TC9600-G/ TC9600/TC6000-G/TC6000
Operation Manual
Ver.1.4
Labnet International, Inc.
PCR Disclaimer:
MultiGene Thermal Cyclers can be used for a variety of molecular biology applications
and techniques, including the Polymerase Chain Reaction (PCR). PCR and other
molecular biology processes and techniques may be protected by patents. Purchase of
this product does not convey a license or right to perform patented PCR processes.
Applied Biosystems holds some of the basic PCR patents that may or may not be
currently valid. For additional information on PCR patents held by Apllied Biosystems
and information about acquiring any required licensing for using PCR techniques, please
contact Applied Biosystems at 800-327-3002 or 650-638-5800 or their website:
www.appliedbiosystems.com .
Table of Contents
1. SAFETY PRECAUTIONS........................................................................................1
3. GETTING STARTED................................................................................................6
3.1 UNPACKING.................................................................................................................6
3.2 INITIAL OPERATION .....................................................................................................6
3.3 HEATED LID ................................................................................................................6
3.4 LID OPENING/CLOSING ...............................................................................................7
3.5 FUNCTION MAP .........................................................................................................10
3.6 KEY ENTRY SYSTEM ................................................................................................. 11
3.7 LOADING THE REACTION VESSEL ............................................................................. 11
5. PROTOCOL MANAGEMENT..............................................................................15
5.1 RUNNING A PROTOCOL ..............................................................................................15
5.2 PAUSE A PROTOCOL ...................................................................................................16
5.3 VIEWING RUNNING PROTOCOL INFORMATION...........................................................16
5.4 TERMINATING A PROTOCOL .......................................................................................17
7. EDITING(VIEWING) A PROTOCOL..................................................................23
1. Safety Precautions
Before using the MultiGene Gradient thermal cycler for the first time, please read this
entire operating manual carefully. To guarantee problem free, safe operation of the
MultiGene thermal cycler, it is essential to observe the following points.
1. Do not use the machine in a potentially explosive environment or with potentially
explosive chemicals.
2. Install the machine in location free of excessive dust.
3. Avoid placing the machine in direct sunlight.
4. Choose a flat, stable surface capable of withstanding the weight of the machine.
5. Install the machine in the room temperature 15 ~ 30°C, relative humidity 20
~80%.
6. DO NOT block the air vents.
7. Keep the side and rear of the machine at least 30cm from the wall or other
machine.
8. Make sure the power source conforms to the required power supply specifications.
9. To avoid electric shock, make sure the machine is plugged into a grounded
electrical outlet.
10. Do not allow water or any foreign objects to enter the various openings of the
machine.
11. Switch off the machine before cleaning or performing service on the machine,
such as replacing the fuses.
12. Repairs should be carried out by authorized service personnel only.
Warning Label: Please be aware of the danger of electric shock or other dangers.
1
2. General Description
The MultiGene Gradient thermal cycler is a powerful, reliable and affordable thermal
cycler. It is equipped with a gradient function that enables the user to set a maximum
24 °C temperature gradient between the left and right rows in the sample block at each
temperature.
2.1 Features
Easy-to-operate slide and lever type lid which prevents accidental skin burn
Large LCD panel enhances visibility and ease-of-operation.
Graphic display makes programming quick and simple.
Simple and easy to use graphical interface.
Extended Temperature return function that makes touchdown possible
Gradient function enables optimization of annealing temperatures.
Preset protocols that make starting amplification easier.
Air intake vents on the front and sides and exhaust vents on the rear that reduce
heat-based interference with other equipment.
Interchangeable blocks.
Automatically creates operating history and error logs.
Heated lid to prevent the formation of condensation.
Robust and attractive design.
2
MultiGene Thermal Cycler
Name Function
Function Keys Use the function keys (F1 to F5) to activate the “button” displayed on
(F1 to F5) the screen above the key; for example, F1 for RUN.
Directional The four Directional Arrow keys can move the cursor into different
Arrow Keys directions.
ENTER key The current selection is activated or accepted when the ENTER key is
pressed.
Alphanumeric The alphanumeric keys are usually used as number keys for the entry
Keys (0~9) of time, temperature, password and cycle numbers. When entering a
user name or protocol name, a small screen will pop out to display the
letters after pressing the Alphanumeric keys. Press the selected key
until the desired letter is highlighted. Press the ENTER key to select
the letter or wait 5 seconds.
CLEAR key Press CLEAR key will erase the entry in an input field.
(CE)
STOP key By pressing STOP, the Protocol current running will be terminated.
3
Figure 2. Front View
Name Function
Lid The lid with heating pad is designed to protect against evaporation
resulting from heating of the reaction tube and reaction plate.
Lid Knob Rotate Lid Knob in clockwise will lower the heating plate to place
pressure on the cap of tubes. This forces the tubes to sit firmly in the
block for better contact. Rotate Lid Knob in counter-clockwise
direction until the Lid pops up. Only after the Lid pops, can users
slide the Lid to the rear.
Sample Block Section of the unit in which sample tubes and plates are placed.
Key-Pad Used to enter the various Protocols and settings.
Control Panel which includes Display and Key-Pad
Display Displays current status of the various system features and functions.
Air Vents (Front) for air output
Air Vents (Left for air output
and Right Sides)
Air Vents for air intake
(Bottom)
4
MultiGene Thermal Cycler
Name Function
5
3. Getting Started
3.1 Unpacking
Open the MultiGene Gradient/non-Gradient thermal cycler package and confirm that all
items below are included:
MultiGene Gradient/non-Gradient Operation manual
thermal cycler unit
Electric fuses (7A & 3.15A) AC power cord
Warranty card 0.5/0.2 ml Labnet tubes
If there are any items missing, damaged, or incorrect items, please contact your
distributor or sales representatives immediately.
Note: Please use 7A fuse in power source 100~120/50~60Hz countries, while 3.15A fuse
is for power source 200~220V/50~60Hz countries.
Press Power On/Off Button to switch on the machine. The screen will show the software
version for 10 seconds and then enter User Folder screen. Pressing any key will go to
User Folder screen immediately.
Note:
6
MultiGene Thermal Cycler
before heated lid is at operating temperature.
2. To avoid heat dissipation, the Heated Lid should be closed and tighten during a
protocol run. Keeping the heated lid open during a protocol run might cause
“HEATER ERROR”, because the heated lid fails to reach the set temperature.
3. To evaporate the condensation water inside sample block, and to avoid “HEATER
ERROR”, users should turn off the Lid Temperature under SYSTEM mode before
heating the sample block and keeping the lid open.
Rotate Lid Knob in a clockwise direction to lower down the heating plate and press down
on the cap of tubes, tighten until you feel resistance. This adjustment will provide better
contact between the tube or plates and the block for better performance. Rotate the Lid
Knob in counter-clockwise direction as shown in Figure 4 until the Lid pops out. Only
when the Lid pops out, users can slide the Lid to the rear as shown in Figure 5.
To lift up the lid for heating plate cleaning, please refer Figure 6 ~ 8.
7
Figure 5. Slide the lid backward to access the block
Figure 6. Slide the lid to align the lid hinge with notches of top case
8
MultiGene Thermal Cycler
9
3.5 Function Map
The User Screen includes 5 function options; 1) VIEW folders under selected User folder,
2) NEW(Create) a User, 3)EDIT a User, 4)DELETE a User, 5)SYSTEM. Once a user
folder is selected and the password is approved, users can; 1) RUN a Protocol,
2)NEW(Create) a protocol, 3) EDIT(VIEW) a Protocol, and 4) DELETE a Protocol.
The function map can is shown in Figure 6. Various functions, including Gradient, can be
added to the selected/created Protocol.
Main Menu
User Folder
Lid Temp.
Insert
Beeper
Delete
Administrator
Advanced
Service Mode
Gradient
Increment
Pause
Save
10
MultiGene Thermal Cycler
Note: To avoid deformation of the sample tubes or Strips, it’s important to load at least
one tube in each of the four corners and at least four tubes in the center of the block, this
will distribute the pressure evenly throughout the surface.
11
4. User Folder Management
After the unit is turned on, the initial screen which contains software version and
manufacturer’s name will show up for 10 seconds and then enter User Folder screen.
Pressing any key will go to User Folder screen immediately.
A sample screen is shown below.
F1 F2 F3 F4 F5
Users can use the function keys(F1~F4) above to view, create, edit and delete the User
folders. F5:SYSTEM is used for system settings.
Note:
1. The [ PRESET ] and [ PUBLIC ] folders are pre-registered folders which require no
Personal Identification Number(PIN) to enter.
12
MultiGene Thermal Cycler
To view the contents of the User, will require you to enter the 4-digit Personal
Identification Number(PIN) as shown in the screen below. After the PIN is entered,
press F1:ACCEPT or ENTER key to confirm the PIN. To run a Protocol, please refer
Section 5.1.
F1 F2 F3 F4 F5
13
4.4 Deleting a User Folder
Use directional arrow keys to select a User folder. Press F4:DELETE to delete the
selected User folder. You are required to enter the PIN and press F1:ACCEPT or
ENTER . The screen will prompt “Do you want to delete”. Press F1:ACCEPT to
confirm the deletion. Press F5:EXIT to abort the deletion.
Note: User folders which contain any protocols can’t be deleted except by administerator.
14
MultiGene Thermal Cycler
5. Protocol Management
After entering the selected User Folder, users can press F1~F4 keys to Run,
New(Create), Edit(View) and Delete a selected Protocol as shown below.
F1 F2 F3 F4 F5
F1 F2 F3 F4 F5
15
2. Users are required to select a Temperature Control Mode before running the Protocol.
Mode 1(Tube Control): Enable system operation with simulated temperature
control. If Mode 1 is selected, users are required to input the sample volume
(5-100ul). In this mode, temperature control is based on Block temperature
and an algorithm which estimates sample temperature based on the block
temperature and the sample volume. Recommended for more accurate
temperature control.
Mode 2( Block Control): Enable system operation with maximum temperature
control based on the actual block temperature. Recommended for normal
operation.
Note:
1. The highlighted temperature (Temp) value represents the current block temperature
during Protocol running. The highlighted time (Time) value below represents the
remaining running time in the temperature segment. Once the countdown is completed,
the highlighted section will continue to move to the next temperature or cycle segment
and start counting down the time.
2. The Lid temperature will flash on the status bar when the Protocol starts and remain
on until it reaches 50ºC. When the lid reaches 50°C “HOT!” will appear in the right
hand corner and remain until the temperature falls below 40°C The sample block will
start to heat up when the lid temperature reaches its set temperature.
16
MultiGene Thermal Cycler
F1 F2 F3 F4 F5
F1 F2 F3 F4 F5
17
6. Creating a New Protocol
A new Protocol can be created by going to the User Folder and selecting and editing a
basic Protocol format, and then assigning it a new name.
1. Select User name on User screen. Press F1:OPEN or ENTER button to enter User
Folder.
2. Key in PIN to enter the User’s folder.
3. Press F2:NEW to create a new Protocol by using a standard template, shown below.
Note: Users can create a new Protocol under Public folder without a PIN. Users can
create a new Protocol inside Preset folder, BUT can not save these changes to the
original protocol settings.
F1 F2 F3 F4 F5
18
MultiGene Thermal Cycler
19
Mike>PROTOCOL>NEW>ADV. 8/21/2006 11:00
1 Temp 3 Temp 25 Cyc 2 Temp
F1 F2 F3 F4 F5
20
MultiGene Thermal Cycler
Note: The temperature values between 2nd and 11th rows are calculated values and
represent an algorithmic based on the heat distribution though out the block.
Row 01 02 03 04 05 06
Temp 84.0 84.6 85.2 85.8 86.4 87.0
Row 07 08 09 10 11 12
Temp 87.5 88.0 88.5 89.0 89.5 90.0
F1 F2 F3 F4 F5
21
Temperature Increment
CLEAR + - EXIT
F1 F2 F3 F4 F5
Note:
Time Increment
CLEAR + - EXIT
F1 F2 F3 F4 F5
22
MultiGene Thermal Cycler
7. Editing(Viewing) a Protocol
An editing Protocol can be saved under a new name (newly created Protocol) or under the
same name (edited Protocol).
Under the User’s Folder, choose the Protocol to be edited. Select F3:EDIT to edit the
selected Protocol. This displays the existing Protocol and enables editing of the
temperature segment and numerical values as shown below.
Please refer Chapter 6 Creating a New Protocol to edit the temperature segment and
numerical values, and other advanced functions. Press F1:SAVE to save the edited
Protocol under the original name or F2:SAVE A to as a new protocol.
Program name:A
Key0(0+=<>&()%)
Key1(1.,-?’#:;/)
Maximum: 12 characters
F1 F2 F3 F4 F5
23
8. Deleting a Protocol
1. An existing Protocol can be deleted under a User’s Folder.
2. Under the User’s Folder screen, choose a Protocol. Select F4:DELETE to delete the
selected Protocol.
3. If F4:DELETE is selected the screen will prompt you “Are you sure you want to
delete?” Press F1:YES to confirm the deletion. Press F5:NO to abort the deletion.
YES NO
F1 F2 F3 F4 F5
Note: Protocols in the Preset Protocols Folder can’t be deleted, while Protocols in the
Public Folder can be deleted by any user if not password protected.
24
MultiGene Thermal Cycler
9. System Setup
System Setup contains 7 functions:
1. Data Log: contains the history of the latest-run Protocol.
2. Date&Time: Users can change the Date and Time settings.
3. Lid Temperature: Users can choose the preferred lid temperature as default setting
during running a Protocol.
4. Beeper: Users can turn on/off the audible beep.
5. Administrator: Users can delete unwanted Usernames and their Protocols.
6. Service Mode: is reserved for Service engineers.
SELECT EXIT
F1 F2 F3 F4 F5
25
9.1 Data Log
The Data Log records a log of activity for the most recently run Protocol. Executing a
new Protocol updates the history log by overwriting previous log. To export log to a PC a
Hyper Terminal communication protocol must be installed. Hyper Terminal
communication protocol comes free with Windows. Please call for Windows based link.
Once installed to export information to your PC press F1:PC (upload reports to PC).
Please refer Appendix D: Hyper Terminal and USB Driver Setup.
PC Exit
F1 F2 F3 F4 F5
26
MultiGene Thermal Cycler
ACCEPT EXIT
F1 F2 F3 F4 F5
ACCEPT EXIT
F1 F2 F3 F4 F5
27
9.4 Beeper
Users can turn ON or OFF the sound of this unit by selecting the Beeper screen as shown
below under SYSTEM screen. Please remember to press F1:ACCEPT to store your
settings.
ACCEPT EXIT
F1 F2 F3 F4 F5
9.5 Administrator
The password of Administrator is “9600”. The Administrator of this unit has rights to
delete any User Folder and any Protocols inside a User’s Folder.
28
MultiGene Thermal Cycler
29
5. Gently slide the heating block until the plug is exposed then disconnect.
6. Slide the new block into place and connect the plug.
7. Gently slide the block until the two screw holes are lined up.
8. Replace the two screws and plugs.
9. Shut the face panel and secure it with two screws.
11. Troubleshooting
Problem Cause Action
The display is off even when 1. Power is not reaching the 1. Check power source
power is switched on. system. voltage.
2. Power cord is not 2. Reconnect the power
plugged into the socket cord.
properly. 3. Replace the power fuse.
3. Power fuse is blown out. 4. Return the unit for
4. Block fuse is blown out. service.
Can’t reach 4°C. 1. Operating temperature 1. Operate the unit in a
environment may be environment temperature
unsuitable. between 15 to 30 °C.
2. Electronic cooling 2. Return the unit for
element may be damaged service.
or aged.
Cycle time is too long 1. Operating environment 1. Operate the system in a
temperature may be temperature environment
unsuitable. between 15 to 30 °C.
2. Electronic cooling 2. Return the unit for
element may be damaged service.
or aged. 3. Return the unit for
3. Faulty temp. sensor. service.
Lid heater does not work. 1. Lid heater is set off. 1. Check Lid Temperature
2. Lid sensor problem setting in System Mode.
2. Return the unit for
service.
No key entry sound. 1. Sound may currently be 1. Check Beeper setting in
set off. System Mode.
2. Faulty keys or keypad 2. Return the unit for
wiring. service.
The display goes off. 1. Faulty backlight. 1. Check the LCD contrast.
30
MultiGene Thermal Cycler
2. Faulty LCD panel 2. Return the unit for
service.
Display is too dark or light Backlight brightness is not Adjust Display Brightness
adjusted properly. Potentiometer.
Lid will not close. 1. Foreign matter, between Remove the foreign object or
lid heater and sample matter.
block.
2. Faulty lid lock
mechanism.
Error message appears. Refer to list of error Check the nature of the error
messages below. and take the suggested
actions.
31
Appendix A: Technical specifications
Model Name TC9600-G TC9600 TC6000-G TC6000
Sample Capacity 0.2ml x 96 tubes 0.5ml x 60 tubes
Temperature Ranges 4 ~ 99.9 °C 4 ~ 99.9 °C
Temperature Accuracy +/- 0.5 °C ( 30 to 99.9 °C) +/- 0.5 °C ( 30 to 99.9 °C)
+/- 2 °C ( 4 to 15 °C) +/- 2 °C ( 4 to 15 °C)
Temperature +/- 0.5°C +/- 0.5°C
Uniformity
Max. Temp. 3°C / 2 °C 3°C / 2 °C
Increment/Decrement
Rate
Gradient Temperature 30 ~ 99 °C N/A 30 ~ 99 °C N/A
(Max. temp. (Max. temp.
difference: 24 °C) difference: 24 °C)
Max. Lid Temperature 100~115 °C 100~115 °C
Protocol Capacity 100 sets 100 sets
Operating Temperature 15 ~ 30 °C 15 ~ 30 °C
Operating Humidity 20 ~ 80% RH 20 ~ 80% RH
Networking 1 x RS-232, 1 x USB 1 x RS-232, 1 x USB
Power Supply 100 to 240V AC, 50/60 Hz, 100 to 240V AC, 50/60 Hz,
500VA 500VA
Dimensions (WxDxH) 24 x 25 x 42cm 24 x 25 x 42cm
Weight (Net) 9 Kg 9 Kg
32
MultiGene Thermal Cycler
F1 F2 F3 F4 F5
Gradient
F1 F2 F3 F4 F5
33
Touch down
- 0.5
F1 F2 F3 F4 F5
Time Increment
- 0:05
F1 F2 F3 F4 F5
34
MultiGene Thermal Cycler
Appendix C: CE Declaration
Declaration of Conformity
35
Appendix D: Hyper Terminal and USB
Driver Setup
Users are required to purchase the USB or RS-232 cable to connect the MultiGene and
the PC. For USB connection, users are required to install the USB driver.
36
MultiGene Thermal Cycler
Sealing Tape
Clear sealing tape will seal 96 and 384 well Plates when used with Labnets compression
mat this polypropylene film eliminates well to well contamination and cross-over in PCR
applications
Functional Temperature range -40ºC to +104ºC Cat# TC96-TT
Sealing Tape
Aluminum sealing tape will seal 96 and 384 well Plates when used with Labnet’s
37
compression mat this polypropylene film eliminates well to well contamination and
cross-over in amplification applications. Ideal for light sensitive samples and its
uniformly applied adhesive provides optimum well sealing
Functional Temperature range -80ºC to +104ºC Cat# TC-AS-100.
Silicone Compression Mat is used to seal 96 well plates when using the MultiGene
Thermal Cycler heated lid function. These silicone mats can be cleaned and sterilized
by either 10% hypochlorite solution and ddi water rinse, or by autoclaving at 121ºC for
15 minutes. Cat# TC96-CM-100