Operating Instructions For Offline Software of Phascan
Operating Instructions For Offline Software of Phascan
Version 1.1.4
MENU
1
1 Software Icons Description
1.1columns icons
:Probe / Wedge, parameters reading and setting for Probe and wedge;
2
Doppler Electronic Technologies Co., Ltd
: Switch language;
: screen shot.
3
2. Data file calling
2.1 Open data file
Click , Select the data file needed analyzing from the folder (This software only
supports *.Data files saved in Phascan). When recalling multi-group of test data, the first
four groups will be displayed by default; while more than four groups displayed on the right
Software interface
If you want to analyze any group, clicking "Group *" beside "Com Groups". Also, you can
Group selection
4
Doppler Electronic Technologies Co., Ltd
Click display menu, Select the mode (this software provides totally 18 kinds of
display modes) tap and hold the left mouse button on the selected mode, drag it to the left
area.
Display modes
5
2.3 Add and delete the display window
If there are more than one group data in a single file or a single group data needs to be
viewed in different display modes, you can click on the top display area, right
Each time add a display window, you need to select the display group on the right area
first, for example, select ,then choose one mode under DISPLAY menu.
Up to 8 display windows
To delete a display window, click the left mouse button to select the window first, and then
double-click the right mouse button, the following dialog box appears:
Note: Assuming there are eight displays on show, if delete one display, such
automatically.
6
Doppler Electronic Technologies Co., Ltd
UT Setting menu involves the saved ultrasonic parameters, such as Gain, Range,
Velocity etc., the gray box is read-only data can not be written, data in not greyed out
UT Setting menu
The gain in grey in Gain Menu is the instrument gain the moment the data is being saved,
the gain in not greyed out frame can be adjusted according to the later data analysis by
scrolling the mouse or direct input dB value to increase or decrease the gain.
7
3.2 Probe/Wedge Menu
Probe/Wedge Menu involves parameters of PAUT probe and wedges used in the test the
moment the data is being saved, all the parameters in the gray are read only.
Probe/Wedge menu
Focal law menu involves focal law parameters the moment the data is being saved, all the
8
Doppler Electronic Technologies Co., Ltd
Gate and Sizing Curves menu involves the parameters of Gate and TCG curve the
moment the test data is being saved, the gray frame is read-only data and can not be
Show selection can be used to switch the Gate in "Show" and "hidden" state, same as the
DAC/TCG curve.
The drop-down select frame use to select Gate A, Gate B and Gate I.
9
3.5 Geometry menu
Geometry menu calls the work piece parameters, scanning position the moment the data
is being saved. and can recall weld geometry in this menu. The gray frame is read-only
Geometry menu
3.5.1 Position data frame is used to change the probe position, by adjusting the Index
Offset value, you can change the position of Weld geometry offset according to setup.
10
Doppler Electronic Technologies Co., Ltd
below:
There are "I-, V-, double V-, U-,diff_V,J,VY" seven kinds of weld types available.
There are "symmetrical, single slope mouth left and right single slope mouth," three kinds
According to the actual situation of the weld, adjust the parameters below:
11
Click And the Weld setup parameters are finished.
Show selection can be used to switch the Weld Illustration in "Show" and "hidden" state.
You can choose to display or hide the LW and BW wave, B scan window shows two blue
12
Doppler Electronic Technologies Co., Ltd
It is non-parallel scan by default for TOFD mode, data processing can select "whole" or
"Wedge separation" refers to the actual distance between the wedges, if we find the
distance is not accurate, it can be fine-tuned here. "Probe center distance" is calculated
automatically, without modification according to the forefront of the wedge spacing and
probe input.
"Wedge delay correction" can be properly adjusted if the parameters is not accurate,
customers can revise it so as LW line intersect with the first positive peak position. This
method is suitable for high precision measurement, general users can use hereinafter the
When the test piece is thick, you need to set more than one group to detect different depth
range, here refers to "layer". "Depth Start" and "Depth end" is respectively corresponding
to the minimum and maximum depth for each layer. If you do not need to set layer, the
default setting of "Depth Start" and "Depth end" are "0" and "work piece thickness"
respectively.
13
Depth calibration can be performed using LW or a known defect, move the red cursor to
the desired position, press calibrate and the depth value calibrated to the entered
thickness.
Multiple measured parameters can be set in Measurement menu , such as A%, DA, SA,
etc., and display in A-Scan, B-Scan, C-Scan, S-Scan simultaneously, the software
Show selection on the right can be used to switch the Measurement in "Show" and
"hidden" state. "Background" is used to add background color box for measured value,
Cursor menu bar involves totally 8 measurement cursor values of % (r),% (m), U (r), U (m),
S (r), S (m), I (r), I (m). In A-Scan, B-Scan, C-Scan, S-Scan, move these measurement
cursors, Cursor values will change. Similarly, change the Cursor values, the cursors in
14
Doppler Electronic Technologies Co., Ltd
Show selection on the right can be used to switch the Cursor in "Show" and "hidden" state.
"Cursor Sync" means the scan axis moves with the cursor correspondingly, enabling
In display menu, you can select the display mode for each display window, change the
3.7.1 There are totally 18 kinds of display modes for selecting in Display Mode Menu.
3.7.2 Can change the display mode and color of C-Scan in ColorSetting menu.
15
(1)There are totally 8 kinds of C-Scan display modes for selecting.
C Scan 的 8 种显示方式
Click on the "scan" icon next to the menu bar to bring down the scanning parameters
column, you can view the scan parameters related settings. Under this menu, you can
16
Doppler Electronic Technologies Co., Ltd
adjust the current location of scanning lines (ie, the yellow dotted line in C-scan and
B-scan).
4. Data analysis
4.1 TOFD data analysis
4.1.1 LW straightening
17
Step 2: Drag the LW blue line to be in front of lateral wave.
Step 3: Click below the icon in red box, if it is "whole" process, it will achieve alignment
effect. If you choose "area" processing, you need to drag the mouse to select the area to
be straightened. For revocation process, click on the shortcut icon in blue box, as
illustrated below.
Step 2: Click below the icon in red box, if it is "whole" process, it will achieve lateral wave
removing effect. If you choose "area" processing, you need to drag the mouse to select
18
Doppler Electronic Technologies Co., Ltd
the area. For revocation process, click on the shortcut icon in blue box.
Before removal
After removal
1. Layer setting: When the test piece is thick, you need to set more than one group and
the work piece need to be layered. At this point, check standard and input "depth start"
and "depth end" value. If you only use one group of TOFD, skip this step.
19
2. Depth calibration: when LW can be seen, align the red cursor to the first positive peak of
lateral wave, set the depth calibration value is 0 then click "calibration". If a second layer is
detected and LW can not be seen, use a known defect to complete calibration.
3. Usually after depth calibration, we can measure the depth of defect. Drag the horizontal
and vertical cursor to frame the defect, then click on the shortcut icons measuring length,
height and depth, then click Save. A red box will appear in B scan and the defect will be
recorded with location and size information, as shown below. When there are multiple
20
Doppler Electronic Technologies Co., Ltd
Defect in B scan
Note: after depth calibration, LW line will coincide with the lateral wave, if there is some
distance , it may be because the thickness of the work piece or the wedge separation is
not accurate and should be carefully checked; if both accurate, you can fine-tune "wedge
delay correction " to make LW line at the first positive peak position of lateral wave, so as
4. SAFT
Synthetic aperture focusing can partly eliminate beam scattering, which helps to improve
Phased array analysis operation is relatively simple, here after summarizes simple
1. Select a set of data, the display window is adjusted to ASC (or ASBC, etc.), drag the
scanning line (yellow dotted line) to the defect to be measured, make sure the reference
and measurement cursors be on left and right sides of the defect respectively, the
21
2. In S scan, drag the purple angle line, find the highest echo of defect, and then drag the
angle line around until the echo height drops -6dB, record the cursor position, click .
Below is saved image information and the corresponding measured values, also you can
22
Doppler Electronic Technologies Co., Ltd
3. TOFD and PA can be analyzed at the same time. If you need to modify the scan offset
value of a group, find the "geometry- Position- scan offset" menu and input value wanted.
NOTE: After setting the offset value, you need to press the enter key.
Scan offset=0
23
Scan offset=50mm
1. Click , set part name, part No., part position, report name, date, Besides, the Group
parameters "shown" or "hidden" in the report.
2. After TOFD data analyzed, click (see 4.1), the system will automatically save the
current screen image (each time press "save" ,a defect is saved with screen image); if the
defect need to save more pictures, click to increase, click to delete the last
image sequentially.
3. PA data analysis (see 4.2), save defect together with screen image. click to
24
Doppler Electronic Technologies Co., Ltd
25
Report example without group parameters
In any A-Scan, B-Scan, C-Scan, S-Scan, the image can be zoomed in quickly to meet the
Press and hold the left mouse button in the area you want to zoom in, drag the mouse to
the bottom right to get a selected rectangle, then release the mouse button, then the
selected rectangle area will be expanded to the entire scan window. Right click on the
enlarged image, the enlarged area will restore to the original proportion.
26
Doppler Electronic Technologies Co., Ltd
Move the mouse cursor to the center of the area want to zoom in, scroll up and the image
will be enlarged, scroll down to zoom out the image or click the right mouse button to
In any A-Scan, B-Scan, C-Scan, S-Scan, you can do quick quantitative measurement by
double clicking. Take S-Scan for example to show how to operate, specific operation
method is as follows:
Move the mouse cursor to the top left corner of the defect image, left double-click, a green
crossed reference cursors will be quickly located at the double-click position. Move the
cursor to the bottom right of the defect image, left double-click, a red crossed reference
cursors will be quickly located at the double-click position. To make measurements more
27
accurately, you can drag the purple angle cursor to view A-Scan waves in different angles.
By observing the A-Scan amplitude, drag the four measurement cursors separately to do
fine-adjustment.
The four measurement values of I (m), I (r), U (m), U (r) are displayed above the horizontal
line or on the left of vertical line, they will also be displayed in the Measurement menu.
You can call I(m-r) or U(m-r) in Measure Value in Measurement menu, then directly get
This software provides S-A-A-A Scan and S-A-A-A-C Scan display modes, it supports up
In S-Scan the purple, respective yellow,red and purple angle cursors are displayed.
28
Doppler Electronic Technologies Co., Ltd
29