PCM CMX User Guide
PCM CMX User Guide
User Guide
By Beamex
BEAMEX OY AB
Ristisuonraitti 10
FIN - 68600 Pietarsaari
FINLAND
Tel: +358-10-5505000
Fax: +358-10-5505404
E-mail: [email protected]
Internet: http://www.beamex.com
Feedback xi
Introduction 1
General ........................................................................................... 1
CMX Family .......................................................................... 2
SSA and Support .................................................................. 4
About This Document...................................................................... 5
Typographic conventions...................................................... 5
A Short Presentation of the Sections in This Document ....... 6
Other Documents ................................................................. 8
What’s New .......................................................................... 9
About Calibration........................................................................... 10
What is Calibrated? ............................................................ 10
The Three Steps ................................................................. 10
Index 226
Notes 232
Company:
Contact person:
Street Address:
City:
State:
Zip Code:
Country:
E-mail:
Telephone:
Fax:
3. Any comments when receiving the product. Did the package contain all required
items and was it as expected?
_________________________________________________________________
_________________________________________________________________
4. For how long have you been using the product? __________________
8. Did anything in the product exceed your expectations? In that case, what was it?
_________________________________________________________________
_________________________________________________________________
9. Did anything in the product disappoint you? In that case, please specify.
_________________________________________________________________
_________________________________________________________________
10.Any ideas You want to propose to Beamex so that we can improve our products,
operations and/or services.
_________________________________________________________________
_________________________________________________________________
General
Beamex was founded in 1975 by people with experience in
calibration and quality maintenance of instruments. CMX is
developed based on that knowledge. CMX also represents
Beamex’s 3rd generation of calibration management soft-
ware.
CMX has an “Explorer-like” user interface. This enables the
software to be simple to use by any operator familiar with
Windows®.
CMX communicates with Beamex's communicating calibra-
tors (excluding PC105). Several calibrators by other manu-
facturers also communicate with CMX.
The main features are:
Automated calibration and documentation
Versatile calibrator communication
Support for several database platforms
Security and change management control
Pocket PC interface
Extensive services and support
Typographic conventions
This document uses the following typographic conventions:
All words related to a CMX window are shown us-
ing bold font style. Specifically:
o Menu and submenu commands. A comma
separates the menu and submenu command
from each other.
Example: File, Exit.
o Command buttons:
Example: Click OK.
o Window names:
Example: the Calibrators window.
File names and directories are in capital letters.
Example: D:\CMX\CMX.EXE.
References to other parts in this guide are shown
in bold.
Example: See the Calibration section in this
guide.
Introduction
Basic Information about CMX, this manual and calibration.
Starting CMX
A short section telling you how to start CMX and how to se-
lect a database.
Basic Settings
Contains information that should be reviewed when CMX is
taken into use. Many of the settings described in this section
are usually done only once, like the Plant Structure data.
Certain parts, like the User and User Group settings are up-
dated whenever changes occur.
Calibration
Here are the ways to utilize CMX for calibration presented.
Communication with a calibrator as well as manual data en-
try. This section also includes a presentation on how to view
the calibration results in CMX.
Pocket PC Interface
Availability: Pocket PC Interface allows you to send instrument data to a
–– CMX Light Pocket PC, manually enter calibration results and receive the
CMX Professional
results to CMX. In other words, the Pocket PC Interface is a
CMX Enterprise
"mobile manual entry utility".
Maintenance Inspection
Availability: Maintenance Inspection allows you to perform check proce-
–– CMX Light dures according to IEC 60079-17 standard and also accord-
CMX Professional
CMX Enterprise
ing to your own needs. The actual check procedure is done
using a compatible tablet with bmobiletm software.
Additional Information
Contains useful information of, e.g. calibrators communi-
cating with CMX.
Note.
These help files are also available pdf format in the Docu-
ments folder of the CMX installation media.
What is Calibrated?
The calibration may be carried out for a process Position
(sometimes also called a "Tag") with an installed Device, or
the calibration may be for an uninstalled Device; more pre-
cisely, the calibration is for a particular Function of the Posi-
tion/Device.
Note.
When downloading the installation instructions, remember to
select the instructions that suit the CMX version at hand.
Notes.
All databases mentioned above have the same Supervisor
User ID and Password (shipped with CMX installation me-
dia).
More info of CMX versions is in Introduction section's chap-
ter CMX Family.
Note.
CMX includes a possibility to utilize either SQL Server Au-
thentication (default) or Windows Authentication. Use the
Authentication selection list to choose the one you want.
Notes.
PARAMETER DEFINITION
/DB:DATABASE NAME The name of the database to be opened.
/DB: parameter is mandatory
/LANG:LANGUAGE The user interface language. Language
name should be written as is it seen in lan-
guage selection list during normal start/login,
e.g. "FINNISH".
If no /LANG: parameter is specified, CMX
uses the language chosen in a previous
start.
If no /LANG: parameter is specified and no
previous starts (with current version of CMX)
have been made, English language is used.
/USER:USER ID User ID to be used when opening the data-
base.
If no /USER: parameter is specified, CMX
login dialog is displayed.
/PWD:PASSWORD Password to be used when opening the da-
tabase.
If password is required and no /PWD: pa-
rameter is specified, CMX login dialog is dis-
played.
/P:POSITION ID The Position Property window is opened and
the data of the specified position is dis-
played.
If the Position ID is invalid, the Position
Property window is not opened.
/D:DEVICE ID The Device Property window is opened and
the data of the specified device is displayed.
If the Device ID is invalid, the Device Proper-
ty window is not opened.
If both a /P: and a /D: parameter is entered, the /P: parame-
ter takes precedence and the /D: parameter is ignored.
Wildcards can also be used. For example "/P:101%" displays
all positions that begin with "101". Wildcards used in a Mi-
crosoft SQL Server database are presented in section Que-
ries, Filters and Sets, chapter Wildcards.
Note.
The same colors indicating the need of calibration are used
for Calibrators, Devices, Calibrator Modules and Users in
their respective trees.
Notes.
CMX supports deactivating items shown in the Tree Struc-
ture (not available in CMX Light). Deactivation means that a
Position/Device etc. is no longer in use. A deactivated item
has a small red “x” in the lower right corner of the icon.
A checked out Position/Device has a red border surrounding
the icon. More of checking out in section Check Out /Check
In.
Work Order Handling is described in section Work Order
Handling.
File
Database … Opens a dialog for viewing/changing
the database connection.
Availability:
–– CMX Light
CMX Professional(*
–– CMX Enterprise
*) with no floating server
license option.
Log Off … Opens a dialog for viewing/changing
the database host connection.
Availability:
–– CMX Light
CMX Professional(+
CMX Enterprise
+)
with floating server
license option.
Import, Docu- A possibility to import Certificate,
ment Layouts… Report and Label layouts from an-
other database/source.
Availability:
–– CMX Light
CMX Professional
CMX Enterprise
Export, Docu- A possibility to export Certificate,
ment Layouts… Report and Label layouts to make
them available for another data-
base.
Availability:
–– CMX Light
CMX Professional
CMX Enterprise
Exit Closes CMX
View
Properties Opens/closes the property window
Devices Opens/closes the Devices window
History Trend Opens/closes the History Trend
window.
Availability:
–– CMX Light
CMX Professional
CMX Enterprise
Position Sets Opens/closes the Position Sets
window
Device Sets Opens/closes the Device Sets win-
dow
Refresh Refreshes the active tree.
*) Availability:
–– CMX Light
CMX Professional
CMX Enterprise
Help
CMX Help… Opens CMX’s main help file.
CMX Calculations… Opens CMX Calculations help file
which describes the calculations
that CMX performs.
CMX Report Opens CMX Report Variables help
Variables… file which describes CMX variables
available in CMX's Report Design
tool.
Availability of this menu option:
–– CMX Light
CMX Professional
CMX Enterprise
Index… Opens CMX help file's index tab.
Search… Opens CMX help file's search tab.
About CMX… Opens the window where CMX’s
license and copyright info is dis-
played.
Accessibility
Availability: CMX's Accessibility feature allows limiting access of data on
–– CMX Light user group level. Define users in groups according to the de-
CMX Professional
partment/plant etc. The accessibility for the following data-
CMX Enterprise
base items may be defined:
Plant Structure Nodes,
Calibrators,
Calibrator Modules,
Positions,
Devices and
Saved Filters.
The accessibility setting is done for each item individually
(available among the item's properties). The setting is hierar-
chical; if the access of a plant node is denied for certain user
groups, access to all nodes, positions and devices below that
node are also denied.
This feature is useful, e.g. when a single database contains
data of several departments/plants etc. Then users arranged
to groups according to their department/plant see only the
database items that concern them.
Notes.
Only users with the proper authorization can view and/or edit
accessibility settings.
The possibility to save filters is not available in CMX Light.
See also:
Users and User Groups and Site User Maintenance both
in section Basic Settings.
Common Features
This chapter and its subchapters describe some fea-
tures/utilities that is used not just in one window but in sever-
al CMX windows.
Activation/Deactivation
Availability: CMX's Activation/Deactivation feature allows activation and
–– CMX Light deactivation of Positions, Devices as well as most of the oth-
CMX Professional
er items shown in a Tree Structure. Deactivation of, e.g. a
CMX Enterprise
Device means that the Device is no longer used. The deac-
tivated item has a small red “x” in the lower right corner of
the icon shown in the Tree Structure.
Deactivation is hierarchical. When you, e.g. deactivate a
Plant Structure node, all Positions, Devices, Functions and
Procedures below that node are also deactivated. Further-
more: (re)activation is also hierarchical.
Note.
Deactivation is not possible for a Device that is installed in a
Position. Unlink the Device and then deactivate it.
Note.
The number of links for each item is not limited. Additionally,
there are no file type restrictions. Any file will do just as long
as you have software suited to view the file.
To make the User Fields visible you must have the user
rights to customize CMX’s User Interface. Customizing is de-
scribed in Basic Settings section’s chapter Customizing
the User Interface.
The text fields and check boxes may be freely used. The la-
bel text for the free field may be edited using either the User
Interface customizing tools or the Direct Editing of the User
Interface tool also described in the Basic Settings section.
Pocket PC option supports the use of all User Fields. Cer-
tain modern calibrators support calibration related User
Fields and Checks. Any data entered into these fields are re-
trieved to CMX and can be seen among calibration results.
To activate these fields in CMX (and compatible calibrators),
customize the Calibration Result Properties window with
Customizing the User Interface tool.
General
The Basic Settings section contains information of CMX’s
settings that should be checked/edited before taking CMX in-
to full use. All settings can be changed later on too, but going
through the settings at first makes the daily work more effi-
cient.
Note.
If you utilize the Weighing Instrument Calibration feature, see
chapter Basic Settings for Weighing Instruments in sec-
tion Calibrating Weighing Instruments. It contains infor-
mation of Weight Sets, Weights and other Weighing Instru-
ment Calibration related Basic Settings.
Note.
Certain option pages may also be opened directly from a
configuration window, e.g. the Security page may be
opened from the Tools menu of the Users window.
General Settings
As the name implies, the page contains settings that define
what is done while starting up CMX (e.g. Position Query) or
when opening certain windows, like the Calibrator window.
Field level information is available in the on-line help deliv-
ered with CMX.
Notes.
Red text color for the Position ID / Device ID indicates that
the calibration of one or more of the Position’s/Device's
Functions are overdue (the scheduled calibration date has
expired).
The same colors indicating the need of recalibration are used
for Calibrators and Calibrator Modules in their respective
trees.
ID Generation Options
Each Position and Device need to have a unique ID. It is
used for identifying Positions and Devices.
The Position ID and/or the Device ID can be manually en-
tered or created automatically, if activated on this page. Also
the format of the automatically created IDs (prefix-number-
suffix) can be defined.
Field level information is available in the on-line help deliv-
ered with CMX.
Security Options
On this page you can define whether users need a password
as well as various other password related settings.
Changing these settings is allowed to database administra-
tors and users with the rights to change security options.
Field level information is available in the on-line help deliv-
ered with CMX.
LDAP Options
Note.
You may override LDAP authentication by using CMX's Li-
cense Key number as the User ID and Master Password as
the password. Both the License Key number and the Master
Password were shipped with the original delivery of CMX.
Note.
Users entered into CMX’s database need not be users that
actually do calibrations. Users responsible for the calibra-
tion/maintenance duties may also be entered. They for in-
stance sign calibrations and accept changes made to the da-
tabase.
Note.
CMX automatically places the current Windows® User ID as
the default user for logging into CMX. For ease of use, enter
the same User IDs into CMX.
Supervisors
Supervisors are by default allowed to do anything that can
be done with CMX, e.g. add Users, reactivate a User ac-
count, define User Groups etc. A Supervisor need not be as-
signed to any Group, Supervisors have all rights anyhow.
Whether a user is a Supervisor or not, can be seen (or edit-
ed, if you have the permission) in the property data shown on
the right hand side of the Users window.
There is no upper limit for the amount of supervisors but at
least one Supervisor is required. CMX does not allow delet-
ing the last Supervisor.
See also note in chapter User Deactivation and Lock Out
on page 46.
Users
Each person using the database may be assigned a unique
User ID (assigned by a Supervisor) and optionally a pass-
word. A User can belong to one or more User Groups de-
pending on the duties of that particular User.
Field level information is available in the on-line help deliv-
ered with CMX.
Notes.
The permissions are defined for User Groups. Once a User
is assigned to a Group, he/she gets the permissions defined
for that Group.
If LDAP authentication is enabled, for all CMX users:
Enter the same CMX User IDs as is used as User Names in
Windows network. See also LDAP Options on page 42.
Note.
If LDAP authentication is enabled, passwords are not
changed in CMX. With LDAP, the handling of passwords is
maintained in Windows. See LDAP Options on page 42.
Note.
User's expiry date does not apply to Supervisors. A Supervi-
sor can access the database as long as he/she has Supervi-
sor rights.
Permissions
The permissions tab shown for User Groups contains a list of
commands and tasks available in CMX.
A command or a task that is unchecked in the list may not be
performed by users belonging to that group. They are disa-
bled in CMX’s menus and toolbars.
A list of available permissions is available in CMX on-line
help.
See also:
Accessibility in section General Presentation of CMX.
Note.
Site User Management affects Accessibility, Owner Group
and Owner selection lists. More of this in chapter Working
with Site User Maintenance.
Notes.
Adding standard User Groups is recommended, since they
can be used to define different permissions for different Us-
ers. Each User is then assigned to (at least) one standard
User Group that defines the User's permissions and also to a
Site Group which defines the User's location in the organiza-
tion.
Any non-supervisor Users added before Site User Mainte-
nance was enabled may (inadvertently) be unassigned to a
Site Group. Users not assigned to any Site Group are not
able to log into CMX. Make sure all non-supervisor Users are
assigned to at least one Site Group when Site User Mainte-
nance is active.
Note.
Deleting a node also deletes all items hierarchically below
the node to be deleted.
Note.
There is a separate window for Calibrator Manufacturers
presented on page 63, so only data for instruments to be cal-
ibrated should be entered here.
Notes.
A Device Manufacturer is also added to the list of Calibrator
Manufacturers. This minimizes the workload if a single
manufacturer delivers both devices that need to be calibrated
and calibrators.
Note.
Deleting a Device Manufacturer is not possible if there is a
reference to the Manufacturer in the database.
Notes.
The Device Model will be added to the selected Manufactur-
er on the Tree Structure. If a Device Model is selected, the
new Device Model will be added to the same Device Manu-
facturer as the selected Device Model.
Currently there is no possibility to move a Device Model from
one Manufacturer to another.
Note.
Deleting a Device Model is not possible if there is a refer-
ence to the Device Model in the database.
Notes.
If you have a Calibrator that is not modular (or has only one
module), just repeat the Calibrator data for the module and
then add the Module Range(s) for the calibrator.
Adding Calibrator Manufacturers and Models can be done
while adding Calibrators. Just select the “<New…>” option
from the list of Models/Manufacturers.
Note.
Adding a Calibrator manually while viewing the Tree Struc-
ture of the Modules tab gives seemingly no results. Since
the new calibrator has no installed modules it is only shown
in the Tree Structure of the Calibrators tab.
Note.
This chapter explains how to “install and remove” Modules
from a Calibrator in CMX software. Refer to the Manuals of
the Calibrators on how to do it for the actual Calibrator (if it is
allowed for calibrator users).
Notes.
Deleting is not possible if there is a reference to the Calibra-
tor/Module/Module Range in the database.
If a calibrator to be deleted is on a list of Calibrators to Use,
CMX asks for confirmation before the calibrator is deleted.
Calibrators to Use field can be found among Calibration
Procedure data.
When deleting a Calibrator/Module/Module Range, all items
hierarchically below the item to be deleted will also be delet-
ed.
Note.
Possible document links are not copied.
Note.
There is a separate window for entering Device Manufac-
turers presented on page 54, so only data of calibrators
should be entered here.
Note.
A Manufacturer of Calibrators is also added to the list of De-
vice Manufacturers. This minimizes the workload if a single
manufacturer delivers both calibrators and devices that need
to be calibrated.
Note.
Deleting a Calibrator Manufacturer is not possible if there is a
reference to the Manufacturer in the database.
Notes.
The Calibrator Model will be added to the selected Manufac-
turer on the Tree Structure. If a Calibrator Model or a Module
Model is selected, the new Calibrator Model will be added to
the same Calibrator Manufacturer as the selected Calibrator
Model or Module Model.
Currently there is no possibility to move a Calibrator Model
from one Manufacturer to another.
Note.
Deleting a Calibrator Model is not possible if there is a refer-
ence to the Model in the database.
Notes.
The Calibrator Module Model will be added to the selected
Manufacturer on the Tree Structure. If a Calibrator Model or
a Module Model is selected, the new Module Model will be
added to the same Calibrator Manufacturer as the selected
Calibrator Model or Module Model.
Currently there is no possibility to move a Calibrator Module
Model from one Manufacturer to another.
Note.
Deleting a Module Model is not possible if there is a refer-
ence to the Model in the database.
What is a List?
Most devices have a process connection, e.g. a ¼” thread for
that particular purpose. If you enter a lot of device data it will
be an arduous work to write the same connection info over
and over. To minimize the workload, CMX offers the Lists.
Enter the frequently repeated data into the lists. Then when-
ever the data is needed you just pick it from the list of pre-
entered options. No need to write it over and over again.
The Lists window is primarily meant for editing existing list
data. Adding list items can easily be done while selecting da-
ta from a list seen in other windows. There you just enter the
missing item to the list. There is of course the possibility to
add list items in the Lists window too, if you want it.
Currently CMX provides the following lists:
- Calibrator Hierarchy,
- Process Connection(*,
- Process Medium(*,
- Signature Status(+,
- Audit Trail Description(+ and
- User Defined Units.
*) Not available in CMX Light
+)
Not available in CMX Light and
optional in CMX Professional
Note.
Adding User Defined Units is not done in Lists window. They
can only be edited/deleted in Lists window.
A User Defined Unit is added using one of the following
methods:
1. Create a new instrument and for its Function,
select Value as Input and/or Output Quantity.
Then enter the Range values and instead of
selecting an existing User Defined Unit, just
write your own unit.
2. Create a new Function Template. Set Input and/or
Output Quantity as Value and enter a new User
Defined Unit as part of the measurement/keyed
Range.
3. Edit an existing Function with Value as Input
and/or Output Quantity. Enter a new User
Defined Unit as part of the measurement/keyed
Range.
When a User Defined Unit is added using one of the above
mentioned methods, the new unit is included in the list of
User Defined Units shown in Lists window.
Notes.
When adding Functions using the wizard:
The first active Calibration Procedure Template of a Function
Template is always selected for use when creating a Posi-
tion/Device/Function.
When adding Functions using the Tree Structure's pop-
up menu:
The new Function contains all Procedures created for the se-
lected Function Template.
Important!
If you have activated Window's UAC (User Account Control)
feature do as follows:
Run CMX in Run as administrator mode when editing the
User Interface. This is done, e.g. by clicking the secondary
mouse button above CMX's icon and selecting Run as ad-
ministrator from the pop-up menu. Otherwise you will lose
all your edits.
General
CMX supports four ways of adding instruments to its data-
base:
Using the Wizard is the quick way of adding single
Function Positions with an installed Device.
Using the Tree Views seen in the main window. This
method leaves all control to you. Positions, Devices,
Functions and Calibration Procedure definitions are all
added according to your own needs.
Copying a Device and/or a Position. This possibility is
found in the context menus of Tree Structures. More
of this in chapter Copying Positions and Devices.
Receiving new instruments from a calibrator support-
ing this feature (e.g. Beamex’s communicating MC se-
ries calibrators do support this feature). More of this
possibility in Calibration section’s chapter Receiving
Calibration Result Data.
Max. amount of
CMX version Max. amount of Functions in
Positions/Devices Positions/Devices
CMX Light 300/300 400/400
1.000/1.000, 1.300/1.300,
CMX Professional 5.000/5.000 6.500/6.500
10.000/10.000 13.000/13.000
or unlimited or unlimited
CMX Enterprise Unlimited Unlimited
Notes.
The wizard creates a Position/Device with a single Function.
To add more functions use the method described in chapter
Adding Functions on page 79.
For maximum amount of Positions, Devices and Functions,
see chapter Maximum Amount of Positions, Devices and
Functions on page 74.
Notes.
When entering the path manually, remember to use the path
separator defined in the General Settings.
You cannot create the Plant Structure here, but select/enter
the location of a previously added Plant Structure.
Note.
If a certain function need to be entered several times and
there is no suitable Function Type available, consider creat-
ing a custom Function Type Template for it. See section
Basic Settings, chapter Function Templates.
Note.
When you have added an instrument using custom User De-
fined templates in the wizard, please check all Function and
Procedure fields in the property window. The wizard may
have omitted some fields. In that case, enter necessary data
manually.
Adding Positions
There are three ways of adding a Position in CMX:
Click the secondary button on the mouse in the up-
permost row of the Position Tree.
Click the secondary button on the mouse above a
Plant Structure node in the Plant Structure Tree.
When viewing the properties of an existing Position
there is a button for creating a New Position in the
property window.
Position ID is the only required field, all others are optional.
The Position ID field may have a default value based on your
database settings. If it is empty, you need to enter a unique
ID.
Field level information for Positions is available in the on-line
help delivered with CMX.
Notes.
A new position may also be created using the copy option of
a context menu of a tree. Open the menu with your mouse
pointer above a position icon and select the Copy option.
For maximum amount of Positions, see chapter Maximum
Amount of Positions, Devices and Functions on page 74.
Notes.
Adding Device Manufacturers and Models can be done while
adding Devices. Just select the “<New…>” option from the
list of Models/Manufacturers.
A new device may also be created using the copy option of a
context menu of a tree. Open the menu with your mouse
pointer above a device icon and select the Copy option.
For maximum amount of Devices, see chapter Maximum
Amount of Positions, Devices and Functions on page 74.
Note.
Adding a Function to a Position with an installed Device au-
tomatically adds the new Function to the installed Device too.
If the Position has several Devices installed to it, the new
Function is added to the Position but not to any of the in-
stalled Devices.
Notes.
If a certain function need to be entered several times and
there is no suitable Function Template available, consider
creating a custom Function Template for it. See section
Basic Settings, chapter Function Templates.
For maximum amount of Functions, see chapter Maximum
Amount of Positions, Devices and Functions on page 74.
HART® Devices
When adding a HART instrument into CMX's database, the
Analog Output (AO) and the Digital Output (PV) should be
added as two Functions with the following settings:
The Analog Output (AO) as a transmitter with the
output category analog variable selected.
The Digital Output (PV) as a transmitter with the
output category digital variable selected. Select
the same Quantity both for the input and the out-
put. Select HART as the Output Method.
IMPORTANT!
First input and output pair (point) must be same as
input and output 0 % range values.
Enter absolute input and output values in increasing
order (percentage of span point of view).
Last input and output values must be equal or
greater than input and output 100 % range values.
Notes.
Notes.
Use some sort of prefix or suffix, e.g. "#", in the unit name.
This is to indicate that this is a custom pressure unit, not to
be confused with standard units available in CMX.
Notes.
To minimize the risk of confusion that may lead to serious
accidents, avoid using sensor names that resemble "re-
served names", i.e. names of standard sensors available in
CMX.
Use some sort of descriptive text, e.g. sensor's serial num-
ber, in the sensor name. This is to indicate that this is a cus-
tom PRT sensor, not to be confused with standard sensors
available in CMX.
User defined PRT sensors are supported by modern calibra-
tors, e.g. Beamex® MC4 Documenting Process Calibrator
and later.
More of Callendar - van Dusen equation in a separate help
file called CMX_Calculations.chm. Open it from CMX Main
Windows' Help menu. Select option CMX Calculations.
Look for main topic PRT Sensor Calculations and from
there sub topic Callendar - van Dusen Equation.
Notes.
CMX always uses the first available Procedure for the Func-
tion (the uppermost active Procedure shown in the Tree). If a
Function has several Procedures, deactivate all Procedures
preceding the one you want to use for calibration.
There are no limitations for the amount of Calibration Proce-
dures, except the available memory in the comput-
er(s)/server(s) used to run CMX and store its database.
Note.
This method of linking several Devices to a Position is not
the only way. See the next chapter to find out how to link
several already existing Devices to an existing Position.
There you can also find out how to link a single Device to
several Positions.
Note.
When linking a Device with existing Function(s)/Procedure(s)
to a Position, the Device will automatically inherit Position's
Function/Procedure data with the following exceptions: Initial
Calibration Date, Initial Calibration Count and Total Cali-
brations fields stay unique for Position and Device.
Note.
This method of linking applies to situation when both the De-
vices and the Position already exist in the instrument data-
base. See chapter Linking While Creating the Instrument
Database on page 85 to learn how to add several new De-
vices to a Position.
Unlinking Devices
To unlink a Device open the Position Tree (alternatively, the
Plant Structure Tree), expand the tree at the Position where
the Device to be unlinked is located. Open the Device’s Con-
text Menu (pop-up menu) and select the Unlink option.
This method applies to both installed and Spare Devices.
Note.
Assigning a Device as a spare Device does not require that
the Functions of the spare Device and the Position match.
Note.
Copying a Position/Device also copies all Functions and Cal-
ibration Procedures the original Position/Device had.
Note.
Editing linked Functions or Calibration Procedures of linked
Positions and Devices always apply to the Proce-
dure/Function of both the Position and Device.
General
This section describes tools that help you sort, group and
view the instrument data according to you own needs.
What Is a Query
A query is a tool found in all database software. It contains
search criteria that select the data to be displayed.
Available Queries
CMX has queries available for Positions, Devices, Calibra-
tors and Calibrator Modules. Queries are part of the software
and therefore available at any time. Use filters for temporary
filtering the instrument data.
Calibrator/
Query Position Device Module
Description Queries Queries Queries
Does not filter out All Positions All Devices All Calibra-
anything. Everything tors/ Mod-
is shown. ules
Shows only the items Due for Due for Due for
due (or overdue) for Calibration Calibration Calibration
calibration.
Shows the items due Due within… Due with- Due with-
within a specified pe- in… in…
riod.
Shows unlinked Posi- Empty Posi- Free De- -
tions/Devices only. tions vices
Shows all currently New Posi- - -
added Positions. See tions
also the note below.
Notes.
The “New Positions” query is not available in the list of Posi-
tion Queries. It is automatically created when adding Posi-
tions and also disappears automatically when you choose
another query from the list.
Keep in mind that the Position Query list is the secondary
level query. The Plant Structure List, shown above the Po-
sition Query list, is the primary level filtering for the Positions
shown in the Position Tree.
To activate a filter:
When all filter conditions are set, press the Apply button in the
Filter Conditions window. Then the Position or Device Tree
only shows the Positions/Devices that fit the filter parameters.
Both the Position Tree and the Device Tree may have their
own independent filters active.
Filters are not case sensitive. TIRCA and Tirca returns the
same results.
You may also search for empty fields. Enter IS NULL as the
search criteria. This is the exception to case sensitivity: IS
NULL need to be written with capital letters.
Saving Filters
Availability: You may give a name to a filter, set its accessibility and enter
–– CMX Light a description for it using the uppermost fields in the Filter
CMX Professional
Conditions window.
CMX Enterprise
Note.
If no user group is selected in the filter's Accessibility list,
CMX treats the filter as private, i.e. no other user has access
to it. If you want other users to have access to a filter you
created, check as many user groups as you want in the Ac-
cessibility list.
To save a new filter, click the New button in the toolbar and
then enter a name for the filter. After all settings are as re-
quired, save the filter by clicking the Save button in the
toolbar.
To edit an existing filter, click the Edit button in the toolbar.
Then do the necessary edits and click the Save button in the
toolbar.
Note.
Your edits may create a query that does not give any results
at all. Make sure you type all SQL parameters correctly.
Editing a Set
To add an item to a set, drag-drop it from a Tree containing
similar items as the set.
Devices can only be dragged from the Device Tree
seen in the Devices window. If the window is not visi-
ble, open it from the main window’s toolbar or the
View menu.
Positions can be dragged from the Position Tree or
the Plant Structure Tree. One of the Trees is always
available on the left side of the main window.
Removing an item from the set is done from the item’s Con-
text Menu (pop-up menu). Similarly, the whole set is deleted
from the Context Menu opening from the set name (the up-
permost row in the set Tree).
General
Check Out / Check In feature is a utility that allows you to
"lock" the Positions/Devices currently under calibration (or
being edited). The data of Positions/Devices that are
Checked Out may not be calibrated/edited by other CMX us-
ers than the one who did the Check Out. When Checking In,
the Position/Device is "unlocked" and may again be calibrat-
ed/edited by all CMX users.
Check Out / Check In feature is part of CMX's standard tools
starting from CMX V2, revision 2.1.
Note.
You may still calibrate Positions/Devices without utilizing the
Check Out / Check In feature.
Notes.
Note that the option is available only for users with per-
mission to edit Position and Device data.
Note.
When doing multiple Check Outs, CMX displays a dialog
asking to confirm the operation.
Note.
When doing multiple Check Ins, CMX displays a dialog ask-
ing to confirm the operation.
General
Calibrations can be done after the necessary instrument data
is entered into the database.
A Position must have an installed Device (with Func-
tions) before a calibration can be performed, but a De-
vice (with Functions) can be calibrated before it is in-
stalled into a Position. If a Device is installed into a Posi-
tion, then only a Position calibration may be carried out,
not a Device calibration.
CMX supports the following methods of calibration data en-
try:
1. Off-line Calibration (sometimes also called Batch
Calibration), where Instruments/Functions to be
calibrated are sent from CMX to a Calibrator able
to communicate with CMX. Then the calibrator is
moved to the field/laboratory where the instrument
is located. After the calibration is performed the
calibrator is brought back and reconnected to
CMX. The results of the calibration are then re-
ceived to CMX.
For more information, see Selecting Instruments
for Calibration on page 110 and Receiving Cali-
bration Result Data on page 113.
2. Manual Entry is for situations, where a calibrator
unable to communicate with CMX is used for cali-
bration.
For more information, see Manual Entry of Cali-
bration Results on page 115.
Manual Entry supports multiple As Found and As Left test. In
Off-line Calibration, the amount of As Found and As Left re-
peats depends on the communicating calibrator. When in
doubt, refer to the calibrator manual.
Notes.
As Found Calibration
The calibrator is moved into the field, and an As Found cali-
bration is performed using the setup information in the
memory. The results are saved in the memory of the calibra-
tor.
As Left Calibration
The instrument function is calibrated again, to check that the
maximum error is at the desired level. The results are stored
in the calibrator memory, and another scheduled calibration
can be performed.
Printing of Results
CMX can then e.g. produce calibration certificates reports
based on saved calibration data.
About CWSI
Calibration Web Service Interface, CWSI, is a tool for com-
Availability of CWSI: munication between calibrators and CMX in virtualized envi-
–– CMX Light
ronments.
CMX Professional
CMX Enterprise CWSI introduces two new components into CMX: CWSI
Server and CWSI Client. CWSI Server is installed on a
server, e.g. the same server as CMX Server is installed.
CWSI Client is installed on all computers where CMX main
software is run. More info of CWSI installation is in CMX In-
stallation Manual and in CWSI Help file included in CWSI
Server and CWSI Client. The help file is also available as a
pdf file on CMX's installation disc.
Starting CWSI
Before using CWSI, CWSI Client must be started and con-
nected to CWSI Server.
In Windows 7 and 10, search for CWSI Client from the
menu opening from the Windows logo, typically found in the
lower left corner:
The context menu of the system tray icon allows you to con-
figure the connection to CWSI Server and connect to CWSI
Server.
Configuring CMX Server is done similarly, but on the com-
puter it resides in, provided you have the rights to make
changes to the computer. The configuration software is
called CMX Server Configurator.
Note.
If CWSI is configured to require a "Secret" (a password
unique to CWSI), The "Secret" has to be entered before
sending and receiving commences.
Notes.
Note.
When selecting a Position/Device and choosing Calibration,
Manual Entry, only the first (the uppermost) Function of the
Position/Device is sent to the Manual Entry window. This is
because CMX’s Manual Entry currently only supports the cal-
ibration of one Function at a time.
To select any other Function (as well as the first Function) in
a Position/Device for Manual Entry, click the secondary but-
ton on the mouse above the Function instead of the Posi-
tion/Device.
Reminder.
To be able to calibrate a Position, a Device (containing Func-
tions) needs to be linked to it. A Device (containing Func-
tions) may be calibrated although it is not linked to a Posi-
tion. When a Device is linked to a Position, Manual Entry
may be invoked for the Position only.
Notes.
For unstable readings, you may add the “~” character (tilde)
after or before the calibration point values, e.g. 12.045~. If
you enter the “~” character before the numbers, CMX moves
the character at the end of the number. This is done to avoid
mix-up between the minus sign and the “~” character.
Input values outside the calibration point's Max. Deviation
limit are indicated in red. This does however not affect any
calculations.
New Repeat
To start entering the results of the next calibration repeat,
press the New button in the Manual Entry window’s toolbar.
If you have entered any data into the Conditions tab, CMX
automatically copies them as part of the new repeat’s data.
Edit them, if needed.
To view previously entered results the click icons on the left
side of the Manual Entry window.
Fields:
Max Found Error is the repeat's found maximum error
value.
Most Significant Error (value) is the numeric value of
the found error that is closest to corresponding error limit.
Most Significant Error (percentage) indicates how close
to the calibration point's error limit the most significant er-
ror value is. The percentage value is less than or equal to
100 % if the calibration passed and more than 100 % if
the calibration failed.
When the relative component of the Reject if Error limit is used,
the Max Found Error may differ from Most Significant Error.
More of Error Limit Function's components in CMX's on-line
Help. Another source for information of Error Limit Function is in
CMX Calculations Help file. Open it from CMX Main Windows'
Help menu. Select option CMX Calculations.
Note.
Depending on your calibration settings, results may be save
"as both" (the results are saved both as As Found and as As
Left results). This setting is among the Calibration options
in the Options window.
Note.
If an input or output value ends in a “~” character (tilde), e.g.
12.045~, the reading was unstable. Adding this character is
possible in CMX's manual entry window, in CMX for Pocket
PC and when appropriate, the character is automatically
added in a reading when using certain modern calibrators.
Note.
Approving/signing calibration results is allowed only if you
have the appropriate user rights.
Note.
Deleting calibration results is allowed only if you have the
appropriate user rights.
Note.
For Weighing Instrument, the History Trend is available for
the actual Weighing Test only.
General
CMX allows the printing of the following document types:
Calibration Certificates,
Reports,
Labels and
History Trend graphs(* and
Audit Trail reports(*.
*) Not available in CMX Light.
To print documentation from CMX, use the Documents but-
ton found in the main window’s toolbar or select the docu-
ment type to be printed from the Documents Menu. A third
possibility is using the Context Menu. In all cases: A more
detailed print selection window will be open for that docu-
ment type.
Exception:
Audit Trail Reports are printed from the Audit Trail window.
More of this in chapter The Audit Trail Window.
Note.
Only Positions/Devices/Functions with saved calibration re-
sults may be dragged to the Print Certificate window.
Note.
One of the available reports is called Manual Entry Form.
Print it to have a practical form for manually entering calibra-
tion results.
Note.
Only Positions/Devices/Functions with saved calibration re-
sults may be dragged to the Print History Trend window.
Note.
If you receive a standalone document layout, sent by mail
from, e.g. Beamex, start from phase 4 in the list above. Just
replace the word "exported" with "standalone".
Exception:
Editing Audit Trail Reports is invoked in the Audit Trail win-
dow. More of this in section The Audit Trail Window. Also
note that you can have only one Audit Trail Report Layout for
each CMX database. You may however edit the existing one
to suit your needs, provided you have the rights and tools to
edit document layouts.
Important!
If you have activated Window's UAC (User Account Control)
feature, you'll need to run CMX as administrator to be able to
save the edited documents. Otherwise you will get an error
message and your edits are not saved.
General
Availability: Change Management and Audit Trail includes tools for keep-
–– CMX Light ing track of changes made to CMX's database. Additionally,
CMX Professional
whoever makes a change to the database may also be
CMX Enterprise
asked to give a description for the change.
Users with the proper authorization may review the change
log (Audit Trail) and mark the changes reviewed.
The feature also includes a possibility to utilize electronic
signatures
Briefly, CMX’s Change management feature includes all the
tools required to comply with regulations of which most im-
portant are GAMP4 (Good Automated Manufacturing Prac-
tices) and 21 CFR Part 11 (Electronic records; Electronic
signatures).
How to Change or check your Change Management and Au-
dit Trail settings is described in section Basic Settings,
chapter The Options Window.
Note.
If the same CMX database is contacted from several time
zones, consider setting the time format in CMX Options
windows' Change Management Options to suit your needs.
Electronic Record
Electronic Record is the name used for the group of data that
has been approved with the Electronic Signature. If anything
that is part of a certain Electronic Record is changed after it
was signed, the data requires a new approval using the Elec-
tronic Signature.
The following list describes what type of data belongs to the
available Electronic Records:
General
Availability: CMX's Pocket PC Interface consists of software installed into
–– CMX Light a Pocket PC and communication between CMX and the
CMX Professional
Pocket PC. Together they allow you to transfer data of in-
CMX Enterprise
struments to be calibrated to a Pocket PC, perform the cali-
bration wherever it is suitable for you, enter the calibration
data into the Pocket PC and then reconnect to CMX and up-
load the results to CMX's database.
Thus CMX for Pocket PC is a "portable manual entry tool"
Note.
Editing the language file requires knowledge of XML files and
how to edit them.
Note.
Before attempting to remove CMX for Pocket PC, ensure
that it is not running in the background. If you're unsure, do
as follows:
Open Pocket PC's Start menu,
select Settings and from the opened window,
select System tab.
Select Memory icon and from the opened window,
select Running Programs tab.
Now you are able to see a list of running programs. If CMX
for Pocket PC is on the list, select it and tap "Stop". Newer
Pocket PC operating systems may also have a shortcut for
opening the memory window directly from the desktop.
Starting
When CMX for Pocket PC is started for the first time or it is
not used for a while, the CMX for Pocket PC icon is not seen
in the Start menu. In that case, select Programs from the
Start menu and locate the CMX icon in the list of available
programs. CMX icon:
When CMX for Pocket PC has recently been used, the CMX
icon is available in the Start menu's list of recently used pro-
grams. Use the CMX icon to start CMX for Pocket PC.
When CMX for Pocket PC starts, a list of sent instruments is
shown.
Note.
If you want to simultaneously enter calibration results of mul-
tiple instruments, see chapter Calibrating Groups of In-
struments on page 146.
Note.
If the instrument is already calibrated, only the Environment
and Note data may be edited in Pocket PC.
When both the input and output values are edited according-
ly, either tap the [Next>] button or press the middle button in
a 5-way navigation button (if applicable).
The fields in the middle of the window as well as the graph
below the fields are updated. Also: The large numeric fields
have pre-entered values for the next calibration point.
Note.
Input values outside the calibration point's Max. Deviation
limit are indicated in red. This does however not affect any
calculations.
During the calibration run and even when viewing results, be-
fore the results are saved, you can correct any mistakes by
tapping the [<Prev] button until you see the point/data with
the erroneous data. Correct the data and tap the [Next>] but-
ton until you are back where you were.
Calibrator References
In the calibrator reference window you may enter calibrator
references for the calibration. If the Points field is set to All,
then the entered calibrator reference applies for all points. If
the Points field is set to a single point, e.g. 3/9, the entered
calibrator reference is valid from that point onwards either to
the last point or to the following point with a new calibrator
reference.
Note.
The Calibrator reference window may also be opened while
entering the calibration point values. Open the Tools menu
and select option Set Calibrator Reference.
Environment Data
You may enter environment at the end of the calibration or at
any time during the calibration too. Open the Tools menu
and select option Environment Information.
Note.
Environment data history is not saved. Whenever you rewrite
the environment data, they are the ones that are saved at the
end of calibration.
Notes.
To get back to "normal" Instrument List and calibrating a sin-
gle instrument at a time, uncheck Create Group.
Groups cannot be saved and only one group may be active
at a time.
Maximum amount of items in a group is 10.
Entering calibration
results is done mostly
as it is done for a sin-
gle instrument as de-
scribed in chapter En-
tering Calibration
Data and its subchap-
ters starting from page
144.
A row in the Calibra-
tion window indicates
which instrument is
calibrated in the Group
and at which calibra-
tion point you are.
Notes.
The Tools menu includes the option Cancel Calibration. It
removes the current instrument from the group.
Selecting Reject All cancels all open instruments and the
Group calibration ends.
Note.
If you have the Calibration Options field Update Calibrator
data from instruments received from Pocket PC checked, the
following happens when you receive a calibration that refers
to a calibrator that is in the Pocket PC's memory, but not in
CMX's database:
CMX opens a dialog asking whether the calibrator should be
added to CMX's database.
How different types of calibrators are received to CMX
database:
- A Weight Set is fully received, including specifications.
- A Beamex calibrator is received and specifications
are added to it from CMX's database.
- A non Beamex calibrator is received without
specifications since no specifications are available
in Pocket PC's memory.
General
Availability: CMX's Weighing Instrument Calibration feature includes a
–– CMX Light possibility to add Weight Sets and Weights as well as Weigh-
CMX Professional
ing Instruments to be calibrated into CMX’s database.
CMX Enterprise
Weighing Instruments can then be calibrated by using either
the Manual Entry window or by using the Pocket PC inter-
face.
CMX's Weighing Instrument Calibration feature is based on:
OIML international recommendation OIML R 76-1:
1992
(OIML = Organization Internationale de Métrologie
Légale or International Organization of Legal Me-
trology),
European standard EN45501:1992 + AC:1993,
NIST Handbook 44 – 2004 / 2.20. Scales
(NIST = The National Institute of Standards and
Technology, US),
EURAMET cg-18, Version 3.0 (03/2011)
Previously: EA-10/18 EA Guidelines on the calibration
of non-automatic weighing instruments (EA = Europe-
an co-operation for accreditation).
U.S. Pharmacopeia Convention’s Chapter 41 (brief-
ly USP 41).
This section presents Weighing Instrument Calibration spe-
cific information. Where the functionality of CMX’s Weighing
Instrument Calibration is similar to other type of instruments
and calibrators, this section refers to the more general
presentation elsewhere in this document.
When adding a new Weight Set, enter the set’s name as the
Model. Also, assure that you check the Weight Set check
box. Then CMX treats the new Calibrator Model as a Weight
Set, not a Process Instrument Calibrator.
Full field level information of Calibrator Model properties are
available in the on-line help delivered with CMX.
When ready, press the OK key to return to the Add New
Calibrator window.
Hint!
Enter the Weight Set name as the Serial Number. This “ties”
the weight to the set it belongs to. If you have several
weights with the same nominal weight, include the identifying
symbol (*) or number (1, 2 …) to the serial number.
Full field level information of Calibrator Module properties are
available in the on-line help delivered with CMX.
Hint.
You can also make you own Weighing Instrument Function
Templates and Calibration Procedures using the Function
Templates utility presented in Basic Settings section. CMX
automatically adds the Weighing Instrument specific fields if
the new Function’s Input/Output Quantity field is set to
“Mass”.
Field Description
Partial Weighing Rang- Select the amount of unique partial
es weighing ranges found in the
Weighing Instrument at hand. Re-
quired field for Weighing Instru-
ments.
*d1 … d4 Actual Scale Interval.
OIMLR76-1:
“the difference between the values
corresponding to two consecutive
scalemarks, for analogue indication,
or the difference between two
consecutive indicated values, for
digital indication.”
The “Partial Weighing Ranges” set-
ting determines the amount of Ac-
tual Scale Intervals seen.
Required field(s) for Weighing In-
struments.
e1 … e4 Verification Scale Interval. The
“Partial Weighing Ranges” setting
determines the amount of Verifica-
tion Scale Intervals seen.
Required field(s) for Weighing In-
struments.
Note.
Editing the Scale Intervals affect
the MPE values found among a
Weighing Instrument's Calibration
Procedure data, unless you are
using a custom Accuracy Class.
Scale Range 1 … Defines the Scale's Partial Weigh-
Scale Range 4 ing Range.
Visible only for Weighing Instru-
ments with more than one weighing
range.
Required field(s) for Weighing In-
struments.
Note.
Make sure both input and output methods are set to
“Keyed”.
Field Description
Accuracy Class A list of standard Weighing Instru-
ment accuracy classes and a possi-
bility to make a customized accura-
cy class. Available standard accura-
cy classes:
- HB 44 I
- HB 44 II
- HB 44 III
- HB 44 IIII
- HB 44 IIIIL
- OIML I
- OIML II
- OIML III
- OIML IIII
HB 44: NIST Handbook 44: ‘Specifica-
tions, Tolerances and Other Technical
Requirements for a Weighing and
Measuring Devices’. Used in USA.
OIML: ‘International Recommendation
OIML R 76-1’, 1992. Used in Europe.
Note.
Selecting a standard Accuracy
Class affects the MPE Range and
MPE fields. New MPE limits are
calculated based on the Weighing
Instrument's Verification Scale In-
terval data found among Function
data.
MPE Range Maximum Permissible Error
range(s) and their corresponding
Maximum Permissible Error (MPE).
The amount of MPE ranges de-
pends on the selected Accuracy
Class. Also note that selecting a
standard Accuracy Class automati-
cally defines the MPE Ranges. Edit
the MPE Ranges and MPE limits
only when using custom Accuracy
Classes.
Available error calculation methods:
% of accuracy class limit
% of load
absolute error
Hint.
You can also make you own Weighing Instrument Function
Templates and Calibration Procedures using the Function
Templates utility presented in Basic Settings section. CMX
automatically adds the Weighing Instrument specific fields if
the new Function’s Input/Output Quantity field is set to
“Mass”.
Terminology
This section presents some Weighing Instrument Calibration
related keywords. The focus is on the different types of cali-
brations that are part of Weighing Instrument Calibrations.
The descriptions are based on the standard mentioned in the
beginning of this section and “EA-10/18 EA Guidelines on
the calibration of non-automatic weighing instruments”
(EA = European co-operation for accreditation).
Eccentricity Test
The test consists of placing a test load in
different positions on the load receptor in
such a manner that the centre of gravity
of the load takes positions indicated in the
adjacent picture.
The test load should not be below 10 % of full scale. Other-
wise the uncertainty component of the eccentricity test may
be too large for uncertainties close to full scale.
Repeatability Test
The test searches the difference between the results of sev-
eral weighings of the same load. The conditions should, as
far as possible, be identical. This applies to the handling of
the load and the instrument as well as environmental condi-
tions.
U.S. Pharmacopeia Convention’s Chapter 41 (briefly USP
41), official in late 2013, defines a new procedure for doing
the repeatability and minimum weight tests. CMX supports
calculating both the old Repeatability and Minimum Weight
tests and the new test according to USP 41. Which one is
used is defined in Weighing Instrument's Calibration Proce-
dure.
Notes.
Immediately to the left of the field where the output reading is en-
tered is an area for indicating/saving that the reading was unsta-
ble. Tap the area and the "unstable" symbol ( ) becomes visible.
For Weighing Test: Input values outside the calibration
point's Max. Deviation limit are indicated in red. This does
however not affect any calculations.
For Eccentricity, Repeatability, and Minimum Weight test, CMX
remembers the selected weights. Thus when you press [Next>],
you do not need to re-select the same weight(s). In Weighing
Test CMX lets you choose the weights during upscale calibra-
tion. During downscale calibration CMX pre-selects the weights
used for corresponding points in upscale calibration.
For situations where a Tare Weight is entered: Enter the
true reading (including both the calibration weights and the
Tare Weight). CMX automatically subtracts the Tare Weight
from the reading before performing any error and uncertainty
calculations.
Remember to check
the Envir. tab for
entering the envi-
ronment data.
If automatic certifi-
cate number gen-
eration is disabled,
you are able to enter
the certificate num-
ber here.
Notes.
You can do one As Found test (AsF) and one As Left test
(AsL) for Eccentricity, Repeatability, and Weighing tests. You
may however do several As Found and As Left Minimum
Weight tests.
After the first As Left test is saved, you may only do As Left
tests.
General
Maintenance Inspection allows you to run check procedures
Availability: either according to IEC 60079-17 standard or by creating
–– CMX Light
custom check procedures. The checks and check procedure
CMX Professional
CMX Enterprise definitions are saved in CMX's database. When checks are
scheduled to be done, the checks are sent to a compatible
tablet with bmobiletm software. After the checks are done,
the results are returned to CMX and when needed forwarded
to an asset management software.
Beamex bmobiletm software is available at this address
(URL):
http://www.beamex.net/download/bMobile/
Note that downloading apps from unknown sources, i.e.
others than Google Play Store, must be enabled in your tab-
let.
Download the app and install it just like any Android App.
The tablet's requirements are:
Android tablet, operating system version 4.4 (Kit-
Kat), or later.
Recommended screen resolution: 1280 x 720 or
higher.
Recommended screen size: 8 inch or higher.
USB port, for communication with CMX.
Procedure
Each procedure contains a single Checklist. A Checklist con-
sists of an unbounded amount of Checklist items. The
Checklist items may be arranged to unbounded amount of
groups. The picture below presents a Checklist consisting of
three Checklist items in a single group.
Checklist Reports
Reports for checklists include instrument data as presented
in regular certificates. The checklist results are shown below
in table format.
Checklists
bmobile opens in Home View (Checklist). It presents all
Checklists sent from CMX. The icons in front of checklists
describe the status of the each checklist:
In Progress, i.e. currently going through the
checks that are part of this check list.
Open, which means this check list hasn't been run
through as yet.
Failed, i.e. the checklist has been run through and
master pass/fail status is "Failed".
Passed, i.e. the checklist has been run through
and the master pass/fail status is "Passed".
To run through a checklist, tap on it.
Note.
bmobile's Home View includes tools for sorting and filtering
checklists. Open bmobile's help file for detailed information
on the available options.
When the last point of the checklist run is done, the current
checklist window's Details tab is opened.
The Details tab allows you to select tester, add environmen-
tal information and enter master Pass or Fail status to the
test.
Notes.
Previously not existing testers cannot be added into bmo-
bile, but you can select one of the available testers sent from
CMX.
Working hours are entered for each tester in the pop-up win-
dow for adding testers. The Details tab presents the total
hours as read-only data.
General
Enhanced Work Order Handling is an option in CMX. To fully
Availability: utilize it, an ERP (Enterprise Resource Planning) or CMMS
–– CMX Light
(Computerized Maintenance Management System) software
CMX Professional
CMX Enterprise together with Beamex® Business Bridge software should
be part of the system. The latter does the communication be-
tween the ERP/CMMS and CMX.
Enhanced Work Order Handling eliminates re-typing calibra-
tion data/information to an ERP/CMMS software. Business
Bridge takes care of transferring information between CMX
and ERP/CMMS software.
Work Order Start Date and Work Order End Date are not
obligatory fields. They may also be empty fields.
If the start dates and/or end dates vary for a multiple instru-
ment work order, the earliest date is shown with a tilde (~) in
front to the date.
Notes.
Note.
The pop-up menu described in chapter Changes in CMX
User Interface, on page 182, allows you to change the
Work Order State, when necessary.
Emergency Calibration
An emergency calibration means that there is an acute need
to calibrate an instrument but there is no time or no means to
create a work order in the ERP/CMMS software first. CMX
supports emergency calibrations as follows:
Do the calibration normally and save the results in CMX's
database.
Afterwards, in ERP/CMMS software:
Create a Work Order Number and set the Work Order
Start Date to a date prior to the moment the emergency
calibration was performed. Send the work order to CMX.
A person with Calibration - Modify permission edits the
instrument's results to add the Work Order Number sent
from ERP/CMMS software to the calibrated instrument's
Calibration Result page.
Depending on Work Order State's update settings in
CMX, the state is either set to "Done" automatically or
CMX asks you whether to update the state or not.
When the Work Order State is "Done", Business Bridge
communicates the Work Order Number and the Work
Order State "Done" to the ERP/CMMS software.
This allows doing emergency calibrations without a pre-made
work order. However, the work order state will still be includ-
ed afterwards to ERP/CMMS software just as for "normal"
calibrations.
Note.
Emergency calibration of free devices is not supported.
Note.
While in Work Order Mode: Group calibration is possible for
instruments belonging to the same work order only. To do
group calibration for a group of instruments not being part of
the same work order, use "normal mode".
General
CMX Database Manager is a tool for doing backups of your
databases as well as restoring databases using a backup da-
tabase as the source. Additionally you can also in-
stall/uninstall, create and copy CMX databases.
Note that certain CMX Database Manager features are not
available in CMX Light.
Warning!
CMX Database Manager should be used by experienced
database and database server users. Using this tool
without knowing what you are doing may result in loss
of valuable data.
Note.
Start the CMX Database Manager in same computer where
the CMX database server is.
Server Connection
The upper part of CMX Database Manager’s window con-
tains server connection data. The data is pre-entered as
needed for CMX. If you log into another server, you need to
enter the server login and password accordingly. Click Log-
in, to log into the server.
Note.
CMX Database Manager includes a possibility utilize either
SQL Server Authentication (default) or Windows Authen-
tication. Use the Authentication selection list to choose the
one you want.
Notes.
Uninstalling a Database is not possible if the same database
is in use in CMX. Close CMX before starting the uninstall
procedure.
If you have entered your own data into either of the da-
tabases shipped with CMX, uninstalling that database
results in data loss!
Copying a Database
Availability: This feature allows you to copy a backup database and cre-
–– CMX Light ate a new CMX database of it.
CMX Professional
CMX Enterprise
Give the new database a name. Standard Windows® file
naming conventions apply, except that you do not need to
enter the file extension. Then click the Copy button.
See the end of chapter Creating a New Database for infor-
mation on how to take use of the copied (new) database.
Note.
There is no way of retrieving a deleted database. Perform
the delete only if you are sure that the database to be delet-
ed is no longer of use at all.
Moving a Database
Select the database to be moved and the destination server.
You may also rename the database to be moved.
Note.
The source and destination databases have to be in the
same computer when moving. If not, use CMX Database
Manager's Backup and Restore tools.
General
This section includes additional information not related to
CMX alone, but useful when working with calibrators com-
municating with CMX.
Note.
If absolute pressure is always measured using the same pair
of standard accuracy and high accuracy modules, consider
editing the absolute pressure measuring range specifications
for that particular gauge module. The field in question is
Constant Error. More of specifications in chapters Adding
Module Ranges and Calibrator Module Range Fields.
When performing this customization, you should record the
change for future needs. If you later on purchase more pres-
sure modules / calibrators and maybe use a different combi-
nation to measure absolute pressure, you need to assure the
specifications are valid for that combination.
Component Description
Display Accuracy The specification of the internal
measurement.
Stability The instability of the block's temper-
ature
Axial Uniformity The vertical temperature gradient in
the block
Radial Uniformity The horizontal temperature gradient
in the block
Load Effect Measurement probe(s) leaking heat
to/from the block to the ambient
Hysteresis The variation in readouts when
reaching a certain temperature from
highter or lower temperatures
Reference Connection From reading the reference sensor
connected to the Temperature Block
Notes.
Not all components are valid in all applications using a Tem-
perature Block.
Detailed specification data is in the Temperature Block's Us-
er Guide.
Method 1:
Temperature Block With Internal Temperature
Measurement
A Temperature Block is used for creating the necessary tem-
perature and the Temperature Block's internal temperature
sensor is used for measuring the Block's temperature. The
Input Method setting of the instrument to be calibrated is
“Controlled/Measured” and no external reference sensor is
connected to the calibrator or the Temperature Block.
The following specification components of the Temperature
Block are combined in CMX's Constant Error and Relative
Error data:
Display Accuracy
Stability
Axial Uniformity
Radial Uniformity
Load Effect
Hysteresis
Method 2:
Reference Sensor Connected to R-type Temperature
Block
An R-type Temperature Block controls the temperature and a
reference sensor connected to the Temperature Block is
used for measuring the temperature. The Input Method set-
ting of the instrument to be calibrated is “Con-
trolled/Measured”.
In this case, the following specification components of the R-
type Temperature Block are combined in CMX's Constant Er-
ror and Relative Error data:
Stability
Axial Uniformity
Radial Uniformity
Load Effect
Reference Connection
Note.
For Beamex smart sensors, the specifications are already
available in CMX. For other sensors and devices, the model
and its specifications need to be added to CMX's database
prior to using them in calibrations.
Method 3:
Reference Sensor Connected to an External Device
Temperature Block controls the temperature and an external
reference sensor is in the block, but the sensor is connected
to an external temperature measuring device. The Input
Method setting of the instrument to be calibrated is “Con-
trolled”.
In this case, the following specification components of the
Temperature Block are combined in CMX's Constant Error
and Relative Error data:
Stability
Axial Uniformity
Radial Uniformity
Load Effect
Where:
utb is the standard uncertainty of a Temperature
Block.
DA is the Display Accuracy specification.
Zero when using methods 2 and 3.
St is the Stability specification.
AU is the Axial Uniformity specification.
RU is the Radial Uniformity specification.
LE is the Load Effect specification.
Hy is the Hysteresis specification.
Zero when using methods 2 and 3.
RC is the Reference Connection specification.
Zero when using methods 1 and 3.
Notes.
All specifications have a rectangular probability distribution.
That is why they are divided by the square root of three.
In CMX, the uncertainties are saved as expanded uncertain-
ties, i.e. the standard uncertainties are multiplied by two.
General
party libraries included in CMX, CWSI Server
CMX, CWSI Server and CWSI Client include a
and CWSI Client is provided further on. By ac-
number of third-party libraries that are used to
cepting the terms of the Beamex software li-
provide certain features. Additional copyright and
cense agreement, You also accept the third par-
legal notices and license terms applicable to the
ty terms.
third-party libraries may apply. A list of the third-
WebSocket4Net
Copyright(C) 2014 Kerry Jiang. Unless required by applicable law or agreed to in
writing, software distributed under the License is
Licensed under the Apache License, Version 2.0
distributed on an "AS IS" BASIS, WITHOUT
(the "License"); you may not use this file except
WARRANTIES OR CONDITIONS OF ANY
in compliance with the License. You may obtain
KIND, either express or implied. See the License
a copy of the License at
for the specific language governing permissions
http://www.apache.org/licenses/LICENSE-2.0 and limitations under the License
log4net
Copyright(C) 2004-2013 Apache Software Unless required by applicable law or agreed to in
Foundation. writing, software distributed under the License is
distributed on an "AS IS" BASIS, WITHOUT
Licensed under the Apache License, Version 2.0
WARRANTIES OR CONDITIONS OF ANY
(the "License"); you may not use this file except
KIND, either express or implied. See the License
in compliance with the License. You may obtain
for the specific language governing permissions
a copy of the License at
and limitations under the License.
http://www.apache.org/licenses/LICENSE-2.0
SecondLanguage
Copyright(C) 2013, 2015 James F. Bellinger. 1. The origin of this software must not be mis-
represented; you must not claim that you
This software is provided 'as-is', without any ex-
wrote the original software. If you use this
press or implied warranty. In no event will the
software in a product, an acknowledgement
authors be held liable for any damages arising
in the product documentation would be ap-
from the use of this software.
preciated but is not required.
Permission is granted to anyone to use this
2. Altered source versions must be plainly
software for any purpose, including commercial
marked as such, and must not be misrepre-
applications, and to alter it and redistribute it
sented as being the original software.
freely, subject to the following restrictions:
3. This notice may not be removed or altered
from any source distribution.
LICENSE - This product (the “Program Product”) the additional microcomputers (or additional us-
ers on a network) are purchased.
is only licensed, not sold. It remains the property
of Beamex Oy Ab (“Beamex”), which grants a CONCURRENT LICENSE (NETWORK) - You
non-exclusive, non-transferable license to the may use the Program Product in accordance
individual and/or corporate licensee to use the with this Agreement on a licensed local area
enclosed Program Product. The Program Prod-
network. A computer network is any combination
uct includes all computer programs, machine- of two or more terminals that are electronically
readable and printed materials as provided by linked and capable of sharing the use of a single
Beamex in accordance with the terms and condi-
Program Product. The Program Product is deliv-
tions set forth in this Software License Agree- ered with purchased amount of licenses. You
ment. The Program Product includes all codes,
may not use the Program Product on more com-
techniques, software tools, formats, designs, puters or terminals than the number licensed, as
concepts, methods and ideas associated with shown on your invoice or other documents from
this Program Product, and any updates which
Beamex. You may install the Program Product
Beamex may provide to you and backup copies on computers attached to a network, or remove
which you may make. it from one computer and install it on another
TERM - This license is effective from the date of one, provided that there is no possibility of simul-
taneous use of the Program Product on more
receipt of this Program Product and shall remain
in force until terminated. You may terminate this computers than the number licensed.
Agreement at any time by returning all contents COPYRIGHT AND RESTRICTIONS - The Pro-
of the Program Product and destroying any por- gram Product, including documentation, is copy-
tions of the Beamex Program Product merged in
righted under the laws of Finland and interna-
any form. This Agreement is automatically termi- tional treaty provisions. Except as specifically
nated if you violate any provision of this Agree- authorized in the product documentation, the
ment. Upon termination you must return all con-
copying, disassembly, or duplication of any part
tents of the Program Product and documentation of the Program Product or documentation, shall
together with all copies, and destroy any portions constitute a violation of copyright law. According-
of Program Product merged in any form. No re-
ly, except as expressly otherwise provided (in
fund is given unless expressively otherwise the Instruction Manual or Quick Guide or other-
agreed in writing by Beamex.
wise), you may not copy, duplicate, modify,
USE OF PROGRAM PRODUCT- This Agree- translate, disassemble or decompile any part of
ment limits the number of simultaneous users the Program Product or documentation, or cre-
who may use the Program Product. The use of ate or attempt to create, by reverse engineering
or otherwise, the source code from the object
the Program Product is limited to the building or
plant area where or to the company to which it is code of the Program Product supplied hereun-
licensed. The limitations depend on whether you der. Further, you may not reproduce, sublicense,
distribute, loan, rent, transfer or otherwise dis-
purchased an individual copy of the program or a
network license. close the Program Product or documentation to
others, in whole or in part, or remove or alter any
INDIVIDUAL LICENSE (LOCAL) - You may use ownership or copyright notices on the cd (s),
the Program Product in accordance with this diskette(s) or the documentation except as oth-
Agreement on a single microcomputer at any erwise expressly permitted by Beamex.
one time. The Program Product may be placed
on a network as long as there is no possibility of OWNERSHIP - Ownership of the Program Prod-
it being used simultaneously by more than one uct (including all adaptations), documentation
and all copies thereof and all copyrights, trade
user at one location. You may not use the Pro-
gram Product on more than one microcomputer secrets, know-how, trademarks, trade names
at any given time unless additional licenses for and all other proprietary rights and interests re-
INFRINGEMENT - Beamex does not warrant BEAMEX DOES NOT AND CAN NOT
that the Program Product will not infringe any WARRANT THAT THE PERFORMANCE OF
copyright, patent or any other rights owned or THE PROGRAM PRODUCT OR THE
possessed by any third parties and Beamex FUNCTIONS CONTAINED IN THE PROGRAM
shall not be required to protect, indemnify, or PRODUCT WILL MEET YOUR
hold you harmless against, or be liable to you REQUIREMENTS, OR THAT THE PROGRAM
for, any liabilities, losses, expenses, or damages PRODUCT WILL OPERATE IN THE
which may be suffered or incurred by you as a COMBINATIONS WHICH MAY HAVE BEEN
result of such infringement or allegation thereof SELECTED FOR USE BY YOU, OR THAT THE
by any third party, Beamex shall not be under OPERATION OF THE PROGRAM PRODUCT
any obligation to defend or participate in the de- WILL BE UNINTERRUPTED OR ERROR-
fense by you against any claim or suit alleging FREE, OR THAT DEFECTS IN THE PROGRAM
such infringement. PRODUCT WILL BE CORRECTED.
Beamex warrants that the media on which the FURTHERMORE, BEAMEX DOES NOT
Program Product is furnished will be free from WARRANT OR MAKE ANY
defects in materials and workmanship under REPRESENTATIONS REGARDING THE USE
normal use for a period of ninety (90) days from OR THE RESULTS OF THE USE OF THE
the date of delivery. Beamex’s only obligation PROGRAM PRODUCT OR DOCUMENTATION
under this warranty is to, at its sole discretion, IN TERMS OF THEIR CORRECTNESS,
either to replace or to repair the defective media. ACCURACY, RELIABILITY, OR OTHERWISE,
NO ORAL OR WRITTEN INFORMATION OR
THE LIMITED WARRANTIES SET FORTH ADVICE GIVEN BY BEAMEX OR AN
HEREIN SHALL NOT BE VALID IF (A) THE AUTHORIZED REPRESENTATIVE OF
PROGRAM PRODUCT HAS BEEN SUBJECT BEAMEX SHALL CREATE A WARRANTY OR
TO UNAUTHORIZED ALTERATION OR IN ANY WAY INCREASE THE SCOPE OF THIS
REPAIR OR MODIFICATION NOT EXPRESSLY WARRANTY. SHOULD THE PROGRAM
CONSENTED BY BEAMEX IN WRITING (e.g. in PRODUCT PROVE DEFECTIVE, YOU (AND
the Instruction Manual or Quick Guide) OR NOT BEAMEX OR AN AUTHORIZED
ABUSE, MISUSE, ACCIDENT OR NEGLECT; REPRESENTATIVE OF BEAMEX) ASSUME
OR (B) THE PROGRAM PRODUCT HAS BEEN THE ENTIRE COST OF ALL NECESSARY
EXPOSED TO CONDITIONS BEYOND SERVICING, REPAIR OR CORRECTION,
BEAMEX'S ENVIRONMENTAL, POWER OR EXCEPT AS EXPRESSLY STATED IN THIS
OPERATING CONSTRAINTS. AGREEMENT.
DISCLAIMER OF WARRANTY - The Program LIMITATION OF LIABILITY - UNDER NO
Product and documentation are intended for pro- CIRCUMSTANCES SHALL EITHER BEAMEX
fessional use by properly trained personnel only. NOR ANYONE ELSE WHO HAS BEEN
As the Program Product may be used for various INVOLVED IN THE CREATION, PRODUCTION,
purposes unknown to Beamex you (and not OR DELIVERY OF THIS PROGRAM PRODUCT
Beamex or any authorized representative of BE LIABLE FOR ANY INCIDENTAL,
Beamex) assume responsibility for the selection CONSEQUENTIAL, SPECIAL, DIRECT OR
of the Program Product to achieve your intended INDIRECT DAMAGES, INCLUDING, BUT NOT
results, and for the installation, use and results LIMITED TO, LOSS OF ANTICIPATED
obtained from the Program Product. You must PROFITS OR BENEFITS, LOSS OF USE OF
always verify proper the performance of the Pro- THE PROGRAM PRODUCT, LOSS OF DATA,
gram Product prior to any particular use. COSTS OF RECREATING LOST DATA
AND/OR THE COST OF ANY SUBSTITUTE
Therefore, you expressly acknowledge and EQUIPMENT OR PROGRAM THAT RESULT
agree that THE ENTIRE RISK AS TO THE USE, FROM THE USE OR INABILITY TO USE THE
RESULTS AND PERFORMANCE OF THE PROGRAM PRODUCT OR DOCUMENTATION,
IMPORTANT—READ CAREFULLY: This End- age device, such as a network server, used only
User License Agreement (“EULA”) is a legal to install or run the Software on computers used
agreement between you (either an individual or a by licensed end users in accordance with Sec-
single entity) and Microsoft Corporation (“Mi- tion 1.1. A single license for the Software may
crosoft”) for the Microsoft software that accom- not be shared or used concurrently by multiple
panies this EULA, which includes computer end users.
software and may include associated media,
1.4 Visual Studio—Effect of EULA. As a
printed materials, “online” or electronic docu-
suite of development tools and other Microsoft
mentation, and Internet-based services (“Soft-
software programs (each such tool or software
ware”). An amendment or addendum to this
program, a “Component”), Components that you
EULA may accompany the Software. YOU
receive as part of the Software may include a
AGREE TO BE BOUND BY THE TERMS OF
separate end-user license agreement (each, a
THIS EULA BY INSTALLING, COPYING, OR
“Component EULA”). Except as provided in
OTHERWISE USING THE SOFTWARE. IF YOU
Section 4 (“Prerelease Code”), in the event of
DO NOT AGREE, DO NOT INSTALL, COPY,
inconsistencies between this EULA and any
OR USE THE SOFTWARE; YOU MAY RETURN
Component EULA, the terms of this EULA shall
IT TO YOUR PLACE OF PURCHASE (IF
control. The Software may also contain
APPLICABLE) FOR A FULL REFUND.
third-party software programs. Any such soft-
ware is provided for your use as a convenience
and your use is subject to the terms and condi-
MICROSOFT SOFTWARE LICENSE
tions of any license agreement contained in that
1. GRANTS OF LICENSE. Microsoft grants software.
you the rights described in this EULA provided
2. ADDITIONAL LICENSE RIGHTS --
that you comply with all terms and conditions of
REDISTRIBUTABLE CODE. In addition to the
this EULA. NOTE: Microsoft is not licensing to
rights granted in Section 1, certain portions of
you any rights with respect to Crystal Reports for
the Software, as described in this Section 2, are
Microsoft Visual Studio .NET; your use of Crystal
provided to you with additional license rights.
Reports for Microsoft Visual Studio .NET is sub-
These additional license rights are conditioned
ject to your acceptance of the terms and condi-
upon your compliance with the distribution re-
tions of the enclosed (hard copy) end user li-
quirements and license limitations described in
cense agreement from Crystal Decisions for that
Section 3.
product.
2.1 Sample Code. Microsoft grants you a
1.1 General License Grant. Microsoft grants
limited, nonexclusive, royalty-free license to: (a)
to you as an individual, a personal, nonexclusive
use and modify the source code version of those
license to use the Software, and to make and
portions of the Software identified as “Samples”
use copies of the Software for the purposes of
in REDIST.TXT or elsewhere in the Software
designing, developing, testing, and demonstrat-
(“Sample Code”) for the sole purposes of design-
ing your software product(s), provided that you
ing, developing, and testing your software prod-
are the only individual using the Software.
uct(s), and (b) reproduce and distribute the
If you are an entity, Microsoft grants to Sample Code, along with any modifications
you a personal, nonexclusive license to use the thereof, in object and/or source code form. For
Software, and to make and use copies of the applicable redistribution requirements for Sam-
Software, provided that for each individual using ple Code, see Section 3.1 below.
the Software within your organization, you have
2.2 Redistributable Code—General. Mi-
acquired a separate and valid license for each
crosoft grants you a limited, nonexclusive, royal-
such individual.
ty-free license to reproduce and distribute the
1.2 Documentation. You may make and use object code form of any portion of the Software
an unlimited number of copies of any documen- listed in REDIST.TXT (“Redistributable Code”).
tation, provided that such copies shall be used For general redistribution requirements for Re-
only for personal purposes and are not to be re- distributable Code, see Section 3.1 below.
published or distributed (either in hard copy or
2.3 Redistributable Code—Microsoft Merge
electronic form) beyond your premises.
Modules (“MSM”). Microsoft grants you a lim-
1.3 Storage/Network Use. You may also ited, nonexclusive, royalty-free license to repro-
store or install a copy of the Software on a stor- duce and distribute the content of MSM file(s)
Availability
Symbols used for a feature's availability in different CMX
versions:
–– Not available
Optionally available
Standard feature
Business Bridge
Beamex® Business Bridge is a software that communicates
between CMX and ERP/CMMS Asset Management soft-
ware. When purchased, Beamex's support helps customizing
Business Bridge to work between CMX and the Asset Man-
agement software in use.
Calibration Points
Calibration Points are the input signal values (sometimes the
output signal values) used in the calibration of a function.
Calibration Procedure
The procedure includes information on how the calibration
should be accomplished, like:
- Can the calibration be performed automatically or is it done
manually
- Can the calibration be carried out on the field or not.
- Recommended calibrator(s) to be used in the calibration.
ERP/CMMS
ERP = Enterprise Resource Planning
CMMS = Computerized Maintenance Management System
Both are asset management software.
Function
A single task done with an instrument, e.g. temperature
transmitter, pressure indicator etc. Instruments capable of
performing several tasks are called multifunction instru-
ments.
Function Type is the description of the function (input/output
quantities and ranges) and information concerning the cali-
bration of the function (e.g. calibration points and error lim-
its).
See also Multi-function.
Input Method
How the input signal of the instrument is obtained into CMX.
Examples of Input Methods:
Measured by the calibrator, Sourced by the calibrator, Con-
trolled by a pressure controller and measured by the calibra-
tor etc.
Instrument
For CMX, any Position seen on the main windows’ Positions
Tree are instruments. No matter what kind of function the in-
strument contains.
Instrument Collection
A collection of instruments located in the same database.
Output Method
How the output signal of the instrument is obtained into
CMX.
Position
A Position is the location in a process where a device can be
installed. Positions often have symbolic names (tags) like
TTI 215 in instrumentation drawings.
A Position can be calibrated only when a device is installed
because a location cannot be calibrated.
Secondary Button
The mouse’s secondary button is the rightmost one, if you
are using the mouse with your right hand.
Sets
Sets are groups of Device's or Position's Functions that the
user may create with the Set Maintain utility. Usually Devices
or Positions belonging to a set have something in common,
e.g. they are all calibrated during the same week, they are all
located in the same unit of plant etc.
How the sets are devised depend on the user application re-
quirements.
User ID
An abbreviation for identifying each user of CMX. Needed,
e.g. when starting CMX. Using the same User IDs as is used
on your network login is recommended.
A
About CMX ........................................... 1 Automatic Check Out /
Accessibility .................................. 31, 51 Check In ........................................... 101
Activating Users ................................. 46 Average Calculation ......................... 121
Activation/Deactivation ....................... 33
Adding B
Calibration Procedure
Templates .............................................. 70 Backing up a CMX
Calibration Procedures ............... 75, 76, 84
database .......................................... 188
Calibrators .............................................. 58
Devices ............................................ 75, 78
Basic Settings .................................... 39
Function Templates ................................ 70 Calibrators ............................................. 57
Functions.................................... 75, 76, 79 Change Management ............................ 43
Instruments ...................................... 75, 77 CMX Options ......................................... 40
Items to a Set ......................................... 98 Device Manufacturers ............................ 54
List Items ................................................ 68 Function Templates ............................... 69
Manufacturers .................................. 54, 63 LDAP ..................................................... 42
Models........................................ 55, 64, 65 Lists ....................................................... 67
Module Ranges ...................................... 60 Passwords ............................................. 46
Modules.................................................. 59 Plant Structure ....................................... 52
Plant Structure Nodes ............................ 52 Security .................................................. 42
Positions........................................... 75, 77 Supervisors ............................................ 45
User Groups .......................................... 47
Additional Information....................... 193
User Interface ........................................ 72
Absolute Pressure
Users ............................................... 44, 45
Measurement ....................................... 197
Weighing Instrument
Temperature Blocks ............................. 199
Calibration............................................ 152
Adjusting .................................... 10, 107
Business Bridge 179, 180, 183, 184, 185
Administrator .....................42, 44, 45, 47
Approving Results ............................ 121
As Found calibration................... 10, 106
As Left calibration ....................... 10, 107
Assigning a Spare Device .................. 89
Audit Trail ................................. 131, 132
Time Format ......................................... 132
Authorization ................................ 45, 47
W
Weighing Instrument
Calibration Procedure
Fields.................................................... 160
Function Fields ..................................... 159
Minimum Weighing
Capability Test ..................................... 166
Minimum Weight Test .......................... 166
Weighing Instrument
Calibration ........................................ 151
Basic Settings ...................................... 152
Eccentricity Test ................................... 166
Function Templates .............................. 158
Linearity Test ........................................ 166
Manual Entry ........................................ 168
Minimum Sample Weight ..................... 166
Repeatability Test ................................ 166
Saving Results ..................................... 172
Sending ................................................ 168
Weight Sets .......................................... 153
Weight Specifications ........................... 157
Weights ................................................ 155