Calibration / Validation The Official Calibration of The SAS (Surface Air System) Family of Air Samplers

The document discusses the importance of periodically validating the flow rate of air samplers to ensure accurate monitoring. It recommends performing validation every 6-12 months or if damage occurs. Validation is done using an impeller anemometer to measure the air flow and confirm performance.
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
121 views

Calibration / Validation The Official Calibration of The SAS (Surface Air System) Family of Air Samplers

The document discusses the importance of periodically validating the flow rate of air samplers to ensure accurate monitoring. It recommends performing validation every 6-12 months or if damage occurs. Validation is done using an impeller anemometer to measure the air flow and confirm performance.
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

CALIBRATION / VALIDATION

The official calibration of the SAS


The flow rate of any air sampler should be (Surface Air System) family of air
periodically checked to evaluate that the volume samplers:
of aspirated air is consistent.
POLITECNIC
There are two types of check to adopt: 32879
• an “on site” monthly or quarterly quality
control check to confirm continued
performance and identify any irregularity,
possibly caused by damage
• bi-annual or annual “official” calibration to
re-validate the sampling rate and satisfy
regulatory bodies that the instrument is Why
operating correctly. and when
you need air
The “on site” Quality Control sampler validation?
monitoring of the SAS (Surface Air It is important to verify that
System) Family of air samplers: the volume of air you are sampling
with your air sampler is correct to avoid errors in
Impeller anemometer with monitoring the contamination of an environment.
LCD display for the routine Validation is an important step to assure that the
evaluation of "SAS (Surface final calculation is based on an accurately
Air System)" air sampler sampled volume of air. Validation of the aspirated
performance volume should be performed every 6 months or
• Shows the 12 months. Re-validation is also recommended if
the air sampler has potentially been damaged or
air flow of any
the flow rate has been compromised.
"SAS (Surface
Air System)"
microbiological
air samplers
PYRAMID • Readings in either SERVICING
89389 litres per minute (L/min)
or cubic feet per minute (CFM) International PBI guarantees a comprehensive
• Just position it on top of the calibration and service program in its Milan (Italy)
aspirating head of the air premises or through local authorized distributors.
sampler and switch on
• Measuring time 30 seconds
397/05 uk - © International pbi S.p.A.

• According to Good Laboratory


Practice and Good Manufacturing
Practice
• Standard Operating Procedure
• Include carrying case

page 10