Super+ Manual PDF

The document introduces the URiSCAN Super+ fully automated urine analyzer. It is designed to analyze urine test strips placed in cassettes, processing between 100-1000 samples per day. The analyzer uses image processing and measurement of absorbance/refractive index to test 13 urine parameters from the strips. It can also read specific gravity, color, and clarity of samples. Key functions include automatic sample handling, dispensing of precise sample volumes, storing results, and transferring data to external systems.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
1K views

Super+ Manual PDF

The document introduces the URiSCAN Super+ fully automated urine analyzer. It is designed to analyze urine test strips placed in cassettes, processing between 100-1000 samples per day. The analyzer uses image processing and measurement of absorbance/refractive index to test 13 urine parameters from the strips. It can also read specific gravity, color, and clarity of samples. Key functions include automatic sample handling, dispensing of precise sample volumes, storing results, and transferring data to external systems.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 67

URiSCAN Super+ Fully automated urine analyzer Rev.

02

CONTENTS

Chapter1 System Overview 2


1.1 Preface 3
1.2 Precaution 4
1.3 Overview of the URiSCAN Super+ 5
1.4 Overview and principle of Urinalysis test 7
1.5 Main Functions 8
1.6 Caution 9
1.7 Symbols 10

Chapter2 Installation 11
2.1 Technical specification 12
2.2 Hardware 13
2.3 Hardware Installation 18
2.4 Software installation 20
2.5 Cassette Installation 35

Chapter3 Operation 38
3.1 Quick manual 39
3.2 Routine operation 40

Chapter4 Maintenance 51
4.1 Periodic Maintenance 52
4.2 Maintenance menu 53

Chapter5 Trouble Shooting 54


5.1 Alarm message 55

Chapter6 Appendix 59
6.1 Communication 60
6.2 Initial value of Range table in each parameter 62
6.3 Program for the sensitivity of each parameter 64
6.4 Measurement density 66

Operator Manual / 1
URiSCAN Super+ Fully automated urine analyzer Rev.02

Chapter 1 System Overview

Introduce Uriscan Super+ Fully automated urine analyzer in this chapter

1.1 Preface ----------------------------------------------------------------------- 3


1.2 Precaution (including international warranty certificate)---------- 4
1.3 Overview of the Uriscan Super+ ---------------------------------------
--------------------------------------------------------- 6
1.4 Overview and principle of urinalysis test -----------------------------
------------------ 7
1.5 Main features ---------------------------------------------------------------
---------------------------------- 8
1.6 Caution -----------------------------------------------------------------------
--------
------------------------------------------------------------- 9
1.7 Symbols ---------------------------------------------------------------------- 10
-

Operator Manual / 2
URiSCAN Super+ Fully automated urine analyzer Rev.02

1.1 Preface
URiSCAN Super+ is a fully automated urine chemistry analyzer, which is designed for
reading urine test strips placed in URISCAN Super+ cassette.

URiSCAN Super+ is optimized for workloads of between 100 and 1000 urine samples per
day. The analyzer is for professional use so that only trained personnel working in a
professional laboratory environment can operate the analyzer. Also, the analyzer is
designed to assay urine samples only.

It is important to read this manual thoroughly before operating URiSCAN Super+. Any
disregard of the instructions in the operator’s manual may result in a safety risk.

How to use this manual


This manual is designed to help you perform all tasks required as part of your work with
the URISCAN Super+.

The manual is divided into the following parts:

 Chapter 1 – Overview: This section contains the general information about


URISCAN Super+

 Chapter 2 - Installation: This section describes the specification, the hardware


components and installation, the instruction for use, and how to set up the cassette.

 Chapter 3 – Operation : This section defines the quick operation and routine
operation process of URISCAN Super+ analyzer in a step-by-step guide.

 Chapter 4 – Maintenance : This section provides details on all of the general


maintenance actions.

 Chapter 5 – Troubleshooting : This section provides the information how to handle


the warning sign, error, and the alarms which require solutions by trained YD technical
support personnel are indicated.

 Chapter 6– Appendix : This section contains additional information

Operator Manual / 3
URiSCAN Super+ Fully automated urine analyzer Rev.02

1.2 PRECAUTION

YD International Warranty Certificate

The URiSCAN Super+ is guaranteed by YD Diagnostics Corp., for a period of 1(one) year
after the date of purchase against defects due to faulty workmanship or materials. The
guarantee covers both piece parts and labor. Service under guarantee is only provided
upon presentation of reasonable evidence (completed guarantee card or purchase receipt)
that date of the claim is within the guarantee period.
The warranty is not valid if the defect is due to accidental damage, misuse or neglect and
in case of alternations or repair carried out by unauthorized persons. Service (during and
after guarantee) is available in all countries where the URiSCAN Super+ is officially
distributed by YD Diagnostics Corp., the local YD service organizations also arrange
service although there may be a delay if the required spare parts are not readily available.

Installation, relocation and customer service

(1) Customer shall not attempt installation of this instrument to other place after
its delivery. Technical service engineer is needed to install the instrument
for its safe and precise operation.
(2) Before installation of this instrument, the customer shall make reparations
for satisfying the installation requirements in accordance with this manual.
(3) When relocation of this instrument is required, please notify the sales
representative or dealer.

Technical seminars and training for users

Users can take a technical seminar or training for using this instrument safely
and exactly. For seminar/training procedure, contact sales representative or
dealer.

Handling of strips and quality control sample

(1) Strips and quality control samples shall be handled, stored and discarded
according to the instructions given by the respective dealers.
(2) Samples and quality control samples to be used by the customer in analytical
tasks with this instrument shall be handled, stored and discarded by the
customer’s responsibility in accordance with the predetermined laboratory
standards.

Operator Manual / 4
URiSCAN Super+ Fully automated urine analyzer Rev.02

1.3 Overview of the URiSCAN Super+

1.3.1 Basic Information of the URiSCAN Super+

URiSCAN Super+ is a fully automated urine chemistry analyzer which works


based on the principle of image processing technology and absorbance/refractive
index measurement. The urine test strip provides reliable urine test results with
high accuracy and reproducibility by adopting high technology.
The cassette manufactured by YD Diagnostics, containing urine test strips which
test maximum of 13 parameters for chemistry analysis: Blood, Bilirubin, Creatinine,
Urobilinogen, Ketones, Protein, Microalbumin, Nitrite, Glucose, pH, Leucocytes,
Ascorbic Acid and compensation pad.
The instrument determines the sample’s specific gravity through CCD module and
converting it to the corresponding clinical value. It can also read urine color and
clarity.

The main function of the URiSCAN Super+ analyzer includes.

 Identification of the sample


 Automatic dispensing on the urine test strips
 Automatic sampling
 Accurate result by specific sample dispensing volume
 Memory of the sample measurement value
 Data management

The 400 test strips are placed in the URiSCAN Super+. The test strips are moved
to the pipette area automatically through the mechanical arrangement. One test
strip is used for analyzing each sample. The test results are measured through the
principles of the CCD color image sensor and the equivalent clinical values
obtained.

This instrument has an additional feature of transfer data to a host PC or laboratory


information and management system (LIMS). This instrument has a built-in
barcode scanner for effective and easy identification of the sample ID, or it can be
also connected with an external barcode reader.

Operator Manual / 5
URiSCAN Super+ Fully automated urine analyzer Rev.02

1.3.2 Urine Test Principle

URiSCAN Super+ is a fully automated urine analyzer for In Vitro Semi –


Quantitative measurements of urine test strips. Each element of the urine will be
absorbed into each test portion of the strip and the subsequent chemical and
enzymatic reactions will change the color of the test paper accordingly. The degree
of color will be proportional to the concentrations of each element in the urine. The
CCD will play the role of measuring and indicating the extents and routes of such
coloring.

CCD is the most advanced optical element that can analysis 3 primary color of light
and read accurate value of the color development of the strip. It can get more
accurate and objective result compare with light and shade detection by photo
diode which used generally.

Below is procedure of test principle in the URiSCAN Super+

Analog-to Digital
CCD Image Sensor Converter

LED
CPU

Test Strip

 The results that are stored in the memory can be sent to host PC by
RS232C or Ethernet or sent to USB. The results can be printed on the
external printer via parallel port. In additional, the instrument can be
connected to an external keyboard or mouse by PS/2/USB port.

* CCD : Charge Coupled Device


Color Image Sensor / R.G.B(3 lines) x 5400 = 5400 pixel / line
* R, G, B : Three primary colors of light (Red, Green, Blue)
* Wave Length: Red – 630nm / Green – 540nm / Blue – 460nm

Operator Manual / 6
URiSCAN Super+ Fully automated urine analyzer Rev.02

1.4 Overview and principle of Urinalysis test

1.4.1. Overview
URiSCAN Super+ is an analyzer which can test the urine components such as
Blood, Bilirubin, Creatinine, Urobilinogen, Ketones, Protein, Microalbumin, Nitrite,
Glucose, pH, Leucocytes, Ascorbic Acid, color, turbidity and specific gravity(with 13
parameter strip), showing the result values via screen and printing. With 11
parameter strip, it tests Blood, Bilirubin, Urobilinogen, Ketones, Protein, , Nitrite,
Glucose, pH, Leucocytes, Ascorbic Acid, color, turbidity and specific gravity.

1.4.2. Principle of Urinalysis


Urinalysis is the general and basic method that executed in laboratories for the
detection and confirmation of early disease along with other diagnostic references.

The three basic principle method of urinalysis

 The visual evaluation of the color and clarity of the urine sample
 Chemical evaluation by using urinalysis strips
 Sediment evaluation by using microscopic examination

Out of above, the test method of urinalysis strips is widely used as a test method
for accurate and rapid diagnosis comparatively of each substance’s level of urine
specimens and clinical aspects.

The URiSCAN Super+ cassette strips manufactured by YD Diagnostics Corp. are


available for 13 parameters Blood, Bilirubin, Creatinine, Urobilinogen, Ketones,
Protein, Microalbumin, Nitrite, Glucose, pH, Leucocytes, Ascorbic Acid and
compensation.

Belows are the figure of URiSCAN Super+ cassette strip.

URiSCAN Super+ Cassette Strip (13 parameters)


Compensation Vt. C Leucocytes pH Glucose Micro Albumin Nitrite Protein Ketones Urobilinogen Creatinine Bilirubin Blood

URiSCAN Super Cassette Strip (11 parameters)


Compensation Vt. C Leucocytes pH Glucose Nitrite Protein Ketones Urobilinogen Bilirubin Blood

Operator Manual / 7
URiSCAN Super+ Fully automated urine analyzer Rev.02

1.5 Main Functions


 Test Capacity: 200 Samples/hour

 Sample volume: 1.5~2.0mL

 Sample mixing function is available

 100 samples can be loaded at once

 LCD Display Monitor


Review data (test result + strip image)
LCD color press screen (Size 8.4 Inches)
Resolutions: 800 X 600 (true color)

 Auto Memory function


1,000,000 test results, 500 emergency and quality control records

 Easy access to review test results stored as database.


Easy to search and access data using search functions in data manage menu
and easy search from [Main] screen.

 Data Reading through CCD Image Sensor


LED for strip reading: White
LED for SG reading: Red
LED for clarity reading: Red

 AC Power with 100-240V


AC 100-240V (50/60Hz)

 Communication Interface
Ethernet – direct connection to network
USB – external memory drive, etc
Parallel Communication - external printer
Serial Communication – host computer
PS/2 – barcode reader, keyboard, mouse, etc

Operator Manual / 8
URiSCAN Super+ Fully automated urine analyzer Rev.02

1.6 Caution

1) The instrument must be installed and used in the environment where meets the
following conditions:
 Free from excessive dust
 Well-ventilated area
 Not exposed to direct sunlight
 Flat and secured place
 Sturdy floor bearing the weight of this product(68Kg,149.91lb)
 Ambient temperature is ranging from 15 to 30°C (59-86°F), with a variation of
±2°C (±3°F) during measurement.
 Relative humidity should be ranging from 20% to 80%, without moisture
condensation.
 Free from sensible vibrations
 Use the power cable supplied together which length is about 3m(9,8ft)
 Place apart from a machine using high voltage
 Free from electromagnetic wave interference
 Free from abrupt power fluctuation (rated voltage ±10V max)
 Place apart from a machine generating a high frequency (centrifuge, electric
discharge machine, etc.)
 Space requirement
① Rear: longer than 20cm (7.9 inches)
② Left: longer than 50cm(19.7inches)
③ Right: longer than 50cm(19.7inches)

2) Power Supply
Operating instruction for AC 100-240V
1. Before using the device, please check if the local operating voltage complies
with AC 100-240V.
2. Plug the power connecting cable into the power input socket of the instrument
and connect the other end with power source.
*It is recommended to use the power cable supplied together. The power switch
and power supply socket are located in the right side.

3) Water Supply
The fresh distilled water should be stored in the water supply tank and used and its
conductivity is 1µS/cm or less.

Operator Manual / 9
URiSCAN Super+ Fully automated urine analyzer Rev.02

1.7 Symbols

Symbol Description

CE Mark

IN VITRO DIAGNOSTIC MEDICAL DEVICE

CONSULT INSTRUCTIONS FOR USE

BATCH CODE

SERIAL NUMBER

CATALOGUE NUMBER

BIOLOGICAL RISKS

CAUTION, CONSULT ACCOMPANYING DOCUMENTS

AUTHORISED REPRESENTATIVE IN THE EUROPEAN COMMUNITY

MANUFACTURER

DATE OF MANUFACTRUE

Waste Electrical and Electronic Equipment(WEEE) mark

Operator Manual / 10
URiSCAN Super+ Fully automated urine analyzer Rev.02

Chapter 2 Installation
2.1 Technical specification --------------------------------------------------- 12
2.2 Hardware ------------------------------------------------------------------- 13
2.3 Hardware installation ----------------------------------------------------- 18
2.4 Software installation ------------------------------------------------------ 20
2.5 Install the strip cassette ------------------------------------------------- 35

Operator Manual / 11
URiSCAN Super+ Fully automated urine analyzer Rev.02

2.1 Technical Specifications


◈ Manufacturer YD Diagnostics Corporation
◈ Type of product Urine chemistry analyzer
◈ Model Name URiSCAN Super+
◈ Materials Plastic & metal
◈ Size 551mm(W) x 687mm(D) x 535mm(H)
◈ Weight 68 Kg
◈ Rated Voltage AC 100-240[V]
◈ Rated Frequency 50/60Hz
◈ Rated Power 300 [VA]
◈ Operating Conditions
Temperature 15 ~ 30 ℃
Humidity 20 ~ 80%
◈ Storage Conditions
Temperature 5 ~ 40 ℃
Humidity 10 ~ 85%
◈ Measurement Method
Test type CCD color image sensor +
Clarity : Absorbance measurement
SG : refractive index measurement
Test velocity 200 tests/ hour
Test cycle 18 seconds
Memory Capacity 1,000,000 patient test results
500 quality control results
Sensor CCD color image sensor
Light Source LED
Result Display LCD color press screen (size 8.4 inches),
Resolutions: 800 X 600 (True Color)
◈ Interface Ethernet – direct connection to network
USB – external memory card, etc
Parallel communication - external printer
Serial communication – host computer
PS/2 – barcode reader, keyboard, mouse, etc

Operator Manual / 12
URiSCAN Super+ Fully automated urine analyzer Rev.02

2.2 Hardware

2.2.1 Hardware components


The major parts of the URiSCAN Super+:
 Embedded computer with dedicated software
 LCD press screen with GUI
 Barcode reader
 Pump device
 CCD color image sensor
 Automatic transporting system for test tubes
 Room for the test strip cassette
 Automatic test strip loading mechanism
 Automatic pipette system
 Automatic liquid handling system

Operator Manual / 13
URiSCAN Super+ Fully automated urine analyzer Rev.02

2.2.2 Layout and structure– exterior

2.2.2.1 Exterior

2.2.2.2 Part name

NO ITEM NO ITEM NO ITEM


1 [Main] TRAY 8 CASSETTE DOOR 15 BOTTOM CASE A
2 RACK GUIDE M 9 PC I/O PORT 16 RUBBER FOOT
3 FRONT TOP COVER 10 MIDDLE WINDOW 17 SYRINGE DOOR
4 DOOR HANDLE 11 FRONT TOP PANEL 18 RIGHT COVER DECO
5 REAR TOP COVER 12 LCD DECO 19 WASTE DOOR
6 COVER FAN 13 PRESS LCD 20 [Main] POWER SWITCH
7 LEFT COVER DECO 14 SAMPLE RACK 21 TUBE FITTING

[Table 2.2.1 Part Name ]

Operator Manual / 14
URiSCAN Super+ Fully automated urine analyzer Rev.02

Top view Left side view

Front view Right side view

Operator Manual / 15
URiSCAN Super+ Fully automated urine analyzer Rev.02

2.2.3 Layout and structure - description


1) Size : 551mm(W) x 687mm(D) x 535mm(H)
2) Weight : 68 kg

Cross View

Operator Manual / 16
URiSCAN Super+ Fully automated urine analyzer Rev.02

2.2.4 Standard packing list


 Uriscan Super+ equipment
 Liquid level sensor
 3 Air flow tubes (4φ, 6φ, 8φ)
 Cable tie (20ea)
 Rack (6ea)
 Rinse solution
 Power cable
 User manual
 Short operating manual
 Barcode Labels for QC / END / RINSE
 Gray Pad

*Water supply tank supplies distilled water into the Super+. The maximum volume to store
is up to 20 liter of D.W. which is sufficient to perform up to 1,000 tests.

2.2.5 Optional accessories


* The below optional accessories are not included in the standard package. These
items can be purchased separately.

 Water supply tank


 Waste water tank with/without optional overflow sensor
 Conical tube
 Uriscan Super+ strip cassette (Code # U800 for 13 parameters, U811 for
11 parameters)
 External thermal printer
 Serial interface(host) cable
 External barcode reader

Operator Manual / 17
URiSCAN Super+ Fully automated urine analyzer Rev.02

2.3 Hardware Installation


It is recommended to read carefully this operating manual before installation in order
to install and operate this instrument properly.

Warning: URiSCAN Super+ should not be installed where an ambient


temperature is ranged below 5’C or above 40°C, or where the relative humidity is
over 70%, without moisture condensation. The optimal atmospheric pressure is
between 900~1,600bar.

The instrument must be installed where flat and free from the
electromagnetic wave interference and vibrations. The floor should be
cleaned without excessive dust, bearing the weight of this product. please
do not attempt the shock while transporting. It is important to keep this
analyzer clean and ventilated. Please install the analyzer apart from high
frequency and electromagnetic wave interference where is free from abrupt
power fluctuation. It is recommended to pull out the power plug after use.

2.3.1 Installation of each part

1) Setup the water supply tank with distilled water


It is to clean the operating parts of the instrument. It should store up to 20
liter (21 quarts)

▶ Installing the water supply tank


1. Fill up with D.W. (distilled water).
2. Connect the liquid level sensor with the terminal of the water level
sensor..
3. Connect water supply tube with the inlet connector.
4. Install the airflow tube into the water supply tank and fix it when the
end of the airflow supply tube reaches bottom of the water tank in
order to consume all distilled water in the tank.
5. Install the liquid level detector. The level detector with rubber
stopper should align vertically to the tank.

2) Liquid waste tank

The liquid waste tank is located outside the analyzer. The capacity of the
tank depends on the specific tank which is used. The tank must have
volume of at least 20 L (21 qts). The overflow sensor is available as an
option. 5 meter (16.4 ft) liquid waste tube is supplied for the connection of
the liquid waste tank to the analyzer.

Operator Manual / 18
URiSCAN Super+ Fully automated urine analyzer Rev.02

 How to set up the liquid waste tank


1. Ensure the liquid waste tank is empty before use. The liquid waste tank
must be positioned lower than the level of the analyzer. It is for reliable
flow of the liquid waste into the waste tank.
2. Connect the liquid waste tube to the drain socket of the instrument. If
necessary, shorten the liquid waste tube to ensure smooth waste liquid
flow in to the waste tank.
3. If an overflow sensor is used, connect it to the drain level connector on
the analyzer and install it into the liquid waste tank.

3) Guide plate & strip waste tray

1. Open the door located in the right side of instrument


2. Insert and push the guide plate until it reaches the end point.
3. Place the strip waste tray properly in its position

2.3.2 External Printer

An external printer can be connected via standard parallel interface port or


USB port. A network printer can be used after connecting Ethernet and
installing proper printer driver.

2.3.3 External host connection


The interface program installed in the instrument can transfer data to the
host PC or LIS (via host PC). Also a serial interface using RS232
communication is possible.

Operator Manual / 19
URiSCAN Super+ Fully automated urine analyzer Rev.02

2.4 Software Installation

2.4.1 Software Installation


The software installed at the instrument controls all functions of URiSCAN Super+.
Two types of modes (i.e. User, Admin) can be applied in accordance with user’s
authority and purpose. The authorization level for accessing the software program
can be decided by the authorized service engineer.

2.4.1.1 How to set up


The English software is pre-installed as default language. The HDD(hard disc
drive) includes an image file for back-up/restore utilities, and this can be used in
the case of the critical system error of windows OS (the user will be able to reset
the system using this program). The updated version of software system will be
available from YD Diagnostics’ website. (http://www.yd-diagnostics.com) For more
details, contact the authorized YD Diagnostics’ local distributor.

2.4.1.2 Press Screen


The press screen is designed to select and operate each menu. The
characters can be input by the digital keyboard (which appears on the screen) or
by external keyboard and mouse. Operator can easily use various functions
through this friendly GUI. It is also designed for operators those who wore
rubber gloves.

[ Fig. 2.1] Main Screen

Operator Manual / 20
URiSCAN Super+ Fully automated urine analyzer Rev.02

2.4.1.3 Display Status

1) Display

Status : [Ready] ready to test, [Pause] sampling stop, [Continue] continue to


test the sample after [Pause], [STOP], [Change Cassette], [Routine] testing
routine sample, [EMER] testing emergency sample, [QC test] testing quality
control material

Remained Strip : number of remaining strip in the cassette is shown on the


screen. This number is deducted by the number of test, and it automatically
resets when changing the strip cassette.

Next Routine Serial# : the routine test sequence number starts from 0001 to
9999. It is automatically increased by the number of routine test.

Next Emergency Serial# : the emergency test sequence number is shown


from 0001 to 9999. It is automatically increased by the no. of emergency test.

Next Rack Serial# : the rack sequence number is shown from 01 to 99. It is
automatically increased by the number of rack loaded for testing. It is
automatically reset the number by pressing [Start] button on the [Main] screen.

Next Tube Position Serial# : this indicates the position number of tube in the
rack. The maximum of 10 tubes (positions) can be loaded so that it indicates
from 1 to 10.

Operator ID : show operator information.

2) Result Table

The test result is shown with the image output of tested strip. Test parameters,
real picture of developed color, result and raw data (maintenance mode) are
shown on screen from the latest result. Press [Prev] and [Next] button for real
time search option.

3) Queue Process

This indicates the test procedure with the sample ID number.


Processing procedure : sampling/No Sample/[Pause] → Dispensing/Row
change → Wait reaction → Image processing

Operator Manual / 21
URiSCAN Super+ Fully automated urine analyzer Rev.02

2.4.1.4 Menu

Option
M
Manage Data
E

N ID Register

U Range Table Quality Control Config

Quality Control QC Management Program

Calibration Reset

[Main]tenance Air Purge

Mode Change Rinse

Error History

1) Option
[Fig. 2.2]

*Important Notice
The following options can be selected or adjusted only in [Admin Mode]. Go to page #
33~34 of this manual [Admin Mode] for detailed instruction before starting the test.
1.Language select 2. Standalone 3. Test order 4. Strip select

Operator Manual / 22
URiSCAN Super+ Fully automated urine analyzer Rev.02

1. Press the [Menu] button at [Main] window.


2. Press the [Option] button to see [Option] window, this allows the users
to set options for Serial Number, Mixing, Rinse, Print etc.
3. Following are 4 options for setting serial#. User can set operator ID.
 Serial Number Auto Reset
 Next Routine Serial #
 Next Emergency Serial #
 Next Control Serial #
- Serial Number Auto Reset: press [Change] button to select
Normal or Auto Daily Reset mode.
- Next Routine Serial #: the serial number is selectable from 1 to
9999. The serial number is reset to 0000 by pressing [Reset]
button. User also can input certain number by pressing
[Manual] button.
- Next Emergency Serial # : Same as Routine Serial #.
- Next Control Serial # : Same as Routine Serial #.
- Operator ID : Press [Change] button to input the Operator’s ID.
4. Representative unit: select the unit(SI unit, or CONV.unit) by pressing
[Change] button.
5. Printer Driver: select the connected printer. The BIXOLON STP-103II
should be selected when the thermal and dot printer are parallel
connected.
6. Result Print: press [Change] button to set printing option for
patient data.
7. Calibration Print: press [Change] button to set printing option for
calibration data.
8. Sample Mixing: press [Change] to set sample mixing option.
9. Rinse Counter: press [Change] to select washing times from 1 to 5
times, or [Until Empty].
10. QC Test Tube Barcode: this is to identify the control tube by inputting
Barcode number.
11. Rinse Tube Barcode: this is to identify Rinse tube by inputting
Barcode number.
12. End Tube Barcode: this is to identify the end tube by inputting
barcode number.
13. Water Supply Tank Alarm : press [Change] button to
On/Off the water level sensor
14. Waste Liquid Tank Alarm: press [Change] button to On/Off
the water level sensor.
15. NUM. of Strip for Tray-Full Warning : press the [Change] button to set
up alarm for remaining strips.
16. NUM. of Strip for Few-Strip Notice : press the [Change] button to
modify the quantities of remaining strips.
17. Control data auto transfer: press [Change] button to set up auto
transfer for QC data Serial (RS232C).

Operator Manual / 23
URiSCAN Super+ Fully automated urine analyzer Rev.02

2) Manage DATA
[ Fig. 2.3 ] [ Fig. 2.4 ]

This menu provides the functions to search, print, delete, transfer all
measurement results. It allows users to export data to the external memory
devices.

1. Press [New Search] button in [Fig. 2.3] to display [Fig. 2.4], [Input
Search Condition] window. User can change the condition for below
parameters.
Type: select desired type: [Routine], [Emergency], [QC Test] or [All].
Date(Start): select the first date.
Date(End) : select the last date.
Serial #(Start) : select the start number from 0001 to 9999 for serial
number area.
Serial #(End) : select the start number from 0001 to 9999 for serial
number area.
ID # : select the ID number.
Flag : this is to do conditional search for the test results: [All], [Normal],
[Abnormal], [Sieve], [Error]. User can set the ranges for abnormal and
sieve in the range table.
2. Press [Recent Data] to list up the latest data(100 samples)
3. The searching results will be displayed on the [Data List] window.
Review the selected sample result with its color development by
pressing [Image View]button.
*At the [Image View], user can see other results by pressing [Previous]
or [Next] button as well as print-out by pressing [Print] button.

Operator Manual / 24
URiSCAN Super+ Fully automated urine analyzer Rev.02

[ Fig. 2.5 ]

4. User can save all data or the selected data in external device, print,
delete, transfer to interface by pressing [Export], [Transfer], [Print] and
[Delete]

[ Fig. 2.6 ]

3) ID Register
[ Fig. 2.7 ] [ Fig. 2.8 ]

This is to register the ID number (sample number) before measurement.


1. Add: input ID number by using numeric keys or barcode reader.
2. Insert: insert ID number into specific spot.
3. Delete One: delete the ID number one by one.
4. Delete Range: delete the ID numbers together as selected.
5. Delete All: this is to delete all registered ID

Operator Manual / 25
URiSCAN Super+ Fully automated urine analyzer Rev.02

4) Range Table

[ Fig. 2.9 ]

This window shows the information about each test item, the contents are
as follows.
1. Grade : concentration of each grade
2. Conv / Unit : unit of result (Conversion unit) ex) mg/dL
3. SI / Unit : unit of result (International system of unit) ex) mmol/L
4. Range Low, High : the min. and max. raw data of each grade.
5. Abnormal : if the blank is marked with a tick, the marked grade is
displayed as abnormal data.
6. Sieve: if the blank is marked with a tick, the marked grade is
displayed as [Sieve Data].

YD standard which is default value cannot be changed. Press [Change] to


adjust the range table if user wants to change it.

Operator Manual / 26
URiSCAN Super+ Fully automated urine analyzer Rev.02

5) Quality Control
Go to [Menu] → [Quality Control] to set QC(quality control).
At the [Quality Control] window, there are two menus: [Quality Control
Config] & [QC Management Program].
The procedure is the same as menu 2.4.5, 2)[Manage Data].

[ Fig. 2.10 ]

1. Quality Control Config.: this is to manage the setting of control material


[Fig. 2.11], to reset name, num. of level, lot number, expiry date, and set
acceptable grade of control material for each level.
2. QC management program: this is to manage or to make a chart for
control data [Fig 2.12].

Operator Manual / 27
URiSCAN Super+ Fully automated urine analyzer Rev.02

[Fig. 2.11]

[Fig. 2.12]

Operator Manual / 28
URiSCAN Super+ Fully automated urine analyzer Rev.02

6) Calibration

[ Fig. 2.13 ] [ Fig. 2.14 ]

[ Fig. 2.15 ] [ Fig. 2.16 ]

[ Fig. 2.17 ] [ Fig. 2.18 ]

[ Fig. 2.19 ]

Operator Manual / 29
URiSCAN Super+ Fully automated urine analyzer Rev.02

It is to check that the raw data of the strip for calibration whether it is within the
normal range (page 47) and to reset the standard value for each test portion.

1) Go to [Main] → [ Menu ] → [Calibration]


2) Press [Change] to input lot no. of strip cassette as shown in [Fig
2.15]
3) Press [Strip Calibration], and place a sample rack with DW sample
in STAT position as shown in [Fig. 2.18].
4) Press [Strip Calibration] on the screen
5) Press [OK] button after displayed [Notice] and do calibration
6) After calibration, [Calibration Data List] will be displayed on the left
side of the table, and the result will be appeared on the right side
[Fig 2.19]
7) Calibration results can be printed or deleted

6.1) Calibration result


a) After calibration, the result is listed on the screen. Also the image of
selected data can be reviewed at the right side of screen.
b) The data list consists of date, lot#, flag, history and expiration date. The flag
is shown (E) if the calibration data is out of range. The latest & right data is
used as calibration value.
c) Expiration date shows the expiration date for calibration. If the date has expired,
the expiration date is indicated in red.

Calibration Range
BLD: 500~515, BIL: 500~515, CRE: 500~515, URO: 500~515, KET: 500~515,
PRO: 500~515, ALB : 500~515, NIT: 500~515, GLU: 500~515, pH: 500~515,
LEU: 500~515, VTC: 500~515, COL: 500~515

Operator Manual / 30
URiSCAN Super+ Fully automated urine analyzer Rev.02

7) Maintenance

[ Fig. 2.20 ]

The followings are the menu of [Maintenance] at the user mode.


*The menu of [Maintenance] is different from each mode.

 Reset : this is to initialize or reset the operating status.


 Air Purge : this is to remove the air bubble in the syringe.
 Rinse : this is to wash tube and pipet with washing solution.
 Error History : this is to review the date, type, error number,
message, etc about error.

[ Fig. 2.21 ] [ Fig. 2.22 ]

Operator Manual / 31
URiSCAN Super+ Fully automated urine analyzer Rev.02

[ Fig. 2.23 ] [ Fig. 2.24 ]

8) Mode Change

[ Fig. 2.25 ] [ Fig. 2.26 ]

[ Fig. 2.27 ] [ Fig. 2.28]

Operator Manual / 32
URiSCAN Super+ Fully automated urine analyzer Rev.02

URiSCAN Super+ is available in two modes. – [User] and [Admin mode]

Two operating mode are available which are [User] &[ Admin mode]. The
default setting is [User mode]. When changing [User mode] to [Admin
mode], press [Admin Mode] and input the password. To change the
password, press [Change Admin mode] in admin mode and input the new
password. At [Admin mode], user can modify the stored result, select the
language which set in [Option] menu and set the analyzer type, strip order
and strip type.

1) Modify test result

[ Fig. 2.29] [ Fig. 2.30]

2) Language: user can select the language by pressing [Change]

[ Fig. 2.31] [ Fig. 2.32]

3) Stand alone: user can select [Yes](working alone) / [No](working with


other analyzer) by pressing [Change]
4) Test order: user can change the test order. The user needs to create
the name of order and then change the order.

Operator Manual / 33
URiSCAN Super+ Fully automated urine analyzer Rev.02

[ Fig. 2.33]

5) Strip select : user can select either 10 parameter strip (U81) or 12


parameter strip including microalbumin and creatinine tests (U80) by
pressing [Change]

[ Fig. 2.34] [ Fig. 2.35]

Operator Manual / 34
URiSCAN Super+ Fully automated urine analyzer Rev.02

2.5 Cassette Installation

2.5.1 Unpack and install the cassette


The cassette is wrapped in polyethylene-coated aluminum bag, which is tightly placed
in an outer package box. This secure packaging ensures shelf life of the strip cassette
in unopened package as printed on the package box.

The test strips are stable in the tightly closed cassette compartment for fourteen
days. After this time, a warning window is displayed on the screen, and the cassette
must be replaced by a new one.

How to replace the cassette are as follows:

1) Press [Change Cassette] button on the [Main] window. [Fig. 2.36]


2) Display [Change Cassette OK?] on the screen and press [Yes] button to
change the cassette as shown in [Fig. 2.37]
3) When the [Open the Cassette Door and press YES] message appears, open
the cassette room door and press the YES button. The cassette will come out
from the cassette room. [Fig. 2.38]
4) When the [Change the Cassette and press YES] message appears, input the
new cassette and press the YES button. The cassette will be moved into the
cassette room as shown in [Fig. 2.39]
5) When the [Close the Cassette Door and press YES] message appears, close
the opened door of the cassette room and press [Yes] button to complete the
change of the cassette. [Fig. 2.40]

[ Fig. 2.36 ] [ Fig. 2.37 ]

Operator Manual / 35
URiSCAN Super+ Fully automated urine analyzer Rev.02

[ Fig. 2.38 ] [ Fig. 2.39 ]

[ Fig. 2.40 ] [ Fig. 2.41 ]

2.5.2 Precautions when using the cassette


When using the strip cassette, follow the precautions as follows:

1) Only use the strip cassette for the Super+ which has not exceeded its expiration
date as printed on the label.
2) Store the strip cassette at the temperature between 2-30°C (36-86°F).
3) Check the package for damage before use.
4) After opening the cassette from the package, check whether the test strip layers
are correctly aligned and the color of pads is OK to use.
5) Do not directly touch the test pads on the test strip.
6) Do not let the strips come in contact with water or with any volatile, strongly
acidic, or alkaline chemicals.
7) Install the cassette into the analyzer within 3 minutes after opening the aluminum
bag.
8) Keep the cassette in the firmly locked cassette compartment until the test strips
are completely used up.

Operator Manual / 36
URiSCAN Super+ Fully automated urine analyzer Rev.02

9) Do not open the cassette compartment door as long as the cassette is not empty
or its on-board stability has not expired.
10) Follow the instruction for use provided with the URISCAN Super+ cassette and
read carefully all information in the package insert.

Operator Manual / 37
URiSCAN Super+ Fully automated urine analyzer Rev.02

Chapter 3 Operation
This chapter describes how to operate the URiSCAN Super+

3.1 Quick Manual -------------------------------------- 39


3.2 Routine Operation --------------------------------- 40
3.2.1 Initial preparation ------------------------ 40
3.2.2 System check ----------------------------- 40
3.2.3 Sample preparation and measurement start --------------- 42
---- 3.2.4 End sample measurement ------------------------- 44
3.2.5 Additional sample test ------------------------- 45
3.2.6 Testing emergency sample -------------------------- 46
3.2.7 Change the strip cassette ---------------- 47
3.2.8 Quality control test ----------------------------- 48

Operator Manual / 38
URiSCAN Super+ Fully automated urine analyzer Rev.02

3.1 Quick Manual


Initial Preparation
Check the water level of distilled water in the supply tank & liquid waste tank, check the strip
waste tray, and clean the guide plate
Switch ON / System Checking
Turn the power switch on, then embedded Windows XP is executed and the operating program
runs automatically. When the program runs, the analyzer performs system initialization.
Calibration (Main→ Menu→ Calibration)
Input Lot number of test strip cassette ▶ Prepare the sample tube filled up D.W. at the STAT
position ▶ Run strip calibration by following the steps indicated in the calibration menu.
ID Management (Main→ Menu→ ID Register)

The built-in barcode reader automatically scans the barcode of each sample. Also, the
specimen ID can be manually registered by pressing LCD screen, or using a keyboard or an
external barcode reader.

Measurement (Main→ Start)


Control measurement: take a measurement using the control which is registered before
measuring specimens
Specimen measurement: the maximum of 10 racks can be loaded, and the additional rack can
be loaded consecutively. In addition, the emergency specimen can be tested by pressing
[Emergency] button during the measurement or ‘Ready’ status.
Data Management (Main→ Menu→ Data Manage)
The measurement result can be monitored together with the direct visual image. The data can
be printed, deleted, trasnfer and stored.
Maintenance
Rinse: rinse the probe arm (pipette) and inner tube everyday for daily maintenance. Clean and
empty the guide plate, waste water tank and strip waste-tray.
Change Cassette (Main→ Change Cassette)
Change Cassette ▶ [Change Cassette OK] and press [Yes] ▶ [Open the Cassette Door] and
press [YES] ▶ Open cassette room and press [Yes] ▶ the cassette eject from the cassette
room.▶ Change the cassette and press [YES] : Input the cassette and press [Yes] ▶ The
cassette is inserted into the cassette room ▶ Close the cassette door and press [YES]

Operator Manual / 39
URiSCAN Super+ Fully automated urine analyzer Rev.02

3.2 Routine Operation

3.2.1 Initial preparation


Users must check below before turning the power switch on:
1) Fill up the D.W in water supply tank and adjust the position of the liquid level
detector.
2) Empty the liquid waste tank.
3) Check the guide plate is in the right position.
4) Empty the strip waste tray and return it in the right position.

3.2.2 System check


 Power switch is at the right side of the instrument and Off mode is indicated as [O]
1) When the power switch is on, embedded Windows XP is executed and the
operating program runs automatically.
2) After the program runs, [Main] screen appears as shown in [Fig. 3.1] and runs
initialization.

[ Fig. 3.1 ]

Operator Manual / 40
URiSCAN Super+ Fully automated urine analyzer Rev.02

[ Fig. 3.2 ]

1) Check point before measurement


[Routine] & [Emergency] serial number on [Main] screen can be changed at
[Option] menu

[ Fig. 3.3 ] Option Menu

Operator Manual / 41
URiSCAN Super+ Fully automated urine analyzer Rev.02

1. Serial Number Auto Reset : under the normal mode, the sequential
identification number is automatically incremented with each sample.
under [Daily Auto Reset], the sequential identification number is
automatically reset every single day
2. Routine / Emergency / Control Serial #
The serial number can be set for each mode.
3. Operator ID : input the operator ID
4. Representative unit : set unit (SI unit, CONV.unit)
5. Printer Driver : select the connected printer. The BIXOLON STP-103II
should be selected when the thermal and dot printer are parallel connected.
6. Result Print : set the print option for patient results
7. Calibration Print : set the printer option for calibration results
8. Sample Mixing : set the specimen mixing option
9. Rinse Counter : set the number of rinsing from 1 to 5 times, or until empty.
10. QC test tube Barcode : set the control tube by inputting barcode number.
11. Rinse Tube Barcode : set the rinse tube by inputting barcode number.
12. End Tube Barcode : set the end tube to finish the test by inputting
barcode number.
13. Water Supply Tank Alarm : press the change button to On/Off the water
level sensor.
14. Waste Liquid Tank Alarm : press the change button to On/Off the water
level sensor.
15. NUM. of strips for Tray-full warning : press the change button to
modify the quantities of remaining strips.
16. NUM. of strips for few-strip Notice : press the change button to set up
alarm for remaining strips.
17. Control data auto transfer : press the change button to set up auto
transfer the quality control data(RS232C) to an interface program.

3.2.3 Sample preparation and measurement start


1) Preparation of Sample Rack and Sample Tube
Use the standard 10-position rack provided with the URiSCAN Super+ and any
tubes which meet below specification. Below is the specification of
recommended sample tube type.
 Diameter : 13 – 16.5 mm
 Height : 70 – 120 mm
 Using the round bottom tubes with minimum 2.0 ml of urine sample
should be filled.
Using the corn shaped tubes with minimum 1.5 ml of urine sample
should be filled.
If the barcode label attached on the sample tube, align the label with the
barcode reader.

Operator Manual / 42
URiSCAN Super+ Fully automated urine analyzer Rev.02

2) Installation of sample rack


User can load the maximum of 10 racks in the Input line.
The position #1 of the rack should be positioned at the left side
Install the rack inside of the hook which guides the rack to its input line

[ Picture 3.1 Sample loading]

3) Start measurement
Check the expiry date of the installed cassette before starting the measurement.
※ Note: the strip cassette is stable for 14days after installing in the instrument. If the
expiry date of the cassette has passed, the alarm sign automatically pops up.

Press [Start] button on [Main] screen in order to start the test.


Status changes from [Ready] to [Routine Test].
Sample rack moves from Input line to cross-feed line. During the rack is moving
to the dispensing area, the barcode label will be automatically scanned.

4 cm

Range : 6 cm

2 cm

[Picture 3.2 ] Barcode labeling

Operator Manual / 43
URiSCAN Super+ Fully automated urine analyzer Rev.02

[Picture 3.3 ] Adjust Barcode label

Once the strip is loaded from the cassette, the strip sensor detects the existence
of strip, and the routine or emergency serial number is incremented one by one.
The number of remaining strip is decreased one by one. The processing queue
on [Main] screen shows the operating status of each sample. The measurement
process is as follows:

Sampling / No Sample / Pause → Dispensing / Row change → Wait for


reaction → Image processing

The test result of processed sample shows on left side of [Main] screen with the
image output.
Pressing [Prev] and [Next] button enables user to search the previous and next
result.

3.2.4 End sample measurement

1) End measurement
If there is no sample rack is detected by the rack input sensor of input line, the
analyzer will be changed to [End Routine] to finish its measurement after
finishing the aspiration for the last test strip. When the 10 racks are loaded on
the output line after measurement, alarm message will appear by the rack
output sensor, and automatically changes to [Pause].

2) Emergency stop measurement


When user press [Pause] button from [Main] screen, the instrument will go to
sampling stop mode and then the sampling test will be stopped. The samples
which have not finished aspiration process will be stopped at the same place.

Operator Manual / 44
URiSCAN Super+ Fully automated urine analyzer Rev.02

The samples which are finished the Aspiration will go to next step, sample
dropping, and measurement will be continued. [Pause] means that the sample
rack stops and the test strips on which the samples are already dropped
continue measurement.

3) Restart measurement
Press [Continue] button to run test. The measurement process will restart.

4) Stop measurement by pressing [Stop] button


Press [Stop] button and the rack moves on the cross-feed line to the output
line and measurement will be stop after testing up to the last aspirated sample
on the guide plate.

5) Stop measurement by using [End Tube Barcode]


When the analyzer reads the barcode on the tube which matches with entered
[End Tube Barcode] in [Option] menu, the measurement automatically stops.

3.2.5 Additional sample test

The sample can be tested additionally by [Routine Test], [Pause], [Ready mode].

1) Routine Test Mode


* Place additional sample rack in the rack input line.
* The sample rack moves from cross-feed line to rack output line. When rack
input sensor detects rack, one of sample rack which attached moving link will
be moved to cross-feed line.

2) [Pause] Mode
* Place additional sample rack in the rack input line.
* Press [Continue] on [Main] screen, [Pause] mode is changed to operation
mode and measurement mode is restarted. The sample rack is moved to
cross-feed line.

3) Ready Mode
* Prepare the additional sample rack in the rack loading line.
* The test procedure is the same as starting the measurement.

Operator Manual / 45
URiSCAN Super+ Fully automated urine analyzer Rev.02

3.2.6 Testing emergency sample

Place the sample in emergency area and press [Emergency] button on [Main]
screen. The emergency sample will be tested first while the routine sampling
(measurement) is performing.

The routine sampling test will be stopped if you press [Emergency] button.
The specimen ID can be input by manually or using the connected barcode reader.
The routine test is started continually once the measurement is finished

1) Testing urgent sample


a) Place the urgent sample on the position #1.
b) The position# 1 should be faced to probe and place the sample on
emergency position and press [Emergency] button.
c) Input the specimen number by manual or barcode reader when the [Input
Emer. ID] is displayed.
d) When the emergency test is completed, the routine test will be started.
e) One urgent sample can be tested for each time.

[ Picture 3.4] Emergency sample

Operator Manual / 46
URiSCAN Super+ Fully automated urine analyzer Rev.02

3.2.7 Change URiSCAN Super+ Cassette

The Instrument will have alarm message and go to [Pause] mode if strip runs out
while the testing or the stability (14 days) has been passed. In this case, the
instrument will automatically display [Change. Cassette] mode and you need to
change new strip cassette. The alarm message will appear when the remaining
strip number is smaller than the set amount in NUM of strips for few-strip notice in
[Option], so user needs to prepare the new strip cassette.

1) Change used cassette:


a) Press [Chg.Cassette] button on the [Main] screen.
b) Displays [Change Cassette OK?]. Press [OK] buttons to change the
Cassette.
c) When the [Open the cassette door] message appears, open the cassette
room door and press [OK] button then the used cassette will come out from
the cassette room.
d) Remove the old cassette and insert the new cassette when [Change the
Cassette] message shows up. Once the new cassette is inserted, press
[Yes] buttons.
e) When [Close the Cassette Door] and [Press YES] message show up,
close the door and press [Yes] button, and then the cassette will be
automatically moved into the analyzer
f) Once the change is completed, press [Continue] to test continually. If the
cassette lot is changed or need to perform calibration, do calibration.

[ Picture 3.5 ] Change cassette

Operator Manual / 47
URiSCAN Super+ Fully automated urine analyzer Rev.02

3.2.8 Quality control test

To run QC(Quality Control) go to [Main]→ [QC test].

1) QC Test

The processing of measurement will be same as [Routine test] mode. Load the
quality control materials on the Rack as the number of control level. In
URiTROL case, place the Level 1(Normal) on the tube position 1, the Level
2(Abnormal) on the tube position 2, and Level 3(Abnormal) on the tube position
3. Match the number of tube position and quality control level, the quality
control test results will be shown and stored with its image.

[ Fig. 3.4 ]

2) Quality Control
Control Data. (Same as [Manage DATA] menu)
Set in [Menu] → [Quality Control]

a) New Search : Review the Data. The options are same as follows:
Date (Start) – search by date (from the start date)
Date (End) – search by date (from the end date)
Serial # (Start) – search by serial number (from the start #)
Serial # (End) – search by serial number (from the end #)
ID No. – search by control ID

Operator Manual / 48
URiSCAN Super+ Fully automated urine analyzer Rev.02

Flag – search by All, Normal, Abnormal, Sieve or Error result

b) Recent Data : review the recent 100 data


c) Image View : shows the data image and result
Previous – review previous data
Next – review the next data
Print – print out the data
Close - close [View Strip]
Export : export the data to external USB memory drive.
d) Transfer: send data to Serial port.
e) Print : print out the selected data
f) Delete : delete the data

[ Fig. 3.5 ] [ Fig. 3.6 ]

[ Fig. 3.7 ]

Operator Manual / 49
URiSCAN Super+ Fully automated urine analyzer Rev.02

3) Quality Control Config

Set the material information of quality control or mange the allowed acceptable
range for each items.
a) Name – set the brand/model name of quality control materials (e.g:
URiTROL, Bio-Rad)
b) Num. of Level – set the number of Quality Control level and its quantity
(eg: URiTROL has 3 different levels)
c) Lot # – set the LOT number of Quality Control materials
d) Expiration Date – set the expiry date of Quality Control materials.
e) Test item – set the allowed test limitation for each items and levels.

4) QC management program

Recall [Control Data] management / statistics program.

[ Fig. 3.8 ]

[Control Management] program shows up the control data via report type and chart
type, and it can be printed out to default printer (A4 size only).

Operator Manual / 50
URiSCAN Super+ Fully automated urine analyzer Rev.02

Chapter4 Maintenance

4.1 Periodic maintenance -------------------------- 52


4.1.1 Daily maintenance ----------------------------------- 52
4.1.2 Weekly maintenance ----------------------------------- 52
4.1.3 Monthly maintenance ----------------------------------- 52
4.1.4 Quarterly maintenance ----------------------------------- 52
4.2 Maintenance menu ------------------------------ 53
4.2.1 Reset ---------------------------------- 53
4.2.2 Air purge ------------------------------- 53
4.2.3 Rinse ---------------------------------- 53
4.2.4 Initialization ------------------------------ 53

Operator Manual / 51
URiSCAN Super+ Fully automated urine analyzer Rev.02

4.1 Periodic Maintenance


URiSCAN Super+ requires the minimal maintenance. Handle the instrument with care.
Keep the exterior of the instrument free of dust and protect from contaminant such as
DW and urine by wiping periodically with a damp cloth and mild detergent. Do not use
any kind of spray cleaners since the aerosol may enter and damage the optical
system.
The cassette door and strip waste tray door must be closed all times, and the interior
of analyzer should not be exposed to the outside. In addition, the periodic
Maintenance is required to protect the analyzer from failure and maintain the reliability
of the test result.

4.1.1 Daily maintenance


The daily maintenance is as follows:

1) The sample pipette and internal tube should be rinsed once a day by Rinse
additive.
2) Empty the strip waste container.
3) Fill up the distilled water in the water supply tank.
4) Check the waste liquid tank

4.1.2 Weekly maintenance


The weekly maintenance is as follows:

1) Clean the guide plate. If it has contaminated with urine sample, it needs to be
cleaned.
2) Clean the rack transfer base. It must be free from dust and urine contaminant.
3) Check whether or not there is an infection of bacteria or algae in the water
supply tank (distilled water).

4.1.3 Monthly maintenance


The monthly maintenance are as follows:

1) Clean the housing of analyzer and press screen LCD.


2) Clean the cooling fan filter on left and right side of the analyzer.

4.1.4 Quarterly maintenance


The Quarterly maintenance is as follow:
1) Check the syringe seal.

Operator Manual / 52
URiSCAN Super+ Fully automated urine analyzer Rev.02

4.2 Maintenance menu


Press [maintenance] button on the [Menu] to go [maintenance].
There are four options – [Reset], [Air Purge], [Rinse], [Error History] on the
[Maintenance] menu. Each option performs the following function.

4.2.1 Reset
It is to initialize the operation. Pressing [RESET] button removes the rack of cross-
feed line. Locate the pipette in the right position and makes the syringe measure
mode.

4.2.2 Air purge


This option is to refill the water supply tank with distilled water and also to remove
the air bubble in the syringe. You can remove air bubble in syringe, tube and needle
with this option.
Perform it repeatedly until air bubble is removed completely

4.2.3 Rinse
At the end of each working day, the sample needle and inner tube must be rinsed in
order to protect from the contamination by urine specimen.

▶ To rinse at the emergency position;

1) Place a tube with rinsing solution at the position No.1 of the rinse rack
2) Select the number of repetition for rinse on [Menu] → [Option].
3) Place the Rinse rack at the emergency position. (※ The position No.1 of the
rinse rack, where the tube with rinsing solution is placed, should face toward
the pipette.)
4) Press [Rinse] button on the [Menu] → [Maintenance].
5) The system performs rinsing.
6) The Rinse rack should not be removed during the job.

4.2.4 Error History

You can check the details of error in the past: date, classification, number, etc

Operator Manual / 53
URiSCAN Super+ Fully automated urine analyzer Rev.02

Chapter 5 Trouble shooting

5.1 Alarm Message ------------------------------------------------------------ 55

Operator Manual / 54
URiSCAN Super+ Fully automated urine analyzer Rev.02

5.1 Alarm Message


5.1.1 WARNING
CODE Message Solution

Warning 02-003 Please check the Waste-box installation! Place Waste-box

Warning 03-003 Please check the Strip-sensing area! Check Strip-sensor.

Warning 04-006 Please check the Waste-tank liquid level! Check the waste-tank

Check the sensor in Rack-input


Warning 06-002 Please check the Rack-input tact sensing area! Forward.

Check the sensor in Rack-input


Warning 07-006 Please check the Rack-Full sensing area! backward.

Please check the Rack-input approach sensing Check the sensor in Rack-input
Warning 07-007 area! Forward.

Check the amount of D.W. in the


Warning 09-006 Please check the DW-tank liquid level! D.W. tank.

Warning 16-002 Waste-tray Full! Empty Waste-tray.

Warning 16-007 Take out the Waste-box and dump trash! Empty Waste-tray.

Warning 16-008 Install Waste-box! Place Waste-box.

5.1.2 ERROR

CODE Message Cause & Solution


Communication Error
ERROR 00-000 Serial Communication ERROR!! Contact Technical support team of YD

When Motor does not reach the set


ERROR 00-009 Motor moving time out!! position in time.
Contact Technical support team of YD.
When Input command is not executed in
ERROR 00-012 Command time out!! time.
Contact Technical support team of YD

Pipette Dispensing_position When Parameter value is not same as the


ERROR 01-001 Set value ERROR!
value in PC.
Contact Technical support team of YD

Pipette Moving_position When Parameter value is not same as the


ERROR 01-002 Set value ERROR!
value in PC.
Contact Technical support team of YD

Pipette Washing_position When Parameter value is not same as the


ERROR 01-003 Set value ERROR!
value in PC.
Contact Technical support team of YD

Pipette Bottom-limit_position When Parameter value is not same as the


ERROR 01-004 Set value ERROR!
value in PC.
Contact Technical support team of YD.
When Pipette does not reach the position
ERROR 01-005 Pipette Top-Goal_position ERROR! of Top sensor, or faulty sensor. Contact
Technical support team of YD.

Operator Manual / 55
URiSCAN Super+ Fully automated urine analyzer Rev.02

CODE Message Cause & Solution


When Pipette Crash sensor has a
ERROR 01-006 Pipette Crash_sensor ERROR! problem.
Contact Technical support team of YD.

ERROR 01-007 Pipette Level-detect_sensor ERROR! Contact Technical support team of YD.

When Pipette does not move to the set


ERROR 01-008 Pipette Goal_position ERROR! value.
Contact Technical support team of YD.
When Pipette detects the substance at
ERROR 01-009 Pipette Waste pot overflow ERROR! Waste Position
Contact Technical support team of YD.
When Pipette does not detect the
ERROR 01-010 Pipette Washing pot empty ERROR! substance at Washing Position.
Contact Technical support team of YD.
When Belt does not move to the location
ERROR 02-001 Belt Goal_position ERROR! set with value.
Contact Technical support team of YD.
When Belt does not reach to the location
ERROR 02-002 Belt Moving ERROR! of Belt Sensor.
Contact Technical support team of YD.

Feeder Ready_position When Parameter value is not same as the


ERROR 03-001 Set value ERROR!
value in PC.
Contact Technical support team of YD.
When Feeder does not reach Right sensor
ERROR 03-002 Feeder Right-Goal_position ERROR! properly, or sensor problem.
Contact Technical support team of YD.
When Feeder does not reach Left sensor
ERROR 03-004 Feeder Left-Goal_position ERROR! properly, or sensor problem..
Contact Technical support team of YD.

Elevator Start_position When Parameter value is not same as the


ERROR 04-001 Set value ERROR!
value in PC.
Contact Technical support team of YD.

Elevator Ready_position When Parameter value is not same as the


ERROR 04-002 Set value ERROR!
value in PC.
Contact Technical support team of YD.

Elevator Top_position When Parameter value is not same as the


ERROR 04-003 Set value ERROR!
value in PC.
Contact Technical support team of YD.
When Elevator does not reach Bottom
ERROR 04-004 Elevator Bottom-Goal_position ERROR! sensor, or sensor problem.
Contact Technical support team of YD.
When Elevator does not reach Top sensor,
ERROR 04-005 Elevator Top-Goal_position ERROR! or sensor problem.
Contact Technical support team of YD.

Cassette Out_position When Parameter value is not same as the


ERROR 05-001 value in PC.
Set value ERROR! Contact Technical support team of YD.

Cassette In_position When Parameter value is not same as the


ERROR 05-002 Set value ERROR!
value in PC.
Contact Technical support team of YD.
When Cassette does not reach Limit
ERROR 05-003 Cassette Limit-Goal_position ERROR! sensor, or sensor problem.
Contact Technical support team of YD.

Sensor problem
ERROR 05-004 Cassette Row_sensor ERROR! Contact Technical support team of YD.

When Cassette does not reach Row


ERROR 05-005 Cassette Row_Goal position ERROR! sensor, or sensor problem.
Contact Technical support team of YD.
When Rack input does not reach Front
ERROR 06-001 Rack input Front-Goal_position ERROR! sensor, or sensor problem.
Contact Technical support team of YD.
When Rack input does not reach Rear
ERROR 06-003 Rack input Rear-Goal_position ERROR! sensor, or sensor problem.
Contact Technical support team of YD.

Operator Manual / 56
URiSCAN Super+ Fully automated urine analyzer Rev.02

CODE Message Cause & Solution


Rack shifter Bring-rack_position When Parameter value is not same as the
ERROR 07-001 Set value ERROR !
value in PC.
Contact Technical support team of YD.

Rack shifter Barcode_position When Parameter value is not same as the


ERROR 07-002 Set value ERROR !
value in PC.
Contact Technical support team of YD.

Rack shifter Sampling_position When Parameter value is not same as the


ERROR 07-003 Set value ERROR !
value in PC.
Contact Technical support team of YD.

Rack shifter Rack-push_position When Parameter value is not same as the


ERROR 07-004 Set value ERROR !
value in PC.
Contact Technical support team of YD.
When Rack Shifter does not reach Limit
ERROR 07-005 Rack Shifter Limit-Goal_position ERROR ! sensor, or sensor problem.
Contact Technical support team of YD.
When Rack Pusher does not move to the
ERROR 08-001 Rack Pusher Goal_position ERROR ! right position set by value.
Contact Technical support team of YD.
When Rack Pusher does not move to the
ERROR 08-002 Rack Pusher Moving ERROR ! right position set by value.
Contact Technical support team of YD.

Arm Sampling_position When Parameter value is not same as the


ERROR 09-001 Set value ERROR !
value in PC.
Contact Technical support team of YD.

Arm Waste_position When Parameter value is not same as the


ERROR 09-002 Set value ERROR !
value in PC.
Contact Technical support team of YD.

Arm Washing_position When Parameter value is not same as the


ERROR 09-003 Set value ERROR !
value in PC.
Contact Technical support team of YD.

Arm 1’st Pad_position When Parameter value is not same as the


ERROR 09-004 Set value ERROR !
value in PC.
Contact Technical support team of YD.
When Arm does not reach Limit sensor, or
ERROR 09-005 Arm Limit-Goal_position ERROR ! sensor problem.
Contact Technical support team of YD.
When Arm does not move to the right
ERROR 09-007 Arm Goal_position ERROR ! position set by value.
Contact Technical support team of YD.

Syringe Bottom_position When Parameter value is not same as the


ERROR 10-001 Set value ERROR !
value in PC.
Contact Technical support team of YD.

Syringe Ready_position When Parameter value is not same as the


ERROR 10-002 Set value ERROR !
value in PC.
Contact Technical support team of YD.
When Syringe does not reach Top sensor,
ERROR 10-003 Syringe Top-Goal_position ERROR ! or sensor problem.
Contact Technical support team of YD.
When Syringe does not reach Bottom
ERROR 10-004 Syringe Bottom-Goal_position ERROR ! sensor, or sensor problem.
Contact Technical support team of YD.

When wrong Barcode is read.


ERROR 13-001 Barcode illegal data reading ERROR ! Try with the new barcode label.

When communication of Barcode data is


ERROR 13-002 Barcode reading data packet ERROR ! broken.
Contact Technical support team of YD.
When Barcode module is not connected
ERROR 13-003 Barcode reader ERROR ! properly.
Contact Technical support team of YD.

When unknown Barcode system is read.


ERROR 13-004 Barcode unknown data ERROR ! Try with the other Barcode system.

When Image Scanner module is not


ERROR 14-001 Scanner Device_Error ! connected well.
Contact Technical support team of YD.

Operator Manual / 57
URiSCAN Super+ Fully automated urine analyzer Rev.02

CODE Message Cause & Solution


Check if there is a cassette in the cassette
ERROR 15-008 No Cassette ! room.
Press [Change cassette] on the Menu.
When there is no Distilled Water in the
ERROR 16-003 DW-tank Empty ! DW-Tank.
Supply Distilled Water to the DW-tank..

When the waste is full in the waste Tank.


ERROR 16-004 Liquid Waste-tank Fill ! Empty Waste-tank.

When there is no Waste-tray.


ERROR 16-005 No Waste-tray ! Install Waste-tray.

When Rack shifter does not work properly.


ERROR 16-006 Rack transfer ERROR ! Contact Technical support team of YD.

Operator Manual / 58
URiSCAN Super+ Fully automated urine analyzer Rev.02

Chapter 6 Appendix

6.1 Communications -------------------------------------------------------- 60


6.2 Value of Initial Standard Range table (%R) ---------------------- 62
6.3 Program for the Sensitivity of each parameter ----------------- 64
6.4 Measurement density --------------------------------------------------- 66

Operator Manual / 59
URiSCAN Super+ Fully automated urine analyzer Rev.02

6.1 Communications
This section is to explain the interfacing with host computer, barcode reader, external
printer.

6.1.1 Host Communication

There are many different ways of interfacing the system. URiSCAN Super+ is using
serial communication which is widely used.
► RS232C cable is used for the communication between host computer and
URiSCAN Super+.
► Once data communication is open, [Start Code] will be transmitted to host
computer from URiSCAN Super+, and then data will be transmitted to host
computer in ASCii code.
► Data Transmission order follows printing-out order. [Line Feed] (hereafter ‘LF’)
and [Carriage Return] (hereafter ‘CR’) will be transmitted at the end of each
Line.
► Start Code, end, LF, CR are in hexa decimal, while other data transmit in ASCii
code.

1) Data Format
 Baud Rate (bps) : 1200, 2400, 4800, 9600, 19.2K, 38.4K, 57.6K, 115.2K
 Data Bit (bit) : 7 or 8-bit code
 Parity Bit (bit) : None, Odd, Even
 Stop Bit (bit) : 1 or 2 stop bits

2) Hardware Connection

TX 2 2 TX
RX 3 3 RX
GND 5 5 GND

3) Download Protocol Structure


 Start character
 Data field (optional)
 CR (OxOd)
 LF (OxOa)

Operator Manual / 60
URiSCAN Super+ Fully automated urine analyzer Rev.02

6.1.2 Barcode Reader Communication

► With the internal barcode reader in URiSCAN Super+ , the external barcode
reader also can be used.
► External barcode reader to URiSCAN Super+ can be connected only through
PS/2(keyboard) port and USB port.

6.1.3 Parallel Communication

► Communication between URiSCAN Super+ and external printer: parallel


communication.
► Parallel port: DB-25 (Adapter with 25 pins.)
► URiSCAN Super+ is possible to compatible with deskjet or bubble-jet printer.

1) Printer Interface-data transfer to a printer


2) Hardware Connection
SYROBE 1 1 SYROBE
DATA 1 2 2 DATA 1
DATA 2 3 3 DATA 2
DATA 3 4 4 DATA 3
DATA 4 5 5 DATA 4
DATA 5 6 6 DATA 5
DATA 6 7 7 DATA 6
DATA 7 8 8 DATA 7
DATA 8 9 9 DATA 8
BUSY 11 11 BUSY
DS 12 12 DS
S-GND 18 18 S-GND

Printer URiSCAN Super+

Operator Manual / 61
URiSCAN Super+ Fully automated urine analyzer Rev.02

6.2 Value of the Initial Standard Range Table (R%)

The following is the initial standard R% (Reflection rate) of URiSCAN Super+. If


necessary, the below values can be adjusted by the authorized service person. However,
the value should not be changed more than +/- 20..

1. Blood 2. Bilirubin 3. Creatinine

Change %R for BLD. Change %R for BIL.


- neg <= [ 15] - neg <= [ 35]
+- 5 <= [ 30] + 0.5 <= [ 50]
After applying
+ 10 <= [ 60] ++ 1.0 <= [ 65]
++ 50 <= [ 90] +++ 3.0 > [ 65]
+++ 250 > [ 90]

4. Urobilinogen 5. Ketones 6. Protein

Change %R for URO. Change %R for KET. Change %R for PRO.


+- norm <= [ 40] - neg <= [ 25] - neg <= [ 20]
+ 1.0 <= [ 55] +- 5 <= [ 35] +- 10 <= [ 30]
++ 4.0 <= [ 70] + 10 <= [ 50] + 30 <= [ 50]
+++ 8.0 <= [ 85] ++ 50 <= [ 70] ++ 100 <= [ 70]
++++ 12.0 > [ 85] +++ 100 > [ 70] +++ 300 <= [ 85]
++++ 1000 > [ 85]

7. Micro Albumin 8. Nitrite 9. Glucose

Change %R for NIT. Change %R for GLU.


- neg <= [ 15] - neg <= [ 30]
+ pos > [ 15] +- 100 <= [100]
After applying
+ 250 <= [185]
++ 500 <= [230]
+++ 1000 > [230]

Operator Manual / 62
URiSCAN Super+ Fully automated urine analyzer Rev.02

10. pH 11. Leucocytes 12. Ascorbic acid

Change %R for pH. Change %R for LEU. Change %R for VTC.


pH 5.0 <= [ 30] - neg <= [ 10] - neg <= [ 20]
pH 5.5 <= [ 40] +- 10 <= [ 20] + 10 <= [ 70]
pH 6.0 <= [ 60] + 25 <= [ 35] ++ 25 <= [120]
pH 6.5 <= [110] ++ 75 <= [ 55] +++ 50 > [120]
pH 7.0 <= [160] +++ 500 > [ 55]
pH 7.5 <= [195]
pH 8.0 <= [210]
pH 8.5 <= [230]
pH 9.0 > [230]

13. COLOR 14. Clarity

Change %R for COL. Change %R for COL. Change %R for CLA.


(Concentration) (hue circle) Clear 0 ~ 100
. SL.Cloudy 101 ~ 170
LT.YELLOW 0 ~ 15 RED 300 ~ 25 Cloudy 171 ~ 350
NORMAL 16 ~ 40 ORANGE 26 ~ 54 Turbid 351 ~ 650
DK. 41 ~ 80 YELLOW 55 ~ 75 Bloody 651 ~
GREEN 76 ~ 100
OTHER 101 ~ 160

Operator Manual / 63
URiSCAN Super+ Fully automated urine analyzer Rev.02

6.3 Program for the sensitivity of each parameter


The following tables describe the sensitivity of each test parameter according to its
concentration.

「*」stands for the sensitivity, it is the first positive result detected by URiSCAN
Super.

Blood
Printout Arbitrary - ± + ++ +++
Printout Conv. unit neg 5 10 50 250 RBC/µL
Printout SI unit neg 5 10 50 250 RBC/µL
Concentration Ranges 0~4 5*~7 8~29 30~140 141[

Bilirubin
Printout Arbitrary - + ++ +++
Printout Conv. unit neg 0.5 1.0 3.0 mg/dL
Printout SI unit neg 9 17 50 µmol/L
Concentration Ranges 0~0.4 0.5*~0.7 0.8~1.9 2.0[

Creatinine
Printout Arbitrary ± + ++ +++ ++++
Printout Conv. unit mg/dL
Printout SI unit µmol/L
Concentration Ranges

Urobilinogen
Printout Arbitrary ± + ++ +++ ++++
Printout Conv. unit 0.1 1.0 4.0 8.0 12.0 mg/dL
Printout SI unit 1.6 16 66 131 197 µmol/L
Concentration Ranges 0.1~1.0 2.0*~2.9 3.0~6.0 7.0~10.0 11.0[

Ketones
Printout Arbitrary - ± + ++ +++
Printout Conv. unit neg 5 10 50 100 mg/dL
Printout SI unit neg 0.1 1 5 10 mmol/L
Concentration Ranges 0~4 5*~7 8~29 30~74 75[

Protein
Printout Arbitrary - ± + ++ +++ ++++
Printout Conv. unit neg 10 30 100 300 1000 mg/dL
Printout SI unit neg 0.1 0.3 1 3 10 g/L
Concentration Ranges 0~20 21*~50 51~80 81~200 201~600 601[

Operator Manual / 64
URiSCAN Super+ Fully automated urine analyzer Rev.02

Micro Albumin
Printout Arbitrary - ± + ++ +++ ++++
Printout Conv. unit neg mg/dL
Printout SI unit neg g/L
Concentration Ranges 0~29

Nitrite
Printout Arbitrary - +
Printout Conv. unit neg pos
Printout SI unit neg pos
Concentration Ranges 0~0.04 0.5[*

Glucose
Printout Arbitrary - ± + ++ +++
Printout Conv. unit neg 100 250 500 1000 mg/dL
Printout SI unit neg 5.5 14 28 55 mmol/L
Concentration Ranges 0~49 50*~175 176~375 376~750 751~1499

pH
Printout Conv. unit 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0
Printout SI unit 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0
Concentration
]5.2 5.3~5.7 5.8~6.2 6.3~6.7 6.8~7.2 7.3~7.7 7.8~8.2 8.3~8.7 8.8[
Ranges

Leucocytes
Printout Arbitrary - ± + ++ +++
Printout Conv. unit neg 10 25 75 500 WBC/µL
Printout SI unit neg 10 25 75 500 WBC/µL
Concentration
0~9 10*~17 18~60 61~300 301[
Ranges

Ascorbic acid
Printout Arbitrary - + ++ +++
Printout Conv. Unit neg 10 25 50 mg/dL
Printout SI unit neg 0.6 1.4 2.8 mmol/L
Concentration
0~9 10*~16 17~35 36[
Ranges

Operator Manual / 65
URiSCAN Super+ Fully automated urine analyzer Rev.02

6.4 Measurement Density


Test Measurement
Display Unit
Parameter
Quantitative Qualitative

Blood BLD RBC/L - 5 10 50 250 - ± + ++ +++

mg/dL - 0.5 1.0 3.0 - + ++ +++


Bilirubin BIL
mol/L - 9 17 50 - + ++ +++
mg/dL
Creatinine CRE
mol/L
mg/dL 0.1 1 4 8 12 ± + ++ +++ ++++
Urobilinogen URO
mol/L 1.6 16 66 131 197 ± + ++ +++ ++++
mg/dL - 5 10 50 100 - ± + ++ +++
Ketones KET
mmol/L - 0.5 1 5 10 - ± + ++ +++

Protein PRO mg/dL - 10 30 100 300 1000 - ± + ++ +++ ++++

mg/dL
Microalbumin ALB
mol/L
Nitrite NIT - +

mg/dL - 100 250 500 1000 2000 - ± + ++ +++


Glucose GLU
mmol/L - 5.5 14 28 55 111 - ± + ++ +++

pH pH 5 5.5 6 6.5 7 7.5 8 8.5 9

Leucocytes LEU WBC/L - 10 25 75 500 - ± + ++ +++


mg/dL - 10 25 50 - + ++ +++
Ascorbic acid VTC
mmol/L - 0.6 1.4 2.8 - + ++ +++

LT.Yellow, Red, DK.Red, Amber, Brown, DK.Brown,


Color COL Orange, DK.Orange, Yellow, DK.Yellow, Green,
DK.Green, Other

Turbidity CLA Clear, SL.Cloudy, Cloudy, Turbid, Bloody


Specific
SG 1.000 ~ 1.050
Gravity

Operator Manual / 66
URiSCAN Super+ Fully automated urine analyzer Rev.02

76 Seori-ro, Idong-myeon, Cheoin-gu,


Yongin-si, Gyeonggi-do, Korea (Post Code: 449-834)
TEL: 82-31-329-2000 FAX: 82-31-329-2007

Care diagnostica
46562 Voerde, Weseler Str. 110, Germany

Operator Manual / 67