Carestream DRX-1 System - Carestream DRX-Mobile Retrofit Kit - Software Version 5.3 - DICOM Conformance Statement
Carestream DRX-1 System - Carestream DRX-Mobile Retrofit Kit - Software Version 5.3 - DICOM Conformance Statement
July 9, 2010
Document # 8J4601
INTRODUCTION .......................................................................................................................................................4
1.1 EXECUTIVE OVERVIEW ................................................................................................................................4
1.2 SCOPE AND FIELD OF APPLICATION ...........................................................................................................4
1.3 IMPORTANT CONSIDERATIONS FOR THE READER ....................................................................................5
1.4 ACCESSING THIS CONFORMANCE STATEMENT ON THE WORLD WIDE WEB.........................................5
1.5 DEFINITIONS, ACRONYMS, ABBREVIATIONS .............................................................................................6
2 IMPLEMENTATION MODEL .......................................................................................................................7
2.1 FUNCTIONAL DEFINITIONS ..........................................................................................................................7
2.2 SEQUENCING OF REAL-WORLD ACTIVITIES ..............................................................................................8
3 APPLICATION ENTITY SPECIFICATIONS ..............................................................................................8
3.1 ASSOCIATION ESTABLISHMENT POLICIES .................................................................................................8
3.1.1 General ..................................................................................................................................................8
3.1.2 Number of Associations .........................................................................................................................8
3.1.3 Asynchronous Nature ............................................................................................................................9
3.1.4 Implementation Identifying Information................................................................................................9
3.2 ASSOCIATION INITIATION POLICY ..............................................................................................................9
3.2.1 Associated Real-World Activity .............................................................................................................9
3.2.2 Presentation Context Table ...................................................................................................................9
3.2.3 SOP Specific Conformance..................................................................................................................11
3.3 ASSOCIATION ACCEPTANCE POLICY ........................................................................................................11
3.4 BASIC PRINT MANAGEMENT META SOP CLASS......................................................................................11
3.4.1 Basic Film Session SOP Class.............................................................................................................11
3.4.2 Basic Film Box SOP Class ..................................................................................................................12
3.4.3 Basic Image Box SOP Class ................................................................................................................14
3.4.4 Printer SOP Class................................................................................................................................15
3.5 STORE SERVICE CLASSES ...........................................................................................................................17
3.5.1 Digital X-Ray Image Storage...............................................................................................................18
3.5.2 Computed Radiography Image Storage...............................................................................................26
3.5.3 Grayscale Softcopy Presentation State IOD........................................................................................31
3.6 BASIC WORK LIST SERVICE .......................................................................................................................36
3.6.1 Modality Work List IOD ......................................................................................................................36
3.6.2 Procedure Mapping .............................................................................................................................39
3.7 MODALITY PERFORMED PROCEDURE STEP NOTIFICATION SOP CLASS ..............................................39
3.7.1 Modality Performed Procedure Step Notification IOD .......................................................................39
3.8 STORAGE COMMITMENT PUSH MODEL SOP CLASS ...............................................................................42
3.8.1 DIMSE Service N-ACTION .................................................................................................................43
3.8.2 DIMSE Service N-EVENT-REPORT ...................................................................................................43
4 COMMUNICATION PROFILES..................................................................................................................44
4.1 SUPPORTED COMMUNICATIONS STACKS .................................................................................................44
4.2 PHYSICAL MEDIA ........................................................................................................................................44
5 EXTENSIONS/SPECIALIZATIONS/PRIVATIZATIONS ........................................................................44
5.1 COMPUTED RADIOGRAPHY IOD................................................................................................................44
5.1.1 Specialization of Pixel Spacing ...........................................................................................................44
5.1.2 Private Tags.........................................................................................................................................44
5.1.3 Alternate Image Display Spaces ..........................................................................................................45
5.1.4 Extension for P-Value Encoding .........................................................................................................45
MPPS Server
MPPS Notificaton
(DirectView DR system is SCU)
Kodak DirectView DR System
Basic Work List
(DirectView DR system is SCU)
HIS/RIS Service
Remote Printer
The DirectView DR System is a digital radiographic image acquisition product utilizing a 3rd
party X-Ray detector, which combines amorphous silicon (a-Si) and cesium iodide (CsI)
technologies. The system allows the technologist to acquire and identify images, review images
for placement and motion, add markers, adjust image processing, manipulate images (flip, rotate,
etc.), pan/zoom and route images to their final destinations. The system is capable of obtaining
patient demographic information manually from the user interface, the Remote Patient Data
Entry System (RPDES) or via a DICOM Work List. The RPDES allows patient demographic
data to be entered from remote computers for sites that do not support the DICOM Work List.
General Study
Study Instance UID (0020,000D) 1
Study Date (0008,0020) 2 Refer to Addendum B
Study Time (0008,0030) 2 Refer to Addendum B
Referring Physician Name (0008,0090) 2 Refer to Addendum B
Study ID (0020,0010) 2
Accession Number (0008,0050) 2 Refer to Addendum B
Patient Study
Admitting Diagnoses (0008,1080) 3
Description
Patient’s Age (0010,1010) 3
Patient’s Size (0010,1020) 3
Patient’s Weight (0010,1030) 3
Occupation (0010,2180) 3
Additional Patient’s History (0010,21B0) 3
General Series
Modality (0008,0060) 1 DX
Series Instance UID (0020,000E) 1
Series Number (0020,0011) 2
Laterality (0020,0060) 2C L
R
Series Date (0008,0021) 3
Series Time (0008,0031) 3
Series Description (0008,103E) 3 Same value as tag
(0018,5101)
Operator Name (0008,1070) 3 Refer to Addendum B
Body Part Examined (0018,0015) 3 Refer to Addendum B
Performed Procedure Step (0040,0244) 3
Start Date
Performed Procedure Step (0040,0245) 3
Start Time
Performed Procedure Step (0040,0253) 3
ID
Performed Procedure Step (0040,0254) 3 Refer to Addendum B
Description
DX Series
Modality (0008,0060) 1 DX
Referenced Performed (0008,1111) 1C
Procedure Step Sequence
>Referenced SOP Class UID (0008,1150) 1C
>Referenced SOP Instance (0008,1155) 1C
UID
Presentation Intent Type (0008,0068) 1 FOR PRESENTATION
General Equipment
Manufacturer (0008,0070) 2 Carestream Health or Kodak
Institution Name (0008,0080) 3 For this value to be
populated and sent, the
DirectView user needs to
enter the value on the
Delivery Preferences screen.
Institution Address (0008,0081) 3 For this value to be
populated and sent, the
DirectView user needs to
enter the value on the
Delivery Preferences screen.
General Image
Instance Number (0020,0013) 2
Patient Orientation (0020,0020) 2C
Image Type (0008,0008) 3
Content Date (0008,0023) 2C
Content Time (0008,0033) 2C
Acquisition Date (0008,0022) 3
Acquisition Time (0008,0032) 3
Derivation Description (0008,2111) 3
Acquisition Number (0020,0012) 3
Source Image Sequence (0008,2112) 3 Note: This is sent for the
FOR_PRESENTATION
image, with the UID to the
FOR_PROCESSING image,
even if the
FOR_PROCESSING image
was not sent to store
destination.
>Referenced SOP Class UID (0008,1150) 1C
>Referenced SOP Instance (0008,1155) 1C
UID
Images In Acquisition (0020,1002) 3
Image Comments (0020,4000) 3 Refer to Addendum B
Lossy Image Compression (0028,2110) 3
00 = No Compression
Image Pixel
Samples per Pixel (0028,0002) 1 1
Photometric Interpretation (0028,0004) 1 MONOCHROME1
MONOCHROME2
Contrast/Bolus
Contrast/Bolus Agent (0018,0010) 2 Refer to Addendum B
DX Anatomy Imaged
Image Laterality (0020,0062) 1 L
R
Anatomic Region Sequence (0008,2218) 2 Refer to Context Identifier
4031 in Addendum A for
supported SNOMED Code
Values
>Code Value (0008,0100) 1C
>Coding Scheme Designator (0008,0102) 1C
>Code Meaning (0008,0104) 1C
>Mapping Resource (0008,0105) 1C
>Context Group Version (0008,0106) 1C
>Context Identifier (0008,010f) 3
DX Image
Image Type (0008,0008) 1
Samples per Pixel (0028,0002) 1 1
Photometric Interpretation (0028,0004) 1 MONOCHROME1
MONOCHROME2
Bits Allocated (0028,0100) 1 16
If FOR_PROCESSING
image, INVERSE
Lossy Image Compression (0028,2110) 1 00 = No Compression
DX Detector
Detector Type (0018,7004) 2 DIRECT
Detector Configuration (0018,7005) 3 AREA
Detector Description (0018,7006) 3
Detector Mode (0018,7008) 3
Detector ID (0018,700A) 3
Date of Last Detector (0018,700C) 3
Calibration
Time of Last Detector (0018,700E) 3
Calibration
Exposures on Detector Since (0018,7010) 3
Last Calibration
X-Ray Collimator
Collimator Shape (0018,1700) 1 POLYGON
Vertices of the Polygonal (0018,1720) 1C
Collimator
DX Positioning
View Position (0018,5101) 3
View Code Sequence (0054,0220) 3 Refer to Context Identifier
4010 in Addendum A for
supported SNOMED Code
Values
>Code Value (0008,0100) 1C
>Coding Scheme Designator (0008,0102) 1C
>Coding Scheme Version (0008,0103) 1C
>Code Meaning (0008,0104) 1C
>Mapping Resource (0008,0105) 1C
>Context Group Version (0008,0106) 1C
>Context Identifier (0008,010f) 3
X-Ray Acquisition
Dose
KVP (0018,0060) 3
X-Ray Tube Current (0018,1151) 3
Exposure Time (0018,1150) 3
Exposure (0018,1152) 3
Distance Source to Detector (0018,1110) 3
X-Ray Filtration
X-Ray Grid
VOI LUT
>LUT Descriptor (0028,3002) 1C
>LUT Explanation (0028,3003) 1C
>LUT Data (0028,3006) 1C
Window Center (0028,1050) 1C 2048
Window Width (0028,1051) 1C 4096
Acquisition Context
Acquisition Context (0040,0555) 2 Empty Sequence
Sequence
SOP Common
SOP Class UID (0008,0016) 1
SOP Instance UID (0008,0018) 1
Specific Character Set (0008,0005) 1C
General Study
Study Instance UID (0020,000D) 1
Study Date (0008,0020) 2 Refer to Addendum B
Study Time (0008,0030) 2 Refer to Addendum B
Referring Physician Name (0008,0090) 2 Refer to Addendum B
Study ID (0020,0010) 2
Accession Number (0008,0050) 2 Refer to Addendum B
Study Description (0008,1030) 3 Refer to Addendum B
Referenced Study Sequence (0008,1110) 3
>Referenced SOP Class UID (0008,1150) 1C
>Referenced SOP Instance (0008,1155) 1C
UID
Procedure Code Sequence (0008,1032) 3
> Code Value (0008,0100) 1C Refer to Addendum B
> Coding Scheme (0008,0102) 1C
Designator
> Coding Scheme Version (0008,0103) 1C
> Code Meaning (0008,0104) 1C
Patient Study
Admitting Diagnoses (0008,1080) 3
Description
Patient’s Age (0010,1010) 3
General Series
Modality (0008,0060) 1 CR
Series Instance UID (0020,000E) 1
Series Number (0020,0011) 2
Laterality (0020,0060) 2C L
R
Series Date (0008,0021) 3
Protocol Name (0008,1030) 3
Series Time (0008,0031) 3
Operator Name (0008,1070) 3 Refer to Addendum B
Performed Procedure Step (0040,0244) 3
Start Date
Performed Procedure Step (0040,0245) 3
Start Time
Performed Procedure Step (0040,0253) 3
ID
Performed Procedure Step (0040,0254) 3 Refer to Addendum B
Description
Performed Protocol Code (0040,0260) 3
Sequence
>Code Value (0008,0100) 1C
>Coding Scheme Designator (0008,0102) 1C
>Coding Scheme Version (0008,0103) 1C
>Code Meaning (0008,0104) 1C
Request Attributes Sequence (0040,0275) 3
>Requested Procedure (0032,1060) 3 Refer to Addendum B
Description
>Scheduled Procedure Step (0040,0007) 3 Refer to Addendum B
Description
>Scheduled Protocol Code (0040,0008) 3
Sequence
>> Code Value (0008,0100) 1C Refer to Addendum B
>> Coding Scheme (0008,0102) 1C
Designator
>> Coding Scheme Version (0008,0103) 1C
CR Series
Body Part Examined (0018,0015) 2
View Position (0018,5101) 2
General Equipment
Manufacturer (0008,0070) 2 Carestream Health or Kodak
Institution Name (0008,0080) 3 For this value to be
populated and sent, the
DirectView user needs to
enter the value on the
Delivery Preferences screen.
Institution Address (0008,0081) 3 For this value to be
populated and sent, the
DirectView user needs to
enter the value on the
Delivery Preferences screen.
Station Name (0008,1010) 3
Institutional Department (0008,1040) 3 Refer to Addendum B
Name
Manufacturer Model Name (0008,1090) 3 DR XXXX – where X is
equal to the model number
of the device (i.e. DR 7500).
Device Serial Number (0018,1000) 3
Software Versions (0018,1020) 3
General Image
Instance Number (0020,0013) 2
Patient Orientation (0020,0020) 2C
Image Type (0008,0008) 3
Content Date (0008,0023) 2C
Content Time (0008,0033) 2C
Acquisition Date (0008,0022) 3
Acquisition Time (0008,0032) 3
Derivation Description (0008,2111) 3
Acquisition Number (0020,0012) 3
Images In Acquisition (0020,1002) 3
Image Comments (0020,4000) 3 Refer to Addendum B
Image Plane
Pixel Spacing (0028,0030) 3 “0.139\0.139”
Image Pixel
Samples per Pixel (0028,0002) 1 1
Photometric Interpretation (0028,0004) 1 MONOCHROME1
MONOCHROME2
Planar Configuration (0028,0006) 1C
Rows (0028,0010) 1 3056
(if the user crops the image,
the supported rows value
may be less than 3056
pixels)
Columns (0028,0011) 1 2544
(if the user crops the image,
the supported columns value
may be less than 2544
pixels)
Pixel Aspect Ratio (0028,0034) 1C
Bits Allocated (0028,0100) 1 16
Bits Stored (0028,0101) 1 12
High Bit (0028,0102) 1 11
Pixel Representation (0028,0103) 1 0
Smallest Image Pixel Value (0028,0106) 1 0
Largest Image Pixel Value (0028,0107) 1 4095
Pixel Data (7FE0,0010) 1
CR Image
KVP (0018,0060) 3
Plate ID (0018,1004) 3
Distance Source to Detector (0018,1110) 3
Distance Source to Patient (0018,1111) 3 Refer to Addendum B
Exposure (0018,1152) 3
Contrast/Bolus
Contrast/Bolus Agent (0018,0010) 2 Refer to Addendum B
VOI LUT
VOI LUT Sequence (0028,3010) 3
>LUT Descriptor (0028,3002) 1C
>LUT Explanation (0028,3003) 1C
>LUT Data (0028,3006) 1C
Window Center (0028,1050) 1C 2048
Window Width (0028,1051) 1C 4096
SOP Common
SOP Class UID (0008,0016) 1
SOP Instance UID (0008,0018) 1
Specific Character Set (0008,0005) 1C
General Study
Study Instance UID (0020,000D) 1
Study Date (0008,0020) 2 Refer to Addendum B
Study Time (0008,0030) 2 Refer to Addendum B
Referring Physician Name (0008,0090) 2 Refer to Addendum B
Study ID (0020,0010) 2
Accession Number (0008,0050) 2 Refer to Addendum B
Study Description (0008,1030) 3 Refer to Addendum B
Referenced Study Sequence (0008,1110) 3
>Referenced SOP Class UID (0008,1150) 1C
>Referenced SOP Instance (0008,1155) 1C
UID
Procedure Code Sequence (0008,1032) 3 Refer to Addendum B
> Code Value (0008,0100) 1C
> Coding Scheme (0008,0102) 1C
Designator
> Coding Scheme Version (0008,0103) 1C
> Code Meaning (0008,0104) 1C
Patient Study
Admitting Diagnoses (0008,1080) 3
Description
Patient’s Age (0010,1010) 3 Refer to Addendum B
Patient’s Size (0010,1020) 3
Patient’s Weight (0010,1030) 3
Occupation (0010,2180) 3
Additional Patient’s History (0010,21B0) 3
General Series
Presentation Series
Modality (0008,0060) 1
View Position (0018,5101) 2
General Equipment
Manufacturer (0008,0070) 2 Carestream Health or Kodak
Presentation State
Instance Number (0020,0013) 1
Content Label (0070,0080) 1
Content Description (0070,0081) 2
Presentation Creation Date (0070,0082) 1
Presentation Creation Time (0070,0083) 1
Content Creator’s Name (0070,0084) 2
(0008,1115)
Reference Series Sequence 1
> Series Instance UID (0020,000E) 1
> Referenced Image (0008,1140) 1
Sequence
>> Referenced SOP Class (0008,1150) 1
UID
>> Referenced SOP Instance (0008,1155) 1
UID
>> Referenced Frame (0008,1160) 1
Number
Shutter Presentation Value (0018,1622) 1C
Display Shutter
Shutter Shape (0018,1600) 1 POLYGONAL
Vertices of the Polygonal (0018,1620) 1C
Shutter
Overlay Plane
Overlay Rows (60xx,0010) 1
Overlay Columns (60xx,0011) 1
Overlay Type (60xx,0040) 1
Overlay Origin (60xx,0050) 1
Overlay Bits Allocated (60xx,0040) 1
Overlay Bit Position (60xx,0102) 1
Overlay Data (60xx,3000) 1C
Displayed Area
Displayed Area Selection (0070,005A) 1
Sequence
> Referenced Image (0008,1140) 1C
Sequence
> > Referenced SOP Class (0008,1150) 1C
UID
>> Referenced SOP Instance (0008,1155) 1C
UID
>> Referenced Frame (0008,1160) 1C
Number
> Displayed Area Top Left (0070,0052) 1
Hand Corner
> Displayed Area Bottom (0070,0053) 1
Hand Corner
> Presentation Size Mode (0070,0100) 1
> Presentation Pixel Spacing (0070,0101) 1C
> Presentation Pixel Aspect (0070,0102) 1C
Ratio
Spatial
Transformation
Image Rotation (0070,0042) 1 Always 0 (no rotation)
Image Horizontal Flip (0070,0041) 1 Always N (no flip)
SOP Common
SOP Class UID (0008,0016) 1
SOP Instance UID (0008,0018) 1
Specific Character Set (0008,0005) 1C
This is intended to produce a series of responses from the Work List SCP for all matching
Scheduled Procedures on the said Station.
The system can be configured to send this request periodically at an interval of 1 to 720 minutes.
The user may also issue a non-periodic C-FIND request to obtain information for a patient
recently added by the site’s information system.
Scheduled
Procedure Step
Scheduled Procedure Step (0040,0100) Required
Sequence
> Scheduled Station AE (0040,0001) Required
Title
> Scheduled Procedure Step (0040,0002) Required
Start Date
> Scheduled Procedure Step (0040,0003) Required
Start Time
> Modality (0008,0060) Required
>Requested Contrast Agent (0032,1070)
> Scheduled Performing (0040,0006) Required
Physician’s Name
>Scheduled Procedure Step (0040,0007)
Description
>Scheduled Station Name (0040,0010)
>Scheduled Protocol Code (0040,0008)
Sequence
>>Code Value (0008,0100)
>>Coding Scheme (0008,0102)
Designator
>>Code Meaning (0008,0104)
>Scheduled Procedure Step (0040,0009)
ID
Imaging Service
Request
Referring Physicians Name (0008,0090)
Accession Number (0008,0050)
Requesting Service (0032,1033)
Visit Status
Visit Status ID (0038,0008)
Patient’s Institution (0038,0400)
Residence
Patient
Identification
Patient’s Name (0010,0010)
Patient ID (0010,0020)
Other Patient Ids* (0010,1000)
Other Patient Names (0010,1001)
Patient
Demographic
Patient’s Birth Date (0010,0030)
Patient’s Birth Time (0010,0032)
Patient Medical
Additional Patient’s History (0010,21B0)
* Accept only one value (the first value) from the Modality Work List with a maximum length
of 64 characters
Performed Procedure
Step Relationship
Patient Name (0010,0010) 2
Patient ID (0010,0020) 2
Patient Birth Date (0010,0030) 2
Patient Sex (0010,0040) 2
Scheduled Step Attribute (0040,0270) 1
Sequence
> Study Instance UID (0020,000D) 1
> Referenced Study (0008,1110) 2
Sequence
>> Referenced SOP Class (0008,1150) 1C
UID
>> Referenced SOP Instance (0008,1155) 1C
UID
> Accession Number (0008,0050) 2
> Requested Procedure ID (0040,1001) 2
> Requested Procedure (0032,1060) 2
Description
> Scheduled Procedure Step (0040,0009) 2
ID
> Scheduled Procedure Step (0040,0007) 2
Description
> Scheduled Protocol Code (0040,0008) 2
Sequence
>> Code Value (0008,0100) 1C
>> Coding Scheme (0008,0102) 1C
Designator
>> Coding Scheme Version (0008,0103) 3
>> Code Meaning (0008,0104) 3
Referenced Patient Sequence (0008,1120) 2 Empty Sequence
Image Acquisition
Results
Modality (0008,0060) 1
Study ID (0020,0010) 3
Performed Protocol Code (0040,0260) 2
Sequence
> Code Value (0008,0100) 1C
> Coding Scheme (0008,0102) 1C
Designator
Radiation Dose
Anatomic Structure (0008,2229) Refer to Context Identifier
Sequence 4031 in Addendum A for
supported SNOMED Code
Values
>Code Value (0008,0100)
>Coding Scheme Designator (0008,0102)
>Coding Scheme Version (0008,0103)
>Code Meaning (0008,0104)
>Mapping Resource (0008,0105)
>Context Group Version (0008,0106)
>Context Identifier (0008,010f)
Total Number of Exposures (0040,0301)
Attribute Tag
Transaction UID (0008,1195)
Referenced SOP Sequence (0008,1199)
>Referenced SOP Class UID (0008,1150)
> Referenced SOP Instance UID (0008,1155)
Attribute Tag
Transaction UID (0008,1195)
Failed SOP Sequence* (0008,1198)
> Referenced SOP Class UID (0008,1150)
> Referenced SOP Instance UID (0008,1155)
> Failure Reason (0008, 1197)
Referenced SOP Sequence (0008,1199)
> Referenced SOP Class UID (0008,1150)
The DirectView DR System normally issues a network echo prior to delivering a job to a
qualified DICOM Print or DICOM Store destination to ensure that the device is active.
5 Extensions/Specializations/Privatizations
Tables 5.1 and 5.2 list a number of extensions to the DICOM Store Service Class, which the
DirectView DR System supports. The extensions are all optional attributes.
Attribute Tag VR VM
privateDirectviewGroup (0029, 0010) LO 1
privatePtoneWindowCenter (0029, 1002) IS 1
privatePtoneWindowWidth (0029, 1003) IS 1
privateBodyPartExamined (0029, 1004) LO 1
privateViewPosition (0029, 1005) LO 1
• Density
Pixel data is linearly related to optical film density. With a photometric interpretation of
MONOCHROME1, a value of 0 represents minimum density and a value of 4095 represents
maximum density.
• Luminance
Pixel data is perceptually linear in brightness according to a proprietary model of the human
visual system response developed by Kodak. Despite the name used, the pixel data is not
linearly related to luminance or log luminance.
This option should only be used in support of legacy Kodak display products.
The DR supports the use of the Density, Luminance or P-Value display spaces with the CR IOD,
DX IOD.
5.1.5 Extension to Body part Examined, View Position and Patient Position
The user may define additional labels that may be used to populate the Body Part Examined
(0018,0015), View Position (0018,5101) attributes. The software enforces the use of valid
characters for the additional labels created by the user.
6 Configuration
Table 6.1 lists the network attributes that are configurable by a qualified service provider.
The value set in the tag Specific Character Set (0008,0005) will be either:
a) “ISO_IR 100” (Latin 1),
b) “ISO_1R 13” (Katakana),
c) “ISO 2022 IR 13\ISO 2022 IR 87\ ISO 2022 IR 159” (Katakana, Hiragana, Kanji),
d) “ISO 2022 IR 149” (Hangul and Hanja),
e) “GB 18030” (Simplified and Traditional Chinese)
f) or blank (ISO-IR 6 is the default character set).
Note: The Extended Character Sets mentioned above do not apply to the Work List.
8 Error Handling
If communication errors are encountered, retries will be attempted based on configurable retry
intervals and maximum number of retries. If the communication cannot be achieved, the job will
be considered failed and set for retrial after a configurable time period (default of 5 minutes).
Table A.1 describes the SNOMED Code Values that are supported by the DirectView DR.
Context Identifier
19
Coding Scheme Designator (0008,0102) Coding Scheme Version (0008,0103)
SNM3 20020904
Code Value (0008,0100) Code Meaning (0008,0104)
F-10440 erect
F-10450 recumbent
F-10460 semi-erect
Context Identifier
20
Coding Scheme Designator (0008,0102) Coding Scheme Version (0008,0103)
SNM3 20020904
Code Value (0008,0100) Code Meaning (0008,0104)
F-10310 prone
F-10316 semi-prone
F-10317 right lateral decubitus
F-10318 lateral decubitus
F-10319 left lateral decubitus
F-10320 standing
F-10326 anatomical
F-10330 kneeling
F-10336 knee-chest
R-10202 frontal
R-10206 antero-posterior
R-10204 frontal oblique
R-10208 antero-posterior oblique
R-10210 right posterior oblique
R-10212 left posterior oblique
R-10214 postero-anterior
R-10216 postero-anterior oblique
R-10218 right anterior oblique
R-10220 left anterior oblique
R-10222 sagittal
R-10224 medial-lateral
R-10228 lateral-medial
R-10232 right lateral
R-10234 right oblique
Context Identifier
4015
Coding Scheme Designator (0008,0102) Coding Scheme Version (0008,0103)
SNM3 20050321
R-102D6 Magnification
Context Identifier
4031
Coding Scheme Designator (0008,0102) Coding Scheme Version (0008,0103)
SNM3 20020904
Code Value (0008,0100) Code Meaning (0008,0104)
Table B.1 describes the DICOM Image attributes and Patient/Exam information flow through the
product. The first column describes the Patient/Exam information fields supported by the DR
products. The second column indicates the MWL IOD Attribute that the system will use to
populate the field if a DICOM Work List SCP is present. The third column identifies the CR
Image IOD Attribute containing the data in the system’s output.
Table B.1 Patient/Exam Input and MWL/CR Image Attributes Cross Reference
User Interface Associated DICOM MWL Associated DICOM IOD
Input field Label IOD Element Element
Cassette ID NA Plate ID
(0018,1004)
input via bar code scanner
Contrast Agent Requested Contrast Agent Contrast/Bolus Agent
(0032,1070) (0018,0010)
Date Of Birth Patient’s Birth Date Patient’s Birth Date
(0010,0030) (0010,0030)
Department Name Requesting Service Institutional Department Name
(0032,1033) (0008,1040)
Gender Patient Sex Patient Sex
(0010,0040) (0010,0040)
Image Comments NA Image Comments
(0020,4000)
Laterality NA Laterality
(0020,0060)
Modality Modality Modality
(0008,0060) (0008,0060)
Orientation NA Cassette Orientation
(0018,1402)
Patient Age Patient’s age Patient’s age
(0010,1010) (0010,1010)
Read only value
Patient Comments NA Patient Comments (0010,4000)
Patient First Name Patient’s Name Patient’s Name
Patient Last Name (0010,0010) (0010,0010)
Patient Middle Name
Patient Location Patient’s Institution Residence Patient’s Institution Residence
(0038,0400) (0038,0400)
Patient Position NA NA
or
or
or