasonic waves as the
energy source and taking advantage of its penetration and focus, the system will send out
ultrasonic energy focusing in the the lamina and muscle fiber layer in a predetermined depth. A
higher intensity of ultrasonic region, called focus region, is formed. In 0.1 second, the temperature
of the region can reach to above 65 ℃ , so the collagen is reorganized and the normal tissue
outside the focal region is undamaged. Therefore, the desired depth layer can obtain the ideal
effect of collagen concentration, reorganization and regeneration. Ultimately, the mysterious effect
of vagina tightening is achieved.
2.Parameter
Frequency 50-60Hz
Maximum power 200w
Treatment cartridges 1.5mm, 3.0mm 4.5mm
Voltage 110v-220v
Packing size 69*64*34cm
Vaginal tightening
HIFU handle handle
Vaginal tightening
interface
Vents
Power cord/Switch
Vaginal Handle
connector Airbag detection
connector HIFU Handle
Connector
4.Interface description
Power
Total length Rotation angle Shot length spacing
Number of
rows played
5.Pressure interface
1.Tighten the switch of the pressure before test, Extrusion air bag to increase the value, and then
adjust the switch of the pressure until the value on screen around 20.
2.Put the test head into vaginal ( If needed use some Lubricants). 3.When
test, separate the lags with the distance less than shoulder width.
4.Relax 6 seconds , Natural breathing, press save ,this value will be the before treatment value.
5.After treatment , test again from the step 1 , you will get another value after put the test head
into vaginal, this value will be the after treatment value.
Pressure value Condition of vaginal Sensitivity
tightening
Under 60 Very bad No
60-80 Relatively bad Low
80-120 Normal general
120-160 Good Squirting
Above 160 Very good nymph
6.Treatment heads
The machine equipped with two different depth of the treatment head.
功能头 图片 治疗皮肤的深度及功效
3.0mm of functional head The power reaches the
mucosa tissue directly to
speed the cell activity, making
the collagen newborn and
reorganized and increasing the
mucosa elasticity and contract
the vagina muscle.
4.5mm of functional head The power reaches the SMAS
layer to make heat coagulation
of SMAS, performing the
pulling and SMAS suspension
well, having improvement
from deep to shallow so as to
support the muscle structure,
better helping muscle layer
recover elasticity; and it is
more compact.
Pressure test handle Using the principle of balloon
pressure measurement, test
the vaginal tightness.
7.Operation introduction
1. Before using the apparatus, disinfect and wipe the probe by the detergent; and clean the
water drop on the face;
2. Install the handle in the rotary bracket; and connect the connection wire to the host;
3. Electrify the host, press the power switch to start the apparatus;
4. Select suitable probe and access to the handle; and the display screen can display the
corresponding parameter of the probe;
5. Choose “ROTATING”
6.If needed, enter the next step and select “F.AUTO” function;
7. Coat the ultrasonic wave conductor to the probe energy emitting area (black area);
8. Sleeve the protection coat on the probe; and coat a layer of lubrication and function product
on the outer surface of the protection coat;
9. Click “START” button to enter the working state;
10. Stretch the probe in the end of the vagina; and “0” scale is at the top;
11. Press the button on the handle to send out one wire;
12. Repeat the above step until to finish the step; and reset the handle;
13.After completing the required steps, turn off the power,Remove the protective cover.Clean
the probe and handle with detergent.
10. Troubleshooting
1. No display on the on the apparatus interface:Please check if the power plug is inserted well
or the power switch is turned on.
2. The apparatus operates normally but the press key malfunctions: Please check if the power
supply is connected well.
3. When the probe is not flatted with the face, the apparatus shall be automatically closed; and it
can be operated as long as restarting the apparatus.