Method Validation Protocol Genial Capsules 150mg
Method Validation Protocol Genial Capsules 150mg
Page 1 of 9
Hygeia Pharmaceuticals
Quality Control Department
Document No.HP/MVP/001
Method Validation Protocol
1.0 Objective:
Analytical monitoring of a pharmaceutical product is necessary to ensure its efficacy through out all
phases of its shelf life; such monitoring is in accordance with the specifications elaborated during product
development. Method validation ensures that the selective method will give reproducible, reliable and
consistent results adequate for intended purpose, it is therefore necessary to define precisely both the
conditions in which the procedure is to be used and the purpose for which it is intended.
The objective of this validation study is to
Ensure and justify, through extensive testing, that the precision and consistency are in accordance with
already established Acceptance Criteria.
Assess the effect of variables (within already set operating limits) on the testing method
Identify and solve the problem(s), if any, encountered during testing.
Establish confidence on existing testing process.
Find ways and means to increase productivity and improve quality.
Assure that complete process is under control.
2.0 Scope:
Validation of testing method, being used for the quantitative determination (Assay) of Fluconazole in
Genial Capsules 150mg.
Page 2 of 9
Hygeia Pharmaceuticals
Quality Control Department
Document No.HP/MVP/001
Method Validation Protocol
In case there is no change or modification in the Validated Method, the Revalidation will be performed
after every 5 years.
Any modification or changes in the Validated Test Method HP/QC/SAP-0013being used for the
determination of Fluconazole in Genial Capsules 150mg should be controlled and will be entered into the
change control form in accordance with the change control procedure.
For changes in existing method and proposing new method or proposing procedure for new product,
following data should be established.
6.1 Specificity:
Ability to assess unequivocally the analyte in the presence of components which may be expected to be
present (impurities, degradants, matrix etc.)
The linearity of an analytical procedure is its ability to produce results that are directly or indirectly
proportional to the concentration of the analyte in the sample within a given range, response to be linear
on at least 5-6 points.
Range is the lowest or highest level of analyte that the method can determine with reasonable accuracy
and precision in the range of 80%, 90%, 100%, 110% and 120 % of claim.
Linearity and range for an analytical procedure can be determine either by
i. Use of reference standard/working standard.
ii. Analysis in duplicate
iii. Determination by least square method/ regression line, slope of curve
iv. Response-concentration curve (Beer’s Law)
Page 3 of 9
Hygeia Pharmaceuticals
Quality Control Department
Document No.HP/MVP/001
Method Validation Protocol
6.3 Accuracy:
It relates to the closeness of the test results to true value i.e. measure of exactness of analytical method. It
is expressed as % recovery by the assay of known/added amount of analyte in the linearity range. The
accuracy of the test procedure can usually be determined by applying the procedure to the quantitatively
prepared samples of the material to be analyzed. Such samples also include to which analyte have been
added in quantities both above and below the expected range of validation. The usual range being 10 %
above or below the expected range of claim.
6.4 Precision:
The precision of the analytical method is the degree of agreement among individual test results and how
individual test results are scattered from the mean value usually expressed as Standard Deviation or as the
Relative Standard Deviation (Coefficient of Variation). These results are based when the complete
procedure is applied repeatedly to multiple sampling from the same homogenous bulk.
The precision is usually expressed as Standard Deviation or Relative Standard Deviation and is determine
by assaying a sufficient number of portions of homogenous samples and to be subjected to statistical
evaluation. The assays for such cases are independent analysis of samples that have been carried out by
applying the complete analytical procedure from sample preparation to final test results. Precision may
also be explained by the following terms,
i. Repeatability
ii. Intermediate Precision
iii. Reproducibility.
Page 4 of 9
Hygeia Pharmaceuticals
Quality Control Department
Document No.HP/MVP/001
Method Validation Protocol
Precision of the method when repeated by the same analyst, same test method and under same set of
laboratory conditions (reagents, equipment etc.) within a short interval of time, the only difference being
the sample. The repeatability of any test procedure is required to be assured by carrying out complete
separate determination on separate sample of the same homogenous batch under normal laboratory
conditions.
When the procedure is carried out by different analysts, in different laboratories using different
equipment, reagents by using the sample from the same homogenous batch in each case. Evaluate the
results obtained from the variability of analyst–to–analyst, day-to-day, laboratory-to-laboratory,
equipment-to-equipment etc., this analytical data will provide information about the reproducibility of the
test procedure under validation.
6.5 Robustness:
Robustness is the measure of the capacity of an analytical method to remain unaffected by small but
deliberate variation in procedure. It provides an indication about variability of the method during normal
laboratory conditions.
6.6 Ruggedness:
Ruggedness is the degree of reproducibility of test result obtained by analyzing the same sample under
variety of normal test conditions such as different
i. Analysts
ii. Instruments/ Equipment
iii. Days
iv. Reagents etc.
Page 5 of 9
Hygeia Pharmaceuticals
Quality Control Department
Document No.HP/MVP/001
Method Validation Protocol
X = ∑x
n
δ = n ∑ x² - (∑ x) ²
n(n – 1)
Page 6 of 9
Hygeia Pharmaceuticals
Quality Control Department
Document No.HP/MVP/001
Method Validation Protocol
Linearity and Range Solutions of 5 different concentrations (80%, 90%, Response should
100%, 110% and 120%) should be used for linearity, be linear.
one or two concentrations below and above linearity
will cover the range.
Accuracy and recovery Minimum nine determinations over a minimum of three RSD ≤ 2%
concentration levels covering the specified range
Intermediate Precision
Standard and sample solutions of concentration (100%)
(Ruggedness)
should be scanned with different analysts.
Page 7 of 9
Hygeia Pharmaceuticals
Quality Control Department
Document No.HP/MVP/001
Method Validation Protocol
Description:
Size “2” hard gelatin shell Yellow Body and maroon color cap containing white powder.
Composition:
Each capsule contains:
Fluconazole ------------------150mg
Limit (Corresponding to 90.0 – 110.0% of the label amount).
Procedure: Take reading of both sample and ref solutions in a 1 cm cell on a U.V/ Visible
spectrophotometer at 261 nm
Calculation :
% Age = abs.test × dilution of std × 100 x 190
abs.std ×dilution of sample × 150
Page 8 of 9
Hygeia Pharmaceuticals
Quality Control Department
Document No.HP/MVP/001
Method Validation Protocol
11.1 Facilities:
The validation of the method of Determination of Fluconazole in Genial Capsules 150mg will be carried
out in the Chemical Lab. Hygeia Pharmaceuticals ISLAMABAD, PAKISTAN.
Page 9 of 9