Test On CT-Analyzer-for OMICRON
Test On CT-Analyzer-for OMICRON
REQUIREMENTS
The requirements as specified in the following standards:
- IEC 60044-1 (1996) + amendment 1 (2000) + amendment 2 (2002)
- IEC 60044-6 (2002)
- ANSI C57.13 (1993)
TEST PROGRAMME
The programme was specified by the client and was as follows:
1 Verification checks of the analyzer test function for determination of winding resistance
2 Verification checks of the analyzer test function for the determination of current transformer
transformation errors
3 Verification checks of the analyzer test function for the determination of magnetization curves of
current transformers
4 Verification checks of the some additional analyzer test functions
©
Copyright: Publication or reproduction of the contents of this report in any other form than a complete copy to the letter, is not allowed without
our written consent.
Electronic copies in e.g. PDF-format or scanned version of this report may be available and have the status "for information only". The bound
version of this report is the only valid version.
Utrechtseweg 310, 6812 AR Arnhem. Telephone +31 26 3 56 31 85. Telefax +31 26 4 43 38 43
-2- 70570064-HVL 06-1071
MATERIAL DATA
CT Analyser
Manufacturer OMICRON, Klaus, Austria
Serial No. BB 174 U
Option not applicable
Input 110-240 V/50/60 Hz/6A
Output 1 x 130 V/5 A/350 VA
Sampling procedure by the Manufacturer
For the purpose of the tests the following current transformers were utilized:
CT OMICRON 1: KSO1811
- Manufacturer: Ritz
- CT 2000/5A and 2000/1A
- Class 0,02% at 2,5 VA, Class 0,1% at 15 VA
- Core Type: Non gapped, measurement CT
- Omicron Ser. No.: 05 – 305
- Ser. No.: 05/596717
CT OMICRON 2: KSO1812
- Manufacturer: Ritz
- CT 1000/1A
- Class 0,2% at 1 VA
- Core Type: Non gapped, measurement CT
- Omicron Ser. No.: 02 – 195
- Ser. No.: 05/560150
CT OMICRON 5: KSO1811
- Manufacturer: Ritz
- CT 2000/5A and 2000/1A
- Class 0,02% at 2,5 VA
- Core Type: Non gapped, measurement CT
- Ser. No.: 03/582927
CT KEMA 01:
- Manufacturer: Ritz
- NTRO 320, CT 5000/5 or CT 10000/5
- Class 0,01% at 10 VA
- Core Type: Non gapped, measurement CT
- KEMA HSL No 12.09 ORS 075392
- Ser. No.: 83/440308
CT KEMA 02:
- Manufacturer: Ritz
- NTTS
- Class 0,01% at 1 VA
- Core Type: Non gapped, measurement CT
- KEMA HSL No 12.07
- Ser. No.: 540 627
Purpose of the test was to verify whether the material complies with the specified requirements.
-4- 70570064-HVL 06-1071
SUMMARY
- General
The functional operation and applicability of the OMICRON CT ANALYZER for testing of current
transformers was verified by checks during determination of main characteristics of several current
transformers. The OMICRON CT ANALYZER was under operation and control by the manufacturer,
under supervision of KEMA. The results are valid only for the checks carried out together with the
selected testing modes and should not be treated as results of type tests, routine tests, acceptance
tests, or similar.
- Verification checks of the analyzer test function for determination of winding resistance
The functional operation during determination of winding resistance using the OMICRON CT
ANALYZER did not give rise to remarks.
The numerical evaluation, during measuring of KEMA shunt resistor resistance values, showed
differences of 0,1% or less, when compared with the calibration certificate values of the shunts.
- Verification checks of the analyzer test function for the determination of current transformer
transformation errors
The functional operation during determination of current transformation errors using the OMICRON CT
ANALYZER did not give rise to remarks. Tests were carried out in both selection modes, respectively
for measurement CT’s with non-gapped cores and for protection CT’s with gapped cores.
- Verification checks of the analyzer test function for the determination of magnetization
curves of current transformers
The functional operation during determination of current transformation magnetization curves using
the OMICRON CT ANALYZER did not give rise to remarks.
CONCLUSION
From the tests, verifications and measurements it has become obvious that the OMICRON CT-
ANALYZER does a very accurate current ratio measurement on continuous ring cores (non gapped)
(better than 0.02%/0.5min) and an adequate accurate current ratio measurement on gapped cores
(TPY better than 0,2%/3min; TPZ better than 1%/5min).
Furthermore, the OMICRON CT-ANALYZER is able to determine in a fast and accurate way the
magnetization curves and the related parameters.
-5- 70570064-HVL 06-1071
MEASUREMENT UNCERTAINTY
The last page of this report contains a table with measurement uncertainties. Unless otherwise
indicated in the report, the measurement uncertainties of the results presented are as indicated in this
table.
INTRODUCTION
The applicability and output results of the OMICRON CT ANALYZER with its functions for testing of
current transformers were compared with requirements and results of testing on the same current
transformers however using other test equipment, that is applied normally by KEMA for the tests to be
carried out on current transformers.
Based on the results of the comparison, conclusions may be drawn as to the applicability of the
OMICRON CT ANALYZER for the tests on current transformers.
Although the OMICRON CT ANALYZER has capabilities for many other testing functions, the tests
and checks, described in this report are related to the functions given in the test program only.
The testing activities (tests, checks, comparison and or verifications) cannot be considered as type
tests or routine tests or acceptance tests or calibration tests. The results may serve as an important
impression as to the applicability of several functions and results of the OMICRON CT ANALYZER for
testing of current transformers.
First in April 2005 and in October 2005 several checks were carried out in order to verify whether the
OMICRON CT ANALYZER was able to perform its function: determination of winding resistance.
For the purpose of this verification some resistance values were determined in two ways: using the
OMICRON CT ANALYZER and using a KEMA laboratory instrument.
-6- 70570064-HVL 06-1071
Initially this verification was performed in a functional way on several current transformers, during
resistance measurements of windings of some available current transformers. The results were
compared with results obtained using other instruments at KEMA. It was not the intention to carry out
accurate tests. The results of these checks are presented in appendix 1 page 1 of this report.
From the results it was concluded that the OMICRON CT ANALYZER did not cause any functional
problem during the checks.
Subsequently, in order to gather more information as to the accuracy of the measurements, this
verification was performed on two standard resistors, used as standard shunt in the KEMA laboratory.
Since the calibration results of these shunts are available, a better check of this function of the
OMICRON CT ANALYZER could be made. For this purpose the (known) resistance values of these
shunts were determined using the OMICRON CT ANALYZER.
From the results it can be concluded that the deviation between the results of the tests carried out
using the analyzer and those carried out using other instruments is small.
Although the checks are not systematically covering the full range of the possibilities of the analyzer
for the determination of winding resistance, it can be concluded that the results that were produced by
the OMICRON CT ANALYZER were correct and precise for those resistance values that were
determined.
First in April 2005 and in October 2005 several checks were carried out in order to verify whether the
OMICRON CT ANALYZER was able to perform its function: determination of current transformer
transformation errors.
For this purpose accuracy tests were carried out in two ways: using the OMICRON CT ANALYZER
and using a KEMA laboratory accuracy measuring bridge. The accuracy tests were carried out on
several current transformers of various design and ratings. For the presentation and results four
examples were selected from the many tests carried out.
The transformation errors of four current transformers were measured using 1) the KEMA
transformation error measuring bridge and 2) the OMICRON CT ANALYZER.
The transformation errors of the current transformers generally were determined at 50 Hz.
-7- 70570064-HVL 06-1071
The current error (%) and the phase displacement (min) were determined at rated current (2000 A)
and respectively with burdens of 2,5 VA, COS(β) = 1,0 and of 1,25 VA, COS(β) = 1,0. In this case the
measurements were carried out at both 50 Hz and 60 Hz.
The current error (%) and the phase displacement (min) were determined at rated current (2000 A)
and respectively with burdens of 2,58 VA, COS(β) = 1 and of 15,25 VA, COS(β) = 0,8 (15 VA is rated
burden). The measurements were carried out at 50 Hz.
The current error (%) and the phase displacement (min) were determined at rated current (1000 A)
and respectively with burdens of 1,29 VA, COS(β) = 1 and of 14,9 VA, COS(β) = 0,8 (1 VA is rated
burden). The measurements were carried out at 50 Hz.
The current error (%) and the phase displacement (min) were determined at rated current (1000 A)
and respectively with burdens of 1,28 VA, COS(β) = 1 and of 14,8 VA, COS(β) = 0,8 (1,5 VA is rated
burden). The measurements were carried out at 50 Hz.
The current error (%) and the phase displacement (min) were determined at rated current (1200 A)
and respectively with burdens of 5,16 VA, COS(β) = 0,8 and of 7,50 VA, COS(β) = 0,8 and of 9,86 VA,
COS(β) = 0,8 (5 VA is rated burden). The measurements were carried out at 50 Hz.
- The results of the determination of transformation errors, as carried out, on the current
transformers and for the values of current and burden as indicated in this report, and measured
using the OMICRON CT ANALYZER, do not differ significantly from those determined by the
KEMA measuring bridge and instrumentation.
In October 2005 several checks were carried out in order to verify whether the OMICRON CT
ANALYZER was able to perform its function: determination of current transformer excitation curves.
For this purpose curves measuring tests were carried out in two ways: using the OMICRON CT
ANALYZER and using a KEMA laboratory instruments. The measurements were carried out on
several current transformers of various design and ratings. For the presentation and results four
examples were selected from the many tests carried out.
The excitation curves of four current transformers were measured using 1) the KEMA laboratory
instruments and 2) the OMICRON CT ANALYZER.
The excitation curves of the current transformers were determined at the rated frequency 50 Hz.
It has to be noted that the accuracy of the measurements by KEMA may have been affected by noise
coming from neighboring tests. This, together with the nature of the selected instrumentation for
measurement of voltage and current may have resulted in a higher level of uncertainty than normal.
It has to be noted too that the values in the tables, for the excitation current were taken at the voltage
levels selected by KEMA. From the OMICRON CT ANALYZER the excitation currents at these voltage
levels could be taken only after interpolation between other voltage levels, as selected by the
OMICRON CT ANALYZER, within the appropriate range.
-9- 70570064-HVL 06-1071
- The results of the determination of excitation curves, as carried out, on the current transformers
and for the values of voltage as indicated in this report, and measured using the OMICRON CT
ANALYZER, do not differ significantly from those determined by the KEMA measuring
instruments.
- Some differences, in individual numbers of current and voltage, in several cases, are mostly related
to saturation effects that are visible in the curves and do not affect the overall curve and still allow
the correct derivation of further quantities from these curves.
In April and in October 2005 several checks were carried out in order to verify whether the OMICRON
CT ANALYZER was able to perform other test functions, additional to those mentioned in paragraphs
1, 2 and 3 of this report.
Since, as a result of the automatic and integrated operation of the OMICRON CT ANALYZER, several
additional test functions came into operation during the checks and verifications performed, these
functions resulted in output parameter values such as for Fs, Ts, knee-point data, etc.
The presentation of the results by the OMICRON CT ANALYZER seems trustworthy; however not all
results were verified or checked by KEMA.
Since the magnetization curves had been determined for several current transformers, by KEMA and
by the OMICRON CT ANALYZER as well, a comparison could be made for the resulting knee-point
data.
It has to be noted that the accuracy of the measurements by KEMA may have been affected by noise
coming from neighboring tests. This, together with the nature of the selected instrumentation for
measurement of voltage and current may have resulted in a higher level of uncertainty than normal.
- that the knee-point data, determined by the OMICRON CT-ANALYZER, are quite similar to or
better than those following from the measurements carried out by KEMA
Appendix 1 page 1
Atmospheric conditions
Ambient temperature 20 °C Ambient air pressure 1018 hPa
Object temperature 20 °C Humidity 14 g/m3
Appendix 1 page 2
Atmospheric conditions
Ambient temperature 20 °C Ambient air pressure 1015 hPa
Object temperature 20 °C Humidity 14 g/m3
Appendix 2 page 1
Atmospheric conditions
Ambient temperature 20 °C Ambient air pressure 1014 hPa
Object temperature 20 °C Humidity 12 g/m3
Appendix 2 page 2
Atmospheric conditions
Ambient temperature 20 °C Ambient air pressure 1018 hPa
Object temperature 20 °C Humidity 14 g/m3
Appendix 2 page 3
Atmospheric conditions
Ambient temperature 20 °C Ambient air pressure 1018 hPa
Object temperature 20 °C Humidity 14 g/m3
Appendix 3 page 1
Atmospheric conditions
Ambient temperature 20 °C Ambient air pressure 1016 hPa
Object temperature 20 °C Humidity 13 g/m3
Appendix 3 page 2
Atmospheric conditions
Ambient temperature 20 °C Ambient air pressure 1016 hPa
Object temperature 20 °C Humidity 13 g/m3
100
Terminal Voltage [Vrms]
Omicron
10
KEMA
1
0,0001 0,001 0,01 0,1 1 10
Appendix 3 page 3
Atmospheric conditions
Ambient temperature 20 °C Ambient air pressure 1016 hPa
Object temperature 20 °C Humidity 13 g/m3
Appendix 3 page 4
Atmospheric conditions
Ambient temperature 20 °C Ambient air pressure 1016 hPa
Object temperature 20 °C Humidity 13 g/m3
1000
Terminal Voltage [Vrms]
100
OMICRON
KEMA
10
1
0,0001 0,001 0,01 0,1 1 10
Excitation Current [Arms]
-20- 70570064-HVL 06-1071
Appendix 3 page 5
Atmospheric conditions
Ambient temperature 20 °C Ambient air pressure 1016 hPa
Object temperature 20 °C Humidity 13 g/m3
Appendix 3 page 6
10000
1000
Terminal Voltage [Vrms]
OMICRON
100
KEMA
10
1
0,001 0,01 0,1 1 10 100
Excitation Current [Irms]
-22- 70570064-HVL 06-1071
Appendix 3 page 7
Atmospheric conditions
Ambient temperature 20 °C Ambient air pressure 1016 hPa
Object temperature 20 °C Humidity 13 g/m3
Appendix 3 page 8
Atmospheric conditions
Ambient temperature 20 °C Ambient air pressure 1016 hPa
Object temperature 20 °C Humidity 13 g/m3
10000
1000
Terminal Voltage [Vrms]
OMICRON
100
KEMA
10
1
0,001 0,01 0,1 1 10 100
Excitation Current [rms]
-24- 70570064-HVL 06-1071
The measurement uncertainties in the results presented are as specified below unless otherwise
indicated.
measurement of impedance ≤ 1%
ac-resistance measurement
measurement of losses ≤ 1%
measurement of dc resistance 1 µΩ - 5 µΩ : 1%
5 µΩ - 10 µΩ : 0,5%
10 µΩ - 200 µΩ : 0,2%
measurement of conductivity 5%
tensile test 1%
Appendix 4 page 1
Atmospheric conditions
Ambient temperature 20 °C Ambient air pressure 1016 hPa
Object temperature 20 °C Humidity 13 g/m3
APPENDIX A
(4 pages)
100
Terminal Voltage [Vrms]
Omicron
10
KEMA
1
0.0001 0.001 0.01 0.1 1 10
Excitation Current [Arms]
Appendix A page 2
0.40%
0.30%
0.20%
0.10%
pos. border
neg. border
0.00%
OMICRON
0% 50% 100% 150% 200% 250%
KEMA
-0.10%
-0.20%
-0.30%
-0.40%
-0.50%
Appendix A page 3
70
60
50
Terminal Voltage [Vrms]
40 Tolerance +
Tolerance -
Diff KEMA/Omic
30 Knee Point
20
10
0
-0.005 -0.004 -0.003 -0.002 -0.001 0 0.001 0.002 0.003 0.004 0.005
Current [Arms]
Appendix A page 4
100
Terminal Voltage [Vrms]
10
Omicron
KEMA
pos Tolerance
neg Tolerance
120 % Inom border
1
0.000001 0.00001 0.0001 0.001 0.01 0.1 1 10
0.1
excitation current [Irms]
APPENDIX B
(4 pages)
1000
Terminal Voltage [Vrms]
100
OMICRON
KEMA
10
1
0.0001 0.001 0.01 0.1 1 10
Excitation Current [Arms]
The graph below shows the current error within the class 0.1 % tolerance borders according to
IEC 60044-1. It shows the difference of the current error calculated from the excitation graph
measured with the OMICRON CT Analyzer and the KEMA test equipment. The calculation was done
under the assumption only the excitation current would influence the ratio error, therefore, no
excitation current would result in a ratio error of zero.
The used burden was 15 VA, Inom = 1 A. The graph is supposed to show the influence of the
measurement error of the excitation measurement and not the absolute ratio error.
0.50%
0.40%
0.30%
0.20%
Current Error [%]
-0.20%
-0.30%
-0.40%
-0.50%
% Nominal Current
Appendix B page 3
The graphs are shown up to the knee point only. Above the knee point, the deviation between the two
curves cannot be calculated properly anymore because of the steep current rise. In this area, a small
deviation in voltage would result in a large deviation in current.
Only two tolerance borders were used: a tolerance of 0.2 % below 20 % of Inom, and a tolerance of
0.1 % above 20 % of Inom.
140
120
100
Terminal Voltage [Vrms]
80 Tolerance +
Tolerance -
Diff KEMA/Omic
60 Knee Point
40
20
0
-0.008 -0.006 -0.004 -0.002 0 0.002 0.004 0.006 0.008
Current [Arms]
Appendix B page 4
The graph below shows the positive tolerance border (yellow) and the negative tolerance border as
defined in IEC 60044-1 for class 0.1 as well as the two measured excitation curves. The logarithmic
scale does not allow the negative tolerance border to be shown properly, therefore the negative values
were limited to 10µA. IEC60044-1 only defines the tolerances up to 120% of nominal current.
The 0.1% tolerance was used for all currents above 20% Inom with a burden of 15 Ohm.
U Ter min al
The output current was calculated by Is =
RCT + RBurden
The error border was calculated as in the following example with 100% Inom.
I posTolerance = I s + I s * 0.001
1000
Terminal Voltage [Vrms]
100 Omicron
KEMA
pos Tolerance
neg Tolerance
10 120 % Inom border
1
0.00001 0.0001 0.001 0.01 0.1 1 10
excitation current [Irms]
APPENDIX C
(5 pages)
Comment: The OMICRON CT-Analyzer does only guarantee high accuracy for non gapped cores but
the measurement results below show that also on gapped protection cores reasonable results can be
reached.
10000
1000
Terminal Voltage [Vrms]
OMICRON
100
KEMA
10
1
0.001 0.01 0.1 1 10 100
Excitation Current [Irm s]
The ratio error measured by KEMA was -0,97% the ratio error measured from CT-Analyzer was -
1,076% (difference 0.106%). The used burden was 14,75VA, Inom = 1 A.
The graph is supposed to show the influence of the measurement error of the excitation measurement
and not the absolute ratio error.
1.50%
1.00%
0.50%
Ratio Error in %
pos. border
0.00% neg. border
0% 200% 400% 600% 800% 1000% 1200% deviation
-0.50%
-1.00%
-1.50%
% of Nom inal Current
1.50%
1.00%
0.50%
Ratio Error in %
pos. border
neg. border
0.00%
OMICRON
0% 200% 400% 600% 800% 1000% 1200%
KEMA
-0.50%
-1.00%
-1.50%
% of Nom inal Current
The graph is shown up to the knee point only. Above the knee point, the deviation between the two
curves cannot be calculated properly anymore because of the steep current rise. In this area, a small
deviation in voltage would result in a large deviation in current.
1400
1200
1000
Terminal Voltage [Vrms]
400
200
0
-0.8 -0.6 -0.4 -0.2 0 0.2 0.4 0.6 0.8
Current [Arm s]
Yellow line: Deviation of the output current measured with KEMA equipment and OMICRON CT-
Analyzer. No ratio error is considered.
Dark blue line: +1% error border
Pink line: -1% error border
Appendix C page 5
10000
1000
Terminal Voltage [Vrms]
Omicron
KEMA
100
pos Tolerance
neg Tolerance
10
1
0.00001 0.0001 0.001 0.01 0.1 1 10 100
excitation current [Irm s]
APPENDIX D
(5 pages)
Comment: The ratio measurement of the TPZ core was done according IEC 60044-1 to get
comparable results between the KEMA equipment and the CT-Analyzer. The OMICRON CT-Analyzer
does only guarantee high accuracy for non gapped cores but the measurement results below show
that also on gapped protection cores reasonable results can be reached.
10000
1000
Terminal Voltage [Vrms]
OMICRON
100
KEMA
10
1
0.001 0.01 0.1 1 10 100
Excitation Current [rms]
The ratio error measured by KEMA was -1,94% the ratio error measured from CT-Analyzer was -
1,658% (difference 0.28%). The used burden was 5,16VA, Inom = 1A.
The graph is supposed to show the influence of the measurement error of the excitation measurement
and not the absolute ratio error.
1.50%
1.00%
0.50%
Ratio error in %
pos. border
0.00% neg. border
0% 200% 400% 600% 800% 1000% 1200% rel. deviation
-0.50%
-1.00%
-1.50%
% of nom inal Current
The ratio error measured by KEMA was -1,94% the ratio error measured from CT-Analyzer was -
1,658% (difference 0.28%). The used burden was 5,16VA, Inom = 1A.
1.50%
1.00%
0.50%
0.00%
Ratio error in %
-1.50%
-2.00%
-2.50%
% of nom inal Current
160
140
120
Output Current [Arms]
100
Tolerance +
80 Tolerance -
Diff. KEMA/OMICRON
60
40
20
0
-2 -1.5 -1 -0.5 0 0.5 1 1.5 2
Error current [Arm s]
Yellow line: Deviation of the output current measured with KEMA equipment and OMICRON CT-
Analyzer. No ratio error considered.
Dark blue line: +1% error border
Pink line: -1% error border
Appendix D page 5
10000
1000
Terminal Voltage [Vrms]
Omicron
KEMA
100
pos Tolerance
neg Tolerance
10
1
0.001 0.01 0.1 1 10 100
excitation current [Irm s]