Technical Data: SUPERIX RL180N USER MANUAL - UM - 18005 - 0698 MECALL SRL 1998 All Rights Reserved

The document provides technical specifications for a medical imaging device. It details power requirements, weights and dimensions of components, performance characteristics of the table, compression cone, spot film device, x-ray tube column, collimator, and tomography capabilities. Technical displays and compliance with regulations are also summarized.

Uploaded by

Raklo Yukhnin
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
181 views

Technical Data: SUPERIX RL180N USER MANUAL - UM - 18005 - 0698 MECALL SRL 1998 All Rights Reserved

The document provides technical specifications for a medical imaging device. It details power requirements, weights and dimensions of components, performance characteristics of the table, compression cone, spot film device, x-ray tube column, collimator, and tomography capabilities. Technical displays and compliance with regulations are also summarized.

Uploaded by

Raklo Yukhnin
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 14

2.

TECHNICAL DATA

2.1 POWER

VOLTAGE SUPPLY: 220V-440V (+/-10%) 3 ∼


(set-up at installation;
see installation manual)

FREQUENCY: 50 - 60 Hz

MAX. ABSORPTION POWER: 4.0 kVA

2.2 WEIGHT AND DIMENSIONS

The unit is supplied according to the packaging list (see Installation Manual)

TABLE WEIGHT. 1200 kg.


(Without X-ray tube, without II)

ELECTRONIC CABINET 170 kg.


(Equipped only with
Electronic control table)

CONSOLLE WEIGHT 50 kg.

COLLIMATOR WEIGHT 14 kg.

The unit dimensions are shown in the figure of the following page:

SUPERIX RL180N USER MANUAL - UM -18005 - 0698 MECALL srl 1998 All rights reserved
SUPERIX RL180N USER MANUAL - UM -18005 - 0698 MECALL srl 1998 All rights reserved
SUPERIX RL180N USER MANUAL - UM -18005 - 0698 MECALL srl 1998 All rights reserved
SUPERIX RL180N USER MANUAL - UM -18005 - 0698 MECALL srl 1998 All rights reserved
SUPERIX RL180N USER MANUAL - UM -18005 - 0698 MECALL srl 1998 All rights reserved
SUPERIX RL180N USER MANUAL - UM -18005 - 0698 MECALL srl 1998 All rights reserved
PERFORMANCE CHARACTERISTICS

2.3.1 TABLE-TOP

DIMENSIONS 240 cm. x 72 cm.

RADIOTRANSPARENT AREA 228 cm. x 51 cm.

MAX.PATIENT’S WEIGHT (I.E.C. 601-1) 150 kg.

X-RAY-ABSORPTION - CARBON FIBER TABLE TOP 0.9 mm. Al.

TABLETOP - FILM DISTANCE 76 mm.(+/- 1mm)


(Without load on the table)

TABLE TOP - I.I. DISTANCE


(Without load on the table) 111 mm.(+/- 1mm)

TRANSVERSAL TRAVEL +/-150mm.(+/-2.5mm)


(From centered position)

MINIMUM TABLE-TOP HEIGHT ABOVE THE FLOOR


(Table in horizontal position; depends on Image System
and on base plate) 850 mm

MAX. ANGLE OF TRENDELENBURG -90° (-0°/+1°)

MAX. ANGLE OF TILTING +90° (+0°/-1°)

TILTING SPEED 0 - 4,5 °/sec

ACCESSORIES footrest,hangrips,
shoulder-rest, compression band

ELEVATION TRAVEL > 380 mm

2.3.2 COMPRESSION CONE

TRAVEL RANGE 30 cm.(± 1cm.)

ANGULATION OPERATIVE RANGE 0° - +20°


± 5° from starting angle

MAX. COMPRESSION FORCE 10.0 kg. (± 10%)

SUPERIX RL180N USER MANUAL - UM -18005 - 0698 MECALL srl 1998 All rights reserved
2.3.3 SPOT FILM DEVICE

Accepts all cassette formats (millimeter and inches) from 18X24 to 35x43 in both directions.

Vertical division (all formats) 1-2

Vertical division (film ≥ 24 format) 1-2-3

Vertical division (film ≥ 30 format) 1–2-3-4

Vertical division (40 and 43 formats) 1-2-3-4-5

Rapid sequence

Fluoroscopy to Radiography change over time 0.6 s.

Radiography and tomography are possible on the same film

Full SFD indications available on the control desk

Predisposition for AEC detector

Standard grid Smit r=12:1 Sid 120

Grid vibration speed 120 c/min. (± 10%)

Operative Modes radio/tomo, rapid sequence

Longitudinal travel (column centred) 1600 mm. (±5 mm.)

Shutters/film distance 34 mm. (± 1mm.)

Hazards protections optical detector and warning label

SUPERIX RL180N USER MANUAL - UM -18005 - 0698 MECALL srl 1998 All rights reserved
2.3.4 X-RAY TUBE COLUMN

Longitudinal travel (tube only) 1750 mm ± 10mm

Longitudinal travel with SFD 1600 mm ± 10mm

Max longitudinal speed Stepless from 0 to 150 mm/sec.

Oblique projection with


parallax correction ± 40 °

Focus / Film distance min 1050 mm

Focus / Film distance max 1500 mm

Sid movement speed 45 mm/sec.

Beam/floor min. distance (tilt=+90) 475 mm ± 10mm

Beam/floor max. distance (tilt=+90) 2075 mm ± 10mm

Beam/table edge min. Distance 400 mm ± 5mm

X-Ray tube rotation with 90° detents ± 180 °

SUPERIX RL180N USER MANUAL - UM -18005 - 0698 MECALL srl 1998 All rights reserved
2.3.5 COLLIMATOR

MAXIMUM KV 150 Kvp

SQUARE FIELD COVERAGE


AT FFD=100cm 43 X 35 (both direction)

OPERATIVE FFD 105 ÷ 150 cm

MINIMUM COLLIMATION AREA < 5 X 5 cm


(IEC601-1-3) (in image plane)

INHERENT FILTRATION (IEC601-1-3) > 1.7 mmAl/Eq

LEAKAGE RADIATION (IEC601-1-3) < 1 mGy/h with compatible tubes

MOTORIZED FILTERS 0.1mm Cu + 0.5 mm Al


0.2mm Cu + 0.5 mm Al

COLLIMATOR LAMP 24V AC / 150W

AVERAGE ILLUMINATION (IEC601-1-3) > 160 Lx FFD=100 cm

LIGHT FIELD EDGE CONTRAST RATIO > 4 FFD = 100 cm

LIGHT FIELD VS X-RAY FIELD ACCURACY < 1.8% FFD

X-RAY FIELD VS FILM SIZE ACCURACY < 2.8% FFD

SCALE VS X-RAY FIELD ACCURACY < 1.8% FFD

SID INDICATION ACCURACY < 1.8% FFD

SUPERIX RL180N USER MANUAL - UM -18005 - 0698 MECALL srl 1998 All rights reserved
2.3.6 TOMOGRAPHY

Parallel plane tomography

Focus / film distance 1050 mm

Angles 8° - 20° - 30° - 40°

Tomo Exposure times 8° < 0,5 s


20° < 1,5 s
30° < 2,0 s
40° < 2,5 s

Fulcrum adjustment range 5 - 280 mm

Adjustment step ± 1 mm

Fulcrum automatic increment/decrement sequences.

SUPERIX RL180N USER MANUAL - UM -18005 - 0698 MECALL srl 1998 All rights reserved
2.3.7 UNIT DISPLAYS

REMOTE CONSOLE

INCIDENCE ANGLE ± 1°

TILTING ANGLE ± 1°

TOMO LAYER ± 2 mm

TOMO ANGLE ± 1°

RX-TOMO TIME layer and angle function

FOCUS-FILM DISTANCE ± 1 cm.

CASSETTE SIZE ± 1 cm.

AVAILABLE EXPOSURES ON FILM numerical indication

NUMBER OF PROGRAMMED LAYER numerical indication

NUMBER OF ACTUAL LAYER numerical indication

COLLIMATOR

TILTING ANGLE ± 1°

LONGITUDINAL X-RAY BEAM SIZE


on the image receptor plane ± 1 cm.

TRANSVERSAL X-RAY BEAM SIZE


on the image receptor plane ± 1 cm.

DIAMETER OF THE CIRCULAR X-RAY BEAM


on the image receptor plane ± 1 cm.

FOCUS TO FILM DISTANCE ± 1 cm.

SUPERIX RL180N USER MANUAL - UM -18005 - 0698 MECALL srl 1998 All rights reserved
2.4 REGULATIONS COMPLIANCE

The SUPERIX RL180N is classified as:

MEDICAL EQUIPMENT CLASS IIb (Directive 93/42 CEE).

The SUPERIX RL180N has been designed to comply with the following regulations:

EN 60601-1 Amendment 1, 2
EN 60601-1-1 Amendment 1
EN 60601-1-2
EN 60601-1-3
EN 60601-1-4
EN 60601-2-28
EN 60601-2-32

THE SUPERIX RL180N IS CERTIFIED CE (0051).

SUPERIX RL180N USER MANUAL - UM -18005 - 0698 MECALL srl 1998 All rights reserved
2.5 COMPATIBILITY WITH OTHER ACCESSORIES

The SUPERIX RL180N has been designed and tested in order to fully comply with the regulations
specified in section 2.4 when connected with the following radiological subsystems:

X-RAY TUBES AND TUBE HOUSING

Varian → Tube G192 B130 Housing


Varian → Tube G192HS B130 Housing

Varian → Tube G292 B130 Housing


Varian → Tube G292HS B130 Housing

X-RAY GENERATORS

Mecall → CASTOR HF ( All Models )

IMAGE SYSTEMS

CCD IMAGE SYSTEMS ALL MODELS

EDS DIGITAL IMAGE SYSTEMS ALL MODELS

HIRIS DIGITAL IMAGE SYSTEMS ALL MODELS

WARNING:

In case the Table is used in combination with other units, whose compatibility is not declared in the
unit documentation, the operator is responsible to verify that no dangerous situations may occur for the
patients and the personnel.
For this scope it is good practice to contact the manufacturer of the unit or an expert.

WARNING:

THE X-RAY TUBE AND THE IMAGE INTENSIFIER SUPPORTS HAVE BEEN DESIGNED IN
ORDER TO WITHSTAND THE ACCESSORIES LISTED IN THIS SECTION.

DO NOT ADD OTHER COMPONENTS THAT INCREASE THE MECHANICAL STRESS


OF THE SUPPORTS

DO NOT STEP ON THE IMAGE SYSTEM

SUPERIX RL180N USER MANUAL - UM -18005 - 0698 MECALL srl 1998 All rights reserved