TOC Application - TOC-L - USP-Specified TOC System Suitability Test by TOC-LCPH
TOC Application - TOC-L - USP-Specified TOC System Suitability Test by TOC-LCPH
Application
Total Organic Carbon Analysis
No. O44
The United States Pharmacopeia (USP) specifies the use n TOC System Suitability Test Data by USP Method
of Total Organic Carbon (TOC) for management of The TOC system suitability test was conducted using
organic impurities in purified water (PW) and water for the Shimadzu TOC-LCPH combustion catalytic oxidation
injection (WFI). According to the USP, the TOC analyzer type analyzer by the procedure outlined in Table 1. The
to be used for these analyses must satisfy the TOC instrument was calibrated beforehand using aqueous
system suitability testing requirement, and must be solutions of potassium hydrogen phthalate with carbon
capable of detecting TOC at concentrations below concentrations of 0 and 0.5 mgC/L, respectively. The
0.05 mg/L. TOC system suitability test data are shown in Fig. 1.
Here, using the Shimadzu TOC-L CPH combustion According to the USP, the detection rate is to be
catalytic oxidation type analyzer, we introduce examples evaluated using the analyzer response values, but here,
of TOC system suitability testing and measurement of the measured concentrations were used instead.
TOC at a concentration below 0.05 mg/L. The result indicated a 100.1 % detection rate with
re s p e c t t o t h e s y s t e m s u i t a b i l i t y t e s t s o l u t i o n
(1,4-benzoquinone aqueous solution), thereby satisfying
n TOC System Suitability Test Specified in USP the system suitability test requirement. (Table 2)
The TOC system suitability test indicated in the USP
specifies the use of two types of USP reference
standards (sucrose and 1,4-benzoquinone). Sucrose is Measurement Conditions
used as a test solution standard, and 1,4-benzoquinone Analyzer : Shimadzu TOC-LCPH Combustion Catalytic
is used as the system suitability test solution. Oxidation Type Analyzer
In addition, calibration of the TOC analyzer is specified Catalyst : High-sensitivity catalyst
Injection Volume : 816 µL
to be conducted using a method that is suitable for Measurement Item : TOC (= NPOC: TOC by acidification/sparging)
that instrument. The test procedure is shown in Table 1. Calibration Curve : 2-point calibration curve using 0 – 0.5 mgC/L
potassium hydrogen phthalate aqueous solution
Table 2 Results of TOC System Suitability Test Table 3 Measurement Data for TOC Concentrations Below
0.05 mg/L (50 µg/L)
(1) Distilled water TOC value rw= 0.0205 mg/L
Coefficient of
(2) Sucrose standard solution TOC value rs = 0.5173 mg/L Sample Name TOC Value [mgC/L]
Variation (CV) [%]
(3) System suitability test solution (1,4-benzoquinone aqueous Potassium hydrogen
solution) TOC value rss = 0.5176 mg/L phthalate aqueous 0.047 2.66
solution
(4) System suitability test solution detection rate
= 100 (rss - rw) / (rs - rw)
= 100 (0.5176 - 0.0205) / (0.5173 - 0.0205)
= 100.1 %
Measurement Conditions