3. Technical Specification:
Model AM2302
Power supply 3.3-5.5V DC
Output signal digital signal via 1-wire bus
Sensing element Polymer humidity capacitor
Operating range humidity 0-100%RH; temperature -40~80Celsius
Accuracy +-2%RH(Max +-5%RH);
humidity +-2%RH temperature +-0.5Celsius
Resolution or sensitivity humidity 0.1%RH; temperature 0.1Celsius
Repeatability humidity +-1%RH; temperature +-0.2Celsius
Humidity hysteresis +-0.3%RH
Long-term Stability +-0.5%RH/year
Interchangeability fully interchangeable
Email: [email protected]
Your specialist in innovating humidity & temperature sensors
4. Dimensions: (unit----mm)
6. Operating specifications:
(1) Power and Pins
Power's voltage should be 3.3-5.5V DC. When power is supplied to sensor, don't send any instruction to the sensor
within one second to pass unstable status. One capacitor valued 100nF can be added between VDD and GND for
wave filtering.
(2) Communication and signal
1-wire bus is used for communication between MCU and AM2302. ( Our 1-wire bus is specially designed, it's
different from Maxim/Dallas 1-wire bus, so it's incompatible with Dallas 1-wire bus.)
Illustration of our 1-wire bus:
Email: [email protected]
Your specialist in innovating humidity & temperature sensors
When MCU send start signal, AM2302 change from standby-status to running-status. When MCU finishs sending
the start signal, AM2302 will send response signal of 40-bit data that reflect the relative humidity and temperature
to MCU. Without start signal from MCU, AM2302 will not give response signal to MCU. One start signal for one
response data from AM2302 that reflect the relative humidity and temperature. AM2302 will change to standby
status when data collecting finished if it don't receive start signal from MCU again.
See below figure for overall communication process, the interval of whole process must beyond 2 seconds.
------------------------------------------------------------------------------------------------------------------------------------
1) Step 1: MCU send out start signal to AM2302 and AM2302 send response signal to MCU
Email: [email protected]
Your specialist in innovating humidity & temperature sensors
Data-bus's free status is high voltage level. When communication between MCU and AM2302 begins, MCU will
pull low data-bus and this process must beyond at least 1~10ms to ensure AM2302 could detect MCU's signal, then
MCU will pulls up and wait 20-40us for AM2302's response.
When AM2302 detect the start signal, AM2302 will pull low the bus 80us as response signal, then AM2302 pulls
up 80us for preparation to send data. See below figure:
------------------------------------------------------------------------------------------------------------------------------------
2). Step 2: AM2302 send data to MCU
When AM2302 is sending data to MCU, every bit's transmission begin with low-voltage-level that last 50us, the
following high-voltage-level signal's length decide the bit is "1" or "0". See below figures:
Attention:
If signal from AM2302 is always high-voltage-level, it means AM2302 is not working properly, please check
the electrical connection status.
Email: [email protected]
Your specialist in innovating humidity & temperature sensors
7. Electrical Characteristics:
Items Condition Min Typical Max Unit
Power supply DC 3.3 5 6 V
Current supply Measuring 1 1.5 mA
Stand-by 40 Null 50 uA
Collecting Second 2 Second
period
8. Attentions of application:
(1) Operating and storage conditions
We don't recommend the applying RH-range beyond the range stated in this specification. The AM2302 sensor
can recover after working in abnormal operating condition to calibrated status, but will accelerate sensors' aging.
(2) Attentions to chemical materials
Vapor from chemical materials may interfere AM2302's sensitive-elements and debase AM2302's sensitivity.
(3) Disposal when (1) & (2) happens
Step one: Keep the AM2302 sensor at condition of Temperature 50~60Celsius, humidity <10%RH for 2 hours;
Step two: After step one, keep the AM2302 sensor at condition of Temperature 20~30Celsius, humidity
>70%RH for 5 hours.
(4) Attention to temperature's affection
Relative humidity strongly depend on temperature, that is why we use temperature compensation technology to
ensure accurate measurement of RH. But it's still be much better to keep the sensor at same temperature when
sensing.
AM2302 should be mounted at the place as far as possible from parts that may cause change to temperature.
(5) Attentions to light
Long time exposure to strong light and ultraviolet may debase AM2302's performance.
(6) Attentions to connection wires
The connection wires' quality will effect communication's quality and distance, high quality shielding-wire is
recommended.
(7) Other attentions
* Welding temperature should be bellow 260Celsius.
* Avoid using the sensor under dew condition.
* Don't use this product in safety or emergency stop devices or any other occasion that failure of AM2302 may
cause personal injury.