Sensepoint XCD: Calibration Guide
Sensepoint XCD: Calibration Guide
Sensepoint XCD
Calibration Handbook
Sensepoint XCD Calibration
It is recommended to periodically carry out a gas response check on the Sensepoint XCD to ensure correct
operation. This may be done in two ways;
1. A simple Response Check often referred to as a “BUMP TEST” is a test using calibration gas applied to
the sensor via the nozzle of the Weather Protection or using the Sensepoint XCD Gassing Cap.
If a BUMP TEST is done via the Weather Protection nozzle it may be necessary in windy conditions to
increase the flow rate of the test gas by a further 1 LPM, OR, to shelter the weather protection from the
wind.
2. A full gas calibration of the sensor as described in the following section, using ONLY the Sensepoint XCD
Gassing Cap (Part No.: S3KCAL).
4. Put the magnet over the ‘✓’ switch again and move to enter the Calibration menu.
5. The display will show the current gas reading, and the ‘ ’ icon flashes.
6. When the zero gas reading is stable use ‘✓’ to confirm zero calibration.
7. If successful the display shows ‘ZEro PASS’ (if not successful, the display shows ‘ZEro FAIL’ and returns
to configuration mode).
10. If span calibration is required use ‘✓’ proceed to the next step. If span calibration is not required, use ‘ !’
to select ‘No’ and ‘✓’ to return to configuration mode.
12. The display will show the current gas reading, and the ‘ ’ icon flashes.
13. Connect the regulator to the span gas cylinder.
14. Apply the span gas to the sensor using the Sensepoint XCD Gassing Cap. The live gas reading is displayed.
When the reading is stable, use ‘✓’ to confirm span calibration.
15. If the sensor has been replaced the following display may be shown.
16. Use ‘ !’ to select ‘YES’ if the sensor has been replaced or ‘No’ if it has not been replaced.
17. If the span calibration is successful the instrument will briefly display ‘SPAn PASS’ (if fails ‘SPAN FAIL’
displayed and returns to configuration mode).
NOTE
J/<.>)/+.(%&*-$&I/)%.%#&2(-%+$)&.'&)$'$+&/@+$)&/&'-22$''@-<&2/<.>)/+.(%3
18. The display alternates between “Purg gAS” and the gas reading to indicate that the unit is expecting the
span gas to be removed from the sensor.
19. Promptly switch off the calibration span gas and remove the Sensepoint XCD Gassing Cap from the
sensor to allow the gas to disperse.
NOTE
K$A$A>$)&+(&/<I/"'&)$1</2$&+?$&L$/+?$)&M)(+$2+.(%&/%*&(+?$)&/22$''().$'3