Memory Hicorder: High Speed Oscilloscope and Multi-Channel Logger - All in One Powerful Instrument
Memory Hicorder: High Speed Oscilloscope and Multi-Channel Logger - All in One Powerful Instrument
O scillos 64 or 128
DigNiteawlModels for up
to
Two
8860, 8861 MEMORY HiCORDER
Recorders
8861
8860
The sampling rate for the memory recorder can be controlled by the timing of
an external clock signal (10 MS/s). This is useful for example to collect data
Sampling rate Output waveform synchronized to the running cycle of an engine.
Time
How long can I record to the internal The following table shows the maximum recording time when measuring 1 channel on Model 8860 using the built-in preset recording lengths and the respective MEMORY BOARDs.
Recording lengths can be increased manually in 1DIV steps to extend the recording time, e.g., up to 320,000 DIV with the 32MW MEMORY BOARD.
direct access memory? 1-channel setting, 32 megawords 1-channel setting, 128 megawords 1-channel setting, 512 megawords 1-channel setting, 1 gigawords
Time axis Sampling memory capacity memory capacity memory capacity memory capacity
The 9715 MEMORY BOARD offers 32MW period Recording length of 200k divisions Recording length of 1000k divisions Recording length of 5000k divisions Recording length of 10, 000k divisions
8995 A4 Printer
Sample and log temperature and other parameters over long periods
- High-speed 50msec scanning across a maximum of 128 channels -
Sheet 1
Four-split screen vertical scroll Sheet 3
Sheet function Six-split screen continuous scroll
Sheet 2
Four-split screen horizontal scroll Sheet 4
Timeline and XY-axis composite waveform
Trigger during capturing and search after capturing (supported from Version 2.00 and onwards) Set multiple triggers on a single channel
The trigger function allows you to set diverse parameters to detect Unlike with conventional MEMORY HiCORDERs, the 8860
a particular waveform anomaly during capturing. Setting the pre- series allows you to set multiple trigger parameters on a particular
trigger mode allows you to monitor the pre-trigger waveform. This single channel.
is useful for analyzing the cause of the anomaly. T h is a l lows you t o set , for
On the other hand, the search function allows you to detect an example, the slope trigger, level
anomaly after all data is captured. This allows you to search for t r igger, p er iod t r igger, a nd
and display an anomaly in the same manner as with the trigger window-in trigger on the same
function. input waveform to monitor it.
When a waveform is unpredictable and setting a parameter during (Eight parameters in the 8860 and 16 in
measurement is difficult, it is recommended to use the search func- the 8861 can be set.)
tion to locate an anomaly after capturing.
A simple level trigger
causes the start trigger to
be activated
■ Slope trigger
Power waveform noise
Unlike with conventional MEMORY HiCORDERs,
a slope trigger is supported. This allows you Rapid change in temperature
Automatic saving during measurement Redundancy against errors in the storage destination
Unlike with conventional MEMORY HiCORDERs, the 8860 The 8860 series allows you to set up to two storage destinations.
series allows automatic saving during measurement. Even if, for example, an overflow error occurs on a PC card during
automatic saving, switching to the second backup storage destina-
tion takes place automatically to ensure saving continues.
Automatic saving during measurement
shortens the waiting time for saving
after measurement ends
■ Restrictions
Automatic saving during measurement is restricted to 1ms/division
or more of the time axis. Also, restrictions to the time axis settings
differ depending on the scanner module, number of channels,
storage media, and setting conditions for real-time printing.
2000, 5000 and 10000). The calculation speed for the same condition (when
Frequency domain
performing the most time-consuming analysis) is about ten times faster than with
the conventional 8855 model.
Scaling in “dB”
The long desired capability to scale in dB is supported. You no longer
need to perform logarithmic calculation holding a calculator in one
hand. The 8860 and 8861 models allow you to enter the overall value
(sum of power spectrum values) in Before scaling
dB, thus making scaling easier.
T his enables signa ls to be
easily read directly from, for
example, a noise meter.
After scaling
9
■ Product Specifications
Basic specifications 8860 (max. 4 input modules) 8861 (max. 8 input modules) Interchangeable input modules
Plug-in input modules Plug-in input modules The slot design using plug-in type modules offers
Input type/number Max. 16 analog channels (max. 64 Max. 32 analog channels (max. 128 superior flexibility for measuring all types of signals
of channels channels with scanner unit) + 16 logic channels with scanner unit) + 16 logic
including voltage, current, frequency, temperature,
channels (standard configuration) channels (standard configuration)
acceleration and more.
Memory (high-speed recording)
Recorder (real-time recording)
Measurement
REC & MEM (real-time recording + high-speed recording, from version 2.00)
functions
FFT (frequency analysis, from version 1.07)
Real-time data save (from version 1.10)
Maximum 20 MS/second (50 ns, all channels simultaneously, using the 8956 ANALOG UNIT)
sampling rate External sampling (10 MS/second, 100 ns)
1 unit: Voltage 2ch, 20 MS/s, 12-bit resolution (8956)
Customized direct access memory capacity
1 unit: Voltage 2ch, 2 MS/s, 16-bit resolution (8957) Determine the amount of direct access memory you will
1 unit: Voltage / Thermocouple scan 16ch, max. 50 ms refresh rate, 1/1000 of range
need based on your application and budget requirements
Types of resolution for temperature axis - (8958)
1 unit: Voltage / RMS, 1 MS/s, 12-bit resolution (8959) and build your own unique measurement system. Install
measurement signals
1 unit: Voltage 2ch, 1 MS/s, 12-bit resolution (8936/8938) either one 32MW, 128MW, 512MW, or 1GW board in the
Highest sampling rate and 1 unit: Voltage / Thermocouple 2ch, 4 kS/s, 12-bit resolution (8937) 8860, and two of the same capacity board in the 8861.
resolution
(Model number of input 1 unit: Strain gauge 2ch, 1 MS/s, 12-bit resolution (8939)
module shown in parentheses) 1 unit: Strain gauge 2ch, 200 kS/s, 16-bit resolution (8960)
1 unit: Frequency / Integration / Current / Voltage 2ch, 1 MS/s, 12-bit resolution (8940)
1 unit: Accelerometer 2ch, 1 MS/s, 12-bit resolution (8947)
1 unit: Voltage 4ch, 1 MS/s, 12-bit resolution (8946) Save in real-time to internal hard disk or MO drive
Choose to save valuable measurement data to the
32 Mega-words (9715 MEMORY BOARD × 1) 64 Mega-words (9715 MEMORY BOARD × 2) internal large-capacity 60 GB hard disk or 2.3GB-
Direct access internal (analog 12-bit + logic 4-bit) × 32 Mega-words/ (analog 12-bit + logic 4-bit) × 32 Mega-words/
memory ch (using 1 channel) to (analog 12-bit + logic ch (using 2 channels) to (analog 12-bit + logic compatible MO drive (choose one option for factory pre-
installation), the two standard PC Card drives, your
*1 Factory installation only: select 4-bit) × 2 Mega-words/ch (using 16 channels) 4-bit) × 2 Mega-words/ch (using 32 channels)
1 board for the 8860, and 2 of own external storage media via the USB port, or to
the same capacity for the 8861 1 Giga-word (9715-03 MEMORY BOARD × 1) 2 Giga-words (9715-03 MEMORY BOARD × 2) the optional floppy disk drive. Use the secondary
when ordering. (analog 12-bit + logic 4-bit) × 1 Giga-word/ch (analog 12-bit + logic 4-bit) × 1 Giga-word/ch
(using 1 channel) to (analog 12-bit + logic (using 2 channels) to (analog 12-bit + logic storage feature as backup in case of writing error
9715: 32 Megawords
9715-01: 128 Megawords 4-bit) × 64 Mega-words/ch (using 16 channels) 4-bit) × 64 Mega-words/ch (using 32 channels) to the primary disk to insure that your important
9715-02: 512 Megawords
9715-03: 1 Gigawords data is properly saved. The internal HD, MO or
Note: 1 word = 2 bytes (12-bits or 16-bits), therefore 1 giga-word = 2 giga-bytes.
PC card drives can also be used to save data in
Note: Internal memory is allocated depending on the number of channels used.
real-time to facilitate long-term recording.
PC Card Type II slot (standard) × 2: up to 4 GB (Flash ATA), FAT, or FAT-32 (Real-time save available from version 1.10.)
Data storage media format supported
3.5" Floppy disk drive (optional external drive): 1.44 MB (2HD), 720 KB (2DD),
*2 Only one slot is available FAT format, via USB connection (external)
in the main unit for either a
built-in MO drive or built-in 3.5" Magneto-optical drive (optional internal drive *2) × 1: Max. 2.3 GB (128
hard disk drive. MB, 230 MB, 540 MB, 640 MB, 1.3 GB), FAT format
2.5" Hard disk drive (optional internal drive *2) × 1: 60 GB, FAT-32 format
The following items are preserved on the memory board(s) even after power off:
Backup functions Clock and parameter setting backup (standard): at least 10 years; at
*3 Factory installation only - reference temperature (25°C) ■ External dimensions
please specify upon order 9719 Waveform backup function (using optional Model 9719 *3): 10 hours (8860) or
MEMORY BACKUP UNIT
5 hours (8861), after full charge, at reference temperature (25°C) 8860
184.5mm
(7.26 in)
Standard external
USB: USB1.1 compliant (for 9716, keyboard/mouse/memory)
interfaces
LAN: RJ-45 connector, Ethernet 100BASE-TX, 10BASE-T
272.5mm
(10.73 in)
*4 Using PC Card slot and Functions: HTTP server, FTP server, File sharing, DHCP compatible
(9.84 in)
250mm
Environmental conditions Temperature and humidity range for use: 0°C to 40°C, 20% to 80% rh
(no condensation) Temperature and humidity range for storage: -10°C to 50°C, 20% to 90% rh 330mm
30 184.5mm
(12.99 in) (7.26 in)
(1.18 in)
Compliance Safety: EN61010,
standard EMC: EN61326-1, EN61000-3-2, EN61000-3-3
■ External dimensions
Power 100 to 240 V AC (50/60 Hz)
requirements 12 V DC (use the 9684 DC POWER UNIT : option, factory installation only) 8861
Power 140 VA max. (printer not used) 190 VA max. (printer not used) Note: MO Drive, thermal printer and input units
284.5mm
(11.20 in)
Supplied
(9.84 in)
250mm
330mm 30 284.5mm
(12.99 in) (1.18 in) (11.20 in)
10
■ Product Specifications
Print/display section *6 Printer functions are available when optional printer unit is installed Recorder functions (X-Y function available from version 2.00)
Display 10.4 inch TFT color LCD (SVGA, 800 × 600 dots) 10 ms to 200 ms *8/division, 500 ms to 1 hour/division with 18
9231 RECORDING PAPER: 216 mm (8.50 in) × 30 m (98.43 ft), ranges, time axis resolution 100 points/division, time axis zoom:
thermal paper roll (when using A4 size 8995 printer unit) ×2 to ×5 in 2 stages, compression: 1/2 to 1/500 in 8 stages
*6 Recording paper Time axis *8:possible,
Virtual record function: At 10 ms - 200 ms/division, printing in real time is not
9234 RECORDING PAPER: 112 mm (4.41 in) × 18 m (59.06 ft), but waveform data are stored in memory and can be monitored on screen.
thermal paper roll (when using A6 size 8995-01 printer unit) Data are stored for 10,000 divisions before the end of measurement. At recording
length settings other than "Continuous", the printer can be used simultaneously, for
9231 RECORDING PAPER: 200 mm (7.87 in) , full scale 20 divisions, 1 follow-up printing of waveforms.
*6 Recording width division = 10 mm (0.39 in) (when using A4 size 8995 printer unit)
Sampling rate 100 ns to 1 sec in 8 stages (selectable in 1/100 of time axis range)
9234 RECORDING PAPER: 100 mm (3.94 in) , full scale 10 divisions,
1 division = 10 mm (0.39 in) (when using A6 size 8995-01 printer unit) 32 MW memory: manual setting in 1-division steps (max. 5,000 *9
10 lines/mm (when using A4-size the 8995 printer unit), 8 lines/mm divisions) or built-in presets of 25 to 5,000 divisions, continuous *8
*6 Paper feed density (when using A6-size the 8995-01 printer unit) 128 MW memory: manual setting in 1-division steps (max. 20,000 *9
* 20 lines/mm with "smoothed printing" memory function (when using A4-size the 8995 printer unit) divisions) or built-in presets of 25 to 20,000 divisions, continuous *8
*6 Recording speed Max. 20 mm (0.79 in)/sec 512 MW memory: manual setting in 1-division steps (max. 50,000 *9
Recording length
divisions) or built-in presets of 25 to 80,000 divisions, continuous *8
Trigger functions 1 GW memory: manual setting in 1-division steps (max. 100,000 *9
divisions) or built-in presets of 25 to 160,000 divisions, continuous *8
Turn on/off independently for each trigger source of analog/logic *8 At time axis 10 ms to 200 ms/division and printer ON, Continuous setting cannot be selected
A – D, external trigger (a rise of 2.5V or terminal short circuit); *9 Memory of 8861 is twice than shown above, but recording length is the same.
timer trigger, inter-source AND/OR, forced trigger,
Trigger sources standard mode (trigger source to all analog channels settable), X-Y sampling rate 300 μs fixed (dot), 300 μs to 25 ms (line)
extend mode (multiple analog sources to a single analog 20 dots/division (LCD, 1 screen), horizontal 80 dots/division × vertical
channel settable, up to 8 for 8860, and up to 8 on channels/units X-Y axis resolution
80 dots/division (printer)
1 – 4, and up to 8 on channels/units 5 – 8 for 8861 settable)
Store data for most recent 5,000 *10 divisions (with 32 MW memory) in
Level: Triggering occurs when preset voltage level is crossed Waveform memory memory. Backward scrolling and re-printing available.
(upwards or downwards). *10 20,000 divisions with 128 MW, 80,000 divisions with 512 MW, 160,000 divisions with 1
GW. Memory of 8861 is twice that of 8860, but recording length is the same.
Window: Triggering occurs when window defined by upper and
lower limit is entered or exited. Split screen (1 to 8), X-Y screen (1, 2, 4 screens, max. 8 combined), sheet
Period: Rising edge or falling edge cycle of preset voltage value display (max. 32 channels per sheet), logging (print/display measurement
is monitored and triggering occurs when defined cycle range is Screen and printing
data as digital values), voltage axis zoom (×2 to ×100), compression (×1/2
exceeded. to ×1/10), variable display
Trigger types Glitch: Triggering occurs when pulse width from rising or falling
(analog)
edge of preset voltage value is underrun.
Slope: Triggering occurs when preset change degree (slope) is
exceeded or underrun. REC & MEM function (function available from version 2.00)
Voltage drop: Triggering occurs when voltage drops below peak
voltage setting (for 50/60 Hz AC power lines only). 10 ms to 200 ms *11/division, 500 ms to 1 hour/division, 18 ranges,
Event setting: Event count is performed for each source, and Time axis time axis resolution 100 points/division, sampling rate: same as
triggering occurs when a preset count is exceeded. (REC) sampling rate for MEM function
*11 Not available for virtual recording at 10 ms to 200 ms/division
Level setting resolution 0.1% of full scale (full scale = 20 divisions)
Time axis 10 μs to 5 min/division, 25 ranges, time axis resolution 100 points/
Trigger types 1, 0, 0⎥1, ×, pattern setting, AND/OR setting for groups of 4 channels, (MEM) division, sampling rate: 1/100 of time axis
(logic) level or edge detect selectable (0⎥1: changing to any value activates trigger)
REC: 25 to 2,000 *12 divisions, max. 50,000 divisions *12, continuous
Trigger filter OFF, setting range 0.1 to 10.0 divisions in 0.1 division steps (MEM, REC Recording length MEM: 25 to 5,000 *12 divisions, max. 100,000 divisions *12
(analog/logic) & MEM function), ON (10 ms)/OFF (REC function)
*12 Depends on installed memory 32 MW to 1GW (free setting in 1-division steps also possible)
Pre-trigger function to capture pre- and post-trigger waveform, Store data for most recent 2,000 *13 divisions (with 32 MW memory) in
trigger output (active Low with BNC terminal and open collector memory, allowing backscroll and reprinting
Other functions Waveform memory
5 voltage output). Level display while waiting for trigger, Start/
(REC) *13 10,000 divisions with 128 MW, 20,000 divisions with 512 MW, 50,000 divisions with 1
stop trigger conditions independently selectable GW. Memory of 8861 is twice that of 8860, but recording length is the same.divisions
with 1 GW.
Fixed: 1/100 of time axis range, Variable: external sampling Divided use of memory space (up to 4096 divisions), sequential
Memory divide
Sampling rate Sampling period can be used to set time axis save, block search
Two different sampling rate settings are possible
Add-on recording (retain previous data at start and resume recording
Other functions from previous data)
32 MW memory : manual setting in 1-division steps (max. 320,000
divisions *7) Or built-in presets of 25 to 200,000 divisions *7
128 MW memory : manual setting in 1-division steps (max. 1,280,000
divisions *7) Or built-in presets of 25 to 1,000,000 divisions *7
Recording length 512 MW memory : manual setting in 1-division steps (max. 5,120,000
divisions *7) Or built-in presets of 25 to 5,000,000 divisions *7 FFT function (function available from version 1.07)
1 GW memory : manual setting in 1-division steps (max. 10,240,000 Storage waveform, linear spectrum, RMS spectrum, power
divisions *7) Or built-in presets of 25 to 10,000,000 divisions *7 spectrum, power spectrum density, cross power spectrum,
*7 Maximum recording length or built-in preset length when using 1 channel (8860) or 2 channels Analysis mode auto-correlation function, histogram, transfer function, cross-
(8861). Memory of 8861 is twice that of 8860, but recording length is the same.
correlation function, impulse response, coherence function,
Record data from before the trigger point, -100 to +100% of octave analysis
Pre-trigger
recording length (free setting in 1% steps) Analysis channels 1-channel FFT, 2-channel FFT in selected channels (up to 8 analysis functions)
Split screen (1 to 8), X-Y screen (1, 2, 4 screens, max. 8 combined), sheet Frequency range 133 mHz to 8 MHz, resolution 1/400, 1/800, 1/2000, 1/4000
Screen and display (max. 32 channels per sheet), logging (print/display measurement data as
printing digital values), voltage axis zoom (×2 to ×100), compression (×1/2 to ×1/10), Number of
1000, 2000, 5000, 10000 points
overlay, zoom, variable display, vernier display sampling points
Memory splitting Divided use of memory space (up to 4096 divisions), sequential save Selected from: Newly loaded data / MEM function waveform
Analysis data
data / MEM waveform of REC & MEM function
Four arithmetic operations, absolute value, exponentiation, common
logarithm, square root, moving average, differentiation once and twice, Window Rectangular, Hanning, Exponential, Hamming, Blackman,
Waveform calculation functions Blackman-Harris, Flat-top
integration once and twice, parallel displacement along the time axis,
trigonometric functions (sin, cos, tan, arc-sin, arc-cos, arc-tan) Screen and Split screen (1/2/4), Nyquist, array, logging (print/display measurement
printing data as digital values), frequency axis zoom and left/right scrolling
(Numerical calculation by specifying calculation area with cursors A and B,
numerical calculation judgment, automatic saving of numerical calculation Time axis / frequency axis simple averaging, exponential
results, saving of any existing numerical calculation results) Averaging
averaging, peak hold
Average value, effective (rms) value, peak to peak value, maximum
Numerical calculation value, time to maximum value, minimum value, time to minimum
value, period, frequency rise time, fall time, standard deviation, area
value, X-Y area value, time to level, pulse width, duty ratio, pulse
count, four arithmetic operations for results of numerical calculation
11
■ Product Specifications
Real-time save function (function available from version 1.10) ■ Options specifications (sold separately) For 8860/8861 only
10 ms to 200 ms *14/division, 500 ms to 1 hour/division, 18 ranges,
Time axis time axis resolution 100 points/division, sampling rate: same as
(REC: compressed data) sampling rate for MEM function Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 183 (7.20in) D mm,
*14 Not available for virtual recording at 10 ms to 200 ms/division approx. 385 g (13.6 oz) Accessories: Flathead screwdriver × 1, short bar × 2
(Accuracy at 23 ±5°C/73 ±9°F, 30 to 80 % rh after 1 hour of warm-up
Time axis 100 μs to 5 min/division, 20 ranges (limited depending on store target and number of 8958 16ch SCANNER UNIT time and adjustment; accuracy guaranteed for 1 year)
(MEM: real-time data) channels), time axis resolution 100 points/division, sampling rate: 1/100 of time axis
Measurement functions Number of channels: 16, for voltage measurement/temperature measurement with thermocouple
Save to Magneto-optical disk, hard disk, LAN, PC Card
Voltage input/Thermocouple input: screw-type terminal strip, recommended
Depending on available space on storage media / file system / wire diameter *1, detachable terminal block (with cover) *1 Recommended cable, single-
Recording length number of channels / REC time axis wire: 0.14 to 1.5 mm2, braided wire 0.14 to 1.0 mm2 (conductor wire diameter min. 0.18 mm), AWG 26 to 16
Selectable in division steps up to maximum recording length Input connectors Input impedance: 1 MΩ, 850 kΩ with line fault detection ON, Max. rated
voltage to earth: 33 Vrms or 70 V DC (with input isolated from the unit, the
During measurement: REC waveform, after measurement: toggle maximum voltage that can be applied between input channel and chassis and between input
REC/MEM waveform display, simultaneous display of REC/MEM channels without damage)
Screen and
waveform with split screen, split screen (1 to 8), sheet display (max. 32
printing Voltage 5m, 50m, 500m, 2 V/division, 4 ranges, full scale: 20 divisions, measurement
channels per sheet), logging (print/display measurement data as digital values),
measurement range: ±100% of full scale, digital filter: 10/50/60 Hz, measurement resolution
zoom, variable display
range 1/1600 of measurement range (using 16-bit A/D conversion; installed in 8860/8861)
Memory transfer MEM function data can be transferred Temperature 10°C/division (-100°C/ to +200°C), 50°C/division (-200°C/ to +1000°C),
measurement range 100°C/division (-200°C/ to +2000°C), 3 ranges, full scale: 20 divisions,
Additional features (Some functions available from version 2.00) (Upper and lower limit values digital filter: 10/50/60 Hz, measurement resolution 1/1000 of
depend on measurement input
Measurement parameter printing, cursor measurement, scaling, range of sensor) measurement range (using 16-bit A/D conversion; installed in 8860/8861)
current clamp setting, comment input, screen hard copy, list/
Thermocouple K: -200 to 1350°C, J: -200 to 1200°C, E: -200 to 1000°C, T: -200 to
gauge, start condition hold, auto setup, auto save, remote control
range 400°C, N: -200 to 1300°C, R: 0 to 1700°C, S: 0 to 1700°C, B: 400 to
(start/stop/print control), auto range, over-range indication, VIEW
General (JIS C 1602-1995) 1800°C, W (WRe5-26): 0 to 2000°C, reference junction compensation:
function, online help, key lock, level monitor, vernier function, (ASTM E-988-96) internal/ external (switchable), line fault detection ON/OFF switchable
offset cancel, event marker input, waveform search function,
report printing, file save of printing image, save & load of 50 ms/all channels (digital filter OFF), 300 ms/all channels (digital
channel selection, 8835/8855 waveform data load Data refresh rate
filter 50/60 Hz), 1.4 s/all channels (digital filter 10 Hz)
Voltage: ±0.2% of full scale, thermocouple (K, J, E, T, N): ±0.05%
of full scale ±1°C, (R, S, B, W): ±0.05% of full scale ±2°C (400°C
Accuracy or more), ±0.05% of full scale ±3.5°C (less than 400°C), reference
junction compensation accuracy: ±1°C (added to measurement accuracy
with internal reference junction compensation)
Max. allowable input 40 V DC (the maximum voltage that can be applied across input pins without damage)
Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm,
■ Options specifications (sold separately) For 8860/8861 only
approx. 290 g (10.2 oz) Accessories: None
(Accuracy at 23 ±5°C/73 ±9°F, 30 to 80 % rh after 30 minutes of warm-up
8959 DC/RMS UNIT time and zero-adjust; accuracy guaranteed for 1 year)
Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm, Measurement functions Number of channels: 2, for voltage measurement
approx. 290 g (10.2 oz) Accessories: None Isolated BNC connector (input impedance 1 MΩ, input capacitance 30 pF), Max. rated
(Accuracy at 23 ±5°C/73 ±9°F, 30 to 80 % rh after 30 minutes of warm-up Input connectors voltage to earth: 370 V AC, DC (with input isolated from the unit, the maximum voltage that
8956 ANALOG UNIT time and zero-adjust; accuracy guaranteed for 1 year) can be applied between input channel and chassis and between input channels without damage)
Measurement functions Number of channels: 2, for voltage measurement 5 mV to 20 V/division, 12 ranges, full scale: 20 divisions, AC
Measurement
Isolated BNC connector (input impedance 1 MΩ, input capacitance 40 pF), Max. rated voltage for possible measurement/display using the memory
range
Input connectors voltage to earth: 300 V AC, DC (with input isolated from the unit, the maximum voltage that function: 280V rms, low-pass filter: 5/500/5k/100k Hz
can be applied between input channel and chassis and between input channels without damage)
Measurement resolution 1/80 of measurement range (using 12-bit A/D conversion; installed in 8860/8861)
5 mV to 20 V/division, 12 ranges, full scale: 20 divisions, AC Highest sampling rate 1 MS/s (simultaneous sampling in 2 channels)
Measurement
voltage for possible measurement/display using the memory
range DC amplitude: ±0.4% of full scale (with filter 5 Hz), zero position: ±
function: 280 V rms, low-pass filter: 5/500/5k/1M Hz Accuracy
0.1% of full scale (with filter 5 Hz, after zero adjustment)
Measurement resolution 1/100 of measurement range (using 12-bit A/D conversion; installed in 8860/8861)
RMS amplitude accuracy: ±1% of full scale (DC, 20 Hz to 1 kHz), ±3% of full scale (1 kHz to
Highest sampling rate 20 MS/s (simultaneous sampling in 2 channels) RMS
100 kHz), response time: SLOW 5 s (rise time from 0 to 90% of full scale), MID 800 ms (rise time
measurement
DC amplitude: ±0.4% of full scale (with filter 5 Hz) from 0 to 90% of full scale), FAST 100 ms (rise time from 0 to 90% of full scale), crest factor: 2
Accuracy
Zero position: ±0.1% of full scale (with filter 5 Hz, after zero adjustment) Frequency characteristics DC to 400 kHz ±3 dB, with AC coupling: 7 Hz to 400 kHz ±3 dB
Frequency characteristics DC to 10 MHz ±3 dB, with AC coupling: 7 Hz to 10 MHz ±3 dB Input coupling DC, GND, AC
Input coupling DC, GND, AC Max. allowable input 400 V DC (the maximum voltage that can be applied across input pins without damage)
Max. allowable input 400 V DC (the maximum voltage that can be applied across input pins without damage)
Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm, Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm,
approx. 310 g (10.9 oz) Accessories: None approx. 290 g (10.2 oz) Accessories: Conversion cable × 2, cable length 50cm (19.69in)
(Accuracy at 23 ±5°C/73 ±9°F, 30 to 80 % rh after 30 minutes of warm-up (Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh after 1 hour of warm-up
8957 HIGH-RESOLUTION UNIT time and zero-adjust; accuracy guaranteed for 1 year) 8960 STRAIN UNIT time and auto-balance; accuracy guaranteed for 1 year)
Measurement functions Number of channels: 2, for voltage measurement Measurement functions Number of channels: 2, for distortion measurement (electronic auto-balancing, balance adjustment range within ±10000 με)
Isolated BNC connector (input impedance 1 MΩ, input capacitance 40 pF), Max. rated Via conversion cable, TAJIMI PRC03-12A10-7M10.5, Max. rated voltage to
Input connectors voltage to earth: 300 V AC, DC (with input isolated from the unit, the maximum voltage that Input connectors earth: 33 Vrms or 70 V DC (with input isolated from the unit, the maximum voltage that
can be applied between input channel and chassis and between input channels without damage) can be applied between input channel and chassis and between input channels without damage)
5 mV to 20 V/division, 12 ranges, full scale: 20 divisions, AC Suitable Strain gauge converter, bridge impedance: 120 Ω to 1 kΩ (bridge voltage 2 V),
Measurement
voltage for possible measurement/display using the memory transducer 350 Ω to 1 kΩ (bridge voltage 5 V, 10 V), bridge voltage 2, 5, 10 ±0.05 V
range
function: 280 V rms, low-pass filter: 5/50/500/5k/50k Hz
Measurement 20 με to 1000 με/division, 6 ranges, full scale: 20 divisions, low-
Integrated filter for suppressing aliasing distortion caused by FFT range pass filter: 5/10/100/1k Hz
Anti-aliasing filter
processing (automatic cutoff frequency setting/OFF) Integrated filter for suppressing aliasing distortion caused by FFT
Anti-aliasing filter
Measurement resolution 1/1600 of measurement range (using 16-bit A/D conversion; installed in 8860/8861) processing (automatic cutoff frequency setting/OFF)
Highest sampling rate 2 MS/s (simultaneous sampling in 2 channels) Measurement resolution 1/1600 of measurement range (using 16-bit A/D conversion; installed in 8860/8861)
DC amplitude: ±0.2% of full scale (with filter 5 Hz) Highest sampling rate 200 kS/s (2-channel simultaneous sampling)
Accuracy
Zero position: ±0.1% of full scale (with filter 5 Hz, after zero adjustment) DC amplitude: ±(0.4% of full scale +2 με), zero position: ±(0.1% of full scale
Accuracy After auto-balancing
Frequency characteristics DC to 200 kHz ±3 dB, with AC coupling: 7 Hz to 200 kHz ±3 dB +2 με) (at 5 Hz filter ON)
Input coupling DC, GND, AC Frequency characteristics DC to 20 kHz +1/-3 dB
Max. allowable input 400 V DC (the maximum voltage that can be applied across input pins without damage) Max. allowable input 10 V DC (the maximum voltage that can be applied across input pins without damage)
* Available from main unit 8860/8861 version 1.06
12
9318 CONVERSION CABLE (to connect 9270 to 9272, 9277 to 9279 and 8940)
9319 CONVERSION CABLE (to connect 3273, 3273-50 and 8940)
Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm, Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm,
approx. 290 g (10.2 oz) Accessories: None approx. 300 g (10.6 oz) Accessories: None
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh after 30 minutes of warm-up (Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh after 30 minutes of warm-up
8936 ANALOG UNIT time and zero-adjust; accuracy guaranteed for 1 year) 8940 F/V UNIT time and zero-adjust; accuracy guaranteed for 1 year)
Measurement functions Number of channels: 2, for voltage measurement Measurement Number of channels: 2, for voltage input based frequency measurement, integration,
functions pulse duty ratio, current (with optional clamp-on sensor), and voltage measurement
Isolated BNC connector (input impedance 1 MΩ, input capacitance 30 pF), Max. rated
Input connectors voltage to earth: 370 V AC, DC (with input isolated from the unit, the maximum voltage that Metallic BNC connector (input impedance 1 MΩ, input capacitance 60 pF), sensor
can be applied between input channel and chassis and between input channels without damage) connector (dedicated connector for clamp-on sensor via conversion cable, common
Input connectors ground with recorder), Max. rated voltage to earth: 30 Vrms or 60 V DC (with
5 mV to 20 V/division, 12 ranges, full scale: 20 divisions, AC input isolated from the unit, the maximum voltage that can be applied between input
Measurement channel and chassis and between input channels without damage)
voltage for possible measurement/display using the memory
range
function: 280 V rms, low-pass filter: 5/500/5k/100k Hz Compatible current sensors 9270, 9271, 9272, 9277, 9278, 9279, 3273, 3273-50
Measurement resolution 1/80 of measurement range (using 12-bit A/D conversion; installed in 8860/8861) Frequency: DC to 100 kHz, with 0.05 Hz to 5 kHz/division, 11 ranges, 5
1 MS/s (simultaneous sampling in 2 channels) (r/min) to 500 (r/min)/division, 5 ranges, P50 Hz (40 to 60 Hz), P60 Hz (50
Highest sampling rate
to 70 Hz) * Power line frequency measurement requires 9322 DIFFERENTIAL PROBE or 9303
Accuracy DC amplitude: ±0.4% of full scale, zero position: ±0.1% of full scale (after zero adjustment) PT.,Accuracy: ±0.2% of full scale (except 5 kHz/division range), ±0.7% of
Frequency characteristics DC to 400 kHz ±3 dB, with AC coupling: 7 Hz to 400 kHz ±3 dB full scale (5 kHz/division range), ±0.032 Hz (P50/P60 Hz range)
Measurement Integration: DC to 90 kHz, with 5 counts to 500 k counts/division, 11
Input coupling DC, GND, AC range ranges
Max. allowable input 400 V DC (the maximum voltage that can be applied across input pins without damage) Pulse duty ratio: 10 Hz - 100 kHz, with 100% of full scale, 1 range,
* When using Model 8936 with serial number earlier than 041018234 on Models 8860 or 8861, residual noise will be 850 μVp-p.
Accuracy: ±1% of full scale (10 Hz - 10 kHz)
Threshold: -10 to +10 V (settable in 0.2V steps)
Full scale: 20 divisions, Max. allowable input: 30 Vrms or 60 V DC (the
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh after 30 minutes of warm-up
8938 FFT ANALOG UNIT time and zero-adjust; accuracy guaranteed for 1 year)
maximum voltage that can be applied across input pins without damage)
Measurement functions Number of channels: 2, for voltage measurement Voltage: 0.5 mV to 2 V/division, 12 ranges
Integrated filter for suppressing aliasing distortion caused by FFT Current: 5 mA to 100 A/division, 10 ranges, using current sensor (powered
Anti-aliasing filter from the 8940, max. 4 sensors total)
processing (automatic cutoff frequency setting/OFF)
Measurement DC amplitude accuracy: ±0.4% of full scale, zero position ±0.15% of full
Other functions Other specifications same as 8936 ANALOG UNIT range scale (current measurement accuracy dependent on sensor accuracy/characteristics)
* When using Model 8938 with serial number earlier than 041132532 on Models 8860 or 8861, residual noise will be 1.4 mVp-p. Frequency characteristics: DC to 400 kHz ±3 dB
Full scale: 20 divisions, Max. allowable input: 30 Vrms or 60 V DC (the
maximum voltage that can be applied across input pins without damage)
Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm, Measurement resolution 1/80 of measurement range (installed in 8860/8861, excluding current range when using 9279)
approx. 300 g (10.6 oz) Accessories: None Highest sampling rate 1 MS/s (2-channel simultaneous sampling), (frequency/duty ratio measurement: 1.125 μs cycle)
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh after 1 hour of warm-up
8937 VOLTAGE/TEMP UNIT Voltage input pull-up: ON (10 kΩ)/OFF, input coupling: DC, GND,
time and zero-adjust; accuracy guaranteed for 1 year) Other functions
AC (voltage/current), DC (others), low-pass filter: 5/500/5k/100k Hz
Measurement functions Number of channels: 2, for voltage measurement/temperature measurement with thermocouple
Voltage input: metallic BNC connector (input impedance 1 MΩ, input capacitance 50 pF),
Input connectors thermocouple input: plug-in connector (input impedance min. 5.1 MΩ), Max. rated
voltage to earth: 30 Vrms or 60 V DC (with input isolated from the unit, the maximum voltage Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm,
that can be applied between input channel and chassis and between input channels without damage) approx. 310 g (10.9 oz) Accessories: None
500 μV to 2 V/division, 12 ranges, full scale: 20 divisions, low-pass filter: 5/500/5k/100k Hz (Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh after 30 minutes of warm-up
Voltage measurement
Measurement resolution: 1/80 of measurement range (using 12-bit A/D conversion; installed in
8946 4ch ANALOG UNIT time and zero-adjust; accuracy guaranteed for 1 year)
range
8860/8861) Measurement functions Number of channels: 4, for voltage measurement
10°C to 100°C/division, 4 ranges, full scale: 20 divisions, low-pass filter: 5/500 Hz Metallic BNC connector (input impedance 1 MΩ, input capacitance 15 pF), Max. rated
Temperature
Measurement resolution:1/80 of measurement range (using 12-bit A/D conversion; Input connectors voltage to earth: 30 Vrms or 60 V DC (with input isolated from the unit, the maximum voltage
measurement range installed in 8860/8861) that can be applied between input channel and chassis and between input channels without damage)
K: -200 to 1350°C, E: -200 to 800°C, J: -200 to 1100°C, T: -200 to 400°C, Measurement 10 mV to 2 V/division, 8 ranges, full scale: 20 divisions, low-
Thermocouple
N: -200 to 1300°C, R: 0 to 1700°C, S: 0 to 1700°C, B: 300 to 1800°C, range pass filter, 5/500/5k/50k Hz, input coupling: DC, GND
range
Reference junction compensation: internal/ external (switchable) Measurement resolution 1/80 of measurement range (using 12-bit A/D conversion; installed in 8860/8861)
Highest sampling rate Voltage input: 1 MS/s, Temperature measurement: 4 kS/s (2-channel simultaneous sampling) Highest sampling rate 1 MS/s (4-channel simultaneous sampling)
Voltage input: DC amplitude ±0.4% of full scale, zero position ±0.15% of full scale,
Accuracy DC amplitude: ±0.5% of full scale, zero position: ±0.15% of full scale (after zero adjustment)
Temperature measurement (K, E, J, T, N): ±0.1% of full scale ±1°C, ±0.1% of
Accuracy full scale ±2°C (-200 to 0°C), (R, S): ±0.1% of full scale ±3°C, (B): ±0.1% of full Frequency characteristics DC to 100 kHz ±3 dB
scale ±4°C (400 to 1800°C) ,Reference junction compensation accuracy: ±0.1%
Max. allowable input 30 Vrms or 60 V DC (the maximum voltage that can be applied across input pins without damage)
of full scale ±1.5 °C (internal reference junction compensation)
Frequency Voltage input: DC to 400 kHz + 1/-3 dB
characteristics Temperature measurement: DC to 1 kHz + 1/-3 dB
Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm,
Input coupling DC, GND, AC approx. 310 g (10.9 oz) Accessories: None
Max. allowable input 30 Vrms or 60 V DC (the maximum voltage that can be applied across input pins without damage) (Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh after 1 hour of warm-up
8947 CHARGE UNIT time and zero-adjust; accuracy guaranteed for 1 year)
* When using Model 8937 with serial number earlier than 041135257 on Models 8860 or 8861, residual noise will be 150 μVp-p.
Measurement functions Number of channels: 2, for acceleration measurement
Voltage input/integrated preamplifier input: metallic BNC connector
(for voltage input: input impedance 1 MΩ, input capacitance 200 pF or less)
Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm, Input connectors Charge input: miniature connector (#10-32 UNF)
approx. 250 g (8.8 oz) Accessories: Conversion cable × 2 Max. rated voltage to earth: 30 Vrms or 60 V DC (with input isolated
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh after 1 hour of warm-up from the unit, the maximum voltage that can be applied between input channel and
8939 STRAIN UNIT time and auto-balance; accuracy guaranteed for 1 year) chassis and between input channels without damage)
Measurement functions Number of channels: 2, for distortion measurement (electronic auto-balancing, balance adjustment range within ±10000 με) Suitable Charge input: Charge-output type piezoelectric acceleration pick-up sensor
transducer Internal preamp input: Acceleration pick-up sensor with an internal preamp
Via conversion cable, TAJIMI PRC03-12A10-7M10.5, Max. rated voltage to 50m (m/s2)/division to 10k (m/s2)/division, 12 ranges × 6 types, charge input
Input connectors earth: 30 Vrms or 60 V DC (with input isolated from the unit, the maximum voltage that Measurement
sensitivity: 0.1 to 10 pC/(m/s2), integrated pre-amplifier input: 0.1 to 10 mV/(m/
can be applied between input channel and chassis and between input channels without damage) range
Charge input s2), amplitude accuracy: ±2% of full scale, frequency characteristics: 1 to 50
Suitable Strain gauge converter, bridge impedance: 120 Ω to 1 kΩ, bridge (miniature connector) kHz, +1/-3 dB, low-pass filter: 500/5k Hz, pre-amplifier drive power source: 2
transducer voltage 2 ±0.05 V Internal pre-amp input mA ±20%, +15 V ±5%, maximum input charge: ±500 pC (high-sensitivity setting,
(BNC connector)
6 ranges), ±50000 pC (low-sensitivity setting, 6 ranges)
Measurement 20 με - 1000 με/division, 6 ranges, full scale: 20 divisions, low-
range pass filter: 10/30/300/3k Hz Measurement 500 μV to 2 V/division, 12 ranges, DC amplitude accuracy: ±0.4% of full scale,
range frequency characteristics: DC to 400 kHz, +1/-3 dB, low-pass filter: 5/500/5k/100k Hz,
Measurement resolution 1/80 of measurement range (using 12-bit A/D conversion; installed in 8860/8861) Voltage input (BNC connector) input coupling: DC, GND, AC, Max. allowable input: 30 Vrms or 60 V DC
Highest sampling rate 1 MS/s (2-channel simultaneous sampling) Measurement resolution 1/80 to 1/32 of measurement range (depending on measurement sensitivity; installed in 8860/8861)
Accuracy After auto-balancing DC amplitude: ±(0.5% of full scale +2 με), zero position: ±0.5% of full scale Highest sampling rate 1 MS/s (2-channel simultaneous sampling)
Frequency characteristics DC to 20 kHz +1/-3 dB Integrated filter for suppressing aliasing distortion caused by FFT
Anti-aliasing filter
Max. allowable input 10 V DC + AC peak (the maximum voltage that can be applied across input pins without damage) processing (automatic cutoff frequency setting/OFF)
* When using Model 8947 with serial number earlier than 040933650 on Models 8860 or 8861, residual noise will be 200 μVp-p.
13
Dimensions and mass: approx. 315.8 (12.43in) W × 29 (1.14in) H × 244.4 (9.62in) D mm,
approx. 1.25 kg (44.1oz) Accessories: None
Note: Factory-installed option, build in on the bottom case of 8860/8861
Dimensions and mass: approx. 315.8 (12.43in) W × 18.2 (0.72in) H × 244.4 (9.62in) D mm,
■ PC Software Specifications Note: Wv ver 1.20 or later, and 8860/8861 main unit ver 1.03 or later
approx. 570 g (20.1oz) Accessories: None Wave Viewer (Wv) Software (Application disk CD-R, bundled accessory)
Note: Factory-installed option, build in on the bottom case of 8860/8861
• Simple display of waveform file
9687 PROBE POWER UNIT • Text conversion: convert binary data file to text format,
No. of powerd channels 8, rated output voltage : ±12 V with selectable space or tab separators in addition to CSV,
and specifiable section, thinning available
Functions • Display format settings: scroll functions, enlarge/reduce
3273 (15 A Max.) : less than 8 units
3273 -50 (35 A Max.) : less than 6 units display, display channel settings
Compatible probes 3274 (150 A Max.) : less than 6 units • Others: voltage value trace function, jump to cursor/trig-
3275 (500 A Max.) : less than 5 units ger position function
3276 (30 A Max.) : less than 6 units
Compatible PC operating systems Windows 95/98/Me, Windows NT 4.0 (SP3 or later), 2000, XP
14
Configuration of options Note: Options described as "factory-installed options" must be specified when ordering and cannot be installed by the user. Note: Product names appearing in this catalog are trademarks or registered trademarks of various companies.
Input modules Logic signal measurement
Install by inserting into the main unit. Can be replaced by user.
8956 ANALOG UNIT
8957 HIGH RESOLUTION UNIT
8958 16ch SCANNER UNIT
8959 DC/RMS UNIT
8960 STRAIN UNIT
8936 ANALOG UNIT 9323 CONVERSION CABLE
8937 VOLTAGE/TEMP UNIT Used for connecting the 9320/9321 and
9320-01 LOGIC PROBE 9321-01 LOGIC PROBE 9327-01 LOGIC PROBE 8860 series MEMORY HiCORDERs,
8938 FFT ANALOG UNIT
4-channel type, for voltage/contact signal 4 isolated channels, ON/OFF detection 4-channel type, for voltage/contact signal because the terminal shapes are different.
8939 STRAIN UNIT
ON/OFF detection (miniature terminal of AC/DC voltage (miniature terminal ON/OFF detection (high-speed, miniature * This cable is not required for the small-
7
8940 F/V UNIT terminal types 9327, 9320-01, and
7 56 4 for use with the 8861/8860, 8855, for use with the 8861/8860, 8855, terminal type)
58 59 895 89 89 8946 4ch ANALOG UNIT 9321-01.
89 89 8807-01/8808-01) 8807-01/8808-01)
8947 CHARGE UNIT
Note: Input cables are not supplied. Please purchase the appropriate cable for the intended application.
Current measurement * The 3274, 3275, and 3276 cannot be used for the 8940 F/V unit.
Voltage measurement 8940 F/V UNIT 9279 Not CE certified
9277 3273-50
9318 CONVERSION CABLE (to connect 9277 - 9279 and 8940) 9278 9319 CONVERSION CABLE (to connect 3273-50 and 8940)
9277 UNIVERSAL CLAMP ON CT 9318 CONVERSION CABLE
Observe waveforms from DC to distorted to connect 9277/9278/9279 and 8940
9197 CONNECTION 9322 DIFFERENTIAL AC. DC to 100 kHz response, input 20 A /
CORD PROBE 3273-50 CLAMP ON PROBE
output 2 V AC DC to 50 MHz wideband response,
For high voltage (up to For up to 2 kV DC or 1 kV AC, 9418-15 9717 MO UNIT
500 V) AC adapter required, or 9687 Probe power Factory-installed option. Either MO 9278 UNIVERSAL CLAMP ON CT mA-class current up to 30 A rms
Observe waveforms from DC to distorted
unit and 9248 Power cord required for DRIVE UNIT or HDD UNIT can be 9319 CONVERSION CABLE
operation. selected. (2.3 GB to 128 MB) AC. DC to 100 kHz response, input 200 A / to connect 3273-50 and 8940
output 2 V AC
9248 POWER CORD 9555 SENSOR UNIT
9718 HD UNIT 9279 UNIVERSAL CLAMP ON CT Connect and power up to one
to connect 9322 Factory-installed option. Either MO Observe waveforms from DC to distorted UNIVERSAL CLAMP ON CTs
Differential probe and DRIVE UNIT or HDD UNIT can be AC. DC to 20 kHz response, input 500 A / 9555 to use in combination with voltage
9687 Probe power unit selected. (60 GB) output 2 V AC input modules
9198 CONNECTION
CORD 9418-15 AC ADAPTER 9719 MEMORY
For low voltage (up to 300 V) For powering 9322 Differential probe, 3272
100 to 240 V AC, 12 V/2.5 A Note: Memory boards are not built in as a standard BACKUP UNIT 3274 3273-50
feature. Choose from the following memory boards Factory-installed option. Internal 3275 3276 3269
for factory pre-installation - one board for Model 8860, type.
and two of the same capacity for the 8861.
3273-50 CLAMP ON PROBE 3272 POWER SUPPLY
*power
Either the 9684 DC
DC to 50 MHz wideband response, Connect and power up to two CLAMP
9715 MEMORY BOARD unit or the 9687
probe power unit may mA-class current up to 30 A rms ON PROBEs to use in combination
(32 Megaword capacity)
be used. If you want with voltage input modules
9715-01 MEMORY BOARD 3274 CLAMP ON PROBE
9665 10:1 PROBE 9217 CONNECTION
(128 Megaword capacity)
to use both at the same DC to 10 MHz wideband response, 3269 POWER SUPPLY
Max. rated voltage to earth is CORD time, consult us. mA-class current up to 150 A rms Connect and power up to four CLAMP
same as for input module, max. Cord has insulated BNC connectors at 9715-02 MEMORY BOARD
input voltage 1 kV rms (up to 1 both ends, and connects to insulated (512 Megaword capacity) 9684 DC POWER UNIT 3275 CLAMP ON PROBE ON PROBEs to use in combination
Factory-installed option, not user DC to 2 MHz wideband response, with voltage input modules
MHz) BNC connectors on input module 9715-03 MEMORY BOARD
(1 Gigaword capacity)
installable, build in on the bottom case. mA-class current up to 500 A rms *power
Either the 9684 DC
unit or the 9687
12 VDC drive.
3276 CLAMP ON PROBE probe power unit may
DC to 100 MHz wideband response, be used. If you want
mA-class current up to 30 A rms to use both at the same
9666 100:1 PROBE 9165 CONNECTION 8860 MEMORY HiCORDER (main unit only) time, consult us.
Max. rated voltage to earth is
same as for input module, max.
CORD
Cord has metallic BNC connectors at
8861 MEMORY HiCORDER (main unit only) 9687 PROBE POWER UNIT
Factory-installed option, not user installable, build
input voltage 5 kV peak (up to 1 both ends, and connects to metallic BNC *optional
The 8860/8861 MEMORY HiCORDER cannot operate alone. You must install one or more in on the bottom case. For up to five or more
8860, and two memory boards*of the same capacity for the 8861.
MHz) connectors such as trigger terminals input modules in the unit. When ordering, specify one memory board for the
3273-50/ 3274/ 3275/ 3276 clamp on probes.
Not CE certified
Printing width 200 mm (7.87 inch).
Other options Supplied with PC
Card adapter 9726 PC CARD 128M
(128 MB capacity)
220H PAPER WINDER 9723 CARRYING CASE 9727 PC CARD 256M
Paper width 70 to 220 mm (2.76 to For only the 8860, hard trunk
8.66 inch), 100V AC Only type (256 MB capacity)
PC Card Precaution
9728 PC CARD 512M
U s e o n l y P C C a rd s s o l d (512 MB capacity)
9724 CARRYING CASE by HIOKI . Compatibility
9234 RECORDING PAPER 8995-01 A6 PRINTER UNIT 9716 FD DRIVE and performance are not
(for 8995-01 printer only) Factory-installed option. Either 8995 External type with USB For only the 8861, hard trunk
guaranteed for PC cards made
9729 PC CARD 1G
A6 width 112 mm (4.41 in) × 18 m connection type by other manufacturers. You (1 GB capacity)
or 8995-01 printer can be installed.
(59.06 ft), 10 rolls Printing width 100 mm (3.94 inch). may be unable to read from or
save data to such cards.
■ Combination example: 8860 (for high-speed applications) 8860 (for logging applications)
Main unit Memory 32 MW 2ch 4ch 6ch 8ch 16ch 32ch 48ch 64ch
Model number × quantity 8860×1 9715×1 8956×1 8956×2 8956×3 8956×4 8958×1 8958×2 8958×3 8958×4
Input cable 9198×2 9198×4 9198×6 9198×8 − − − −
■ Combination example: 8861 (for high-speed applications) 8861 (for logging applications)
Main unit Memory 64 MW 4ch 8ch 12ch 16ch 32ch 64ch 96ch 128ch
Model number × quantity 8861×1 9715×2 8956×2 8956×4 8956×6 8956×8 8958×2 8958×4 8958×6 8958×8
Input cable 9198×4 9198×8 9198×12 9198×16 − − − −
DISTRIBUTED BY
All information correct as of Mar. 20, 2006. All specifications are subject to change without notice. 8860E4-63M-03P Printed in Japan